Subscribe to RSS
DOI: 10.1055/a-1806-1990
A novel sphincterotome facilitates endoscopic sphincterotomy in patients with Roux-en-Y gastrectomy
Endoscopic retrograde cholangiopancreatography (ERCP) using a short-type single-balloon enteroscope (short SBE) (SIF-H290; Olympus Medical Systems, Tokyo, Japan), with a working length of 152 cm and a working channel diameter of 3.2 mm, has been reported to be effective in patients with surgically altered anatomy (SAA) [1] [2] [3]. However, endoscopic sphincterotomy (ES) in such patients is challenging, as the appearance of the papilla is inverted and the position is frequently tangential, which makes visualizing the correct incision direction difficult. Despite the effectiveness of reported devices, they do not always achieve the correct incision direction [4] [5].
To overcome this limitation, a new sphincterotome (New CleverCut3V; KD-V410V-0720; Olympus Medical Systems), with a working length of 240 cm, was developed for balloon enteroscopy-assisted ERCP in patients with SAA ([Fig. 1]). The blade of the sphincterotome can be easily adjusted around the 5 o’clock position, which indicates the bile duct direction in patients with SAA. Moreover, the blade can be stretched and loosened to modify the incision direction ([Fig. 2]). We report a case of successful ES using this novel sphincterotome in a patient with SAA.
A 47-year-old woman presented with cholangitis due to suspected debris in the common bile duct (CBD). She had undergone total gastrectomy with Roux-en-Y for gastric cancer. ERCP was performed using a short SBE ([Video 1]). As there was a native papilla ([Fig. 3 a]), ES was needed to clear the debris from the CBD. The new sphincterotome was inserted into the papilla after selective biliary cannulation. The blade was easily adjusted to the 5 o’clock position ([Fig. 3 b]) before ES was safely performed in the correct incision direction ([Fig. 3 c]). Clearing the CBD was completed smoothly using a balloon catheter ([Fig. 3 d]).
Video 1 Successful endoscopic sphincterotomy using a novel sphincterotome in a patient with Roux-en-Y gastrectomy.
Quality:
This novel sphincterotome achieved effective and safe ES in patients with SAA. It may aid in the standardization of ES in patients with SAA.
Endoscopy_UCTN_Code_CCL_1AZ_2AI
Endoscopy E-Videos is an open access online section, reporting on interesting cases and new techniques in gastroenterological endoscopy. All papers include a high quality video and all contributions are freely accessible online. Processing charges apply (currently EUR 375), discounts and wavers acc. to HINARI are available.
This section has its own submission
website at
https://mc.manuscriptcentral.com/e-videos
#
Competing interests
The authors declare that they have no conflict of interest.
Acknowledgment
We would like to thank Editage (www.editage.com) for English language editing.
-
References
- 1 Yane K, Katanuma A, Maguchi H. et al. Short-type single balloon enteroscope-assisted ERCP in postsurgical altered anatomy: potential factors affecting procedural failure. Endoscopy 2017; 49: 69-74
- 2 Tanisaka Y, Ryozawa S, Mizuide M. et al. Status of single-balloon enteroscopy-assisted endoscopic retrograde cholangiopancreatography in patients with surgically altered anatomy: systematic review and meta-analysis on biliary interventions. Dig Endosc 2021; 33: 1034-1044
- 3 Tanisaka Y, Ryozawa S, Itoi T. et al. Efficacy and factors affecting procedure results of short-type single-balloon enteroscopy-assisted ERCP for altered anatomy: a multicenter cohort in Japan. Gastrointest Endosc 2022; 95: 310-318
- 4 Takenaka M, Yoshikawa T, Okamoto A. et al. Novel sphincterotomy device that orientates blade along the axis of the bile duct in patients with Roux-en-Y anastomosis. Endoscopy 2019; 51: E132-E134
- 5 Zhu F, Guan Y, Wang J. Efficacy and safety of the rotatable sphincterotome during ERCP in patients with prior Billroth II gastrectomy (with videos). Surg Endosc 2021; 35: 4849-4856
Corresponding author
Publication History
Article published online:
08 April 2022
© 2022. The Author(s). This is an open access article published by Thieme under the terms of the Creative Commons Attribution-NonDerivative-NonCommercial License, permitting copying and reproduction so long as the original work is given appropriate credit. Contents may not be used for commercial purposes, or adapted, remixed, transformed or built upon. (https://creativecommons.org/licenses/by-nc-nd/4.0/)
Georg Thieme Verlag KG
Rüdigerstraße 14, 70469 Stuttgart, Germany
-
References
- 1 Yane K, Katanuma A, Maguchi H. et al. Short-type single balloon enteroscope-assisted ERCP in postsurgical altered anatomy: potential factors affecting procedural failure. Endoscopy 2017; 49: 69-74
- 2 Tanisaka Y, Ryozawa S, Mizuide M. et al. Status of single-balloon enteroscopy-assisted endoscopic retrograde cholangiopancreatography in patients with surgically altered anatomy: systematic review and meta-analysis on biliary interventions. Dig Endosc 2021; 33: 1034-1044
- 3 Tanisaka Y, Ryozawa S, Itoi T. et al. Efficacy and factors affecting procedure results of short-type single-balloon enteroscopy-assisted ERCP for altered anatomy: a multicenter cohort in Japan. Gastrointest Endosc 2022; 95: 310-318
- 4 Takenaka M, Yoshikawa T, Okamoto A. et al. Novel sphincterotomy device that orientates blade along the axis of the bile duct in patients with Roux-en-Y anastomosis. Endoscopy 2019; 51: E132-E134
- 5 Zhu F, Guan Y, Wang J. Efficacy and safety of the rotatable sphincterotome during ERCP in patients with prior Billroth II gastrectomy (with videos). Surg Endosc 2021; 35: 4849-4856