Introduction
Over the past decade, digital radiography has continued to replace conventional film-screen
radiography of the chest [1 ]. Storage phosphor image receptors and flat panel detectors not only match the quality
of the former technique but also open possibilities for overcoming traditional limitations
[2 ]. One significant challenge are the major differences in X-ray transmittance of the
thorax. Systematic studies showed that conventional X-ray films with optimum contrast
in mid-density areas frequently fail to provide adequate diagnostic quality in the
apices, the retro-cardiac region and behind the diaphragm due to underexposure [3 ]
[4 ]. Current solutions including wide-latitude films, dual image receptor screen-film
combinations or fixed or flexible compensation filters placed close to the source
of the X-rays are either achieved at the cost of a considerable loss in edge contrast
or are technically complicated and expensive [5 ]
[6 ]
[7 ].
Digital systems cover a much wider range of exposures than a conventional film. However,
frequently the old contrast/latitude trade-off is re-introduced when a limited range
of exposures is selected from the raw data to display diagnostic images that look
like a conventional film radiogram. Additional information poorly displayed on the
far tails of the display curve is then sacrificed. Hence, different post-processing
techniques to widen the dynamic range of digital images have been developed [8 ]
[9 ]
[10 ]
[11 ]. EVP (enhanced visualization processing, Carestream Health, Inc., Rochester, NY,
USA) as one example, separately reduces the contrast of the low image frequencies,
thereby increasing the latitude [8 ]
[12 ]. Simultaneously, the high frequencies are enhanced to preserve detail contrast.
It was hypothesized that latitude enhancement of digital projection images would improve
the detection of artificial air space opacities behind the diaphragm, without significant
effects on other regions. The objective of this study was to perform an ex vivo experiment
with EVP as one of the current equalization post-processing techniques to determine
whether it could enhance the accuracy of the detection of artificial air space opacities
in parts of the lung that are superimposed by the diaphragm.
Materials and Methods
The ex vivo setup
The “artificial thorax” consists of a double-walled container holding a freshly excised,
inflated porcine heart-lung explant [13 ]
[14 ]. The artificial chest walls have a thickness of 2 – 5 cm and are filled with saline
to achieve realistic X-ray absorbency. The silicone “diaphragm” was filled with water.
19 lung specimens were harvested from regular-weight pigs (80 – 100 kg) at a local
slaughterhouse. No animals were sacrificed for the purposes of this study. A 6.5 mm
tracheal tube (Portex; SIMS Portex Ltd., Hythe, Kent, UK) was introduced into the
trachea and connected to an outlet through the phantom wall. The lungs were then inflated
by continuous evacuation of the artificial pleural space at –20 to –30 hPa. Lung collapse
and shifting during transfer from CT to the X-ray unit were prevented by maintaining
the evacuation. For chest X-ray the phantom was collocated onto a wooden support ([Fig. 1 ]).
Fig. 1 For chest X-ray studies, the artificial thorax a was collocated onto a wooden support b in an upright position. This made it possible to take chest X-ray exposures with
the direct radiography detector c as in patient studies. Lung collapse was prevented by keeping the system constantly
evacuated during positioning. The margins of the lung inside the posterior recesses
are indicated by black arrow heads.
Abb. 1 Für die digitale Direktradiografie wie an einem stehenden Patienten wurde das Thoraxphantom
a auf einem Holzgestell b in aufrechter Position fixiert. Der Kollaps der Lungen während der Positionierung
wurde durch einen konstanten Unterdruck im künstlichen Pleuraspalt verhindert. Die
Lungengrenzen in den posterioren Zwerchfellrippenwinkeln sind mit schwarzen Pfeilspitzen
markiert
Simulation of artificial air space opacities
Each of the 19 lungs was prepared with simulated pulmonary opacities by injecting
30 – 50 ml of a gelatin-stabilized liquid into the tracheobronchial system (7.5 g
of cold prepared food gelatin (Homann, Dissen, Germany) per 100 ml of water (aqua
ad injectabilia, Braun, Melsungen, Germany) [14 ]). Gelatin served to increase the viscosity and to prevent fast drying. To achieve
distribution into the distal air spaces, the evacuation was temporarily interrupted
to deflate the lungs. Then the liquid was instilled via a soft silicone tube during
re-inflation of the lung explants. The sites of liquid deposition included areas in
the posterior costophrenic angles (superimposed by the artificial diaphragm) and areas
above the diaphragmatic dome (non-superimposed). Per definition from the latest guidelines
of the Fleischner Society, the generated fluid accumulations inside the air spaces
are referred to as artificial air space opacities rather than infiltrates [15 ].
