Keywords
zirconia implants - cantilever - finite element analysis - rehabilitation
Introduction
The choice of the material used in implants, its positioning and installation, are determining factors in the rehabilitation of the anterior maxillary region.[1]
[2] Ceramic materials such as yttrium-stabilized zirconia are an aesthetic alternative to the titanium implant[2]
[3]
[4] especially in patients with thin peri-implant gingival tissue,[4] gingival smile, or high aesthetic line, and in cases of bone resorption followed by gingival retraction and exposure of part of the implant.[3] Besides, when there is a crack propagation, the yttrium-stabilized zirconia presents a volumetric expansion, which submits the crack under compressive tension and against its progression.[4]
The presence of anatomical structures, such as nasal fossa and bone deficiencies associated with a low density of the cancellous bone, promotes a restriction in the number, position, length, and width of the implants in the anterior region of the maxilla.[5]
[6] Thus, the cantilever prosthesis is a viable option due to the possibility of preserving the bone crest necessary for the maintenance of the interdental papilla.[1] Moreover, the cantilever prostheses should be used in clinical situations where patients do not have severe occlusal forces, such as in cases of parafunctional habits and places with adequate bone density, especially in previous rehabilitation.[1]
[7]
[8] The use of cantilevered prostheses is a good alternative in places with bone defects, avoiding the need for surgery to place a bone graft prior to dental implantation. In addition, these prostheses are effective when the patient has anatomical limitations, such as extension of the maxillary sinus in previous maxillary rehabilitation.[1]
Another treatment option in the case of anterior edentulous areas is the placement of two implants in the region of the lateral incisors,[9]
[10] as this allows the dental technician to manufacture central bridges with the necessary aesthetic and functional dimensions, in addition to favoring aesthetics and manipulation of soft tissue.[1]
[10] Scientific studies that address this theme show divergent results in several factors, which hinders the comparison between studies.[1]
[9]
[10]
Currently, the advances in image processing and analysis techniques through software have been changing the reality in Dentistry. Digitization is widely used in studies as well as dental clinic in several areas, such as in radiology, orthodontics, implantology, and prosthodontics. Finite element analysis (FEA) is used in Dentistry to study how an implant and prosthodontics structures will respond to complex stress loading and to predict any possibility of structural failure.[2]
[9]
[11]
[12]
[13]
[14]
Thus, the aim of the present study was to verify the stress distribution in the implant, crown, and bone tissue in zirconia implants (single-body configuration) placed in different positions in the anterior region of the maxilla (12–22) in rehabilitation with and without a cantilever. The tested hypothesis is that implants placed on central incisors (11–21; with cantilever) would have lower mechanical performance compared with the other groups.
Materials and Methods
The bone model was acquired from images obtained from the toothed maxilla using computed tomography (i-CT Cone Beam 3D Dental Imaging System, Imaging Sciences International; Hatfield, Pennsylvania, United States). This model consisted of a real geometric representation of the upper maxillary arch (18.0 mm height × 19.0 mm mesiodistal width × 15.0 mm buccolingual depth) with type three bone (thin layer of cortical bone—1.0 mm—around a cancellous bone) concerning its density. The purpose of tomography was to provide a bone-based model for the positioning of implants and subsequent manufacture of prostheses for simulation.
Three models were made in finite elements using a software (SolidWorks; SolidWorks Corporation, Concord, Massachusetts, United States). For all groups, a monotype zirconia implant (single-body zirconia configuration; 4.1 × 12.0 mm implant) with prosthetic lithium disilicate crowns (0.5 mm thickness) was simulated using images obtained by computed microtomography. The drawings were standardized and three-dimensional computational models were generated, following the standards of conventional fixed partial and cantilever prostheses ([Fig. 1]). The models were divided according to the placement of the implants in three groups:
Fig. 1 Drawings (mesh and geometric) of the groups tested (A) Zr (11–21), (B) Zr (12–22), and (C) Zr (11–22).
Zr (11–21): implants in regions 11 and 21 with cantilever.
Zr (12–22): implants in regions 12 and 22 without cantilever.
Zr (11–22): intercalated implants in regions 11 and 22.
