Background and study aims: Tubular stents have been used to accomplish endoscopic transluminal drainage, but do not impart lumen-to-lumen anchorage. We evaluated a novel lumen-apposing stent designed for enteric drainage of nonadherent lumens.

Material and methods: Ex vivo benchtop testing was performed to quantify various physical and performance metrics of the stent (Axios). A simulator was developed to test the stent deployment through an echoendoscope. Survival experiments were performed on four pigs. Under endosonographic guidance, a cholecystogastrostomy tract was created and the stent was deployed across the lumens. Direct cholecystoscopy was performed. Surveillance gastroscopy was performed at weekly intervals for up to 8 weeks. Measured outcomes were procedural success, safety, and device durability and patency.

Results: In benchtop testing, the Axios stent withstood various vector forces of movement, yet allowed easy removability from the simulated tissue. The stent was successfully deployed across the stomach and gallbladder lumens in all four animals without complication. Direct cholecystoscopy and contrast injection documented the absence of tissue trauma and leakage. Gastroscopy at weekly intervals showed the stent in stable position without dislodgment. The stent remained patent in all animals. The covering remained intact and there was no hyperplastic tissue ingrowth or overgrowth, or tissue injury. One stent was removed at 4 weeks. On necropsy, the gallbladders showed focal adherence to the stomach at the site of cystogastrostomy with a negative leak test.

Conclusions: The Axios stent enables the creation of a robust and reliable conduit between nonadherent lumens around the gastrointestinal tract.

References

  • 1 Varadarajulu S, Trevino J M. Review of EUS-guided pancreatic duct drainage (with video).  Gastrointest Endosc. 2009;  69 S200-S202
  • 2 Itoi T, Coelho-Prabhu N, Baron T H. Endoscopic gallbladder drainage for management of acute cholecystitis.  Gastrointest Endosc. 2010;  71 1038-1045
  • 3 Piraka C, Shah R J, Fukami N et al. EUS-guided transesophageal, transgastric, and transcolonic drainage of intra-abdominal fluid collections and abscesses.  Gastrointest Endosc. 2009;  70 786-792
  • 4 Fritscher-Ravens A, Mosse C A, Mills T N et al. A through-the-scope device for suturing and tissue approximation under EUS control.  Gastrointest Endosc. 2002;  56 737-742
  • 5 Belle S, Collet P, Post S et al. Temporary cystogastrostomy with self-expanding metallic stents for pancreatic necrosis.  Endoscopy. 2010;  42 493-495
  • 6 Fritscher-Ravens A, Mosse C A, Mukherjee D et al. Transluminal endosurgery: single lumen access anastomotic device for flexible endoscopy.  Gastrointest Endosc. 2003;  58 585-591
  • 7 Cope C, Faigel D O, Ginsberg G G et al. Creation of a gastroenteric anastomosis with endoscopy and percutaneous gastrostomy in pigs.  J Vasc Interv Radiol. 2008;  19 124-128
  • 8 Cope C, Ginsberg G G. Long-term patency of experimental magnetic compression gastroenteric anastomoses achieved with covered stents.  Gastrointest Endosc. 2001;  53 780-784
  • 9 Chopita N, Vaillaverde A, Cope C et al. Endoscopic gastroenteric anastomosis using magnets.  Endoscopy. 2005;  37 313-317

K. F. BinmoellerMD 

Interventional Endoscopy Services
California Pacific Medical Center

2351 Clay St, 6th Floor
San Francisco
CA 94115
USA

Email: BinmoeK@sutterhealth.org