Keywords
complication - endoscopic surgical procedure - minimal invasive - procedure - neuroendoscopy
- training program
Introduction
The indications for endoscopic surgeries have increased for various cranial and spinal
problems such as atlantoaxial dislocation,[1] lumbar disk,[2]
[3]
[4] colloid cyst,[5] cerebellopontine angle epidermoid,[6] Arnold-Chiari malformation,[7] craniopharyngioma,[8] hydrocephalus,[9] deep-seated brain tumors,[10] trigeminal neuralgia and hemifacial spasm,[11]
[12] and arachnoid cyst.[13] Endoscopic techniques have many advantages such as good visualization, a panoramic
view, and minimal invasiveness leading to less pain and early return to work. Although
microendoscopic spine techniques have many advantages, there are also some obstacles
that should be recognized and avoided for developing good endoscopic skills and reducing
complications.[14]
[15]
[16]
[17]
[18]
This article discusses the practical aspects of microendoscopic techniques and reviews
ways to avoid complications in spine surgeries. We also suggest how to manage complications
based on the personal experience of 1,574 endoscopic spinal procedures ([Table 1]) performed by the senior author and a review of relevant literature. Topic search
was conducted on PubMed, Google Scholar, and Medline using terms “neuroendoscopy,”
“complications and neuroendoscopy,” “neuroendoscopy and complication avoidance,” “complication
avoidance and endoscopy,” “endoscopic neurosurgery,” and “minimally invasive neurosurgery.”
Full texts of the relevant articles were analyzed after study of the abstracts and/or
topics.
Table 1
Surgical spinal procedures and related complications or problems
|
Endoscopic spine surgeries
|
Complications or problems during surgery, no. of patients (%)
|
|
Lumbar disk (
n
= 970)
|
Minor dural punctures: 12 (1.2)
Facet injury: 5 (0.5)
Postoperative diskitis: 5 (0.5)
Recurrence: 4 (0.4)
Persistent paresthesia: 3 (0.3)
Root prolapse: 2 (0.2)
Root injury: 2 (0.2)
Conversion to open surgery: 2 (0.2)
|
|
Lumbar canal stenosis
(
n
= 87)
|
Minor dural punctures: 5 (5.7)
Persistent paresthesia: 3 (3.4)
Facet injury: 2 (2.3)
Root prolapse: 1 (1.1)
Postoperative diskitis: 1 (1.1)
Root injury: 1 (1.1)
|
|
Intradural extramedullary spine tumor
(
n
= 42)
|
CSF leak: 1 (2.4)
Difficulty in dural repair: all patients (100)
|
|
Anterior partial corpectomy for cervical myelopathy
(
n
= 27)
|
Bleeding from epidural vessels: 7 (25.6)
Blind area and difficulty in bimanual dissection in limited space: 27 (100)
|
|
Transcervical approach for atlantoaxial dislocation
(
n
= 23)
|
Difficulty due to oblique angle: 3 (13.0)
|
|
Transoral approach for atlantoaxial dislocation
(
n
= 68)
|
CSF leak: 1 (1.5)
|
|
Anterior approach for cervical disk
(
n
= 222)
|
Minor bleeding from epidural vessels: 16 (7.2)
Transient difficulty in swallowing: 2 (0.9)
Incomplete decompression: 2 (0.9)
Esophageal injury: 1 (0.5)
Temporary C5 root paresis: 1 (0.5)
Temporary hoarseness of voice: 1 (0.5)
|
|
Posterior cervical approach for multilevel compression
(
n
= 91)
|
Minor bleedings from epidural vessels: 5 (5.5)
Minor dural tear: 2 (5.5)
C5 root injury: 2 (5.5)
|
|
Arnold-Chiari malformation withor without syringomyelia
(
n
= 31)
|
CSF leak: 1 (3.2)
Hydrocephalus: 1 (3.2)
|
|
Posterior cervical approach for facet hypertrophy
(
n
= 13)
|
Minor bleeding: 2 (15.4)
|
Abbreviation: CSF, cerebrospinal fluid.
