Breast MRI at Very Short TE (minTE): Image Analysis of minTE Sequences on Non-Fat-Saturated, Subtracted T1-Weighted Images
Brust MRT mit sehr kurzer TE Zeit (minTE): Bildanalyse der minTE Sequenzen auf nicht fettunterdrückten, subtrahierten T1 gewichteten BildernCorrespondence
Publication History
17 January 2016
05 October 2016
Publication Date:
31 January 2017 (online)
Abstract
Purpose The aim was to evaluate a minimum echo time (minTE) protocol for breast magnetic resonance imaging (MRI) in patients with breast lesions compared to a standard TE (nTE) time protocol.
Methods Breasts of 144 women were examined with a 1.5 Tesla MRI scanner. Additionally to the standard gradient-echo sequence with nTE (4.8 ms), a variant with minimum TE (1.2 ms) was used in an interleaved fashion which leads to a better temporal resolution and should reduce the scan time by approximately 50 %. Lesion sizes were measured and the signal-to-noise ratio (SNR) as well as the contrast-to-noise ratio (CNR) were calculated. Subjective confidence was evaluated using a 3-point scale before looking at the nTE sequences (1 = very sure that I can identify a lesion and classify it, 2 = quite sure that I can identify a lesion and classify it, 3 = definitely want to see nTE for final assessment) and the subjective image quality of all examinations was evaluated using a four-grade scale (1 = sharp, 2 = slight blur, 3 = moderate blur and 4 = severe blur/not evaluable) for lesion and skin sharpness. Lesion morphology and contrast enhancement were also evaluated.
Results With minTE sequences, no lesion was rated with “definitely want to see nTE sequences for final assessment”. The difference of the longitudinal and transverse diameter did not differ significantly (p > 0.05). With minTE, lesions and skin were rated to be significantly more blurry (p < 0.01 for lesions and p < 0.05 for skin). There was no difference between both sequences with respect to SNR, CNR, lesion morphology, contrast enhancement and detection of multifocal disease.
Conclusion Dynamic breast MRI with a minTE protocol is feasible without a major loss of information (SNR, CNR, lesion morphology, contrast enhancement and lesion sizes) and the temporal resolution can be increased by a factor of 2 using minTE sequences.
Key points
-
Increase of temporal resolution for a better in-flow curve.
-
Dynamic breast MRI with a shorter TE time is possible without relevant loss of information.
-
Possible decrease of the overall scan time.
Citation Format
-
Wenkel E, Janka R, Geppert C et al. Breast MRI at Very Short TE (minTE): Image Analysis of minTE Sequences on Non-Fat-Saturated, Subtracted T1-Weighted Images. Fortschr Röntgenstr 2017; 189: 137 – 145
#
Zusammenfassung
Einleitung Das Ziel war es in der Brustmagnetresonanztomografie einen Vergleich zwischen zwei Sequenzen mit einer minimalen TE-Zeit (minTE) und mit einer Standard TE Zeit (nTE) bei Patienten mit Brustläsionen anzustellen.
Methoden 144 Frauen wurden mit einem 1,5-Tesla-MRT untersucht. Zusätzlich zu den Standard-Gradienten-Echo-Sequenzen mit normaler TE (4,8 ms) wurde eine Variante mit minimaler TE (1,2 ms) verwendet was zu einer besseren zeitlichen Auflösung führt und die Scanzeit um ca. 50 % reduzieren sollte. Die Läsionsgrößen wurden gemessen und das Signal-Rausch-Verhältnis (SNR) sowie das Kontrast-Rausch-Verhältnis (CNR) wurden berechnet. Die subjektive Diagnosesicherheit wurde mit Hilfe einer 3-Punkte-Skala (1 = sehr sicher, dass ich eine Läsion identifizieren und klassifizieren kann, 2 = ziemlich sicher, dass ich eine Läsion identifizieren und klassifizieren kann, 3 = auf jeden Fall benötigt man eine nTE Sequenz zur abschließenden Bewertung) und die subjektive Bildqualität aller Untersuchungen wurde unter Verwendung einer vierstufigen Skala (1 = scharf, 2 = leichte Unschärfe, 3 = mäßige Unschärfe und 4 = starke Unschärfe / nicht auswertbar) sowohl für die Läsion als auch die Schärfe der Haut evaluiert. Ebenso wurden die Herdbeschaffenheit und die KM-Aufnahme evaluiert.
Ergebnisse Mit minTE Sequenzen wurde keine Läsion mit „auf jeden Fall benötigt man eine nTE Sequenz zur abschließenden Bewertung“ bewertet. Der Längs- und Querdurchmessers unterschied sich nicht signifikant (p > 0,05). Mit minTE Sequenzen wurden Läsionen und Haut deutlich verschwommener bewertet (p < 0,01 für Läsionen und p < 0,05 für Haut). Es gab keinen signifikanten Unterschied zwischen den beiden Sequenzen in SNR, CNR, Herdbeschaffenheit, KM-Aufnahme und in der Erkennung von Multifokalität.
Zusammenfassung Ein dynamisches Brust-MRT ist mit minTE Sequenzen ohne größeren Informationsverlust durchführbar (SNR, CNR, Herdbeschaffenheit, KM-Aufnahme und Läsionsgröße) und die zeitliche Auflösung kann durch minTE Sequenzen um einen Faktor 2 erhöht werden.
