Rofo 2018; 190(12): 1141-1151
DOI: 10.1055/a-0643-4553
Technique and Medical Physics
© Georg Thieme Verlag KG Stuttgart · New York

Whole-Body CT in Multiple Trauma Patients: Clinically Adapted Usage of Differently Weighted CT Protocols

Article in several languages: English | deutsch
Stefan Ulrich Reske
1   Department of Diagnostic Imaging and Interventional Radiology, BG-Klinikum Bergmannstrost Halle, Halle (Saale), Germany
2   Department of Diagnostic and Interventional Radiology and Neuroradiology, Heinrich-Braun-Klinikum, Zwickau, Germany
,
Rainer Braunschweig*
1   Department of Diagnostic Imaging and Interventional Radiology, BG-Klinikum Bergmannstrost Halle, Halle (Saale), Germany
5   Steinhorst, Germany
,
Andreas Wolfgang Reske
3   Department of Anaesthesiology, Intensive Care, Emergency Medicine and Pain Therapy, Heinrich-Braun-Klinikum, Zwickau, Germany
,
Reinhard Loose
4   Institute of Medical Physics, Nuremberg General Hospital, Paracelsus Medical University, Nuremberg, Germany
,
Michael Wucherer
4   Institute of Medical Physics, Nuremberg General Hospital, Paracelsus Medical University, Nuremberg, Germany
› Author Affiliations
Further Information

Correspondence

Stefan Ulrich Reske
Institut für Diagnostische und Interventionelle Radiologie und Neuroradiologie, Heinrich-Braun-Klinikum gemeinnützige GmbH
Karl-Keil-Straße 35
08060 Zwickau
Germany   
Phone: ++ 49/3 75/51 55 48 38   

Publication History

24 December 2017

07 May 2018

Publication Date:
12 November 2018 (online)

 

Abstract

Purpose Whole-body CT (wbCT) has been established as an internationally accepted diagnostic modality in multiple trauma. Until 2011, a uniform CT scanning protocol was used for all multiple trauma patients (pat.) at our hospital (OLD protocol = OP). In 2011, 2 new differently weighted protocols were introduced: TIME protocol (TP) for hemodynamically unstable pat. and DOSE protocol (DP) for pat. with stable vital parameters. The aim of this study was to compare the original “One-fits-all-concept” with the new, clinically oriented approach to wbCT.

Materials and Methods This study retrospectively evaluated 3 distinct wbCT protocols, looking at automatic exposure control variation (AEC; OP/TP) and arm positioning close to the body/overhead (TP/DP). The analysis included waist circumference (WC, cm), injury severity score (ISS), examination time (ET, min), image noise (IN), and effective dose (E, mSv). Normality of distribution was assessed with the Kolmogorov-Smirnov test. Data are given as median and range. Test of significance with Kruskal-Wallis test or Mann-Whitney-U-test. Level of significance: 0.05.

Results 308 pat. were included in the study (77 % m; age: 46 a, 18 – 90 a; WC: 93 cm, 66 – 145 cm). ISS was 14 (OP; n = 104; 0 – 75), 18 (TP; n = 102; 0 – 75) and 9 (DP; n = 102; 0 – 50). ET was 3.9 min (OP; 3.3 – 5.6 min), 4.1 min (TP; 2.8 – 7.2 min) and 7.7 min (DP; 6 – 10 min). IN showed no significant differences when comparing OP/TP but was significantly reduced in DP. For a wbCT (vertex to ischium), E could be reduced from 49.7 mSv to 35.4 mSv by optimizing AEC (OP/TP). Through the overhead repositioning of the arms in DP, a further reduction to 28.2 mSv was achieved.

Conclusion AEC and arm repositioning have a crucial influence on image quality and dose. The presented clinical approach is superior to the original concept.

Key Points:

  • The use of 2 differently weighted wbCT protocols allows a more flexible approach to the patient’s clinical presentation.

  • The clinically adapted concept presented in this study allows trauma care centers to reduce the collective dose.

  • Whole-body CT is leading to exposure to relevant radiation doses – further multicenter research is required.

Citation Format

  • Reske SU, Braunschweig R, Reske AW et al. Whole-Body CT in Multiple Trauma Patients: Clinically Adapted Usage of Differently Weighted CT Protocols. Fortschr Röntgenstr 2018; 190: 1141 – 1151


#

Introduction

Accidents and injuries are the most common cause of death and permanent disability among Germans under the age of 45 [1]. Optimal care of trauma patients plays a significant socioeconomic role. Efficient emergency room care in the acute phase has a major effect on the survival and long-term prognosis of patients with multiple injuries. Exact knowledge of the pattern of injuries is essential for treating in a targeted manner with defined priorities. Computed tomography (CT) has optimal characteristics for trauma diagnosis due to the short examination time and high sensitivity and specificity both for injuries as well as for trauma-independent, treatment-relevant incidental findings. Whole-body CT (wbCT) is considered as an independent predictor for the survival of trauma patients and has therefore become internationally established as the diagnostic method of choice [2] [3] [4]. Consequently, a rapid increase in the number of whole-body CT examinations in trauma patients has been observed in the last decade (in 2016 79 % of 33 374 patients entered in the TraumaRegister DGU® underwent whole-body examination) [5] [6].

