Key words
image manipulation/reconstruction - liver - dual energy CT - metastases - window setting
Introduction
Several studies have demonstrated the benefits of dual-energy CT for the detection of hypervascular abdominal lesions, especially of the liver [1]
[2]
[3]
[4]. The value of dual-energy CT for contrast-enhanced abdominal organ imaging is an exciting topic. The improved CT contrast resolution obtained at low kVp settings has been shown to provide higher detection rates of hypervascular lesions in phantom and clinical hepatic imaging, provided that the noise in the low kVp image is not excessive [4]
[5]. Further attention should be paid to the visibility of hypodense hepatic metastases [6]
[7]
[8]
[9]. 80 kVp data acquired from a DECT scanner demonstrate a greater attenuation difference between metastatic disease and normal liver, thus increasing lesion conspicuity and potentially improving assessment of early metastatic disease [9].
It is unclear whether window settings (WSs) can influence the performance of low kVp images. While liver windows (50/350 HU) are commonly used in clinical practice, a study concluded that the use of liver windows has limited added value in the detection of liver disease, as evidenced by an only 3.1 % increase in lesion detection in 1175 consecutive patients, resulting in a change in diagnosis in only 1.7 % and a change in follow-up recommendation in 0.85 %. [9]
[10]
[11]
[12]
[13]. Therefore, an optimization of image parameters would be useful here.
The aim of this study was to investigate DECT of hypovascular liver metastases in patients with colorectal cancer with a special focus on adjusted window settings (WSs) in addition to the use of low energy data sets.
Materials und Methods
Patient Population
Between October 2009 and December 2011, 30 consecutive patients with CRC and LM (17 men, 13 women; mean age 65 years +/– 14 years) underwent dual-energy CT (DECT) for regular staging. The inclusion criterion was an abdominal diameter measured ≤ 35 cm on the AP or lateral topogram [9].
Liver metastasis was determined by comparison with monoenergetic pre- and follow-up scans (pre 5.3 months ± 2 (30/30 patients); post 4.8 months ± 1.9 (30/30 patients).The study was approved by our institutional review board and all patients provided written informed consent before participation.
Imaging Protocol and Scan Parameters
All CT examinations were obtained using a dual-source multi-detector CT unit (Somatom Definition™ Dual Source; Siemens Medical Solutions, Forchheim, Germany). This system consists of two X-ray tubes mounted on one gantry at a 90° angle [14] and two corresponding 64-section detectors. One detector (detector A) covers a 50-cm field of view and the other detector (detector B) covers a 26-cm field of view [14]. Patients were positioned supine on the table. Patients were positioned slightly off center to the left to ensure complete coverage of the liver by the smaller field of view of detector B. After intravenous injection of a non-ionic contrast agent (1.5 mL per kilogram of body weight, Xenetix 300™, Guerbet, Sulzbach, Germany), bolus tracking was started in the abdominal aorta at the level of the celiac trunk. Dual-energy datasets were acquired in the portal-venous (bolus tracking and 80 sec) phase from the dome of the liver to the iliac crest by operating tube A at 140 kVp and a reference value of 96 mAs and tube B at 80 kVp and a reference value of 404 mAs. For both tubes, online dose modulation (Care DOSE 4D™, Siemens Medical Solutions) was used. For the dual-energy CT, collimation was set to 14 × 1.2 mm on both detectors. For all imaging, the gantry rotation speed was 0.5 seconds.
Post-Processing and Image Reconstruction
Axial post-contrast images were reconstructed using a section thickness and an increment of 3 mm. Dual-energy CT generates three different series of images: 80 kV images, 140 kV images and weighted-average images, which are based on attenuation information on images obtained from both detectors, using 70 % information from the high kVp and 30 % from the low kVp images. Weighted-average images are similar to 120-kVp imaging of the abdomen. For post-processing, images were loaded onto a dedicated dual-energy post-processing workstation (syngo MMWP; Siemens Medical Solutions, Forchheim, Germany).