Kernaussagen
-
Erhöhung der zeitlichen Auflösung für eine bessere KM-Anflutungskurve.
-
Dynamische Brust-MRT mit einer kurzen TE Zeit ist ohne relevanten Informationsverlust möglich.
-
Mögliche Reduktion der gesamten Scandauer.
#
Introduction
Magnetic resonance imaging (MRI) has become an accepted method in breast imaging, joining mammography and ultrasound. Several indications are widely accepted for its use, like screening women with a high risk of breast cancer, occult primary breast cancer or searching for recurrent disease in inconclusive mammography and ultrasound [1] [2] [3] [4] [5] [6] [7] [8] [9]. The role of MRI in the preoperative staging of tumor extent is still under debate [10] [11]. MRI image analysis is based on lesion enhancement patterns in dynamic breast MRI and on morphologic changes [12] [13] [14] [15] [16] [17] [18] [19]. With these criteria, breast MRI has a sensitivity of about 85 % to 99 % in detecting malignant breast lesions, as has been shown in several studies [1] [2] [16] [17] [18] [19] [20] [21] [22] [23] [24]. Recently, the implementation of MRI exams in regular screening settings has been discussed as well [19] [25] [26]. If MRI of the breast is to be used under screening conditions, the shortening of protocols would be preferable. In non-fat-suppressed dynamic breast imaging, it is a well-accepted recommendation to acquire data at or close to echo times that fulfill the in-phase condition for fat and water, such as 4.8 ms at 1.5 Tesla, in order to avoid chemical shift artifacts that lead to signal cancellation at fat/water interfaces. This effect was described and analyzed, e. g., by Bedrosian et al.[1] and Fischer et al. [1] [4] [27] with the recommendation of using either in-phase echo time (TE) or “TE less than 1.2 ms” resulting in a phase difference of below 90° [28]. With otherwise comparable parameter settings, this would result in a decrease of 50 % of the acquisition time due to the resulting shorter repetition time (TR) of approximately 8 ms vs. 4 ms. Meanwhile, with current gradient systems and fast imaging sequences (now rapid changes of the magnetic gradient due to a better scanner geometry and improved materials are possible and better coils lead to an improved signal) [28] [29] [30] as well as better software-based post-processing, it has become possible to achieve such short echo times without compromising the matrix size or the bandwidth. In this study, we have set up an interleaved protocol approach to achieve a direct comparison of a minimum TE (minTE) vs. standard TE (nTE) acquisition within a clinical standard protocol. The purpose of our study was to evaluate the image quality on non-fat-saturated, subtracted T1-weighted images of the minTE acquisition compared to conventional standard TE protocols (nTE) in patients with breast lesions. To our knowledge, this is the first study in which the feasibility of minTE breast MRI has been evaluated in a prospective manner.
#
Methods
Patients
The study was approved by the local institutional review board and all patients gave written consent to the study and the MRI exam. The study complies with the Declaration of Helsinki. The parameters of the study sequence ranged within the limits of breast MRI performance guidelines [31] [32].
From January 2014 to July 2014, 231 women underwent breast MRI at our institution for various indications. A total of 144 women were randomly scheduled for the scanner with the additional minTE sequences dependent on scanner slot availability. MRI indications for these women were: to exclude preoperatively multifocal disease (n = 11) of histologically verified breast cancer, positive family history for breast cancer (n = 40), to exclude recurrent breast cancer (n = 39), monitoring neo-adjuvant chemotherapy (n = 2), unclear findings in mammography (BI-RADS 0), ultrasound or clinical breast complaints (n = 47), and cancer of unknown primary (n = 5).
#
MRI protocol
Data were acquired on a 1.5 Tesla MRI scanner (MAGNETOM Avanto, Siemens, Erlangen, Germany) using a bilateral dedicated phased-array breast coil (4-channel breast array coil, Siemens, Erlangen, Germany). Besides the dynamic T1 protocol, an axial 2 D T2-weighted turbo spin echo (TSE) pulse sequence (TR/TE, 5250/113 ms, FOV 380 × 380 mm, matrix 512 × 358, slice thickness 3 mm, in-plane resolution 0.4 × 0.4 mm) and one pre- and one post-dynamic contrast sagittal 3 D T1-weighted spoiled gradient echo pulse sequence (TR/TE, 21/4.8 ms, FOV 380 × 380 mm, matrix 512 × 512, slice thickness 2 to 2.5 mm, in-plane resolution 0.4 × 0.4 mm) for morphologic details were used. The dynamic sequence was a bilateral, axial 3 D spoiled gradient echo pulse sequence (TR/TE, 7.3/4.8 ms, FOV 340 × 340 mm, matrix 512 × 430, slice thickness 2 mm, in-plane resolution 0.8 × 0.8 mm) and consisted of one pre- and four post-contrast repetitions of the whole volume of both breasts. Contrast agent (Gadobutrol, Bayer, Leverkusen, Germany, 0.1 mmol per kilogram body weight) was applied after the first dynamic acquisition with an MR-compatible power injector (Spectris; Medrad, Pittsburgh, PA, USA) with a flow of 1 ml/sec followed by a 20 ml saline flush.