The main point of criticism regarding wbCT is the relevant radiation exposure [7] [8] [9]. Particularly in young patients, there is an increased stochastic risk of developing a radiation-induced malignancy [10]. This is to be offset against the risk of death or permanent damage as a result of the trauma [11]. Therefore, indication determination plays an important role in the selection of the diagnostic concept (wbCT vs. diagnostic imaging performed in steps including ultrasound, X-ray and targeted CT). The main problem regarding indication determination is the high heterogeneity of the patient population in relation to the mechanism of trauma, pattern of injuries and clinical presentation. Young patients in particular can temporarily compensate for their circulatory situation despite serious injury thus masking the actual severity of the injuries prior to sudden worsening of their condition. To prevent the risk of undertriage (underestimation of the injury severity) with possibly fatal consequences for the patient or medicolegal consequences resulting from mistreatment, the indication for wbCT is often determined quickly at many hospitals in the case of a corresponding mechanism of trauma. However, experience in the daily radiology routine has shown that only some patients examined with wbCT meet the definition of polytrauma according to Tscherne (“injuries to multiple regions of the body suffered at the same time with at least one or the combination of the individual injuries being life-threatening”). If the inclusion criteria of the TraumaRegister DGU® (admission of a trauma patient to the emergency room and further treatment at an intensive care unit or intermediate care unit; emergency room admission with exitus letalis prior to admission to the intensive care unit; secondary transfer of a trauma patient with TraumaRegister DGU® documentation having been started at the hospital providing primary care) are taken into consideration, it can be concluded that the number of patients examined with wbCT and not entered in the TraumaRegister DGU® is significantly greater than the number stated above [12].

There are various approaches to indication determination in the literature: Some authors use the white paper or S3 guideline criteria for admitting a patient to the emergency room ([Table 1]) while others developed special score systems [13] [14] [15]. To date, there is no final interdisciplinary consensus. This is also shown in particular by the extensive discussion in the study by Gupta regarding trauma surgery and emergency room doctors [16]. There is a risk of an increased rate of wbCT without detection of injury in the case of generalized use of wbCT based solely on the mechanism of trauma. Therefore, almost 51 % of patients in the Gupta study did not have any injury that could be detected on CT and the injury severity score (ISS) was above 15 points in only 20 % of cases [16]. According to the definition of overtriage by Wurmb (ISS≥ 16 = polytrauma; ISS< 16 = overtriaged for wbCT), the overtriage rate was 80 % [17].

Table 1

Criteria for emergency room activation.

grade of recommendation A

grade of recommendation B

unstable vital parameters

apparent injuries

type of accident

  • systolic blood pressure less than 90 mmHg (age-adapted in children)

  • GCS less than 9

  • respiratory problems/need for intubation

  • penetrating injuries/gunshot wounds of the body/neck region

  • fractures of > 2 proximal long bones

  • unstable thorax

  • pelvic fractures

  • amputation injury proximal to hands/feet

  • spinal cord injury

  • open cranial injury

  • burns covering > 20 % of the body surface and degree ≥ 2b

  • fall from a height of greater than 3 m

  • traffic accident with:

    • front collision with intrusion of > 50 – 75 cm

    • change in speed of delta > 30 km/h

    • pedestrian/bicycle collision

    • death of a vehicle occupant

    • ejection of a vehicle occupant

Summary taken from the S3 guideline “Polytrauma/Schwerverletzten-Behandlung”. If at least one of these criteria is fulfilled, the trauma patient should be treated in the pre-warned emergency room of a trauma care hospital. The grades of recommendation A, B and 0 correspond to the level of evidence present during the development of the guideline and take the form of “shall”, “should” and “can” recommendations, respectively [13]. Many hospitals use these criteria to decide whether or not to perform a whole-body CT scan.

In many hospitals one single wbCT protocol is used for all trauma patients. This “one-fits-all” concept was abandoned in our hospital in 2011. Instead two differently weighted protocols are used adapted to the patient’s clinical condition. The goal of the present study is to evaluate three different wbCT protocols in relation to examination time, image quality, and dose. The original “one-fits-all” concept is compared to the new clinically adapted concept.


#

Materials and Methods

The present retrospective, monocentric and anonymized study was approved by the responsible ethics committee.

All presented data were acquired in the period from 6/1/2010 to 6/1/2013 in the clinical routine of a tertiary care hospital for trauma management with the status of a Level 1 trauma center. There are 2 emergency rooms for treating seriously ill/injured patients with one of the two CT units at our hospital being installed directly in one of the emergency rooms. Patient care is performed directly at the CT table, with the patient being positioned in the feet-first position for spatial-logistic reasons. The patient is repositioned by the emergency room team from the transportation means used by emergency medical services to a CT-compatible spine board. wbCT performed early after the primary survey has been the established primary diagnostic method at our hospital since 2003. Emergency room admission of trauma patients – and also determination of the indication for wbCT – were performed according to the criteria of the S3 guidelines ([Table 1]) [13].

The inclusion and exclusion criteria and the wbCT scans examined in the study period are shown in [Fig. 1]. 308 evaluable datasets were included in the study. All evaluated wbCT scans were acquired on the CT unit installed in the emergency room (Toshiba Aquilion 32®, software version V3.20GR011, Toshiba Medical Systems Europe, Zoetermeer, The Netherlands). Compliance tests and regular consistency checks were performed according to § 16 of the X-Ray Ordinance. Iterative reconstructions were not available in the study period.

Zoom Image
Fig. 1 This figure shows the number of whole-body CT scans performed in the study period and assessed by inclusion and exclusion criteria. Furthermore, the number of included and excluded scans is given. Included in the study were all primary care patients who underwent whole-body CT after severe trauma on our emergency room scanner using one of the established protocols. The following data sets were excluded: underage patients, missing dose report in the PACS system, discrepancy between used and established CT protocol, technical problems occurring during the scan and relevant deviation from the primarily defined scan length of the scan regions ([Fig. 3]).

Three different wbCT protocols were evaluated in the study period. The protocol referred to in the following as the “OLD protocol” was used to 2011 for all trauma patients (group O). With the goal of dose optimization, various changes were made in 2011 and two new differently weighted protocols were created (“TIME protocol” – group T and “DOSE protocol” – group D). These were presented to the hospital’s own trauma committee and have been used since 8/1/2011. The decision as to which of the two new protocols was to be used was made as a function of the patient’s clinical condition based on the decision algorithm shown in [Fig. 2].

Zoom Image
Fig. 2 During emergency room treatment, the decision as to which CT protocol should be used (TIME optimized versus DOSE optimized Protocol) was made according to this scheme. In hemodynamically unstable patients the TIME protocol was used and in patients with stable vital parameters the DOSE protocol was used. If injuries of one or both arms were present in the initial body check, either the modified DOSE protocol with the injured arm alongside the body or the TIME protocol with both arms alongside the body was used.

