Keywords
bedside ultrasound - POCUS - program development - ultrasound program
In critical care medicine, point-of-care ultrasound (POCUS) has been shown to provide rapid diagnostic capability, enhance patient care, and improve procedural accuracy.[1] POCUS is used readily in numerous medical specialties and as a result of growing utilization, ultrasound is now a key component of medical school curriculum and graduate medical education training programs in emergency medicine and critical care.[2]
[3]
[4] In neonatology, training in ultrasound and the use of ultrasound for diagnosis and management is increasing, but overall its use is still uncommon compared with other critical care fields.[5]
[6] A growing body of literature demonstrates the potential benefits of POCUS for several diagnostic and procedural applications in the neonatal intensive care unit (NICU). This includes rapid assessment of lung disease, cardiac function, and cranial pathology in neonates.[7]
[8]
[9]
[10] In addition, ultrasound can guide and confirm the placement of central lines and can improve the accuracy of arterial line placement and lumbar punctures.[11]
[12]
[13]
[14]
Given the numerous applications for POCUS in neonatology, the establishment of ultrasound programs in NICUs is increasing. Currently, no formal guidelines exist in neonatology for training, competency standards, or certification in POCUS. In this article, we report our initial efforts, structure, and progress in establishing a POCUS program in an academic NICU. We describe our training program and curriculum development, program structure, data management and quality assurance (QA) process, as well as the initial impact of POCUS on patient care.
Materials and Methods
Program Implementation and Structure
The NICU at Lucile Packard Children's Hospital (LPCH) is a Level IV regional center.[15] Experience with ultrasound in our NICU was initially limited to two faculty members with prior training in echocardiography (ECHO) and POCUS. An additional faculty member, one fellow and two neonatal nurse practitioners (NNP) received immersive bedside and hands-on training in bedside ultrasound at an intensive national course in November 2018. This group formed a core LPCH NICU POCUS team with the intent of establishing a program beginning in December 2018. The team eventually grew to include four faculty, two fellows, two NNPs, and one research data coordinator. The program received a grant in August 2019 from the Lucile Packard Foundation for Children's Health to support equipment, personnel, and training. The NICU received two ultrasound machines from other areas of the hospital and purchased a second more advanced machine with ultrasound probes more applicable for the neonatal population in December 2019.
Prior to program initiation, we utilized a multidisciplinary approach to integrate bedside ultrasound into the unit. We collaborated with radiology and cardiology divisions to determine the scope of bedside ultrasound in our unit as an adjunct to clinical care. Both divisions agreed to assist in establishing guidelines and protocols for image acquisition and review images for QA. At our institution, bedside ultrasound was already readily utilized by providers in the pediatric critical care unit (PICU). Therefore, we collaborated to share resources, create a joint database, and establish educational and training opportunities for providers in both divisions. Team meetings occurred weekly and included discussion on program development, research opportunities, and imaging review for QA. At least one provider from the team was available in the NICU 3 days a week to serve as a resource to the clinical teams and be available for POCUS when needed.
Training and Curriculum Development
Following the ultrasound course attended by the initial core team, which included education on basic ultrasound principles and terminology; didactic and hands-on training sessions on cardiac, lung, abdomen, spine and cranial diagnostic imaging; and ultrasound-guided procedures, the team began performing regular scans in the LPCH NICU. We established an ultrasound curriculum at our institution that included formal 2-day immersive ultrasound courses held at least annually, in addition to joint bimonthly didactic and hands-on training sessions with the PICU. The first formal immersive ultrasound course was held in June 2019 for NICU fellows, attendings, and NNPs with a goal of expanding the number of providers with knowledge of ultrasound basics and skills in bedside ultrasound. Faculty from radiology, cardiology, and PICU helped facilitate the course, in addition to neonatology providers. A second course was held for the first and second-year fellows in neonatal-perinatal medicine and pediatric critical care medicine in August 2020. Bimonthly didactic and bedside practical sessions were organized in conjunction with pediatric critical care faculty and focused primarily on training fellows.
NICU POCUS leadership met with the lead physician for the Stanford Hospital ultrasound program to understand what institutional precedents, criteria, and resources existed for establishing an equivalent program in the children's hospital. Based on their feedback, learners were expected to achieve competency through performing and interpreting scans under the supervision of or reviewed by a core team member for diagnostic exams, based on recommendations from the American College of Emergency Physicians statement for diagnostic applications.[16] Learners are required to perform 25 scans in each core diagnostic application, including 10 required for bladder, before scans could be used to influence clinical management. Core team members were also available routinely on the unit or remotely to review scans and provide feedback on image acquisition and interpretation in real time.
