Key words guidelines - liver - fibrosis - chronic liver disease - hepatitis - chronic viral
hepatitis - shear wave elastography - strain elastography - alcoholic hepatitis -
NAFLD
Basic Principles and Technology
Basic Principles and Technology
Introduction
In these guidelines, the term elastogram refers to any image of any elastic property
of tissue. This article also describes non-imaging measurements of such characteristics
over a defined region. These are referred to as point elastography methods. Elastometry,
on the other hand, is the measurement of an elastic characteristic of tissue, which
may be obtained from either imaging or non-imaging methods.
The basic principles of elastography have not changed since they were outlined in
the first part of the original EFSUMB guidelines and WFUMB guidelines on this subject
[1 ]
[2 ]. This paper therefore aims to provide an update of technology changes as of 2017
in this rapidly evolving field. Sufficient recapitulations are provided to allow the
present paper to be understood without reference to the earlier paper, although the
purpose is not to reproduce the material of the 2013 paper in detail.
Basic principles of elastography
We recommend that elastography be considered as a type of remote palpation that allows
measurement and display of biomechanical properties associated with the elastic restoring
forces in the tissue that act against shear deformation. This view unifies the different
types of elastography and explains why they all display images with contrast to obtain
the same underlying information, associated with the shear elastic modulus defined
below. It is explained in detail in [1 ], along with different types of shear deformation. Shear deformation is generated
by a force applied to a single location or broadly across the body surface. Importantly,
transient shear deformations will propagate as shear waves, as discussed below.
A force may be applied by pressing on or vibrating the body surface, by naturally
occurring internal physiological motion or by electronically controlling the ultrasound
transducer to create focused acoustic radiation force at a controlled depth. The acoustic
radiation force is: proportional to the time-averaged intensity I of the sound beam
at the push location and the amplitude absorption coefficient α at the push location,
inversely proportional to the speed of sound cι at the push location and increases with any reflection or scattering of the sound
at the push location. In the absence of reflection or scattering, the magnitude of
the radiation force acting at each point in the volume of the medium is F = 2αI / cι [3 ]. All ultrasound beams generate such force but the force integrated over the length
of a diagnostic pulse (< 2 μs) is too small to generate a measurable displacement.
Beams designed to create a measurable displacement with acoustic radiation force use
deliberately longer pulses (50 – 1000 μs) to generate micron-level tissue displacements.
All available ultrasound elastography methods ([Fig. 1 ]) employ ultrasound to measure the internal tissue shear deformations resulting from
an applied force. The type of force ([Fig. 1 ], column 2) is important. If the force varies slowly relative to the shear propagation
time to the depth of interest, as is the case for probe palpation or physiological
motion, it is considered quasi-static. It is difficult to acquire images that are
quantitative for tissue properties using a quasi-static applied force although quantification
relative to a reference tissue or phantom material may be possible. Dynamic forces
permit absolute quantification of tissue property. They include impulses and continuous
vibrations, and may be produced mechanically at the body surface or deep within the
body using acoustic radiation force. Only impulsive dynamic forces are shown in [Fig. 1 ] because current commercial ultrasound elastography systems still do not use a continuously
vibrating force, although there are examples of this in the research reported in the
ultrasound elastography literature [4 ]
[5 ]
[6 ]. This is unlike magnetic resonance elastography (MRE), where a continuous harmonic
vibration induced at the body surface is the most common approach [7 ]
[8 ]
[9 ].
Fig. 1 The types of elastography described in this paper.
The signal processing within the scanner for all current commercial ultrasound elastography
methods begins with the measurement of tissue displacement as a function of spatial
position and time, which is done using cross-correlation tracking, Doppler, or other
signal processing. The various elastography methods differ with respect to what they
do with these displacement data to create an elastogram or elasticity measurement.
There are three options for the property to be displayed ([Fig. 1 ], column 4):
Display of displacement without further processing ([Fig. 1 ], row 2), as in acoustic radiation force impulse (ARFI) imaging.
Tissue displacement is associated with shear deformation. The greater the force, the
greater the displacement, but stiff tissues are displaced less than soft tissues.
Stiffness is defined as applied force divided by the generated tissue displacement.
If the force (not normally known) is assumed to be the same for all image locations,
an image of displacement can be thought of as an inverse relative stiffness map (it
is the inverse, because the stiffer the tissue, the smaller the displacement). Note
that stiffness is not only governed by the tissue’s mechanical properties but is also
governed by geometrical factors such as the area of the cross-section of the tissue
sample over which the force is applied. Although displacement itself is a quantitative
measurement (units of µm), image brightness is typically scaled between soft (bright)
and hard (dark).
Display of tissue strain or strain rate ([Fig. 1 ], row 1), calculated from the spatial gradient of displacement or velocity, respectively,
as in strain elastography (SE).
Displacement imaging works well when a highly localized force is applied internally
at a depth, as in ARFI imaging. However, when the force is applied across a large
region of the surface of the body, strain provides a better measure of deformation,
as, unlike displacement, it does not decay with depth. Young’s modulus E is a true
tissue property, measurable by axial compression of a tissue sample with a constant
cross-sectional area along its length. It is equal to the applied stress σ divided
by the measured strain ε, i. e., the well-known Hooke’s law E = σ / ε, where stress
is the applied force per unit area and strain is the change in length of the sample
divided by its original length. If the stress (not normally known) is assumed to be
the same for all image locations, an image of strain can be thought of as an inverse
relative Young’s modulus map (it is the inverse, because the higher the Young’s modulus,
the smaller the strain). Although strain itself is a quantitative measurement (units
of %), image brightness is typically scaled between soft (bright) and hard (dark).
Display of shear wave speed ([Fig. 1 ], rows 3 – 7), calculated by using the time varying displacement data to measure
the arrival time of a shear wave at various locations. All such methods are grouped
under the heading shear wave elastography (SWE), and include transient elastography
(TE), point shear wave elastography (pSWE) and multidimensional shear wave elastography
(2D‑SWE and 3D-SWE).
This is only possible when the force is applied dynamically. The shear deformation
generated locally and momentarily within tissue will propagate as a shear wave and,
so long as it is not completely attenuated as it travels, will reach a distant location
after a time determined by the shear wave speed cs . Shear wave speed may be displayed in units of ms–1 . Alternatively, assuming that it does not vary with the magnitude or frequency of
the applied force, or with position and direction in the tissue, cs may be converted to either Young’s modulus E or shear modulus G, using the equations
E = 3ρcs
2 and G = E / 3 (assuming the tissue is incompressible), where ρ is the density of
the tissue. E and G are expressed in units of kilopascal (kPa). The choice of whether
to display speed or modulus may be under user control, or such a choice may be unavailable
(sometimes determined by regulatory authorities for the region of the world in which
the scanner or measurement system is manufactured or sold).
Explanation: It is typical of the MRE literature that data are provided for shear
modulus G, whereas in the ultrasound elastography literature it is more common to
provide values for Young’s modulus E. The reader should be wary when comparing data
across published papers, as both quantities are expressed in the same units (kPa),
and yet E = 3G. When publishing results of studies, it is important to state which
modulus has been measured when using units of kPa, avoiding non-specific terms such
as stiffness, and to say how it was calculated.
The reasons why ultrasound can be used to make elastograms
Ultrasound is the propagation of a transient density deformation. In soft tissues,
it travels at speeds in the range of 1350 – 1600 ms–1 , whereas shear deformation travels much slower, in the range of 1 – 10 ms–1 [10 ]. This speed difference means that ultrasound may be used to measure tissue displacements
at precise phases of shear deformation. Ultrasound also suffers relatively little
attenuation, allowing high-resolution (mm) measurements of displacements of a few
microns at depths of many centimeters. Furthermore, the speckle in ultrasound images
allows measurements of tissue displacements even when there are no resolved tissue
structures to track.
The differences between elastograms and ultrasound images
The tissue contrast in all types of elastograms is potentially very high because it
is related to differences in shear modulus G, which varies by more than five orders
of magnitude across all soft tissues, as compared with bulk modulus K which varies
by about 12 %. Ultrasound is able to propagate through fluids, and in tissue the molecular
composition largely determines ultrasound speed and absorption, whereas shear will
not propagate in a non-viscous pure fluid and is transferred largely via the high-level
tissue structure. Although ultrasound scattering occurs at the large-scale structural
level, it also arises at the cellular level and does not require the structure to
be connected. Shear on the other hand will not propagate across zero friction breaks
in the structural matrix. For example, a dilute suspension of cells in saline will
scatter ultrasound but will not support a shear deformation or a shear wave. Finally,
and very importantly, unlike the situation for ultrasound propagation and scattering,
the shear modulus of most tissues increases with vascular and interstitial pressure.
These features may make elastography sensitive to the presence of disease in situations
where conventional ultrasound imaging fails to detect it.
Technology of ultrasound elastography
The types of elastography referred to above and commercially available systems are
shown in [Fig. 1 ]. Note that they may be complementary in many respects for general clinical use.
For applications in the liver, methods that display shear wave speed (SWS) are the
most common, followed by displacement imaging (for lesions in the liver) and finally
by strain imaging.
Strain elastography (SE) and strain rate imaging (SRI): quasi-static strain imaging
([Fig. 1 ], row 1)
Strain elastography, although it remains the most widely implemented elastography
method on commercial systems, is the least used method for the liver. As motion occurs
due to either palpation with the ultrasound probe or to physiological motion, images
of axial displacement between sequentially acquired ultrasound echo images are calculated
using either radiofrequency echo correlation tracking or Doppler processing (when
Doppler processing is used the method is often called strain rate imaging). A moving
window axial-gradient estimator converts the axial displacement images to strain images.
Little has changed concerning the available systems, artefacts and advantages and
disadvantages of strain imaging, although the trend is towards increased robustness,
using more frame and pixel quality selection methods, and increased sensitivity allowing
useful elastograms to be obtained with little or no palpation, relying on involuntary
hand motions or physiological motion, on some systems.
Advice concerning semi-quantitative analysis of strain images using, for example,
strain ratios and strain histograms, remains as described in [1 ].
Acoustic radiation force impulse (ARFI) imaging: imaging tissue displacement induced
by radiation force ([Fig. 1 ], row 2)
The axial displacement for locations down a given image line at and around a single
pushing focus is measured by comparing the locations of tissue echoes and/or speckle
for imaging pulses emitted before and immediately (< 1 ms) after the push. A displacement
image within a small region of interest (ROI) is then acquired by repeating the pulse
sequence “image-push-image” for a sequence of A-lines in the ROI. ARFI remains the
proprietary imaging technology Siemens Virtual Touch™ as described in 2013, and the
advantages and disadvantages and the various image artefacts remain unchanged [1 ]. It is not used for assessment of diffuse liver conditions.
Shear wave elastography (SWE); methods that use shear wave speed
Transient elastography (TE): shear wave elastometry by measurement of the speed of
a shear wave that has been generated using a surface impulse ([Fig. 1 ], row 3)
Despite the name, this is not the only method that uses a transient force. All of
the dynamic methods in [Fig. 1 ] do so. This device is designed only for liver elasticity measurement and for use
also by persons who are not imaging specialists. An automated movement of a piston,
which is also a disc-shaped ultrasound transducer, applies a single cycle 50 Hz push
to the body surface with controlled applied force. The transient shear deformation
created in this way, propagates into the tissue. Its near constant speed for about
4 cm in the liver (before being rendered non-detectable due to attenuation) is measured
by a straight line automatically fitted to the displacement M-mode shown on the right
in [Fig. 2 ], obtained by processing the ultrasound radio frequency A-mode echo signal versus
time.
Fig. 2 Transient elastography with the Fibroscan™. Sample display showing the echo M-scan
on the left, single-line amplitude A-scan in the middle, and the displacement M-mode
after a vibration-controlled impulse push on the surface (see [1 ] for a full explanation).
Point shear wave elastography (pSWE): shear wave elastometry at a location by measurement
of the speed of a shear wave generated using acoustic radiation force ([Fig. 1 ], row 4)
When an ARFI is applied at a depth within tissue, a shear wave is generated. Its average
speed of propagation from the focal point positioned on one lateral boundary of a
measurement ROI to another on the opposite lateral boundary of the ROI may be measured
by detecting its time of arrival at that point, relative to that of the ARFI. As with
TE, no elastograms are created. Only a regional average of SWS is measured. Ultrasound
imaging is used, however, to guide placement of the ROI, and measurement may be made
even when it is necessary to propagate through ascites. Other advantages and disadvantages
are described in [1 ]. The main update since [1 ] is that, as listed in [Fig. 1 ], more manufacturers are now providing this type of measurement, which was known
as ARFI quantification when first introduced by Siemens. Currently, this option is
also offered by Esaote, GE, Hitachi, Philips and Samsung.
The quality of the SWS estimate is judged by the SWS estimation algorithms in pSWE
systems and may warn the user if it is not adequate.