Computed tomography and documentation of findings
Injection of the liquid was carried out on top of the CT table of a commercial 16-row
detector scanner (Sensation 16, Siemens Healthcare, Erlangen, Germany). CT scans with
a standard chest protocol were acquired prior to and after the injection (100 mAs,
120 kVp, slice collimation 16 × 0,75 mm, table feed 15 mm, reconstructed slice thickness
1 mm, reconstruction increment 0.7 mm, B 70f kernel, FOV 350 mm, matrix 512 × 512
pixel). The location of the artificial air space opacities was recorded by two board
certified radiologists using a standard display (window width 1600 HU, center at –600
HU) on a commercial workstation (CT WIZARD®, syngoCT, Siemens Healthcare, Erlangen,
Germany; [Fig. 2 ]). Documentation was produced for eight regions of the image defined by lines at
the level of the carina, mid-way between carina/diaphragm and at the diaphragmatic
dome. The area below the top of the diaphragmatic dome was divided into the non-superimposed
lateral recesses and the parts that were superimposed by the hemidiaphragm. Finally,
each region was divided by the midline into a right (numbered from 11 – 14) and left
section (numbered from 21 – 24; [Fig. 3 ]). The retrocardium was excluded since the position of the heart was variable. To
match CT and radiography, the scout view of the CT and the radiograms were compared
side by side. These data served as the gold standard.
Fig. 2 CT cross section just above the diaphragm showing a dense artificial air space opacity
in the lateral left lower lobe (circled). Nb.: Chest tubes in the posterior recesses
for the evacuation of the artificial pleural space.
Abb. 2 CT-Querschnitt unmittelbar oberhalb der Zwerchfellkuppe mit Abbildung einer dichten
intraalveolären Verschattung im lateralen linken Lungenunterlappen (Kreis). Nb.: Drainagen
in den posterioren Abschnitten zur Absaugung der Luft aus dem künstlichen Pleuraspalt.
Fig. 3 Standard tone scale image of the phantom taken at 4 mAs, 125 kVp. The grid indicates
the margins of the 8 regions that were read for the presence of air space opacities
(the same lung as in Fig. 2 ).
Abb. 3 Aufnahme des Phantoms mit 4 mAs, 125 keV, Nachverarbeitung mit der Standard-Grauwertskala.
Die überlagerten Linien zeigen die Grenzen der 8 Regionen, die für das Vorhandensein
von Verschattungen bewertet wurden (gleiche Lunge wie in Abb. 2 ).
Digital radiography and image post-processing
The direct radiography system (KODAK DIRECTVIEW DR 9000 System, Rochester, NY, USA)
was equipped with a 35 × 43 cm flat panel detector consisting of an amorphous selenium
semiconductor X-ray absorber coating over a thin-film transistor array of amorphous
silicon. The matrix of 2560 × 3072 elements corresponded to a single pixel size of
139 µm. The image bit depth was 14. All images were acquired in the postero-anterior
projection at a distance of 180 cm at 125 kVp, once with a fixed time-current product
of 4 mAs and once with automatic exposure control (cut-off dose for the detector was
set to 4 µGy). This was compliant with the 5 µGy detector dose limit for 400 speed
systems defined by law in the country where the study was performed. The fixed 4 mAs
images were further processed for the study. The average tube load with automatic
exposure was only obtained as a measure of the phantoms overall X-ray transmittance
(4.49 mAs ± SD 0.61). For comparison, we retrospectively analyzed the average tube
load from automatic exposures in 70 consecutive routine examinations of the same unit
(3.88 mAs ± SD 1.04 for adult male patients). Raw image data (2456 × 2968 pixel) for
each image was archived to CD-ROM and transferred to a personal computer equipped
with a copy of the fully functional image processing software of the DR unit.
