Key words urolithiasis - dose-reduction - CT - ureteral stent - tin-filter
Introduction
In Germany, well over 100 000 patients are treated every year as full inpatients due
to renal colic associated with nephrolithiasis [1 ]. The underlying nephrolith is usually confirmed by imaging and evaluated for further
therapeutic options. Knowledge of the exact calculus (stone, urolith, concrement)
location and size is an important basis for therapy planning and, if necessary, modification.
In advance of performing interventional therapies such as extracorporeal shock wave
lithotripsy (ESWL), ureterorenoscopy (URS), or percutaneous nephrolithotomy (PCNL),
reliable up-to-date information (often repetitive) is needed for both estimating the
likelihood of spontaneous or assisted stone passage and procedure selection [2 ]. Follow-up controls document a progressing stone passage or the success of fragmentation.
Native computed tomography (CT) is the most important diagnostic tool in urolithiasis
in addition to ultrasonography of the urogenital tract, which is usually performed
initially. In addition to localization of the stone, CT can clarify and document its
morphology, possible fragmentation and complications such as urinary retention, fornix
rupture or a therapy-related hematoma both initially and in possible follow-up examinations.
The advantages of CT compared to sonography are the independence of the examiner,
overall lower susceptibility to artifacts, complete imaging of the urinary tract,
as well as the option of determining the stone composition based on its density. Ureteral
stents directly adjacent to the calculus can thereby affect the measurability of the
size and density of the urolith [3 ]. Ureteral stents are regularly indicated to relieve urinary retention, making this
constellation a common challenge for the diagnostician [2 ]. A major limitation of the method is the radiation dose associated with CT, especially
when it must be applied repeatedly, for example, due to recurrent episodes of ureteral
colic [4 ].
The option of spectral filtering has been introduced into routine diagnostics with
the current generation of computer tomographs. This is associated with the possibility
of a significant dose reduction with good diagnostic image quality, as has been demonstrated
in numerous studies in examinations with a variety of indications [5 ]
[6 ]
[7 ]
[8 ] including urolithiasis [9 ]
[10 ]
[11 ]. However, effects of dose-reducing spectral filtering on image quality with a ureteral
stent directly adjacent to a calculus were not explicitly investigated. In this regard,
there is the particular problem of the narrow spatial position of the calculus to
the ureteral stent, which is itself very radiopaque, leading to difficult delineation
of a concrement next to an ureteral stent due to the blurring of the individual image
elements according to the point-spread function (PSF) inherent in any non-ideal imaging
system. Compared to the conventional protocol, using spectral filtering changes the
PSF with the possible result of poorer delineation of the concrement.
The aim of this study was to evaluate the effect on urolith delineability by using
a low-dose protocol with a tin filter versus a conventional examination protocol without
a tin filter, characterizability, and dose in patients with an implanted ureteral
stent who presented for stone location checks.
Materials and Methods
Patients
In this prospective, randomized, 2-arm comparative study, 65 patients (48 men, 17
women, 18–90 years, mean age 55.0 ± 15.2 years) received 84 native CT scans of the
abdomen. Study participants were randomly assigned to two groups, one in which CT
acquisition was performed with a tin filter and the other without (study design: [Fig. 1 ]).
Fig. 1 Study design.
After informed consent had been given, over a period of 11 months, patients registered
for CT stone scans at the urology clinic with ureteral stents (only initial and second
in-house examinations), of legal age and able to give consent, were included in this
study approved by the responsible ethics committee (ethics vote 389/21). Two study
arms were defined and patients were distributed equally. In the case of a second examination
of the same person, this patient was assigned to the respective other study arm; additional
examinations of the same person were not included in the study. The specific patient
characteristics are listed in [Table 1 ].
Table 1
Patient characteristics, continuous variables are reported as mean ± standard deviation,
dichotomous variables as absolute frequencies with percentages in parentheses. Gender
distribution, BMI, and effective body diameter related to examinations.