Multidimensional shear wave elastography (2D-SWE, 3D-SWE): quantitative SWE imaging
(and elastometry) by measuring the speed of shear waves generated using acoustic radiation
force ([Fig. 1 ], rows 5 – 7)
The method described above for pSWE may be extended to make quantitative images of
SWS in a large ROI by placing the ARFI focus (push) at multiple sequential locations
and, at each, detecting the shear wave arrival time at multiple lateral locations.
This creates patches of small SWS images that may be mosaicked to create a large ROI
2D-SWE image, which is displayed in color or grayscale. This may be overlaid with
adjustable transparency on the B-mode image or, optionally, the SWE image and B-mode
may be displayed separately, side-by-side. In addition to visual judgment of such
elastograms against a color scale (calibrated in ms–1 or kPa), they may be quantitatively analyzed using measurement boxes. These ROIs
can be placed at desired locations to obtain statistical quantities such as the mean,
standard deviation, minimum and maximum values of the SWS or Young’s modulus in kPa
(although the minimum is often not useful because of missing data values). It takes
time to push sequentially and measure at multiple push locations. Therefore, in some
implementations, a static SWE image is created (in about 700 ms on the Siemens S3000™),
which may be repeated upon user command, after allowing time for the transducer to
cool. In the SuperSonic Imagine Aixplorer™ ‘real-time’ 2D-SWE images (up to several
frames per second) are created using a combination of an axially swept push focus,
which generates a push line in the SWE image ROI and a shear wave that should decay
less rapidly than that from a single push focus. The echo images are generated at
thousands of frames per second by using plane transmitted ultrasound waves and highly
parallel receive beam-forming to allow detection of shear wave arrival times at many
locations without the need for many repeated ARFI pushes.
The main update since [1 ] is that other manufacturers, including Toshiba, GE, Philips and Mindray, have now
released 2D-SWE products that are broadly similar but differ with respect to the details
of the method and sampling rate.
Toshiba, Philips and Mindray seem to have implemented a conventional ARFI pushing
method similar to Siemens. Mindray employs highly parallel receive beam-forming and
a zoned transmit wave to sample the travelling shear wave at more than 1000 frames
per second within areas up to 4 cm wide. Philips also employs highly parallel processing,
allowing free-running 2D-SWE over areas up to 5 cm high by 7 cm wide with an abdominal
curved array transducer at a frame rate of 0.4 – 1.6 Hz, while the underlying B-mode
runs at 20 Hz for visual guidance ([Fig. 3b ]).
Fig. 3 Example of 2D-SWE in the liver as implemented on the Philips EPIQ™ b , which may be displayed concurrently with the B-mode image a and a confidence map c described in the text.
GE has implemented the “comb push” technique [11 ], which (as illustrated in [Fig. 4 ]) pushes along a number of ARFI lines simultaneously. As a consequence, the propagating
shear waves travelling away from different push lines pass each other. To avoid confusion
of detected times of arrival for waves travelling in different directions, the system
analyzes the arrival time at a constant depth as a function of lateral position and
uses Fourier domain filtering to separate waves travelling to the left from those
travelling to the right. In combination with time-interleaved interpolation of the
displacement tracking results, this allows an acceptably large 2D-SWE ROI to be formed
in about 100 ms using a conventional (hardware) beamformer.
Fig. 4 The comb push technique in the GE LOGIQ E9™, which includes shear wave directional
filtering for SWE using a conventional (hardware) beam former [12 ].
Note that direction filtering is also employed by other manufacturers, because it
has the additional benefit of reducing SWS measurement artefacts that are, for example,
due to shear wave reflections at shear impedance boundaries (where either the tissue
density or the shear wave speed changes). It is also becoming common on the recent
2-SWE systems (e. g. Toshiba, GE, Mindray) to provide a choice between a continuously
refreshed image (almost real time) and a more slowly acquired, said to be of higher
quality, single frame.
All 2D-SWE systems have built-in indicators of the quality of the shear-wave speed
estimate and take appropriate action to adjust the display when the quality falls
too low, such as dropping out the offending pixel of the 2D-SWE image overlay to allow
the B-mode image to show through, or setting that pixel to black. Typically, this
will occur as the signal-to-noise ratio deteriorates with depth, limiting the penetration
capability of 2D-SWE, as in the example shown in [Fig. 5 ].
Fig. 5 Example of side-by-side display of B-mode (left) and SWE image (right), from a phantom
containing a spherical inclusion that has little or no echo contrast against the background
but is softer than the background, illustrating that SWE pixel dropout as poor signal-to-noise
ratio at a large depth reduces the quality of the shear wave speed estimate below
that regarded as acceptable, thus limiting the SWE penetration depth.
An additional approach to quality judgement is provided by Toshiba as shown in [Fig. 6 ] (right), which optionally shows the detected times of arrival in a “wavefront” style
of display. The concept is that if consecutive lines are parallel (even if irregular),
the quality is acceptable, but it would not be acceptable if they are not parallel.
In the image shown, it can be seen that time-of-arrival lines become spread further
apart when the speed increases (c.f. image on the left). It appears as if the shear
wave has propagated from left to right, becoming irregular in shape as it emerges
on the right of the relatively stiff inclusion, where the elastogram appears to show
a stiffer region in the background but the non-parallel time-of-arrival lines in the
image on the right indicate that this is an artefact.
Fig. 6 Illustration of the SWE color overlay (left) and “wavefront” style display of detected
shear wave time of arrival (right) on the Toshiba Aplio 500™ displaying artefacts.
A biopsy training phantom containing cyst-mimicking inclusions that are echo-free
but solid and slightly stiffer than the background was used.
Philips provides an additional optional confidence map ([Fig. 3c ]) that combines multiple image quality metrics, which may be displayed concurrently
with the elastogram, and that guides the operator with respect to making measurements
(either online or retrospectively on stored images) in areas where confidence in the
SWS assessment is high.
ROIs with a side length of many centimeters may usually be set and 2D-SWE image penetration
depth may be at least 8 cm, depending on the system and transducer used. Extensions
to make 3 D quantitative SWE images (3D-SWE) have to date been limited to the use
of 3 D probes that contain mechanically swept 1 D transducer arrays. Such a feature
is available on the SuperSonic Imagine Aixplorer™, which is able to take advantage
of its real-time 2D-SWE capability to acquire a 3D-SWE image stack.
The majority of advantages, disadvantages and artefacts for SWE imaging remain as
discussed in [1 ]. Pressure of the probe may cause superficial tissues to stiffen because of the nonlinear
stress-strain character of tissue, but the resulting high shear wave speed artefact
is generally not a problem for liver imaging because it only occurs near the surface.
Note that an absence of echo signal on the B-mode image does not always produce an
absence of SWE signal, as should happen in a cystic structure (e. g. [Fig. 6 ], left). The artefactual SWE signal may result from an interaction between the relatively
large slice thickness typical of most SWE images and limited elevational dimensions
of the cystic structure. Similar influences of the background on quantitative shear
wave speed values for any small structures, whether solid or cystic, stiff or soft,
can be expected.
Penetration depth of the various technologies
When SE relies on the application of a quasi-static surface deformation rather than
internal physiological motion, the limited area of the probe (among other factors)
causes the strain to decay rapidly with depth. This limits the use of SE to only the
most superficial regions of the liver, although the use of a probe footprint extender
may increase the depth over which strain information is useful.
For the dynamic deformations employed by the SWE methods, as with ultrasound waves,
the attenuation of shear waves in tissue increases with frequency, except that the
attenuation coefficient for shear waves is about 1014 times that for ultrasound at the same frequency [13 ]. As a result, very low force-vibration frequencies must be used to generate shear
waves for elastography. When surface mechanical sources are used to create the shear
waves (as in TE), shear wave penetrations of many centimeters are needed, requiring
the use of frequencies of tens of Hz. When focused ARFI is used to create the shear
waves, the shear waves do not need to reach the depth of interest by propagation from
the surface and higher shear wave frequencies (hundreds of Hz) may be employed because
the penetration is achieved by the ultrasound beam used to generate the shear wave.
Even though the shear wave may then travel only a short distance (because of its higher
frequency), another shear wave may easily be generated at a new source position using
another ARFI beam.
Although the wavelength is about a thousand times shorter for shear waves than for
ultrasound waves of the same frequency, the fact that the frequencies are so low for
SWE means that wavelengths are in the region of many centimeters (wavelength = speed
divided by frequency). Shear wave sources such as surface thumpers or focused acoustic
radiation force impulses are typically therefore smaller than a wavelength, producing
diverging shear waves. Nevertheless, the method described above for ultrasonically
sweeping the ARFI focus faster than the SWS, which has been used to create a line-like
source deep in the tissue, could be used to create a source that is to some extent
focused.
kPa or m s–1 ?
As mentioned in 1.2c) some SWE systems allow a choice for the image scale and the
measurements to be reported in units of either ms–1 or kPa. There is more than one reason why it is preferable to report results in units
of ms–1 rather than kPa. First, this is the unit of shear wave speed cs , which is the quantity measured by the scanner. If the user requests the results
in kPa, the unit of elastic modulus, the scanner must convert the measured data to
an elastic modulus, such as Young’s modulus E. This conversion typically makes use
of a very simple equation, E = 3ρcs
2 . As with most simple equations, many assumptions are made and these are generally
not valid. In this case, these assumptions include that:
The tissue density is always 1000 kg m–3 . This may not always be true, although the error resulting from the assumption may
be small.
There is no variation of the elastic modulus with the magnitude of applied force (i. e.,
either the push magnitude that generates the shear wave or any pre-stress, such as
probe pressure or portal hypertension, upon which the push is superimposed). This
is equivalent to assuming that the tissue’s elastic response is linear. In general,
this is not true and indeed, in the future it is expected that the amount of non-linearity
will be found to be an important tissue property.
There is no variation of the elastic modulus with the shear wave frequency. This is
equivalent to assuming that the tissue is purely elastic, although in fact all tissues
are viscoelastic, causing the tissue to become stiffer the faster it is strained,
i. e., the higher the frequency.
There is no variation of the elastic modulus with direction, i. e., the tissue is
mechanically isotropic, although almost all tissues are on some length scale anisotropic.
There are no tissue boundaries or structures nearby which may cause the wave to travel
as another kind of wave (e. g., a Lamb or Rayleigh wave), or shear wave scatterers,
all of which will invalidate the assumed relationship between speed and elastic modulus.
The fact that these assumptions may not be correct for many tissues makes the elastic
modulus an indirect measurement. Accurate conversion of measured speed to Young’s
modulus requires much more complex equations and many more measurements, and is in
general impossible. Shear wave speed is the direct measurement, and the U.S. Food
and Drug Administration has thus far only approved systems for sale that use numerical
elastography displays in units of SWS.
Second, tissue elastic modulus values in the literature are often given for G, the
shear modulus, rather than Young’s modulus. In particular, this is the convention
in the MRE literature. Both moduli have units of kPa, which may mislead into thinking
that data can be pooled, whereas the numerical values for the two moduli are different.
Conversion using the relation E = 3G is possible, under the assumption that the tissue
is incompressible, but to apply this relationship authors must state which modulus
has been measured.
Reporting results in terms of SWS (ms–1 ) would appear to resolve these difficulties. However, it does not free the scientist
from having to provide full details of all of the above experimental variables (a-e),
particularly the shear wave frequency, since the SWS will vary with each of them.
Finally, although it is advisable to report SWS rather than elastic modulus, there
are complexities. First, measurements of viscoelasticity are likely to be provided
on ultrasound elastography systems. This is already the case for MRE, where results
are often cited in the form of a complex shear modulus with real and imaginary components.
This means that the situation for ultrasound elastography will have to be kept under
review, although a preferred method (rather than in the future switching to an indirect
measurement such as the complex shear modulus) may be to report direct measurements
such as SWS and shear wave attenuation coefficient, and their frequency dependences.
It is important to realize that because of the squared relationship between modulus
and speed, means and standard deviations calculated for data recorded in ms–1 and then converted to kPa will not be equal to the means and standard deviations
calculated after first converting the original data values to kPa. This will be true
across patient datasets and for means and standard deviations calculated over ROIs
for inhomogeneous tissues. Consideration of this issue is essential when trying to
pool data.
Comparability of data between the different SWE technologies
As already indicated, measured values of SWS will vary with a number of system factors,
in particular shear wave vibration frequency and bandwidth. In addition, measurement
bias may occur due to the software method employed, for example, to calculate relative
shear wave arrival time and speed. Thus, although it is always a good idea to report
as many of the experimental and system variables as possible, at present, corrections
for the effects of these variables is not possible. Therefore, unless a specific study
is undertaken to derive an experimental factor in the subject population of interest,
to correct for the biases of different equipment, data should not be pooled when acquired
using more than one product. Importantly, stiffness thresholds for clinical use known
for specific equipment should not be utilized for other equipment. Unless proven otherwise,
this recommendation should be taken to apply not only to equipment from different
manufacturers but even to different equipment from the same manufacturer or different
settings for the same equipment. For example, depth-dependent biases in SWS measurement
have been reported. Thus, depth setting standardization should be employed when accumulating
data within a population.