General Principles in Microendoscopic Spine Surgery
General Principles in Microendoscopic Spine Surgery
The abdomen should be free to avoid venous bleeding. Light cable, suction tube, and
camera cable should be properly tied so they do not come into the field during the
surgery.[16] The endoscope can be damaged due to poor handling during surgery or by a running
drill that can be avoided by shielding it with a sheath. The endoscope should be held
from the eyepiece side. The endoscope can be damaged by lifting it from the tip side,
especially when it is attached to a camera. The endoscope should be stationed at the
corner and as far away as possible from the surgical target to avoid an obstruction
in instrument manipulation. The endoscope should be retracted when an instrument is
introduced to avoid accidental injury to the structures.[17] The side rail of the table, where the endoscope holder is attached, should not be
loose.[17] Instruments should fit properly in the working channel of the endoscopic system.
The tissue to be removed should be moved to the sides (cranial, caudal, medial, or
lateral).[17] It should not be pulled toward the endoscope tip to avoid soiling of the lens tip
and blinding of vital structures (such as nerve, vessel, dura mater, etc.) in the
bed of the tissue.[17]
Proper Endoscopic Instruments
Proper Endoscopic Instruments
There should be a proper selection of instruments. A slender instrument with a single
limb is better because it occupies less space.[15]
[17]
[18] Proper functioning of the instruments should be checked (e.g., scissor should open
smoothly without any jerking) before surgery.[15]
[17] Instrument tip should be slightly curved to allow better visualization.[15]
[17]
[18] Round shaft instruments are preferred rather than flat ones.[17]
[18] Instruments with a precision grip are preferred compared with a power grip for better
control in microendoscopic surgery.[16]
[17]
[18] Straight instruments are preferred to a bayonet shape.[17]
[18] The proper length of an instrument (preferably as short a length as possible) is
better for precision.[16] Dual-use instruments that, for example, can perform suction and coagulation, or
suction and irrigation, or drilling and irrigation simultaneously can be very helpful
in difficult situations like working in a narrow space.[17]
[18]
Bimanual Dissection
Bimanual surgical techniques are better than one-handed procedures. Bimanual dissection
makes tissue dissection, retraction, drilling, achieving hemostasis, and cutting easier.
These bimanual techniques are hard to achieve in some situations such as when the
surgeon is holding the endoscope or when there is only one working channel in the
endoscopic set.[15]
[17]
[18] The surgeon can use an endoscope holder or ask the assistant to hold the endoscope
to circumvent this limitation.
Proper Adjustment of the Operative Table Height and Use of a Platform
Proper Adjustment of the Operative Table Height and Use of a Platform
Although height adjustment is possible in most tables, use of a platform can be complementary.[16]
[17]
[18] Proper height adjustment allows surgery to be performed without shoulder abduction
that can cause fatigue and physiologic tremors during surgery.[16]
[17]
[18] Use of a platform or a lower position of the table also allows observation of the
operative area and permits introduction of the instrument in the blind area when an
endoscope holder is used.[17]
Differences between Endoscopic and Microsurgical Techniques
Differences between Endoscopic and Microsurgical Techniques
Endoscopy has advantages such as better illumination, small incision, and improved
visualization, especially in the corners; however, there are some difficulties associated
with it compared with microscopy.[15] Straight instruments are preferred in endoscopic technique compared with the bayonet
shape in a microscopic one.[17] There may be an obstruction in instrument manipulation when the endoscope is placed
at the center. In such situations, the surgical field is kept in the corner in the
endoscopic procedure, whereas it is in the center in microscopic surgery.[17] There is a need to create extra space to station an endoscope in an already limited
surgical space in microendoscopic surgery, whereas in microscopic surgery, the whole
space is available for manipulation of the instrument.[17] Other limitations of endoscopic surgery compared with microscopic ones include a
blind area (space proximal to lens tip) and a possibility of disorientation due to
accidental rotation of the camera.[15]
[16]
[17] The surgeon should learn to overcome these difficulties for a better clinical outcome
in endoscopic procedures.
