Key words
stroke - thrombectomy - thrombolysis - stent retriever
Purpose
The effect of i. v. thrombolysis seems to be limited in patients with a higher clot burden [1]. Mechanical thrombectomy and large vessel recanalization using stent-based devices are increasingly used for the treatment of patients with acute embolic stroke. Stent retrievers can achieve fast recanalization even with large thrombi. Thus it is conceivable that patients with small thrombi can be treated efficiently with systemic thrombolysis whereas a high clot burden requires thrombectomy. Non-contrast thin-slice CT is able to detect thrombus location and size in patients with acute intracranial vessel occlusion sufficiently [2] and is readily available in the routine clinical diagnostic work-up of acute stroke patients.
We retrospectively compared the clinical outcome in patients, who received thrombolytic therapy only with patients who underwent neurothrombectomy with or without additional thrombolysis. Patients in both groups had an acute anterior circulation occlusion and a thrombus length of at least 8 mm. This thrombus length has been recently proposed as the threshold for treatment allocation to systemic thrombolysis or neurothrombectomy because of a probability below 1 % for successful recanalization by i. v. thrombolysis with thrombi above this size [1]. The aim of the study was to test if neurothrombectomy with the Trevo® device is beneficial compared to standard stroke treatment in high clot burden stroke with a thrombus length of ≥ 8 mm.
Methods
Patient selection
The predefined selection criteria for the retrospective inclusion of patients with intraarterial thrombectomy were: Occlusion of a large vessel within the anterior circulation by a thrombus with a length of ≥ 8 mm, NIHSS ≥ 4, age of ≥ 18 and ≤ 85. Thrombectomy had to be started within 8 hours after symptom onset. Patients were excluded if they had had a previous stroke within 30 days, seizure at the onset of stroke, signs of intracranial hemorrhage, aneurysm or early ischemic changes larger than one third of the middle cerebral artery territory.
40 consecutive patients (median age 71, range 20 – 83; 27 females) fulfilling the inclusion criteria were treated with neurothrombectomy using the Trevo® device in addition to standard stroke treatment between January 2011 and March 2012 and were included in the study group ([Table 1]). The median National Institutes of Health Stroke Scale (NIHSS) on admission was 14 (12 – 20). The duration of symptoms on admission was 148 min (58 – 268). We compared to 42 patients (median age 72.5, range 35 – 86; 21 females) of a historical control group with a thrombus length of ≥ 8 mm that received i. v. thrombolysis only between July 2007 and January 2009 ([Table 1]). During this period neurothrombectomy was not implemented in our clinical routine. The National Institutes of Health Stroke Scale (NIHSS) on admission was 13 (10 – 16), and the duration of symptoms was 70 min (55 – 90). The clinical data after 90 days was accessible through a local stroke registry.
Table 1
Patient characteristics at baseline.
Tab. 1 Patientencharakteristika vor Behandlungsbeginn.
characteristics
|
neurothrombectomy
|
thrombolysis only
|
p-value
|
number of patients, n
|
40
|
42
|
|
demographic data
|
|
13:27 (33 vs. 68 %)
|
21:21 (50 vs. 50 %)
|
|
|
71 (54 – 76)
|
72.5 (66 – 80)
|
0.15
|
clinical deficit before treatment
|
|
5 (4 – 5)
|
5 (4 – 5)
|
0.21
|
|
14 (12 – 20)
|
13 (10 – 16)
|
0.05
|
site of occlusion
|
|
16/24
|
20/22
|
|
|
13
|
10
|
|
|
26
|
30
|
|
|
1
|
2
|
|
thrombus length1, mm
|
13 (10 – 20)
|
12.4 (9 – 17)
|
0.22
|
i. v. thrombolysis
|
|
15 (37.5 %)
|
42 (100 %)
|
|
|
118 (98 – 151)
|
133 (104 – 166)
|
0.89
|
|
148 (58 – 259)
|
70 (55 – 90)
|
< 0.01
|
M1 / M2, M1 / M2 segment of middle cerebral artery; ICA-T carotid T; IVT i. v. thrombolysis. Time from onset to i. v. treatment: Patients were allocated from external hospitals and partially received i. v. thrombolysis before admission at our site.
