Key words portosystemic shunt - transjugular intrahepatic balloon angioplasty - paclitaxel
Introduction
Transjugular intrahepatic portosystemic shunt (TIPS) is the therapeutic method of
choice to reduce the complications of portal hypertension. Because of its feasibility
and low mortality rate, TIPS has become a clinical standard [1 ]
[2 ]
[3 ]. Some investigators have pointed out that shunt dysfunction occurs in 50 – 85 %
of patients within the first year of TIPS insertion in patients with bare metal stents
[4 ]. Early TIPS dysfunction (within 30 days) is most commonly caused by thrombosis linked
to perioperative trauma and bile leak, while in-stent intimal pseudohyperplasia is
a frequent cause of long-term TIPS dysfunction [1 ]
[5 ]
[6 ], which can lead to frequent re-interventions in a subgroup of TIPS patients with
a long clinical follow-up and TIPS patency.
Plain optimal balloon angioplasty (POBA) is an established method for vascular in-stent
dilation and functional recovery. Paclitaxel coated balloons (PCB) are used in arteries
and dialysis shunts to prospectively inhibit intima proliferation to prevent recurrent
stenosis and therefore optimize secondary patency after PTA [7 ]
[8 ]
[9 ]
[10 ].
This is the first-in-man use of PCB for patients with cirrhotic liver disease and
recurrent TIPS in-stent stenoses. In-stent patency after treatment was evaluated as
well as the short-term tolerability, the residual drug content of used balloon catheters
and sheaths, and systemic venous plasma paclitaxel concentrations.
Materials and Methods
Patients
All patients gave informed consent to this compassionate use of treatment. After TIPS
insertions, patients in our hospital are monitored with clinical examination as well
as ultrasound and invasive angiographic follow-up examinations every 6 months, according
to our hospital standard. This analysis included six patients (3 men, 3 women; mean
age: 65 ± 10 years) who were diagnosed with recurrent TIPS dysfunction due to in-stent
restenoses. Restenoses occurred during a mean period of 76 ± 32 months (mean: 7 ± 3
restenoses per patient) after implantation of TIPS. All patients had a clinical indication
for PTA and were treated between January 2010 and December 2012. The mean TIPS duration
between stent insertion and the first uncoated PTA was 18 ± 7 months ([Table 1 ]). Following TIPS, all patients underwent angiographic examination at 6-month intervals,
according to our in house follow-up protocol, and additional examination if there
was a clinical need.
Table 1
Individual patient data.
patient
age[1 ]
gender
initial stent type
covering
TIPS-duration before paclitaxel (months)
number of previous stenoses (n)
first PTA after TIPS insertion (months)
in-stent stenosis side
1
46
male
Viatorr (10/60 mm, W.L. Gore)
covered
72
9
22
hepatic vein
2
73
female
Luminexx (10/60 mm, C.R. Bard)
non-covered
84
9
12
hepatic vein
3
72
male
Memotherm (10/50 mm, 10/70 mm, C.R. Bard)[2 ]
non-covered
90
8
11
hepatic vein
4
70
female
Memotherm (10 × 60 mm, C.R. Bard)
non-covered
102
6
25
hepatic vein
5
65
female
Memotherm (10 × 60 mm, C.R. Bard)
non-covered
92
8
25
portal vein
6
66
male
Luminexx (10/50 mm, 10/80 mm, C.R. Bard)2
non-covered
13
1
13
portal vein
mean ±SD
65 ± 10
76 ± 32
7 ± 3
18 ± 7
1 Age at time of PTA with paclitaxel-coated balloon catheter.
2 Patients 3 and 6 received two overlapping stents.
Shunt dysfunction was clinically defined as a portosystemic pressure gradient (PPG)
exceeding 15 mmHg and in-stent restenosis and/or in-stent lumen narrowing to less
than 50 % of the stent diameter combined with new ascites or rebleeding. Mean PPG
was determined before and after the interventions.
Patients with tumor disease were not treated with PCB. Results of PCB treatment were
compared on an intra-patient basis with the results of earlier treatments with uncoated
balloons (the latter thus served as controls).
Initial TIPS Insertion
Initial TIPS insertion was performed under combined fluoroscopic and sonographic guidance
in all patients. Six patients received eight initial stents: Luminexx (C.R. Bard,
Murray Hill, USA) 10/50 mm (n = 1), 10/60 mm (n = 1), and 10/80 mm (n = 1); Memotherm
(C.R. Bard, Murray Hill, USA) 10/50 mm (n = 1), 10/60 mm (n = 2), and 10/70 mm (n = 1).
