Am J Perinatol 2014; 31(12): 1105-1110
DOI: 10.1055/s-0034-1371707
Original Article
Thieme Medical Publishers 333 Seventh Avenue, New York, NY 10001, USA.

The Use of Electrical Cardiometry for Continuous Cardiac Output Monitoring in Preterm Neonates: A Validation Study

R. Song
1   Division of Neonatology, Department of Pediatrics, University of California San Diego, San Diego, California
,
W. Rich
1   Division of Neonatology, Department of Pediatrics, University of California San Diego, San Diego, California
,
J. H. Kim
1   Division of Neonatology, Department of Pediatrics, University of California San Diego, San Diego, California
,
N. N. Finer
1   Division of Neonatology, Department of Pediatrics, University of California San Diego, San Diego, California
,
A. C. Katheria
2   Neonatal Research Institute, Sharp Mary Birch Hospital for Women & Newborns, San Diego, California
› Author Affiliations
Further Information

Publication History

22 November 2013

27 January 2014

Publication Date:
28 March 2014 (online)

Abstract

Background Electrical cardiometry (EC) is a continuous noninvasive method for measuring cardiac output (CO), but there are limited data on premature infants. We evaluated the utility of EC monitoring by comparing the results obtained using EC to measurements of CO and systemic blood flow using echocardiography (ECHO).

Methods In this prospective observational study, 40 preterm neonates underwent 108-paired EC and ECHO measurements.

Results There were correlations between EC-CO and left ventricular output (LVO, p < 0.005) and right ventricular output (RVO, p < 0.005) but not with superior vena cava (r = 0.093, p = 0.177). Both RVO and LVO correlated with EC with and without a hemodynamically significant ductus arteriosus (p = 0.001 and 0.008, respectively). The level of agreement was decreased in infants ventilated by high-frequency oscillation ventilators (HFOV). The bias in HFOV was also positive compared with the negative biases found in other modes of ventilation.

Conclusion Given the correlation of EC with LVO, RVO, and lack of confounding effects of the ductus, our results suggest that EC has promise for trending CO in the preterm infant. However, given the limitations with mode of ventilation and the lack of correlation at low LVO values, further study is needed before this technology can be for routine use.

 
  • References

  • 1 Kluckow M, Seri I, Evans N. Functional echocardiography: an emerging clinical tool for the neonatologist. J Pediatr 2007; 150 (2) 125-130
  • 2 Osypka MJ, Bernstein DP. Electrophysiologic principles and theory of stroke volume determination by thoracic electrical bioimpedance. AACN Clin Issues 1999; 10 (3) 385-399
  • 3 Bernstein DP, Osypka MJ. Apparatus and method for determining an approximation for stroke volume and cardiac output of the heart. US Patent No. 6,511,438
  • 4 Osthaus WA, Huber D, Beck C , et al. Comparison of electrical velocimetry and transpulmonary thermodilution for measuring cardiac output in piglets. Paediatr Anaesth 2007; 17 (8) 749-755
  • 5 Norozi K, Beck C, Osthaus WA, Wille I, Wessel A, Bertram H. Electrical velocimetry for measuring cardiac output in children with congenital heart disease. Br J Anaesth 2008; 100 (1) 88-94
  • 6 Grollmuss O, Demontoux S, Capderou A, Serraf A, Belli E. Electrical velocimetry as a tool for measuring cardiac output in small infants after heart surgery. Intensive Care Med 2012; 38 (6) 1032-1039
  • 7 Katheria AC, Leone TA, Woelkers D, Garey DM, Rich W, Finer NN. The effects of umbilical cord milking on hemodynamics and neonatal outcomes in premature neonates. J Pediatr 2014; pii :S0022-476(14)00030–000304
  • 8 Katheria A, Leone T. Altered transitional circulation in infants of diabetic mothers with strict antenatal obstetric management: a functional echocardiography study. J Perinatol 2012; 32 (7) 508-513
  • 9 Katheria AC, Leone TA. Changes in hemodynamics after rescue surfactant administration. J Perinatol 2013; 33 (7) 525-528
  • 10 Cecconi M, Rhodes A, Poloniecki J, Della Rocca G, Grounds RM. Bench-to-bedside review: the importance of the precision of the reference technique in method comparison studies—with specific reference to the measurement of cardiac output. Crit Care 2009; 13 (1) 201
  • 11 Noori S, Drabu B, Soleymani S, Seri I. Continuous non-invasive cardiac output measurements in the neonate by electrical velocimetry: a comparison with echocardiography. Arch Dis Child Fetal Neonatal Ed 2012; 97 (5) F340-F343