In addition to the “normal” standard TE of 4.8 ms (nTE), a sequence with a “minimal” TE of 1.2 ms (minTE) was used ([Table 1]). The very short TE of 1.2 ms was achieved by a gradient-echo sequence (VIBE) that allows asymmetric readout windows. Thus, only an adjustable percentage of the gradient echo is sampled. The clinical reference protocol with nTE was set up in a way that all scan parameters were identical with the minTE protocol, except for the echo asymmetry and the flip angle, which were adapted to the TR. The spatial resolution was 0.8 × 0.8 × 2 mm3 with a bandwidth of 445 Hz; parallel imaging was used (GRAPPA acceleration factor of 2). This resulted in an acquisition time of 60 s for the nTE scans and 30 s for the minTE scans, thus the resulting temporal resolution for both techniques was 90 s. [Fig. 1] shows the interleaved scan protocol, which allows a direct comparison and [Fig. 2] shows an example of the complete protocol.
Post-processing included subtractions of the second post-enhanced sequence from the sequence without contrast medium using the scanner software.
#
Image analysis
The breast MRI images were evaluated in consensus by one senior breast radiologist with more than 10 years of experience in breast MRI and one radiologist with 2 years of experience in breast and MR imaging. For the comparison of the standard TE sequence (nTE) and the minimum TE sequence (minTE), only examinations with identifiable enhancing lesions in the contrast-enhanced sequences were chosen. The study evaluation was performed on the second post-enhanced sequence. Dignity of the lesions was either confirmed by core or excisional biopsy or on the basis of the clinical examination, ultrasound, mammography and MR findings.
-
Image quality was assessed on standard PACS workstations on images in the axial plane. For each examination the regions of interest (ROI) were placed in the enhancing lesion, in the pectoral muscle and in the air between the right and left breast. The mean signal intensity and the standard deviation were obtained for each measurement. Standard deviation of the air was considered as image noise. Signal-to-noise ratio (SNR = mean density/standard deviation of image noise) and contrast-to-noise ratio (CNR = (mean density of the lesion – mean density of the pectoral muscle)/standard deviation of image noise) were calculated for the ROIs of the enhancing lesions and the pectoral muscle in the native and in the second post-contrast sequence of both the nTE and minTE dynamic series.
-
The lesion size (longitudinal and transverse diameter) was measured in millimeters.
-
The subjective confidence of the minTE sequences was evaluated on a subjective 3-point scale before looking at the nTE sequences (1 = very sure that I can identify a lesion and classify it, 2 = quite sure that I can identify a lesion and classify it, 3 = definitely want to see nTE for final assessment).
-
The subjective image quality of all examinations (nTE and minTE sequences) was evaluated using a subjective four-grade scale (1 = sharp, 2 = slight blur, 3 = moderate blur and 4 = severe blur/not evaluable) for lesion (= lesion surrounding) and skin sharpness (= skin evaluation).
-
Multifocal disease was detected on both sequences (more than one suspicious lesion: yes or no).
-
To describe single lesions, we used the ACR BI-RADS® Atlas (fifth edition 2013). Initially we distinguished the lesions between focus (less than 5 mm in diameter) and mass/non-mass enhancement. In masses, the shape (oval, round, irregular), the margins (circumscribed, not circumscribed (irregular/spiculated)) and the internal enhancement patterns (homogeneous, heterogeneous, rim enhancement, dark internal septations) were evaluated. In non-mass enhancement we checked for the distribution (focal, linear, segmental, regional, multiple regions, diffuse) and for the internal enhancement patterns (homogeneous, heterogeneous, clumped, clustered ring). Histologically proven multifocality was evaluated for both the nTE and the minTE sequences.
#
Statistical analysis
Statistics were calculated using Excel 2007 (Microsoft, USA) and Graph-Pad Prism 4.03, 2005 (Graph-Pad Software, San Diego, CA). The difference in lesion surroundings and skin evaluation was evaluated with the Wilcoxon signed rank test and the signal intensity in both nTE and minTE sequences was evaluated with the paired t-test. A p-value < 0.05 was considered as significant, and a p-value < 0.01 was considered as highly significant.
#
#
Results
Patients
A total of 144 MRI exams with minTE sequences were performed. 121 patients did not require histological interventions (78 patients with BI-RADS 1, 43 patients with BI-RADS 2). A total of 46 lesions were identified in the MRI examinations in 23 patients in 26 breasts.
MRI indications for these 23 patients were multifocal breast cancer (n = 11); family history of breast cancer (n = 4); unclear findings in mammography (BI-RADS 0), ultrasound or clinical breast complaints (n = 6); and exclusion of local recurrence (n = 2). Histology revealed 32 malignant and 14 benign lesions ([Table 2]).
invasive ductal carcinoma (IDC), invasive lobular carcinoma (ILC), ductal carcinoma in-situ (DCIS), fibrocystic disease (fibrocy. dis.), fibroadenoma (FA), atypical epithelial proliferation (atyp. prol.), papillomatosis (pap.), inflammatory cyst (infl. cyst).
Comparison of patient images is presented in [Fig. 3], [4], [5], [6].
The median age of all patients included in the study was 47.6 years (range: 22 to 72 years).
#
ROI measurements
The mean SNR of the lesions before contrast medium was 22.5 ± 9.1 in minTE and 23.3 ± 7.6 in nTE sequences (p = 0.6) and after contrast medium 42.5 ± 13.1 in minTE and 41.4 ± 14.4 in nTE sequences (p = 0.45). The mean CNR of the lesions was 42.8 ± 23.6 in minTE and 47.6 ± 21.8 in nTE sequences (p = 0.17) ([Fig. 7]).