Measurements of Metastases and Image Quality
Two experienced radiologists read all cases in consensus. On a dedicated workstation (syngo MMWP, Siemens Medical Solutions), 3-mm CT images were displayed side-by-side (window setting (WS): 25/200 HU, 50/200 HU, 75/200 HU, 25/350 HU, 50/350 HU, 75/350 HU, 25/500 HU, 50/500 HU, 75/500 HU). CT attenuation was determined in 6 regions using regions of interest of 0.2 – 1.0 cm² in size: hepatic metastases; normal hepatic parenchyma; aorta at the level of the coeliac trunk; psoas muscle; perihepatic fat; and air in front of the patient. The signal (liver)/noise ratio and liver/LM ratio were calculated (SNR/LLMR). As images could be easily discriminated, radiologists were not blinded as to which image represented the 80 kVp dataset and the virtual 120 kVp dataset. For both CT image sets, the radiologists rated the conspicuity of the LMs on a five-point scale: 1: excellent; 2: good; 3: fair; 4: poor; and 5: not interpretable. Both readers reported all LMs separately for the previous best proven and regular (50/350 HU) WSs of the 80 kVp images and the virtual 120 kVp images according to size (< 1 cm; 1 – 2 cm; > 2 cm), total number and number per patient.
Statistical Analysis
CT attenuation, the number of detected LMs and the image quality parameters were tested for differences using the Wilcoxon test for paired samples. The Kruskal-Wallis test was used to investigate statistically significant differences in the qualitative scores. If there was a statistically significant difference among the groups, pairwise comparisons were performed using the Steel-Dwass test [8]. For all studies, a difference with a P-value of less than 0.05 was considered significant. All calculations were performed using statistical software (SPSS, version 21, SPSS Inc., Chicago, USA).
Results
The best conspicuity of LMs on 80 kVp images was documented with the WS 25/350 HU compared to the best proven WS on virtual 120 kVp images with 75/200 HU (1.2 vs. 2.5; p = 0.01) ([Fig. 1]). The mean conspicuity of all WSs of the 80 kVp images was 2.6 and significantly better compared to the mean conspicuity of all WSs of the virtual 120 kVp images with 3.9 (p < 0.05). The best mean window width for 80 kVp and virtual 120 kVp images was 350 HU (1.5/p < 0.05; 3.3/p < 0.05) ([Table 1]).
Table 1
Conspicuity of LMs depending on window settings (HU) and tube voltage.
Tab. 1 Erkennbarkeit der LM abhängig von den Fensterwerteinstellungen (HU) und der Röhrenspannung.
|
80 kVp
|
120 kVp
|
conspicuity of LMs/each WS (HU)
|
25/200 = 4.5
|
25/350 = 1.2
|
25/500 = 2.8
|
25/200 = 5.0
|
25/350 = 3.5
|
25/500 = 4.5
|
50/200 = 4.1
|
50/350 = 1.6
|
50/500 = 2.7
|
50/200 = 4.5
|
50/350 = 3.1
|
50/500 = 4.6
|
75/200 = 2.1
|
75/350 = 1.8
|
75/500 = 2.8
|
75/200 = 2.5
|
75/350 = 3.2
|
75/500 = 4.6
|
conspicuity of LMs/window width 200/350/500 HU
|
3.6
|
1.5
|
2.8
|
4
|
3.3
|
4.6
|
conspicuity of LMs/all WSs
|
2.6
|
3.9
|
Fig. 1 Conspicuity of single LM with proven superior WS for 80 kVp (25/350) and virtual 120 kVp (75/200) compared to regular abdominal WS (50/350) for each tube voltage.
Abb. 1 Erkennbarkeit einer einzelnen LM mit verbesserten Fensterwerteinstellungen für 80 kV (25/350) und virtuellen 120 kV (75/200) verglichen mit regulären abdominellen Fensterwerteinstellungen (50/350) und jeweiligen Röhrenspannungen.
The best WSs with regard to the conspicuity of LMs (25/350 HU for 80 kVp; 75/500 HU for virtual 120 kVp) were compared to the regular abdominal WSs (50/350 HU for 80 kVp/virtual 120 kVp) ([Table 1], [Fig. 2]). The highest total number of LMs (218; p = 0.01), the mean number of LMs/patient (7.0; p = 0.02), the mean diameter of LMs/patient (20 ± 11mm; p < 0.01), the conspicuity of LMs (1.2; p < 0.01) and the total number of LMs < 1 cm (64; p < 0.01) were better detected with WS 25/350 HU on 80 kVp images. This is in comparison to the best proven WS 75/200 HU on virtual 120 kVp images and the regular WS 50/350 HU on 80 kVp images and virtual 120 kVp images. 80 kVp 50/350 HU images, which represent the regular abdominal image parameters, were superior to virtual 120 kVp 50/350 HU with regard to the total number of LMs, the mean number of LMs/patient and the total number of LMs< 1 cm ([Table 2]).