Investigator education
EFSUMB is working to promote high quality in ultrasound education and sustain excellent
professional standards in elastography training and practice [14 ]. To ensure the lowest possible intraoperator variability, EFSUMB recommends that
ultrasound elastography be performed by operators that have passed competence Level
1. However, it may also be possible to train dedicated personnel to perform elastographic
measurements only [15 ]. It is known from TE [16 ]
[17 ] that the required learning period is short and this has also been observed in studies
using ultrasound elastography [18 ]
[19 ]. Nevertheless, it appears that experience in liver ultrasonography plays a role
in the ability to perform stiffness measurements, at least by means of shear wave
imaging, and leads to more reliable elastographic measurements, particularly in obese
subjects [20 ].
Operator experience significantly influences the reliability of liver stiffness measurement
(LSM) performed with TE, with an operator with less than 500 procedures representing
a risk of 2.6 fold for unreliable results [21 ]
[22 ]. TE can be readily learned by a nurse or a technician after approximately 100 examinations
[16 ]
[17 ]. 100 examinations should be considered the minimum required training, and training
including > 500 examinations yields an experienced TE operator.
There is no agreement as to what constitutes an experienced operator for pSWE and
2D-SWE. Proposed definitions for 2D-SWE include having performed > 300 abdominal ultrasound
scans or > 50 supervised 2D-SWE examinations [20 ]
[23 ].
A major question remains: Who should be allowed to acquire and interpret elastography
data? To date, there is insufficient data in the literature to answer this question
adequately. However, TE acquisition can be performed by nurses or dedicated persons.
Training for non-clinicians should insist on reporting standards. Interpretation should
be performed by a physician with knowledge of the clinical condition of the patient.
The key issue is to obtain adequate competence and training and to perform elastography
in accordance with the medico-legal regulations of the relevant country. The following
recommendations are in accordance with the EASL guidelines.
The basic principles and technology for elastography were developed by the academic
research community before commercial translation, and it remains a heavily researched
and rapidly developing field. EFSUMB recommends that users maintain an awareness of
this field.
The operator must acquire appropriate knowledge and training in ultrasound elastography
(LoE 5, GoR C). Strong consensus (13/0/0, 100 %)
Data acquisition should be undertaken by dedicated and specially trained personnel.
For pSWE and 2D-SWE, experience in B-mode ultrasound is mandatory (LoE 5, GoR C).
Strong consensus (13/0/0, 100 %)
Shear wave elastography (TE, pSWE and 2D-SWE), general technical comments
Shear wave elastography (TE, pSWE and 2D-SWE), general technical comments
Introduction
The liver is an important target organ for the use of elastography; stiffness correlates
with the degree of fibrosis and indirectly with portal hypertension (see liver application).
Examination procedure
Subjects should be examined in a supine position with the right arm in maximal extension.
The transducer is positioned in a right intercostal space to visualize the right liver
lobe in A or B mode. Artifacts and large vessels on the A-mode (TE) or B-mode image
(pSWE and 2D-SWE) should be avoided. Optimal measurement quality of pSWE and 2D-SWE
occurs with the ROI placed a minimum of 1 – 2 cm and a maximum of 6 cm beneath the
liver capsule [24 ]
[25 ]
[26 ]
[27 ]. In the authors’ opinion, the focus should be placed at the level of the ROI.
Deep inspiration affects the measurement. A transient breath hold in a neutral position
is optimal. Details of the examination techniques are documented in the specific sections.
Measurements in the left liver lobe are significantly higher and show more variability
than those in the right lobe [28 ]
[29 ]. 2D-SWE LSMs correlate best with the stage of liver fibrosis when obtained in the
upper right lobe of the liver [30 ]. LSM in the left liver lobe should be avoided. The time required for the measurement
with TE, pSWE and 2D-SWE is usually < 5 minutes [31 ].
Measurement of liver stiffness by SWE should be performed through a right intercostal
space in supine position, with the right arm in extension, during breath hold, avoiding
deep inspiration prior to the breath hold (LoE 2b, GoR B) [1 ]
[32 ]. Strong consensus (18/0/0, 100 %)
Measurement of liver stiffness by SWE should be performed by experienced operators
(LoE 2b, GoR B) [1 ]
[32 ]. Strong consensus (18/0/0, 100 %)
Measurement of liver stiffness by pSWE and 2D-SWE should be performed at least 10 mm
below the liver capsule (LoE 1b, GoR A) [24 ]
[25 ]
[26 ]
[27 ]
[32 ]
[33 ]
[34 ]
[35 ]. Strong consensus (18/0/0, 100 %)
Fasting and resting
Food ingestion increases measurement readings (independent of fibrosis) for an estimated
120 – 180 minutes after the meal [36 ]
[37 ]
[38 ]. Moreover, food intake significantly increases the median interquartile range (IQR)
values at 30 min and 120 min compared with the baseline IQR [39 ]. It can cause misclassification of subjects by overestimating the stage of fibrosis,
as reported in subjects infected with the hepatitis C virus (HCV). It may even result
in overstaging of subjects as pre-cirrhotic or cirrhotic, with a normal fasting value
being confirmed by liver biopsy. There are reports that the measurement decreases
to true values about 180 minutes after eating [36 ]. The examination should ideally be performed after an overnight fast, while abstaining
from food/drinks (especially caffeine) and smoking. In addition, since exercise increases
liver stiffness, subjects should be examined after a minimum of 10 – 20 minutes of
rest [40 ].
Patients should fast for a minimum of 2 hours and rest for a minimum of 10 minutes
before undergoing liver stiffness measurement with SWE (LoE 2b, GoR B) [1 ]. Majority consensus (13/2/3, 72 %)
Factors influencing liver stiffness independent of liver fibrosis (confounders)
Liver stiffness does not solely reflect liver fibrosis, but can reflect many other
physiological or pathological conditions. The majority of these variables have been
described in studies addressing TE, but should be the same for other SWE methods.
Liver stiffness is increased with hepatic inflammation (often but not exclusively
shown by an elevated transaminase level) [41 ]
[42 ]
[43 ]
[44 ], obstructive cholestasis [45 ] and hepatic congestion [46 ]
[47 ]. Also acute toxic hepatitis (phenprocoumon-induced liver failure) increases measurements
[48 ]. For patients with falsely elevated LSMs due to alcoholic hepatitis, liver stiffness
decreases following 1 – 4 weeks of abstinence [49 ]
[50 ]
[51 ]. Other diseases, which cause increased liver stiffness, independent of liver fibrosis
include amyloidosis, lymphomas and extramedullary hemopoiesis. Presently, it is uncertain
whether hepatic steatosis modulates liver stiffness [52 ]
[53 ] or does not [54 ]
[55 ].
The major potential confounding factors (liver inflammation indicated by AST and/or
ALT elevation > 5 times the normal limits, obstructive cholestasis, liver congestion,
acute hepatitis and infiltrative liver diseases) should be excluded before performing
LSM with SWE, in order to avoid overestimation of liver fibrosis (LoE 2b, GoR B),
and/or should be considered when interpreting the SWE results (LoE 1b, GoR B) [41 ]
[42 ]
[43 ]
[44 ]
[45 ]
[46 ]
[49 ]
[50 ]
[51 ]
[56 ]
[57 ]
[58 ]
[59 ]. Broad consensus (15/0/1, 94 %)
Normal values
TE measurements of Young’s modulus in healthy people vary between 4.4 and 5.5 kPa
(95th percentile 6.7 kPa) [60 ]
[61 ]
[62 ]
[63 ]. LSMs are generally higher in men than in women [60 ]
[61 ] and may be affected by steatosis [60 ] but are not influenced by age [62 ]
[63 ].
pSWE measurements using Virtual Touch Quantification (VTQ® ) in healthy populations range between 1.01 and 1.59 m/s, but in most studies the
range is 1.07 – 1.16 m/s [28 ]
[29 ]
[33 ]
[34 ]
[64 ]
[65 ]
[66 ]
[67 ]. Age has no apparent influence on the SWS assessed by VTQ® [28 ]
[29 ]
[67 ]. All but one study [33 ] similarly found no correlation between gender or body mass index (BMI) and SWS values.
Depth as assessed by the skin-to-liver capsule distance may influence the SWS values
assessed by VTQ® [28 ]. In healthy children, the mean SWS obtained in the right liver lobe was 1.07 ± 0.10 m/s
in one study [68 ] and 1.12 m/s (range: 0.73 to 1.45 m/s) in another[69 ]. Regarding the influence of age, there are conflicting results, since one study
reports a statistically significant difference of SWS values obtained in children
< 6 years compared to those obtained in older children, while another study did not
find any difference [68 ]
[69 ]. A depth of SWS measurements of up to 5 cm beneath the skin did not influence accuracy
[68 ]
[69 ].
Values obtained with Elastography point quantification (ElastPQ® ) in healthy people are comparable to those obtained with VTQ® [70 ]
[71 ]
[72 ], although in contrast to VTQ® findings, measurements using ElastPQ® were 8 % higher in healthy men than in healthy women [72 ].
2D-SWE measurements of Young’s modulus using supersonic shear imaging (SSI) in healthy
subjects cover the range 4.5 – 5.5 kPa (95th percentile 6.2 kPa) [73 ]
[74 ]. Healthy men may have higher LSMs than healthy women, while BMI and age do not seem
to influence LSM in subjects without liver disease [74 ].
For all equipment, an SWE measurement within the normal range, in a subject without
other clinical or laboratory evidence of liver disease, may exclude significant liver
fibrosis with a high degree of certainty.
The current literature has been recently summarized [75 ].
SWE within the normal range can rule out significant liver fibrosis when in agreement
with the clinical and laboratory background (LoE 2A, GoR B) [60 ]
[61 ]. Broad consensus (17/0/1, 94 %)
Transient elastography (TE)
Transient elastography (TE)
Procedure
Transient elastography uses an ultrasound displacement M-mode and A-mode image produced
by the system. The operator locates a portion of the liver at least 6 cm thick and
free of large vascular structures. By pressing the acquisition button, the machine
displays the median of the measured Young’s modulus in kPa, the IQR (the difference
between the 75th and the 25th percentile), IQR/median (IQR/M), the value of the current
measurement and, only in the old version of the system, the success rate (the ratio
between valid and total number of acquisitions). The system displays a result only
if the acquisition is valid, since the software automatically rejects acquisitions
without correct vibration shape or a correct follow-up of the vibration propagation
[1 ]
[32 ]
[76 ].
The new version of the system suggests using the M or the XL probe in adult subjects
based on calculations of the skin-to-liver capsule distance. For children as well
as in adults with a thoracic circumference ≤ 75 cm, the S probe is recommended, either
S1 for a thoracic circumference < 45 cm or S2 for 45 – 75 cm [77 ].
How to measure?
Following the manufacturer’s recommendation, assessment is reliable when 10 valid
readings and an IQR ≤ 30 % of the median (IQR/M ≤ 30 %) are obtained. The majority
of studies have used these reliability criteria as well as a success rate ≥ 60 %.
Areas under the receiver operator characteristics (AUROCs) curves were compared for
the prediction of cirrhosis between the median values of the 3 first, the 5 first
and the 10 first successful acquisitions of TE [78 ]. No significant loss in performance was reported if only 5 valid acquisitions were
obtained. It is likely that IQR/M is the key factor for discordance between TE and
histological fibrosis stage [79 ]
[80 ]. In a recent large population study, new reliability criteria based on IQR/M and
median LSMs were proposed [81 ]. According to this study, TE results should be classified as “very reliable” if
IQR/M ≤ 10 %, regardless of LSM; “reliable” if 10 % < IQR/M ≤ 30 %, or IQR/M > 30 %
with Young’s modulus < 7.1 kPa and “poorly reliable” if IQR/M > 30 % with Young’s
modulus ≥ 7.1 kPa. However, these criteria have not been externally validated. A reliable
TE assessment can be achieved in over 90 % of adults, when both the M and XL probes
are used as required [82 ]
[83 ]
[84 ]
[85 ]. Because the M probe takes measurements between 25 and 65 mm from the probe, to
increase viability, those patients with a skin-to-liver capsule distance (SCD) of
> 25 mm should be assessed with the XL probe. In practice, < 8 % of patients with
a body mass index (BMI) < 30 kg/m2 have an SCD > 25 mm. The rate of SCD > 25 mm increases to 50 % for patients with
a BMI of 35 – 40 kg/m2 [86 ].
Although the problem of overweight patients is not completely resolved with the XL
probe [85 ]
[87 ], the applicability of TE increases when both the M and XL probes are available [82 ]
[83 ]
[84 ]
[85 ]
[88 ]
[89 ]
[90 ]
[91 ]. The diagnostic accuracy of the XL probe appears similar to that of the M probe
but the Young’s modulus values are lower than those obtained with the M probe by a
mean of 1.5 kPa (range of 0.8 – 2.3 kPa) [82 ]
[83 ]
[84 ]
[85 ]
[92 ]. The lower values obtained with the XL probe can be explained by the fact that the
validation of the XL probe in overweight patients was undertaken in comparison with
the M probe. In these patients there is a high probability of SCD > 25 mm and the
measurements obtained with the M probe are likely to be more sub-capsular and, thus,
higher [93 ].