Study participants (examinations)
65 (84)
Age (in years)
55.0 ± 15.2
Men/Women
48 (74 %)/17 (26 %)
Body Mass Index (BMI) [kg/m²]
27.3 ± 5.4
– BMI ≥ 35 kg/m²
11 (7/65)
Effective body diameter
30.9 ± 4.2
Examination method
All examinations were performed on a clinical 64-slice whole-body CT scanner (Somatom
Go.Top, Siemens, Erlangen, Germany) with the possibility of adding a tin filter. Participants
in the study were randomized to a conventional low-dose protocol without tin filtering
or a low-dose protocol with additional spectral filtering using tin filters. The following
parameters were held constant in both study arms:
Automatic tube voltage selection (Care kV, Siemens, Erlangen, Germany)
Tube current modulation (Care Dose 4 D, Siemens, Erlangen, Germany) at constant quality
index Qref @ 120 kV of 30 mAs underlying the tube current modulation.
Collimation 0.6 mm
Pitch 0.8
The area of examination was defined according to the clinical question based on the
topogram of the area from the upper margin of the left kidney to the middle of the
pubic symphysis. Image reconstruction was performed with a slice thickness of 3 mm
in the transverse, coronary and sagittal planes. Corresponding to a standard abdominal
core, Br31 was used as the reconstruction core in connection with an iterative reconstruction
(SAFIRE – strength 3, Siemens, Erlangen, Germany).
Quality assessment
Both the general image quality and distinction of the stone from the ureteral stent
were evaluated independently by two experienced radiologists who were blinded to the
underlying examination protocol. The results were quantified using a 5-point Likert
scale (1: insufficient – stone can only be guessed at if the location is known; 2:
poor – stone hardly distinguishable, low diagnostic certainty; 3: moderate – stone
moderately distinguishable after intensive review of the image material; 4: good –
stone clearly distinguishable when reviewing the data set; 5: excellent – stone clearly
visible even on cursory review) ([Fig. 2 ]) [12 ].
Fig. 2 Example of quality assessment of stone delineation A , B ) Level 2: poor – stone barely delineable; C , D ) Level 3: moderate – stone moderately delineable on intensive review of the image
material; E , F ) Level 4: good – stone well delineated on review of the data set; G , H ) Level 5: excellent – stone readily visible even on cursory review.
Due to the expected better delineation of large compared to small stones from the
ureteral stent, visual assessment was differentiated with regard to stone size. The
stone size was estimated by approximating the stone morphology by an ellipsoid with
corresponding semi-axes a1 , a2 , a3 and from this an effective stone diameter was determined, which corresponds to the diameter of an equatorial spherical section
(circle) with the same cross-section. The use of the effective diameter in contrast
to a relation to the cut surface of the calculus, which is also conceivable in this
regard, is based on fundamental investigations into the delimitation of round structures
in the presence of overlying image noise [13 ].
Quantitative criteria of image quality
For quantitative assessment of image quality, measurements of X-ray densities in Hounsfield
units (HU) and their standard deviations were made using placed regions-of-interest
(ROI) within both the urinary stone and the ureteral stent, as well as within five
other regions (liver parenchyma, abdominal aorta, psoas major muscle, subcutaneous
adipose tissue, extracorporeal air space). The standard deviation (SD) of the HU values
of the extracorporeal airspace was used as a measure of image noise. The signal-to-noise
ratio (SNR) and the contrast-to-noise ratio (CNR) were calculated from the respective
mean density values in HU for the individual ROIs using SNR = mean valuemeasurement object /SDair and CNR = (mean valuemeasurement object – mean valuefattissue )/SDmeasurement object . To determine the dose efficiency, a figure of merit (FOM) was calculated for each
ROI with FOM = CNR/effective dose2 .
Evaluation of radiation exposure
The computed tomography scanner reported the following indices of radiation exposure:
volume-based CT dose index (CTDIvol ) based on a standard 32 cm phantom, dose-length product (DLP).
The effective dose was calculated as the product of DLP and an abominopelvic conversion
factor of 0.015 mSv/mGy cm [14 ]. Size-specific dose estimates (SSDE) were determined as a product of the CTDIvol with conversion factors dependent on anthropometric measurements of the respective
patient (anteroposterior and transverse body diameter with the calculated effective
) [15 ].