Magnification
High magnification is preferred in most endoscopic surgeries for better visualization.[16]
[17] Lower magnification is required for an overview of anatomy and orientation.[16] This can be achieved by using the zoom function or by moving the scope away from
the area of interest. The anatomy should be identified first by inspecting the surrounding
structures (medial, lateral, anterior, and posterior relations) before starting the
procedure. Proper knowledge of anatomy is very important because most of the time
only a small portion of a structure is seen.[15] Neuronavigation facility[15]
[16] could be useful, especially when the landmarks are not properly visualized, such
as in a repeat surgery.
Orientation and Position of the Camera
Orientation and Position of the Camera
Camera orientation should be checked by anterior, posterior, and side-to-side movements
before the procedure.[15]
[16] Camera head (buttons) should be toward the monitor so the image orientation is the
same as seen in open surgery.[17] The camera can rotate during surgery causing disorientation and should be checked
frequently for proper orientation throughout the surgery.
Straight versus Triangular Arrangement
Proper Planning of the Size and Site of an Incision
Proper Planning of the Size and Site of an Incision
Proper size of the incision should be planned at the beginning of the surgery.[17] A large incision has a disadvantage in spine endoscopic surgery, especially when
performed without an endoscope holder.[17] Proper size of an incision keeps the system snugly fit and stable even when it is
unsupported, and it also helps reduce oozing by the tamponade effect of the tubular
retractor.[17] The site of an incision should be planned properly.[17] The trajectory should be in such a way that the angle of approach is perpendicular
to the surgical target. If there is excessive angulation, soft tissues tend to protrude
inside the retractor causing difficulty in surgery. The second option (microscope)
should be ready if it is difficult to perform endoscopic surgery.[17]
Proper Position of the Endoscope
Proper Position of the Endoscope
Proper position of the endoscope is essential to prevent obstruction in the instrument
movements.[15] The endoscope needs to be placed as far away as possible and at the corner of the
operative field to prevent obstruction in instrument handling.[15]
[17] It is desirable to make a triangular arrangement between the scope, the surgical
target, and the instrument, as explained earlier.[15] When a straight arrangement cannot be avoided (in a narrow and deep space) ([Figs. 4] and [5]), such as in partial corpectomy for cervical myelopathy,[19] the distal instrument is not visualized. In such a situation the distally intended
instrument (suction in the top line of [Fig. 5]) cannot be introduced due to nonvisualization caused by the presence of the Kerrison
punch between the endoscope and the suction. In such settings, the distal instrument
(suction) should be introduced first (middle line of [Fig. 5]), should consciously be kept at that depth in the blind area, and followed by an
introduction of the Kerrison punch in between the suction and the endoscope ([Fig. 5] the bottom line). Using this trick, one can continue to remove bone in the oozing
field even when all three instruments are in a straight line ([Fig. 5]). As another option, the endoscope can be placed in the center and the instruments
(the Kerrison punch and the suction) can be passed from either side of the scope ([Fig. 6]).
Fig. 4 Visualization at the bottom of the partial corpectomy is not possible when the endoscope
is in the first or second locations (A). It is better when an endoscope is placed
at the third or fourth positions (B). It is not possible to introduce the Kerrison
punch (K) due to nonvisualization when suction (S) is already placed between the endoscope
(T) and the punch (C). Suction (S) should first be introduced at a more distal site
in the operative field, and the Kerrison punch should be advanced after that (D).
Fig. 5 Suction (S) could not be introduced due to nonvisualization when the Kerrison punch
(K) is between the endoscope (T) and the suction (top line). In such a situation,
the suction should be introduced first and placed at a distal location under proper
visualization (middle line), followed by the introduction of the Kerrison punch between
the suction and the endoscope (bottom line). The surgeon should consciously keep the
suction at that depth in the blind area.
Fig. 6 The distal instrument (Kerrison punch [K]) will not be visualized when suction (S)
is placed between the punch and the endoscope (top line). The endoscope (T) should
be placed in the center so that both the Kerrison punch and suction can be well visualized
(bottom line).