10 measurements should be obtained. An IQR/M ≤ 30 % of the 10 measurements is the
most important reliability criterion (LoE 1b, GoR A) [79 ]
[80 ]. Strong consensus (17/0/0, 100 %)
Values obtained with the XL probe are usually lower than with the M probe. Therefore,
no recommendation on the cut-offs to be used can be given (LoE 2B, GoR B) [82 ]
[83 ]
[84 ]
[85 ]
[87 ]. Broad consensus (13/1/3, 77 %)
Reproducibility
The intra- and interobserver agreements are excellent, with reported intraclass correlation
coefficients (ICC) above 0.90 [94 ]
[95 ]
[96 ]. The agreement decreases in overweight patients or in early stages of fibrosis [94 ]
[95 ]. Although the LSM seems to be reproducible at different examination sites, the best
examination site is the median axillary line on the first intercostal space under
the liver percussion dullness upper limit, with the patient lying in dorsal decubitus
[95 ].
Point shear wave elastography (pSWE)
Point shear wave elastography (pSWE)
Experience with point shear wave elastography (pSWE) has been mainly acquired with
the VTQ® product, because it was the first method available, subsequently followed by ElastPQ® and, more recently, by pSWE methods from many companies.
Procedure (how to measure?)
The operator can select the depth at which liver elasticity is evaluated by placing
a “measuring box” (size depending on the manufacturer) in the right liver lobe (segment
V, VIII or VII), via an intercostal approach and with the transducer at 90° in relation
to the liver capsule, in an area free of large vessels. Published studies [97 ]
[98 ]
[99 ] show that reliable measurements can be obtained in > 90 – 95 % of patients. In a
pSWE study using VTQ® to measure SWS [24 ], the best correlation with histological fibrosis was observed for measurements performed
1 – 2 cm and 2 – 3 cm beneath the liver capsule (0.675 and 0.714, respectively), but
in up to 15 % of cases, measurements could not be obtained if performed 2 – 3 cm under
the liver capsule.
Adequate B-mode liver image is a prerequisite for pSWE and 2D-SWE measurements (LoE
5, GoR D). Strong consensus (18/0/0, 100 %)
How many measurements?
Most studies perform 10 valid measurements by pSWE and report the median of these
values. A few studies have used only 5 [99 ]
[100 ] or 6 [101 ] valid measurements. Another study [102 ] calculated the mean and standard deviation (SD) of 10 valid measurements. A high
SD correlated with misclassification of fibrosis. Additionally, higher stages of fibrosis
were associated with a higher SD, indirectly indicating that "more" measurements should
be obtained in patients with suspected fibrosis.
The median value of at least 10 measurements should be used for liver elastography
by pSWE (LoE 2b, GoR B) [102 ] Strong consensus (18/0/0, 100 %)
Reproducibility
pSWE has excellent intra- and inter-operator reproducibility for liver elastography
assessment in both healthy subjects and patients with chronic liver disease [66 ]
[70 ]
[72 ]
[103 ]
[104 ]
[105 ].
Healthy subjects
In a study on SWS measurement, reproducibility was assessed by VTQ® in 50 healthy volunteers. A very good ICC (0.86) was found [66 ]. Age, gender and BMI did not influence the reproducibility of this method. Similar
results (ICC = 0.87) were obtained [103 ] in 20 healthy subjects. In a cohort of 69 healthy subjects, ElastPQ® showed very good intra- (ICC = 0.96) and interobserver (ICC = 0.93) reproducibility
[70 ].
Patients
The reproducibility of SWS assessed by VTQ® was evaluated in 61 patients with chronic hepatopathies [104 ]. The following cut-off values for liver fibrosis staging using SWS were applied:
F≥ 2 – 1.37 m/s; F≥ 3 – 1.45 m/s and F4 = 1.75 m/s. There was 87 % agreement between
repeated measurements for different stages of fibrosis. For the differentiation between
patients with significant fibrosis (F≥ 2 Metavir) and those with fibrosis stage F< 2,
the interoperator agreement was 90 %. In another study [19 ] excellent agreement (ICC = 0.84) was observed in a cohort of 50 patients with various
etiologies of chronic liver disease. Other authors [101 ] assessed the correlation of SWS values obtained with VTQ® by 2 operators in 41 patients with chronic hepatopathies. A good interoperator correlation
was obtained (Spearman correlation coefficient r = 0.874, p < 0.0001) with similar
results [99 ]
[100 ]
[105 ]. VTQ® reproducibility for SWS values and influencing factors were assessed in 33 and 58
patients, respectively. The overall intraoperator agreement was better than the interoperator
agreement: ICC 0.90 vs. ICC 0.81. For both intra- and interoperator reproducibility,
the ICCs were lower in women vs. men (0.88 vs. 0.91 and 0.67 vs. 0.86, respectively),
in patients with a high BMI (≥ 25 kg/m²) vs. BMI < 25 kg/m² (0.88 vs. 0.91 and 0.79
vs. 0.82, respectively), in patients with ascites vs. no ascites (0.80 vs. 0.93 and
0.78 vs. 0.84, respectively) and in non-cirrhotic vs. cirrhotic patients (0.77 vs.
0.82 and 0.70 vs. 0.83, respectively) [106 ]. The interobserver reproducibility of the ElastPQ® technique was assessed in 291 consecutive patients with chronic hepatitis B (CHB)
who underwent partial hepatectomy or liver biopsy due to hepatocellular carcinoma
(HCC). Very good reproducibility of this technique was achieved (ICC = 0.798) with
a similar conclusion reached in a further study [70 ].
Quality criteria
When VTQ® was first introduced, no recommendations regarding the use of quality criteria for
measurements were defined. One study [107 ] evaluated factors that influenced the correlation of SWS assessed by VTQ® with histological fibrosis in a cohort of 106 chronic hepatitis C (CHC) patients.
In univariate and multivariate analysis, an IQR/M ≥ 30 % was associated with a discordance
of at least 2 stages of fibrosis between SWS and histological fibrosis. The importance
of IQR as a quality criterion was further demonstrated in another study [108 ]. Others studied the importance of standard deviation (SD) as a quality criterion
[102 ]. 10 valid SWS measurements were adopted in this study and the mean value was calculated.
The authors obtained the following accuracy for predicting the presence of F≥ 2, F≥ 3
and F = 4 in all cohorts of patients: 80.7 %, 86.2 % and 88.7 %, respectively. After
the exclusion of patients with SD > 30 %, the accuracy of VTQ® for predicting the same stages of liver fibrosis increased significantly up to 86.1 %,
91.2 % and 91.5 %, respectively. Using ElastPQ® , a recent study has suggested that an IQR/M ≤ 30 % is the most important quality
criterion, whereas the number of measurements seems not to affect the performance,
provided that they are at least five [109 ]. Thus, the compliance with quality criteria may increase the diagnostic accuracy
of pSWE [102 ]
[107 ]. Quality parameters have been described for other manufacturers as well [110 ].
2D-SWE
Almost all 2D-SWE studies for liver applications have been carried out using SSI,
because other companies have only recently introduced 2D-SWE products. This description
is therefore limited to the SSI system, but the principles may be applied to other
2D-SWE products.
Procedure
Obtaining an elastogram
2D-SWE evaluation should be performed in a well-visualized area of the right liver
lobe, free of large vessels, liver capsule, ligaments and the gallbladder [30 ]. Since movement greatly influences results, the subject is asked to suspend breathing.
With 2D-SWE working in continuous and not with single shot emissions, the SWE acquisition
is continued for 4 – 5 seconds (can be longer for other scanning systems) once a stable
SWE image is obtained. The operator should aim to achieve homogeneous color filling
of the SWE ROI. With SSI, the scale of the elastogram can be modified. Usually a Young’s
modulus scale of up to 30 kPa is sufficient, but a higher scale of up to 150 kPa can
be adopted on a case-by-case basis. The operator freezes the image (and optionally
saves the clip for further post-processing) and an analysis box (Q Box, for SSI) is
placed on the most homogeneous, stable elastogram for a few seconds to measure Young’s
modulus (SWS, if the scanner is set to that mode). Experience seems to play a role
in reliable measurement, particularly in obese patients [20 ]. However, there is no unified understanding of what constitutes an experienced 2D-SWE
operator. Adequate ultrasound knowledge with at least 300 abdominal ultrasound investigations
may be considered a minimum requirement [20 ]. 2D-SWE measurement of liver elasticity should be performed by experienced operators
after achieving a stable, homogeneous elastogram [20 ]
[23 ].
How to measure?
Analysis box size and shape
For 2D-SWE measurements, the analysis box should be set to at least 10 mm, preferably
15 mm or more. A round shape is usually chosen [57 ]
[58 ]. The ROI should be placed over an isoechoic area of liver parenchyma, as seen on
the grayscale image (no vessel, no nodule, no other structure), in priority in the
middle line of the elastogram (avoiding positioning the Q Box on the edges of the
elastogram), while also avoiding SWS artefactual areas (reverberation, noisy areas
from rib shadowing).
Valid and invalid measurements
There is no agreement on objective quality criteria. Some authors suggest that a minimal
Young’s modulus value of ≤ 0.2 kPa in the analyzed region is useful to identify invalid
measurements as indicated by a lack of concordance with TE [111 ], while others use a minimal Young’s modulus < 1 kPa. Furthermore, among valid measurements
an IQR/M ≤ 30 % is recommended by other studies mimicking TE reliability criteria.
The latter criterion, however, is not supported by evidence. For 2D-SWE with Logiq
E9 (GE), the manufacturer recommends an IQR/M value below 30 % as a quality criterion.
Temporal stability of the elastogram for three seconds or more during breath hold,
in combination with placement of the analysis box in a homogeneous area with complete
filling results in high accuracy, high reliability and low variance of measurements
with SSI [112 ]
[113 ]
[114 ]. The new software version of the Aixplorer® system also shows the stability index (SI) and according to the manufacturer a reliable
LSM should exclude measurements with an SI < 90 %. Aplio 500 (Toshiba) provides a
display of shear waves travelling within the box, allowing selection of areas not
affected by artefacts for analysis. For 2D-SWE with the Philips system, a confidence
map guides the operator to perform measurements in areas where the signal-to-noise
ratio of the SWS assessment is high.
Data regarding the best criterion for judging a measurement as reliable are limited.
Hence, following the manufacturer’s advice for acquisition is suggested in the absence
of better evidence [56 ]
[112 ]
[113 ]
[114 ].
How many measurements?
From 3 to 15 measurements are used in published studies, but data from several studies
suggest that 3 measurements suffice to obtain consistent results for the assessment
of liver fibrosis and portal hypertension, and for optimal correlation with TE [27 ]
[58 ]
[59 ]
[114 ]
[115 ]
[116 ]. Some studies report the median, some the mean, of the available measurements. There
is no convincing evidence to suggest superiority of the mean versus median of the
SSI measurements. However, since the median and IQR are robust against non-normal
distributed data, they should be preferred for reporting.
For 2D-SWE a minimum of three measurements should be obtained; the final result should
be expressed as the median together with the interquartile range (LoE 2b, GoR B) [27 ]
[117 ]. Strong consensus (18/0/0, 100 %)
Reproducibility
Reproducibility in healthy subjects
In three studies, the intraobserver reproducibility of SSI during the same session
was excellent (ICC ranged from 0.92 to 0.95) [23 ]
[73 ]
[118 ]. The interobserver agreement on different days was affected by operator experience
and ranged from 0.63 to 0.84 [23 ]
[118 ].
Reproducibility in patients
The intraobserver reproducibility of 2D-SWE for liver stiffness assessment in liver
fibrosis patients is excellent, with the ICC ranging from 0.90 to 0.95 in published
studies [56 ]
[57 ]
[117 ]. The intra-subject reproducibility (evaluated over periods of 2 days to 4 weeks)
ranges between 0.83 and 0.90 [31 ]. Interobserver reproducibility on the same day ranges from 0.83 [31 ] to 0.94 [112 ]. Intra- and interobserver variance may be inferior to pSWE using VTQ® [31 ].
Limitations
Failures
Failure – most often defined as the “inability to obtain an adequate signal for all
acquisitions” – occur in up to 10 % of cases reported in the published series. Common
causes of failure are: depth below 4 – 5 cm [26 ], poor ultrasound window, reverberations, pulsatile movement, poor breath hold, large
amounts of ascites [117 ], intercostal wall thickness ≥ 25 mm [119 ], BMI ≥ 30 kg/m2 , histological steatosis and waist circumference ≥ 102 cm [57 ]
[59 ].
Unreliable assessments
Unreliable measurements are randomly defined across studies and comparison is limited.
Reported predictors of unreliable results with SSI are similar to those reported for
TE and pSWE [56 ]
[117 ]. In one study, a lower SD and larger analysis boxes independently correlated with
correct classifications of severe fibrosis and cirrhosis [114 ].
The main factors limiting the applicability of 2D-SWE include obesity, poor acoustic
window or presence of artefacts and inability of the subjects to hold their breath
[26 ]
[56 ]
[57 ]
[59 ]
[117 ].
