Endoscopy 2009; 41(2): 143-148
DOI: 10.1055/s-0028-1119454
Original article

© Georg Thieme Verlag KG Stuttgart · New York

Magnetic retraction in natural-orifice transluminal endoscopic surgery (NOTES): addressing the problem of traction and countertraction

M.  Ryou1 , C.  C.  Thompson1
  • 1Division of Gastroenterology, Brigham and Women’s Hospital, Harvard Medical School, Boston, Massachusetts, USA
Further Information

Publication History

submitted 14 September 2008

accepted after revision 5 November 2008

Publication Date:
12 February 2009 (online)

Background and study aims: Because of their reliance on the flexible endoscope, most current procedures in natural-orifice transluminal endoscopic surgery (NOTES) suffer from the inability to vigorously grasp and move tissue or to retract organs. We aimed to assess the use of internal and external magnets that might allow the vigorous multiaxial traction/countertraction required in more complicated NOTES procedures.

Methods: Ex vivo and in vivo porcine model. Study components were: (1) Evaluation of force–distance relationship of this magnetic retraction system using a digital tensiometer. (2) Application of this magnetic retraction system to two procedures in the porcine model: (a) Liver retraction during transcolonic cholecystectomy in five nonsurvival pigs. Procedure time was recorded and compared to historical controls. (b) Mesh positioning for implantation into the anterior abdominal wall for ventral hernia repair in three survival pigs.

Results: Over a distance of 5 cm to 0.25 cm, the magnetic force of our system increased from 3 to 90 gramforce (29.42 to 882.60 mN.) In vivo, the magnet system provided robust liver retraction, shortening the procedure time of NOTES cholecystectomy from a historical mean of 68 minutes (range 42 – 90 minutes; n = 5) to 49.6 minutes (range 33 – 61 minutes; n = 3). The magnetic system also greatly enhanced mesh positioning and stability, and these animals survived for 2 weeks without complications.

Conclusions: Our basic system provided critical liver retraction during NOTES cholecystectomy and was also instrumental in moving and stabilizing mesh for implantation during NOTES ventral hernia repair. Magnets can potentially provide the vigorous traction and countertraction required to advance NOTES procedures.

References

  • 1 Park P O, Bergstrom M, Ikeda K. et al . Experimental studies of transgastric gallbladder surgery: cholecystectomy and cholecystogastric anastomosis (videos).  Gastrointest Endosc. 2005;  61 601-606
  • 2 Swanstromm L L, Kozarek R, Pasricha P J. et al . Development of a new access device for transgastric surgery.  J Gastrointest Surg. 2005;  9 1129-1136
  • 3 Chopita N, Vaillaverde A, Cope C. et al . Endoscopic gastroenteric anastomosis using magnets.  Endoscopy. 2005;  37 313-317
  • 4 Arana A, Hauser B, Hachimi-Idrissi S. et al . Management of ingested foreign bodies in childhood and review of the literature.  Eur J Pediatr. 2001;  160 468-72
  • 5 Chow D R, Garretson B R, Kuczynski B. et al . External versus internal approach to the removal of metallic intraocular foreign bodies.  Retina. 2000;  20 364-369
  • 6 Noar J H, Evans R D. Rare earth magnets in orthodontics: an overview.  Br J Orthod. 1999;  26 29-37
  • 7 Pai R D, Fong D G, Bundga M E. et al . Transcolonic endoscopic cholecystectomy: a NOTES survival study in a porcine model (with video).  Gastrointest Endosc. 2006;  64 428-434
  • 8 Fong D G, Pai R D, Thompson C C. Transcolonic endoscopic abdominal exploration: a NOTES survival study in a porcine model.  Gastrointest Endosc. 2006;  65 312-318
  • 9 Fong D G, Ryou M, Pai R D. et al . Transcolonic ventral wall hernia mesh fixation in a porcine model.  Endoscopy. 2007;  39 865-869

C. C. ThompsonMD, MHES 

Brigham and Women’s Hospital
Division of Gastroenterology

75 Francis St
Boston, MA 02115
USA

Email: ccthompson@partners.org