Radiographies without EVP post-processing were produced with the default automated
tone scale algorithm of the Kodak DirectView systems. It is based on a “perceptually
linear” tone scale that properly incorporates the characteristics of the human visual
system [16 ]
[17 ]
[18 ]. The essential elements of the algorithm, i. e., image analysis, tone scale generation,
and tone scale application, produce a display-ready image that is similar in appearance
to optimum screen-film imaging [19 ]. The purpose of enhanced visualization processing (EVP) is to increase the latitude
of radiographic images while preserving or enhancing the contrast of image detail.
Latitude is defined as the difference between the lowest and the highest exposure
value covered by the grayscale of the image. Exposures below this window are displayed
in white, exposures above in black tones. This is accomplished by analyzing the image
as low and high-frequency component images. The contrast of the low-frequency image
is reduced, thereby increasing the latitude (range of exposures visible in the image).
The contrast of the high-frequency image may be enhanced to preserve the visual appearance
of image detail. Finally, the altered low- and high-frequency image components are
recombined and the tone scale mapping is applied. Hence the effect of EVP is to lower
the overall or global contrast of the image, thereby increasing the latitude of the
displayed image. In this way image features that would have been very light or very
dark are made darker or lighter, respectively. In addition, the contrast of image
details smaller than the kernel size is increased. EVP thus extends the latitude of
the displayed image, without any loss of contrast for details at mid-range exposures.
A larger fraction of the potentially useful diagnostic information is presented without
the need for image manipulation by the radiologist. For a detailed description of
the technique, refer to [8 ]
[12 ].
In this study, renderings of each image were prepared with the default automated tone
scale algorithm as well as with five levels of latitude enhancement having EVP gain
values (α-1) from 1.0 to 3.0 in increments of 0.5. An EVP gain of 2.0 doubles the
latitude of an image. The EVP kernel size was set to the manufacturer’s default value
for chest imaging, 1 / 20 of the short dimension of the image (17.5 mm). The high-frequency
gain (β) in these experiments was set to the recommended default of 1.1. The images
were extracted in six different processing conditions and saved for review as DICOM
files ([Fig. 4 ]). Then the original image was mirrored in left-right direction and another set of
six images was produced. These mirrored images were included to reduce recall bias
since all images appeared six times with different image processing conditions. Additional
X-rays were obtained prior to the injection of liquid in 4 of 19 lungs and included
as negative controls. Thus, a total of 23 acquisitions and a total of 276 rendered
images were available for reading.
Fig. 4 Example demonstrating the effects of image post-processing with increased latitude
on an image of the phantom taken at 4 mAs, 125 kVp (same lung as in Fig. 2, 3 ). The artificial air space opacity in the lateral left lower lobe (indicated by white
arrow heads in image d ) is hardly visible at standard tone scale a and becomes clearly visible with increased latitude with log.-factors of 1.0 b , 2.0 c and 3.0 d. The difference between c and d is only marginal.
Abb. 4 Ein Beispiel zur Demonstration der Bildnachverarbeitung mit erweiterter Latitude
an einer Aufnahme bei 4 mAs und 125 keV (gleiche Lunge wie in Abb. 2, 3 ). Die artifiziellen intraalveolären Verschattungen im lateralen linken Lungenunterlappen
(markiert mit weißen Pfeilspitzen in d ) sind bei Nachverarbeitung mit der Standard-Grauwertskala schwer erkennbar a , demarkieren sich aber deutlicher mit Erweiterung der Latitude um Log.-Faktoren von
1,0 b , 2,0 c und 3,0 d. Hierbei ist der Unterschied zwischen c und d nur noch marginal.