#
Lesion size
The mean longitudinal diameter was 1.8 cm (range: 0.3 to 8.6 cm) and the transverse diameter was 1.4 cm (range: 0.3 to 6.8 cm) in the nTE (4.8 ms) sequences. In the minTE (1.2 ms) sequences, the mean longitudinal diameter was 1.8 cm (range: 0.3 to 8.4 cm) and the transverse diameter was 1.4 cm (range: 0.3 to 6.9 cm). The difference was not significant using the paired t-test (p = 0.96 for longitudinal diameter and p = 0.99 for transverse diameter).
#
Subjective confidence
With the minTE sequence, 26 lesions were rated with 1 = very sure that I can identify a lesion and classify it and 20 lesions with 2 = quite sure. No lesion was rated with 3 = definitely want to see nTE sequences for final assessment.
#
Subjective image quality (lesion surroundings, skin evaluation) and multifocality
In minTE sequences, lesions and the skin were rated to be significantly more blurry than in the normal TE sequences with a p-value of 0.0003 (lesions) and 0.047 (skin); see [Table 3], [4]. Nevertheless, neither the skin nor the lesion was rated as having severe blurring in nTE or in minTE sequences. There was no difference between the nTE and the minTE sequences in the detection of histologically confirmed multifocal disease.
Using the Wilcoxon signed rank test, there was a significant difference between the two types of sequences (p < 0.01). Relative frequency is shown in brackets [..].
Mit Hilfe des Wilcoxon Rangsummentest konnte ein signifikanter Unterschied festgestellt werden (p < 0.01). Die relative Häufigkeit wird in den eckigen Klammern angegeben [..].
Using the Wilcoxon signed rank test, there was a significant difference between the two types of sequences (p < 0.05). Relative frequency is shown in brackets [..].
Mit Hilfe des Wilcoxon Rangsummentest konnte ein signifikanter Unterschied festgestellt werden (p < 0.05). Die relative Häufigkeit wird in den eckigen Klammern angegeben [..].
#
Lesion characterization
7 lesions were rated as a focus in nTE and in minTE sequences. No non-mass enhancement was seen in both sequences. Therefore, 39 lesions were evaluated as mass enhancement. In both sequences the shape of this mass enhancement was rated as round in 10 lesions, as oval in 6 lesions and as irregular in 23 lesions. There was also no difference in the evaluation of the margins (12 × circumscribed, 13 × not circumscribed (irregular), 14 × not circumscribed (spiculated)) and the internal enhancement (11 × homogenous, 25 × heterogeneous, 1 × rim enhancement, 2 × dark internal septations). Overall no difference in lesion characterization was detected between nTE and minTE sequences.
#
#
Discussion and conclusion
In our study we compared the subtracted images of two different TE sequences without fat saturation, a minTE sequence (TE 1.2 ms) and a normal TE sequence (TE 4.8 ms), for the enhancement kinetics. The study suggests that visualizing breast disease and its extent is feasible with a minTE MRI sequence. Compared to the classic approach, there was no significant difference in the detection, morphology and size determination of breast lesions. In our approach we used a TE of 1.2 ms, which reduced the scan time per time point by about 50 % with no significant SNR or CNR changes of the lesions. This short TE did not lead to false-negative or false-positive findings. As there was no significant difference in the longitudinal and transverse lesion diameter, the minTE sequence should not lead to a clinically important under- or overestimation of the lesion size. The drawback of the minTE measurement is an increase in skin and lesion blurring, which can be explained by the asymmetric readout that does not detect the full echo signal. Therefore, there could be a loss of signal information. Opposed phase cancellation might also play a role in small lesions [28] [33]. Also in the literature ultrashort TE sequences acquire an echo signal from the central to the outer parts of k-space with ramp sampling and therefore are very sensitive to small k-space trajectory errors [34]. There was no significant difference in the detection of multifocal disease in minTE and nTE. We were also able to detect tumor vessels in both sequences which is important because the adjacent vessel sign was significantly associated with malignancy [35] [36].
To our knowledge, no breast MRI study has dealt with short TE sequences before. There were some studies that showed the feasibility of ultrashort TE sequences for the clinical routine but, unlike our study, the TE time ranged between 0.07 ms and 0.14 ms. For example, Robson et al. showed that contrast enhancement can be identified in tissues using ultrashort TE pulse sequences (TE of 0.08 ms) [37] or the periosteum can be visualized with ultrashort echo time pulse sequences in health and disease [38]. In addition, ultrashort TE pulse sequences provided anatomical detail not apparent with conventional sequences and showed patterns of both increased and decreased enhancement in tendinopathy [39]. However, in contrast to our study, they did not use a dynamic protocol. Yamashita et al. demonstrated that with a 3.0 Tesla MRI scanner the middle ear ossicles could be clearly visualized on short TE images, while they were not visible in long TE images [40]. In contrast to this study, we used a 1.5 Tesla MRI scanner and were not able to find a difference in lesion detection in breast imaging.
