Endoscopy 2018; 50(12): 1236-1237
DOI: 10.1055/a-0677-1531
E-Videos
© Georg Thieme Verlag KG Stuttgart · New York

Chronic tracheoesophageal fistula successfully treated using Amplatzer septal occluder

Mario Traina
1   Endoscopy Service, Department of Diagnostic and Therapeutic Services, IRCCS – ISMETT, Palermo
,
Michele Amata
1   Endoscopy Service, Department of Diagnostic and Therapeutic Services, IRCCS – ISMETT, Palermo
,
Lavinia De Monte
2   Thoracic Surgery, Department of Diagnostic and Therapeutic Services, IRCCS – ISMETT, Palermo
,
Antonino Granata
1   Endoscopy Service, Department of Diagnostic and Therapeutic Services, IRCCS – ISMETT, Palermo
,
Dario Ligresti
1   Endoscopy Service, Department of Diagnostic and Therapeutic Services, IRCCS – ISMETT, Palermo
,
Ilaria Tarantino
1   Endoscopy Service, Department of Diagnostic and Therapeutic Services, IRCCS – ISMETT, Palermo
,
Caterina Gandolfo
3   Cardiac Cath Lab, Department of Diagnostic and Therapeutic Services, IRCCS – ISMETT, Palermo
› Author Affiliations
Further Information

Corresponding author

Michele Amata, MD
Endoscopy Service, IRCCS-ISMETT
Via Tricomi, 5
90127 Palermo
Italy   
Fax: +39-091-2192400 (specify Endoscopy Service)   

Publication History

Publication Date:
10 September 2018 (online)

 

Tracheoesophageal fistula (TEF) is a serious life-threatening condition that appears in critically ill patients with a prolonged history of mechanical ventilation. Enteral feeding and dietary support combined with surgery is the gold standard. In selected patients, TEF healing can be ensured by a mini-invasive approach using an Amplatzer septal occluder (AGA Medical Corporation), intended for cardiac septal defect closure [1]. The Amplatzer septal occluder, which is composed of a nitinol mesh, has two self-expandable disks connected by a thin diameter waist ([Fig. 1]) and ensures mechanical closure of the two sides of the fistula, making a potential platform for subsequent tissue ingrowth [2].

Zoom Image
Fig. 1 Longitudinal and transverse profile of the Amplatzer septal occluder showing its typical dumbbell shape. Source: Federico Amata.

This technique was used in a 44-year-old man with tracheostomy and a history of protracted invasive lung support. He had been diagnosed as having a TEF after numerous episodes of aspiration pneumonitis and had subsequently undergone anterior cervicotomy with surgical closure of the fistula. After 7 months, his dysphagia relapsed. Endoscopy confirmed recurrence of the TEF ([Fig. 2]), which was initially treated unsuccessfully by submucosal injection of acrylic glue.

Zoom Image
Fig. 2 Endoscopic diagnosis of tracheoesophageal fistula recurrence was made by the finding of: a a fissure (red dashed line) in the anterior wall of the esophagus; b in communication with the tracheal lumen.

Given the poor clinical condition of the patient and the failure of both surgical and endoscopic therapy, it was decided to try positioning of an Amplatzer septal occluder ([Video 1]). The procedure was performed using a gastroscope (GIF-1TH190; Olympus Europe) and a bronchoscope (BF-1T180; Olympus Europe). The TEF was cannulated using a papillotome (TRUETome; Boston Scientific) and a 0.025-inch guidewire (Jagwire; Boston Scientific) was inserted into the bronchial segment. The wire was then grasped with a biopsy forceps (Endo-Jaw; Olympus Europe) passed through the bronchoscope, providing countertraction by maintaining a straightened position. The septal occluder catheter was introduced and the two ends were released into the trachea and esophagus, respectively ([Fig. 3]). Successful closure of the TEF was confirmed by contrast medium injection ( [Fig. 4]). The procedure was uneventful.