Statistics
The data were evaluated using the R 3.6.1 software package (R Foundation for Statistical
Computing) in conjunction with R Studio 1.3 based on a significance level of 0.05.
Representation is in the format mean ± standard deviation. Tests for normal distribution
were performed using the Shapiro-Wilk test and visually by analyzing histograms.
Statistical testing for non-inferiority (NI) of the studies with tin filter was performed
by defining an NI margin and comparing the confidence interval with tin filter with
the limit for rejecting the NI hypothesis (mean image quality without tin filter –
NI margin) [16 ]
[17 ]
[18 ]. A conservative NI margin of 0.5 quality levels was chosen for the 5-point Likert
scale. Confidence intervals were calculated using the bootstrapping method [19 ] when the subjective quality ratings were not normally distributed; in this case
BCa confidence intervals were used.
The necessary sample size was calculated according to the procedure given in [16 ], assuming a statistical power of 90 % and a one-sided error of the first kind of
2.5 %. Testing for statistical significance was performed depending on the normal
distribution of the data using a t-test or Mann-Whitney-Wilcoxon test. Correlation
tests were performed using Pearson’s correlation coefficient. Cohen’s kappa was used
to quantify interrater agreement.
Results
Subjective quality assessment
Regarding general image quality, the mean rating by both raters was almost identical
across all stone sizes, 4.0 ± 0.4 without and 4.2 ± 0.4 with tin filter (κ = 0.53).
Using an NI margin of 0.5 rating levels, the NI hypothesis resulted in a rejection
limit of 3.5. This led to a confirmation of the NI hypothesis taking into consideration
that this was not undercut by the lower limit of the 95 % confidence interval of the
assessments with tin filter of 4.1.
The average quality assessment of the stone delineation (image examples: [Fig. 3 ]) across all stone sizes was 3.9 ± 0.8 without a tin filter and 4.3 ± 0.7 with a
tin filter (statically significant, p < 0.05).
Fig. 3 Image example of the same patient with right-sided ureteral stent and proximal calculus
adjacent to the stent from the medial side, examination at two different times with
no intermediate change in findings (temporal difference 14 days, no intermediate therapeutic
measures, enlarged image section in the lower right of each image). A , B coronary in the soft tissue window, C , D transverse in the bone window. A , C without tin filter; B , D with tin filter.
The limit size of perfect stone delineation (Λ) corresponds to the effective stone
diameter below which the subjective quality assessment of stone delineation (averaged
over both examiners) falls below the maximum value of 5. In the data set presented,
this corresponded to an effective stone size of 3.8 mm. Above the cutoff size, there
is equivalence of delineability across both study protocols.
Subjective delineation of uroliths from stents showed a high correlation with effective
stone diameters below the cutoff size (Pearson’s correlation coefficient = 0.69).
Due to this high correlation, the range of effective stone sizes below Λ was partitioned
into two areas to allow statistical analysis in groups of similar stone sizes each.
The subdivision was based on the effective stone sizes into size groups A (effective
stone diameter [1–2.4 mm]) and B (effective stone diameter [2.4–3.8 mm]), respectively,
including the lower limit and excluding the upper limit).
Using an NI margin of 0.5 evaluation levels, the NI hypothesis rejection limit in
Group A resulted in 2.73 and in Group B in 3.65, each based on the average stone separability
score in the protocol without tin filter. The 95 % confidence intervals of stone separability
with tin filter in group A were [3.39–4.12] and in group B [4.09–4.47]. The respective
discard limit was not undercut by the lower limit of the confidence interval with
tin filter in any of the groups, consistent with the assumption of NI of the study
protocol with tin filter. The interrater agreement κ of stone delineability ratings
across all stone sizes was 0.64, corresponding to substantial agreement ([Fig. 4 ]).
Fig. 4 NI testing of the delimitability of stone and stent. 95 % confidence intervals of
the individual size groups with tin filter, relative to the mean without tin filter
(centered); the rejection limit of the NI hypothesis is not reached in any group.