Recently, it was reported that the maximum curve slope in an ultrafast protocol performed even better in the differentiation of benign and malignant lesions compared to BI-RADS curve types [25]. This study reported that the initial phase of the enhancement curve was more important compared to late-phase characteristics. Typically, such ultrafast protocols are realized using view-sharing protocols that lead to temporal blurring, while in this study, all time points are acquired separately. The minTE technique allows acquisition of more time points in dynamic scans because minTE last 30 s and nTE last 60 s (therefore two minTE sequences are acquired in the time it takes to acquire one nTE sequence). Therefore, this could lead to more detailed dynamic curves and/or to a shortened protocol so that one option for minTE application could be a short-scan protocol with few minTE dynamic time points for lesion detection and to perform the dynamic early curve-phase evaluation. This is a possible scenario for a screening population in which detection is more important than detailed morphologic lesion characterization [26]. For more detailed morphologic information, a high-resolution non-dynamic sequence could be added to the scan protocol. Mango et al. showed that an abbreviated breast MRI protocol allows detection of breast carcinoma with only one pre- and post-contrast T1-weighted sequence [41]. Whether a shortening of the examination time with minTE also shows this effect needs to be clarified in a further study.
There are some limitations to our study. The number of individuals undergoing breast MRI and showing a suspicious lesion was small. We only evaluated the subtracted images and not the kinetic curves because the different TE sequences were acquired consecutively and therefore would result in slightly different kinetic curves. The impact of more measurements on the kinetic curve has to be evaluated in the future in a separate study. Lesion detection, classification and time curves are not only dependent on sequence parameters but also on the contrast medium used [42]. We used just one type of contrast medium because the comparison of different contrast mediums was not part of the study. The diagnoses of suspicious lesions smaller than 5 mm were all made in breasts with another obvious pathological lesion. Opposed-phase lesion cancellation might play a role in affecting curve shape in small lesions, but this was not part of the study. Malignant lesions sometimes show central necrosis and the rim of the lesion may be too thin for correct ROI placement. In our study we evaluated one rim-enhancing cyst where it was impossible to settle a correct ROI, but the enhancing cyst could be detected in both sequences, and it was rated as suspicious in both sequences. Interestingly there was no non-mass enhancement in our study despite some findings as DCIS usually shows non-mass enhancement. We can only speculate about this finding. Maybe the number of patients was too small in this study. Nevertheless no non-mass enhancement was seen in both sequences and so there was no difference between these two sequences. We focused our study on the image quality of subtracted images. Therefore, there could be a bias in the evaluation of the skin and lesion surroundings due to motion artifacts. Nevertheless, motion artifacts were in both sequences, and therefore should not be statistically relevant. Another limitation was that a subjective scale was used for image analysis, but images were evaluated in consensus by two breast radiologist.
In conclusion, dynamic breast MRI with a shorter TE time than the in-phase condition (nTE) is possible. Shorter TE times increase the temporal resolution, which leads to a better in-flow curve and might decrease the overall scan time.
-
Increase of the temporal resolution for a better in-flow curve
-
Dynamic breast MRI with a shorter TE time than the in-phase condition is possible
-
Possible decrease of the overall scan time
#
#
-
References
- 1 Bedrosian I. Mick R. Orel SG. et al. Changes in the surgical management of patients with breast carcinoma based on preoperative magnetic resonance imaging. Cancer 2003; 98: 468-473
- 2 Bluemke DA. Gatsonis CA. Chen MH. et al. Magnetic resonance imaging of the breast prior to biopsy. Jama 2004; 292: 2735-2742
- 3 Duygulu G. Oktay A. Bilgen IG. et al. The role of breast MRI in planning the surgical treatment of breast cancer. Diagnostic and interventional radiology 2012; 18: 460-467
- 4 Fischer U. Kopka L. Grabbe E. Breast carcinoma: effect of preoperative contrast-enhanced MR imaging on the therapeutic approach. Radiology 1999; 213: 881-888
- 5 Schnall M. MR imaging evaluation of cancer extent: is there clinical relevance?. Magnetic resonance imaging clinics of North America 2006; 14: 379-381, vii
- 6 Schnall M. Orel S. Breast MR imaging in the diagnostic setting. Magnetic resonance imaging clinics of North America 2006; 14: 329-337, vi
- 7 Young P. Kim B. Malin JL. Preoperative breast MRI in early-stage breast cancer. Breast cancer research and treatment 2012; 135: 907-912
