J Knee Surg 2012; 25(02): 151-154
DOI: 10.1055/s-0032-1313916
Original Article
Thieme Medical Publishers 333 Seventh Avenue, New York, NY 10001, USA.

Comparison of Kneelax and KT-1000 Knee Ligament Arthrometers

Russ Paine
1   Department of Sports Medicine, Memorial Hermann Sports Medicine Institute, Houston, Texas
,
Walter Lowe
2   Department of Orthopedic Sports Medicine, University of Texas Medical School, Memorial Hermann Sports Medicine Institute, Houston, Texas
› Author Affiliations
Further Information

Publication History

16 September 2011

22 January 2012

Publication Date:
23 May 2012 (online)

Abstract

Two commercially available knee ligament arthrometers (KT-1000 MEDmetric Corporation, San Diego, California and Kneelax MR Systems, Haarlem, The Netherlands) were used to measure anterior tibial translation. In this study 91 subjects with no history of knee pathology were tested using both devices. Both normal knees were tested and comparison of recorded anterior translation (millimeters) was performed. Absolute numbers at forces of 67N, 89N, 134N, and manual maximum displacements were recorded. Side-to-side differences were calculated and data were compared. There was a significant difference found between the absolute values of each device. The KT-1000 was found to record greater values in millimeters of translation. Although the absolute values were significantly different, the side-to-side differences were not significantly different between devices.

 
  • References

  • 1 Boniface RJ, Fu FH, Ilkhanipour K. Objective anterior cruciate ligament testing. Orthopedics 1986; 9 (3) 391-393
  • 2 Pugh L, Mascarenhas R, Arneja S, Chin PY, Leith JM. Current concepts in instrumented knee-laxity testing. Am J Sports Med 2009; 37 (1) 199-210
  • 3 Daniel DM, Malcom LL, Losse G, Stone ML, Sachs R, Burks R. Instrumented measurement of anterior laxity of the knee. J Bone Joint Surg Am 1985; 67 (5) 720-726
  • 4 Daniel DM, Stone ML, Sachs R, Malcom L. Instrumented measurement of anterior knee laxity in patients with acute anterior cruciate ligament disruption. Am J Sports Med 1985; 13 (6) 401-407
  • 5 Daniel DM, Stone ML, Dobson BE, Fithian DC, Rossman DJ, Kaufman KR. Fate of the ACL-injured patient. A prospective outcome study. Am J Sports Med 1994; 22 (5) 632-644
  • 6 Sernert N, Kartus J, Köhler K, Ejerhed L, Karlsson J. Evaluation of the reproducibility of the KT-1000 arthrometer. Scand J Med Sci Sports 2001; 11 (2) 120-125
  • 7 Wroble RR, Van Ginkel LA, Grood ES, Noyes FR, Shaffer BL. Repeatability of the KT-1000 arthrometer in a normal population. Am J Sports Med 1990; 18 (4) 396-399
  • 8 Wiertsema SH, van Hooff HJ, Migchelsen LA, Steultjens MP. Reliability of the KT1000 arthrometer and the Lachman test in patients with an ACL rupture. Knee 2008; 15 (2) 107-110
  • 9 Daniel DM, Stone ML. Diagnosis of knee ligament injury: tests and measurements of joint laxity. In: Feagin JA, ed. The Crucial Ligaments. 1st ed. New York: Churchill Livingstone; 1988: 287-300
  • 10 Paine RM, Johnson R. Instrumented examination. In: Ellenbecker, ed. Knee Ligament Rehabilitation, revised edition. New York: Churchill Livingstone; 2000: 40-65
  • 11 Kennedy JC, Fowler PJ. Medial and anterior instability of the knee. An anatomical and clinical study using stress machines. J Bone Joint Surg Am 1971; 53 (7) 1257-1270
  • 12 Daniel DM. Assessing the limits of knee motion. Am J Sports Med 1991; 19 (2) 139-147
  • 13 Daniel DM, Stone ML, Barnett P, Sachs R. Use of the quadriceps active test to diagnose posterior cruciate-ligament disruption and measure posterior laxity of the knee. J Bone Joint Surg Am 1988; 70 (3) 386-391