Strain elastography (SE): quasi-static strain imaging
Strain elastography (SE): quasi-static strain imaging
Procedure
The best B-mode ultrasound image is required since real-time strain elastography (SE)
depends on the quality of the B-mode images. Acquisition can be improved as follows
[32 ]:
Image the right liver lobe through an intercostal space with the patient supine and
the right arm elevated to widen the intercostal space;
Place the transducer lightly on the skin without moving it since intrinsic, mainly
cardiac, movement is often enough to displace the tissue. However, sometimes the operator
needs to induce repeated pressure movements with the probe to create the necessary
stress needed for strain measurements;
Select a region of interest in which B-mode images are free from interfering structures
and artefacts;
Obtain images displaying axial, not lateral or elevational, movement by pointing the
transducer toward the heart;
Have the patient perform a short breath hold to ensure that SE images are displayed
consistently [120 ].
Analysis region placement and size
Similarly to pSWE and 2D-SWE, the analysis box should be placed at least 1 cm below
the liver capsule [120 ]
[121 ]. Measurement should not be performed too deep in the parenchyma as acoustic shadows,
reverberation artefacts, and a lack of sufficient penetration will generate incorrectly
high readings. Some researchers include the surrounding tissues, such as the subcutaneous
and muscle layers [122 ]
[123 ]. However, placing the ROI entirely inside the liver is key to generating uniform
images to calculate the liver fibrosis (LF) index [124 ]. Large vessels should be avoided by using a 2.5 × 2.5 cm ROI [120 ]
[125 ]. Elimination of artefacts requires attention to technique. Trialling the placement
of the transducer between the ribs will lead to optimal positioning. When an examination
is difficult, attempt another intercostal space, perhaps selecting one that is softer
and with a thinner subcutaneous layer. Other subcutaneous structures, such as the
ribs and lungs, should not be included in the image. For the analysis, frames with
strain generated in the depth direction with no artefacts should be selected. Good
images may be obtained at the end of diastole with electrocardiographic gating or
at the largest downward wave.
Evaluation methods
An examiner’s experience influences the outcome of visual assessment. To overcome this,
various quantitative and semi-quantitative methods have been developed to assess tissue
strain more objectively. In the research literature “image pattern recognition” indices
obtained by analyzing grayscale, histogram, and binarization transformations of SE
images have been shown to correlate with liver fibrosis. The strain image is converted
to numerical values, for example using color gradations, with blue being 0 and red
being 255. Some authors have reported that mean strain values inversely correlated
with liver stiffness and fibrosis in patients with CHC. On the other hand, the standard
deviation of values of strain, the percentage of area of low strain and its complexity
were positively correlated with liver stiffness and fibrosis [126 ].
Scores
Different scores for semi-quantitative interpretation of SE have been proposed: the
German Elasticity Score [121 ], the Japanese Elasticity Score [126 ] and the LF Index [127 ]. The LF Index (an algorithm based of nine features evaluated directly by software)
is the only one recommended by the manufacturer [32 ]. However, evidence is not sufficient to allow recommendation for clinical use, at
least in European patients [121 ]
[125 ].
Reproducibility (published literature)
The interobserver consistency among four separate locations in the liver using a semi-quantitative
method was almost perfect (ICC 0.97) [128 ]. On the other hand, the intraobserver variability and intraobserver agreement of
SE for the assessment of liver fibrosis have been criticized in several studies [121 ]
[125 ], but no new data are available.
Methods to objectively assess strain are being developed, but currently cannot be
recommended in clinical practice (LoE 5, GoR D). (Consensus 18/0/0, 100 %).
Comparison of results between systems
Comparison of results between systems
Introduction
Different US-based SWS technologies are available for the noninvasive assessment of
liver fibrosis and the measurements produced can be slightly to moderately different
between systems from different manufacturers. Even systems that use the same technique
but were developed by different manufacturers can yield different values due to different
and proprietary methods to measure the SWS. Moreover, the units of measurement may
be meters per second if the speed of the shear wave is expressed as m/s or kPa if
used to calculate the Young’s or shear modulus. There is another potential difference:
it is assumed that the SWS is independent of the shear wave frequency which is incorrect
since higher frequencies generate shear waves that travel faster. The ‘dispersion
error’ could render results from different studies incomparable.
Studies on phantoms
The Ultrasound Shear Wave Speed technical committee of the Radiological Society of
North America, Quantitative Imaging Biomarker Alliance (QIBA) has quantified the differences
between commercially available systems. An inter-laboratory study comparing shear
wave velocities obtained with four different machines (Fibroscan, Philips iU22, ACUSON
S2000 and Aixplorer) was conducted [129 ]. Working on elastic phantoms, a statistically significant difference in the SWS
estimates among systems and depth of measurement in the phantom was shown, whereas
no statistically significant differences were found among operators using the same
or equivalent systems under the same conditions [129 ]. Similar results were obtained using phantoms with viscoelastic properties similar
to those observed in normal and fibrotic liver [130 ]. The measurements were performed at multiple focal depths (3.0, 4.5 and 7.0 cm).
The deepest focal depth (7.0 cm) produced the greatest inter-system variability for
each phantom (up to 17.7 %) as evaluated by the interquartile range. Inter-system
variability was consistent across all phantoms and was not related to stiffness.
Sources of variability
Several sources of variability are detailed in published studies, including technical
and patient-dependent factors that could affect comparability between systems.
Technical factors
Measurement depth
As shown by the studies on phantoms [129 ], the influence of depth on the estimation of elastic properties is not negligible.
Furthermore, with curved transducers used in liver imaging, the angle affects the
readings, with the best results being achieved when the ROI is straight ahead. Using
VTQ® , it has been shown that the results with the lowest variability are obtained at a
depth of 4 – 5 cm with a convex transducer (1 – 4 MHz; mean push pulse: 2.67 MHz)
and at a depth of 2 – 3 cm with a linear transducer (4 – 9 MHz; mean push pulse: 4 MHz)
[131 ]. The acoustic push pulse is progressively attenuated as it traverses the tissue.
Attenuation is higher in a stiffer liver. Thus, measurements are more variable in
cirrhotic patients [25 ].
Frequency of the transducer
In a prospective study on 89 patients with CHC, pSWE (based on VTQ® ) was performed using both available transducers (4C1 and 9L4) [132 ]. The linear transducer gave higher values (1.91 ± 0.87 m/s vs. 1.70 ± 0.67 m/s).
However, the results were correlated to each other (r = 0.70). Using the same method
in a phantom and a series of eight volunteers, it was found that the convex transducer
showed values that were significantly higher than those obtained with the linear transducer
[131 ].
Position of the transducer
The highest intra- and interobserver agreement was obtained for the measurements performed
through the intercostal space rather than the subcostal approach [72 ]
[103 ]
[133 ].
Operator experience
Methods using pSWE have shown excellent interobserver agreement, with concordance
ranging from 0.80 to 0.97 for measurements performed via an intercostal approach,
and independent of operator experience, suggesting that operators require only a short
period of training to perform reliable LSMs [66 ]
[70 ]
[103 ]
[106 ]. Using 2D-SWE (SSI) an expert operator had higher reproducibility of measurements
over time than a novice operator [23 ]. It is suggested that at least 50 supervised 2D-SWE measurements should be performed
by a novice operator in order to obtain consistent measurements.
It was also demonstrated that a novice was able to achieve reliable results with TE
after a single learning session. However, the success rate progressively increased
with the experience of the operator [16 ]. On the contrary, it was reported that operator experience increases the applicability
and diagnostic accuracy of TE [92 ]. Echosens, the manufacturer of the FibroScan device, recommends that TE be performed
by an experienced operator (> 100 examinations).
Using liver biopsy as the reference standard, a recent study has shown that ElastPQ® matched TE for accuracy after the operator had performed at least 130 examinations
[134 ].
Equipment
Proprietary elastographic technologies generally give different estimates of the SWS
within the same liver. This translates into the need to define threshold values for
fibrotic stages for each specific equipment model.
Patient-dependent factors
The most important patient-dependent factors for variability between different elastographic
methods are: position (supine position has been used in most published studies), respiration
phase (deep inspiration affects the measurement; it has been suggested that a breath
hold for a few seconds during quiet breathing may lead to the best results) [25 ]
[29 ]
[135 ], non-fasting state [36 ]
[37 ]
[38 ], fibrosis stage (lower stages of liver fibrosis seem to reduce TE reproducibility
[94 ]
[136 ]
[137 ] or the presence of steatosis [52 ].
Morbid obesity limits the applicability of elastography methods (TE XL probe, pSWE)
and impairs measurement accuracy [138 ]. The theoretical possibility that values are different rather than identical in
different parts of the liver, even in the same region, may also contribute to the
variability between different equipment reported in many studies. Unfortunately, there
is no indication in most studies as to whether the same sampling position and the
same intercostal space were predetermined and utilized.
Clinical studies
Diagnostic accuracy
The AUROCs of ElastPQ® and VTQ® for predicting the presence of liver fibrosis were similar (0.85 and 0.82, p = 0.48)
[71 ]. Other authors compared 2D-SWE (SSI), VTQ® , and TE in 349 consecutive patients with chronic liver diseases [57 ]. They found that 2D-SWE, TE and VTQ® correlated significantly with the histological fibrosis score (r = 0.79; r = 0.70;
r = 0.64 respectively, all p < 0.00 001). The AUROCs of 2D-SWE, TE, and VTQ® were 0.89, 0.86, and 0.84, respectively, for the diagnosis of mild fibrosis; 0.88,
0.84, and 0.81, respectively, for the diagnosis of significant fibrosis; 0.93, 0.87,
and 0.89, respectively, for the diagnosis of severe fibrosis; and 0.93, 0.90, and
0.90, respectively, for the diagnosis of cirrhosis. 2D-SWE had a higher accuracy than
TE for the diagnosis of severe fibrosis (≥F3) (p = 0.0016), and a higher accuracy
than VTQ® for the diagnosis of significant fibrosis (≥F2) (p = 0.0003). No significant difference
was observed for the diagnosis of mild fibrosis and cirrhosis. Similar comparison
was made in patients with NAFLD [57 ]: 2D-SWE had a higher accuracy than VTQ® for diagnoses of significant fibrosis (≥F2) (p = 0.004).
In a study on 132 patients with chronic liver disease, 2D-SWE, pSWE (VTQ® ) and TE were compared. The reference methods were liver biopsy for the assessment
of liver fibrosis (n = 101) and magnetic resonance imaging/computed tomography for
the diagnosis of liver cirrhosis (n = 31). No significant difference in diagnostic
accuracy was found between the three elastography methods for the diagnosis of significant
and advanced fibrosis and liver cirrhosis in the "per protocol" and "intention to
diagnose" cohorts [57 ].
Several studies have compared the results obtained with B-mode-based SWE techniques
(pSWE, 2D-SWE) to those of TE and have found that the accuracy of these techniques
is similar to that of TE [59 ]
[139 ]
[140 ].
Some comparative studies show similar accuracy for different elastographic systems.
Larger prospective studies are necessary to find if there are differences in accuracy
between each different system for liver stiffness evaluation.
The results with the lowest variability in comparing different pSWE or 2D-SWE systems
were obtained at a depth of 4 – 5 cm from the transducers (with convex transducers)
(LoE 4, GoR C) [131 ]. Accordingly, this location is recommended if it is technically suitable. Broad
consensus (17/0/1, 94 %)
Liver diseases
Introduction
The assessment of fibrosis in chronic liver diseases is pivotal for prognosis and
guiding management, including whether to commence antiviral treatment. Liver biopsy
is considered the “gold standard” for fibrosis assessment and stage classification
and can also grade necro-inflammatory activity. However, liver biopsy is limited by
its invasiveness, with potential severe complications in up to 1 % of cases, sampling
error, since the specimen represents roughly only 1/50 000 of the liver volume, and
the inter- and intraobserver variability at microscopic evaluation. Therefore, noninvasive
methods for liver fibrosis assessment including ultrasound elastographic methods have
been an intense field of research [1 ].
Due to the differences among elastography methods outlined in previous sections, cut-off
values for fibrosis are system-specific and cannot be equated across machines. Currently,
TE is the most widely studied method outside of radiology practice, but a considerable
amount of evidence regarding other methods has been published. The results obtained
by pSWE (mainly VTQ® ) and 2D-SWE (mainly SSI) show similar accuracy as compared to TE for the assessment
of liver stiffness, with a potentially higher applicability [59 ]
[112 ].
It is becoming increasingly clear that the best cut-off values of the different elastography
techniques used to evaluate the presence and severity of liver fibrosis depend upon
the etiology of the underlying liver disease, and upon the prevalence of the condition
under study in the target population. Although different cut-off values have been
proposed for cirrhosis according to etiology (ranging, for instance, from a Young’s
modulus of 11 kPa in CHB to 22.6 kPa in alcoholic liver disease), it should be kept
in mind that these cut-off values have been defined in a single population using ROC
curves in order to maximize sensitivity and specificity – and have not been applied
to a validation cohort. Differences between cut-offs may be simply related to differences
in cirrhosis prevalence and severity in the studied populations, known as the spectrum
bias. Therefore, elastography values should be interpreted by a liver specialist aware
of the clinical aspects of the liver disease to be assessed and aware of the peculiarities
of elastography in general and each elastography technique in particular.