Table 2
Comparison of 4 image parameter constellations (80 kVp 25/350 HU; 80 kVp 50/350 HU; virtual 120 kVp 75/200 HU; virtual 120 kVp 50/350 HU) with regard to the total number of LMs, mean number of LMs/patient, mean diameter of LMs/patient, total number of LMs < 1 cm, total number of LMs 1 – 2 cm, total number of LMs > 2 cm, conspicuity of LMs.
Tab. 2 Vergleich verschiedener Bildparameterkombinationen (80 kV 25/350 HU; 80 kV 50/350 HU; virtuell 120 kV 75/200 HU; virtuell 120 kV 50/350 HU) unter Berücksichtigung der Gesamtanzahl der LM, durchschnittlichen Anzahl der LM/Patient, durchschnittlichen Durchmesser der LM/ Patient, Gesamtanzahl der LM < 1 cm, Gesamtanzahl der LM 1 – 2 cm, Gesamtanzahl der LM > 2 cm und Erkennbarkeit der LM.
tube voltage
WS
|
80 kVp
25/350 HU
|
80 kVp
50/350 HU
|
virtual 120 kVp
75/200 HU
|
virtual 120 kVp
50/350 HU
|
p-value in multiple comparison
|
total number of LMs
|
218
|
201[1]
|
1981
|
1881
|
< 0.01[2]
|
mean number of LMs/patient
|
7.0
|
6.31
|
6.01
|
5.31
|
< 0.022
|
mean diameter of LMs (mm)
|
20 ± 11
|
23 ± 101
|
24 ± 71
|
26 ± 61
|
< 0.012
|
total number of LMs < 1 cm
|
64
|
531
|
451
|
401
|
< 0.012
|
total number of LMs 1 – 2 cm
|
71
|
68
|
71
|
68
|
< 0.8
|
total number of LMs > 2 cm
|
83
|
80
|
82
|
80
|
< 0.3
|
conspicuity of LMs
|
1.2
|
1.61
|
2.51
|
3.11
|
< 0.012
|
1 Value shows a statistically significant difference in pairwise comparison of protocol “80 kVp 50/350 HU”, “virtual 120 kVp 75/200 HU” and “virtual 120 kVp 50/350 HU” with protocol “80 kVp 25/350 HU” using Steel-Dwass test.
2 Value shows a statistically significant difference compared with the value of four protocols using Kruskal-Wallis test.
Fig. 2 Conspicuity of single LM with 9 different window settings, each for 80 kVp images and virtual 120 kVp images. Superior image quality was seen with the WS 25/350 for 80 kVp images and the WS 75/200 for virtual 120 kVp images.
Abb. 2 Erkennbarkeit einer einzelnen Lebermetastase mit 9 verschiedenen Fensterwerteinstellungen jeweils für 80 kV- und virtuelle 120 kV-Bilddaten. Verbesserte Bildqualität mit WS 25/350 für 80 kV- und WS 75/200 für virtuelle 120 kV-Bilddaten.
The attenuation (HU) of normal liver and aorta as well as the SNR and LLMR differed significantly between 80 kVp images and virtual 120 kVp images (128.1 vs. 93.6; p < 0.05/192.8 vs.131.4; p < 0.05/10.3 vs. 8.1; p < 0.05/2.8 vs. 2.1; p < 0.05) ([Table 3], [4]). The attenuations of LM, muscle, fat and air were not significantly different.
Table 3
Attenuation (HU) of LM, normal liver, muscle, fat, air and aorta depending on tube voltage.
Tab. 3 CT-Dichtewerte (HU) der LM, des gesunden Lebergewebes und von Muskel, Fett, Luft sowie Aorta abhängig von der Röhrenspannung.
|
80 kVp
|
virtual 120 kVp
|
p-value
|
LM
|
45.4 ± 10.2
|
43.2 ± 11.4
|
0.2
|
liver
|
128.1 ± 9.2
|
93.6 ± 8.5
|
< 0.05
|
muscle
|
52.4 ± 14.7
|
49.4 ± 15.9
|
0.09
|
fat
|
–117.3 ± 23.1
|
–104.2 ± 34.6
|
0.06
|
air
|
–1012.4 ± 132.7
|
–1000.5 ± 103.5
|
0.5
|
aorta
|
192.8 ± 23.1
|
131.4 ± 26.3
|
< 0.05
|
Table 4
Background noise (BN), signal (liver)/noise ratio (SNR) and liver/LM ratio depending on tube voltage.