Neonatal Ultrasound Core Applications and Scanning Protocols
Based on the existing literature and experience in other critical care fields, diagnostic core applications for ultrasound use in the NICU were determined to include imaging for umbilical line assessment, cardiac, lung, cranial, and abdomen/bladder. Procedural applications included umbilical line placement, lumbar puncture, and vascular access. Bedside ultrasound was not intended to replace formal ultrasound or ECHO when otherwise indicated, and the specific application of POCUS in the NICU was discussed with cardiology and radiology divisions. The team established scanning protocols for each core application that included guidelines on how to obtain a complete study, including probe type, patient position, necessary views, and sample images for reference. These protocols provide a framework for standardization to ensure images are properly acquired. For each core application, indications were established to aid providers in targeting the scan to answer a defined question.
Data Management and Quality Assurance
We initially tracked only ultrasound indication, interpretation, and impact before creating a secure, web-based database in October 2019 using Research Electronic Data Capture (REDCap) that was Health Insurance Portability and Accountability Act (HIPAA) compliant and Institutional Review Board approved. The Stanford REDCap platform (http://redcap.stanford.edu) is developed and operated by Stanford Medicine Research IT team. The REDCap platform services at Stanford are subsidized by a) Stanford School of Medicine Research Office, and b) the National Center for Research Resources and the National Center for Advancing Translational Sciences, National Institutes of Health, through grant UL1 TR001085. Indications for the ultrasound and other key metrics for each core application were predetermined and included in the datasheet ([Fig. 1]). After each study, providers completed an online data sheet in REDCap that is accessed by a Quick Response code on the machines. Providers documented the type of image obtained, clinical questions, findings and interpretation, and what actions were taken based on the ultrasound findings. Along with the review of the images, the database serves to primarily aid in QA and data tracking. In addition, it assists in clinical research and program development, acting as a logbook to track an individual's ultrasound numbers.
Fig. 1 Example of REDCap online datasheet.
Images from the ultrasound machines were acquired and archived on a HIPAA-compliant cloud storage service available through Stanford University. QA sessions with imaging review were held weekly, where the team reviewed all archived images and REDCap logs. Image sequences were reviewed for completeness and image quality based on predefined metrics from the protocols for each application. Sessions incorporated multiple disciplines including radiology faculty and other critical care faculty with ultrasound expertise. As part of the REDCap log, providers could request feedback regarding their images and interpretation.
Statistical Analysis
The total number of ultrasounds performed in the NICU for clinical indications was evaluated from December 1, 2018 to December 30, 2020. We describe the ultrasounds performed in each core application, the types of providers performing the ultrasounds, and the indications for each scan. Finally, through reviewing the REDCap logs, based on two independent assessments, we evaluated whether the bedside ultrasound prompted a change in clinical management.
Results
In addition to the core team of 6 providers initially trained in bedside ultrasound, an additional 22 providers attended the institutional 2-day POCUS course in June 2019. Sixteen neonatology and pediatric critical care providers attended the course in August 2020.
A total of 508 (76.8% diagnostic and 23.2% procedural) bedside ultrasounds were performed in the NICU on 290 patients from December 1, 2018 to December 31, 2020 by 4 attendings, 12 fellows, 4 NNPs, and 3 neonatal hospitalists. Scans were performed in five diagnostic applications: umbilical line visualization, cardiac, lung, abdomen (including bladder), and cranial. Procedural applications included ultrasound-guided umbilical line placement, vascular access, and lumbar puncture. The number of scans and their indications are detailed in [Table 1].