Hand Support
Hands should be properly supported during surgery for good precision.[15] An unsupported hand can increase physiologic tremors and could give rise to fatigue,
especially during a long surgery.[15]
[16]
[17] Routine hand support devices used in microsurgery may not be suitable for microendoscopic
operations. Gentle hand support on the endoscopic sheath or surrounding structures
can be a great help.[15]
[17]
Use of Precision Grip and Avoidance of Power Grip
Use of Precision Grip and Avoidance of Power Grip
Precision grip is preferred over power grip in endoscopic surgery.[16]
[17] There is no hand support in a power grip, and there is involvement of long muscles
and multiple joints that results in less precision.[17] Precision grip permits hand support and use of intrinsic muscles of the thumb and
index finger that helps in controlled movements during the surgery.[16]
[17] It is reasonable to have a power grip in a Kerrison punch because the desired function
of this instrument needs power. But some instruments are poorly designed, unfortunately,
and they are too long for a power grip design, for example, fine scissors. If the
power grip is unavoidable, the precision grip should be added with the other hand.[16]
[17]
[18]
[20] A quiet hand technique should be used where the ulnar side of the hand is well supported.[18] The muscles of the arm should be quiet with the least number of motor units of thumb
and index finger in use.[18]
Poor Visualization during Microendoscopic Surgery and How to Overcome This Limitation
Poor Visualization during Microendoscopic Surgery and How to Overcome This Limitation
Poor vision in endoscopic surgery could be due to lens staining by blood, bone dust,
fluid, or any tissue ([Table 2]).[15]
[16] For example, high humidity in the air medium, a damaged endoscope, and an out-of-focus
camera can be responsible for poor image quality.[17] Povidone-iodine scrub and Savlon (GlaxoSmithKline, Mumbai, India) are very effective
anti-fogging agents. The lens tip can be cleaned by the mechanical method or by using
various commercially available lens cleaners that clean the endoscope tip quickly
without causing any interruption in the procedure. The Clearvision II system from
Karl Storz (Tuttlingen, Germany) and InstaClear lens cleaner from Olympus are a few
examples. The lens cleaner sheath attached on the endoscope can be easily autoclaved.
Saline irrigation and removal of liquid drops from the lens tip by the suction is
also an effective manual cleaning technique. An improper connection from the endoscope
to the monitor may be responsible for poor image quality.[17] The endoscope should be directed toward the area of interest. For example, if the
surgical field is toward the cranial side of the operative field, the endoscope should
be angulated cranially.[17]
Table 2
Causes of poor visualization and steps to improve it
|
Causes of poor visualization
|
How to improve visualization
|
|
Blood, bone dust, drop of fluid, and tissue staining the lens tip
|
1. Anti-fogging agents, commercially available lens cleaner, or manual irrigation
by saline and suction
2. Bone dust staining of lens can be avoided by keeping endoscope as far away as possible,
using comparatively larger size endoscope and by zooming
3. Intermittent irrigation in between the short period of drilling
|
|
Improper connection from endoscope to the monitor and damaged lens
|
Proper connection; protect endoscope by sheath
|
|
Endoscope out of focus
|
Proper focusing
|
|
Unwanted tissue in front of the endoscope lens
|
Remove or retract tissue; move endoscope
|
|
Excessive moisture content in the air medium
|
Suction of humid air
|
|
Straight arrangement
|
Triangular arrangement
|
|
Endoscope in wrong (opposite) direction
|
Direction of the endoscope toward the object
|
|
Difficulty in visualization of the instrument tip
|
Use angled-tip instrument or introduce instrument at some angle
|
|
Pulling tissue toward the endoscope
|
Move tissue to the side (cranial, caudal, medial, or lateral). Do not pull toward
the endoscope
|
|
Disorientation due to the camera rotation during the surgery
|
Keep checking the orientation during the surgery
|
|
Bleeding
|
1. Prevent bleeding
2. Avoid large and vascular lesions
3. Working in air media is better for visualization of bleeding point and for cauterization
|
The sword-fighting effect (the instrument is pointing in one direction and the endoscope
in the opposite direction) can lead to poor image quality and instrument collision
and should be avoided.[17] The straight arrangement of the endoscope, instruments, and the target tissue should
be avoided as explained earlier.[15]
Unwanted tissue in front of the endoscope lens can impair proper visualization. Slight
withdrawal of the endoscope away from the obstructive tissue, retraction or removal
of the obstructing tissue, suctioning of humid air, and proper focusing can improve
visualization.[15]
[17]
Frequent soiling of the lens tip can be partly prevented by using a comparatively
larger diameter endoscope with a longer focal length. A large-diameter endoscope with
a longer focal length (10-mm diameter endoscope with 5-cm focal length) can permit
placement of the endoscope away from the drilling site, for example in the transoral
removal of the odontoid process.[17] Intermittent irrigation in between the short periods of drilling and using lower
revolutions (∼ 50,000 RPM) can be helpful in avoiding lens soiling.[17]
Limited Space for Instrument Manipulation
Limited Space for Instrument Manipulation
Instrument manipulations can be difficult because some space is utilized by the endoscope
in an already narrow area.[18] The endoscope itself can obstruct instrument handling in microendoscopic surgery
([Table 3]).[21] Utilization of slender shaft instruments is better because such devices occupy less
space.[13]
[15]
[16] Instrument handling during surgery can be difficult when the endoscope is close
to the surgical target.[15] The endoscope and instruments should be pointed in the same direction to avoid sword
fighting, which will help use the limited space adequately.