1 Data are presented as median (interquartile range).
Imaging and thrombus length measurement
In order to access clot burden, the length of the hyperdense arterial segment was measured on multiplanar reformatted thick-slab maximum intensity projections of 0.625 mm thin NECT slices [1] aligned with the thrombus axis. All images were obtained using a 64-detector row CT scanner (Brilliance 64; Philips, Best, The Netherlands) with a tube voltage of 120 kV and a tube current of 320 mAs, an incremental scanning mode and a smooth reconstruction kernel. Measurements were done by an experienced neuroradiologist blinded to the patient data using the EMPR 3 D plugin postprocessing capability of the IMPAX® EE PACS workstation (AGFA®Health Care N.V., Mortsel, Belgium). With this plugin, a 3 D curved line was drawn along the HMCAS for measurement of the thrombus length in each patient. Occlusion of the artery was then verified by contrast-enhanced CT angiography.
Acute stroke management
Patients eligible for i. v. thrombolysis (n = 15 in the thrombectomy group, all in the control group) were treated with intravenous rtPA (0.9 mg/kg, max. 90mg; Actilyse®; Boehringer Ingelheim, Germany) within 4.5 hours after symptom onset according to the European guidelines for stroke treatment [3]. Patients eligible for thrombectomy were treated using a bridging concept with i. v. thrombolysis followed by the angiographic procedure which was initiated within 8 hours after symptom onset. 25 patients were treated exclusively with mechanical thrombectomy. Intravenous GP2b3a antagonists were given in 4 patients due to additional stenting of internal carotid artery stenosis and due to stenting of an MCA stenosis in one case.
Mechanical thrombectomy
Prior to mechanical thrombectomy, patients were evaluated by non-contrast-enhanced incremental CCT as described above followed by CT perfusion and CT angiography. All patients received general anesthesia. A flexible 6F sheath (Super Arrow-Flex®; Arrow International, Inc.; Reading, PA, USA) was placed into the proximal internal carotid artery via a transfemoral approach. In all patients, a distal guiding/aspiration catheter (either DAC® 057, iDAC® 070 (Concentric Medical, Inc.; Mountain View, CA, USA) or ReFlex® A+ 058 (Reverse Medical Corporation; Irvine, CA, USA)) was used for clot aspiration. Using a 2.7F 0.021” microcatheter (Trevo® Pro 18; Concentric Medical Inc.; Mountain View, CA, USA), the Trevo® stent retriever was released penetrating the thrombus and was left there for three to five minutes. Clot retrieval was combined with aspiration through a 20 ccm syringe. The procedure was repeated when necessary and the TICI score was recorded prior to and after thrombectomy.
Treatment evaluation
Successful vessel recanalization by thrombectomy was defined by a recanalization rate of TICI 2b or 3 confirmed by angiography. The clinical status was scored using the NIHSS and modified Rankin Scale (mRS) on admission, mRS at discharge from hospital, and on day 90. The latter was accessible through a regional prospective stroke registry, in which the data was acquired by a questionnaire. Control CT scans were used to determine infarct location and occurrence of hemorrhage. A symptomatic hemorrhage was defined as intracranial bleeding with a neurological deterioration of ≥ 4 points in NIHSS.
Statistics
Data are presented as median and interquartile range. For group data comparison the Mann-Whitney U-test was used. P-values < 0.05 were regarded as statistically significant. Pearson correlation was used to assess correlations between time to revascularization and clinical outcome (mRS). χ²-test was used to assess significant differences in frequencies.
Results
Demographic and clinical baseline data
The demographic and clinical data are shown in [Table 1]. Groups (thrombectomy vs. i. v. thrombolysis) differed in NIH-SS on admission (14 [12]
[13]
[14]
[15]
[16]
[17]
[18]
[19]
[20] vs. 13 [10]
[11]
[12]
[13]
[14]
[15]
[16]; p = 0.05), but did not differ in age, mRS on admission ([Fig. 1]) and thrombus length. The groups differed significantly in the time from symptom onset until the time of admission (148 min [58 – 259] vs. 70 min [55 – 90]; p = 0.001), because some patients for thrombectomy were referred from other hospitals. There was no difference concerning the time from symptom onset to i. v. thrombolytic therapy (118 min (98 – 151) vs. 133 min (104 – 166)).
