J Knee Surg 2019; 32(08): 710-713
DOI: 10.1055/s-0039-1681030
Special Focus Section
Thieme Medical Publishers 333 Seventh Avenue, New York, NY 10001, USA.

A Cam–Post Mechanism Is No Longer Necessary in Modern Primary Total Knee Arthroplasty

R. Michael Meneghini
1   Department of Orthopaedic Surgery, Indiana University Hip & Knee Center, Indiana University School of Medicine, Fishers, Indiana
,
Michael D. Stefl
1   Department of Orthopaedic Surgery, Indiana University Hip & Knee Center, Indiana University School of Medicine, Fishers, Indiana
,
W. Andrew Hodge
2   Department of Orthopaedic Surgery, Institute for Mobility and Longevity, Phoenix, Arizona
,
Scott A. Banks
3   Department of Mechanical and Aerospace Engineering, University of Florida, Gainesville, Florida
› Author Affiliations
Further Information

Publication History

23 December 2018

10 January 2019

Publication Date:
01 March 2019 (online)

Abstract

Conclusive evidence supporting a clinical difference between posterior-stabilized (PS) and cruciate-retaining (CR) designs in total knee arthroplasty (TKA) does not currently exist, despite both designs being used for decades. Proponents of PS designs that employ a cam-and-post mechanism cite in vivo fluoroscopic data that demonstrate improved posterior rollback; however, optimal rollback has never been correlated to superior clinical outcomes. Recently, anterior-lipped and more conforming CR bearings, such as ultracongruent, medial pivot, and dual-pivot designs, have been used to substitute for the posterior cruciate ligament and obviate the need for the cam–post mechanism of a traditional PS design. Advantages of avoiding a PS TKA include eliminating the risk of box cut induced femoral condylar fracture, improving operative efficiency by removing procedural steps, removing the articulation that is a source of wear, postdeformation, breakage, or dislocation, and eliminating patellar clunk. Health care reform efforts mandate cost reduction, and procedural efficiencies and minimizing inventory through the removal of unnecessary bearing options foster that initiative.

 
  • References

  • 1 Dennis DA, Komistek RD, Colwell Jr CE. , et al. In vivo anteroposterior femorotibial translation of total knee arthroplasty: a multicenter analysis. Clin Orthop Relat Res 1998; (356) 47-57
  • 2 Cates HE, Komistek RD, Mahfouz MR, Schmidt MA, Anderle M. In vivo comparison of knee kinematics for subjects having either a posterior stabilized or cruciate retaining high-flexion total knee arthroplasty. J Arthroplasty 2008; 23 (07) 1057-1067
  • 3 Dennis DA, Komistek RD, Walker SA, Cheal EJ, Stiehl JB. Femoral condylar lift-off in vivo in total knee arthroplasty. J Bone Joint Surg Br 2001; 83 (01) 33-39
  • 4 Banks S, Bellemans J, Nozaki H, Whiteside LA, Harman M, Hodge WA. Knee motions during maximum flexion in fixed and mobile-bearing arthroplasties. Clin Orthop Relat Res 2003; (410) 131-138
  • 5 Alden KJ, Duncan WH, Trousdale RT, Pagnano MW, Haidukewych GJ. Intraoperative fracture during primary total knee arthroplasty. Clin Orthop Relat Res 2010; 468 (01) 90-95
  • 6 Scott DF, Smith RR. A prospective, randomized comparison of posterior stabilized versus cruciate-substituting total knee arthroplasty: a preliminary report with minimum 2-year results. J Arthroplasty 2014; 29 (9, Suppl) 179-181
  • 7 Lee CS, Chen WM, Kou HC, Lo WH, Chen CL. Early nontraumatic fracture of the polyethylene tibial post in a NexGen LPS-Flex posterior stabilized knee prosthesis. J Arthroplasty 2009; 24 (08) 1292.e5-1292.e9
  • 8 Lim HC, Bae JH, Hwang JH, Kim SJ, Yoon JY. Fracture of a polyethylene tibial post in a Scorpio posterior-stabilized knee prosthesis. Clin Orthop Surg 2009; 1 (02) 118-121
  • 9 Lee SC, Jung KA, Nam CH, Hwang SH, Lee WJ, Park IS. Anterior dislocation after a posterior stabilized total knee arthroplasty. J Arthroplasty 2012; 27 (02) 324.e17-324.e20
  • 10 Frye BM, Floyd MW, Pham DC, Feldman JJ, Hamlin BR. Effect of femoral component design on patellofemoral crepitance and patella clunk syndrome after posterior-stabilized total knee arthroplasty. J Arthroplasty 2012; 27 (06) 1166-1170
  • 11 Biyani RK, Ziemba-Davis M, Ireland PH, Meneghini RM. Does an anterior-lipped tibial insert adequately substitute for a post-cam articulation in total knee arthroplasty. Surg Technol Int 2017; 30: 341-345
  • 12 Parsley BS, Conditt MA, Bertolusso R, Noble PC. Posterior cruciate ligament substitution is not essential for excellent postoperative outcomes in total knee arthroplasty. J Arthroplasty 2006; 21 (06) (Suppl. 02) 127-131
  • 13 Laskin RS, Maruyama Y, Villaneuva M, Bourne R. Deep-dish congruent tibial component use in total knee arthroplasty: a randomized prospective study. Clin Orthop Relat Res 2000; (380) 36-44
  • 14 Sur YJ, Koh IJ, Park SW, Kim HJ, In Y. Condylar-stabilizing tibial inserts do not restore anteroposterior stability after total knee arthroplasty. J Arthroplasty 2015; 30 (04) 587-591
  • 15 Scott DF. Prospective randomized comparison of posterior-stabilized versus condylar-stabilized total knee arthroplasty: final report of a five-year study. J Arthroplasty 2018; 33 (05) 1384-1388
  • 16 Schmidt R, Komistek RD, Blaha JD, Penenberg BL, Maloney WJ. Fluoroscopic analyses of cruciate-retaining and medial pivot knee implants. Clin Orthop Relat Res 2003; (410) 139-147
  • 17 Samy DA, Wolfstadt JI, Vaidee I, Backstein DJ. A retrospective comparison of a medial pivot and posterior-stabilized total knee arthroplasty with respect to patient-reported and radiographic outcomes. J Arthroplasty 2018; 33 (05) 1379-1383
  • 18 Harman MK, Bonin SJ, Leslie CJ, Banks SA, Hodge WA. Total knee arthroplasty designed to accommodate the presence or absence of the posterior cruciate ligament. Adv Orthop 2014; 2014: 178156
  • 19 Abdel MP, Morrey ME, Jensen MR, Morrey BF. Increased long-term survival of posterior cruciate-retaining versus posterior cruciate-stabilizing total knee replacements. J Bone Joint Surg Am 2011; 93 (22) 2072-2078
  • 20 Vertullo CJ, Lewis PL, Lorimer M, Graves SE. The effect on long-term survivorship of surgeon preference for posterior-stabilized or minimally stabilized total knee replacement: an analysis of 63,416 prostheses from the Australian Orthopaedic Association National Joint Replacement Registry. J Bone Joint Surg Am 2017; 99 (13) 1129-1139