Image reading
The images were presented in a room with low ambient lighting on the diagnostic 1280
× 1024 pixel TFT monitor of a commercial workstation (monitor: SMD 1879-M, Siemens,
Erlangen, Germany; graphic processor: Matrox Millennium G 450 DVI, Matrox Graphics,
Dorval, Quebec, Canada; standard viewing software: Magic View 300, Siemens, Erlangen,
Germany; monitor calibration for grayscale display was performed by visual assessment
of display bit-depth with the TG 18-MP pattern of the AAPM (American Association of
Physicists in Medicine TG#18). All images appeared individually in randomized order
and in full size on a black background with fixed default window settings of 2048
(center)/ 4095 (width). The initially fixed monitor settings and ambient light were
kept constant throughout the study. The images were randomized and presented to six
independent observers who were familiar with chest X-ray reading from the daily routine
(2, 2, 3, 4, 6 and 8 years of chest radiology experience, respectively).
The images were read for the presence or absence of opacifications in the above defined
regions. To reduce the learning bias, the readers were instructed during a separate
session with 25 training images and subsequent presentation of the correct information
(these images were not presented again in the main study). Reading time was limited
to 6 min per image and four hours per day. Breaks could be taken whenever desired,
but at least every hour for 10 min.
Statistics
The sensitivity and specificity for the artificial air space opacities were calculated.
The six regions 11 – 13 and 21 – 23 were regarded as representative of non-superimposed
lung parenchyma above the diaphragm, while the two regions 14 and 24 represented the
superimposed parts of the lung (anterior and posterior pleural recesses, [Fig. 2 ]). The significance of different latitude levels upon sensitivities and specificities
for each individual reader was calculated with McNemar’s test. A p-value of 0.05 was
defined as statistically significant. For statistical evaluation of diagnostic accuracy,
the results for sensitivities and specificities were averaged over the six readers.
As suggested by van den Hout, the mean sensitivity and specificity at different processing
levels were summarized as Az = 1 / 2 (sensitivity + specificity), which corresponds
to the area under an ROC curve with a single data point (non-parametric calculation/trapezoidal
rule under the assumption of a concave receiver operating characteristic (ROC) curve)
[20 ]
[21 ]. Positive and negative predictive values were not calculated since according to
Bayes’ theorem they depend on the incidence of a finding and could be manipulated
in the experiment. Interobserver agreement was described with Cohen’s kappa. A kappa
value smaller that 0.10 was rated as no agreement (0.10 – 0.40 “weak”, 0.41 – 0.60
“moderate” and 0.61 – 0.80 “good” agreement). A correlation of results with the density
of findings on CT was not performed since this data was only descriptive. Calculations
were made with standard software (Excel 97, Microsoft, Redmond, WA, USA; SPSS, 10.0,
SPSS Inc., Chicago, IL, USA).
Results
Radiographic presentation of the artificial air space opacities
The radiographic presentation of the opacifications varied from dense and easily detectible
to very fine, almost invisible attenuations. This corresponded to the appearance in
CT where diffuse, confluent and dense opacifications were seen at segmental and subsegmental
levels. 23 CT scans, each with 8 subregions were included in the statistical evaluation.
Of the resulting 184 subregions, 84 presented with opacifications, 11 were partially
collapsed (characterized as atelectasis) and 4 had both partial atelectasis and artificial
air space opacities. 85 subregions presented without any finding. For the distribution
of findings behind and above the diaphragm, refer to [Table 1 ].
Table 1 Results from 23 CT scans (19 of 23 after generation of artificial air space opacities).
The table contains the totaled results for all regions (left column) and separately
for regions behind the diaphragm (middle) and above the diaphragm (right; for the
definition of regions Fig. 3 ). NB: The number of regions in Table 2 is twice the number in this table since all radiograms were presented twice: one
time as original, one time mirrored.