Besides increasing evidence regarding liver stiffness measured by different techniques
to detect liver fibrosis and cirrhosis (described in detail in the following paragraphs),
other new applications of elastography are being tested in the field of liver diseases.
They include spleen stiffness assessment for portal hypertension and the evaluation
of the stiffness of focal liver lesions to differentiate between benign and malignant
nodules. These applications appear promising, but remain under development and cannot
yet be recommended in clinical practice.
Clinical needs
The ranges for intermediate fibrosis stages (F2-F3) are quite narrow, in the order
of a Young’s modulus of 2 – 3 kPa (over a total range spanning 2 to 75 kPa with the
TE), so that small differences in outputs could shift the assessment of patients from
one stage to another [141 ]. However, in “the real life” situation, attention should be focused on the patient
and what is appropriate from a clinical point of view. Following the availability
of novel antiviral agents, the European Association for the Study of the Liver (EASL)
together with the Asociación Latinoamericana para el Estudio del Hígado (ALEH) have
produced guidelines for the clinical use of noninvasive tests for the evaluation of
liver disease severity and prognosis [76 ]. These guidelines have outlined that the two clinically relevant endpoints in patients
with viral hepatitis are the detection of significant fibrosis (F≥ 2) and the detection
of cirrhosis (F = 4), and the most important endpoint is the detection of cirrhosis,
because it guides treatment (A1 recommendation). In patients with NAFLD and with chronic
liver diseases of other etiologies, the detection of cirrhosis is also the most important
clinical endpoint (A1 recommendation).
Clinical decision-making before elastography
Elastography is designed as a noninvasive tool to identify and stage liver fibrosis,
and it should be used in an appropriate manner. Specifically, it should be interpreted
along with a diagnostic workup aimed at identifying the cause of liver disease. This
is important since the interpretation of liver elastography values depends on the
specific clinical scenario, the prevalence of disease in the population under investigation,
the current patient’s comorbidities and the etiology of the liver disease.
Liver stiffness evaluation is not part of the screening process and initial evaluation.
The main reason for this is that a rational stepwise approach to the diagnosis and
staging of liver disease should follow the same rationale used for any other disease
in order to optimize the use of resources, limit costs and minimize the risk of iatrogenic
injury. The backbone of any workup is anamnesis, physical examination and laboratory
tests, followed by the use of noninvasive diagnostic methods, ideally from less to
more expensive, and following a balanced view of the cost and accuracy of each for
the specific clinical question. The use of invasive methods, which always imply a
certain risk for patients, should be limited when the previous phases did not provide
sufficient information. In addition, despite elastography methods being increasingly
used and available in reference centers, they are not yet universally available and
require experience. Therefore, elastography is part of a clinical decision-making
process.
Anamnesis, clinical examination and abdominal ultrasound examination are recommended
prior to elastographic liver evaluation. This strategy detects the presence of ascites,
tumoral masses in the liver, obstruction of the biliary tree, and signs of right heart
dysfunction. Direct liver elastographic evaluation reduces the time of examination,
decreases the cost and can be performed by trained technicians (TE).
Not recommending elastography as a first-line screening approach for liver disease
allows an appropriate diagnostic workup in patients without significant fibrosis at
elastography. These patients may have a liver disease prone to progression, despite
a normal elastography measurement.
Screening
According to the WHO principles for screening for disease [142 ], a screening test should offer high validity, reliability and acceptance in the
screened population, in addition to a positive balance between yield and costs and
available follow-up services in the case of a positive test. Thus, SWE may be well
suited as a screening method for asymptomatic liver disease in low prevalence populations.
However, there is insufficient evidence to evaluate the value of SWE in the general
population, in high-risk populations with intravenous drug use, diabetes mellitus,
obesity and/or alcohol abuse or in migrants from areas with a high prevalence of HCV
and HBV [61 ]
[143 ]
[144 ]
[145 ].
TE as a screening procedure in a general population > 45 years has been evaluated
in two large studies, with a total of 4700 apparently healthy subjects [61 ]
[143 ]. One study found that the presence of diabetes mellitus and obesity was associated
with a Young’s modulus above 8 kPa. However, neither of the studies systematically
investigated the causes of elevated liver stiffness in participants with a positive
result.
Initial evaluation of chronic liver disease (prerequisites for SWE)
Current complaints and past medical history
The initial evaluation of a patient with abnormal liver biochemical and/or function
tests includes knowledge (anamnesis) of current complaints, obtaining a past medical
history to identify potential risk factors for liver disease and performing a physical
examination to look for clues to the etiology and for signs of chronic liver disease.
Elastography should not be performed without the knowledge of current complaints,
past medical history and basic laboratory testing
Conventional ultrasound
Conventional ultrasound examination could be part of the initial evaluation [146 ], especially if the equipment is located within the clinic. Conventional B-mode ultrasound
can be used not only to detect the classic features of established cirrhosis but also
to determine the presence of biliary dilation indicating extrahepatic cholestasis
and of signs of cardiac failure or hepatic vein disease, which are conditions limiting
the accuracy of elastography. B-mode US screening for focal liver lesions, splenic
size, venous collaterals, ascites and other signs of portal hypertension can also
be performed. Ascites precludes the use of TE elastography. Knowledge of conventional
B-mode ultrasound findings is mandatory before strain and SWE elastographic examination,
or at least affects interpretation of the elastography results. Doppler ultrasound
is a basic tool for the optimization of B-mode ultrasound and the exclusion of vascular
liver disease and is helpful for determining complications of chronic liver disease,
e. g., portal hypertension.
In conclusion, in patients with liver disease of unclear etiology, elastography should
not be interpreted without a complete ultrasound assessment of the upper abdomen (B-mode
and Doppler ultrasound).
Chronic hepatitis C (CHC)
Fibrosis staging
Transient elastography
In patients with CHC, TE can differentiate absent and mild fibrosis from significant
fibrosis and cirrhosis, but is not accurate enough to distinguish between separate
stages of fibrosis (F1 – F4) [147 ]
[148 ]
[149 ]. A Young’s modulus greater than 6.8 – 7.6 kPa indicates a high probability of significant
fibrosis (F ≥ 2) on biopsy. However, optimal cut-off values vary considerably depending
on fibrosis prevalence and may range between 5.2 kPa – 9.5 kPa as indicated in the
clinical practice guidelines of the EASL [76 ]. Accordingly, the optimal cut-off values for predicting cirrhosis (F = 4) range
between 11 and 15 kPa [76 ]. Therefore, the local fibrosis prevalence and the diagnostic aim (sensitive screening
vs. secure exclusion strategies) must be considered when adapting cut-off values for
clinical use. It should be emphasized that TE gives the best diagnostic performance
in the context of cirrhosis diagnosis, and for this purpose it is better at ruling
out than at ruling in cirrhosis.
In CHC patients with HIV co-infection, TE can be used with similar diagnostic accuracy
as compared to HCV infection alone for fibrosis and cirrhosis detection [150 ]. TE can also be helpful in liver transplant recipients for the staging of recurrent
fibrosis and cirrhosis [151 ].
The use of TE for the diagnosis of cirrhosis and the estimation of fibrosis severity
in CHC has been endorsed in the recommendations for the management of viral hepatitis
by the EASL and the ALEH, ideally in combination with an alternative and unrelated
noninvasive approach such as laboratory tests/serum markers of fibrosis [152 ]. In the case of TE failure or inconclusive noninvasive test results, biopsy is still
recommended when fibrosis staging is relevant for clinical decisions, although a preliminary
attempt with an alternative SWE method could be considered.
TE can be used as the first-line assessment for the severity of liver fibrosis in
patients with chronic viral hepatitis C. It performs best with regard to the ruling
out of cirrhosis (LoE 1b, GoR A) [25 ]
[150 ]
[152 ]. Broad consensus (17/0/1, 94 %)
Point shear wave elastography (pSWE)
As with TE, VTQ® SWS quantification has been studied extensively in patients with CHC. Cut-offs of
1.21 – 1.34 m/s predict significant fibrosis (F ≥ 2) (AUROC 0.85 – 0.89), while VTQ® cut-offs between 1.55 and 2 m/s (AUROC 0.89 – 0.93) predict cirrhosis [104 ]
[153 ]. The diagnostic performance of VTQ® is comparable to TE [154 ] with high accuracy for predicting significant (F ≥ 2, AUROC 0.87) and severe fibrosis
(F ≥ 3, AUROC 0.91) as well as cirrhosis (AUROC 0.93) [155 ]. However, discordance (> one fibrosis stage) between VTQ® and histology occurred in > 30 % in a study including 106 patients infected with
HCV. The discordance was associated with female gender and a high interquartile range
(IQR/M ≥ 30 %) of Young’s modulus [107 ]. Therefore, pSWE results require cautious interpretation. Evidence regarding ElastPQ® is limited [70 ]
[156 ]. The results in the pilot study [70 ] for fibrosis staging are similar to those reported for VTQ® , but more data are needed. Fibrosis biomarkers may help to clarify indistinct cases
[157 ].
pSWE as demonstrated with VTQ® can be used as the first-line assessment for the severity of liver fibrosis in patients
with chronic viral hepatitis C. It performs best with regard to the ruling out of
cirrhosis (LoE 2a, GoR B) [155 ]. Broad consensus (17/0/1, 94 %)
2 D shear wave elastography (2D-SWE)
2D-SWE using SSI was reported in several studies in patients with CHC [139 ]
[158 ]. Diagnostic accuracy was high for the detection of significant and advanced fibrosis
and cirrhosis (AUC> 0.90). In these studies the diagnostic performance for 2D-SWE
was better than for TE [139 ]
[159 ] and serum fibrosis markers (FIB-4 index, APRI and Forns' index) [159 ]. In one study including 102 obese CHC patients, 2D-SWE had excellent diagnostic
accuracy for the detection of severe fibrosis and cirrhosis (AUROC > 0.90 for both)
[91 ].
2D-SWE as demonstrated with SSI can be used as a first-line assessment for the severity
of liver fibrosis in patients with chronic viral hepatitis C. It performs best with
regard to the ruling out of cirrhosis (LoE 1b, GoR A) [139 ]
[158 ]
[159 ]. Broad consensus (17/0/1, 94 %)
Strain Elastography
Among the manufacturers, only Hitachi SE has been evaluated for the assessment of
fibrosis in liver disease and in viral hepatitis. Different scores for semi-quantitative
interpretation of SE in the liver have been proposed [1 ]. While the method produced promising results for the prediction of liver fibrosis
in Asian patients, its performance was inferior to TE in European patients [121 ]
[125 ]. New data on this technique are lacking.
Prediction of hepatic complications
Growing evidence supports the use of TE for risk stratification and the prediction
of clinical endpoints. Liver stiffness predicts 5-year mortality with better accuracy
than histological fibrosis staging (METAVIR) for HCV mono-infected (TE > 9.5 kPa)
[160 ] and HIV co-infected patients (TE > 9.0 kPa) [161 ]. Furthermore, an elevated Young’s modulus indicates an increased risk of HCC development
(TE > 10 kPa) [162 ], hepatic decompensation and variceal bleeding [163 ]. Especially for patients with an established diagnosis of cirrhosis, TE can be used
for grouping patients into different risk classes [162 ]. The combination of liver stiffness (Young’s modulus of > 14 kPa) with platelet
count (< 141 × 103 /μL) and response to antiviral treatment may increase the predictive value of TE for
HCC development [164 ]. In patients with compensated liver cirrhosis, liver stiffness helps to identify
patients with portal hypertension (see section recommendations on portal hypertension).
Data regarding firm endpoint prediction using pSWE and 2D-SWE are lacking and no evidence-based
recommendation can be specified. Two groups evaluated the predictive value of TE in
transplant patients with a recurrence of HCV infection [165 ]
[166 ]. A Young’s modulus > 8.7 kPa at 12 months after liver transplantation was associated
with a significantly reduced five-year graft and cumulative patient survival [165 ].
Role of elastography in the setting of anti-HCV treatments
Role of SWE to identify patients to be treated
In the absence of universal access to direct-acting antiviral agents (DAA) as a consequence
of high cost, different countries have implemented strategies to prioritize patients
for treatment. TE is used as the first-line investigation for the prioritization of
HCV patients for DAA (e. g. Young’s modulus values ≥ 7.1 kPa are considered equivalent
to fibrosis ≥F2, and values ≥ 9.5 kPa are considered equivalent to fibrosis ≥F3).
In countries using interferon-based strategies, TE can help to identify patients with
cirrhosis who have a lower likelihood of achieving a sustained virological response
(SVR: undetectable HCV-RNA, 24 weeks after completion of antiviral therapy).
