Abbreviations
AE:
adverse event
CH-EUS:
contrast harmonic endoscopic ultrasound
ESGE:
European Society of Gastrointestinal Endoscopy
EUS:
endoscopic ultrasound
EUS-TA:
EUS-guided tissue acquisition
FCM:
fluorescence confocal laser microscopy
GI:
gastrointestinal
GRADE:
Grading of Recommendations Assessment, Development and Evaluation
IPMN:
intraductal papillary mucinous neoplasia
LBC:
liquid-based cytology
MIAB:
mucosal incision-assisted biopsy
MOSE:
macroscopic on-site evaluation
nCLE:
needle-based confocal laser endomicroscopy
NET:
neuroendocrine tumor
PCL:
pancreatic cystic lesion
PICO:
population, intervention, comparator, and outcome
RCT:
randomized controlled trial
ROSE:
rapid on-site evaluation
TTNB:
through-the-needle biopsy
VOSE:
visual on-site evaluation
This European Society of Gastrointestinal Endoscopy (ESGE) Technical and Technology
Review addresses the technical aspects of endoscopic ultrasound (EUS)-guided sampling
in gastroenterology, providing updated guidance on the available needles and devices,
and techniques for tissue sampling.
1 Introduction
1.1 Focus and rationale
The focus of this Technical and Technology Review from the European Society of Gastrointestinal
Endoscopy (ESGE) is the technical aspects of endoscopic ultrasound (EUS)-guided sampling
in gastroenterology, including the available types of needle, technical aspects of
tissue sampling, new devices, and specimen handling and processing.
Since the publication of the previous ESGE technical review on EUS-tissue sampling
[11 ], several randomized controlled trials (RCTs) have been published covering many of
the technical aspects and approaches to EUS-guided tissue sampling of gastrointestinal
(GI) solid lesions and pancreatic cysts. Furthermore, new devices have been introduced
and tested in clinical practice.
While the indications, results, and clinical impact of EUS-guided sampling are addressed
in a separate Guideline from the ESGE [22 ]
[33 ] and in the recent ESGE core curriculum, the target audience for this technical and
technology review is endoscopists who perform EUS-guided sampling.
1.2 Methodology and development process
ESGE commissioned this technical and technology review and appointed two co-leaders
(A.F. and M.A.), who invited the listed authors (the panel) to participate in the
development of the project. The key questions were prepared by the two leaders in
cooperation with the chair of the ESGE research committee (L.F.) and then approved
by the other members. The relevant clinical questions were developed according to
the PICO format, which outlined the specific patient population (P), intervention
(I), comparator (C), and outcome (O) for each question. The two leaders then assigned
the key questions to task force subgroups.
The members of the task forces, under the supervision of the task force leaders, performed
a systematic literature search to prepare evidence-based statements to address their
assigned key questions. The literature search was performed on the main scientific
databases up until June 2024, focusing mainly on RCTs and meta-analyses of RCTs. Retrospective
or nonrandomized studies were included only if they addressed topics not covered in
the RCTs. The proposed statements were discussed during an online meeting in June
2024. The Grading of Recommendations Assessment, Development and Evaluation (GRADE)
system was adopted to define the quality of evidence and strength of recommendation
[44 ].
The certainty of the evidence was defined as one of four grades: high, moderate, low,
or very low (Table 1s , see online-only Supplementary material). The assessment of the strength of recommendation
reflected the extent to which the panel was confident that the desirable effects of
an intervention outweighed its undesirable effects, or vice versa, across the range
of patients for whom the recommendation is intended. Recommendations fell into two
categories: strong or conditional.
The final draft was also reviewed by the ESGE Governing Board, and after agreement
on a
final version, the manuscript was submitted to the journal Endoscopy for publication. All authors agreed on the final revised version.
The studies included in the current Technical and Technology review used a variety
of outcomes classified in five main categories: (i) diagnostic accuracy; (ii) sample
adequacy and quality; (iii) safety; (iv) technical performance of the needle/device;
and (v) costs. All of these outcomes were systematically analyzed and are discussed
in this document, with the recommendations listed in [Table 1Table 1 ]. Of note, EUS-guided liver biopsy was not covered in the current document as it
is the subject of another ESGE Technical and Technology review.
Table 1
Table 1 List of recommendations.
Recommendation
Strength of recommendation Quality of evidence
ERCP, endoscopic retrograde cholangiopancreatography; EUS, endoscopic ultrasound;
EUS-TA, EUS-guided tissue acquisition; FNA, fine-needle aspiration; FNB, fine-needle
biopsy; PCL, pancreatic cystic lesion; ROSE, rapid on-site evaluation.
1 ESGE suggests that B-mode and contrast harmonic EUS
perform equally for EUS-guided tissue acquisition in patients with solid pancreatic
lesions
Conditional recommendation Low quality evidence
2 ESGE recommends end-cutting FNB needles over
reverse-bevel FNB or FNA needles for tissue sampling of solid pancreatic lesions;
FNA may still have a role when ROSE is available
Strong recommendation High quality of evidence
3 ESGE suggests using end-cutting FNB needles rather
than reverse-bevel FNB or FNA needles for the histopathologic diagnosis of
autoimmune pancreatitis
Conditional recommendation Very low quality of evidence
4 ESGE recommends EUS-FNB or mucosal incision-assisted
biopsy (MIAB) equally for tissue sampling of subepithelial lesions ≥20 mm in size.
MIAB could represent the first choice for smaller lesions (<20 mm) if proper
expertise is available
Strong recommendation Moderate quality of evidence
5 ESGE suggests end-cutting FNB needles rather than
reverse-bevel FNB or FNA needles for tissue sampling of lymph nodes
Conditional recommendation Very low quality of evidence
6 ESGE recommends use of the fanning technique because
it improves the diagnostic accuracy of EUS-guided tissue acquisition of solid
pancreatic lesions
Strong recommendation Moderate quality of evidence
7 ESGE recommends equally the wet-suction or slow-pull
techniques for EUS-FNB of pancreatic masses because they provide high rates of
adequate samples and tissue integrity, and reduce specimen bloodiness compared with
dry suction
Strong recommendation Moderate quality of evidence
8 ESGE recommends performing two passes with
end-cutting FNB needles and three passes with reverse-bevel FNB needles. At least
four passes are needed with FNA needles when ROSE is not available
Strong recommendation Moderate quality of evidence
9 ESGE suggests performing EUS-TA before ERCP in
jaundiced patients with pancreatic head lesions, especially when self-expandable
metal stents are being used
Conditional recommendation Low quality of evidence
10 ESGE recommends the use of ROSE when EUS-FNA is
performed. ROSE is not needed for EUS-FNB, particularly with end-cutting needles,
of
pancreatic solid lesions and submucosal lesions
Strong recommendation Moderate quality of evidence
11 ESGE recommends the use of macroscopic on-site
evaluation (MOSE) for EUS-FNB of pancreatic masses because this approach leads to
high diagnostic accuracy, with the need for fewer needle passes
Strong recommendation Moderate quality of evidence
12 ESGE suggests the use of glucose level over
carcinoembryonic antigen (CEA) in the cystic fluid for the diagnosis of pancreatic
mucinous cysts
Conditional recommendation Low quality of evidence
13 ESGE suggests EUS-guided through-the-needle biopsy
(TTNB) where the histologic diagnosis may change the clinical management of the
patient. TTNB should be avoided in PCLs with communication with the pancreatic duct
owing to the risk of serious adverse events
Conditional recommendation Very low quality of evidence
14 ESGE suggests the use of needle-based confocal laser
endomicroscopy (nCLE) to discriminate between mucinous and non-mucinous pancreatic
cysts in centers with adequate expertise
Conditional recommendation Low quality of evidence
15 ESGE does not recommend the use of antibiotic
prophylaxis before EUS-guided tissue sampling of solid masses and EUS-FNA of
PCLs
Strong recommendation Moderate quality of evidence
The gold standard for diagnostic accuracy was the histologic diagnosis on the surgical
specimen for resected patients and the clinical follow-up (for at least 6 months/1
year) for nonsurgical patients. Sample adequacy and the quality of the specimen were
inconsistently defined across most of the included RCTs; however, a pooled analysis
was performed when a homogeneous definition was available. Adverse events (AEs) were
reported in the RCTs, although none of them was primarily designed and adequately
powered to evaluate safety issues. The PICOs and search strategies are shown in Tables 2s and 3s , and the table of evidence is shown in Table 4s .
2 Advanced EUS imaging to guide tissue acquisition
2 Advanced EUS imaging to guide tissue acquisition
ESGE suggests that B-mode and contrast harmonic EUS perform equally for EUS-guided
tissue acquisition in patients with solid pancreatic lesions.
Conditional recommendation, low quality of evidence.
Contrast-enhanced EUS is a well-established method that combines the advantage of
high resolution EUS of internal organs with the administration of ultrasound contrast
agents.
Pancreatic adenocarcinoma shows a hypovascular heterogeneous pattern and less intense
enhancement than the surrounding parenchyma on contrast harmonic EUS (CH-EUS), whereas
neuroendocrine tumors (NETs) are typically homogeneously hyperenhancing [55 ]. CH-EUS may be used not only to characterize the mass, but also to suggest where
to target the needle during EUS-guided tissue acquisition (EUS-TA). Indeed, CH-EUS,
according to RCTs, clearly outperforms B-mode EUS in identifying the target area in
solid lesions by avoiding necrotic areas, which appear avascular, and vessels [66 ], and in delineating the lesion [77 ] ([Video 1Video 1 ]).
Contrast harmonic-endoscopic ultrasound can assist in identifying the target area
for tissue acquisition in solid lesions by avoiding necrotic areas, which appear avascular,
and vessels.Video 1Video 1
In spite of the above-mentioned advantages, CH-EUS-guided tissue acquisition (CH-EUS-TA)
did not prove superior over B-mode EUS-TA in the routine setting for solid pancreatic
lesions, as outlined in several RCTs, considered at low risk of bias, and meta-analyses
[88 ]
[99 ]
[1010 ]
[1111 ]
[1212 ]
[1313 ] (Table 5s ). The panel performed a comparative meta-analysis of diagnostic accuracy between
the two approaches and did not find any significant difference (relative risk [RR]
0.99, 95%CI 0.95–1.03) (Figs. 1s and 2s ). The quality of evidence was downgraded owing to imprecision (95%CIs crossing unity)
and indirectness (owing to different contrast agents and needles being used). Therefore,
a conditional recommendation with a low quality of evidence was provided that supports
the equal effectiveness of the two approaches (Table 4s ).
CH-EUS-TA could be indicated in poorly visible lesions on B-mode EUS, such as indeterminate
masses in chronic pancreatitis, isoechoic lesions or suspicion of tumors in the context
of acute pancreatitis. Additionally, CH-EUS allows better visualization and tissue
acquisition of extravascular migratory metastases, which are identified by EUS as
soft-tissue cuffs surrounding the vessel in up to 28% patients [1414 ].
With regard to cystic pancreatic lesions, vascularized mural nodules are considered
to be “high risk stigmata” for malignancy when they are ≥5 mm in size (sensitivity
and specificity for high grade dysplasia/invasive carcinoma of 73%–100% and 73%–85%,
respectively) and a “worrisome feature” when their size is <5 mm [1515 ]. CH-EUS is helpful for assessing vascularity and differentiating mural nodules from
nonenhancing mucin plugs, with a reported sensitivity and specificity of 88% and 79%,
respectively [1616 ].