Table 1
POCUS scans by core application and indication
Core application and indication
|
Number of scans (%)
|
Umbilical line assessment
|
283 (55.7%)
|
Visualization
|
243
|
Placement and visualization[a]
|
40
|
Cardiac
|
15 (3.0%)
|
Function
|
8
|
Inferior vena cava filling
|
6
|
Pericardial effusion
|
4
|
Other
|
3
|
Lung
|
72 (14.2%)
|
Effusion
|
17
|
Pneumothorax
|
9
|
Evaluation of lung pathology while on respiratory support (RDS and TTN)
|
49
|
Cranial
|
7 (1.4%)
|
IVH
|
4
|
Ventricular size assessment
|
2
|
Other (abnormal anatomy)
|
2
|
Abdomen/Bladder
|
53 (10.4%)
|
Bladder volume assessment
|
45
|
Urinary catheter visualization
|
7
|
Ascites
|
2
|
Other
|
3
|
Lumbar puncture[a]
|
31 (6.1%)
|
Peripheral vascular access[a]
|
47 (9.2%)
|
Total
|
508
|
Abbreviations: IVH, Intraventricular hemorrhage; POCUS, point-of-care ultrasound; RDS, respiratory distress syndrome; TTN, transient tachypnea of the newborn.
a Procedural applications included ultrasound-guided umbilical line placement, lumbar puncture, and peripheral vascular access.
In our unit, bedside ultrasound was used both for surveillance of umbilical catheter tip position and during umbilical catheter placement. Of the 243 umbilical lines that were visualized as part of routine surveillance, 64 (26%) were identified to be in an inappropriate position, prompting line manipulation or removal. Based on the review of documentation in REDCap, 5 out of 15 (33%) cardiac, 10 out of 72 (14%) lung, 27 out of 48 (56%) bladder, 1 out of 5 (20%) abdominal (non-bladder), and 3 out of 7 (43%) cranial ultrasounds influenced clinical management. [Table 2] details examples of how POCUS within each of these applications impacted management. Ultrasound-guided peripheral vascular access was used for arterial catheter placement and resulted in successfully placed catheters in 35 out of 47 (74%) attempts. There were 31 scans done to evaluate spinal fluid, and in all scans lumbar puncture was either attempted (26) or deferred (5) based on the presence or absence of adequate fluid or presence of a hematoma. Successful lumbar puncture was performed in 23 out of 26 (89%) of patients who had spinal ultrasounds demonstrating adequate fluid. As we continued to train providers and provide bedside support, bedside ultrasound use increased over the 2-year period ([Fig. 2]).
Fig. 2 Number of POCUS scans by quarter from December 2018 to December 2020.
Table 2
Examples of POCUS scans in the NICU affecting management
|
Patient information and diagnosis
|
Study indication/clinical question
|
Interpretation
|
Intervention
|
Umbilical line
|
35-wk infant with ectopic atrial tachycardia and umbilical venous catheter
|
Confirmation of tip location given arrhythmia while awaiting X-ray
|
Tip in the right atrium, measured 1.5 cm above the junction between the inferior vena cava (IVC) and right atrium (RA)
|
Line retracted under ultrasound guidance with resolution of arrythmia before X-ray was able to be obtained
|
|
27-wk infant with umbilical venous catheter
|
Catheter tip below diaphragm on recent X-ray, so POCUS done to confirm location
|
Tip visualized below IVC/RA junction in ductus venosus
|
Catheter removed and PICC line placed
|
Cardiac
|
Term infant with necrotizing fasciitis
|
Hypotension postoperative, assess cardiac function and filling and need for vasoactives versus volume
|
Poor cardiac function, ventricles not underfilled
|
Initiated pressor support, cardiology consult called for comprehensive echocardiogram
|
|
Term infant with hypoxic respiratory failure on extracorporeal membrane oxygenation support with decreased near infrared spectroscopy values of unknown etiology
|
Assess cannula position
|
Large pericardial effusion
|
Formal ECHO obtained and multidisciplinary discussion regarding intervention
|
Lung
|
Term infant with congenital diaphragmatic hernia on extracorporeal membrane oxygenation support with new opacification on X-ray
|
Evaluate for effusion versus consolidation
|
Large effusion visualized
|
Formal ultrasound ordered to quantify fluid, surgical team notified and chest tube placed
|
Cranial
|
27-wk infant with anemia, acidosis, coagulopathy
|
Evaluate for hemorrhage
|
New large, intraparenchymal hemorrhage
|
Received several blood products. Formal head ultrasound confirmed bilateral intraparenchymal hemorrhages
|
Abdomen/Bladder
|
Term infant with intestinal perforation and bowel resection with purulent fluid at the incisional site
Preterm infant with prenatally identified large ascites, compromised ventilation in the delivery room
|
Evaluate for abdominal fluid
Evaluate for abdominal fluid prior to paracentesis
|
Fluid collection visualized
Large volume of ascites visualized
|
Formal ultrasound obtained and surgical team notified
Guided paracentesis performed in the delivery room with subsequent improvement in respiratory status
|
|
Term infant with hypoxic ischemic encephalopathy, receiving therapeutic hypothermia and on morphine
|
Oliguria
|
Distended bladder visualized
|
Foley placed and 27-mL urine drained
|
|
Term infant with esophageal atresia with foley in place but decreased urine output postoperatively
|
Oliguria
|
Bladder with low volume
|
Normal saline bolus given
|
Abbreviations: ECHO, echocardiography; IVH, Intraventricular hemorrhage; NICU, neonatal intensive care unit; PICC, peripherally inserted central catheter; POCUS, point-of-care ultrasound; RDS, respiratory distress syndrome; TTN, transient tachypnea of the newborn.