Tab. 1 Ergebnisse aus 23 CT-Datensätzen (19 von 23 nach Erzeugung artifizieller intraalveolärer
Verschattungen). Die Tabelle enthält die Gesamtzahl der Ergebnisse für alle Regionen
(linke Spalte) sowie getrennt für die Regionen hinter der Zwerchfellkuppe (in der
Mitte) und oberhalb der Zwerchfellkuppe (rechts; zur Definition der Regionen Abb. 3 ). Nb.: Die Tab. 2 weist eine 2-fach höhere Anzahl an Regionen auf, da die Röntgenaufnahmen 2-fach präsentiert
wurden (im Original und um die Längsachse gespiegelt).
n = 184
total (regions 11 – 14 and 21 – 24; n = 184)
superimposed by the diaphragm (regions 14 and 24; n = 46)
not superimposed by the diaphragm (regions 11 – 13 and 21 – 23; n = 138)
no opacification
air space opacity
no opacification
air space opacity
no opacification
air space opacity
no atelectasis
85
84
22
16
63
68
atelectasis
11
4
6
2
5
2
Accuracy of detecting artificial air space opacities in superimposed parts of the
lung
The diagnostic accuracy for the detection of artificial air space opacities in superimposed
parts of the lung behind the diaphragm was calculated from 3312 single observations
(n = 92 regions, each presented at six latitude levels and read by six observers.
For details, see the left section of [Table 2 ] and [Fig. 5 ]
a ). Behind the diaphragm (artificial air space opacities present in 32 / 92 regions),
the median sensitivity for artificial air space opacities improved from 0.35 over
0.50 at a latitude enhancement gain of 1.0 (changes not significant) to 0.56 at a
gain of 3.0 (changes significant in 3 of the 6 observers). The specificity remained
almost unchanged around 0.95 for latitude enhancement gains of up to 2.0, but decreased
to 0.90 at a gain of 3.0 (changes compared to the standard tone scale significant
in 3 of 6 observers). The Az values improved from 0.66 at the standard tone scale
to 0.73 – 0.74 at any level of enhanced latitude post-processing. The difference with
respect to the median Az at the standard tone scale was significant at all levels.
Interobserver agreement improved from 0.39 to a maximum of 0.48 at a gain of 1.0 ([Table 3 ]). A further increase in latitude resulted in median kappa values between 0.40 and
0.46 (individual values ranging from 0.32 – 0.55, changes not significant).
Table 2 Diagnostic accuracy for the detection of artificial air space opacities behind the
diaphragm (3312 single observations from 46 images, n = 92 regions, six latitude levels
and six observers) and above (9936 single observations from 46 images, n = 276 regions,
six latitude levels and six observers). The median of the observers appears in italic
type, and the range of the individual results and the fraction of observers with significant
changes compared to the standard tone scale appear in brackets.
Tab. 2 Diagnostische Treffsicherheit für die Detektion artifizieller intraalveolärer Verschattungen
hinter dem Zwerchfell (3312 Einzelbeobachtungen der 6 Beobachter an 46 Aufnahmen für
92 Regionen und 6 Stufen der Latitudenerweiterung) und oberhalb der Zwerchfellkuppe
(9936 Einzelbeobachtungen der 6 Beobachter an 46 Aufnahmen für 276 Regionen und 6
Stufen der Latitudenerweiterung). Der jeweilige Median der Beobachter ist in Kursivdruck
wiedergegeben, die Spannweite der individuellen Ergebnisse und der Anteil der Beobachter
mit signifikant von der Standard-Grauwertskala abweichenden Ergebnissen in runden
Klammern.
latitude
behind diaphragm (n = 92 regions)
above diaphragm (n = 276 regions)
level
sensitivity
specificity
Az
sensitivity
specificity
Az
median (range)
median (range)
median (range)
median (range)
0
0.35
(0.14 – 0.64)
0.96
(0.84 – 0.98)
0.66
0.71
(0.60 – 0.75)
0.76
(0.66 – 0.84)
0.74
1
0.50
(0.33 – 0.72; 0 / 6 sign.)