Elastography determines the stiffness of the lesion and pancreatic adenocarcinoma
tends to show high stiffness owing to the desmoplastic reaction of the tissue. While
elastography can be a useful tool for tissue characterization, there is limited evidence
on its use to target tissue sampling. Indeed, despite the promising results of a preliminary
report [1717 ], elastography-guided fine-needle aspiration (FNA) from the hardest (blue) part of
the pancreatic mass ([Fig. 1Fig. 1 ]), which is considered to have the highest suspicion of malignancy, proved not to
have added utility over conventional EUS-FNA (sensitivity 90% vs. 93%) [1818 ].
Fig. 1
Fig. 1 Endoscopic ultrasound images showing elastography-guided fine-needle biopsy of a pancreatic
lesion; the needle was targeted to the “blue” area of the lesion, which represents
an area with higher stiffness that is more likely to be neoplastic.
No data on the use of these ancillary techniques to target EUS-TA in extrapancreatic
lesions are available.
3 Needle choice for the sampling of solid lesions
3 Needle choice for the sampling of solid lesions
3.1 Available needle types
In the last three decades, the evolution of needles has undergone a remarkable change
and they are still constantly under development, with the design of innovative tips
allowing for larger histologic specimens with preserved tissue architecture, and the
use of tissue for genomic profiling and organoid development, which are very important
in the era of oncologic personalized medicine. As a result, the current era has seen
a decisive shift from FNA to fine-needle biopsy (FNB) needles, with the former’s role
now mainly restricted to interventions for pancreaticobiliary access and pancreatic
cysts [1919 ].
EUS-FNA needles have limitations that are well known: their diagnostic performance
is dependent on a cytopathologist to render a rapid on-site evaluation (ROSE); no
histologic core tissue is possible (this is especially relevant to differentiate pancreatic
cancer from autoimmune pancreatitis or chronic pancreatitis); and insufficient tissue
is obtained for risk stratification to tailor anticancer therapy.
Reverse-bevel EUS-FNB needles were developed in 2011 (EchoTip ProCore; Cook Medical,
Indiana, USA). They consisted of an opening at the side of the needle, with a hollow
reverse-bevel architecture, which allows cutting of the tissue during backward retraction
of the needle [2020 ]. Multiple studies, including RCTs and meta-analyses, have been published comparing
reverse-bevel FNB needles with standard EUS-FNA needles, with the former not having
been able to establish their superiority with respect to diagnostic accuracy over
the latter [2121 ]
[2222 ]. These meta-analyses showed that both needles had similar diagnostic accuracy and
sample adequacy, but ProCore FNB needles required fewer passes to establish the diagnosis
(mean difference −1.2 and −0.32, respectively) [2121 ]
[2222 ].
Subsequently, newer “third generation” EUS needles were developed, known as “end-cutting”
needles. These needle types have shown excellent diagnostic accuracy (>90%), even
without ROSE, and are not associated with an increased incidence of AEs [2323 ].
Currently, four needle-tip geometries are available.
3.1.1 Franseen type
The Franseen type, which has three symmetrically distributed needle points and cutting
edges (Acquire, Boston Scientific, Marlborough, USA; EchoTip AcuCore, Cook Medical,
Indiana, USA; SonoTip TopGain, Medi-Globe, Germany).
The Acquire is made of cobalt–chromium and is available in 19G, 22G, and 25G sizes.
The new version, the Acquire S, has a taper-point stylet to ease needle penetration.
The EchoTip AcuCore EUS-FNB needle was introduced in 2024. It is made of cobalt–chromium
combined with a spring-coiled sheath to enhance its flexibility; it is available only
in the 22G size. The SonoTip TopGain needle is made of stainless steel, with a nitinol
stylet, and has a laser engraving over the needle length for better visibility; it
is available in 19G, 22G, and 25G sizes.
3.1.2 Fork-tip type
The fork-tip type has two protruding asymmetrical sharp points and six distal cutting
edges (SharkCore, Medtronic, Minneapolis, USA) [2424 ]. SharkCore needles, available in 19G, 22G, and 25G, are made of stainless steel
and contain a nitinol stylet. They use a Beacon delivery system (Medtronic, Minneapolis,
USA), allowing for needle removal from the sheath, maintaining the sheath in the echoendoscope.
3.1.3 Three-prong asymmetric tip
The three-prong asymmetric tip has one tip longer than the other two (Trident, Microtech,
Nanjing, China). The Trident needle is made of cobalt–chromium, has laser etched markings
for enhanced needle echogenicity, and it is available in 19G, 22G, and 25G sizes.
3.1.4 Menghini tip
The Menghini tip is characterized by a tapered continuous cutting edge (ProCore, Cook
Medical, Indiana, USA; EZ Shot 3 Plus, Olympus, Tokyo, Japan; Sonopsy CY, Hakko Medical,
Nagano, Japan). The 20G ProCore has a forward-facing bevel, antegrade core trap technology,
and ReCoil stylet. The automatic recoiling pattern of the latter helps the technicians
to easily manage the stylet, minimizing contamination [2525 ]. The EZ Shot 3 Plus is made of nitinol with a side hole and a multilayer coil metal
sheath, and is available in 19G, 22G, and 25G sizes. The Sonopsy CY is a 21G needle
with a sharpened stylet that is directly connected to the plunger of the aspiration
syringe and retracted while activating the aspiration.
3.2 EUS-FNA vs. EUS-FNB
ESGE recommends end-cutting FNB needles over reverse-bevel FNB or FNA needles for
tissue sampling of solid pancreatic lesions; FNA may still have a role when ROSE is
available.
Strong recommendation, high quality of evidence.
ESGE suggests using end-cutting FNB needles rather than reverse-bevel FNB or FNA
needles for the histopathologic diagnosis of autoimmune pancreatitis.
Conditional recommendation, very low quality of evidence.
Based on previous meta-analyses that did not show a significant difference in terms
of diagnostic accuracy between FNA and FNB ProCore needles [2121 ]
[2222 ], the latest ESGE Guidelines on tissue sampling recommended FNB and FNA equally in
patients with pancreatic solid lesions with only a weak recommendation for end-cutting
FNB needles when the aim is to obtain core tissue [22 ].
Data across previous RCTs were incorporated in two network meta-analyses, performed
by
Han et al. and Gkolfakis et al. [2626 ]
[2727 ]. Han et al. evaluated data from 29 RCTs and showed that 22G fork-tip FNB needles
had
the best performance in terms of pooled diagnostic accuracy, followed by the 22G Olympus
needle and the 22G Franseen needle (Boston Scientific) [2626 ]. Similarly, Gkolfakis et al. showed that Franseen and fork-tip FNB needles were
the
best ranked needles, especially in the 22G size [2727 ]. Additionally, 25G Franseen or fork-tip needles did not outperform 22G reverse-bevel
needles; of note, FNA needles seemed to still be competitive when ROSE was available,
providing similar results to FNB in this specific setting [2727 ].
The panel updated the literature search of the above mentioned meta-analyses, and
the main characteristics of the 19 RCTs [2828 ]
[2929 ]
[3030 ]
[3131 ]
[3232 ]
[3333 ]
[3434 ]
[3535 ]
[3636 ]
[3737 ]
[3838 ]
[3939 ]
[4040 ]
[4141 ]
[4242 ]
[4343 ]
[4444 ]
[4545 ]
[4646 ] that compared end-cutting FNB needles with each other or with FNA/reverse-bevel
FNB needles in patients with pancreatic solid lesions are listed in the [Table 2Table 2 ]. The panel confirmed the results of the previous network meta-analysis [2727 ], with Franseen and fork-tip FNB clearly outperforming FNA and reverse-bevel FNB
(Franseen vs. reverse-bevel, RR 1.2, 95%CI 1.1–1.4; Franseen vs. FNA, RR 1.2, 95%CI
1.0–1.3; fork-tip vs. reverse-bevel, RR 1.2, 95%CI 1.0–1.3; and fork-tip vs. FNA,
RR 1.1, 95%CI 1.0–1.2). Therefore, end-cutting needles, such as Franseen and fork-tip
FNB needles, should be preferred over FNA and reverse-bevel needles, particularly
when ROSE is not available.
Table 2
Table 2 Randomized controlled trials comparing different end-cutting FNB needles with each
other or versus FNA/FNB needles with reverse-bevel design for tissue sampling of pancreatic
masses.
Author, year country
Study design
Number of patients
Age, years; sex, male, n (%)
Needle used
ROSE
Accuracy
Adequacy
FNA, fine needle aspiration; FNB, fine needle biopsy; RCT, randomized controlled trial;
ROSE; Rapid On-Site Evaluation.
Ardengh, 2020 Brazil
Cross over RCT
FNA: 20
65 9 (45)
22G
No
15/20
19/20
FNB: 20
20G forward-facing bevel
18/20
20/20
Ashat, 2020 USA
Parallel RCT
Franseen: 39
66 56 (42)
64 56 (42)
22G
Yes
36/39
38/39
Fork-tip: 36
22G
34/36
35/36
Asokkumar, 2019 Singapore
Cross over RCT
FNA: 20
64 16 (44)
22G
Yes
18/20
20/20
FNB: 20
22G Franseen
18/20
19/20
Bang, 2017 USA
Cross over RCT
FNA: 46
68 28 (61)
22G
Yes
37/46
38/46
FNB: 46
22G Franseen
43/46
45/46
Bang, 2018 USA
Cross over RCT
Franseen: 50
71 28 (56)
22G
Yes
48/50
48/50
Fork-tip: 50
22G
46/50
46/50
Bang, 2021 USA
Parallel RCT
Fork-tip: 31
64 17 (55)
68 20 (61)
22G
No
30/31
31/31
Menghini: 33
22G
25/33
33/33
Franseen: 32
70 14 (44)
71 18 (53)
22G
30/32
32/32
Reverse bevel: 33
22G
22/33
28/33
Chen, 2021, Canada
Parallel RCT
Fork-tip: 115
70 65 (57)
69 58 (48)
22/25G
Yes
106/115
110/115
FNA: 120
22/25G
112/120
112/120
Cho, 2020 Korea
Parallel RCT
Reverse bevel: 43
64 26 (61)
64 22 (49)
25G
No
34/43
32/43
Forward-facing bevel: 45
20G
40/45
41/45
Crino, 2020 Italy
Parallel RCT
Fork-tip: 96
66 64 (67)
64 46 (48)
22/25G
No
91/96
94/96
Reverse bevel: 96
22/25G
80/96
89/96
Igarashi, 2019 Japan
Cross over RCT
Menghini: 30
74 11 (37)
21G
No
27/30
27/30
Reverse bevel: 30
22G
27/30
27/30
Kandel, 2020 USA
Cross over RCT
Fork-tip: 50
68 25 (50)
19/22G
Yes
50/50
50/50
FNA: 50
25G
50/50
50/50
Karsenti, 2020 France
Cross over RCT
Franseen: 60
69 27 (45)
22G
No
52/60
60/60
Forward-facing bevel: 60
20G
40/60
49/60
Mendoza Ladd, 2022 USA
Parallel RCT
Fork-tip: 22
NR NR
22G
No
NR
22/22
Franseen: 33
22G
NR
33/33
Mizukawa, 2020 Japan
Cross over RCT
Menghini: 97
68 63 (65)
21G
Yes
86/97
93/97
FNA: 97
22G
71/97
89/97
Mohamadnejad, 2023 Iran
Parallel RCT
Franseen: 57
60 36 (63)
61 32 (56)
22G
No
54/57
NR
Three-prong tip: 57
22 3G
56/57
NR
Oh, 2021 Korea
Parallel RCT
25G Franseen: 70
66 39 (56)
62 40 (57)
25G
No
68/70
70/70
22G Franseen: 70
22G
70/70
70/70
Oppong, 2021 UK
Cross over RCT
Fork-tip: 108
69 57 (53)
22/25G
No
85/108
97/108
FNA: 108
22/25G
69/108
99/108
Tomoda, 2021 Japan
Cross over RCT
25G Franseen: 88
71 45 (51)
25G
Yes
76/88
79/88
22G Franseen: 88
22G
79/88
82/88
Yousri, 2022 Egypt
Parallel RCT
Franseen: 50
58 28 (56)
58 29 (58)
22G
No
49/50
NR
FNA: 50
22G
45/50
NR
The quality of evidence was rated as high because the literature was based on a large
number of RCTs, without evidence of heterogeneity, indirectness, publication bias,
inconsistency, or imprecision (Table 4s ). Therefore, a strong recommendation with a high quality of evidence supported the
comparison among the different needle types for pancreatic solid lesions.