Table 3
Various causes of instrument obstructions and steps to overcome limitations
|
Various causes of instrument obstructions
|
Steps to overcome limitations
|
|
Endoscope tip or side can obstruct: endoscope too close to the target in the anteroposterior
direction (depth)
|
Keep endoscope as far away as possible; zoom for better visualization
|
|
Endoscope in center of the field
|
Endoscope placement at corner; use angled endoscope
|
|
Sword effect (direction of endoscope and instrument in opposite side)
|
Endoscope and instruments should be angulated in same direction
|
|
Thick and straight instruments
|
Slender and angled-tip instruments
|
|
Requirement of more number of instruments such as in cases of bleeding
|
Two functions can be incorporated into one instrument such as suction and bipolar
coagulation, suction and irrigation, drilling and irrigation
|
|
Limited space
|
Use slender and single-shaft instruments
|
|
If nothing works: unable to bring instrument at the surgical target
|
Use angled-tip instruments
Withdraw endoscope, take instrument to the desired place, and then move endoscope
toward the object slowly
|
Suturing is difficult in microendoscopy due to the narrow space, and to circumvent
this, a rotation maneuver can be performed.[20] It is sometimes difficult to introduce the instrument in a limited space even after
stationing an endoscope at the corner of an operative corridor and also after keeping
the endoscope as far away as possible from the surgical target. In such a situation,
the instrument should be advanced first toward the required target, and then the endoscope
should follow in the available space.[17] The angled-tip instrument is better to reach to the extreme corner of the operative
field as compared with the straight tip tool.[17]
Endoscopic Blind Spot
Although there are advantages such as better visualization and panoramic view in endoscopic
surgery, it also has a limitation due to an inability to see the pathway proximal
to the endoscope tip.[15] There is a danger that the instruments or the side of the endoscope can damage structures
in the blind area ([Table 4]).[15] Endoscopic surgeons should train themselves to remove an endoscope with the insertion
of each new instrument.[17] This problem can be handled by visualization of the instrument directly by the naked
eye in the blind area until it appears in the endoscopic view.[13]
[15]
[21]
Table 4
Causes of endoscopic blind spot and nonvisualization
|
Causes of nonvisualization
|
Steps to overcome this limitation
|
|
Endoscopic blind spot: The pathway between the skin and the endoscope tip is not visualized
|
• Remove the endoscope with the insertion of each new instrument and follow instrument
into the field under direct endoscopic visualization
• Visualize instrument directly by the naked eye in the operative field; intermittently
look at monitor and the instrument when it is present in the blind area
• Avoid side movement, especially when endoscope has passed through a narrow opening
|
|
Distal object is not visualized when endoscope and the other objects are in a straight
line
|
• Move endoscope or object/instrument in such a way that the arrangement becomes triangular
• Keep endoscope in between the instrument and the surgical target
|
Control of Bleeding
Control of bleeding can be a challenge, especially in a fluid media in which small
amounts of blood can disturb proper visualization.[17] It is better to use the tamponade effect to control bleeding, using gentle pressure
on the hemorrhagic point by the instrument that is already there in the field and
avoiding removal of the existing tool and bringing the cautery forceps.[15] Copious irrigation and intermittent blocking of irrigation fluid allows visualization
of the bleeding point and helps achieve hemostasis.[15] Liquid media can be converted into air media (by suctioning of fluid) if it is difficult
to see a bleeding point.[17] Visualization of a bleeding point and cauterization of the vessel is comparatively
easy in air media. Application of cotton patties can stop bleeding. If nothing works,
the endoscope could be taken out, and brisk bleeding can be controlled by a microscope[20] or an exoscope, especially when a tubular retractor is used.