Discussion
This article describes our initial process of implementing a POCUS program in a large academic NICU. In a 2-year period, we performed 508 scans; expanded the number of providers performing ultrasound in our unit; and established an educational program, ultrasound image database, and QA regimen. Following our initial immersive course, we partnered with the PICU to enhance program development. Successful implementation of a POCUS program requires access to educational programs and competency assessment, a defined scope of practice and established protocols, and imaging review with QA. In addition, collaboration with other fields that utilize bedside ultrasound, including critical care, radiology, and cardiology, is crucial to ensuring success.
We demonstrate that POCUS was able to influence clinical management, including as a prompt to obtain formal ultrasound or radiograph imaging in five core applications (umbilical line assessment, cardiac, lung, abdomen, and cranial). Although it is difficult to define the degree to which POCUS improved care and clinical outcomes, our data highlight clinical scenarios where POCUS can be a crucial diagnostic tool. For example, in an infant who has decreased urine output, POCUS can quickly elucidate whether this is due to urinary retention or oliguria, and that information can guide whether a urinary catheter or fluid administration would be an appropriate intervention. Similarly, in an infant who is hemodynamically unstable, POCUS can identify pathology such as a pericardial effusion that would allow for rapid intervention. POCUS allows for more targeted interventions, can avoid unnecessary procedures and treatments, and can be used to provide rapid diagnostic information when formal imaging may not be immediately available.[1]
[17]
[18]
[19]
Despite the widespread use of bedside ultrasound in clinical medicine throughout the world, POCUS in neonatology is still in an early initiation phase in much of the United States. In Australia, New Zealand, and throughout Europe, training programs for comprehensive bedside cardiac ultrasound have been established and neonatologists regularly use ultrasound in their clinical practice.[20]
[21]
[22] In the United States, however, neonatologists report limited availability of equipment and required skills to perform POCUS.[10] A recent survey of neonatologists in the United States found that although 90% of responders supported the introduction of POCUS in their clinical practice, several barriers prevented them from doing so.[6] A majority of neonatologists cited concerns about the lack of training guidelines, inadequate support from other services, and legal concerns. Additional barriers included skill maintenance, conflict of interest with radiology and cardiology services, and cost to purchase and maintain equipment.[6] In establishing our program, we were fortunate to obtain a grant that supported a program coordinator position, purchasing equipment, and time for faculty to devote to program development. We initiated a method for image archiving and review to ensure a robust QA process. We engaged multiple services early, including partnering with the PICU for joint educational sessions and QA. Our radiology and cardiology divisions were supportive of the use of POCUS, with several meetings prior to the initiation of our program to clarify the applications for POCUS in the NICU to ensure that it would not replace a formal ultrasound or ECHO when those would otherwise be clinically indicated.
Neonatologists considering utilizing POCUS in their institution should first determine appropriate applications in their clinical setting. Large variability currently exists in bedside ultrasound application, training, and oversight in clinical neonatology practice. The Emergency Care Research Institute listed the lack of oversight regarding when and how to use POCUS as one of the top ten health technology hazards in 2020.[23] Many experts agree that as more NICUs invest in ultrasound equipment, it is important to develop evidence-based guidelines for use, develop standards for training, specify the scope of practice, and establish mechanisms for QA.[24] The European Society of Pediatric and Neonatal Intensive Care established an international POCUS working group that created evidence-based guidelines for bedside ultrasound use in the NICU and PICU for the following application: cardiac, lung, abdominal, cranial, and vascular access. These guidelines outline important features of structured POCUS applications in clinical practice, including the estimated level of training required for their use. For example, for those with basic lung POCUS training, there was agreement that POCUS may be helpful with differentiating between respiratory distress syndrome and transient tachypnea of the neonate, assessing for pneumothorax and pleural effusion, and providing guidance for thoracic procedures.[25] As our hospital has not yet implemented credentialing requirements for POCUS, the scanning protocols and REDCap datasheets that we created for our core applications correlate with these guidelines in anticipation of requiring medical staff privileges for POCUS in the future.