The same scan segments were used for all three protocols: one native head/neck scan and one contrast-enhanced body scan. In approximately 1/3 of all patients, a leg scan was additionally performed. However, this is not included in the present publication. While the body and leg scans were planned with an overlap in group O, this overlapping was avoided in the new protocols (T + D) ([Fig. 3]).

Zoom Image
Fig. 3 All of the three protocols evaluated in this study used the same scan sections: An unenhanced head/neck scan planned from vertex to thoracic vertebra 3 was followed by a contrast-enhanced scan of the body region planned from orbital roof to ischium. If severe injuries of the legs were suspected, a facultative subsequent scan of the legs was added. In the OLD protocol, the body and leg scans were planned with an overlap, which was explicitly avoided in both new protocols. The leg scans are not presented in this study. The OLD and TIME protocol had the same protocol design with arm positioning alongside the body and automated consecutively running scans. The DOSE protocol had a modified protocol design with a pause between the head/neck scan and body scan in which the arms were repositioned over the head. To ensure operational reliability of automatic exposure control, it is indispensable to acquire a new scout after repositioning the arms.

All head/neck scans were examined without dose modulation (= automatic exposure control = AEC) and without gantry tilting. All three protocols used the existing AEC (SUREExposure3D™, Toshiba Medical Systems Europe, Zoetermeer, The Netherlands) in the body scan. The exact technical specifications of the three protocols are shown in [Fig. 4].

Zoom Image
Fig. 4 Examination protocols. Collimation: detector rows × slice thickness; MPR: multiplanar reconstructions; AEC: automatic exposure control; FOV: field-of-view.

The arms were positioned close to the body in group O and the complete wbCT scan was planned based on a long whole-body scout. After simultaneous start of the scanner and contrast agent via injector, the wbCT scan was performed automatically: After acquisition of the native head/neck scan, an interscan delay and a venous contrast-enhanced body scan were performed.

Group T used the same automatic examination procedure with the arms close to the body. However, on the basis of a new reference convolution kernel (FC17), the body scan was acquired with a revised AEC setting ([Fig. 4]).

Group D was examined with a divided protocol design: After acquisition of a short scout, the native head/neck scan was planned and acquired. The patient’s arms were then positioned overhead. The body scan was then planned and acquired on the basis of a new long scout. This additional scout acquisition after arm repositioning is absolutely necessary so that the device recognizes the correct patient geometry for the body scan under consideration of the arms and can thus ensure an appropriate AEC.

The sex and age of the patient on the examination day (difference between birth and examination date) were recorded. Waist circumference was defined as the circumference of a region of interest (ROI) at skin level at the height of the navel. ([Fig. 5a]). The injury severity was quantified via the ISS which was determined on the basis of the radiology wbCT findings and the discharge letters [18]. The examination time was calculated as the difference between the parameters “study_time” (DICOM entry 0008|0030 – selection of the protocol at the CT unit) and the “acquisition_time” of the last image of the axial primary acquisition (DICOM entry 0008|0032). Thus, the scout acquisition, examination planning, contrast delay, actual scan time, and repositioning of the arms are taken into consideration. The objective image quality was determined as the standard deviation of the Hounsfield units in an ROI (= “image noise”). One ROI was placed in liver segment VII and one in the aorta ([Fig. 5b]) [19] [20]. The size and shape of the ROI were individually adapted to include as many voxels as possible and rule out vascular calcification, liver vessels, and parenchymal lesions (ROI circumference: Aorta 5.4 ± 1 cm; liver 9 ± 1.6 cm). From the dose reports in the PACS, the scan length (difference between start and end position) and the dose length product (DLP) were documented separately for the scans of the head/neck and body. The maximum CT dose index (CTDImax) documented at that time in the PACS dose reports is not suitable for a comparison to the literature. Therefore, the CTDIvol was calculated as the quotient of DLP and scan length. The effective dose (E) of the individual scans was calculated as the product of the relevant DLP and the corresponding conversion factor (k-factor). The k-factors published by Huda for the head/neck scan (kHead/Neck = 0.0045 mSv/mGy*cm, in relation to 16 cm head phantom) and the body scan (kWhole-Body = 0.0154 mSv/mGy*cm, in relation to 32 cm body phantom) according to ICRP103 [21] [22] were used for this purpose. E for a complete wbCT scan from vertex to ischium was calculated as the sum of the individual values.

Zoom Image
Fig. 5a Waist circumference was defined as the circumference of a freehand ROI at skin level at the height of the navel. b Objective image quality was defined as “image noise” (= standard deviation of Hounsfield units in an ROI). The ROIs were placed in liver segment VII and in the aorta.

The statistical analysis was performed with SPSS (Version 16.0, SPSS Inc., Chicago, IL, US). Extreme values (values outside the 3 rd interquartile range) were excluded from the analysis. The Kolmogorov-Smirnov test was used to test for normal distribution. Since there was no normal distribution in all three groups in any comparison, nonparametric tests were exclusively used. Therefore, results are specified as median and range (min-max). For better comparability with the literature, the mean and standard deviation (mean±SD) are additionally presented. The test for significance was performed with the Kruskal-Wallis test or Mann-Whitney U-test. The level of significance was 0.05.


#

Results

In the study period 308 patients fulfilled the inclusion and exclusion criteria (age: 46 a, 18 – 90 a; 46 ± 19 a). Neither gender distribution (male: n = 238; = 77 %; P = 0.64) nor waist circumference (93 cm, 66 – 145 cm; 94 ± 15 cm; P = 0.59) showed significant differences between the three groups.

The results regarding ISS and examination time and the ISS distribution within the groups are shown in [Fig. 6], [7]. While the examination time varied only slightly between groups O and T, the examination time in group D was 3.6 minutes longer because the automatic examination procedure was not used.