Prevention of Dural Tear
There is an increased risk of a dural tear and cerebrospinal fluid leak while addressing
the contralateral compression in severe spinal canal stenosis ([Table 5]).[15]
[21]
[22] Proper case selection helps prevent the dural injury, especially during the initial
learning curve, and by keeping the ligamentum flavum intact until all bony work is
finished.[22]
[23] Both 45- and 90-degree Kerrison punches should be used for removal of the cranial
and caudal lamina, respectively.[17] Using an eggshell drilling technique is preferred for lamina,[17] and then the thinned-out lamina can be removed.[21] Proper drilling technique should be used. Drilling should be parallel to or away
from the dura mater.[18] Use of a dural guard in a drill or bone shaver can avoid injury to the dura mater.
Utilization of Abgel (Sri Gopal Krishna Labs Pvt. Ltd., Mumbai, India) or bone wax
(no cotton patties) between the lamina and dura helps avoid injury to the dura mater
during the drilling.[17]
Table 5
Other difficulties along with steps to prevent them
|
Limitations
|
Steps to prevent complications
|
|
CSF leak:
More common in severe canal stenosis, large central or extruded disk, dealing with
opposite-side pathology
|
• Simple case selection in the beginning
• Keep ligamentum flavum intact until all bony work is finished
• Use 45-degree Kerrison punch for cranial and 90 degrees for caudal work
• Partially retract Kerrison punch and hold proximal part of nibbled bone or ligament
• Drill away from dura matter or parallel to it
• Dura guard, bone shaver
• Eggshell drilling technique
• Patties, Abgel, bone wax between bone and the dura
• Hold bone or ligament under proper visualization using rotation technique of the
Kerrison punch
|
|
Difficulties in drilling:
Frequent soiling of the lens by bone dust or fluid
|
• Intermittent irrigation in between the short period of drilling and using lower
revolutions
• Keeping suction near drill
• Using comparatively large-diameter endoscope with longer focal length
|
|
Causes of steep learning curve:
• Control of bleeding
• Blind area
• Unique endoscopic skills
• Two-dimensional vision
• More operative time
• Limited space control
|
• Simple case selection in the beginning of the learning curve
• Practice on models
• Multidisciplinary team approach
• Cadaveric dissection
• Live operative workshop
• Laboratory training
• Simulators
|
Stepwise surgery for removal of the lamina by the Kerrison punch should be performed.
Steps should involve (1) separation of the dura and nibbling of a small part of the
bone, (2) partial retraction of the tip of the Kerrison punch, and (3) final removal
of the nibbled piece of the bone ([Fig. 7]). But if the nibbled bone is removed in a single stroke, the dura may be caught
and get injured.
Fig. 7 Steps for removal of the lamina by the Kerrison punch. (A, B) The piece of bone is
disconnected from the rest of the lamina. The nibbled piece should be released from
the Kerrison punch. (C) Punch should be slightly withdrawn, and then the detached
piece should be held from a proximal site and removed.
Bone or ligament should be removed under good visualization with the help of a rotation
technique. Pulling of the part of the bone or ligament toward the lens tip is not
desirable because it prevents proper visualization of the underlying dura that may
have been caught in the Kerrison punch bite.[17]
Excision of the opposite-side protruded lumbar disk tissue from the contralateral
side can cause dural injury. The protruded disk fragment should be pushed in the disk
space (using a 90-degree angled instrument under the dura mater) that can then be
removed from the ipsilateral side.[17]
Dural Repair
Dural suturing in a limited area is a formidable task in endoscopic surgery. Rotation
of the needle holder is preferred over a linear movement because of the constrained
space.[20]
[22] The first knot can be preplaced at the end of the suture to avoid wasting time tying
the knot.[22] A loop for suturing can be performed inside the endoscopic field with the help of
a needle holder and the suction tip. The loop can also be made outside, which can
be carried inside the endoscopic field with the help of the needle holder.[17] Although a specialized endoscopic suturing tool such as the Covidien endo-suturing
instrument is available, it is not suitable in the present form in neurosurgery due
to its big size.