The histologic diagnosis of autoimmune pancreatitis represents a great challenge for
endoscopists given the need to obtain high quality core tissue for the final diagnosis
[4747 ]. The development of EUS-FNB needles has generated a great deal of interest in the
field; however, although FNB needles are supposed to improve tissue capture, previous
meta-analyses found only a suboptimal diagnostic accuracy, up to 63% with FNB, although
this was significantly superior to that of FNA [4848 ]. Of note, this meta-analysis included mainly studies using reverse-bevel FNB needles
as only a few reports using end-cutting needles were available at that time.
The panel updated the meta-analysis including only FNB needles and found 12 noncomparative
series, with 496 patients [4949 ]
[5050 ]
[5151 ]
[5252 ]
[5353 ]
[5454 ]
[5555 ]
[5656 ]
[5757 ]
[5858 ]
[5959 ]
[6060 ]. The characteristics of the included studies are reported in Table 6s .
The overall diagnostic accuracy rate of EUS-FNB in these patients was 75% (95%CI 66%–83%)
(Fig. 3s ), with superiority of end-cutting FNB over reverse-bevel FNB (80% [95%CI 70%–90%]
vs. 49% [95%CI 21%–67%]; P <0.001). FNB provided International Consensus Diagnostic Criteria (ICDC) level 1
histologic diagnosis in 47% of cases (38%–57%) and level 2 in 23% (16%–30%). Specifically,
FNB provided a rate of definitive histopathology in those patients without a diagnosis
with other criteria as high as 77% (95%CI 63%–91%). Therefore, the panel concluded
that EUS-FNB with end-cutting needles should be preferred over FNA or other FNB needles
in the diagnostic management of autoimmune pancreatitis. The quality of evidence was
downgraded because the meta-analysis was based on nonrandomized observational studies
and because of inconsistency (high heterogeneity in the estimates) and indirectness
(indirect comparison between different needles) (Table 4s ). Therefore, a conditional recommendation with a very low quality of evidence was
provided that supported the use of end-cutting needles in patients with autoimmune
pancreatitis.
ESGE recommends EUS-FNB or mucosal incision-assisted biopsy (MIAB) equally for tissue
sampling of subepithelial lesions ≥20 mm in size. MIAB could represent the first choice
for smaller lesions (<20 mm) if proper expertise is available.
Strong recommendation, moderate quality of evidence.
ESGE suggests end-cutting FNB needles rather than reverse-bevel FNB or FNA needles
for
tissue sampling of lymph nodes.
Conditional recommendation, very low quality of evidence.
In terms of data on EUS-TA for lymph nodes and subepithelial lesions, data are somewhat
scarce. A meta-analysis of 10 studies, of which six were RCTs, comparing FNA versus
FNB for subepithelial lesions showed FNB to be superior in terms of pooled sample
accuracy (95% vs. 81%; P =0.007), histologic core (90% vs. 65%; P <0.001), diagnostic accuracy (odds ratio [OR] 4.1, 95%CI 2.5–6.8) with a lower mean
number of passes (mean difference −0.75) [6161 ].
The same group recently published a network meta-analysis comparing all the available
techniques to sample subepithelial lesions [6262 ]. Based on eight RCTs [6363 ]
[6464 ]
[6565 ]
[6666 ]
[6767 ]
[6868 ]
[6969 ]
[7070 ] (Table 7s ), the authors found that EUS-FNB was significantly superior to EUS-FNA in terms of
sample adequacy (RR 1.2, 95%CI 1.1–1.5), and appeared to be the best technique (surface
under cumulative ranking [SUCRA] score 0.90), followed by mucosal incision-assisted
biopsy (MIAB; SUCRA 0.83), whereas endoscopic bite-on-bite biopsy showed the poorest
performance. Interestingly, when considering lesions <20 mm, MIAB (a group of newer
techniques involving unroofing the subepithelial lesion to expose its surface or submucosal
tunneling allowing direct biopsy sampling of the lesion), but not EUS-FNB, showed
significantly higher accuracy rates compared with EUS-FNA (RR 1.7, 95%CI 1.0–2.9)
and ranked as the best intervention in this specific subset of lesions. As observed
with solid pancreatic lesions, when ROSE was available, no difference between EUS-FNB,
EUS-FNA, and MIAB was observed [6262 ].
Therefore, as already outlined in the previous ESGE guidelines [7171 ], the panel recommend the use of EUS-FNB or MIAB for tissue sampling of subepithelial
lesions in the GI tract. MIAB could represent the first choice in the subset of smaller
lesions (<2 cm); however, owing to the risk of bleeding and complications, and the
more challenging technical characteristics, MIAB should be performed in centers with
high case volume and adequate expertise. The quality of evidence was downgraded because
of high imprecision in the estimates owing to wide 95%CIs crossing unity or failure
to reach the optimal information size. Therefore, a strong recommendation with a moderate
quality of evidence was provided (Table 4s ).
As already observed with solid pancreatic lesions, a meta-analysis clearly showed
that end-cutting FNB needles outperform EUS-FNA in terms of diagnostic accuracy for
tissue sampling of lymph nodes (OR 1.9, 95%CI 1.2–3.0), whereas no difference was
observed in the sensitivity analysis comparing reverse-bevel FNB versus FNA (OR 1.0,
95%CI 0.5–1.5) [7272 ]. Of note, a recent large prospective Italian series of lymph nodes sampled with
22G Franseen or 25G fork-tip FNB needles showed an overall sensitivity, specificity,
and accuracy of 96%, 100%, and 97%, respectively [7373 ].
The panel updated the meta-analysis [7272 ], including 10 studies [7474 ]
[7575 ]
[7676 ]
[7777 ]
[7878 ]
[7979 ]
[8080 ]
[8181 ]
[8282 ]
[8383 ] that directly compared EUS-FNB versus EUS-FNA for tissue sampling of lymph nodes
(Table 8s ). Diagnostic accuracy was similar between FNB and FNA (OR 1.5, 95%CI 0.9–2.5) (Fig. 4s ); however, FNB performed with end-cutting needles clearly outperformed FNA (pooled
accuracy with end-cutting FNB needles, 89%; OR 1.9, 95%CI 1.2–3.0).
Therefore, the panel concluded that EUS-FNB with end-cutting needles could be the
best option in these patients, although further data from RCTs are needed. The quality
of evidence was downgraded because the meta-analysis was based mainly on nonrandomized
observational studies and because of inconsistency (high heterogeneity in the estimates),
risk of bias in the literature, and imprecision in the estimates (Table 4s ). Therefore, a conditional recommendation with a very low quality of evidence was
provided.
4 Sampling techniques
4.1 Fanning
ESGE recommends use of the fanning technique because it improves the diagnostic
accuracy of EUS-guided tissue acquisition of solid pancreatic lesions.
Strong recommendation, moderate quality of evidence.
The fanning technique allows sampling of multiple areas within the same lesion with
a single needle pass by changing the needle's direction with the up/down deflection
wheel (big wheel) or the elevator [8484 ]
[8585 ] ([Video 2Video 2 ]). To increase the efficacy of fanning, it is suggested that the direction of the
puncture be changed after withdrawing the needle at the proximal limit of the tumor.
Despite not being widely investigated, two RCTs have demonstrated that, compared with
the standard technique, the fanning technique results in higher diagnostic accuracy,
a lower number of passes to establish the diagnosis, and a higher percentage of patients
in whom a diagnosis is achieved with just the first pass, yet with no difference in
terms of technical failure or complication rates [8484 ]
[8585 ]. A change in the direction of the needle (i.e. fanning the needle into the lesion)
can also be obtained by torquing the scope, with results similar to those of the “classic”
fanning technique [8686 ]
[8787 ] (Table 9s ). The fanning technique also seems equivalent to sampling using contrast-harmonic
EUS guidance [1313 ], but without requiring the additional cost of contrast.
Demonstration of the fanning technique.Video 2Video 2
Notably, all of the available studies included only pancreatic solid lesions, and
the role of the fanning technique in nonpancreatic lesions has not been investigated.
Intuitively, the fanning technique is also supposed to acquire a sample more representative
of the whole tumor, allowing an advantage even for molecular analyses, despite not
yet being adequately investigated.
The panel then performed a meta-analysis of the four RCTs [8484 ]
[8585 ]
[8686 ]
[8787 ] that compared fanning versus standard FNA/FNB and found a significant superiority
of fanning for accuracy (RR 1.21, 95%CI 1.11–1.31) (Fig. 5s , part A), which was confirmed when excluding the torque technique (RR 1.27, 95%CI
1.16–1.38) (Fig. 5s , part B). The quality of evidence was downgraded owing to indirectness related to
different sampling techniques (suction vs. no suction vs. slow pull) in the included
RCTs (Table 4s ). Therefore, a strong recommendation with a moderate quality of evidence was provided.
4.2 Aspiration techniques
ESGE recommends equally the wet-suction or slow-pull techniques for EUS-FNB of
pancreatic masses because they provide high rates of adequate samples and tissue
integrity, and reduce specimen bloodiness compared with dry suction.
Strong recommendation, moderate quality of evidence.
Four main different aspiration techniques have been described: no-suction, slow-pull,
dry-suction, and wet-suction techniques (including a modified wet-suction technique)
([Fig. 2Fig. 2 ]). The no-suction technique just requires removal of the stylet before lesion puncture,
with no application of negative pressure. The slow-pull technique consists of a slow
retraction of the stylet while performing the to-and-fro movements within the lesion,
which appears to apply low negative pressures to the needle tip. Suction techniques
require removing the stylet and using a syringe with a 5-, 10-, or 20-mL negative
pressure attached to the needle to provide a greater negative pressure than the slow-pull
technique. The wet-suction technique also requires flushing of the needle with saline
before performing the puncture to apply the negative pressure onto a fluid that has
no expandability and, therefore, precisely transmits it to the tip of the needle [8888 ].