In two patients there were two overlapping stents. One patient received a 10/60 mm
Viatorr endoprosthesis (W.L. Gore, Flagstaff, USA). In all patients, the underlying
liver disease was alcoholic cirrhosis. Individual data are shown in [Table 1 ].
Procedure Protocol
All invasive interventions were performed under systemic pain medication (5 mg pitriamid)
and local anesthesia, according to the hospital standard. After exclusion of any coagulation
disorders, a single dose of 5000 IU heparin was administered intravenously before
angioplasty.
Conventional Angioplasty in Restenosis
A transjugular, sheath-mediated angioplasty balloon catheter (Passeo-35, Biotronik,
Germany; balloon dimensions: 10/40 mm) was placed in the portocaval stent according
to the identified stenosis and inflated with a constant pressure of 7 atm that was
maintained for 3 minutes. All patients repeatedly received POBA with uncoated balloon
catheters before PCB angioplasty for TIPS restenosis.
PCB Angioplasty in Restenosis
Technical insertion of PCB was performed as described for conventional PTA. Paclitaxel
was passively transferred into the segment of stenosis during inflation of the PTA
balloons (Admiral™, Invatec S.p.A, Italy; balloon dimensions: 10/40 mm). The angioplasty
balloons were inflated with a constant pressure of 7 atm that was maintained for 3
minutes.
During the post-procedural hospital stay, all patients were observed and monitored
for at least 24 hours after the procedure.
Paclitaxel-Coated Balloon Catheters and Paclitaxel Analysis
The balloon surfaces of sterile PTA balloon catheters were coated with paclitaxel
(3 µg/mm²) in a hydrophilic matrix by the hospital pharmacy as described before [11 ]. After balloon dilation, residual paclitaxel on the balloons and sheaths was analyzed
by high-performance liquid chromatography (HPLC).
The difference between the pre- and post-interventional amounts of paclitaxel on the
balloons was defined as the in vivo dose. Systemic paclitaxel exposure was assessed
in venous blood samples before treatment and 0.5, 4, and 24 hours after the intervention.
The venous blood samples were analyzed for residual paclitaxel by HPLC. The limit
of paclitaxel detection was 5 ng/ml for all samples.
Follow-up Protocol
The standard follow-up protocol after PCB included clinical evaluation and ultrasound
imaging as well as invasive pressure monitoring and angiographic imaging every six
months during a two-year follow-up period after single PCB. In any case of recurrence
of shunt dysfunction with clinical signs of portal hypertension after PCB, POBA with
uncoated balloon catheters was performed. The follow-up endpoint during angiographic
examination was the first recurrent need for a further PTA (POBA) after a single PCB-PTA.
In-stent minimal lumen diameter (MLD) at the side of stenosis was measured on a commercially
available workstation. The difference in MLD between two successive angiographic examinations
was given as late lumen loss (percentage decrease in MLD). In cases where a PTA (POBA
or PCB) was performed, the MLD after PTA was taken as the baseline value for the following
time interval. Mean late lumen loss (LLL), defined as MLD at the last assessment minus
MLD at the following assessment in mm (or the difference in percentage of diameter
stenosis) for several time intervals was calculated for each patient.
The percentage of angiographic examinations that needed a PTA was calculated for the
period before (period of recurrent stenosis following POBA) and after a single paclitaxel
PTA. The two percentages were compared with each other.
Statistics
Numerical values are given as means with standard deviations. Statistical analysis
was performed by a two-sided Student’s t test using Graph Pad Prism 6.0 for Macintosh.
Binary and categorical variables such as clinical outcomes were compared using Fisher’s
exact test. A P value below 0.05 was considered statistically significant.
Results
During a 16-month period (from January 2012 until October 2013), 6 of 48 patients
with recurrent re-stenosis and concomitant recurrent ascites were treated with PCB.
A mean of 7 ± 3 restenoses treated with POBA per patient (range 1 – 9) were observed
before considering PCB.
The mean TIPS duration between stent insertion and the first POBA was 18 ± 7 months.
The mean duration of re-intervention with a POBA was 6 ± 0.4 months. A POBA was indicated
for every other angiographic follow-up examination (41 of 77, 53 %).