Drilling
Proper drilling techniques were mentioned earlier. A gentle paintbrush technique should
be used without any force. The drill should be removed only when it is fully stopped
to prevent injury to the vital structures and the scope.[17]
Learning Curve
Endoscopic surgeries are associated with a steep learning curve.[13]
[18]
[21] The views of endoscopic images are often different, leading to confusion when compared
with the microscopic technique, especially during the evolution from microscopic to
endoscopic procedures. The surgeon needs to acquire special endoscopic skills for
the transition from the microscopic procedure to endoscopic ones ([Table 5]).[15]
[16]
[17] The endoscopic surgeon needs to overcome many obstacles (two-dimensional visualization,
more operative time, control of bleeding, difficulty in instrument manipulation in
a limited space, problems associated with the blind area, etc.) to perform effective
endoscopic surgeries.[15]
[16]
[17]
The learning curve of the transforaminal approach is steep but easy to achieve, whereas
for the interlaminar approach it is flat but hard to master.[24] There is usually more operative time, higher technical failures, and increased recurrences
in the early part of the learning curve in the endoscopic transforaminal approach.[25] A learning curve of ∼ 72 cases was needed to reach 90% good/excellent results in
the Morgenstern et al series.[26] The surgeons' training level is an important factor for clinical success in the
endoscopic technique. Better clinical results are observed from a well-trained surgeon.[27] Jhala et al observed more technical problems and complications (durotomy or nerve
injury) during the first 25 cases of 100 patients.[28] Dural tear (n = 6 cases [12%]), facet injury (n = 2 cases [4%]), root injury (n = 1 case [2%]), and conversion to open surgery (n = 1 case [2%]) were more frequent in the initial 50 patients in the Yadav et al series,
whereas there was only one dural tear and no facet or root injury in the last 350
cases.[2]
In another study, the learning curve of microendoscopic technique in lumbar canal
stenosis was analyzed for a single surgeon.[29] Operative time reduced gradually, and the blood loss was stabilized after 30 surgeries.
However, intraoperative complications were observed even after attaining mastery of
this technique.[29] In the Kim et al study, the operative time plateaued after 20 and 19 cases for ligament
flavum and lamina resection, respectively.[30] In other research, 25 to 30 cases were needed to reach the learning curve in microendoscopic
diskectomy (MED).[31] Even during this initial learning period, MED was a safe procedure.[31]
A steady state of the learning curve with respect to operation time was achieved after
40 operations for surgeon 1 and after 16 surgeries for surgeon 2 in the Joswig et
al series.[32] Although complications did not negatively affect the long-term outcome in patients
who underwent surgery before or after the learning phase, guidance under an experienced
surgeon reduced the learning curve.[32] The higher rate of conversions, complications, recurrent lumbar disk herniations,
and a steep learning curve should be considered before a surgeon takes up endoscopic
spine surgery.[32]
The learning curve can be shortened by using a multidisciplinary team approach, simple
case selection in the beginning,[33] cadaveric dissections, practice on models, laboratory training, simulators, and
attending live operative workshops.[15]
[17] The development of a formal program and a peer-review board for endoscopic spine
surgery is advisable to achieve the minimum standard and technical skills.[34]
Complications
Overall complications were 3.6% and 13.8% in lumbar disk surgery and lumbar canal
stenosis, respectively, in our series. Complications varied from 3.6% to 21% in various
series in lumbar disk surgery.[28]
[31]
[35] Complications in the MED group (8.1%) were comparable with microdiskectomy (9.8%).[31] The rate of complications, length of hospital stay, return to daily activities,
and overall patient satisfaction were comparable in MED and conventional diskectomy
in the study by Anichini et al.[36] Overall, 21% of patients developed complications (a significant number were minor
ones) in the Jhala et al series. This included open conversion (1%) for suspected
root damage, perioperative facet damage (5%), minor dural punctures (7%), recurrence
requiring reoperation (3%), persistent paresthesia (1%), and postoperative diskitis
(4%).[28] Overall complications were 12.5% in the Kim et al series that included dural tear
(n = 3 cases [6.2%]), conversion to open surgery (n = 1 [2.1%]), and a requirement for transforaminal lumbar interbody fusion after the
procedure (n = 2 [4.2%]).[35] Surgical indication and experience of the surgeon were crucial factors in determining
the ultimate outcome.[36] We suggest a timely conversion to the microscopic procedure to prevent and treat
complications especially during the initial learning curve.
