Removal of a migrated biliary stent using new digital cholangioscopy retrieval devices in a transplant patient

Francesc Bas-Cutrina
1   Endoscopy Unit, Department of Digestive Diseases, Hospital Universitari de Bellvitge-IDIBELL, University of Barcelona, Catalonia, Spain
,
Albert Garcia-Sumalla
1   Endoscopy Unit, Department of Digestive Diseases, Hospital Universitari de Bellvitge-IDIBELL, University of Barcelona, Catalonia, Spain
,
Julio Velasquez
1   Endoscopy Unit, Department of Digestive Diseases, Hospital Universitari de Bellvitge-IDIBELL, University of Barcelona, Catalonia, Spain
,
Claudia F. Consiglieri
1   Endoscopy Unit, Department of Digestive Diseases, Hospital Universitari de Bellvitge-IDIBELL, University of Barcelona, Catalonia, Spain
,
Laura Lladó
2   Liver Transplant Unit, Department of Surgery, Hospital Universitari de Bellvitge-IDIBELL, University of Barcelona, Catalonia, Spain
,
Joan B. Gornals
1   Endoscopy Unit, Department of Digestive Diseases, Hospital Universitari de Bellvitge-IDIBELL, University of Barcelona, Catalonia, Spain
3   Faculty of Health Sciences, Universitat Oberta de Catalunya, Barcelona, Catalonia, Spain
› Author Affiliations
Further Information

Corresponding author

Joan B. Gornals MD, PhD
Endoscopy Unit, Dept. of Digestive Diseases
Hospital Universitari de Bellvitge-IDIBELL (Bellvitge Biomedical Research Institute)
Feixa Llarga s/n, 08907 L’Hospitalet de Llobregat
Barcelona
Catalonia
Spain   
Fax: +34-93-2607681   

Publication History

Publication Date:
04 June 2019 (online)

 

A 51-year-old man who had undergone liver transplantation developed a symptomatic anastomotic biliary stricture 23 months after surgery. Endoscopic biliary therapy via endoscopic retrograde cholangiopancreatography (ERCP) was planned. Progressive biliary balloon dilation of the stenosis was performed, with placement of three coaxial plastic stents (8.5-Fr × 12 cm, 8.5-Fr × 9 cm, and 10-Fr × 12 cm; Advanix, Boston Scientific, Natick, Massachusetts, USA).

During an endoscopy to replace the stents, fluoroscopy revealed proximal migration of an 8.5-Fr plastic stent at the level of the cystic insertion ([Fig. 1]). Several failed extraction attempts were made using the standard ERCP techniques (i. e. extractor balloon, Lasso technique, and others) [1] [2]. Single-operator peroral intraductal cholangioscopy (SpyGlass DS direct visualization system, Boston Scientific) confirmed impaction of the distal end of the proximally migrated stent, located 3 cm proximally to the duodenal papilla. An attempt to mobilize the migrated stent was made using biopsy forceps (SpyBite, Boston Scientific), without success.

Zoom Image
Fig. 1 Fluoroscopy image during an endoscopic retrograde cholangiopancreatography plastic stent exchange session showing that one 8.5-Fr plastic stent has migrated proximally (yellow arrow), while the other two plastic stents remain well positioned (red arrows).

Finally, use of the new intraductal cholangioscopy retrieval devices (basket and snare) ([Fig. 2]) allowed successful stent removal. Firstly, the stent was disimpacted using the new SpyBasket (Boston Scientific), then the distal end was wrapped by the SpySnare (Boston Scientific) ( [Fig.3]), and the migrated stent was totally extracted ([Video 1]).

Zoom Image
Fig. 2 The new intraductal cholangioscopy retrieval devices: a the basket (SpyBasket); b the snare (SpySnare). Source: Boston Scientific.
Zoom Image
Fig. 3 Fluoroscopy image showing the intraductal cholangioscope (SpyScope DS; orange arrow) and the new retrieval snare (SpySnare; blue arrow) wrapping around the proximally migrated plastic stent (yellow arrow).