PCB-PTA was performed 76 ± 32 months after TIPS insertion ([Table 1 ]). In-stent stenosis occurred at the proximal (venous) segment in four patients and
at the distal (portal) segment of the stent in two patients ([Fig. 1 ]). In an intra-individual correlation, late lumen loss was higher after POBA (2.4 ± 1.5 mm/28 ± 18 %)
than after PCB-PTA (0.5 ± 0.8 mm/7 ± 11 %, P = 0.029) ([Fig. 2 ]). After PTA with a PCB, the percentage of angiographic examinations with the need
of PTA decreased to 19 % (3 of 16, P = 0.014) ([Fig. 3 ]). In two patients, restenosis was diagnosed after six months. Four of six patients
showed no relevant restenosis up to 18 months after PCB PTA, and two of them showed
no relevant restenosis after 24 months post intervention. One of these patients received
a liver transplantation (not attributable to paclitaxel) after 18 months of follow-up
without occurrence of restenosis. One patient with recurrent re-stenosis after PCB
was treated with an endograft (stent in stent, Viatorr, 10/60 mm, W.L. Gore, Flagstaff,
USA) after 20 months ([Fig. 4 ]). Detailed individual data are presented in [Table 2 ].
Fig. 1 Images of transjugular intrahepatic portosystemic shunt (TIPS) in one patient before
and after catheterization with a paclitaxel-coated balloon (PCB) catheter. A Decreased in-stent minimum lumen diameter (MLD) before paclitaxel PTA on day 0 after
recurrent stenosis. B Paclitaxel PTA. C Increased in-stent MLD after PCB treatment. D–F Follow-up images obtained 6 – 18 months after treatment show nearly the same in-stent
MLD at all follow-up times. The arrow indicates the in-stent lumen reduction, which
did not increase over the 24-month follow-up period. Asterisk = portal vein; Bar = 10 mm;
PTA = percutaneous transluminal angioplasty.
Abb. 1 Abbildungen des transjugulären intrahepatischen Shunts des selben Patienten vor und
nach Katheterisierung mit Paclitaxel-beschichteten Ballonkathetern (PCB). A Abnahme des minimalen In-Stent-Lumendurchmessers (MLD) vor Paclitaxel-PTA an Tag
0 nach vorangegangenen rezidivierenden Stenosen. B Paclitaxel-PTA. C Erhöhter In-Stent-MLD nach PCB-Behandlung. D–F Die Abbildung der angiografischen Nachsorge im Zeitraum von 6 – 18 Monaten nach Behandlung
zeigen nahezu die gleichen MLD. Die Pfeile kennzeichnen die In-Stent-Lumen-Reduktionen,
bei welchen kein Anstieg über einen Verlauf von 24 Monaten zu beobachten war. Stern = Portalvene;
Balken = 10 mm; PTA = Perkutane transluminale Angioplastie.
Fig. 2 In-stent minimal lumen diameter (MLD) is affected by local paclitaxel administration.
Late lumen loss was higher after plain optimal balloon angioplasty (POBA) with uncoated
balloons; *P = 0.029, paclitaxel-coated balloon (PCB) vs. POBA.
Abb. 2 Der minimale In-Stent-Lumendurchmesser (MLD) wird durch die lokale Paclitaxel-Applikation
beeinflusst. Das Late lumen loss war nach konventioneller PTA mit unbeschichteten
Ballonen (POBA) erhöht; *P = 0,029, Paclitaxel-beschichtete PTA (PCB) vs. POBA.
Fig. 3 Local paclitaxel delivery from coated balloons decreased the need for clinically
driven repeated percutaneous transluminal angioplasty (PTA) compared with plain optimal
balloon angioplasty (POBA); *P = 0.014, paclitaxel-coated balloon (PCB) vs. POBA.
Abb. 3 Eine lokale Paclitaxelapplikation durch beschichtete Ballone (PCB) verringert die
Notwendigkeit für weitere Angioplastien verglichen mit der konventionellen PTA (POBA);
*P = 0,014, PCB vs. POBA.
Fig. 4 Months without repeat PTA (percutaneous transluminal angioplasty) after treatment
with paclitaxel-coated balloon (PCB). Day 0 corresponds to a single PCB treatment.
No need to repeat PTA in patients 2, 4, and 6. Patient 6 received a liver transplantation
after 18 months. Patient 5 was treated with an endograft (stent in stent) after 20
months. Asterisk = first PTA (POBA) after PCB; PTA = percutaneous transluminal angioplasty;
PCB = paclitaxel-coated balloon.
Abb. 4 Monate ohne eine erneute PTA (perkutane transluminale Angioplastie) nach Behandlung
mit Paclitaxel-beschichteten Ballonen (PCB). Tag 0 entspricht einer einmaligen PCB.