Fig. 1 MRS on admission in the thrombectomy (Trevo) and control group (Control).
Abb. 1 MRS bei Aufnahme in der Thrombektomie- (Trevo) und Kontrollgruppe (Control).
Angiographic outcome and safety
The mean duration from admission to the start of angiographic interventions was 101 min (81 – 144). Successful recanalization was achieved in 33 patients (83 %). The time from symptom onset to recanalization was 309 min (248 – 395) ([Table 2]). The following procedure-related complications occurred: In one patient with M1 occlusion receiving i. v. thrombolysis prior to mechanical recanalization, a symptomatic intracranial hemorrhage occurred within 24 hours. The patient reached good recovery (mRS 2) after 90 days. In another patient embolization of the initially unaffected anterior cerebral artery occurred during successful recanalization of an M1 occlusion. In one patient the advancing microcatheter perforated the target vessel distal to an M1 thrombus. The procedure was terminated and the patient remained asymptomatic for the resulting subarachnoid hemorrhage.
Table 2
Interventional data.
Tab. 2 Daten zur interventionellen Therapie.
characteristics
|
neurothrombectomy
|
thrombolysis only
|
recanalization
|
[1]
|
[2]
|
|
33 (83 %)
|
0 (0 %)
|
|
20
|
n.a.
|
|
13
|
n. a.
|
|
4
|
n. a.
|
|
3
|
n. a.
|
duration to recanalization
|
|
101 (81 – 144)
|
n. a.
|
|
309 (248 – 395)
|
n. a.
|
|
61 (50 – 103)
|
n. a.
|
stenting
|
ICA, n
|
6
|
n. a.
|
MCA, n
|
1
|
n. a.
|
n. a.: not applicable.
1 Controlled during intervention;
2 Controlled with CT or TCD 24 hrs after i. v. rTPA.
3 Data are presented as median (interquartile range);
Clinical outcome
The median mRS at discharge was significantly better in the thrombectomy group compared to the control group (3.5 [1.25 – 5] vs. 5 [4]
[5]
[6], p < 0.001) and remained better on day 90 (3 [1]
[2]
[3]
[4] vs. 5 [4]
[5]
[6], p < 0.001) ([Table 3, ]
[Fig. 2], [3]). There was no improvement of mRS from admission to discharge in 7 (18 %) thrombectomy patients and 24 (57 %) patients of the control group (p < 0.01). On day 90, 17 of the 40 patients (43 %) having been treated with thrombectomy presented a slight disability (mRS ≤ 2), but only 2 of the 42 patients (5 %) in the control group. The time to revascularization did not correlate with the clinical improvement assessed with the mRS (r = -1.9; p = 0.1). After thrombectomy three patients died within 90 days (8 %), while 17 patients (41 %) died in the control group. Symptomatic intracerebral hemorrhage occurred in 3 (8 %) vs. 7 (17 %) patients. 4 vs. 0 patients showed mild non-symptomatic subarachnoid hemorrhage (p < 0.05). In the thrombectomy group 3 patients with hemorrhage and 1 patient with subarachnoidal hemorrhage received thrombolytic therapy within the bridging regime.
Table 3
Clinical outcome and safety data.