Role of SWE during treatment (monitoring)
Limited evidence in the setting of interferon-based therapy postulated that increasing
liver stiffness as measured by TE during the treatment period might indicate a reduced
possibility of achieving a sustained virological response [167 ]
[168 ]. Data in patients undergoing IFN-free antiviral therapies suggest that liver stiffness
rapidly declines during treatment, even in patients with advanced fibrosis and cirrhosis.
This decline appears to reflect the reduction in liver inflammation, restoration of
liver function and the decrease in portal pressure, like an effect of HCV eradication
[169 ]
[170 ].
Role of SWE after treatment (monitoring in follow-up)
Data regarding the usefulness of liver stiffness monitoring during antiviral therapy
are scant. In the largest prospective study published (n = 91), a significant liver
stiffness decrease was observed during therapy with peg-interferon and ribavirin;
the decrease in liver stiffness continued after treatment only in patients who achieved
SVR [171 ]. In the era of DAAs, it is important to remark that after successful HCV eradication,
the use of pre-treatment cut-off values can impair the accuracy of TE [172 ]
[173 ], and it might lead to erroneous conclusions if the SVR status is not carefully taken
into account [170 ]. Several studies have evaluated the use of VTQ® for monitoring IFN-based antiviral therapy in HCV patients: SWE decrease or increase
reflects response or no response to treatment, respectively [174 ]
[175 ]
[176 ]. For IFN-free antiviral therapies, no data are available.
The monitoring of cirrhotic patients after SVR will become the new standard in the
era of DAAs. Although it is tempting to use SWE in this setting to observe the dynamics
of liver stiffness over time, no recommendation can be made at this stage on cut-offs
and the time interval to identify cirrhosis regression.
SWE is not recommended to monitor fibrosis changes during anti-HCV treatment (LoE
3, GOR D) [172 ]
[173 ]. Strong consensus (18/0/0, 100 %)
LSM changes after successful anti-HCV treatment should not affect the management strategy
(e. g. surveillance for HCC occurrence in patients at risk) (LoE 3, GOR D) [76 ]. Broad consensus (16/0/1, 94 %)
Chronic Hepatitis B (CHB)
Introduction
A large amount of evidence regarding elastography in CHB is available. There are > 50
published studies. The majority of them use TE but also validate the elastographic
methods of pSWE and 2D-SWE.
Fibrosis staging
The most important goal of noninvasive diagnostic tools is diagnosis of compensated
cirrhosis that would benefit from treatment regardless of the transaminase level [177 ].
Transient elastography
Transient elastography is the most validated elastographic method for staging CHB
and has similar accuracy in this clinical scenario compared to CHC [76 ]. Three meta-analyses confirmed the good performance of TE in CHB staging [178 ]
[179 ]
[180 ]. Despite LSMs showing a substantial overlap among adjacent stages of fibrosis (particularly
at lower fibrosis stages), LSM may effectively identify patients with ≥F2 and F4.
Recent publications confirmed previous evidence, suggesting that the AUROCs for ≥F2
vary between 0.80 and 0.90 [181 ]
[182 ] with Young’s modulus cut-off values between 6.6 kPa and 8.8 kPa [182 ]
[183 ]. Regarding the identification of cirrhosis (F4), recent data confirm previous evidence,
with AUROCs ranging between 0.81 and 0.97 [182 ] and cut-off values between 9.4 and 13.4 kPa [184 ]
[185 ]. A recent meta-analysis suggested that a value > 11.7 kPa should raise suspicion
of cirrhosis [178 ]. It has been suggested that LSM cut-offs should be adapted to transaminase levels
[185 ] since transaminase levels tend to influence LSM in CHB, and hepatitis flares are
often observed in CHB. However, recent studies showed that ALT-adapted cut-offs do
not influence TE diagnostic performance [186 ] and that the only variable associated with overestimation of F4 stage in CHB is
moderate/severe necro-inflammatory activity without any direct correlation with transaminase
level [187 ]. Interestingly, a Young’s modulus of < 5 kPa in patients with normal ALT and low
serum HBV DNA levels (< 2000 IU/ml) characterize inactive HBV carriers [188 ]
[189 ]. TE can be used to rule out significant fibrosis and cirrhosis in HBV inactive carriers,
which is the best indication for TE in HBV.
TE is useful in patients with CHB to identify those with cirrhosis. Concomitant assessment
of transaminases is required to exclude flare up (elevation > 5 times upper limit
of normal). (LoE 1b, GoR A) [178 ]
[179 ]
[180 ]. Broad consensus (17/1/0, 94 %)
TE is useful in inactive HBV carriers to rule out fibrosis (LoE 2, GOR B) [188 ]
[189 ]. Strong consensus (18/0/0, 100 %)
Point shear wave elastography (pSWE)
VTQ® has the advantage of a lower failure rate and has a similar diagnostic performance
as TE. The discriminative ability for staging fibrosis in CHB is good, with AUROCs
for ≥F2 and F4 of 0.76 – 0.91 [184 ]
[190 ] and 0.72 – 0.97 [155 ]
[184 ], respectively. These findings were confirmed in a meta-analysis that included patients
with several etiologies of liver disease. The analysis of data of patients with CHB
showed an AUROC for ≥F2 of 0.88 and the best cut-off was 1.35 m/s and the AUROC for
F4 was 0.93 and the best cut-off was 1.87 m/s [191 ]
[192 ]. There is limited data about other pSWE methods. ElastPQ® has been used for staging CHB patients in four studies, with good performance for
staging liver fibrosis. Further validation is required [106 ]
[192 ]
[193 ]
[194 ].
2 D shear wave elastography (2D-SWE)
Recently, 2D-SWE (SSI) was tested in patients with CHB and proved to have a lower
failure rate than TE and at least similar performance for fibrosis staging [195 ]. The AUROC for ≥F2 varies between 0.85 and 0.91 [58 ]
[196 ], with Young’s modulus cut-offs between 7.1 and 8.0 kPa [195 ]
[196 ], while the AUROC for F4 varies between 0.92 and 0.98 [195 ]
[197 ], with optimal cut-offs between 10.1 and 11.7 kPa [58 ]
[195 ]. The best indications of 2D-SWE in HBV are inactive carriers to rule out significant
fibrosis and cirrhosis diagnosis.
pSWE as demonstrated with VTQ® is useful in patients with CHB to identify those with cirrhosis (LoE 2a, GoR B) [191 ]. Strong consensus (18/0/0, 100 %)
2D-SWE as demonstrated with SSI is useful in patients with CHB to identify those with
cirrhosis (LoE 3a, GoR C) [196 ]
[197 ]. Broad consensus (17/0/1, 94 %)
Monitoring (evaluation of) response to treatment
Under nucleoside/nucleotide analogs, liver stiffness measured by TE significantly
decreases regardless of the baseline values of ALT [198 ]
[199 ]. The only factors associated with decline of LSM are higher baseline LSM and HBV
DNA levels [198 ]. The diagnostic accuracy and thresholds of liver stiffness using TE may differ in
untreated and treated patients with chronic hepatitis B and C. This aspect should
be taken into account when interpreting the results of elastography. Moreover, even
if liver stiffness declines with antiviral treatment, it is unknown if this reflects
disease regression and LSM changes should not affect management.
LSM changes under HBV treatment should not affect the management strategy (e. g. surveillance
for HCC occurrence in patients at risk) (LoE 2b, GOR B) [198 ]
[199 ]. Strong consensus (16/0/0, 100 %)
Prognostic relevance
Baseline LSM by TE has modest prognostic relevance with an AUROC between 0.70 and
0.73 for liver-related events [200 ]
[201 ], which may be increased by adding spleen diameter and platelet count as a prediction
model called LSPS (= LSM × spleen diameter/platelet count) up to a level of 0.83 [202 ]. TE is a good prognostic marker for HCC development that may occur without cirrhosis.
A Young’s modulus of > 8 kPa might be a value indicating the need to start screening
for HCC, even if a complete virological response was achieved [203 ]
[204 ]. The risk of HCC is even higher if the Young’s modulus by TE is > 12 – 13 kPa, which
also implies an increased risk of decompensation [181 ]
[203 ]
[204 ]
[205 ].
Non-alcoholic fatty liver disease (NAFLD)
In NAFLD patients, noninvasive markers should aim at the following: a) identify the
risk of NAFLD/NASH among individuals with metabolic syndrome; b) identify those with
a worse prognosis; c) monitor disease progression; d) predict response to treatment.
Fibrosis staging
Transient elastography
A particular problem in NAFLD patients is the failure rate of elastography. Different
studies report increased failure rates owing to increased body mass index (BMI > 30 kg/m²)
or waist circumference, which may interfere with the transmission of the push impulses
and the tracking ultrasound, thus preventing correct estimation of liver stiffness.
A second issue in NAFLD patients is that two probes are available (M and XL probes),
with the latter being dedicated to use in overweight subjects. Comparing the diagnostic
performance of the two probes (M and XL probes), the XL probe obtains more reliable
results than the M probe in obese patients. The main limiting factors for the XL probe
are a skin-to-liver capsule distance > 3.4 cm and extreme obesity (BMI > 40 kg/m²).
The third issue is that F3-F4 fibrosis is the most important prognostic factor for
liver-related outcomes and mortality in NAFLD patients.
TE performance is better for cirrhosis than for significant fibrosis [206 ]
[207 ]. TE has a higher rate of false-positive than false-negative results and a higher
negative predictive value (NPV) than positive predictive value (PPV). Therefore, the
ability to diagnose bridging fibrosis or cirrhosis is insufficient for clinical decision-making
[208 ]
[209 ]. A systematic review of TE in patients with NAFLD involved 9 studies and 1047 patients
[210 ]. TE was excellent in diagnosing F3 fibrosis (85 % sensitivity, 82 % specificity)
and cirrhosis (92 % sensitivity, 92 % specificity), but had only moderate accuracy
for F2 fibrosis (79 % sensitivity, 75 % specificity).
With the M probe, patients with steatosis > 66 % at liver biopsy had higher LSM values,
which led to higher false-positive LSM results [52 ]. Thus, in obese patients with a high degree of steatosis, TE (using the M probe)
may be less accurate in diagnosing severe fibrosis in NAFLD, and additional evaluation
may be warranted to avoid overestimation of fibrosis. However, additional studies
on the effects of steatosis on LSM measured with an XL probe are needed. The XL probe
produces lower stiffness values than the M probe. Different cut-offs should be used
[85 ]. With the M probe, at a Young’s modulus cut-off value of 7.9 kPa, the sensitivity,
specificity, PPV and NPV for F3 or greater disease are 91 %, 75 %, 52 %, and 97 %,
respectively [55 ].
In NAFLD patients, the best cut-off for F3 or greater disease is 7.2 kPa. With this
cut-off, the NPV to exclude F3 or greater disease is 89 % (95 % CI 84 – 95 %). Cut-off
values of 5.7 kPa and 9.3 kPa have 90 % sensitivity and specificity to rule out and
rule in F3 disease, respectively [55 ].
TE can be used to exclude cirrhosis in NAFLD patients (LoE 2a, GoR B) [76 ]
[210 ]. Broad consensus (13/0/3, 81 %)
Point shear wave elastography (pSWE)
A systematic review of 7 studies for a total of 723 patients who underwent SWS measurements
with VTQ® technique to evaluate the diagnostic efficacy of pSWE in patients with NAFLD was
recently published [211 ]. The summary sensitivity was 80.2 % for detecting significant fibrosis, which is
not an appropriate endpoint.
2 D shear wave elastography (2D-SWE)
There are only two studies that evaluate the performance of 2D-SWE (SSI) [57 ]
[212 ]. The results are too limited to make recommendations.
Comparison of different elastographic methods for NAFLD
One study recently compared TE (using the M probe), pSWE (VTQ® ) and 2D-SWE (SSI) in 291 patients with NAFLD enrolled in two different hospitals
[57 ]. All methods showed AUROCs ≥ 0.84 for severe fibrosis and cirrhosis and had a similar
performance for the diagnosis of these endpoints. The diagnostic performance of 2D-SWE
was superior to that of VTQ® for the diagnosis of significant fibrosis.
Follow-up of patients
Monitoring of the progression of fibrosis is also necessary in the follow-up of these
patients. Patients who achieved a ≥ 5 % weight loss at the 6-month follow-up showed
a decrease in LSM by TE, independent of the changes in aminotransferase levels [208 ]. No data are available for pSWE and 2D-SWE.
Prediction of liver-related complications
A recent study supports the use of TE for risk stratification and prediction of clinical
endpoints [213 ]. For pSWE and 2D-SWE data on this aspect are lacking and no evidence-based recommendation
can be given.
Alcoholic liver disease (ALD)
Fibrosis staging
Transient elastography in patients with prior or current chronic alcohol overuse can
distinguish absent and mild fibrosis (F0 – 1) from severe fibrosis and cirrhosis,
but similar to other etiologies, there is no evidence to suggest that it can differentiate
absent and mild fibrosis from significant fibrosis. [214 ] Additionally, in the eight published single etiology studies on TE for staging liver
fibrosis, there is no consensus regarding optimal Young’s modulus cut-off values for
significant fibrosis (≥F2), severe fibrosis (≥F3) or cirrhosis (= F4) [49 ]
[56 ]
[215 ]
[216 ]
[217 ]
[218 ]
[219 ]
[220 ]. The optimal cut-off values range from 7.8 [216 ] to 9.6 [56 ] kPa for significant fibrosis, from 8.0 [49 ] to 17.0 [218 ] kPa for severe fibrosis and from 12.5 [49 ] to 22.7 [215 ] kPa for cirrhosis. The considerable discrepancy between cut-off values in individual
studies is likely a consequence of differences in fibrosis stage prevalence with overrepresentation
of cirrhotic patients, place of recruitment, and whether patients with alcoholic hepatitis
or decompensated disease were excluded.