0.95
(0.86 – 1.00; 1 / 6 sign.)
0.73
0.77
(0.72 – 0.85; 3 / 6 sign.)
0.70
(0.50 – 0.78; 2 / 6 sign.)
0.74
1.5
0.53
(0.42 – 0.94; 3 / 6 sign.)
0.95
(0.81 – 1.00; 1 / 6 sign.)
0.74
0.75
(0.72 – 0.88; 3 / 6 sign.)
0.64
(0.30 – 0.76; 3 / 6 sign.)
0.70
2
0.51
(0.47 – 0.94; 4 / 6 sign.)
0.95
(0.77 – 0.96; 1 / 6 sign.)
0.73
0.81
(0.76 – 0.87; 5 / 6 sign.)
0.64
(0.30 – 0.79: 6 / 6 sign.)
0.73
2.5
0.54
(0.40 – 0.92; 4 / 6 sign.)
0.91
(0.73 – 0.96; 2 / 6 sign.)
0.73
0.81
(0.76 – 0.89; 5 / 6 sign.)
0.64
(0.30 – 0.79; 5 / 6 sign.)
0.73
3
0.56
(0.50 – 0.92; 3 / 6 sign.)
0.90
(0.73 – 0.98; 3 / 6 sign.)
0.73
0.82
(0.78 – 0.91; 5 / 6 sign.)
0.62
(0.28 – 0.78; 5 / 6 sign.)
0.72
Table 3
Tab. 3
EVP
interobserver kappa
median range
behind diaphragm
off
0.39
0.27 – 0.52
1.0
0.48
0.38 – 0.59
1.5
0.46
0.38 – 0.55
2.0
0.46
0.38 – 0.54
2.5
0.40
0.32 – 0.48
3.0
0.44
0.37 – 0.52
above diaphragm
off
0.55
0.52 – 0.58
1.0
0.52
0.49 – 0.56
1.5
0.48
0.44 – 0.52
2.0
0.44
0.40 – 0.49
2.5
0.48
0.43 – 0.52
3.0
0.48
0.44 – 0.53
Fig. 5 Graphics demonstrating the contents of [Table 2 ]: Diagnostic accuracy for the detection of artificial air space opacities behind
a and above the diaphragm b . Black marks and the dotted line represent the median of the sensitivity calculated
from the observations of six different readers at given latitude levels from 0 to
3.0. The span of the individual results is indicated by error bars. Accordingly, dark
gray marks and corresponding error bars with a dashed line represent the specificity.
Light gray marks and the straight line represent the Az values calculated from the
medians of sensitivity and specificity.
Abb. 5 Grafische Illustration des Inhalts von [Tab. 2 ]: Diagnostische Treffsicherheit für die Detektion artifizieller intraalveolärer Verschattungen
hinter a und über der Zwerchfellblase b . Die schwarzen Marken und die gepunktete Linie zeigen den Median der Sensitivität
berechnet aus den Aufzeichnungen der 6 Beobachter bei Latituden-Erweiterung von 0
– 3,0. Die Spannweite der Einzelwerte wird durch Fehlerbalken angezeigt. Dunkelgraue
Marken mit Fehlerindikator und einer gestrichelten Linie geben die Spezifität wieder.
Hellgraue Marken und die durchgezogene Linie entsprechen den Az-Werten (Fläche unter
der ROC-Kurve, berechnet aus den Medianwerten für die Sensitivität und Spezifität).
