Subscribe to RSS
DOI: 10.1055/s-0031-1291515
Over-the-scope clip used to control bleeding from a duodenal ulcer
Corresponding author
Publication History
Publication Date:
06 March 2012 (online)
The over-the-scope clip (OTSC) is a new device that was initially developed for the closure of perforations of the gastrointestinal tract, but has since been used for a variety of indications including closure of fistulas [1] [2]. We report the use of an OTSC to control persistent bleeding from a duodenal ulcer.
A 57-year-old woman was transferred to our hospital in February 2011 because of acute renal failure with hyperglycemic acidosis secondary to type I diabetes. During her admission she developed a non-ST elevation myocardial infarction (NSTEMI) and was commenced on double platelet aggregation inhibitor therapy with acetylsalicylic acid and clopidogrel.
She was readmitted to our hospital 3 weeks later with acute gastrointestinal bleeding, presenting as melena and hematemesis. Immediate endoscopy showed a duodenal ulcer (Forrest Ia) with arterial bleeding ([Fig. 1]) that was stopped with injection of saline and adrenaline (1 : 10000) and subsequent application of hemoclips (Olympus, Hamburg, Germany). A Helicobacter pylori quick test was positive and the patient was treated with intravenous pantoprazole twice daily and commenced on eradication therapy. The patient was stable overnight on the intensive care unit and underwent a further endoscopy the next day. During this examination, rebleeding was encountered and was treated with epinephrine, further hemoclips, and injection of fibrin ([Fig. 2]).




The patient chose not to undergo the surgical intervention that was proposed to her at this stage, instead opting strongly for treatment using an OTSC (Ovesco, Tübingen, Germany). This was attempted the following day, by which time the hemoclips had again been lost and an actively bleeding ulcer (now Forrest Ib) was present ([Fig. 3]). Because it was not possible to get the whole ulcer into the OTSC overcap ([Fig. 4 a]), the OTSC was centered on the vessel, which was laterally placed ([Fig. 4 b]). Following application of the OTSC the bleeding stopped immediately. Follow-up endoscopies 3 days and 6 days later showed no further signs of bleeding. The patient was discharged and returned for follow-up endoscopies 4 weeks and 12 weeks later, at which times the OTSC was found to be still in situ ([Fig. 5]).








Endoscopy_UCTN_Code_TTT_1AO_2AD
#
Competing interests: None
-
References
- 1 Kirschniak A, Kratt T, Stüker D et al. A new endoscopic over-the-scope clip system for treatment of lesions and bleeding in the GI tract: first clinical experiences. Gastrointest Endosc 2007; 66: 162-167
- 2 Repici A, Arezzo A, De Caro G et al. Clinical experience with a new endoscopic over-the-scope clip system for use in the GI tract. Dig Liver Dis 2009; 41: 406-410