Zoom Image
Fig. 6 This figure presents the results of examination time and ISS as median and range. For comparison mean value and standard deviation are given additionally. s = significant; ns = not significant. * In the analysis of examination time, some extreme values (values out of 3 rd IQR) were excluded from further analysis (group OLD (O): 3, group TIME (T): 3, group DOSE (D): 2).
Zoom Image
Fig. 7 The ISS was calculated using the findings of the whole-body CT scan and diagnoses from the hospital discharge letter [18]. The OLD protocol was used for all patients in the sense of a “one-fits-all” concept. Since 2011, two new and differently weighted CT protocols have been implemented ([Fig. 2]). From this data it is evident that more severely injured patients were found in the TIME group, and more minorly injured patients in the DOSE group. This reflects the triage carried out in the emergency room.
The box represents the interquartile range (IQR) which contains the middle 50 % of the ISS values. The line across the box indicates the median. The whiskers (lines that extend from the upper and lower edge of the box) represent the highest and lowest values which are no greater than 1.5 times the IQR. Outliers are values beyond the whiskers between 1.5 and 3 times the IQR.

The results regarding objective image quality are specified in [Fig. 8]. There was no difference between groups O and T, while significantly better image quality was seen in group D.

Zoom Image
Fig. 8 The objective image quality was defined as image noise (= IN = standard deviation of Hounsfield units in a ROI). The respective ROIs were placed in one slice at the level of the upper abdomen in liver segment VII and in the aorta. In addition to the results (Median/Range) mean value and standard deviation are presented for better comparison to literature. s = significant; ns = not significant. * In the analysis of image noise, some extreme values (values outside of the 3 rd IQR) were excluded from further analysis (group TIME (T): aorta-2, liver-4).

[Fig. 9] shows the determined dose values of the head/neck and body scans. There were no significant or clinically relevant differences in dose parameters between the three groups for the head/neck scans. In contrast, significant dose reductions could be achieved for the body scan. The comparison between groups O and T shows that the DLP is 37 % lower (927 mGy*cm) under optimized configuration of the AEC with the same image quality. In the comparison between groups T and D, the DLP was able to be reduced by an additional 29 % (466 mGy*cm) by repositioning the patient’s arms.

Zoom Image
Fig. 9 This figure presents the dose evaluations of the single scans as median and range. Mean values and standard deviation are given for better comparison to literature additionally. Scan length and dose length product (DLP) incl. overranging. The effective dose E was calculated using conversion factors according to the ICRP103 [21] [22]. Because of identical settings in the head/neck scans, no differences were found among the three groups (reference to the 16 cm head phantom). In contrast, through optimization of automatic exposure control (comparing group OLD with TIME) and through repositioning of the arms (comparing group TIME with DOSE), significant dose reductions were achieved in the body scan (reference to the 32 cm body phantom). s = significant; ns = not significant.

The average effective dose of a wbCT from vertex to ischium is compared in [Fig. 10] to the dose reference values of the Federal Agency for Radiation Protection [23].

Zoom Image
Fig. 10 In section 1 for each protocol the effective doses of the single scan regions ([Abb. 9]) were added to calculate the mean effective dose for a whole-body CT scan from vertex to ischium. For reference in section 2, the dose reference values (= DRVs) of the German Federal Office for Radiation Protection are shown in column DLP-DRVold and column DLP-DRVnew (updated in 2016). There was no value in the old DRV for the CT scan of the neck region. The corresponding value of the new DRV was used instead. The DRVs were multiplied with conversion factors according to ICRP103 [21] [22] to calculate “effective dose reference values”, shown in column DRVold-EHuda and DRVnew-EHuda. The aim was to generate a maximum limit for an elective whole-body CT scan. All effective doses found in this study exceeded the current DRVs. When using whole-body CT as an emergency diagnostic modality in multiple trauma, it is, however, acceptable to exceed the DRV.

#

Conclusion

  1. An optimized AEC setting can significantly reduce the radiation exposure even in the case of older devices.

  2. Due to the constant further development of CT devices by manufacturers, existing protocols should be regularly reevaluated and optimized.

  3. An examination of the body with the arms overhead has a longer examination time, but is superior with regard to image quality and dose to an examination with the arms positioned close to the body.

  4. The clinically adapted dual concept with 2 differently weighted wbCT protocols is superior to the use of a single protocol for all patients (“one-fits-all concept”).

  5. WbCT scans result in relevant radiation exposure. Multicenter studies quantifying the average radiation exposure of a wbCT scan are needed.


#

Discussion

In the comparison between groups O and T, optimizing the AEC results in a reduction in radiation exposure with a similar examination time and stable image quality. At an AEC setting identical to that of group T, the repositioning of the patient’s arms and the modified protocol design in group D result in a longer examination time with reduced radiation exposure and improved image quality.

The patient group examined in the present study is to be considered representative compared to the TraumaRegister (annual report 2016: MeanAge 51.4 a; 70 % male) [6].

Since CT takes the patient’s habitus into consideration based on scouts when using AEC, it has a direct influence on dose. Since there were no significant differences regarding waist circumference between the three groups, there was no falsification of dose values due to asymmetrical distribution of patients with a different habitus among the groups.

Comparison of the three groups showed a clear injury severity dynamic ([Fig. 7]). In the OLD protocol previously used for all patients, the ISS covers the complete range of 0 – 75 points. In contrast, higher ISSs have been redistributed to group T and lower scores to group D since implementation of the clinically adapted procedure. This reflects the clinical triage done in our emergency room. In total, the ISS was higher than 15 points in 139 patients, theoretically resulting in overtriage of 54.9 % of patients according to Wurmb [17]. The literature shows a broad range of published ISS values: Huber-Wagner: ISS mean 28.8 ± 12.1 [2], Yaniv: mean 13 ± 11.2 [24] or median 5 (IQR 1 – 13) [16]. Differences can be primarily explained by variations in study design. Many studies on polytrauma wbCT neglect this important characteristic of trauma patients. Therefore, future studies should additionally determine the more precise NISS (new ISS).