Accuracy of detecting artificial air space opacities in non-superimposed parts of
the lung
The calculation of diagnostic accuracy for the detection of artificial air space opacities
in non-superimposed parts of the lung (above the diaphragm) was based on 9936 single
observations (n = 276 regions, each presented at six latitude levels and read by six
observers, opacifications present in 136 / 276 regions. For details, see the right
section of [Table 2 ] and [Fig. 5 ]
b ). Within this group, the median sensitivity for artificial air space opacities increased
from 0.71 on standard tone scale images over 0.77 with latitude enhancement at a gain
of 1.0 to 0.82 at a gain of 3.0 (changes compared to the standard tone scale significant
in 5 of the 6 observers). The specificity decreased from 0.76 at the standard tone
scale over 0.70 with latitude enhancement at a gain of 1.0 to 0.62 at a gain of 3.0
(changes significant in 5 of 6 observers). The Az values ranged from 0.70 to 0.74
and at any level of latitude (changes not significant). Median kappa values for interobserver
agreement changed from a maximum of 0.55 at the standard tone scale over 0.52 at a
latitude level of 1.0 to 0.44 – 0.48 at a gain of 3.0 (changes not significant). Differences
between observers related to chest radiology experience were not observed.
Discussion
The ex vivo study demonstrated that post-processing of digital projection chest radiograms
with moderately increased latitude and simultaneously enhanced detail contrast improved
the sensitivity for artificial air space opacities, in particular for parts of the
lung superimposed by the diaphragm. However, depending on the level of EVP gain, this
effect was significant in only 4 of 6 observers. The specificity decreased simultaneously
indicating false positive findings. Hence, overall improvement of diagnostic accuracy
was registered only for superimposed parts of the lung behind the diaphragm and did
not change above the diaphragm. Individual differences in the response to the level
of increased latitude were observed. Interobserver agreement for the regions behind
the diaphragm improved from standard tone scale to latitude enhancement with a gain
of 1.0 but tended to decrease again at higher levels.
As an advantage of the study, the experiment simulated a realistic setting: reading
chest X-rays for pulmonary opacities. The necessary raw image data with CT correlation
could be easily produced. Comparable clinical data with proven infiltrative lung disease
and corresponding CT would have been more difficult to obtain: repeated exposures
would have been prohibitive in patients or at least ethically problematic in a laboratory
animal [22 ]
[23 ]. The phantom size, radiation absorption and the dynamic range of X-ray transmittance
(which is proportional to the required latitude of the imaging system) were equivalent
to those of a large human thorax [13 ]
[14 ]. This was confirmed by retrospective comparison of the tube loads for adult male
patients in automatic exposure control images. By using a fixed exposure of 4 mAs,
variations with respect to the position of the phantom and the inserted material were
excluded.
Realistic artificial air space opacities of variable extent and density were produced
with gelatin-stabilized liquid [14 ]. Unlike for pulmonary nodules, electronic simulation of such “infiltrates” was not
available [24 ]
[25 ]
[26 ]. It was assumed that differences between porcine and human material would not be
relevant for the detection of pulmonary opacifications. Multiple row detector CT provided
the best available standard for the detection of pulmonary opacifications [27 ]
[28 ]
[29 ]
[30 ]. The higher sensitivity of CT for the artificial air space opacities in this experiment
is consistent with studies on immunocompromised patients which demonstrated that the
sensitivity of chest X-ray for subtle pulmonary infiltrates approximates 50 – 70 %
[30 ]
[31 ]
[32 ]. A realistic amount of anatomic noise was represented [23 ]. Independent factors such as heart pulsation or attenuations related to the chest
wall (ribs and spine) were excluded. The influence of overlying bones could not be
studied, but this has already been explicitly assessed by other investigators [23 ]
[25 ].
Limitations of the study
The presented data should not be interpreted as a measure to quantify differences
in the diagnostic accuracy of X-ray and CT, since the density of opacifications was
not standardized. The accuracy of detecting air space opacities behind and above the
diaphragm was not directly compared since the prevalence of opacifications in both
regions differed after exclusion of atelectases (32 / 92 and 136 / 276, respectively).