Bei den Patienten 2, 4 und 6 bestand keine Notwendigkeit einer repetierten PTA. Der
Patient 6 erhielt eine Lebertransplantation nach 18 Monaten. Der Patient 5 erhielt
nach 20 Monaten einen weiteren Stent (Stent in Stent). Stern = Erste PTA (POBA) nach
PCB; PTA = Perkutane transluminale Angioplastie; PCB = Paclitaxel beschichteter Ballon.
Table 2
Individual patient in-stent MLD and LLL before and after paclitaxel-coated balloon
PTA (PCB).
plain optimal balloon angioplasty (POBA)
paclitaxel-coated balloon (PCB)
patient
MLD at last assessment or after reintervention (mm)
MLD at following assessments (mm)
LLL (mm)
LLL (% of original MLD)
MLD after PCB PTA or assessment after PCB PTA (mm)
MLD at following assessment (mm)
LLL (mm)
LLL (%)
1
7.6 ± 1.9
5.7 ± 1.0
2.1 ± 1.9
24 ± 22
6.3 ± 1.0
5.6 ± 0.5
0.6 ± 0.8
9 ± 13
2
7.9 ± 1.1
6.0 ± 0.8
1.9 ± 1.4
23 ± 15
7.1 ± 0.3
7.6 ± 0.7
–0.5 ± 0.6
–7 ± 8
3
7.7 ± 1.2
6.5 ± 0.9
1.2 ± 1.2
14 ± 16
7.5
5.6
1.9
25
4
8.5 ± 1.2
6.4 ± 2.3
2.1 ± 1.9
25 ± 21
6.2 ± 0.2
6.3 ± 0.3
–0.1 ± 0.5
–2 ± 8
5
8.5 ± 0.9
6.8 ± 1.5
1.8 ± 1.5
21 ± 18
7.6
6.9
0.7
9
6
8.6
3.1
5.5
64
7.8 ± 0.1
7.4 ± 0.6
0.4 ± 0.7
5 ± 9
mean ± SD
8.1 ± 0.5
5.8 ± 1.4
2.4 ± 1.5
28 ± 18
7.1 ± 0.7
6.6 ± 0.9
0.5 ± 0.8[1 ]
7 ± 111
All values referring to individual patients are given as mean±SD of repeat observations
in these patients before the use of a paclitaxel-coated balloon (first 4 columns)
and after the use of a paclitaxel-coated balloon (last 4 columns). The line at the
bottom of the table displays the mean±SD of the means of the 6 patients. PTA = percutaneous
transluminal angioplasty; PCB = paclitaxel-coated balloon; LLL = late lumen loss;
MLD = in-stent minimal lumen diameter.
1 P< 0.05, uncoated balloon vs PCB.
Residual paclitaxel on the balloons was 28 ± 9 % of the original 3 µg paclitaxel/mm2 (0.8 ± 0.3 paclitaxel/mm2 ) and 0.2 ± 0.1 % of the calculated dose was found in the extracts of the sheaths
(0.008 ± 0.003 paclitaxel/mm2 ). The corresponding mean in vivo paclitaxel dose was 72 ± 9 % of the total dose on
the balloons. Paclitaxel levels were below 5 ng/ml in all venous blood samples after
treatment. The procedure was well tolerated and no clinical side effects attributable
to paclitaxel were observed.
Discussion
This retrospective analysis investigates the compassionate first-in-man use of PCB
in the treatment of recurrent TIPS dysfunction due to multiple restenoses in six patients.
Our results suggest that secondary patency of TIPS is higher after treatment with
paclitaxel-coated balloons compared with uncoated balloon angioplasty.
Thromboses and bile leaks were identified as main causes for early TIPS dysfunction.
In contrast, one reason for long-term TIPS dysfunction is the unpredictable patency
of the shunts due to pseudointimal hyperplasia within the parenchymal tract or within
the outflow hepatic vein. Histopathologic investigations of in-stent neolumina in
the TIPS tract have revealed the development of an organized in-stent cell layer that
is different from the lining of a normal vessel wall. Continuous lining of the inner
stent with mature endothelial cells of unknown origin has been reported [12 ]
[13 ]. Stimulation of pseudointima formation through migration and proliferation of smooth
muscle cells in the middle layer [14 ], lined by hepatic sinusoidal cells [14 ], and increased collagen type I and II secretion by TIPS cells have been found in
human specimens of TIPS patients [13 ]. Proliferation of smooth muscle cells has also been identified as a major cause
of TIPS stenosis in an animal model [15 ]. Another group has described proliferation of myofibroblasts and fibrin within the
neolumen [12 ]. Paclitaxel is known to have an effect on intimal cell proliferation and also inhibits
cell migration in a concentration-dependent manner [16 ]. Therefore, it might have an effect which is not limited directly to the place of
application [17 ]. The artificial connection of two vessel systems with differences in resistance
causes turbulences and shear stress, which could play a complementary role in focal
pseudointimal hyperplasia. Electron-microscopic findings suggesting that endothelial
cell arrangement depends on the direction of blood flow [12 ] are in line with this hypothesis.