Although recently there have been more opportunities for training in POCUS, options for formal training remain limited and no universal curriculum or accreditation process exists. Establishing a scope of practice for an institution and utilizing existing guidelines are important when first implementing the use of POCUS and determining what training is required. All POCUS group members received initial bedside ultrasound experience through a 2-day intensive course and gained hands-on experience under the guidance of two group members who had prior training in ECHO and POCUS. To introduce POCUS to a broader group of neonatologists, nurse practitioners, and trainees, we subsequently organized courses at our own institution while also soliciting external faculty experts in bedside ultrasound. In addition, we implemented regular didactic and hands-on training sessions in collaboration with PICU providers. Short POCUS courses offer an introduction to bedside ultrasound but regular hands-on training sessions and bedside implementation are necessary to truly master the skill. In implementing a POCUS program, both large and smaller NICUs may find it beneficial to start with a core group of providers who can be trained in POCUS and can subsequently expand the use of the skill to as many providers as is necessary and feasible. At our institution, we have continued to organize POCUS courses with our PICU colleagues to expand the use of POCUS in both academic and community settings in Northern California.
A collaborative approach with the involvement of other clinical imaging services and critical care fields is integral for POCUS implementation. Early acceptance and integration by PICU and emergency medicine providers are a resource for enhancing bedside ultrasound education and imaging review. Similarly, given the overlap that bedside ultrasound may have with the services provided by cardiology and radiology, it is important to define the scope of utilization with these departments and reinforce that bedside ultrasound is a specific tool to answer defined questions rather than a replacement for medical imaging performed or interpreted by cardiologists and radiologists. In particular, cardiac POCUS in neonates should be intended to answer acute and focused clinical questions, and given the possibility of congenital heart disease, a comprehensive ECHO must be performed to confirm normal heart anatomy.[26] Many community NICUs may not have access to other pediatric providers who already utilize POCUS; however, pediatric radiology and cardiology services are generally available and these providers can be valuable resources for education, imaging review, and QA.
Certain programs may require credentialing prior to use of POCUS, with pathway to the creation of a program including discussions with hospital administration to identify a credentialing pathway. Certification and credentialing are currently not established at our institution, but our program aims to develop internal guidelines for credentialing that will require completion of both didactic, hands-on, or simulation programs, as well as a successful demonstration of a requisite number of ultrasounds and interpretations for each application with oversight and assessment of competency through a QA program. Currently, no national or international standards to define competence or credentialing for POCUS in neonatology exist, but standardized pathways to credentialing, as emergency medicine has established, should be pursued.[4]
[27]
[28] Smalley et al outline a standardized POCUS credentialing initiative in a large health care system that demonstrated ability to credential community emergency medicine physicians, including providing educational courses, scanning and documenting educational exams in the electronic medical record, and reviewing images.[29] Mathews and Zwank describe the images necessary to obtain as the process for credentialing as well as the requirement to pass a final written and hands-on exam to assess competency prior to applying for POCUS credentialling. The timeline to achieve competency varied among providers.[30] Outlining a similar process in neonatology is needed to provide a framework and pathway to credentialing with QA processes as programs adopt POCUS in their units. POCUS in the NICU is becoming increasingly ubiquitous, with increasing applications providing further diagnostic tools to aid in the care of our neonatal population. For those in institutions who do not permit the use of POCUS without prior credentialing and certification, we recommend working with hospital administration and following a transparent and careful initiation, leveraging expertise from all areas of the hospital to establish a high quality, high fidelity program to advance the care of newborns. Without this, a mandate for prior credentialing amounts to circular logic that would prevent POCUS from being introduced into the NICU when established literature supports its use.
The development of a NICU POCUS program benefits from a team approach, encompassing educational opportunities and QA with interdisciplinary collaboration. Taking notes from our colleagues in fields with established POCUS standards, such as emergency medicine, POCUS in the NICU would benefit from standardization and guidance including scope of practice, training guidelines, credentialing, and quality assessment to enhance program development and create a mechanism for ensuring optimal roll out and use to impact patient care.