Subscribe to RSS
DOI: 10.1055/s-0042-117712
A forward-pulling technique for complicated Trapezoid basket impaction during retrieval of common bile duct stones
Corresponding author
Publication History
Publication Date:
14 October 2016 (online)
Although basket impaction rarely occurs during retrieval of common bile duct stones, it is a potentially serious event [1]. The Trapezoid RX wire-guided retrieval basket (Boston Scientific Corp., Marlborough, Massachusetts, USA) has an emergency release feature, whereby the tip is designed to disengage, thereby facilitating basket removal and reducing the risk of impaction [2]. However, we encountered a rare case of trapezoid basket impaction that was complicated by difficulties in detaching the outer sheath from the basket wires.
A man in his 50s was admitted with multiple common bile duct stones ([Fig. 1]). After endoscopic sphincterotomy and endoscopic papillary large-balloon dilation, mechanical lithotripsy was attempted using a Trapezoid basket. However, basket impaction occurred because the stone was too hard and the tip of the Trapezoid failed to disengage.
We attempted to remove the duodenoscope and outer sheath simultaneously, but were unable to remove the outer sheath at the stomach because it was entwined with the inner basket wires. As a result, only the duodenoscope was removed, following which a forward-viewing endoscope was inserted along the basket catheter. Using rat-tooth forceps, we grasped, twisted, and pulled the basket wires into the working channel of the scope. Next, we simultaneously pushed the scope to the anal side and pulled the outer sheath to the oral side, dissociating it from the basket wires, so that it could be successfully removed ([Video 1]). The impaction was subsequently successfully reversed using an emergency mechanical lithotriptor handle (Olympus Medical Systems, Tokyo, Japan).
Quality:
Therefore, even when using a Trapezoid retrieval basket, basket impaction can occur and an emergency mechanical lithotriptor handle should be available on standby. Additionally, entanglement between the outer sheath and the inner basket wires can occur because the outer sheath of the Trapezoid is thinner than some other mechanical lithotriptors. The technique described here may be useful in such cases.
Endoscopy_UCTN_Code_CPL_1AK_2AC
#
Competing interests: None
-
References
- 1 Chavalitdhamrong D, Donepudi S, Pu L et al. Uncommon and rarely reported adverse events of endoscopic retrograde cholangiopancreatography. Dig Endosc 2014; 26: 15-22
- 2 ASGE Technology Committee . Adler DG, Conway JD et al. Biliary and pancreatic stone extraction devices. Gastrointest Endosc 2009; 70: 603-609
Corresponding author
-
References
- 1 Chavalitdhamrong D, Donepudi S, Pu L et al. Uncommon and rarely reported adverse events of endoscopic retrograde cholangiopancreatography. Dig Endosc 2014; 26: 15-22
- 2 ASGE Technology Committee . Adler DG, Conway JD et al. Biliary and pancreatic stone extraction devices. Gastrointest Endosc 2009; 70: 603-609