With the same protocol design, there was a similar examination time in groups O and T (approx. 4 min). As a result of the modified protocol design, the examination time in group D increased to 7.7 minutes. A literature comparison is only possible on a limited basis since the definitions of examination time vary. The scan time (= time that radiation is emitted) is unsuitable as a reference value (e. g. 35 – 65 s [25]). The definition of examination time used in this study, i. e., the difference between protocol selection at the CT device and acquisition time of the last axial image of the primary acquisition, takes into account scout acquisition, planning, scan time, contrast delay, and any arm repositioning and is thus a realistic variable. Extreme values (life-saving procedures , reanimation) were excluded from the statistical analysis. The optimal definition of examination time to be preferred by future authors was used by Yaniv and Bayer (difference between acquisition time of the first scout image and the last axial image) [24] [26]. The examination time of 14.1 ± 4.1 to 14.3 ± 9.5 minutes or 7:30 ± 2:52 min:s is comparable with the present study. In the case of unstable patients requiring a short wbCT examination time, the protocol design of groups O and T with an examination time of 4 minutes is a real time-optimized protocol.

The occurrence of typical artifacts with the patient’s arms close to the body is generally known and is common in obese patients ([Fig. 11]). The definition of objective image quality used in this study provides values for organ-specific image noise and is used to compare the protocols evaluated here. In the case of differences regarding CT unit, AEC functionality, and ROI position, comparison of the individual values with other studies is not useful. Other studies with a similar method come to the same conclusion: wbCT scans with the patient’s arms close to the body have worse image quality and a higher dose than scans with the patient’s arms overhead [19] [20].

Zoom Image
Fig. 11 The figure demonstrates different examples of single slices of body scans with arms over the head (group DOSE – ex. a) and arms close to body (group TIME – ex. b – d). Characteristic beamhardening artifacts due to the positioning of the arms are present and can significantly impact the interpretation of the diagnostic procedure, especially in obese patients.

Since the head/neck scans were examined with the same setting, no differences could be identified between the groups as expected. It must be noted here that the neck was overexposed when examining this region with a constant tube current and a dose for intracranial image quality. This is reflected in the dose values that exceed the dose reference values of the Federal Agency for Radiation Protection ([Fig. 9], [10]) [23]. This aspect was identified as an object of future protocol optimization.

Comparing body scans shows similar scan lengths between groups T and D which are shorter than those in group O. This is a reasonable result of the increased planning by technologists in the new protocols since the overlapping of body and leg scans is to be explicitly avoided.

In the group comparison of dose values, significant dose reductions ([Fig. 9]) were achieved both by optimizing the AEC using current software (O-T) and by repositioning the patient’s arms (T-D). Both aspects are considered relevant variables influencing dose and image quality [19] [20] [27] [28]. In the case of an identical AEC setting, the difference in the positioning of the arms between groups T and D is the only possible explanation for the variations in image quality and dose ([Fig. 4]). The dose reduction resulting from the repositioning of the patient’s arms was 7 mSv according to Brink which is comparable with the present study (7.2 mSv) [20]. However, such values cannot be generalized since significant differences can occur when examining the same patient on a device by the same manufacturer simply due to the use of a different AEC setting. It can thus be concluded that existing protocols must be regularly reevaluated and optimized due to the constant further developments made by CT device manufacturers.

The dose reference values of the Federal Agency for Radiation Protection were updated in 2016 and provide guidelines to be observed during typical examinations of standard patients (body weight 70 ± 3 kg) [23]. Depending on the indication, habitus, and special aspects (metal implants) and particularly in the case of emergency diagnostic imaging, it is acceptable to exceed the dose reference values. [Fig. 10] shows that the doses particularly in groups O and T were above the current dose reference values, with the doses published here being acquired in the validity period of the old dose reference values between 2011 and 2013. Additional protocol optimizations have since been performed and are the subject of a planned follow-up study.

The general opinion in the literature is that wbCT from vertex to ischium corresponds to approx. 10 – 20 mSv [13]. The present values as well as the speculation of Ruchholtz (30.05 mSv) contradict this [9]. Harrieder describes a broad range of dose values for wbCT in the literature [29]. From our point of view there are many reasons for this:

  1. There are various wbCT protocols partially with multiphase scans and multiple overlaps.

  2. Many studies publish only dose values for body scans neglecting the head/neck scans.

  3. Some studies provide data regarding radiation exposure [19] [20] [24] [27] [29], while others use phantom studies and others provide estimated values that cannot be reproduced.

  4. Dose values are published in a highly varied and sometimes incomplete manner. In addition, different k-factors based on either tissue weighting factors according to ICRP 60 or the current ICRP 103 are used [21] [22] [30]. Depending on the published, dose-relevant parameters, conversion and thus comparability with other studies can be limited or even impossible. Converting the published values using the k-factors from Huda to values comparable with this study would result in dose values of 41.1 – 49.5 mSv [25], 26.2 – 28 mSv [29], 15.7 – 20.2 mSv [27], 11.4 – 28.2 mSv [20] or 12.4 – 18.2 mSv [24]. It must be explicitly noted here that the last three studies provided values only for body scans but not for head/neck scans. To date, no anonymized retrospective multicenter evaluation of this topic has been performed.

A comparison of a “one-fits-all” concept with the clinically adapted concept using the latest hardware and software yields the following points:

Supplementing the prior exclusive use of a protocol with the patient’s arms close to the body for all patients (TIME protocol) with the addition of a DOSE protocol makes it possible to examine stable patients according to the ALARA principle with an adequate dose and better image quality. This results in reduced collective radiation exposure which particularly benefits slightly injured patients who were overtriaged in the emergency room.

In the case of the previous sole use of a DOSE protocol with repositioning of the patient’s arms overhead for all patients, particularly severely injured, hemodynamically unstable patients benefit from the addition of a TIME protocol due to the reduced examination time and the more time-efficient care. Higher radiation exposure is acceptable for these patients during emergency diagnostic imaging.

By using differently weighted wbCT protocols, each patient’s clinical situation can be approached with greater flexibility on an individualized basis.

