Although basket impaction rarely occurs during retrieval of common bile duct stones,
it is a potentially serious event [1]. The Trapezoid RX wire-guided retrieval basket (Boston Scientific Corp., Marlborough,
Massachusetts, USA) has an emergency release feature, whereby the tip is designed
to disengage, thereby facilitating basket removal and reducing the risk of impaction
[2]. However, we encountered a rare case of trapezoid basket impaction that was complicated
by difficulties in detaching the outer sheath from the basket wires.
A man in his 50s was admitted with multiple common bile duct stones ([Fig. 1]). After endoscopic sphincterotomy and endoscopic papillary large-balloon dilation,
mechanical lithotripsy was attempted using a Trapezoid basket. However, basket impaction
occurred because the stone was too hard and the tip of the Trapezoid failed to disengage.
Fig. 1 Endoscopic retrograde cholangiography revealing multiple common bile duct stones.
We attempted to remove the duodenoscope and outer sheath simultaneously, but were
unable to remove the outer sheath at the stomach because it was entwined with the
inner basket wires. As a result, only the duodenoscope was removed, following which
a forward-viewing endoscope was inserted along the basket catheter. Using rat-tooth
forceps, we grasped, twisted, and pulled the basket wires into the working channel
of the scope. Next, we simultaneously pushed the scope to the anal side and pulled
the outer sheath to the oral side, dissociating it from the basket wires, so that
it could be successfully removed ([Video 1]). The impaction was subsequently successfully reversed using an emergency mechanical
lithotriptor handle (Olympus Medical Systems, Tokyo, Japan).
The outer sheath could not be removed as it was entwined with the basket wires. After
a forward-viewing endoscope was inserted along the basket catheter, rat-tooth forceps
were used to grasp and push the basket wires to the anal side, while the outer sheath
was pulled to the oral side; however, adequate traction could not be obtained because
the rat-tooth forceps slipped. The basket wires were therefore grasped, twisted, and
pulled into the working channel of the scope, which was then pushed to the anal side,
while the outer sheath was simultaneously pulled to the oral side (arrows). This successfully
dissociated the outer sheath from the basket wires (arrow), so that it could be removed.
Therefore, even when using a Trapezoid retrieval basket, basket impaction can occur
and an emergency mechanical lithotriptor handle should be available on standby. Additionally,
entanglement between the outer sheath and the inner basket wires can occur because
the outer sheath of the Trapezoid is thinner than some other mechanical lithotriptors.
The technique described here may be useful in such cases.