Nuklearmedizin 1997; 36(06): 202-204
DOI: 10.1055/s-0038-1629740
Originalarbeiten — Original Articles
Schattauer GmbH

Wie hoch ist die Strahlenexposition in der Umgebung der Patienten nach Applikation von 99mTc-Sestamibi?

Radiation Exposure near of Patients after Administration of 99mTc-Sestamibi
A. Kurtaran
1   Aus der Universitätsklinik für Nuklearmedizin Wien, Österreich
,
J. Preitfellner
1   Aus der Universitätsklinik für Nuklearmedizin Wien, Österreich
,
P. Schaffarich
2   Institut für Biomedizinische Technik und Physik Wien, Österreich
,
Ch. Scheuba
3   Chirurgie, Wien, Österreich
,
B. Niederle
3   Chirurgie, Wien, Österreich
,
I. Virgolini
1   Aus der Universitätsklinik für Nuklearmedizin Wien, Österreich
,
E. Havlik
2   Institut für Biomedizinische Technik und Physik Wien, Österreich
4   L. Boltzmann-Institut für Nuklearmedizin, Wien, Österreich
› Author Affiliations
Further Information

Publication History

Eingegangen: 21 February 1997

angenommen: 11 March 1997

Publication Date:
03 February 2018 (online)

Zusammenfassung

Ziel der vorliegenden Arbeit war es, die zusätzliche Strahlenexposition von Pflegepersonal, Nachbarpatienten und Familienangehörigen in der Nähe von Patienten nach Applikation von 99mTc-Sestamibi (Cardiolite® DuPONT PHARMA) abzuschätzen. Die 99mTc-Sestamibi-Szintigraphie wurde zum Zweck der präoperativen Lokalisation von Nebenschilddrüsenadenomen durchgeführt. Methode: 10 min, 180-200 min und 24 h nach Applikation von 600 ± 50 MBq 99mTc-Sestamibi wurden bei 16 Patienten in 0,5, 1 und 2 m Entfernung die Dosisleistungen mit einem Strahlenmeßgerät (Berthold LB 133®) gemessen. Aufgrund dieser Messungen wurde die mögliche Strahlenexposition anderer Personen in der Umgebung der Patienten abgeschätzt. Ergebnisse: Die Dosisleistungen in der Nähe der Patienten betrugen kurz nach der Applikation 20,5 μSv/h in 0,5 m Entfernung, 5,25 μSv/h in 1 m Entfernung und 1,55 μSv/h in 2 m Entfernung. Die biologische Halbwertszeit wurde mit 54 h ermittelt. Für Personen in der Wartezone wurde die maximale Dosis mit 45,4 μSv ermittelt, für Personen im gemeinsamen Haushalt mit 27,6 (jlSv bzw. für das Pflegepersonal auf einer Bettenstation mit 31,2 μSv Schlußfolgerung: Die zusätzliche Strahlenexposition von anderen Patienten, Personal und Begleitpersonen durch »strahlende« Patienten ist auch unter ungünstigsten Annahmen unbedenklich und liegen weit unter der zulässigen Dosis für nicht beruflich strahlenexponierte Personen.

Summary

Aim: The aim of the present study was to estimate the additional radiation exposure to personnel, other patients and members of the family caused by patients who had been injected with 99mTc-Sestamibi (Cardiolite® DuPONT PHARMA) for preoperative localization of parathyroid adenoma. Methods: Dose rates were measured from 16 patients who had received an intravenous injection of 600 ± 50 MBq 99mTc-Sestamibi. All measurements were performed with a portable dosimeter (Berthold LB 133®) at 3 different distances (0.5, 1 and 2 m) at 10 min, 180-200 min and 24 h after administration of the tracer. Results: The dose rates amounted to 20.5 μSv/h at 0.5 m, 5.25 μSv/h at 1 m and 1.55 μSv/h at 2 m distance from patients respectively. The biological half-life was 54 h. The calculated maximal dose to other persons in the waiting area was 31.2 μSv, to family members 27.6 μSv and to nurses in a ward 31.2 μSv. Conclusion: Our results indicate that the calculated maximal radiation exposure for personnel, family members and other patients even under very unfavourable conditions was below the maximal allowed dose for non-professionally exposed persons.

 
  • LITERATUR

  • 1 Diaz M, Bockisch A, Hahn K. et al. Szintigra-phische Darstellung ektoper Nebenschildrü-senadenome mit 99mTc-MIBI. 2 Fallbeispiele. Nuklearmedizin 1994; 33: 42-5.
  • 2 Fogelman I, McKillop JH, Bessent RG. et al. Successful localization of parathyroid adenomata by Thallium-201 and Technetium-99m subtraction scintigraphy: description of an improved technique. Eur J Nucl Med 1984; 9: 545-7.
  • 3 Ferlin G, Borsato N. Camerani et al. New perspective in localizing enlarged parathyroidy by technetium-thallium subtraction scan. J Nucl Med 1983; 24: 438-41.
  • 4 Coakley AJ, Kettle AG, Wells CP. et al. 99mTc-Sestamibi, a new agent for parathyroid imaging. Nucl Med Comm 1989; 10: 791-4.
  • 5 Bares R, Müller B, Büll U. et al. The radiation dose to surgical personnel during intraoperative radioimmunoscintimetry. Eur J Nucl Med 1992; 19: 110-2.
  • 6 Kurtaran A, Preitfellner J, Kohoutek D. et al. Strahlenexposition in der Umgebung von Patienten nach 201T1-Myokardszintigraphie. Nuklearmedizin 1997; 36: 29-31.
  • 7 Preitfellner J, Kurtaran A, Havlik E. Strahlenexposition in der Umgebung von Patienten nach Gabe von 99mTc-DPD. Nuklearmedizin 1995; 34: 151-5.
  • 8 Harding LK, Bossuyt A, Pellet S. et al. Radiation doses to those accompanying nuclear medicine department patients: a waiting room survey. Eur J Nucl Med 1994; 21: 1223-6.