Fig. 2
Fig. 2 The different techniques for tissue sampling include: a the slow-pull technique, consisting of a slow retraction of the stylet while performing
to-and-fro movements within the lesion, which appears to apply low negative pressures
to the needle tip; b the wet-suction technique, which requires removal of the stylet and flushing of the
needle with saline before performing the puncture so that, when negative pressure
is applied to a fluid, which has no expandability, this is precisely transmitted to
the tip of the needle.
Despite a wide literature on the topic, there is no striking evidence to suggest the
unequivocal superiority of one suction technique over any another. Outcomes commonly
evaluated in available studies were sample adequacy, cellularity, and tissue integrity/core
procurement. Several meta-analyses showed that the wet-suction technique ranked as
the best technique in terms of diagnostic accuracy [8989 ]
[9090 ]
[9191 ]. Similarly, the wet-suction technique is ranked as the best technique regarding
adequacy, followed equally by the slow-pull and dry-suction techniques, with the no-suction
technique being inferior for this outcome. On the other hand, the no-suction and slow-pull
techniques provided very low blood contamination compared with the wet-suction and
dry-suction techniques [9292 ]
[9393 ]. In addition, wet-suction seems better than the dry-suction technique regarding
blood contamination. Notably, the impact of blood contamination has been inherited
from FNA studies as it could impair cytologic evaluation. In contrast, it should not
impact the assessment of histologic samples or the capability to retrieve high quality
histologic tissue [9494 ]. Furthermore, most studies evaluated only pancreatic solid lesions, with a few lymph
nodes and submucosal tumors included, and subgroup analyses were not performed in
the meta-analyses.
The panel updated the meta-analysis of RCTs comparing different techniques for tissue
sampling with EUS-FNB of pancreatic solid lesions and found 10 RCTs [3333 ]
[4444 ]
[9494 ]
[9595 ]
[9696 ]
[9797 ]
[9898 ]
[9999 ]
[100100 ]
[101101 ] (Table 10s ).
As outlined in a previous network meta-analysis [8989 ], the no-suction technique was significantly inferior to the other techniques in
terms of sample adequacy (RR 0.85, 95%CI 0.08–0.92 vs. slow pull; RR 0.85, 95%CI 0.78–0.92
vs. dry suction; RR 0.83, 95%CI 0.76–0.90 vs. modified wet suction), although it led
to less
blood contamination of samples. The highest tissue integrity was observed with modified
wet
suction (RR 1.36, 95%CI 1.06–1.75 vs. dry suction) and the highest rate of blood
contamination was observed with dry suction (RR 1.44, 95%CI 1.15–1.80 vs. slow pull).
Altogether, this evidence favors using the wet-suction technique, with the slow-pull
technique considered a valuable alternative to be used intuitively for hypervascular
pancreatic lesions or where cytologic evaluation is required. With any suction technique,
it
is recommended to neutralize residual negative pressure in the needle before withdrawing
it
from the target lesion to avoid contamination of the sample by GI fluid.
Data on nonpancreatic lesions are scarce. The trial by Crinò et al. found a higher
rate
of tissue core retrieval with the wet-suction rather than slow-pull technique in the
subgroup of nonpancreatic lesions (67% vs. 48%), although the significance threshold
(adjusted for multiplicity) was not reached, probably because the RCT was underpowered
for
this subgroup analysis [9494 ]. It is likely that the application of suction may have a greater impact on the
quantity of tissue aspirated into the needle in softer tissues, such as nonpancreatic
lesions, rather than stiff pancreatic masses, like adenocarcinoma.
The panel downgraded the quality of evidence to moderate owing to indirectness related
to the different study designs and different FNB needles used in the individual studies
(Table 4s ). Therefore, a strong recommendation with a moderate quality of evidence was provided.
4.3 Number of to-and-fro movements
The number of to-and-fro movements, also called the “number of actuations,” is the
number of back-and-forth movements performed inside a lesion within the same needle
pass. Most studies reported performing 10–15 actuations, but only a few have specifically
investigated this topic and only in pancreatic solid lesions, reporting scanty and
heterogeneous data that are difficult to standardize.
An RCT comparing 40 versus 20 actuations performed with a 22G side-fenestrated needle
showed no benefit in doing more movements [102102 ]. A recent RCT performed with FNA needles demonstrated similar accuracy with 10,
15, and 20 actuations when suction was applied, but with higher blood contamination
with 20; in contrast, 10 actuations were associated with lower accuracy if no suction
was used [103103 ]. The authors concluded that 10 or 15 to-and-fro movements should be performed with
and without suction.
Two crossover RCTs using a 22G Franseen needle compared different numbers of actuations.
The first study demonstrated a slightly, though not significantly, higher accuracy
with 15 actuations compared with five; however, 15 actuations were better than five
regarding the area of the tissue core retrieved [104104 ]. Conversely, the second study demonstrated that three actuations provided similar
results to 12 in terms of diagnostic accuracy and microscopic histologic quantity,
but with less blood contamination [105105 ].
Finally, a large retrospective study raised concerns about an increased risk of AEs
when performing more than 15 actuations during EUS-FNA [106106 ], although, considering its retrospective design, the results of this study should
be interpreted with caution.
Given the paucity of data and the discordant results of the RCTs, which tested different
numbers of actuations, thereby making it impossible to standardize the data, the panel
decided to refrain from providing a statement on this topic. Intuitively, the number
of actuations should also depend on the lesion size. In fact, each to-and-fro movement
should ideally cross the entire diameter of the tumor. Consequently, in the case of
small lesions, the range of movement would be very small, and increasing the number
of actuations could result in a better sample. In contrast, the range of movement
in large masses is extended, so allowing for a reduction in the number of to-and-for
movements.
4.4 Number of passes
ESGE recommends performing two passes with end-cutting FNB needles and three passes
with reverse-bevel FNB needles. At least four passes are needed with FNA needles when
ROSE
is not available.
Strong recommendation, moderate quality of evidence.
Overall, several meta-analyses have demonstrated that, compared with FNA, FNB needles
can reduce the number of passes required to achieve a diagnosis [2121 ]
[107107 ]
[108108 ]
[109109 ]
[110110 ]
[111111 ]; however, most of the studies available in the literature reported the diagnostic
yield of EUS-TA in relation to the cumulative number of passes or reported partial
data on each pass outcome, and only a limited number of studies performed a per-pass
analysis with a pre-established (or established on ROSE) number of passes [3636 ]
[4545 ]
[8484 ]
[9696 ]
[112112 ]
[113113 ]
[114114 ]
[115115 ]
[116116 ]
[117117 ]
[118118 ]
[119119 ]
[120120 ]
[121121 ]
[122122 ]. Notably, most studies employing FNA needles used ROSE during EUS-guided sampling.
With the above-mentioned limitation, based on specifically designed studies, when
ROSE is unavailable, to achieve a diagnostic yield of approximately 90%, four passes
should be performed with an FNA needle and three with a reverse-bevel needle. With
the advent of new end-cutting FNB needles, it seems that the number of required passes
could be further lowered; in fact, when using 22G Franseen [122122 ]
[123123 ] or fork-tip needles [3636 ], two passes increased diagnostic yield compared with one and reached the plateau,
thereby suggesting that two passes might be sufficient. A slight advantage of performing
a third pass was observed using 22G three-prong asymmetric tip [4545 ] and 25G fork-tip needles [3636 ].
These findings were also observed in a recent meta-analysis of 19 RCTs that demonstrated,
using any FNB needle, a pooled diagnostic accuracy of 78% (95%CI 73%–84%), 87% (95%CI
83%–91%), 92% (95%CI 88%–95%), 89% (95%CI 81%–96%), and 96% (95%CI 95%–97%) with one,
two, three, four, and five passes, respectively, with significant benefit for performing
up to three passes. Subgroup analysis of end-cutting needles (Franseen and fork-tip)
demonstrated a pooled accuracy of 80% (95%CI 74%–87%), 87% (95%CI 79%–94%), and 91%
(95%CI 84%–97%) with one, two, and three passes, respectively, with no benefit from
adding the third pass [124124 ] ([Table 3Table 3 ]).
Table 3
Table 3 Comparison of numbers of needle passes for tissue sampling of pancreatic masses.
Outcomes
Number of passes
Number of RCTs
Number of patients
OR (95%CI); I
2
FNB, fine-needle biopsy OR, odds ratio.
Any FNB needle
Diagnostic accuracy
Two vs. one
9
1159
1.9 (1.4–2.7); 50%
Three vs. two
9
1159
1.6 (1.2–2.1); 0%
Three vs. one
8
1045
3.2 (2.0–5.3); 61%
Four vs. three
4
398
1.1 (0.7–1.8); 0%
Tissue adequacy
Two vs. one
6
1072
2.1 (1.6–2.8); 36%
Three vs. two
5
962
1.9 (1.3–2.8); 61%
Three vs. one
5
962
4.1 (2.9–5.9); 50%
End-cutting FNB needles
Diagnostic accuracy
Two vs. one
6
674
1.8 (1.2–2.6); 0%
Three vs. two
5
411
1.4 (0.9–2.2); 0%
Tissue adequacy
Two vs. one
4
434
2.2 (1.7–2.9); 0%
Three vs. two
3
544
3.0 (2.0–4.5); 0%
Most of the reported studies included patients with pancreatic solid lesions; however,
two to three passes with an end-cutting needle seem sufficient even for submucosal
lesions and lymph nodes [7373 ]
[120120 ].
The panel downgraded the quality of evidence to moderate owing to the indirectness
related to different availability of ROSE/macroscopic on-site evaluation (MOSE) across
the RCTs (Table 4s ).
Finally, with the progressive advent of molecular analyses to aid personalized treatments
and considering the need to also retrieve material for this precise aim, additional
passes, apart from the diagnostic ones, could be considered. Although the aforementioned
meta-analysis [124124 ] found higher adequacy with three passes, even using end-cutting FNB needles, probably
owing to the higher quality of tissue samples, promising data show that two passes
are not different from three passes in terms of the quantity of retrieved DNA and
RNA for molecular analyses [125125 ]. Nonetheless, the standard technique for EUS-TA should also be established based
on the center’s and endoscopists' individual experiences, as most of the published
literature represents standards in high volume centers with experienced endosonographers
and pathologists.
4.5 EUS-guided tissue acquisition in jaundiced patients needing biliary stenting
ESGE suggests performing EUS-TA before ERCP in jaundiced patients with pancreatic
head
lesions, particularly when self-expandable metal stents are being used.
Conditional recommendation, low quality of evidence.
For patients with obstructive jaundice due to an unresectable pancreatic head tumor,
given the concurrent need for cyto/histologic confirmation that is mandatory before
chemotherapy, performing both the EUS and ERCP procedures during the same session
may represent a valuable option. ERCP is however more widely available than EUS, and
relieving bile duct obstruction is commonly considered more clinically urgent than
tissue acquisition for establishing a pathologic diagnosis [126126 ]. Therefore, in most instances, biliary stenting frequently precedes EUS-TA; however,
some concerns have been raised regarding the impact of biliary stents on the diagnostic
accuracy of EUS-TA, likely owing to poor visualization of the lesion from acoustic
shadowing, reverberation, and/or surrounding inflammation associated with the stent
[127127 ]
[128128 ]
[129129 ]. Additionally, these studies seem to suggest that plastic or self-expandable metal
stents (SEMSs) could have a dissimilar effect because of differences in their material
and diameter.