TE is more suited to rule out than rule in cirrhosis. At a Young’s modulus of 12.5
kPa, TE may rule out cirrhosis with a negative likelihood ratio of 0.07 if the disease
prevalence is 50 % or lower. [214 ].
TE can be used to exclude cirrhosis in patients with alcoholic liver disease, provided
that acute alcoholic hepatitis is not present (LoE 2b, GoR B) [56 ]
[214 ]
[219 ]
[220 ]. Strong consensus (15/0/0, 100 %)
Prognostication of alcoholic liver cirrhosis
There is scant evidence to suggest a role for ultrasound elastography for determining
the prognosis and for the monitoring of patients with alcoholic liver disease [221 ], or for TE to predict esophageal varices [222 ] and the hepatic venous pressure gradient in patients with alcoholic cirrhosis [221 ].
Timing of liver stiffness measurements with regard to alcohol abstinence
In patients undergoing alcohol detoxification, 0.5 to 4 weeks of abstinence causes
a clinically significant decrease in TE [49 ]
[50 ]
[51 ]
[223 ]. However, the decrease is associated with a normalization of transaminases, bilirubin,
alkaline phosphatase and/or gamma-glutamyltransferase. It is, therefore, unclear whether
alcohol alone or alcohol-induced hepatitis and cholangiocyte damage cause the increase
in liver stiffness. [220 ]. One study suggests that TE is accurate for staging in patients with ongoing alcohol
abuse but normal gamma-glutamyltransferase [56 ], while another study suggests that AUC for the diagnosis of cirrhosis in alcoholic
patients diminishes when AST is above 100 – 150 U/L [220 ].
Screening the general population or high-risk groups in primary care
Ultrasound elastography for systematic screening of high-risk populations in primary
care for alcoholic liver disease was performed in one study. TE was offered to primary
care patients with an AST:ALT ratio ≥ 0.8 [224 ]. However, This study did not include biopsy confirmation in patients with elevated
TE. In a diagnostic study, TE had excellent diagnostic accuracy for significant fibrosis
and cirrhosis in a subgroup of 71 patients recruited from primary alcohol rehabilitation
centers [56 ].
With a prevalence of cirrhosis of 2 – 4 % in a background population of at-risk individuals
[225 ], the PPV of TE should be considered low regardless of cut-off values.
Point shear wave elastography (pSWE) and 2D-SWE
There is only one study to support the use of 2D-SWE (SSI) for assessing alcoholic
liver fibrosis [56 ]. There are three small studies on the use of pSWE [226 ]
[227 ]
[228 ], two of which report diagnostic accuracies and test probabilities. The results are
consistent regarding diagnostic accuracy, which suggests that VTQ® may be used to rule out severe fibrosis and cirrhosis. The results regarding cut-off
values are, however, inconsistent. Therefore, there is insufficient evidence to make
recommendations for using VTQ® to distinguish absent and mild fibrosis (F0 – 1) from significant or severe fibrosis
and cirrhosis. Thus, there is still insufficient evidence to evaluate the role of
pSWE or 2D-SWE in alcoholic liver disease.
Cholestatic liver disease and autoimmune hepatitis (AIH)
Risk stratification is a major need in patients with chronic cholestatic diseases
in order to allow personalized management and the selection of candidates for clinical
trials of new drugs. Studies of liver stiffness as a surrogate of liver fibrosis and
prognosis focusing on cholestatic liver disease (primary biliary cholangitis-PBC,
primary sclerosing cholangitis-PSC) or AIH are scarce. Most available data focus on
TE [229 ]
[230 ]
[231 ].
Transient elastography
Transient elastography is currently considered one of the best surrogates of fibrosis
in PBC. High baseline or increasing values over time indicate a worse outcome in this
population [191 ]
[229 ]. Liver stiffness was investigated in 73 patients with PSC, regularly undergoing
clinical and elastographic follow-up [232 ]. LSMs were able to differentiate severe vs. non-severe fibrosis with a high discriminative
accuracy for cirrhosis (AUROC 0.88). There was high reproducibility between two operators.
Higher baseline LSM and an increase of LSM over time were associated with adverse
outcome such as death, liver transplantation, ascites, hepatic encephalopathy, gastrointestinal
bleeding, or HCC [232 ]. Dilatation of the intrahepatic biliary system due to a dominant stricture should
be excluded in PSC before interpreting the LSMs. Cholestasis increases liver stiffness
independent of liver fibrosis.
Data regarding the pediatric population with biliary atresia suggest that liver (and
spleen) elastography could be a valuable tool to predict outcomes before surgery,
and might be used after the Kasai operation to monitor liver disease and portal hypertension
[233 ].
Due to the limited evidence, no recommendation can be given.
Point shear wave elastography (pSWE)
VTQ® was initially performed in 9 patients with AIH, PBC and PSC having higher shear wave
velocities than healthy volunteers [234 ]. In total two studies dealt with VTQ® in AIH and primary biliary cholangitis. In 15 patients with treated AIH, VTQ® could differentiate between the absence of fibrosis and significant fibrosis [235 ]. SWS assessed by VTQ® showed good diagnostic accuracy for detecting cirrhosis (AUROC 0.91) in 61 patients
with primary biliary cholangitis [236 ].
2 D shear wave elastography (2D-SWE)
Data regarding 2D-SWE in AIH, PBC and PSC are not available. Liver stiffness assessment
helped in differentiating between biliary atresia and neonatal hepatitis in one study
[237 ].
Due to the paucity of data, no recommendation can be given.
Portal hypertension
Transient elastography
In patients with compensated advanced chronic liver disease/cirrhosis LSM correlates
with the hepatic venous pressure gradient (HVPG). Even though the correlation between
the two does not allow for an accurate estimate of the exact HVPG value (range: 0.59 – 0.70),
the discriminative ability of liver stiffness for the presence of clinically significant
portal hypertension (CSPH, defined as HVPG ≥ 10 mmHg, threshold for the appearance
of complications) is very high, with a summary AUROC of 0.93 in a recent meta-analysis
[238 ]. However, it should be emphasized that most of the patients included in the studies
concerning HVPG had viral or alcoholic cirrhosis, and evidence regarding other etiologies
is limited. In viral cirrhosis, Young’s modulus values of > 20 – 25 kPa are highly
specific for CSPH, and values of > 21 kPa predict the onset of a first clinical decompensation
with an accuracy similar to that of HVPG > 10 mmHg [239 ].
LSM with TE is useful to identify patients with a high likelihood of having clinically
significant portal hypertension (HVPG ≥ 10 mmHg) (LoE 2b, GoR B) [238 ]
[239 ]. Strong consensus (15/0/0, 100 %)
The accuracy of LSM in predicting the presence and size of gastroesophageal varices
has been the subject of several studies. Despite the fact that it is currently the
best single noninvasive predictor in this field with summary AUROCs of 0.84 for esophageal
varices (EV) and 0.78 for large EV in a recent meta-analysis [238 ], the cut-offs vary widely among the studies, and the accuracy is not sufficient
to replace endoscopy. The accuracy of LSM for the diagnosis of CSPH and varices improves
if it is combined with platelet count and spleen size [240 ]
[241 ]. Recent data indicates that if a combination of a Young’s modulus value of < 20
kPa and a platelet count of > 150 G/L is used, varices needing treatment can be ruled
out with a high accuracy (< 5 % of patients missed) [242 ]
[243 ], and endoscopy can be safely avoided [244 ].
Point shear wave elastography (pSWE)
Point SWE (VTQ® ) has been used in three studies addressing the diagnosis of CSPH [245 ]
[246 ]
[247 ], and showed excellent applicability and very good diagnostic accuracy (AUROC 0.82 – 0.90).
VTQ® has been used in a few studies addressing the diagnosis and severity of esophageal
varices. SWS was higher in patients with esophageal varices of any size, and was even
higher in patients with large varices [245 ]
[246 ]. However, reliable cut-offs are not available yet. No strong recommendation regarding
the cut-offs to be used can be made due to the limited evidence.
2 D shear wave elastography (2D-SWE)
2D-SWE (SSI) has been tested for the diagnosis of CSPH in 4 studies and a further
small series [27 ]
[248 ]
[249 ]
[250 ]
[251 ]. The accuracy of the method was reliable in all of the published studies (AUROC
0.80 – 0.92).
Two studies performed a head-to-head comparison between LSM by TE and 2D-SWE [27 ]
[248 ]. TE was less applicable, and both techniques showed similar accuracy for the diagnosis
of CSPH.
LSM by 2D-SWE is higher in patients with esophageal varices of any size and is further
increased in patients with large varices. However, reliable cut-offs are not available
yet. No strong recommendation regarding the cut-offs for 2D-SWE can be given, and
further evidence is needed.
Liver stiffness using TE combined with platelet count is useful to rule out varices
requiring treatment (LoE 2b, GoR B) [244 ]. Although preliminary results are encouraging, there is insufficient evidence to
recommend pSWE and 2D-SWE in this setting. Broad consensus (13/0/1, 93 %)
Reimbursement
Social health care systems and noninvasive liver fibrosis
The data on the cost-effectiveness of liver elastography are limited to TE, are mainly
derived from UK and Canadian health care systems and depend on local health care economic
conditions. Although TE is generally accepted as a noninvasive alternative for the
staging of liver fibrosis, which helps to avoid a liver biopsy [252 ], it does not represent the most effective approach with respect to resources in
common clinical scenarios [253 ]. TE is considered cost-effective for annual cirrhosis screening in patients infected
with HCV [254 ], with the most cost-effective treatment scenario for patients infected with HCV
not depending on the fibrosis category (“treating all”) [253 ]. However, in countries that cannot afford the “treating all” scenario, TE is the
accepted method to identify patients deserving access to treatment based on the greater
severity of fibrosis, making it competitive in cost-effectiveness in comparison to
biopsy to screen large populations of patients. Inaccurate noninvasive fibrosis staging
may limit the cost-saving value of elastography [255 ]. Weighing the diagnostic limitations and costs of elastography against laboratory-based
fibrosis scores, widely available and cheap scoring systems (e. g. AST-to-platelet
ratio index) are the recommended method in regions with limited health care resources
[256 ]. Considering the emerging prevalence of non-alcoholic fatty liver disease (NAFLD)
in industrialized countries, TE can represent a cost-effective approach for the screening
of patients at risk for disease progression and complications [257 ]
[258 ]. However, a formal study assessing hard outcomes such as survival, hospitalization
rates, etc., has not been performed recently and is unlikely to be performed given
the vast number of patients required.
Current reimbursement in European countries
Reimbursement is often limited to conventional ultrasound, which includes shear wave-based
liver stiffness assessment. For example, no reimbursement is available within the
compulsory health insurance for TE (Fibroscan® ) in Germany. In private practice, 50 – 150 € is usually charged for a Fibroscan® examination.
In many other European countries, no specific reimbursement code is established. Where
ultrasound elastography is reimbursed (e. g. some regions in Italy), it is accepted
under the code of liver ultrasonography. However, the same exam cannot be reimbursed
twice on the same day. Therefore, conventional ultrasonography and ultrasound elastography
of the liver cannot be reimbursed together at the same appointment, thus limiting
the cost-effectiveness of the technique. In Romania, TE is reimbursed in the daily
hospitalization system with the possibility for future reimbursement of pSWE and 2D-SWE.
In private practice, the cost of TE is between 50 and 120 €.
Outlook: what should be done?
One future possibility in liver elastography is to compare the accuracy of different
technologies (TE, pSWE and 2D-SWE) in a large cohort of patients with division of
the etiologies of liver diseases. Probably, as indicated by earlier comparative studies,
we will have some results showing the superiority of new technologies in comparison
with established ones (point or 2D-SWE vs TE). Robust cut-off values must be established
for each of the different systems and diseases. The combination of elastographic methods
using ultrasound with biological tests may improve the accuracy of the evaluation
of the patients.
Christoph F. Dietrich, Jeffrey Bamber, Annalisa Berzigotti et al. EFSUMB Guidelines
and Recommendations on the Clinical Use of Liver Ultrasound Elastography, Update 2017
(Long Version). Ultraschall in Med 2017; doi:10.1055/s-0043-103952
This is to clarify that in the published EFSUMB Guidelines and Recommendations on
the Clinical Use of Liver Ultrasound Elastography, Update 2017 (Long Version) in Figure
1 (“The types of elastography described in this paper”) the availability of Easaote
SpA point shear wave technology (pSWE) was not mentioned. Herewith we present the
corrected figure.