Objective quality assessment
Regarding the measurements of SNR and CNR of the stones, a significantly higher CNR
and a not significantly different SNR were shown using the tin filter ([Table 2 ]).
Table 2
Quantitative image quality parameters.
SNR –Sn
SNR +Sn
CNR –Sn
CNR +Sn
FOM –Sn
FOM +Sn
Stone
68.9 ± 29.4
79.2 ± 33.1
10.6 ± 6.8[* ]
17.0 ± 9.3[* ]
111.8 ± 143.0[* ]
366.5 ± 457.9[* ]
Stent
245.7 ± 78.6
257.9 ± 61.1
22.3 ± 20.4
24.6 ± 19.4
639.1 ± 1253.1[* ]
841.3 ± 1477.9[* ]
Liver
4.1 ± 1.4[* ]
4.8 ± 1.4[* ]
7.2 ± 1.3
7.3 ± 1.3
42.0 ± 30.9[* ]
54.9 ± 36.2[* ]
Aorta
3.7 ± 1.0
4.0 ± 0.8
6.9 ± 1.1[* ]
6.4 ± 1.1[* ]
37.3 ± 25.0
41.4 ± 22.4
Psoas
3.9 ± 1.0[* ]
4.6 ± 0.9[* ]
7.1 ± 1.3
7.0 ± 1.2
39.7 ± 24.7
50.0 ± 30.3
Legend: SNR = Signal-Noise Ratio, CNR = Contrast-Noise Ratio, FOM = Figure of Merit,
–Sn = Protocol without tin filter, +Sn = Protocol mit tin filter.
* = Significant difference across both protocols (p < 0.05).
The FOM resulting in combination with the effective radiation exposure of the corresponding
examinations showed significantly higher average values when using the tin filter
(p < 0.05) ([Table 2 ]).
Radiation exposure
Using spectral filtering by tin filter, there was significantly decreased radiation
exposure of patients, both in terms of effective dose with 1.2 ± 0.4 mSv (without
tin filter 1.5 ± 0.4 mSv, p < 0.05) and in terms of SSDE with 2.33 ± 0.38 mGy (without
tin filter 3.09 ± 0.47 mGy, p < 0.05). In a subgroup analysis related to body mass
index (BMI), the dose reduction did not reach the significance level in the group
with BMI ≥ 35. [Table 3 ] provides an overview of the parameters of the radiation dose in relation to the
use of the tin filter and as a function of BMI.
Table 3
Radiation dose related to the use of a tin filter and as a function of BMI.
–Sn total
–Sn, –BMI
–Sn, +BMI
+Sn total
+Sn, –BMI
+Sn, +BMI
Effective dose [mSv]
1.5 ± 0.4[* ]
1.4 ± 0.4[* ]
2.0 ± 0.5
1.2 ± 0.4[* ]
1.1 ± 0.3[* ]
1.7 ± 0.7
Effective dose/BMI
0.053 ± 0.011[* ]
0.053 ± 0.012[* ]
0.053 ± 0.011
0.043 ± 0.011[* ]
0.042 ± 0.009[* ]
0.046 ± 0.022
SSDE
3.09 ± 0.47[* ]
3.01 ± 0.43[* ]
3.56 ± 0.45
2.33 ± 0.38[* ]
2.26 ± 0.29[* ]
2.89 ± 0.54
SSDE/BMI
0.11 ± 0.02[* ]
0.12 ± 0.02[* ]
0.093 ± 0.012[* ]
0.09 ± 0.02[* ]
0.088 ± 0.014[* ]
0.078 ± 0018[* ]
Legend: SNR = Signal-Noise Ratio, CNR = Contrast-Noise Ratio, FOM = Figure of Merit,
–Sn = Protocol without tin filter, +Sn = Protocol mit tin filter, BMI = Body Mass
Index, +BMI: BMI ≥ 35, –BMI: BMI < 35.
* = Significant difference across both protocols (p < 0.05).