Recurrence Rates
Rate of recurrent/residual compression is the most debated topic, and the results
are extremely variable. It related to the selection of cases, level of training of
the surgeon, and publication bias in the Anichini et al series.[36] The rate of recurrence/residual was higher in MED, compared with the open technique
in the initial part of the learning curve.[36] Although the recurrence rate varied from 0.4% to 10% in various series, it was between
2% and 10% in most of the reported series.[36] There were four recurrences (0.4%) in MED in our series. In another large series,
two recurrences (0.5%) were observed of 400 treated patients.[2] There were four cases (4%) of recurrence in the Jhala et al series.[28] Of these, three patients (3%) required a reoperation.
On the contrary, a 10% recurrence was reported in 344 patients in the Matsumoto et
al series.[37] Kulkarni et al[38] and Sencer et al[39] reported 1.5% and 5% recurrences of a total of 188 and 163 patients, respectively.
Joswig et al reported a high recurrence rate of 28% in MED, which they hypothesized
was due to the younger age of the patients and longer follow-up.[32] Hsu et al reported a total of 10.5% recurrence or residual disk (recurrent disk
in 2/57 cases [3.5%]) and a residual disk in 4 of 57 cases (7%).[24] Choi et al compared transforaminal and interlaminar endoscopic diskectomy and found
the recurrence rate to be 3.3% in the former and 6.7% in the latter, although the
patient selection was different between them.[40] After attaining adequate experience, most of the microendoscopic series reported
comparable recurrence rates with the open ones.[36]
Publication Bias
Publication bias in this review may have occurred if the original literature was compromised
by biased publications.[41] Studies are more likely to be published if they have positive findings and can collect
more citations for the journal. Publication bias may occur due to poor study design,
inconsistency in reporting, and imprecision.[42]
Training in Endoscopic Surgery
Training in Endoscopic Surgery
Good endoscopic training is needed to prevent complications and for a better clinical
outcome. In the beginning, simple cases should be selected to prevent complications.[43] There is usually less opportunity for a young neurosurgeon to get trained in endoscopy
during routine operative hours.[17] The surgeon should spend more time in laboratories for training. Learning by cadaveric
dissection is best but difficult due to a shortage of cadavers. Models and simulators[44]
[45]
[46] can be good alternatives for training, but these are often very costly. Indigenous
and inexpensive models can be used for developing skills like dissection, hand-eye
coordination, cutting, and suturing in a limited space.[47]
[48] Surgical gloves, papaya, silastic tubes, capsicum, low-cost commercially available
camera and a LED light source, and a popsicle stick for lamina simulation can be used
for endoscopic training.[47] The surgeon can practice endoscopy using simple and inexpensive models.[47] A 0- and 30-degree endoscopic view can be obtained by angulating the camera for
training.[48] An exoscope system (video telescopic operating monitor) can also be used as a bridge
for learning endoscopy.[49]
Conclusion
The surgeon needs to learn endoscopic skills in addition to the microsurgical ones
to prevent complications and for good outcomes. The complications are more in the
initial part of the learning curve that can be shortened by attending live workshops,
understanding endoscopic anatomy, proper case selection, practicing on cadavers or
models, and watching operative videos and experienced surgeons. Timely conversion
to the alternative procedure is better to prevent complications, especially during
the initial learning curve.