Video 1 An Amplatzer septal occluder is placed for the treatment of a chronic tracheoesophageal fistula, a mini-invasive approach in a critically ill patient.


Quality:
Zoom Image
Fig. 3 Radiographic image showing the Amplatzer septal occluder following placement (red dashed line).
Zoom Image
Fig. 4 Radiographic image following injection of contrast medium showing successful closure of the tracheoesophageal fistula with no contrast leakage or stenosis of the esophageal lumen (red dashed lines).

Repeat endoscopy 4 weeks later showed that the stent remained in the correct position and the patient has remained asymptomatic during 12 months of follow-up.

In selected tertiary care centers, where advanced endoscopic and catheter lab suites are available, placement of an Amplatzer septal occluder is feasible and safe [3] [4], offering potential fistula closure, especially in patients in a critical condition, with severe comorbidities and recurrence of a TEF.

Endoscopy_UCTN_Code_TTT_1AO_2AI

Endoscopy E-Videos
https://eref.thieme.de/e-videos

Endoscopy E-Videos is a free access online section, reporting on interesting cases and new techniques in gastroenterological endoscopy. All papers include a high quality video and all contributions are freely accessible online.
This section has its own submission website at https://mc.manuscriptcentral.com/e-videos


#

Competing interests

None

Acknowledgments

The authors gratefully acknowledge the contribution of Federico Amata for his graphic support.

  • References

  • 1 Repici A, Presbitero P, Carlino A. et al. First human case of esophagus-tracheal fistula closure by using a cardiac septal occluder (with video). Gastrointest Endosc 2010; 71: 867-869
  • 2 Scordamaglio PR, Tedde ML, Minamoto H. et al. Endoscopic treatment of tracheobronchial tree fistulas using atrial septal defect occluders: preliminary results. J Bras Pneumol 2009; 35: 1156-1160
  • 3 Coppola F, Boccuzzi G, Rossi G. et al. Cardiac septal umbrella for closure of a tracheoesophageal fistula. Endoscopy 2010; 42: E318-E319
  • 4 ASGE Technology Committee. Banerjee S, Barth BA. et al. Endoscopic closure devices. Gastrointest Endosc 2012; 76: 244-251 Erratum in: Gastrointest Endosc 2013; 77: 833

Corresponding author

Michele Amata, MD
Endoscopy Service, IRCCS-ISMETT
Via Tricomi, 5
90127 Palermo
Italy   
Fax: +39-091-2192400 (specify Endoscopy Service)   

  • References

  • 1 Repici A, Presbitero P, Carlino A. et al. First human case of esophagus-tracheal fistula closure by using a cardiac septal occluder (with video). Gastrointest Endosc 2010; 71: 867-869
  • 2 Scordamaglio PR, Tedde ML, Minamoto H. et al. Endoscopic treatment of tracheobronchial tree fistulas using atrial septal defect occluders: preliminary results. J Bras Pneumol 2009; 35: 1156-1160
  • 3 Coppola F, Boccuzzi G, Rossi G. et al. Cardiac septal umbrella for closure of a tracheoesophageal fistula. Endoscopy 2010; 42: E318-E319
  • 4 ASGE Technology Committee. Banerjee S, Barth BA. et al. Endoscopic closure devices. Gastrointest Endosc 2012; 76: 244-251 Erratum in: Gastrointest Endosc 2013; 77: 833

Zoom Image
Fig. 1 Longitudinal and transverse profile of the Amplatzer septal occluder showing its typical dumbbell shape. Source: Federico Amata.
Zoom Image
Fig. 2 Endoscopic diagnosis of tracheoesophageal fistula recurrence was made by the finding of: a a fissure (red dashed line) in the anterior wall of the esophagus; b in communication with the tracheal lumen.
Zoom Image
Fig. 3 Radiographic image showing the Amplatzer septal occluder following placement (red dashed line).
Zoom Image
Fig. 4 Radiographic image following injection of contrast medium showing successful closure of the tracheoesophageal fistula with no contrast leakage or stenosis of the esophageal lumen (red dashed lines).