- 8 Dietzel M. Zoubi R. Burmeister HP. et al. Combined staging at one stop using MR mammography: evaluation of an extended protocol to screen for distant metastasis in primary breast cancer – initial results and diagnostic accuracy in a prospective study. RoFo: Fortschritte auf dem Gebiete der Rontgenstrahlen und der Nuklearmedizin 2012; 184: 618-623
- 9 Spick C. Szolar DH. Preidler KW. et al. Breast MRI used as a problem-solving tool reliably excludes malignancy. European journal of radiology 2015; 84: 61-64
- 10 Pediconi F. Padula S. Dominelli V. et al. Role of breast MR imaging for predicting malignancy of histologically borderline lesions diagnosed at core needle biopsy: prospective evaluation. Radiology 2010; 257: 653-661
- 11 Pediconi F. Miglio E. Telesca M. et al. Effect of preoperative breast magnetic resonance imaging on surgical decision making and cancer recurrence rates. Investigative radiology 2012; 47: 128-135
- 12 Freed M. Effect of protocol parameters on contrast agent washout curve separability in breast dynamic contrast enhanced MRI: a simulation study. Magnetic resonance in medicine 2012; 68: 516-522
- 13 Heller SL. Moy L. Lavianlivi S. et al. Differentiation of malignant and benign breast lesions using magnetization transfer imaging and dynamic contrast-enhanced MRI. Journal of magnetic resonance imaging: JMRI 2013; 37: 138-145
- 14 Jansen SA. Fan X. Karczmar GS. et al. Differentiation between benign and malignant breast lesions detected by bilateral dynamic contrast-enhanced MRI: a sensitivity and specificity study. Magnetic resonance in medicine 2008; 59: 747-754
- 15 Jansen SA. Shimauchi A. Zak L. et al. Kinetic curves of malignant lesions are not consistent across MRI systems: need for improved standardization of breast dynamic contrast-enhanced MRI acquisition. American journal of Roentgenology 2009; 193: 832-839
- 16 Schnall MD. Blume J. Bluemke DA. et al. Diagnostic architectural and dynamic features at breast MR imaging: multicenter study. Radiology 2006; 238: 42-53
- 17 Szabo BK. Aspelin P. Wiberg MK. et al. Dynamic MR imaging of the breast. Analysis of kinetic and morphologic diagnostic criteria. Acta radiologica 2003; 44: 379-386
- 18 Macura KJ. Ouwerkerk R. Jacobs MA. et al. Patterns of enhancement on breast MR images: interpretation and imaging pitfalls. Radiographics: a review publication of the Radiological Society of North America, Inc 2006; 26: 1719-1734 ; quiz 1719
- 19 Dietzel M. Baltzer PA. Vag T. et al. Potential of MR mammography to predict tumor grading of invasive breast cancer. RoFo: Fortschritte auf dem Gebiete der Rontgenstrahlen und der Nuklearmedizin 2011; 183: 826-833
- 20 Heywang-Kobrunner SH. Bick U. Bradley Jr WG. et al. International investigation of breast MRI: results of a multicentre study (11 sites) concerning diagnostic parameters for contrast-enhanced MRI based on 519 histopathologically correlated lesions. European radiology 2001; 11: 531-546
- 21 Kaiser WA. Magnetic resonance tomography of the breast. The results of 253 examinations. Deutsche medizinische Wochenschrift 1989; 114: 1351-1357
- 22 Kinkel K. Helbich TH. Esserman LJ. et al. Dynamic high-spatial-resolution MR imaging of suspicious breast lesions: diagnostic criteria and interobserver variability. American journal of Roentgenology 2000; 175: 35-43
- 23 Kuhl CK. Mielcareck P. Klaschik S. et al. Dynamic breast MR imaging: are signal intensity time course data useful for differential diagnosis of enhancing lesions?. Radiology 1999; 211: 101-110
- 24 Wiener JI. Schilling KJ. Adami C. et al. Assessment of suspected breast cancer by MRI: a prospective clinical trial using a combined kinetic and morphologic analysis. American journal of Roentgenology 2005; 184: 878-886
- 25 Mann RM. Mus RD. van Zelst J. et al. A novel approach to contrast-enhanced breast magnetic resonance imaging for screening: high-resolution ultrafast dynamic imaging. Investigative radiology 2014; 49: 579-585
- 26 Kuhl CK. Schrading S. Strobel K. et al. Abbreviated breast magnetic resonance imaging (MRI): first postcontrast subtracted images and maximum-intensity projection-a novel approach to breast cancer screening with MRI. Journal of clinical oncology: official journal of the American Society of Clinical Oncology 2014; 32: 2304-2310
- 27 Heywang-Kobrunner SH. Wolf HD. Deimling M. et al. Misleading changes of the signal intensity on opposed-phase MRI after injection of contrast medium. Journal of computer assisted tomography 1996; 20: 173-178
- 28 Robson MD. Gatehouse PD. Bydder M. et al. Magnetic resonance: an introduction to ultrashort TE (UTE) imaging. Journal of computer assisted tomography 2003; 27: 825-846
- 29 Tyler DJ. Robson MD. Henkelman RM. et al. Magnetic resonance imaging with ultrashort TE (UTE) PULSE sequences: technical considerations. Journal of magnetic resonance imaging: JMRI 2007; 25: 279-289
- 30 O'Brien KR. Myerson SG. Cowan BR. et al. Phase contrast ultrashort TE: A more reliable technique for measurement of high-velocity turbulent stenotic jets. Magnetic resonance in medicine 2009; 62: 626-636
- 31 Mann RM. Kuhl CK. Kinkel K. et al. Breast MRI: guidelines from the European Society of Breast Imaging. European radiology 2008; 18: 1307-1318