Table 11s shows the data from the studies that have assessed the impact of the presence of
a biliary stent on the diagnostic performance of EUS-FNA/FNB of masses in the pancreatic
head [128128 ]
[129129 ]
[130130 ]
[131131 ]
[132132 ]
[133133 ]
[134134 ]. As reported in the meta-analysis by Facciorusso et al. [135135 ], only SEMSs showed an impact on decreasing the diagnostic accuracy of EUS-TA (OR
0.5, 95%CI 0.2–1.0), whereas plastic stents had no impact (OR 0.9, 95%CI 0.5–1.5),
with the pooled accuracy of EUS-FNB after placement of a plastic stent around 85%
[135135 ]. Therefore, the panel concluded with a conditional recommendation, based on retrospective
studies, in favor of performing EUS-TA before ERCP in patients with jaundice due to
pancreatic head masses, particularly if SEMSs are to be used.
5 On site evaluation
5.1 Rapid on-site evaluation (ROSE)
ESGE recommends the use of ROSE when EUS-FNA is performed. ROSE is not needed for
EUS-FNB, particularly with end-cutting needles, of pancreatic solid lesions and submucosal
lesions.
Strong recommendation, moderate quality of evidence.
ROSE is a cytomorphologic diagnostic procedure that allows assessment of the adequacy
of the material obtained during individual biopsy passes within a few minutes, in
or near to the endoscopy suite (on-site). Once the tissue has been acquired, the section
obtained is printed onto a glass side, stained, and observed under a microscope by
the cytopathologist. The sampling is repeated until the cytopathologist deems the
specimen obtained to be adequate [136136 ].
ROSE may be performed by a cytopathologist, a pathology trainee, or a cytotechnologist.
Given the low availability of these professional figures, training endosonographers
to
evaluate the specimen themselves has been proposed as a potential solution. In the
prospective double-blind RCT by Savoy et al., well-trained endosonographers had inferior
abilities compared with a cytopathologist in terms of identifying specimen adequacy
(P =0.004) and making a preliminary estimate of malignancy (P <0.001) in various lesions (lymph nodes, pancreas, and liver
specimens) [137137 ]. In a more recent RCT, which exclusively analyzed pancreatic solid lesions, the
sensitivity and accuracy of ROSE by the endosonographer for malignancy were high (93%
and
88%, respectively), with a shorter procedural duration, as well as fewer needle passes
in
the ROSE versus non-ROSE group [138138 ]. Zhang et al. compared ROSE performed by endoscopists in 97 patients to 97 patients
without ROSE in a RCT [139139 ]. Self-ROSE improved the diagnostic accuracy (P <0.001)
and sensitivity (P <0.001) compared to the non-self-ROSE group,
with an acceptable consistency between endoscopist and pathologist in the cytopathologic
diagnosis (P <0.05) and in the sample adequacy rate (P <0.001) [139139 ]. In the prospective study by Eloubeidi et al. [140140 ], an excellent agreement between ROSE and the final cytology was demonstrated. Of
the
300 lesions initially reported as malignant on ROSE, 294 (98%) remained malignant
on the
final cytology, while only 2% of the lesions initially reported as being benign on
ROSE were
upgraded to malignant or suspicious for malignancy in the final cytology report. Scant
cellularity was the most frequently reported reason for discrepancies, followed by
the need
for consultation with other specialists, the need for ancillary studies, intraobserver
variability, and challenging diagnoses [140140 ].
Given the rapid advance of artificial intelligence (AI) in the medical field, Lin
et al. [141141 ] reported the validation of an AI-based model (ROSE-AI model) to substitute for manual
ROSE during EUS-FNA; it demonstrated strong performance in detecting cancer cells,
achieving an accuracy rate of 83% in the internal validation dataset and a comparable
result of 89% in the external validation dataset [141141 ].
The cost-effectiveness of ROSE is still under debate. In the cost-analysis study by
Wani et al. [142142 ], EUS-FNB of pancreatic and nonpancreatic solid lesions appeared a more cost-effective
strategy than EUS-FNA with ROSE. On the other hand, in a study aimed at assessing
the cost-efficacy of ROSE performed during EUS-FNA of several different GI lesions,
EUS-FNA with ROSE was associated with a significant cost-effectiveness benefit, resulting
in savings of almost $252 per case of EUS-FNA when ROSE was applied. The economic
savings attributed to ROSE were owing to a lower rate of inadequate specimens and
fewer EUS-FNA sessions needed to achieve the final pathologic diagnosis [143143 ]. In a retrospective study by the same group, the cost-effectiveness analysis was
not different between EUS-FNB and EUS-FNA with ROSE in GI lesions [144144 ]. Therefore, the authors concluded that the choice of strategy depends on availability
and local expertise.
As demonstrated in previous meta-analyses [145145 ]
[146146 ], the inclusion of ROSE in EUS-FNA improves sample adequacy for pathologic evaluation
and enhances the diagnostic performance of the procedure. In particular, the aforementioned
network meta-analysis demonstrated that EUS-FNA with ROSE leads to similar diagnostic
outcomes as compared with EUS-FNB, even with highly performing end-cutting needles,
both in pancreatic solid lesions [2727 ] and subepithelial lesions [6262 ]. On the other hand, conflicting results are reported in the literature concerning
EUS-FNA of lymph nodes [147147 ].
ROSE may not be routinely and widely available; it is especially present in high volume
tertiary care centers. In a global survey published in 2016 [148148 ], ROSE was used by 98% of American endosonographers and only by 48% of European and
55% of Asian respondents.
As previously reported, the development of end-cutting FNB needles led to very high
adequacy and accuracy rates, thereby limiting the need for ROSE, as outlined in several
studies and in a large multicenter RCT enrolling 771 patients with solid pancreatic
masses, which found comparable diagnostic accuracies (96% with ROSE and 97% without
ROSE; P =0.40), but with a significantly shorter mean sampling procedural time (18 [SD 9]
vs. 12 [SD 6] minutes; P <0.001) [149149 ]. Similar results were also reported in the recent meta-analysis of eight studies
(2147 patients), where the difference in terms of sample adequacy of solid pancreatic
lesions was not statistically significant between EUS-FNB alone versus EUS-FNB + ROSE
[150150 ]. The diagnostic accuracy was superior in the EUS-FNB + ROSE group (P =0.03); however, when considering only the new FNB needles, this difference did not
reach statistical significance (P =0.56). The same results were observed in the setting of subepithelial lesions, where
the above network meta-analysis showed that ROSE did not improve the diagnostic yield
of EUS-FNB [6262 ].
Nowadays, with the advancement of digital imaging technology, there is the possibility
of an expert cytopathologist evaluating FNA specimens remotely. In 2014 Khurana et
al. demonstrated that telecytopathology on-site evaluation of EUS-FNA of pancreatic
lesions reduced the nondiagnostic rate, especially for solid lesions [151151 ]. A more recent study by Machado et al. demonstrated the feasibility of a low-cost
instant messenger smartphone application for evaluating specimens of EUS-FNA of pancreatic
solid lesions in terms of sample adequacy, with the possibility of an initial diagnosis
of malignancy in 61% of samples [152152 ]. Telecytopathology offered by high volume centers may in the future serve as an
effective substitute for on-site evaluation, overcoming the issue of a shortage of
cytopathologists.
The studies comparing the diagnostic outcomes of EUS-FNB with ROSE versus EUS-FNB
without ROSE are reported in Table 12s
[7979 ]
[149149 ]
[153153 ]
[154154 ]
[155155 ]
[156156 ]
[157157 ]
[158158 ].
As reported in the study by Facciorusso et al. [150150 ], the meta-analysis of the two RCTs [7979 ]
[149149 ] that compared the two approaches showed no difference (OR 1.1, 95%CI 0.3–3.9), with
moderate evidence of heterogeneity (I
2 =49%).
The high level of indirectness, related to the different study designs (one of the
RCTs was a subanalysis of an RCT) and different FNB needles used, led to downgrading
of the quality of evidence to moderate (Table 4s ). Therefore, a strong recommendation with a moderate quality of evidence was provided.
5.2 Macroscopic on-site evaluation (MOSE) and visual on-site evaluation (VOSE)
ESGE recommends the use of MOSE for EUS-FNB of pancreatic masses because this approach
leads to high diagnostic accuracy, with the need for fewer needle passes.
Strong recommendation, moderate quality of evidence.
Other techniques have been implemented for assessing the quality of tissue samples
during EUS procedures. MOSE, introduced by Iwashita et al. [159159 ], involves evaluating the presence of a macroscopically visible core (a whitish or
yellowish tissue fragment) in the sample dispensed onto a glass slide; if it is ≥4
mm, it
predicts the presence of a histologic core ([Fig. 3Fig. 3 ]). MOSE has been evaluated with both FNA [160160 ] and FNB needles [123123 ], demonstrating excellent diagnostic parameters for both techniques. Moreover,
according to a recent noninferiority RCT [123123 ], MOSE can significantly reduce the number of needle passes in patients with
pancreatic and nonpancreatic solid lesions, with an accuracy that is not inferior
to that of
EUS-FNB with three needle passes. Sundaram et al. showed that, when end-cutting EUS-FNB
needles were used for the evaluation of solid pancreatic lesions, no difference in
sensitivity, specificity, or diagnostic accuracy was detected between ROSE and MOSE
[161161 ].
Fig. 3
Fig. 3 Macroscopic on-site evaluation (MOSE) involves evaluating for the presence of a macroscopically
visible core (a whitish or yellowish tissue fragment) in the sample after it has been
dispensed onto a glass slide.
Visual on-site evaluation (VOSE) might represent an easy alternative to MOSE, consisting
of assessing the EUS sample fully expelled into a graduated formalin vial, which is
shaken to better visualize the presence of a macroscopically visible core. In the
prospective single-center study by Stigliano et al. [162162 ], the endoscopist evaluated the presence of a macroscopic visible core, the color
(white or red/mixed) and the blood contamination; the authors found that a VOSE “red-mixed
specimen” was associated with a higher probability of histologic adequacy.
VOSE and MOSE are quick and simple procedures performed by the endosonographer in
the endoscopy room, without the need for other professional figures, making them highly
cost-effective; however, in comparison with ROSE, they do not provide quality information
or a preliminary diagnosis.
A recent meta-analysis by Mohan et al. showed a pooled accuracy of FNA and/or FNB
specimens in yielding a pathologic diagnosis by MOSE of 91% (95%CI 89%–93%), pooled
sensitivity of 92% (95%CI 89%–94%), and pooled specificity of 99% (95%Ci 97%–100%),
which was also confirmed in the subgroup analysis that was restricted to end-cutting
FNB needles [163163 ]. Based on the single RCT by Mangiavillano et al. [123123 ], no difference was observed in terms of diagnostic accuracy between EUS-FNB with
MOSE versus without MOSE (RR 1.02, 95%CI 0.95–1.10). The panel therefore decided to
downgrade the quality of evidence owing to imprecision (wide 95%CIs crossing unity
and failure to reach the optimal information size) (Table 4s ). Therefore, a strong recommendation with a moderate quality of evidence was provided.