Growing knowledge about stent material characteristics and their effects has led to
several studies investigating modified stent surfaces customized to avoid shunt dysfunction.
The first use of such a modified, polytetrafluoroethylene (PTFE)-covered stent in
humans was reported in 1999. Various studies found a better patency rate and clinical
outcome compared with conventional bare metal stents [3 ]
[18 ]
[19 ]
[20 ]. These data are unambiguous; however, restenosis occurs with both types of stent.
PTFE-covered endoprostheses show stenoses in the hepatic venous outflow tract and
at the junction between the non-covered and covered segments of the stent graft [19 ]
[21 ]. Venous outflow stenosis is one of the most frequent locations of TIPS stenosis,
especially if the stent is not reaching the confluence of the hepatic veins. The outflow
hepatic veins can shrink diffusely to half of their diameter after TIPS implantation
[4 ]. In our patients no venous outflow tract stenoses were observed.
Additional studies to analyze the effect of local paclitaxel delivery to the hepatic
venous outflow tract would be helpful, as paclitaxel proved to show positive effects
on stenosis in arteriovenous fistulas compared to POBA [8 ]. Other strategies to increase TIPS patency in humans include intraluminal (in-stent)
brachytherapy [22 ] and local delivery of motexafin gadolinium [23 ]. First studies in an animal model showed promising effects of local paclitaxel delivery
on pseudointima hyperplasia [24 ]. Paclitaxel is a lipophilic chemotherapeutic taxane that causes irreversible stabilization
of microtubules and thus interrupts the mitosis process. Its cytostatic effects have
been shown in some tumor models [25 ]. Paclitaxel also inhibits the proliferation of smooth muscle cells [26 ]
[27 ], which is one cause of in-stent hyperplasia [14 ]
[26 ]. Lipophilicity of paclitaxel facilitates cell penetration and long residence time,
which may underlie its long-term local efficacy [15 ]
[26 ]
[28 ]. Paclitaxel-mediated intimal hyperplasia inhibition has been reported for human
femoroarterial [9 ]
[10 ] and coronal in-stent stenosis [7 ] as well as for arteriovenous grafts [8 ]. The results presented here indicate that local delivery of a single paclitaxel
dose can reduce neointimal hyperplasia and thus the secondary patency after PTA. Therefore,
paclitaxel-coated balloon PTA may be a novel strategy to improve the outcome after
treatment of TIPS in-stent restenoses. We performed a follow-up according to our standardized
hospital protocol and our results indicate a long-term effect persisting for at least
18 months after paclitaxel-coated balloon PTA in a setting with frequent restenoses.
The positive effect after a single paclitaxel administration is in line with the results
of previous studies, which have shown a long persistence of paclitaxel in arterial
vessel walls [29 ]. We found relatively high residual amounts of paclitaxel on our balloons (28 ± 9 %
of dose) compared with the use of this type of coating in coronary vessels (about
10 % of dose) [11 ], which might be surface dependent. This finding, combined with a negligible proportion
on sheaths (0.2 ± 0.1 %) in our study, still suggests successful local administration
of paclitaxel. Paclitaxel concentrations in venous blood in the first 4 hours after
PTA were below the detection limit of our method (< 5 ng/ml plasma), suggesting systemic
venous blood concentrations below 5 ng/ml. Data from the use of paclitaxel in tumor
treatment indicate that plasma concentrations greater than 80 ng/ml are required for
systemic effects [30 ]. This finding is consistent with those of a previous study [31 ] using PCB in the arterial system.
Our analysis of paclitaxel-coated balloons in the treatment of TIPS restenosis has
several limitations. These include the small number of patients, the lack of blinding
of physicians, the participation of only a single hospital, and the lack of a control
group due to compassionate use of treatment. Also, this study included mostly patients
with bare metal stents and a long follow-up period, which does not match the current
day-to-day standard of patients with endoprosthesis. However, these findings may encourage
prospective and randomized studies to investigate new options in the prophylaxis and
treatment of shunt dysfunction in patients with cirrhotic liver disease.
In conclusion, in our patients with recurrent in-stent stenoses in bare metal stents,
a single PTA with paclitaxel-coated balloons (PCB) prolongs secondary patency without
a systemic effect of paclitaxel.