In patients with a relevant mechanism of trauma, there is an alternative diagnostic concept consisting of the combination of ultrasound, X-rays of the skeleton of the trunk and if necessary subsequent CT of individual body regions. Adding up the literature values yields an effective dose of 3.32 mSv for X-ray imaging of the cervical spine, thoracic spine, lumbar spine, thorax, and pelvis [31]. This corresponds to 1/10 the dose of one wbCT ([Fig. 10]). Imaging of the thorax p. a. corresponds to 0.02 mSv. Therefore, one wbCT at 30 mSv would equal 1500 chest X-rays.

According to the Federal Agency for Radiation Protection, the average annual radiation exposure in Germany from natural and civilization-based sources is 4 mSv. Trauma patients are often very young, and they have often only encountered natural radiation exposure (2.1 mSv). wbCT with 30 mSv would therefore correspond to natural radiation exposure of 14.3 a.


#

Limitations

The basic limitations of this study are its retrospective, monocentric, unblinded, and non-randomized design.

Particularly in obese patients, there is high variability regarding the position of the navel. Therefore, future studies should determine the waist circumference along the X-axis not based on the navel as in this study but based on bony landmarks (e. g. L3).

In the case of examinations with the arms close to the body, the actual body volume exposed to radiation is represented incompletely through the waist circumference. Therefore, no correlations between waist circumference and dose are presented in this study.

The definition of examination time that is used in this study is susceptible to outliers, e. g. due to prolonged emergency room care. The definition used by Yaniv and Bayer is to be given preference for future studies [24] [26].

In the analysis of examination time, no differentiation was made between wbCT with and without a leg scan. This results in inaccuracies in the examination time in patients who underwent a leg scan (frequency: T > O > D) of approx. 20 s (5 s interscan delay with respect to the body scan and a scan time of approx. 15 s for leg scans). This explains the difference in examination time between groups O and T in particular.

The examination of all patients on a CT-compatible spine board is a further limitation as it can have a significant effect on dose and image quality [32]. This was not evaluated in this study. Since all patients were examined in this way, this represents a systematic error and is to be ignored within this study.


#

Conclusion

Acute care of polytraumas is an interdisciplinary challenge for trauma centers. There are usually interdisciplinary guidelines for procedures that can differ significantly between various hospitals. Diagnosis via wbCT is an important component of a total concept and is subject to complex influences and interactions. In addition to technical factors of the CT unit (manufacturer, product, software options, maximum scan length, iterative reconstructions) and the actual device settings (tube voltage, rotation time, AEC), spatial, logistic, and interdisciplinary aspects (CT in the emergency room or separate rooms, emergency room care procedures, head/feet-first position, arm position, contrast regime) also affect the protocol design. Due to this high complexity and based on the focus compromises must always be made compared to elective standard protocols for individual body regions. There are many publications addressing protocol optimization in wbCT, for example in relation to the manufacturer, in relation to the device, in relation to individual aspects of protocol design, and in relation different designs compared to one another. A multicenter study more precisely quantifying the average radiation exposure caused by wbCT is not yet available.

Additional research regarding the topic of indication determination is also needed to develop a practical concept that rules out undertriage to the greatest extent possible and keeps the overtriage rate relatively low. The present study describes a clinically adapted concept that can be integrated between indication determination and the performance of wbCT. It allows a more flexible approach to individual aspects of trauma patients in the emergency room and provides an option for lowering the collective radiation exposure even when performing wbCT.

Clinical relevance of the study
  • The present study describes a concept for wbCT in patients with multiple injuries: The use of 2 differently weighted protocols allows a more flexible approach to the clinical presentation of trauma patients in the emergency room.

  • The time-optimized wbCT protocol targets severely injured patients. The focus is on the shortest possible examination time with compromises being made regarding image quality and radiation exposure.

  • The dose-optimized wbCT protocol targets more slightly injured, possibly overtriaged patients. The focus is on the lowest possible radiation exposure with the use of as many dose reduction options as possible.

  • The clinically adapted concept presented here may be a further step in improving the acute care of trauma patients in Germany with the goal of reducing the collective dose.


#
#

Acknowledgments

The present study was performed between 2010 and 2015 (data acquisition 2010 – 2013) at BG Hospital Bergmannstrost in Halle (Saale). It is the publication of an excerpt from the doctoral thesis of the first author. We would like to thank the Department of Diagnostic Imaging and Interventional Radiology (BG-Klinikum Bergmannstrost, Halle (Saale), Germany) for providing the data that was used.

* Board Member, Working Group Imaging Procedures of the Musculosceletal System of the German Röntgen Society, Berlin, Germany.



Correspondence

Stefan Ulrich Reske
Institut für Diagnostische und Interventionelle Radiologie und Neuroradiologie, Heinrich-Braun-Klinikum gemeinnützige GmbH
Karl-Keil-Straße 35
08060 Zwickau
Germany   
Phone: ++ 49/3 75/51 55 48 38   