Tab. 3 Klinisches Ergebnis und Sicherheitsdaten.
characteristics
|
neurothrombectomy
|
thrombolysis only
|
p-value
|
mRS at discharge[1]
|
3.5 (1.25 – 5)
|
5 (4 – 6)
|
< 0.01
|
mRS after 90 days1
|
3 (1 – 4)
|
5 (4 – 6)
|
< 0.001
|
mRS≤ 2 at 90 days, n (%)
|
17 (43 %)
|
2 (5 %)
|
< 0.001
|
death in hospital, n (%)
|
1 (2.5 %)
|
11 (26 %)
|
< 0.02
|
death in 90 days, n (%)
|
3 (8 %)
|
17 (41 %)
|
< 0.02
|
no clinical improvement, n (%)
|
7 (18 %)
|
24 (57 %)
|
< 0.01
|
symptomatic ICH, n (%)
|
3 (8 %)
|
7 (17 %)
|
0.21
|
non-symptomatic SAH, n (%)
|
4 (10 %)
|
0 (0 %)
|
< 0.05
|
1 Data are presented as median (interquartile range)
Fig. 2 Clinical outcome (mRS) at discharge in the thrombectomy (Trevo) and control group (Control).
Abb. 2 Klinisches Ergebnis (mRS) bei Entlassung in der Thrombektomie- (Trevo) und Kontrollgruppe (Control).
Fig. 3 Clinical outcome (mRS) 90 days after stroke in the thrombectomy (Trevo) and control group (Control).
Abb. 3 Klinisches Ergebnis (mRS) 90 Tage nach Schlaganfall in der Thrombektomie- (Trevo) und Kontrollgruppe (Control).
Discussion
In this study we compared mechanical neurothrombectomy in patients with a high clot burden to a control group treated with i. v. rtPA thrombolysis alone. Clot burden is a major factor for clinical outcome in acute embolic stroke [4] and accordingly thrombus length is a main factor for recanalization after i. v. thrombolysis [1]. Assessment of thrombus length is easily implemented in routine stroke CT scanning without relevant further time consumption. Furthermore, the success of recanalization is a critical factor for clinical outcome after acute embolic stroke. In patients with a thrombus length of more than 8 mm, i. v. thrombolysis achieved recanalization in less than 1 % [1] and usually results in a poor clinical outcome. Accordingly, only 5 % of the patients in our control group reached the favorable outcome of mRS≤ 2. In this study we found a much better outcome of the mechanical thrombectomy group with regard to survival and disability at 90 days and an acceptable risk profile of the intervention.
Our data show that mechanical thrombectomy using the Trevo® device in acute stroke of the anterior circulation is highly effective. The intervention using the Trevo® device showed a remarkably high rate of successful recanalization (TICI 2b or 3) in 83 %. This exceeds the recanalization rates of the Merci trials (69.5 %, TIMI II-III) [5], (44 %)[6], the IMS-III trial (41 %)[7], the MR RESCUE trial (27 %)[8] and the recently published MR CLEAN trial (58.7 %)[9], which was the first positive RCT on intraarterial treatment against standard stroke treatment. Comparable results have been published by other retrospective trials using the Penumbra system (81.6 %, TIMI II-III) [10], the Solitaire device (90 %, 79 %) [11]
[12], the Revive® device (100 %, n = 10) [13], the Trevo® device (68 %, 39 %, 78.3 %) in three trials [6]
[14]
[15] and using the Trevo® or Solitaire® device (83.9 %, TICI 2a or 3) in the case of Pagola et al. [16]. Thus, there is good evidence that mechanical thrombectomy using stent retrievers achieves high rates of successful recanalization. However, none of these previous studies assessed thrombus length.
