Under the title “MR Imaging in Patients with Cardiac Pacemakers and Implantable Cardioverter
Defibrillators”, this issue of Röfo – simultaneously with the journal of the German
Cardiac Society “Der Kardiologe” – presents a consensus paper of the German Roentgen
Society and the German Cardiac Society [1]
[2] that was jointly written by the authors in radiology and cardiology but does not
exclusively address cardiac MRI. This publication relates to MR imaging of all regions
of the body.
In Germany and internationally the number of MRI examinations is increasing: 1,008,944
examinations were performed on patients receiving inpatient treatment in 2005 while
1,767,005 examinations were performed in 2013 (DRG hospitals). This development can
be attributed to our aging population as well as to new indications for MRI including:
Analyses of tissue composition and function, for example in the liver [3]
[4] and the heart [5]
[6]; multiparametric analyses of MR perfusion, e. g. in treated brain tumors [7]; new organs such as the lung [8]; dedicated examinations for intervention planning and operation monitoring [9]
[10]
[11]; as well as MRI-guided interventions [12]
[13]
[14]
[15]
[16].
Implants must always be considered in all of these MRI examinations even if the reason
for the examination request is not related to an implant. The involvement of cardiology
in this case is not based on the medical issue but rather on the type of implant.
Expertise in cardiology is required when dealing with cardiac pacemakers (PM) and
implantable cardioverter defibrillators (ICD).
Background
MR imaging is fundamentally seen as safe because interactions between the electromagnetic
fields needed for MR imaging (static field, high-frequency field and gradient field
in the audio frequency range) and body tissue do not have a lasting effect on the
person being examined. However, if other materials are in the body, e. g. in the case
of implants, interactions affecting patient safety can occur. Safety concerns regarding
these interactions resulted in the presence of an implant being considered an absolute
contraindication for MR imaging. Manufacturers of MRI units therefore stated in their
instructions for use that examination of patients with implants is fundamentally not
allowed. This satisfies the requirements of an international standard for manufacturers
of MRI units regarding safe operation of their devices. According to the (German)
Medical Devices Operator Ordinance [17], operators of MRI units are required to observe (safety) information in the operating
instructions.
The realization that it is ultimately not feasible to exclude all patients with implants
from MR imaging due to developments in medicine prompted manufacturers of implants
and MRI units to examine the possible risks posed by implants more closely. As a result,
conditions under which MR imaging can be performed without harm to the patient were
defined for suitable implants. These implants are labeled as “MR conditional”. The
conditions, i. e., the requirements regarding MR imaging and possibly implant settings,
are included in the labeling [18]
[19]. These requirements can be extensive and complicated so that examinations involving
implants often cannot be performed in the daily routine even when the implants are
labeled as “MR conditional”.
Knowledge of the exact conditions of use and the guarantee that these conditions will
be strictly observed during MR imaging are essential for patient safety in the case
of such “MR conditional” implants.
At the same time, manufacturer standards, which contain specifications regarding the
content of operating instructions, have been updated to state that patients with “MR
conditional” implants can be examined, provided that the specified conditions are
met [20].
Current situation
There is still great uncertainty about implants so that even patients with “MR conditional”
implants are sometimes barred from a permissible MRI examination.
This is particularly true for patients with “MR conditional” cardiac implants. The
conditions to be met are complicated and involve radiology as well as cardiology even
if the examination request is made by a completely different department. Radiologists
and cardiologists must work in close collaboration when preparing for an MRI examination
involving a cardiac pacemaker or ICD. Therefore, this joint statement by radiologists
and cardiologists presenting guidelines regarding the conditions and necessary procedures
for examining patients with pacemakers and ICDs represents an important step forward
for these patients.
The revised standards initially affect patients with implants explicitly labeled as
“MR conditional”. The situation for patients with implants without this labeling is
more complicated. This doesn't just apply to patients with older implants. Implants
that were not specifically designed or tested for safe operation during MR imaging
are still used today because implants that are designed to withstand the fields of
MRI scanners are more expensive.
However, the in-depth investigation of possible interactions between PMs or ICDs and
MR fields allows better assessment of the possible risks that can occur during MR
imaging in patients with conventional, not “MR conditional”, implants. It was shown
that the risks can be significant. However, in some cases the benefits of an MRI examination
can be so great that examination is justified despite the risks. In such a case, the
referring department, cardiology, and radiology (and possibly additional departments
that can evaluate therapeutic consequences) must perform a joint risk-benefit analysis
so that the patient can make a decision regarding a proposed off-label examination.
The paper presented here also describes the decision criteria and processes for examinations
in patients with PMs and ICDs that are not approved for MR imaging.
Cardiology and radiology are responsible for the settings and monitoring of their
respective devices for both in-label and off-label examinations. The radiologist or
the specialized physician with MR expertise conducting the MRI examination is ultimately
responsible for the examination procedure.
It must be noted that “MR conditional” only means that the MRI examination does not
pose an immediate danger to patients and personnel. Whether image artifacts can affect
the diagnostic value of imaging (thus making an examination obsolete) or whether the
implant will resume full functionality after the examination is explicitly omitted
here. The most recent documentation of the implant manufacturer should always be closely
reviewed in this regard. Some manufacturers provide this documentation on their websites.
This consensus paper is based on the current state of the art. It would appear that
newer developments may be able to simplify examinations in patients with implants:
Manufacturers of implants and MRI units have agreed on a set of measurement parameters
to be implemented initially at 1.5 T for MRI scanners as “fixed parameter option:
basic” (FPO:B) [21]. When this option is (hopefully easily) selected, all relevant measurement parameters
of the MRI system will be limited to values that prevent harmful interactions with
appropriately designed or prepared implants. Moreover, there are software options
that can be used to limit a series of parameters beyond basic parameters like SAR
that are otherwise inaccessible and are specified as critical by implant manufacturers
in the list of conditions for MR imaging. Additional settings (besides limitation
of the SAR value) that, for example, reduce the risk of voltage induction in systems
and leads can be made even in conventional PMs and ICDs.
This consensus paper clearly defines procedures for MR imaging in the case of active
cardiac implants. The requirements are high. Such examinations should only be performed
if the necessary radiology and cardiology expertise can be ensured. An attempt should
be made to develop or further develop similar procedures for MR imaging for additional
implant types in order to ensure maximum patient safety.