Given the very limited published data on the use of VOSE, the panel decided to refrain
from providing a statement on VOSE.
5.3 Digital confocal laser microscopy
Ex vivo fluorescence confocal laser microscopy (FCM) represents a new digital tool
allowing for real-time microscopic assessment of fresh unfixed biologic specimens.
Its main advantage, especially in the field of EUS-FNB, is the fact that it produces
digital histologic images without any need for conventional slide preparation. Its
application could lead to shorter procedures, fewer needle passes, and a lower risk
of AEs, while at the same time obviating the need for ROSE.
Currently, the VivaScope 2500M-G4 microscope (MAVIG GmbH, Munich, Germany) has been
evaluated in a limited number of studies. Briefly, the specimen obtained after the
first needle pass is placed directly in a dedicated scaffold and, after a minimum
amount of preparation, is put between two dedicated microscope slides and placed in
the imaging device that, using a dedicated algorithm, converts the acquired image
information into colors that resemble hematoxylin and eosin (H&E) staining [164164 ]. In a prospective study that enrolled 81 patients undergoing EUS-FNB of solid pancreatic
lesions, FCM accurately predicted the adequacy of the tissue sample in 96% of the
samples, while there was significant agreement between FCM evaluation and the final
histologic assessment (Cohen’s k coefficient 0.95, 95%CI 0.89–1.01), with 100% sensitivity and 67% specificity [164164 ].
In a more recent multicenter study, the validity of digital FCM was evaluated by comparing
the interobserver agreement between FCM and standard histologic analysis [165165 ]. The digital images from FCM and digitalized paired whole-slide images from 25 consecutive
patients with pancreatic EUS-FNB were sent to 10 international expert pathologists,
with a high interobserver agreement demonstrated (Cohen’s k 0.79, 95%CI 0.65–0.92) [165165 ]. Based on these studies, FCM could be of value in improving the real-time assessment
of pancreatic solid lesions, although the high costs limit its applicability currently.
6 Tissue acquisition for pancreatic cystic lesions
6 Tissue acquisition for pancreatic cystic lesions
6.1 Cyst fluid cytology and molecular analysis
ESGE suggests the use of glucose level over carcinoembryonic antigen (CEA) in the
cystic fluid for the diagnosis of pancreatic mucinous cysts.
Conditional recommendation, low quality of evidence.
Accurate diagnosis and risk stratification for the development of malignancy in pancreatic
cystic lesions (PCLs) are challenging. An unacceptable number of patients with PCLs
with worrisome features that, according to guidelines, should undergo surgical resection
are eventually diagnosed with benign disease on final histologic examination of the
resection specimen [166166 ]. In order to avoid unnecessary surgery, effective methods to better diagnose and
subclassify PCLs are highly warranted.
EUS characteristics alone are usually insufficient to make a diagnosis, but there
are some occasions where they are relevant (e.g. classic features of a pseudocyst
in the correct clinical circumstances). EUS alone is also accurate in identifying
serous cystic neoplasms, especially when honeycombing is present in a lesion with
clear noncommunication with the pancreatic duct. For most other lesions, there will
be a need for sampling of the most suspicious aspect of the cyst, such as a mass or
nodule, and this should be conducted as for a solid lesion. For those lesions without
a clear suspicious target but where sampling has been deemed necessary, direct sampling
is performed from fluid within the cyst. The use of adjunctive devices such as confocal
microscopy and through-the-needle biopsy (TTNB) forceps are discussed separately.
The technique involves using a standard FNA needle, as opposed to a biopsy needle,
and most operators prefer either a 22G or 19G FNA needle. If a mucinous cyst is suspected,
then a 19G needle is preferable, as the fluid can be very thick and requires a large
needle channel for successful aspiration; however, 19G needles are more challenging
in areas such as the uncinate process. If possible, the whole cyst should be aspirated
to dryness in a single pass but, in multiloculated lesions, targeting of the largest
cystic component and full aspiration is acceptable.
The needle should be positioned in the center of the cyst and the tip observed during
aspiration to prevent contact with the cyst wall as it collapses. This should ensure
complete aspiration and reduces any abrasion of the cyst wall by the tip of the needle.
EUS-guided cyst FNA is low risk, but AEs such as bleeding, infection, cyst rupture,
and pancreatitis have all been described. The overall risk was reported to be between
2% and 6% in a previous meta-analysis [167167 ].
Cyst fluid assessment is comprised of its descriptive appearance and measurement of
various components of the fluid, with the most common being cytology, and levels of
carcinoembryonic antigen (CEA), amylase, and glucose. Few studies have examined the
value of the cyst fluid appearance, although most endosonographers give a subjective
description of the fluid viscosity at aspiration. The most objective measure of this
is the “string sign,” which involves placing a droplet of the aspirate between the
thumb and index finger of a gloved hand and then separating the fingers ([Fig. 4Fig. 4 ]). A positive string sign is where the “string” stretches to 1 cm and persists for
>1 second [168168 ]. This has been assessed prospectively and is specific, but not very sensitive [169169 ].
Fig. 4
Fig. 4 The string sign is shown during endoscopic ultrasound-guided fine-needle aspiration
of a pancreatic cystic lesion. This test involves placing a droplet of the aspirate
between the thumb and index finger of a gloved hand and then separating the fingers.
A positive string sign is where the “string” stretches to 1 cm and persists for >1
second.
Several tumor markers have been investigated to try to determine which cystic lesions
are more likely to be mucinous, as well as those at highest risk of malignant transformation.
Initial studies showed the most accurate results using CEA, and a cutoff of 192 ng/mL
gave an area under the curve (AUC) of 0.79 [170170 ]. More recent studies have shown that lower levels of CEA in the aspirate may be
more accurate [171171 ]
[172172 ]. One of the biggest issues is the lack of use of a standardized assay for measurement
in different institutions, leading to the variance in the accuracy for the different
reported cutoffs. It is worth noting that high CEA levels have also been found in
inflammatory cysts [172172 ].
Measurement of the cyst fluid amylase is a potentially useful adjunct. A raised amylase
level suggests communication of the cyst with the pancreatic duct, with levels <250
U/L largely excluding pseudocysts [173173 ]. A highly elevated amylase level is associated with pseudocysts, but has also been
seen in mucinous lesions [172172 ].
More recently there has been interest in inexpensive, widely available markers that
might reflect the level of metabolic activity within a cystic lesion. In initial studies,
a lower glucose level in the cyst fluid aspirate was reported to have a high sensitivity
for diagnosing mucinous cysts, whether sent to the laboratory or performed using point-of-care
glucometers [174174 ]. A recent meta-analysis showed that a cyst fluid glucose level <50 mg/dL had a sensitivity
of 91% and specificity of 91% when sent to the laboratory and 90% and 84%, respectively,
when using point-of-care glucometers [175175 ]. There are two studies that have examined the utility of combining glucose and CEA
levels, with one showing improved accuracy [176176 ] and the other showing no additional benefit of CEA over glucose alone [177177 ]. A more recent study examined the lactate level within cyst aspirates and showed
significantly lower levels in mucinous cysts. Glucose had superior operating characteristics
to both CEA and lactate in this study [178178 ].
Examination of the cellular content in a pancreatic cyst aspirate is perhaps one of
the oldest methods used to try to determine the nature of the lesion. The sensitivity
of cyst fluid cytology is generally poor owing to the low cellular yield in the fluid
aspirate. One meta-analysis showed a pooled sensitivity of 63% (95%CI 56%–70%) and
a specificity of 88% (95%CI 83%–93%) [179179 ]. This was also reflected in the high positive likelihood ratio (4.5) and low negative
likelihood ratio (0.5). This means that cytology is only reliable if a positive diagnosis
is made as false positives, especially for advanced changes, are uncommon, but negative
cytology cannot currently exclude a mucinous lesion or even malignancy.
Within a pancreatic cyst fluid aspirate, there are other sources of information that
are currently being investigated. These include assessment for the presence of genetic
mutations, such as KRAS and GNAS , which appear to have a high specificity for mucinous cysts [180180 ]. Other studies have used next-generation sequencing to create a genetic profile
for the cyst, combining the presence of different gene mutations [180180 ]. Differing expression of mucin proteins in proteomic studies has suggested higher
risk patterns, but this needs confirmation. Similarly, exploration of proteomics,
metabolomics, and the microbiome may improve the accuracy of fluid sampling [181181 ]. These areas are not currently widely available clinically and remain within the
realm of exploratory research. Table 13s shows the outcomes of seven meta-analyses published in the last 5 years [175175 ]
[177177 ]
[182182 ]
[183183 ]
[184184 ]
[185185 ]
[186186 ].
Glucose level showed similar specificity and significantly higher sensitivity as compared
with CEA; molecular analyses led to very good results, but the high costs and the
limited availability represent major concerns with their use. Therefore, a weak recommendation,
with a low quality of evidence, in favor of the use of glucose levels was provided,
because it was based on nonrandomized studies. The use of CEA might have a complimentary
role, although whether their combined use would increase significantly the diagnostic
accuracy is still uncertain.
6.2 Through-the-needle biopsy (TTNB)
ESGE suggests EUS-TTNB where the histologic diagnosis may change the clinical
management of the patient. TTNB should be avoided in PCLs with communication with
the
pancreatic duct owing to the risk of serious AEs.
Conditional recommendation, very low quality of evidence.
TTNB is performed by a new device that enables retrieval of histologic tissue, comprising
epithelium and stroma, from the cystic wall or septa [187187 ]. The through-the-needle microforceps biopsy technique has shown promising results
for the diagnosis and subclassification of pancreatic cysts, enabling immunohistochemical,
as well as high quality molecular diagnostic examination, far more accurately when
compared with EUS-FNA [188188 ]
[189189 ]
[190190 ]
[191191 ]. Moreover, a good interobserver agreement among pathologists has been demonstrated
[192192 ].
There are currently two commercially available devices: the Moray Micro forceps (US
Endoscopy, Mentor, Ohio, USA) and the Micro Bite forceps (MTW Endoskopie Manufaktur,
Germany), with the former most extensively tested in the published studies.
The Moray Micro forceps has a length of 230 cm, a sheath diameter of 0.8 mm, and a
jaw opening width of 4.3 mm [193193 ]. The single-use biopsy forceps can be passed through a standard 19G FNA needle.
The jaws of the forceps are flat with multiple sharp teeth. The micro forceps can
either be preloaded in the needle or introduced after the cyst has been punctured
by a 19G needle and the stylet removed. When the transducer is positioned in the second
part of the duodenum with the endoscope tip flexed, it may be difficult to advance
the micro biopsy forceps through the needle owing to resistance from within a bent
needle. If resistance is experienced, slight, repetitive, gentle opening and closing
of the forceps during introduction through the needle reduces the friction and often
helps overcome this problem. In addition, aspirating some of the fluid before TTNB
ensures an amount of “nonblood-contaminated” fluid, as some minor bleeding is often
observed after TTNB. Moreover, the fluid aspiration reduces the cystic wall tension
and may facilitate grabbing tissue into the jaws of the microforceps [194194 ]. Following cyst puncture and forceps introduction, the jaws of the forceps are easily
visualized under real-time EUS guidance. The jaws are opened, and the forceps advanced
anteriorly to the needle tip to touch the cyst wall. The forceps should then be closed,
while maintaining a slight pressure, and then retracted, while showing “tenting” of
the wall lining. Observed “tenting” of the wall lining on EUS is usually a good predictor
of an adequate sample; if this tenting effect is not experienced, the procedure should
be repeated until the desired effect is seen or alternatively until clear resistance
is felt when the biopsy forceps is retracted ([Video 3Video 3 ]).