Zoom Image
Fig. 1 This figure shows the number of whole-body CT scans performed in the study period and assessed by inclusion and exclusion criteria. Furthermore, the number of included and excluded scans is given. Included in the study were all primary care patients who underwent whole-body CT after severe trauma on our emergency room scanner using one of the established protocols. The following data sets were excluded: underage patients, missing dose report in the PACS system, discrepancy between used and established CT protocol, technical problems occurring during the scan and relevant deviation from the primarily defined scan length of the scan regions ([Fig. 3]).
Zoom Image
Fig. 2 During emergency room treatment, the decision as to which CT protocol should be used (TIME optimized versus DOSE optimized Protocol) was made according to this scheme. In hemodynamically unstable patients the TIME protocol was used and in patients with stable vital parameters the DOSE protocol was used. If injuries of one or both arms were present in the initial body check, either the modified DOSE protocol with the injured arm alongside the body or the TIME protocol with both arms alongside the body was used.
Zoom Image
Fig. 3 All of the three protocols evaluated in this study used the same scan sections: An unenhanced head/neck scan planned from vertex to thoracic vertebra 3 was followed by a contrast-enhanced scan of the body region planned from orbital roof to ischium. If severe injuries of the legs were suspected, a facultative subsequent scan of the legs was added. In the OLD protocol, the body and leg scans were planned with an overlap, which was explicitly avoided in both new protocols. The leg scans are not presented in this study. The OLD and TIME protocol had the same protocol design with arm positioning alongside the body and automated consecutively running scans. The DOSE protocol had a modified protocol design with a pause between the head/neck scan and body scan in which the arms were repositioned over the head. To ensure operational reliability of automatic exposure control, it is indispensable to acquire a new scout after repositioning the arms.
Zoom Image
Fig. 4 Examination protocols. Collimation: detector rows × slice thickness; MPR: multiplanar reconstructions; AEC: automatic exposure control; FOV: field-of-view.
Zoom Image
Fig. 5a Waist circumference was defined as the circumference of a freehand ROI at skin level at the height of the navel. b Objective image quality was defined as “image noise” (= standard deviation of Hounsfield units in an ROI). The ROIs were placed in liver segment VII and in the aorta.
Zoom Image
Fig. 6 This figure presents the results of examination time and ISS as median and range. For comparison mean value and standard deviation are given additionally. s = significant; ns = not significant. * In the analysis of examination time, some extreme values (values out of 3 rd IQR) were excluded from further analysis (group OLD (O): 3, group TIME (T): 3, group DOSE (D): 2).
Zoom Image
Fig. 7 The ISS was calculated using the findings of the whole-body CT scan and diagnoses from the hospital discharge letter [18]. The OLD protocol was used for all patients in the sense of a “one-fits-all” concept. Since 2011, two new and differently weighted CT protocols have been implemented ([Fig. 2]). From this data it is evident that more severely injured patients were found in the TIME group, and more minorly injured patients in the DOSE group. This reflects the triage carried out in the emergency room.
The box represents the interquartile range (IQR) which contains the middle 50 % of the ISS values. The line across the box indicates the median. The whiskers (lines that extend from the upper and lower edge of the box) represent the highest and lowest values which are no greater than 1.5 times the IQR. Outliers are values beyond the whiskers between 1.5 and 3 times the IQR.
Zoom Image
Fig. 8 The objective image quality was defined as image noise (= IN = standard deviation of Hounsfield units in a ROI). The respective ROIs were placed in one slice at the level of the upper abdomen in liver segment VII and in the aorta. In addition to the results (Median/Range) mean value and standard deviation are presented for better comparison to literature. s = significant; ns = not significant. * In the analysis of image noise, some extreme values (values outside of the 3 rd IQR) were excluded from further analysis (group TIME (T): aorta-2, liver-4).
Zoom Image
Fig. 9 This figure presents the dose evaluations of the single scans as median and range. Mean values and standard deviation are given for better comparison to literature additionally. Scan length and dose length product (DLP) incl. overranging. The effective dose E was calculated using conversion factors according to the ICRP103 [21] [22]. Because of identical settings in the head/neck scans, no differences were found among the three groups (reference to the 16 cm head phantom). In contrast, through optimization of automatic exposure control (comparing group OLD with TIME) and through repositioning of the arms (comparing group TIME with DOSE), significant dose reductions were achieved in the body scan (reference to the 32 cm body phantom). s = significant; ns = not significant.
Zoom Image
Fig. 10 In section 1 for each protocol the effective doses of the single scan regions ([Abb. 9]) were added to calculate the mean effective dose for a whole-body CT scan from vertex to ischium. For reference in section 2, the dose reference values (= DRVs) of the German Federal Office for Radiation Protection are shown in column DLP-DRVold and column DLP-DRVnew (updated in 2016). There was no value in the old DRV for the CT scan of the neck region. The corresponding value of the new DRV was used instead. The DRVs were multiplied with conversion factors according to ICRP103 [21] [22] to calculate “effective dose reference values”, shown in column DRVold-EHuda and DRVnew-EHuda. The aim was to generate a maximum limit for an elective whole-body CT scan. All effective doses found in this study exceeded the current DRVs. When using whole-body CT as an emergency diagnostic modality in multiple trauma, it is, however, acceptable to exceed the DRV.
Zoom Image
Fig. 11 The figure demonstrates different examples of single slices of body scans with arms over the head (group DOSE – ex. a) and arms close to body (group TIME – ex. b – d). Characteristic beamhardening artifacts due to the positioning of the arms are present and can significantly impact the interpretation of the diagnostic procedure, especially in obese patients.
