Email: mwilkinson@notes.cc.sunysb.edu
Publication History
Publication Date:
19 July 2011 (online)
A 72-year-old woman underwent a laparorobotic pancreaticoduodenectomy using the da Vinci Surgical System (Intuitive Surgical Inc., Sunnyvale, California, USA) for removal of a 4-cm ampullary adenoma. Intraoperatively, a nasogastric tube was inserted into the gastric fundus and its proximal end was secured to the patient’s nares. On postoperative day 2, the tube could not be removed at the bedside.
Esophagogastroduodenoscopy revealed that the nasogastric tube was trapped along the gastrojejunostomy anastomotic staple line by a few staples ([Fig. 1]).
Initial attempts to completely liberate this portion of the tube were unsuccessful despite the use of multiple mechanical modalities (such as rat-tooth forceps, loop wire cutters, snares, and standard endoscopic scissors) in combination with monopolar and bipolar electrocautery ([Fig. 2]).
Finally, a pair of flexible endoscopic scissors, FlexShears (Apollo Endosurgery Inc., Austin, Texas, USA) ([Fig. 3]), were inserted into the side hole of the nasogastric tube and used to completely cut though the tube as it was dislodged from the staple line ([Fig. 4]; [Video 1]).
Quality:
No evidence of subsequent disruption or dehiscence at the anastomosis was observed, and the patient continued her postoperative recovery without further complications or the need for further operations.
Comparable scissors-type devices have been used to aid in endoscopic submucosal dissection of both gastric [1] and colorectal [2] neoplasms, but these devices function more as grasping forceps than actual cutting shears. Karmali et al. reported the successful use of the Olympus straight endoscopic scissors device to remove eroded laparoscopically placed gastroplasty bands in nine patients with obstructive symptoms [3]. The FlexShears described in the current report act as single-use flexible cutting shears designed for tissue dissection during flexible endoscopic procedures. This is the first reported case in humans demonstrating safe and effective use of these flexible endoscopic scissors since their approval by the Food and Drug Administration (FDA) in 2009.
Endoscopy_UCTN_Code_TTT_1AO_2AD
Competing interests: None
#References
- 1 Akahoshi K, Honda K, Akahane H et al. Endoscopic submucosal dissection by using a grasping-type scissor forceps: a preliminary clinical study. Gastrointest Endosc. 2008; 67 1128-1133
- 2 Honma K, Kobayash M, Watanabe H et al. Endoscopic submucosal dissection for colorectal neoplasia. Dig Endosc. 2010; 22 307-311
- 3 Karmali S, Snyder B, Wilson E B et al. Endoscopic management of eroded prosthesis in vertical banded gastroplasty patients. Surg Endosc. 2010; 24 98-102
M. WilkinsonMD
Division of Gastroenterology
Stony Brook University
Medical Center
100 Nicolls Road
HSC Level 17, Room 060
Stony
Brook
New York 11794
USA
Fax: +1-631-444-8886
Email: mwilkinson@notes.cc.sunysb.edu
References
- 1 Akahoshi K, Honda K, Akahane H et al. Endoscopic submucosal dissection by using a grasping-type scissor forceps: a preliminary clinical study. Gastrointest Endosc. 2008; 67 1128-1133
- 2 Honma K, Kobayash M, Watanabe H et al. Endoscopic submucosal dissection for colorectal neoplasia. Dig Endosc. 2010; 22 307-311
- 3 Karmali S, Snyder B, Wilson E B et al. Endoscopic management of eroded prosthesis in vertical banded gastroplasty patients. Surg Endosc. 2010; 24 98-102
M. WilkinsonMD
Division of Gastroenterology
Stony Brook University
Medical Center
100 Nicolls Road
HSC Level 17, Room 060
Stony
Brook
New York 11794
USA
Fax: +1-631-444-8886
Email: mwilkinson@notes.cc.sunysb.edu