The clinical outcome was also favorable after thrombectomy. 43 % of the patients lead an independent life (mRS≤ 2) compared to 5 % in the control group despite the more severe stroke on admission in the thrombectomy group. In our study mortality seems to be reduced by thrombectomy as well, as indicated by a decrease from 41 % to 8 %. Previous studies compared intraarterial thrombectomy to standard stroke treatment in acute stroke patients. Three randomized controlled trials failed to show clinical benefit by intraarterial treatment [7]
[8]
[17]. The inclusion criteria are discussed as a reason for this failure, but the main factor seems to be the low recanalization rate of 27 % to 41 % in the IMS-III and MR RESCUE trial. This may be due to the infrequent use of stent retrievers or the usage of old recanalization devices in these studies. For example, in the IMS-III trial, only 4 patients were treated with stent retriever devices. Furthermore, in the SYNTHESIS trial and IMS-III trial, no demonstration of large artery occlusion was required. Therefore, the cohorts of these studies are not comparable to the patients analyzed in our study. The randomized controlled MR CLEAN trial [9] very recently proved a clinical benefit of intraarterial treatment compared to standard stroke treatment. Occlusion of the distal internal carotid artery, the middle cerebral artery (M1, M2) or anterior cerebral artery (A1, A2) was required for inclusion and thrombectomy was carried out using a stent retriever in 81.5 % of the cases. Thus, the findings of the MR CLEAN trial are in accordance with the results of our study, but thrombus length was not assessed in that trial. However, the clinical and procedural outcome in our study even exceeds the results of the MR CLEAN trial and, in contrast, significant reduction of mortality could be observed in our study. Thus, patient identification by thrombus length, not only the site of occlusion, and solely the use of highly effective stent retrievers seem to be important corner points for successful intraarterial treatment. Once more, registry studies in acute stroke using stent-based thrombectomy also revealed an increase in favorable outcome after 90 days: Leker et al. found up to 60 % of patients treated by thrombectomy reaching an mRS of 0 to 2 compared to 37.5 % in patients treated by i. v. thrombolysis [18]. In a recent study a 90-day favorable outcome of even 77 % and a recanalization rate of 90 % could be achieved using the Solitaire® device [19]. Thus, the application of different thrombectomy devices may be the reason for the controversial findings since stent retriever devices showed higher recanalization rates and higher clinical improvement in comparison with other devices [6]
[11]
[14]
[19]
[20]
[21]
[22]
[23]. We conclude that the significantly higher rate of favorable outcome in patients treated with neurothrombectomy in comparison to patients treated with i. v. thrombolysis alone is caused both by the usage of a stent retriever device for mechanical thrombectomy and patient selection based on the criterion of thrombus length. This might be a very robust criterion indicating neurothrombectomy rather than site of occlusion only.
We had three serious complications among 40 patients which finally led to death in one, poor outcome in another and a favorable outcome in another. We consider this an acceptable risk because of the high mortality in this cohort of stroke patients. However, the complication rate was even lower in the neurothrombectomy group compared to i. v. thrombolysis only. In addition, the incidence of symptomatic intracranial hemorrhages after thrombectomy is not higher than with systemic thrombolysis alone. A higher rate of non-symptomatic subarachnoid hemorrhages in the thrombectomy group might be consistent with reperfusion injury after large vessel recanalization. However, the complication rate seems to be lower than with other thrombectomy devices but head-to-head studies are needed. Furthermore, the four randomized controlled trials [7]
[8]
[9]
[17] on intraarterial treatment found no increase in severe adverse events even though three of them did not show superiority of intraarterial treatment. Thus, mechanical thrombectomy seems to be safe in patients with and without previous intravenous thrombolysis.
Conclusion
Patient selection for neurothrombectomy by thrombus length in acute stroke patients and implementation of stent retriever devices for neurothrombectomy provided a clinically highly effective and technically safe method for successful recanalization with an acceptable risk while patients with a comparable clot burden treated with i. v. thrombolysis only had a low probability of reaching a favorable clinical outcome and showed a high rate of mortality. Our study had an even better result than the MR CLEAN study but it was retrospective and open.
Nevertheless, a clot length of at least 8 mm seems to be an additional valuable criterion for indicating neurothrombectomy in acute stroke patients and our results may guide future randomized controlled trials with comparison of neurothrombectomy with systemic thrombolysis only in patients with a thrombus length of more than 7 mm, since recently published controlled trails have not implemented such a criterion. However, further controlled clinical trials on clinical efficacy and safety are warranted and, like the THERAPY trial [NCT01 429 350], are on the way.
Clinical relevance of the study
-
Increasing evidence supports neurothrombectomy in addition to standard stroke treatment in acute embolic stroke.
-
Patients with severe acute stroke syndromes should be carefully evaluated regarding high clot burden and should be considered for neurothrombectomy.
-
Not only site of occlusion but also thrombus length should be taken into account when thrombectomy is considered.
-
Thrombectomy in patients with a high clot burden using the Trevo® stent retriever seems to be safe and effective.