- 32 Sardanelli F. Boetes C. Borisch B. et al. Magnetic resonance imaging of the breast: recommendations from the EUSOMA working group. European journal of cancer 2010; 46: 1296-1316
- 33 Robson MD. Tyler DJ. Neubauer S. Ultrashort TE chemical shift imaging (UTE-CSI). Magnetic resonance in medicine 2005; 53: 267-274
- 34 Takizawa M. Hanada H. Oka K. et al. A robust ultrashort TE (UTE) imaging method with corrected k-space trajectory by using parametric multiple function model of gradient waveform. IEEE transactions on medical imaging 2013; 32: 306-316
- 35 Dietzel M. Baltzer PA. Vag T. et al. The adjacent vessel sign on breast MRI: new data and a subgroup analysis for 1084 histologically verified cases. Korean journal of radiology 2010; 11: 178-186
- 36 Fischer DR. Malich A. Wurdinger S. et al. The adjacent vessel on dynamic contrast-enhanced breast MRI. American journal of Roentgenology 2006; 187: W147-W151
- 37 Robson MD. Gatehouse PD. So PW. et al. Contrast enhancement of short T2 tissues using ultrashort TE (UTE) pulse sequences. Clinical radiology 2004; 59: 720-726
- 38 Reichert IL. Benjamin M. Gatehouse PD. et al. Magnetic resonance imaging of periosteum with ultrashort TE pulse sequences. Journal of magnetic resonance imaging: JMRI 2004; 19: 99-107
- 39 Robson MD. Benjamin M. Gishen P. et al. Magnetic resonance imaging of the Achilles tendon using ultrashort TE (UTE) pulse sequences. Clinical radiology 2004; 59: 727-735
- 40 Yamashita K. Yoshiura T. Hiwatashi A. et al. Ultrashort echo time imaging of normal middle ear ossicles: a feasibility study. Dento maxillo facial radiology 2012; 41: 601-604
- 41 Mango VL. Morris EA. Dershaw D. et al. Abbreviated protocol for breast MRI: are multiple sequences needed for cancer detection?. European journal of radiology 2015; 84: 65-70
- 42 Fallenberg EM. Renz DM. Karle B. et al. Intraindividual, randomized comparison of the macrocyclic contrast agents gadobutrol and gadoterate meglumine in breast magnetic resonance imaging. European radiology 2015; 25: 837-849
Correspondence
-
References
- 1 Bedrosian I. Mick R. Orel SG. et al. Changes in the surgical management of patients with breast carcinoma based on preoperative magnetic resonance imaging. Cancer 2003; 98: 468-473
- 2 Bluemke DA. Gatsonis CA. Chen MH. et al. Magnetic resonance imaging of the breast prior to biopsy. Jama 2004; 292: 2735-2742
- 3 Duygulu G. Oktay A. Bilgen IG. et al. The role of breast MRI in planning the surgical treatment of breast cancer. Diagnostic and interventional radiology 2012; 18: 460-467
- 4 Fischer U. Kopka L. Grabbe E. Breast carcinoma: effect of preoperative contrast-enhanced MR imaging on the therapeutic approach. Radiology 1999; 213: 881-888
- 5 Schnall M. MR imaging evaluation of cancer extent: is there clinical relevance?. Magnetic resonance imaging clinics of North America 2006; 14: 379-381, vii
- 6 Schnall M. Orel S. Breast MR imaging in the diagnostic setting. Magnetic resonance imaging clinics of North America 2006; 14: 329-337, vi
- 7 Young P. Kim B. Malin JL. Preoperative breast MRI in early-stage breast cancer. Breast cancer research and treatment 2012; 135: 907-912
- 8 Dietzel M. Zoubi R. Burmeister HP. et al. Combined staging at one stop using MR mammography: evaluation of an extended protocol to screen for distant metastasis in primary breast cancer – initial results and diagnostic accuracy in a prospective study. RoFo: Fortschritte auf dem Gebiete der Rontgenstrahlen und der Nuklearmedizin 2012; 184: 618-623
- 9 Spick C. Szolar DH. Preidler KW. et al. Breast MRI used as a problem-solving tool reliably excludes malignancy. European journal of radiology 2015; 84: 61-64
- 10 Pediconi F. Padula S. Dominelli V. et al. Role of breast MR imaging for predicting malignancy of histologically borderline lesions diagnosed at core needle biopsy: prospective evaluation. Radiology 2010; 257: 653-661
- 11 Pediconi F. Miglio E. Telesca M. et al. Effect of preoperative breast magnetic resonance imaging on surgical decision making and cancer recurrence rates. Investigative radiology 2012; 47: 128-135
- 12 Freed M. Effect of protocol parameters on contrast agent washout curve separability in breast dynamic contrast enhanced MRI: a simulation study. Magnetic resonance in medicine 2012; 68: 516-522
- 13 Heller SL. Moy L. Lavianlivi S. et al. Differentiation of malignant and benign breast lesions using magnetization transfer imaging and dynamic contrast-enhanced MRI. Journal of magnetic resonance imaging: JMRI 2013; 37: 138-145
- 14 Jansen SA. Fan X. Karczmar GS. et al. Differentiation between benign and malignant breast lesions detected by bilateral dynamic contrast-enhanced MRI: a sensitivity and specificity study. Magnetic resonance in medicine 2008; 59: 747-754
- 15 Jansen SA. Shimauchi A. Zak L. et al. Kinetic curves of malignant lesions are not consistent across MRI systems: need for improved standardization of breast dynamic contrast-enhanced MRI acquisition. American journal of Roentgenology 2009; 193: 832-839
- 16 Schnall MD. Blume J. Bluemke DA. et al. Diagnostic architectural and dynamic features at breast MR imaging: multicenter study. Radiology 2006; 238: 42-53
