CC BY-NC-ND 4.0 · Asian J Neurosurg 2021; 16(03): 507-511
DOI: 10.4103/ajns.AJNS_446_20
Original Article

Superior effectiveness of a newly developed nonadherent polyurethane-coated surgical patty for hemostasis

Yohei Otsuka
Department of Neurosurgery, National Defense Medical College, Tokorozawa, Saitama, Japan
,
Terushige Toyooka
Department of Neurosurgery, National Defense Medical College, Tokorozawa, Saitama, Japan
,
Satoru Takeuchi
Department of Neurosurgery, National Defense Medical College, Tokorozawa, Saitama, Japan
,
Arata Tomiyama
Department of Neurosurgery, National Defense Medical College, Tokorozawa, Saitama, Japan
,
Satoshi Tomura
Department of Neurosurgery, National Defense Medical College, Tokorozawa, Saitama, Japan
,
Kojiro Wada
Department of Neurosurgery, National Defense Medical College, Tokorozawa, Saitama, Japan
› Author Affiliations

Background: Cotton patty is usually used to aspirate blood and cerebrospinal fluid to maintain a dry field. However, the cotton patty easily adheres to the vessels by capillary action, especially in combination with hemostat. Therefore, re-bleeding may be induced by removal of the cotton patty stuck to the vessel despite initial control of the bleeding. Methods: We have developed a new cotton patty (Non-Stina X®, Hakujuji, Co., Ltd., Tokyo, Japan) which does not adhere to the vessels. The newly developed cotton patty is made of 100% cotton, with only the contact surface coated with polyurethane film which prevents capillary action. The coated side includes many holes to allow aspiration from both sides. Results: The characteristics of four different surgical patties including our new patty which are available for surgical use in Japan were investigated. Transverse sections of four different surgical patties were investigated by light microscopy (magnification ×150). Our new cotton patty did not show any fluffing on the polyurethane-coated surface. However, other surgical patties showed some fluffing on their surfaces. The friction coefficients of four different surgical patties were investigated. Our new cotton patty had the lowest of the four neurosurgical patties. We confirmed the nonadherent characteristic using with hemostats of gelatinous sponge or fibrin glue-soaked oxidized cellulose cotton during hemostasis in neurosurgical procedures. The polyurethane-coated cotton patty could be removed easily from the hemostats without re-bleeding. Conclusions: The newly developed polyurethane-coated cotton patty is more effective for bleeding control from vessels with several types of hemostat due to the nonadherent characteristics.

Financial support and sponsorship

Nil.




Publication History

Received: 23 September 2020

Accepted: 14 September 2021

Article published online:
16 August 2022

© 2021. Asian Congress of Neurological Surgeons. This is an open access article published by Thieme under the terms of the Creative Commons Attribution-NonDerivative-NonCommercial License, permitting copying and reproduction so long as the original work is given appropriate credit. Contents may not be used for commercial purposes, or adapted, remixed, transformed or built upon. (https://creativecommons.org/licenses/by-nc-nd/4.0/)

Thieme Medical and Scientific Publishers Pvt. Ltd.
A-12, 2nd Floor, Sector 2, Noida-201301 UP, India

 
  • References

  • 1 Jadhav V, Solaroglu I, Obenaus A, Zhang JH. Neuroprotection against surgically induced brain injury. Surg Neurol 2007;67:15-20.
  • 2 Kinoshita M, Taniguchi M, Takagaki M, Seike N, Hashimoto N, Yoshimine T. Development of a non-tissue adherent neurosurgical patty and an ex vivo testing system to evaluate adherent characteristics. J Neurosurg 2015;122:1180-4.
  • 3 Stratton-Powell AA, Anderson IA, Timothy J, Kapur N, Culmer P. Neurosurgical patties: Adhesion and damage mitigation. J Neurosurg 2015;123:153-60.
  • 4 Dernehl CU. Health hazards associated with polyurethane foams. J Occup Med 1966;8:59-62.
  • 5 Arefin A, Huang JH, Platts D, Hypes VD, Harris JF, Iyer R, et al. Fabrication of flexible thin polyurethane membrane for tissue engineering applications. Biomed Microdevices 2017;19:98.
  • 6 Sugiyama F, Kleinschmidt AT, Kayser LV, Alkhadra MA, Wan JM, Chiang AS, et al. Stretchable and degradable semiconducting block copolymers. Macromolecules 2018;51:5944-9.
  • 7 Sekhar LN, Natarajan SK, Manning T, Bhagawati D. The use of fibrin glue to stop venous bleeding in the epidural space, vertebral venous plexus, and anterior cavernous sinus: Technical note. Neurosurgery 2007;61:E51.
  • 8 Li J, Chen Z, Yang X. State of the art of small-diameter vessel-polyurethane substitutes. Macromol Biosci 2019;19:e1800482.
  • 9 Bhushan B. Adhesion and stiction: Mechanisms, measurement techniques, and methods for reduction. J Vac Sci Technol B Microelectron Nanometer Struct Process Meas Phenom 2003;21:2262-96.
  • 10 Inoue M. Surface friction properties of fabrics and human skin. In: Ghrib T, editor. New Tribological Ways. 1st ed., Ch. 13. Rijeka, Croatia: InTech; 2011. p. 265-72.
  • 11 Menovsky T, De Ridder D, de Vries J. Neurosurgical patties and their intraoperative interaction with neural tissue. Surg Neurol 2009;71:326-9.
  • 12 Wada K, Otani N, Mori K. Usefulness of newly developed non-sticky polyurethane-coated surgical patty for cerebral vein or sinus bleeding control. Jpn J Neurosurg 2014;23:589-96.