The implants were positioned parallel to their long axis on the bone base from tomography, and the implant platform was positioned at the height of the bone crest with standardization of the bone ridge. All implants were considered totally osseointegrated. The materials were considered isotropic, homogeneous, and linearly elastic. The values of the Young’s modulus and the Poisson’s ratio are described in [Table 1].
Table 1
Physical properties of the materials used
Material
|
Modulus of elasticity (GPa)
|
Poisson’s ratio
|
Source: Adapted from Çaglar et al 2011[29] and Ereifej et al 2011[30].
|
Cortical bone
|
13.7[29]
|
0.30[29]
|
Cancellous bone
|
1.37[29]
|
0.30[29]
|
Zirconia
|
200[29]
|
0.31[29]
|
Lithium disilicate
|
68[30]
|
0.24[30]
|
Then, the models were imported into the ANSYS Workbench 14.0 software (Swanson Analysis Systems Inc.; Houston, Pennsylvania, United States) for FEA. For all groups, an oblique load of 150 N at 45 degrees was applied to 11 and 21 axes, at the palatal portion of the prosthetic crowns. The oblique load results in harmful tensile/compression stresses and 150 N is the mean load applied on anterior teeth as well as used in several studies.2,10,15 The results were evaluated qualitatively and quantitatively by maximum principal N stress (tensile) and minimum principal stress (compressive) in the implant, crown, and bone tissue.
Results
In all considered positions, the cervical palatal region of the implant was where the highest concentrations of the maximum principal stress occurred, with the worst result for the Zr (12–22) group where the highest values of the maximum principal stress occurred (2.05 MPa). The best result was found in the Zr (11–21) group where the lowest values of the maximum principal stress (1.53 MPa) occurred ([Fig. 2]).
Fig. 2 Vestibular view of the implants showing the distribution of the maximum principal stress on the implants in the groups (A) Zr (11–21), (B) Zr (12–22), and (C) Zr (11–22).
Concerning bone tissue, the maximum principal stress was greater in the palatal region near the cervical of the implants. Small differences can be observed regarding the distribution of these stresses among the groups evaluated. In addition, the different groups tested showed similar maximum values for the intensity of the maximum principal stress (approximately 8.0 MPa) ([Fig. 3]).
Fig. 3 Distribution of the maximum principal stress on bone tissue in the groups (A) Zr (11–21), (B) Zr (12–22), and (C) Zr (11–22).
In [Fig. 4], it was observed qualitatively and quantitatively that the minimum principal stress in bone tissue was concentrated in the vestibular region of bone tissue close to the implants. Equally to the maximum principal stress, a great similarity can be observed in the intensity (approximately 9.0 MPa) and the form of distribution of the minimum principal stress.
Fig. 4 Distribution of minimum principal stress on bone tissue in groups (A) Zr (11–21), (B) Zr (12–22), and (C) Zr (11–22).
Regarding the ceramic crown, the maximum principal stress was concentrated in the cervical palatal region of the crown. The Zr (11–21) group presented the highest values of the maximum principal stress (8.05 MPa). The Zr (12–22) group showed the lowest values of the maximum principal stress (5.47 MPa). The Zr (11–22) group, on the other hand, showed intermediate maximum stress values between the two groups, but presented similar stress in the two elements, there was a higher concentration of stress in element 21 compared with element 22 ([Fig. 5]).
Fig. 5 Distribution of maximum principal stress on the crown in the groups (A) Zr (11–21), (B) Zr (12–22), and (C) Zr (11–22).
[Table 2] shows the maximum principal stress and minimum principal stress in implant, bone tissue, and crown for all groups tested.
Table 2
Maximum principal stress and minimum principal stress (MPa) in implant, bone tissue, and crown for all groups tested
|
|
Maximum principal stress
|
Minimum principal stress
|
Zr (11–21)
|
Implant
|
1.53
|
–
|
Crown
|
8.05
|
–
|
Bone tissue
|
6.13
|
7.40
|
Zr (12–22)
|
Implant
|
2.05
|
–
|
Crown
|
5.47
|
–
|
Bone tissue
|
8.18
|
9.86
|
Zr (11–22)
|
Implant
|
1.80
|
–
|
Crown
|
7.06
|
–
|
Bone tissue
|
7.16
|
9.86
|
Discussion
The hypothesis tested was rejected since the implants in the region of the central incisors (11–21) had a lower concentration of stress, and the stress transmitted to bone tissue was equivalent between the three groups tested.