Furthermore, the lack of blood inside the lung vessels may have biased the detection
results between the superimposed and non-superimposed regions. Theoretically, the
lack of a vessel filling might facilitate the detection of opacities in the non-superimposed
parts more than in the superimposed parts. The lack of other superimposed structures
such as bones and air-filled parts of the bowel in the upper abdomen or even lung
scars may have contributed to an overestimation of the effects of detail enhancement
as well. These structures would have been displayed with sharper delineation and contrast.
This might have reduced the positive effects on the detection of air space opacities.
Further investigation of this subject should therefore include chest wall structures,
e. g. in an experimental setup using cadavers and superimposed nodules to simulate
pathology. A more sophisticated setup should principally allow for imaging in two
planes as well [31 ]
[32 ]
[33 ]
[34 ]. In clinical practice, the lateral image plane is very helpful for detecting lower
lobe infiltrates. However, the design of the phantom was not suitable for lateral
projections since the flanges and screws are superimposed on the image. Hence, only
posterior-anterior projections were analyzed.
Clinical relevance
Latitude enhancement is a general approach to improving radiograms of subjects with
a large dynamic range. Our results are specific to EVP, but other programs (e. g.
DRC = dynamic range control processing [9 ]) produce similar effects. For DRC it was shown that it improves the visibility of
tubes and lines superimposed on the mediastinal tissues if applied to bedside images
taken with a mobile unit. When used for in-department chest radiography, it was assumed
to provide slight advantages in the evaluation of disease in the mediastinum [9 ]
[11 ]. For the EVP algorithm, this was already shown with ex vivo experiments using artificial
pulmonary nodules, but the effects on the detection of subtle opacifications have
not yet been studied [12 ].
An advantage of EVP over other techniques is that it is separately effective on high
and low frequency components of the full scale raw image [12 ]. This makes the difference to simply applying a larger window width to all image
components. The resulting modified raw image combines a large latitude with good detail
contrast. This image is then subject to further routine processing. Other products,
e. g. DRC [9 ], are applied to image data within the display curve. Principally this data has been
already truncated by preprocessing, and the effectiveness of such procedures is potentially
limited. DRC and EVP have been commercially introduced and studies on storage phosphor
systems have demonstrated their comparable capacity for displaying chest radiograms
with a wide dynamic range [8 ]
[9 ]. In theory, EVP may enhance image noise by enhancing the high frequencies. This
effect was not further evaluated, but may be critical in instances with very high
EVP gain.
A clinical application of enhanced latitude post-processing would be to improve softcopy
image review by sending original (unprocessed) image data to a review station thereby
allowing readers to adjust image latitude and detail contrast interactively [10 ]
[12 ]. Although processing with recalculation of the images on a dedicated workstation
with preset keys would take less than five seconds, it appears questionable whether
this additional effort would be accepted as routine in a busy radiological department.
In clinical practice, image processing methods with automatic tone scale rendering
and latitude and contrast enhancements are accepted for producing display ready hardcopies
or minimizing the time required for image manipulation during softcopy reading [35 ]. These effects on clinical workflow were documented recently [35 ]. Since the recommendations for the latitude enhancement gain can probably not be
directly applied to other products, further studies with other products might be useful.
Finally it was concluded from the present data that image post-processing of digital
p. a. chest radiograms with EVP improves the diagnostic accuracy for artificial air
space opacities in the superimposed parts of the lung (area under the ROC curve).
Above the diaphragm accuracy is not affected due to a trade-off in sensitivity/specificity.
For practical use, a moderate EVP gain of 1.5 appears to be suitable for most readers.
The positive effect of latitude enhancement on sensitivity became statistically significant
for 3 / 6 observers at a level of 1.5 and the negative effects on specificity became
effective for a gain of 1.5 for parts above the diaphragm. Hence, in consistency with
other literature, we suggest a latitude enhancement gain between 1.0 and 1.5 to be
suitable for most readers [12 ]
[36 ]. With respect to optimum detectability of pulmonary opacifications without a significant
loss in specificity, we recommend not exceeding a level of 1.5.