Zoom Image
Abb. 1 Die Abb. zeigt die Anzahl der im Studienzeitraum durchgeführten, auf die Ein- und Ausschlusskriterien geprüften und letztlich ein- bzw. ausgeschlossenen GKCTs. Eingeschlossen wurden alle Patienten, die im Rahmen der Primärversorgung nach Trauma eine GKCT mit einem zum jeweiligen Zeitpunkt etablierten GKCT-Protokoll am Schockraum-CT erhielten. Ausgeschlossen wurden Patienten unter 18 Jahren, bei fehlendem Dosisbericht im PACS, bei Abweichungen vom etablierten Protokoll, bei technischen Problemen während des Scans und wenn die vorab definierten Scanabschnitte über- oder unterschritten wurden ([Abb. 3]).
Zoom Image
Abb. 2 Anhand dieses Algorithmus erfolgte während der Schockraumversorgung von Trauma-Patienten nach Stellung der rechtfertigenden Indikation zur GKCT die Entscheidung über das anzuwendende CT-Untersuchungsprotokoll.
Zoom Image
Abb. 3 Alle 3 evaluierten Protokolle hatten die gleichen Scanabschnitte. Der native Kopf-/Hals-Scan wurde von Scheitel bis BWK 3 geplant. Im Anschluss folgte ein KM-gestützter Scan des Körpers, der vom Orbita-Dach bis unter die Sitzbeine reichte. Bestand klinisch der Verdacht auf relevante Verletzungen der Beine, wurde fakultativ direkt im Anschluss ein venös kontrastierter Bein-Scan durchgeführt. Hierbei bestand beim ALT-Protokoll eine Überlappung zwischen Körper- und Bein-Scan, die jedoch bei den beiden neuen Protokollen explizit vermieden wurde. Die Bein-Scans werden in der vorliegenden Arbeit nicht präsentiert. Während ALT- und ZEIT-Protokoll das gleiche Protokolldesign mit Armlagerung am Körper und automatisch nacheinander ablaufenden Scans hatten, wurde der Ablauf beim DOSIS-Protokoll modifiziert und in einer Pause zwischen den Scans die Arme über den Kopf gelagert. Für die Funktionstüchtigkeit der AEC ist dabei die Akquisition eines neuen Scouts nach Umlagerung der Arme zwingend erforderlich.
Zoom Image
Abb. 4 Ausführliche Auflistung der technischen Spezifikationen der 3 evaluierten GKCT-Protokolle. Kollimation: Angabe in Detektorzeilen × Schichtdicke; MPR: Multiplanare Reformationen; AEC: Dosismodulation; FOV: Field-of-view.
Zoom Image
Abb. 5a Der Bauchumfang (BU) der Patienten wurde als Freihand-ROI im Hautniveau in Höhe des Nabels ermittelt. b Die objektive Bildqualität wurde als „Bildrauschen“ (= Standardabweichung der HU in einer Mess-ROI) ermittelt. Die ROIs wurden im Lebersegment VII und in der Aorta platziert.
Zoom Image
Abb. 6 In der Abbildung sind die Ergebnisse der UZ und des ISS als Median und Range dargestellt. Für den Literaturvergleich werden zusätzlich Mittelwert und Standardabweichung angegeben. s = signifikant; ns = nicht signifikant. * Bei der UZ wurden mehrere Extremwerte (Werte außerhalb des 3. IQR) identifiziert und aus der statistischen Analyse der UZ ausgeschlossen (Gruppe A: 3, Gruppe Z: 3, Gruppe D: 2).
Zoom Image
Abb. 7 Die ISS wurden in der vorliegenden Arbeit anhand der GKCT-Befunde und Entlassungsbriefe ermittelt [18]. Das ALT-Protokoll wurde im Sinne eines „One-fits-all-Konzepts“ für alle Patienten verwendet. Seit 2011 wurden 2 unterschiedlich gewichtete Protokolle vorgehalten und anhand des in [Abb. 2] dargestellten Algorithmus eingesetzt. In den Boxplots ist zu erkennen, dass in der ZEIT-Gruppe häufiger schwerverletzte und in der DOSIS-Gruppe häufiger leichtverletzte Patienten untersucht wurden. Dies spiegelt die im Schockraum durchgeführte Triage wider.
Die Box markiert den Interquartil-Abstand (IQA) der ISS-Werte je Gruppe, d. h. die mittleren 50 %. Die dicke Linie in der Box entspricht dem Median. Die Whisker (an der Ober- bzw. Unterkante der Box ansetzende Linien) stehen für Werte, die um den 1,5fachen IQA über- bzw. unterhalb des IQA liegen. Ausreißer (Werte außerhalb der Whisker (zwischen 1,5 und 3fachem IQA)) sind als Kreise dargestellt.
Zoom Image
Abb. 8 In der Abbildung sind die Ergebnisse der objektiven BQ (Bildrauschen = BR) als Median und Range dargestellt. Diese wurde als Standardabweichung der Hounsfield-Einheiten in einer ROI in einer Oberbauchschicht im Lebersegment VII und in der Aorta ermittelt. Für den Literaturvergleich werden zusätzlich Mittelwert und Standardabweichung angegeben. s = signifikant; ns = nicht signifikant. * Bei der BQ wurden mehrere Extremwerte (Werte außerhalb 3. IQR) identifiziert und aus der statistischen Analyse ausgeschlossen (Gruppe Z: Aorta 2, Leber 4).
Zoom Image
Abb. 9 In der Abbildung sind die Dosisevaluationen der Einzelscans dargestellt. Zusätzlich zu Median/Range werden für den besseren Literaturvergleich Mittelwert und Standardabweichung angegeben. Scanlänge und Dosislängenprodukt (DLP) inkl. Overranging. Die effektive Dosis E wurde mittels k-Faktoren nach ICRP103 berechnet [21] [22]. Bei gleichen Einstellungen des Geräts im Kopf-/Hals-Scan bestanden keine Unterschiede zwischen den 3 Gruppen (Bezug auf 16 cm-Kopfphantom). Im Gegensatz dazu wurden durch Optimierung der Dosismodulation (A – Z) und die Armumlagerung (Z – D) im Körperscan jeweils signifikante Dosisreduktionen erreicht (Bezug auf 32 cm-Körperphantom). s = signifikant; ns = nicht signifikant.
Zoom Image
Abb. 10 Für eine GKCT (Scheitel bis Sitzbein) wurden die gefundenen Dosiswerte der Einzelscans ([Abb. 9]) addiert und in Teil 1 der Abbildung gegenübergestellt. In Teil 2 wurden die alten Dosisreferenzwerte (DRW) und die aktuellen, seit 2016 gültigen DRW des Bundesamtes für Strahlenschutz mit k-Faktoren nach ICRP103 [21] [22] multipliziert und damit „effektive Referenzdosen“ für die Einzelscans berechnet. Da die alten DRW keinen Wert für die Hals-CT lieferten wurde hier stellvertretend der neue DRW verwendet, um vergleichen zu können. Ziel war die Ermittlung eines maximalen DRW-Rahmens für eine entsprechende elektive GKCT. Im Vergleich lagen alle Dosiswerte über den aktuellen DRW. Da es sich bei der GKCT um Notfalldiagnostik handelt, sind Überschreitungen der DRW jedoch legitim.
Zoom Image
Abb. 11 In der Abbildung sind verschiedene Beispiele von Einzelschichten von Körperscans mit Armen über dem Kopf (Gruppe DOSIS – Bsp. a) und mit Armen am Körper (Gruppe ALT/ZEIT – Bsp. b – d) dargestellt. Deutlich zu erkennen sind die typischerweise auftretenden, streifenförmigen Aufhärtungsartefakte durch die Arme. Dadurch kann insbesondere bei adipösen Patienten die Beurteilbarkeit der Untersuchung massiv eingeschränkt sein.