The lower concentration of stress in the implant and crown in Zr (11–21) group can be explained by the presence of the cantilever. The positioning of the implant adjacent to the cantilever does not promote problems in the prosthesis or bone resorption as long as the position and extension of even one dental element from the implant to the cantilever is respected,[15] as in the current study. On other hand, a study[7] showed that these variables (position and extension of the cantilever), since the greater extension and distance of the implants increase the prosthesis lever arm, consequently increases the accumulated stress in the prosthetic components. Thus, the presence of the cantilever does not increase the stress that is transferred to the implants, crown, and bone tissue.
In the implants, the stress concentration was located in the cervical palatal region of the implants. This is due to the macrogeometry of the monotype zirconia implant, which provides a more uniform stress distribution in its body and more concentrated in the region of the implant neck and the alveolar region of the bone tissue.[2]
[16]
[17]
Regarding the load distribution in bone tissue, this study showed that the stress distribution was equivalent in the three groups studied, with the tensile stress being concentrated in the palatal region where the bone is in contact with the neck (first threads) of the implant and compressive stress concentrated in the vestibular region of the bone tissue close to the implants. The positioning of the implants 1 mm below the bone cortex, as in the current study, provides better dissipation of masticatory loads in bone tissue.[18]
[19]
[20]
[21] Thus, the position of the implants within the bone tissue influences the stress distribution in the bone more than the presence of the cantilever or the positioning of the implants in the different anterior regions evaluated (11–21, 12–22, and 11–22). The stress distribution in bone tissue is influenced, in addition to the macrogeometry of the monotype zirconia implant, by the type of bone tissue and the direction of the applied load.[2]
[10]
[22] In low-density bone tissue, stress is distributed over a larger area around the apex, while in high-density bone, stress is more concentrated in the cervical region.[2]
[6]
[19] In the direction of the applied load, the oblique inclination of the load directs the stresses to the areas where there was compression, influencing the biomechanical performance of implant-supported rehabilitation.[22]
[23]
All groups showed adequate levels of bioperformance (ratio <1), with stress values below the critical limit that the cortical bone can receive without suffering damage, such as plastic deformations (beginning of the bone resorption process) or fracture.[2]
[24]
The results are also linked to the type of load applied in the study. Thus, differences found between the results of current study and the literature can also be explained by the intensity and inclination of the load, since different loads form different lever arms in monotype zirconia implants, increasing the stresses in the regions of rotational fulcrum and bone crest.[5]
[20]
[25]
[26]
[27]
[28] Furthermore, the results of this study (monotype zirconia implant) showed lower stress values than several studies[2]
[3]
[17]
[24]
[29] that used titanium implants. The highest Young’s Modulus for zirconia (200 GPa; [Table 1]) in relation to the titanium (115 GPa)[29] could explain this difference. A high Young’s Modulus (elasticity) causes low stress in crown, implant, and bone as well as less bone resorption in the peri-implant region against occlusal load.[29]
There is still controversy regarding the planning and location of implants in cases of anterior maxillary rehabilitation. Despite the values considered in this study to be adequate, it is important to consider the limitations inherent in any academic study. In the current study, we did not consider different occlusal patterns, parafunctional habits, greater extension of the prosthetic space, in addition to the quality of bone tissue and the prolonged use of the prosthesis, which could maximize transferred stress and generate system overload, affecting the prognosis of the rehabilitating treatment.[6]
[8]
[10]
[15]
[16]
[30] Thus, future research on this topic is needed to assess the longevity of anterior maxillary rehabilitation with monotype zirconia implants with and without cantilever.
Conclusion
According to the results obtained, it is concluded that (1) the position of the implants did not interfere in the transmission of stress in bone tissue, (2) the positioning of the implants in regions 11–21 presented lower stress values in the implants and higher stress in the crown, and (3) the use of cantilever does not increase stress on implants, crown, and bone tissue.