Demonstration of the “tenting effect” during through-the-needle biopsy.Video 3Video
3
The tissue obtained should be visually inspected and the procedure repeated until
at least two optimal biopsies have been acquired. It has been shown that extraction
of more than two biopsies does not increase the sensitivity, but raises the risk of
AEs [195195 ].
The endoscopist should be aware of the risks associated with the procedure and try
to minimize the intracystic needle time after obtaining an adequate number of samples.
Following the retraction of the instruments, the lesion and the surrounding area should
be re-examined to exclude any AEs, such as intracystic hemorrhage, although the latter
is self-limited in almost all cases.
After retraction of the micro biopsy forceps, the small specimen is placed onto a
small paper disc and finally introduced into a cassette [196196 ].
According to the recently published evidence-based international Kyoto guidelines
[1515 ], there are no established indications for EUS-TTNB, and it seems too early to reach
any definitive conclusions. It was demonstrated that TTNBs can substantially change
the clinical management strategy in patients with PCLs in 12%–19% of cases, mostly
by providing the diagnosis of a serous cyst adenoma in an oligocystic or unilocular
cyst, and so leading to discontinuation of follow-up [188188 ]. As a general principle, EUS-TTNB should be performed only in patients who are fit
for surgery, with a clear indication for biopsy, and where the biopsy diagnosis may
change the clinical management of the patient [1515 ].
Because EUS-TTNB offers the chance to evaluate a microhistologic sample of the cyst
wall, the key advantage of the procedure is gained when the histotype of the cyst
is unknown, as typically happens in the case of unilocular/oligocystic PCLs. This
cyst “morphology” may encompass a wide spectrum of PCLs, ranging from the more common
intraductal papillary mucinous neoplasia (IPMN), serous cystadenoma, or mucinous cystadenoma
to more rare conditions, including NETs [197197 ]
[198198 ] or schwannomas [199199 ]
[200200 ].
The accurate definition of a PCL is crucial for the management of the patient as,
for example, large mucinous cysts, unlike serous cystadenomas, are frequently indicated
for resection. In contrast, whether EUS-TTNB will play a role in the evaluation of
a PCL’s dysplasia grade remains doubtful. In fact, as a general principle, biopsy
is not the way to define the grade of dysplasia because of its variability within
the same lesion and its patchy distribution along the cystic wall, thereby producing
a risk of a false-negative result owing to inadequate sampling [201201 ].
Several meta-analyses have reported very high technical success rates of 90% and 100%
for TTNB, and high sensitivity and specificity of around 80%, which are superior to
standard FNA cytology [202202 ]
[203203 ]
[204204 ].
Limiting factors include the lack of epithelium in non-mucinous cysts, but also denudation
of the cyst wall, which may yield false-negative results. Moreover, as previously
commented, the evaluation of dysplasia warrants cautious interpretation as the biopsies
represent only a portion of the cyst wall and the results may be falsely negative.
Therefore, any high risk features, such as a thickened cyst wall, nodules, or solid
parts, should be targeted if possible.
AEs associated with the procedure range from 2% to 23%, and the rate seems higher
compared with the standard EUS-FNA procedure [205205 ]
[206206 ]. The most common AE reported is intracystic hemorrhage, which is defined as hyperechoic
changes in the cystic lumen following the tissue acquisition; however, all published
cases have been mild and, in most cases, without any clinical implications [207207 ]. Infections are rarely seen after EUS-TTNB of pancreatic cysts and a multicenter,
retrospective, propensity-score matched study demonstrated that antibiotic prophylaxis
does not influence the infection rate, which appears to be very low [208208 ]. EUS-TTNB can also cause acute pancreatitis, probably owing to destruction of the
cyst wall, edema of the adjacent normal pancreatic tissue architecture, microclot
passage into the main pancreatic duct, and leakage of the pancreatic juice. In a recently
published meta-analysis, the pooled AE rate for pancreatitis was shown to be around
3% (95%CI 1.8%–4.5%) [207207 ]. Most pancreatitis events had a mild clinical course; however, the reported rate
of serious AEs was around 1.1% (95%CI 0.4%–2.7%), mainly due to severe pancreatitis
[202202 ]. Serious concerns about the use of TTNB were raised by a prospective Danish study
that reported an AE rate of 10%, including a case of death due to post-TTNB pancreatitis
[188188 ].
A large multicenter publication that included 506 patients found age (OR 1.3, 95%CI
1.1–2.1), number of TTNB passes (from OR 2.2 [95%CI 1.3–4.3] to OR 3.2 [95%CI 2.0–6.3]
with
increasing number of passes), complete aspiration of the cyst (OR 0.6, 95%CI 0.3–1.0),
and
diagnosis of IPMN (OR 4.2, 95%CI 2.3–7.7) to be independent predictors of AEs [209209 ]. No preventive measures have been clearly demonstrated to reduce the risk of
pancreatitis, although Kovacevic et al. observed a decreasing trend in the incidence
of
pancreatitis when using prophylactic rectal indomethacin/diclofenac [188188 ]. Future studies are needed to evaluate the role of rectal nonsteroidal
anti-inflammatory drugs (NSAIDs) or aggressive hydration in minimizing the risk of
pancreatitis after TTNB.
Table 14s reports the available meta-analyses assessing TTNB performance in patients with PCLs
[202202 ]
[203203 ]
[204204 ]
[205205 ]
[206206 ]
[207207 ]
[210210 ]. In all of these studies, TTNB clearly outperformed cytology in terms of diagnostic
accuracy (ORs ranging from 3.4 [95%CI 1.3–9.0] to 9.4 [5.7–15.4]). Moreover, a recent
network meta-analysis of retrospective studies found TTNB to be the best technique
to differentiate mucinous cysts and to identify the malignant nature of a PCL (superiority
index 13, 95%CI 5–17) [211211 ].
The panel provided a conditional recommendation with a very low quality of evidence
favoring the use of TTNB in PCL patients, because the evidence was based only on retrospective
studies and owing to the high risk of imprecision (failure to reach the optimal information
size) (Table 4s ).
6.3 Confocal laser endomicroscopy
ESGE suggests the use of needle-based confocal laser endomicroscopy (nCLE) to
discriminate between mucinous and non-mucinous pancreatic cysts in centers with adequate
expertise.
Conditional recommendation, low quality of evidence.
Over the last decade, needle-based confocal laser endomicroscopy (nCLE) has emerged
as a promising technology to overcome the limitations of EUS-FNA in the characterization
of PCLs. nCLE enables real-time imaging of tissue and vascular structures.
Cellvizio I.V.E. (Mauna Kea Company, Paris, France) is a confocal laser imaging device
adapted for imaging through a bundle of optical fibers, assembled with other miniaturized
elements, called a Confocal Miniprobe, to allow real-time imaging of the internal
microstructure of tissues ([Fig. 5Fig. 5 ]).
Fig. 5
Fig. 5 Appearances of some of the different patterns observed during needle confocal laser
endomicroscopy of pancreatic cystic lesions. a A serous cystadenoma (SCA), with the superficial vascular network seen as multiple
interconnected black vessels (arrows) and big vessels (stars), in which moving red
cells (small black points) are seen moving in the fluorescein-filled vessel, which
appears white, and neuroendocrine tumor (NET) with irregular clusters of dark cells
(blue ovals). b A pancreatic intraductal papillary mucinous neoplasia (IPMN) contains small papillary
projections (stars) and one large papilla (blue oval) with a vascular core (red star),
and the visible epithelial border (arrows). c A mucinous cystadenoma (MCN) with the epithelial border (blue star) and its limits
marked (arrows), and a pancreatic pseudocyst, with inflammatory cells that can be
seen moving in the liquid (blue ovals).
Prior to each procedure, the AQ-Flex probe is inserted into the 19G FNA needle in
lieu of the stylet and advanced until the probe is positioned with the tip aligned
1 mm beyond the needle tip and then secured with a special locking device. After EUS-guided
puncture, the FNA needle should be carefully positioned within the cyst. A soft contact
between the nCLE probe and the epithelium of the cyst is necessary to obtain relevant
images. After image acquisition, fluid aspiration for cytology, tumor markers, and/or
DNA, and/or TTNB can be performed.
Based on expert opinion, it can be recommended to minimize catheter manipulation and
to stop the examination as soon as specific nCLE criteria are observed, and systematically
when the potentially examinable area inside the cyst has been checked. The recommended
imaging time is <10 minutes (ideally <6 minutes) to decrease the risk of pancreatitis
[212212 ].
Some concerns were previously raised concerning the reproducibility of nCLE results
[213213 ]; however, three other studies conducted with blinded reviewers reported an almost
perfect interobserver agreement (>0.60) for mucinous lesions, serous cystadenomas,
and pseudocysts [214214 ]
[215215 ]
[216216 ]. Intraobserver reliability was addressed by two trials using the INDEX-study population
[215215 ]
[216216 ] and was reported for all nCLE criteria as substantial, kappa ranging from 0.68 to
0.78, for nCLE-naïve blinded reviewers (n=6) and as almost perfect, kappa ranging
from 0.85 to 0.91, among six blinded nCLE experts (experience >30 nCLE cases) [216216 ].
The main AE reported in the studies was post-procedural acute pancreatitis (with only
one severe case reported). The highest rate was reported in the DETECT study (7%),
which combined Spyglass cystoscopy and nCLE imaging in the same procedure [217217 ]. These rates are comparable to the pooled pancreatitis rate of 1.6% (95%CI 0.6%–3.8%)
and global AE pooled rate of 5.5% (95%CI 0.9%–13.6%) associated with EUS-guided FNA
for PCLs performed using a 19G needle [167167 ].
A French health economic evaluation of nCLE for the diagnosis of PCLs showed nCLE
to be cost-effective for the therapeutic management of PCLs for serous cystadenoma
[218218 ]. This analysis showed that the addition of nCLE to EUS-FNA resulted in a reduction
of 23% in the total rate of surgical intervention, which translated to a reduction
in clinical costs of 13% [218218 ].
Table 15s reports the data of five meta-analyses that assessed the diagnostic performance of
nCLE in patients with PCLs [206206 ]
[219219 ]
[220220 ]
[221221 ]
[222222 ]. The pooled sensitivity and specificity ranged between 82% and 90%, and 91% and
99%, respectively, with a reported pooled AE rate ranging from 1% to 5%, mostly due
to pancreatitis. As reported in the meta-analysis by Facciorusso et al. [219219 ], the OR for diagnostic accuracy of nCLE compared with EUS-FNA was 3.9 (95%CI 1.6–9.8).