Video 1 Successful disimpaction and removal of a proximally migrated plastic stent using new retrieval devices for single-operator peroral cholangioscopy (SpyBasket and SpySnare) in a patient who had undergone liver transplantation. Source for stents: Boston Scientific.


Quality:

Proximal migration of biliary stents occurs in around 5 % of ERCP procedures, despite the antimigratory design of the stents [3]. Although in most cases extraction is possible, it may be challenging and technically difficult. Sporadically, the use of single-operator cholangioscopy and biopsy forceps has been reported [4].

This is the first case to report the use of the new intraductal cholangioscopy retrieval devices for the removal of a migrated plastic stent. The appearance of these dedicated tools expands the therapeutic armamentarium in the biliary tract and pancreatic duct and allows for single-operator peroral cholangiopancreatoscopy.

Endoscopy_UCTN_Code_TTT_1AR_2AZ

Endoscopy E-Videos
https://eref.thieme.de/e-videos

Endoscopy E-Videos is a free access online section, reporting on interesting cases and new techniques in gastroenterological endoscopy. All papers include a high quality video and all contributions are freely accessible online.
This section has its own submission website at https://mc.manuscriptcentral.com/e-videos


#

Competing interests

None

  • References

  • 1 Tarnasky PR, Cotton PB, Baillie J. et al. Proximal migration of biliary stents: attempted endoscopic retrieval in forty-one patients. Gastrointest Endosc 1995; 42: 513-520
  • 2 Chaurasia OP, Rauws EA, Fockens P. et al. Endoscopic techniques for retrieval of proximally migrated biliary stents: the Amsterdam experience. Gastrointest Endosc 1999; 50: 780-785
  • 3 Okabe Y, Tsuruta O, Kaji R. et al. Endoscopic retrieval of migrated plastic stent into bile duct or pancreatic pseudocyst. Dig Endosc 2009; 21: 1-7
  • 4 Ogura T, Okuda A, Miyano A. et al. Migrated endoclip removal after cholecystectomy under digital single-operator cholangioscopy guidance. Endoscopy 2018; 50: 74-75

Corresponding author

Joan B. Gornals MD, PhD
Endoscopy Unit, Dept. of Digestive Diseases
Hospital Universitari de Bellvitge-IDIBELL (Bellvitge Biomedical Research Institute)
Feixa Llarga s/n, 08907 L’Hospitalet de Llobregat
Barcelona
Catalonia
Spain   
Fax: +34-93-2607681   

  • References

  • 1 Tarnasky PR, Cotton PB, Baillie J. et al. Proximal migration of biliary stents: attempted endoscopic retrieval in forty-one patients. Gastrointest Endosc 1995; 42: 513-520
  • 2 Chaurasia OP, Rauws EA, Fockens P. et al. Endoscopic techniques for retrieval of proximally migrated biliary stents: the Amsterdam experience. Gastrointest Endosc 1999; 50: 780-785
  • 3 Okabe Y, Tsuruta O, Kaji R. et al. Endoscopic retrieval of migrated plastic stent into bile duct or pancreatic pseudocyst. Dig Endosc 2009; 21: 1-7
  • 4 Ogura T, Okuda A, Miyano A. et al. Migrated endoclip removal after cholecystectomy under digital single-operator cholangioscopy guidance. Endoscopy 2018; 50: 74-75

Zoom Image
Fig. 1 Fluoroscopy image during an endoscopic retrograde cholangiopancreatography plastic stent exchange session showing that one 8.5-Fr plastic stent has migrated proximally (yellow arrow), while the other two plastic stents remain well positioned (red arrows).
Zoom Image
Fig. 2 The new intraductal cholangioscopy retrieval devices: a the basket (SpyBasket); b the snare (SpySnare). Source: Boston Scientific.
Zoom Image
Fig. 3 Fluoroscopy image showing the intraductal cholangioscope (SpyScope DS; orange arrow) and the new retrieval snare (SpySnare; blue arrow) wrapping around the proximally migrated plastic stent (yellow arrow).