- 17 Szabo BK. Aspelin P. Wiberg MK. et al. Dynamic MR imaging of the breast. Analysis of kinetic and morphologic diagnostic criteria. Acta radiologica 2003; 44: 379-386
- 18 Macura KJ. Ouwerkerk R. Jacobs MA. et al. Patterns of enhancement on breast MR images: interpretation and imaging pitfalls. Radiographics: a review publication of the Radiological Society of North America, Inc 2006; 26: 1719-1734 ; quiz 1719
- 19 Dietzel M. Baltzer PA. Vag T. et al. Potential of MR mammography to predict tumor grading of invasive breast cancer. RoFo: Fortschritte auf dem Gebiete der Rontgenstrahlen und der Nuklearmedizin 2011; 183: 826-833
- 20 Heywang-Kobrunner SH. Bick U. Bradley Jr WG. et al. International investigation of breast MRI: results of a multicentre study (11 sites) concerning diagnostic parameters for contrast-enhanced MRI based on 519 histopathologically correlated lesions. European radiology 2001; 11: 531-546
- 21 Kaiser WA. Magnetic resonance tomography of the breast. The results of 253 examinations. Deutsche medizinische Wochenschrift 1989; 114: 1351-1357
- 22 Kinkel K. Helbich TH. Esserman LJ. et al. Dynamic high-spatial-resolution MR imaging of suspicious breast lesions: diagnostic criteria and interobserver variability. American journal of Roentgenology 2000; 175: 35-43
- 23 Kuhl CK. Mielcareck P. Klaschik S. et al. Dynamic breast MR imaging: are signal intensity time course data useful for differential diagnosis of enhancing lesions?. Radiology 1999; 211: 101-110
- 24 Wiener JI. Schilling KJ. Adami C. et al. Assessment of suspected breast cancer by MRI: a prospective clinical trial using a combined kinetic and morphologic analysis. American journal of Roentgenology 2005; 184: 878-886
- 25 Mann RM. Mus RD. van Zelst J. et al. A novel approach to contrast-enhanced breast magnetic resonance imaging for screening: high-resolution ultrafast dynamic imaging. Investigative radiology 2014; 49: 579-585
- 26 Kuhl CK. Schrading S. Strobel K. et al. Abbreviated breast magnetic resonance imaging (MRI): first postcontrast subtracted images and maximum-intensity projection-a novel approach to breast cancer screening with MRI. Journal of clinical oncology: official journal of the American Society of Clinical Oncology 2014; 32: 2304-2310
- 27 Heywang-Kobrunner SH. Wolf HD. Deimling M. et al. Misleading changes of the signal intensity on opposed-phase MRI after injection of contrast medium. Journal of computer assisted tomography 1996; 20: 173-178
- 28 Robson MD. Gatehouse PD. Bydder M. et al. Magnetic resonance: an introduction to ultrashort TE (UTE) imaging. Journal of computer assisted tomography 2003; 27: 825-846
- 29 Tyler DJ. Robson MD. Henkelman RM. et al. Magnetic resonance imaging with ultrashort TE (UTE) PULSE sequences: technical considerations. Journal of magnetic resonance imaging: JMRI 2007; 25: 279-289
- 30 O'Brien KR. Myerson SG. Cowan BR. et al. Phase contrast ultrashort TE: A more reliable technique for measurement of high-velocity turbulent stenotic jets. Magnetic resonance in medicine 2009; 62: 626-636
- 31 Mann RM. Kuhl CK. Kinkel K. et al. Breast MRI: guidelines from the European Society of Breast Imaging. European radiology 2008; 18: 1307-1318
- 32 Sardanelli F. Boetes C. Borisch B. et al. Magnetic resonance imaging of the breast: recommendations from the EUSOMA working group. European journal of cancer 2010; 46: 1296-1316
- 33 Robson MD. Tyler DJ. Neubauer S. Ultrashort TE chemical shift imaging (UTE-CSI). Magnetic resonance in medicine 2005; 53: 267-274
- 34 Takizawa M. Hanada H. Oka K. et al. A robust ultrashort TE (UTE) imaging method with corrected k-space trajectory by using parametric multiple function model of gradient waveform. IEEE transactions on medical imaging 2013; 32: 306-316
- 35 Dietzel M. Baltzer PA. Vag T. et al. The adjacent vessel sign on breast MRI: new data and a subgroup analysis for 1084 histologically verified cases. Korean journal of radiology 2010; 11: 178-186
- 36 Fischer DR. Malich A. Wurdinger S. et al. The adjacent vessel on dynamic contrast-enhanced breast MRI. American journal of Roentgenology 2006; 187: W147-W151
- 37 Robson MD. Gatehouse PD. So PW. et al. Contrast enhancement of short T2 tissues using ultrashort TE (UTE) pulse sequences. Clinical radiology 2004; 59: 720-726
- 38 Reichert IL. Benjamin M. Gatehouse PD. et al. Magnetic resonance imaging of periosteum with ultrashort TE pulse sequences. Journal of magnetic resonance imaging: JMRI 2004; 19: 99-107
- 39 Robson MD. Benjamin M. Gishen P. et al. Magnetic resonance imaging of the Achilles tendon using ultrashort TE (UTE) pulse sequences. Clinical radiology 2004; 59: 727-735
- 40 Yamashita K. Yoshiura T. Hiwatashi A. et al. Ultrashort echo time imaging of normal middle ear ossicles: a feasibility study. Dento maxillo facial radiology 2012; 41: 601-604
- 41 Mango VL. Morris EA. Dershaw D. et al. Abbreviated protocol for breast MRI: are multiple sequences needed for cancer detection?. European journal of radiology 2015; 84: 65-70
- 42 Fallenberg EM. Renz DM. Karle B. et al. Intraindividual, randomized comparison of the macrocyclic contrast agents gadobutrol and gadoterate meglumine in breast magnetic resonance imaging. European radiology 2015; 25: 837-849