The aforementioned network meta-analysis [211211 ] concluded that, for centers with relevant expertise and facilities, EUS-guided nCLE
and EUS-guided TTNB are better choices for the diagnosis of PCLs (OR 88 [95%CI 29–214]
and OR 40 [95%CI 11–104], respectively). Moreover, the indirect comparison between
the two procedures reported in the meta-analysis by Kovacevic et al. [206206 ] showed a slightly higher diagnostic yield with nCLE. Therefore, a conditional recommendation
with a low quality of evidence was provided in favor of the use of nCLE, because it
was based on nonrandomized studies.
7 Antibiotic prophylaxis
ESGE does not recommend the use of antibiotic prophylaxis before EUS-guided tissue
sampling of solid masses and EUS-FNA of PCLs.
Strong recommendation, moderate quality of evidence.
Whether antibiotic prophylaxis should be administered prior to EUS-guided tissue sampling
remains controversial. Incidence rates of bacteremia are reported to be low after
EUS, with or without FNA, of cystic or solid lesions along the upper GI tract and
for solid rectal lesions [223223 ].
The American Society of Gastrointestinal Endoscopy (ASGE) published a guideline on
antibiotic prophylaxis for GI endoscopy in 2015 that recommended against prophylaxis
for solid lesions, while prophylaxis for EUS-guided sampling of cystic lesions (both
pancreatic and mediastinal) was still recommended [223223 ]. In 2017, the ESGE published a technical guideline on EUS-guided tissue sampling,
confirming the indication for antibiotic prophylaxis for PCLs only, but with a low
quality of evidence [11 ].
Over the past years, new evidence regarding antibiotic prophylaxis for EUS-guided
tissue sampling of pancreatic cysts has been published. In 2020, Colan-Hernandez et
al. published a randomized, noninferiority, double-blind, placebo-controlled clinical
trial investigating whether EUS-FNA of pancreatic cysts without antibiotic prophylaxis
increased the infection rate as compared with EUS-FNA with antibiotic prophylaxis
[224224 ]. They found that the risk of infections related to the procedure was overall very
low (<1%) and was not increased when antibiotic prophylaxis was not used (OR 0.3,
95%CI 0.01–8.3) [224224 ]. This finding was confirmed by two systematic reviews and meta-analyses, including
six studies (the RCT and five retrospective series), which confirmed the very low
incidence of infectious AEs (<1%), with no difference between the two groups [225225 ]
[226226 ]. Table 16s reports the studies comparing antibiotic prophylaxis versus no prophylaxis before
EUS-FNA of PCLs [106106 ]
[224224 ]
[227227 ]
[228228 ]
[229229 ]
[230230 ].
Based on the available evidence, the panel recommended against the use of antibiotic
prophylaxis in patients with PCLs undergoing EUS-guided sampling with complete aspiration
of the cyst. A strong recommendation with a moderate quality of evidence was provided
because of the risk of imprecision (wide CIs crossing unity and failure to reach the
optimal information size) (Table 4s ).
Very limited evidence is available about TTNB and the indication for prophylactic
antibiotics. A retrospective propensity score-matched study reported one infectious
event (1%) in the antibiotic prophylaxis group versus 2% in the control group [208208 ]. Despite this preliminary evidence, further studies are needed to provide a definitive
statement on the need for antibiotic prophylaxis before TTNB of PCLs.
8 Specimen handling and processing
8 Specimen handling and processing
Techniques used to process EUS-guided samples include conventional smears, cytospins,
liquid-based cytology (LBC), and formalin-fixed paraffin-embedded tissue (specifically,
cell blocks and biopsies). Cytologic methods (conventional smears, cytospins, and
LBC) assess individual cells or clusters of cells, whereas histology (cell blocks
and biopsies) allows assessment of intact fragments of tissue, including the architectural
arrangements of cells and their relationship to surrounding tissues.
The technique of choice varies upon the nature of the lesion (solid or cystic), the
type of needle used (FNA or FNB needle), the need to perform ROSE during the procedure,
and the practical experience and expertise of the involved pathology personnel.
For solid lesions or for mural nodules of cystic lesions, the specimen obtained with
an FNA needle can be partly spread onto glass slides (conventional smears) and partly
collected in a vial containing saline or fixative. Glass slides must be fixed immediately
by spray fixation or immersion in 95% alcohol for Papanicolaou or H&E staining; alternatively,
they can be allowed to dry for Giemsa or Diff quick stain if ROSE is required. If
ROSE is not required, the entire sample can be collected in the vial without conventional
smear preparations. If the sample is collected fresh in saline solution, it should
be rapidly transported to the laboratory and prepared to limit cell degeneration.
Whatever the medium, in the pathology laboratory, macroscopically visible tissue fragments
and blood clots measuring >2 mm must be removed and processed separately in a paraffin
block as a conventional biopsy, which requires formalin fixation. The remaining sample
or its entirety, if no tissue fragments or blood clots are present, can be processed
as LBC and/or cell blocks, the former technique requiring alcohol-based fixation,
the latter 10% buffered formalin.
Alcohol-based fixatives are indicated for LBC (the two most common types being Cytolyt
for ThinPrep and Cytorich Red for SurePath), which markedly attenuate the background
elements, such as blood, mucus, and necrosis, but guarantee perfect cellular details.
The residual LBC specimen can also be processed as a cell block, with subsequent formalin
fixation.
To obtain a cell block, the sample is spun down with a centrifuge and the cellular
material is concentrated into a cell button that is agglutinated and then processed
as a routine histology specimen. There are several methods of agglutinating the cell
pellet for embedding, the choice depending on which technique is available in the
laboratory [231231 ]. Sometimes no agglutination is necessary when large, clotted material is acquired
with an FNA needle.
When using an FNB needle, the sample is generally corpuscolated with macroscopically
visible tissue fragments that can be collected in formalin and processed as a conventional
histologic biopsy, even if there can be differences among the various needle sizes
and types in term of specimen volume [232232 ]. If tissue fragments are not visible or are very small (e.g. ≤2 mm), a cell block
technique can be employed to avoid loss of material.
If ROSE is required, instead of smearing part of the sample onto glass slides, touch-imprint
cytology can be performed [233233 ]: tissue fragments are gently pushed and partially spread between two glass slides
to avoid smashing, in this way slides suitable for cytologic examination are obtained.
Tissue fragments should then be picked up and collected in formalin.
For cystic lesions, even if ROSE is not indicated, the liquid sample can be smeared
onto glass slides (see above), especially if it is viscous/mucinous, or it can be
entirely collected fresh and sent immediately to the pathology laboratory, where it
can be processed undiluted with the cytospin technique. This is a cell concentration
method using centrifugal force to isolate, concentrate, and deposit a monolayer of
cells onto a circular area on a slide. Cytospin preparations maintain the background,
such as necrotic debris and mucin, but samples that have undergone transportation
stress, such as exposure to abnormally high or low temperatures, may yield poor quality
cells with low overall viability. Alternatively, the liquid sample can be collected
in an alcohol-based fixative for LBC or in 10% buffered formalin for cell block preparation.
When dealing with pancreatic cyst fluid, the liquid should be triaged according to
its quantity. Triage of aspirated PCLs requires an integrated multidisciplinary approach
that involves clinical evaluation, radiologic assessment, cyst fluid biochemistry,
and molecular/genetic testing, as well as cytologic evaluation. In general, if it
is ≤0.5 mL, it is sent fresh for biochemical analysis or collected in ethanol for
molecular studies, avoiding cytologic examination. If >0.5 mL, an aliquot of fluid
can be submitted for cytologic examination and processed as described above. Each
ancillary test requires at least 0.3 mL [234234 ]. An EUS-TTNB will be processed as a standard histology specimen.
All of cytospins, conventional smears, and LBC, as well as formalin-fixed paraffin-embedded
tissue (cell blocks and biopsies) allow for further studies (immunohistochemical stains
and molecular analysis), although cell blocks and any biopsies, which have been formalin
fixed, are considered optimal for immunohistochemistry and additionally allow direct
comparison of the same cells between multiple analyses [235235 ]
[236236 ].
Regarding processing techniques for EUS-FNA samples, the superiority of LBC over conventional
smears is controversial; however, as shown in a recent meta-analysis [237237 ], combining LBC with conventional smears showed a significantly higher pooled sensitivity
compared with conventional smears alone and LBC alone for the clinical evaluation
of pancreatic lesions. The combination of LBC and cell blocks has also been shown
to significantly improve diagnostic accuracy, indicating that these specimen processing
techniques complement each other [238238 ].
With regard to solid lesions around the digestive tract, as demonstrated in the study
by Huang et al. [239239 ], the combination of cytologic and histologic specimen processing methods also contributes
additional diagnostic efficacy to EUS-FNA in solid lesions of the digestive tract.
As for PCLs, it was demonstrated that the diagnostic yield of an EUS-FNA sample improved
when the endosonographer processed the specimen by spreading it onto glass slides,
instead of collecting it directly into the liquid medium for LBC [240240 ]; in contrast, with regards to the presence of mucus, cell block assessment appears
superior to direct smears [241241 ].
With regard to genomic analysis, there is limited literature on sample adequacy for
molecular testing and genomic sequencing for pancreatic ductal adenocarcinoma obtained
via EUS-FNA versus EUS-FNB. Comprehensive genomic profiling using next-generation
sequencing has become an approved diagnostic test and has been successfully performed
on FNA and FNB samples, including cell blocks, in several smaller studies [242242 ]
[243243 ]
[244244 ]. Tumor surface area and tumor cellularity are essential parameters in determining
sample adequacy for molecular testing: up to 2000 cells and 30% tumor cellularity
may be necessary to perform comprehensive genomic analysis, but analysis can be successful
at <10% tumor cellularity [243243 ]. Mohamad et al. found that FNB was more likely to result in sufficient tissue sampling
for genomic and molecular testing [245245 ]. They demonstrated that sample adequacy for molecular testing was achieved in 71%
of patients who underwent FNB compared with 32% of patients who underwent FNA (P <0.001).
In the absence of exact guidelines for material processing, each institution should
thoroughly evaluate the optimal preparation method of EUS-FNA/FNB samples, considering
that pathologists must fill the gap between the high technology endosonographer apparatus
and the low/no technology sample preparation tools that they use [246246 ]. Consequently, collaboration between endoscopists and pathologists is critical for
success, with an understanding that cytologic and histologic processing techniques
often complement each other, but tissue pieces >2 mm need to be processed for histology.
Sample processing techniques should be selected according to an individual center’s
experience and expertise, bearing in mind that samples processed histologically are
optimal for immunohistochemistry, are more likely to be adequate for molecular analyses,
and may be necessary to make certain diagnoses requiring intact tissue architecture.
Disclaimer
ESGE Technical and Technology Reviews represent a consensus of best practice based
on the available evidence at the time of preparation. They may not apply in all situations
and should be interpreted case by case in the light of specific clinical situations
and resource availability. Further controlled clinical studies may be needed to clarify
aspects of the statements, particularly in the absence of high quality of evidence,
and revision may be necessary as new data appear.
ESGE Technical and Technology Reviews are intended to be an educational device to
provide information that may assist endoscopists in providing care to patients. They
are not rules and should not be construed as establishing a legal standard of care
or as encouraging, advocating, requiring, or discouraging any treatment.