Tab. 4 Hintergrundrauschen (BN), Signal (Leber)/Rausch-Verhältnis (SNR) und Leber/LM-Verhältnis (LLMR) abhängig von der Röhrenspannung.
|
80 kVp
|
virtual 120 kVp
|
p-value
|
BN
|
12.4 ± 4.3
|
11.5 ± 3.2
|
0.4
|
SNR
|
10.3 ± 2.6
|
8.1 ± 3.1
|
< 0.05
|
LLMR
|
2.8 ± 0.8
|
2.1 ± 0.5
|
< 0.05
|
Discussion
We demonstrated with our study that hypovascular liver metastases in patients with colorectal cancer could be significantly better detected using low kVp DECT images compared to 120 kVp virtual DECT images. Robinson et al. concluded in their study (11 patients) that pure 80 kVp data acquired from a dual-source dual-energy MDCT scanner demonstrates greater attenuation differences and improved contrast-to-noise between metastatic disease and normal liver when compared with 120 kVp data [9]. We could show that this increased conspicuity depended not only on a lower tube voltage but also on the selection of special window settings. By comparing 80 kVp images and 120 kVp images, the best image quality was reached with different WSs. For 80 kVp, WS 25/350 proved to be superior and for virtual 120 kVp, WS 75/200 was superior over the other tested window settings. Finally, we could show that the regular abdominal WS 50/350 was inferior to our evaluated new parameters. Higher total and mean numbers of detected liver metastases especially < 1 cm were achieved with these adjusted window settings.
Because low kVp DECT increases image noise, contrast enhancement, and soft tissue contrast, Nakaura et al. hypothesized that wide window width display settings would improve the image quality of low kVp CT. Their results also suggest that wide window width display settings are optimal for the inspection of low kVp scans. A wider window width might reduce the contrast in qualitative analysis because it displays a larger CT number range [7]. Our data suggest that low kVp images require a lower window level and width to reach the highest conspicuity.
Chu et al. showed that DECT has the potential to provide added diagnostic value for the evaluation of focal pancreatic tumors. Virtual native images could replace true native images and may reduce radiation dose and the iodine map, while improving the differentiation between cystic or solid nature of lesions and improving lesion conspicuity compared with simulated weighted 120 kVp images. In their study, the differences in CNR between the images with a weighting factor of 0.3 and the images with a weighting factor of 0.5 or 0.7 were statistically significant (different?) among the cystic lesions and solid hypovascular tumors (p: 0.05) [15]. Therefore, this might be another important parameter in addition to WS with influence on the image quality and conspicuity of hypovascular lesions.
Our study has some limitations. No histological samples of liver metastases were available. However, with the help of pre- and follow-up scans, we were able to perform reliable differentiation in our opinion. Non-cystic lesions with a size increase or decrease, completely new lesions or lesions that were unrecognizable on follow-up scans were rated as LM. The evaluation of LM recognizability was a subjective process, which was offset by a consensus decision of two experienced radiologists. Although a certain body circumference was regarded for the inclusion of patients, we did not consider this parameter in the statistical analysis. The most notable disadvantage of low-voltage CT is the potential for increased image noise (4, 5, 7, 9, 17) resulting from the lower tube energy, which results in unacceptable images in large patients. Beyond a certain patient size, the benefit of increased contrast and increased contrast-to-noise ratio (CNR) at 80 kV is offset by the increased noise and beam-hardening artifacts, thus making abdominal 80 kV single-source and dual-energy CT (using 80 kV in the lower energy tube) inadvisable [16]
[17].
In conclusion, our study demonstrates that DECT is a promising tool for detecting hypovascular liver metastases. The diagnostic value of low kVp image data can be significantly increased by using special window settings.
-
DECT is a promising tool for detecting hypovascular liver metastases.
-
The diagnostic value of low kVp image data can be significantly increased by using special window settings.
-
The use of low kVp images leads to reduced radiation exposure.