Discussion
The use of spectral pre-filtering of the X-ray beam (originally used in CT as part
of the optimization of dual-source CT [21 ]), is a common method of reducing radiation dose in X-ray diagnostics. The applicability
of the method has been demonstrated in numerous previous studies, both with respect
to achievable radiation reduction and regarding consistently high image quality. Applications
were made in CT of the trunk including topograms [5 ]
[6 ]
[7 ]
[8 ] as well as in the context of urolithiasis [9 ]
[10 ]
[11 ]. Effects on image quality during stone position checks in urolithiasis and existing
therapeutically implanted ureteral stent, on the other hand, have not yet been studied
in a dedicated manner, but are of particular interest, because in this case, due to
the given PSF of the imaging system, artifacts may locally occur due to the ureteral
stent, and the effect of the tin filter on that constellation is not known.
Analogous to the published results, there were no significant differences across both
study arms with regard to subjective assessment of overall image quality in this study.
Also, with regard to the assessment of the subjective separability of the urinary
stone from the stent, which is relevant in the context of this study, according to
the diagnostic accuracy in the given clinical question, the NI of the protocol with
tin filter was proven in all groups of stone sizes ([1 .. 2.4), [2.4 .. Λ), [Λ .. ∞)).
The average rating of stone delineation from the stent was thereby (with the exception
of the uroliths > Λ) always greater when using the tin filter. Correspondingly, the
objective-quantitative quality parameter CNR was also significantly higher when the
tin filter was used. Thus, when a conservatively chosen NI margin was used, the NI
hypothesis when using a tin filter was clearly demonstrated in terms of both image
quality and delineation of urinary tract stones. The data also suggest the potential
for stone delineation improvement beyond this, although the sample size does not provide
adequate statistical power for this statement.
At the same time, a significant reduction in radiation exposure was demonstrated using
the tin filter. The mean reductions in effective dose and SSDE were approximately
20 % and 25 %, respectively. This effect was also seen to a lesser extent in the subgroup
of particularly dose-exposed patients with a BMI ≥ 35 [m²/kg], but without reaching
the statistical significance level. Compared with some similar published studies,
this corresponds to a similar dose reduction [9 ]
[10 ], although further dose reduction by using higher pitch values seems possible [11 ]. However, the present study did not focus on reducing the radiation dose by applying
special optimizations that might have resulted in maximum dose reduction. The focus
here was rather an investigation of how strong the effect is under conditions that
are commonly used in practice (examination protocol supplied in principle by the device
manufacturer including automatic tube voltage selection and tube current modulation
[22 ]).
This study has some limitations. A larger sample would have been desirable to optimize
statistical power and highlight possible differences between various stone types.
However, instead of considering all patients with nephroliths and implanted stents,
this study focused on nephroliths with direct contact between stone and stent to address
this issue, which is more clinically relevant. A higher number of cases would be required
to demonstrate the improved stone delineation suggested by our data when examining
with tin filters. However, the aim of our study was to test the non-inferiority of
the method employing tin filters. For further conclusions, studies (preferably multicenter)
with higher case numbers and prospective design will be required.
The present study does not provide any information regarding the delineation of stones
with effective diameters below 1 mm, since these did not occur in the sample. However,
this appears to be a rather theoretical problem, since spontaneous stone passage of
these stones can be excpected, and stent placement is not indicated in such cases.
Conclusions
Low-dose CT protocols demonstrate no loss of image quality or limited delineation
of uroliths from ureteral stents when using spectral filtering with tin filters compared
with examinations without tin filters. In addition, there is about a 20 percent further
dose reduction.
Clinical Relevance of the Study
In patients with urinary stones and the associated high probability of follow-up examinations,
reduction of radiation exposure while maintaining image quality is of particularly
high clinical relevance.
The use of spectral filtering by means of tin filters is a method for dose reduction
that has been well researched in general issues and has found its way into routine
clinical diagnostics, but still has evidence gaps in detailed issues.
Stone position controls with an implanted ureteral stent are potentially repetitive
and present a particular challenge due to artifact-related difficulty in delineating
uroliths.