Key words
breast radiography - strain elastography - BI-RADS - breast cancer - strain ratio
Introduction
The most prevalent cancer in the world is breast cancer (4.4 million survivors up to 5 years after diagnosis) [1]. According to the last global data, about 400 000 women died from breast cancer, representing 1.6 percent of all female deaths [2]
[3]. Fortunately, breast cancer mortality is decreasing in many high-risk countries due to intensified early detection, resulting in the diagnosis of smaller, earlier stage tumors, and advances in treatment [4].
In earlier US studies, US diagnosis was made according to differences in the size, shape, and boundary of lesions [5]. In 2003, the standardized BI-RADS-US classification system was proposed by the American College of Radiology (ACR) [6]. Nevertheless, in our clinical observation, BI-RADS-US 3 and BI-RADS-US 4 lesions, especially BI-RADS-US 4A lesions, remain a critical category. In a prospective study by Barr RG, BI-RADS category 3 lesions have a low malignancy rate (0.8 %) and only 0.1 % of the cancers had suspicious changes at 6-month follow-up [7]. In a retrospective study, 3 % of BI-RADS-US 3 lesions were diagnosed as malignant, leading to a delayed diagnosis of cancer in many patients [8], while a considerable percentage of BI-RADS-US 4A lesions was confirmed to be benign, resulting in excessive biopsies. Therefore, a suitable diagnostic predictor of malignancy should be explored in BI-RADS-US 3 and BI-RADS-US 4A lesions.
In the last decade, strain elastography (SE) was found to be an efficient examination method for the diagnosis of liver lesions, thyroid cancer and breast cancer [9]
[10]. In 2013, the revised BI-RADS-US classification system indicated stiffness, as a feature of malignant masses, should be integrated into breast tumor diagnosis [11]. After Waki et al. presented a new diagnosis system using the strain ratio measurement [12], the stiffness of breast lesions could be semi-quantitatively calculated with the same depth of breast tissue as the reference and the subjective bias could be avoided [13].
Therefore, with a focus on BI-RADS-US 3 and BI-RADS-US 4 lesions, our retrospective study scrutinized whether the strain ratio system can filter out malignancy in BI-RADS-US 4 lesions and identify malignancy in BI-RADS-US 3 lesions. The aim of this study was to assess whether the strain ratio could be used as a further characterization of lesions that are initially categorized as BI-RADS-US 3 or BI-RADS-US 4 and whether the strain ratio should be integrated into the BI-RADS-US classification system for precise prediction of malignancy.
Methods
General design
This study was carried out at the Sun Yat-sen Memorial Hospital. Data was collected from April 2010 to September 2015. All patients were examined with B-mode sonography and strain elastography. As we were using a standard of care clinical protocol, the responsible ethics committee did not require additional approval for the non-interventional design of our retrospective study. We obtained consent from each study patient prior to data collection.
Patients
Patients with a sonographically visible lesion classified as BI-RADS-US 2 – 5 were regarded as being suitable for our study. Any patients with radiotherapy and chemotherapy of breast cancer, skin disorders, inflammatory conditions of the breast, a distance between the front boundary of the lesion and the skin of > 4 cm were excluded. A total of 1301 lesions in 1080 patients were tested. Of these, 1071 lesions in 976 patients were included in the final analysis ([Fig. 1]). Histological confirmation was performed in all cases.
Fig. 1 Flowchart of participants on the basis of B-mode US evaluation.
B-mode sonography and strain elastography examinations
Both the B-mode sonography and strain elastography examinations were performed by 3 radiologists with more than 5 years of experience in US and SE. Two ultrasound scanners were used: Hitachi EUB-900 and Hitachi Hi-Vision PREIRUS (Hitachi Medical Corporation, Inc., Tokyo, Japan) equipped with the EUP L74 M linear transducer (5 – 13 MHz, 5.0 cm) or the EUP-L65 linear probe (6 – 14 MHz, 3.8 cm).
Firstly, all patients underwent bilateral whole-breast ultrasound examination. The BIRADS system [11] (BI-RADS category 2: benign; BI-RADS category 3: probably benign; BI-RADS category 4a: low suspicion; BI-RADS category 4b: intermediate suspicion; BI-RADS category 4c: moderate suspicion; BI-RADS category 5: highly suggestive of malignancy) was used to obtain a summative evaluation of the US features of each lesion.
Secondly, on real-time SE, the elastogram box mostly extended from the subcutaneous fat layer to the greater pectoral muscle to scan sufficient breast tissue. To obtain an optimal elasticity image, the process was repeated when the pressure indicator bar displayed an index of 3 – 4 until a stable image was acquired. The SE image was illustrated on a color map, with red, blue and green indicating soft, intermediate and hard elasticity, respectively. The strain ratio measurement was performed with a same-depth area as an internal reference [13].The region of interest (ROI) including the lesion was expressed as A, and another ROI including the same-depth area was expressed as B. The strain ratio was automatically calculated as a B/A ratio.
Further diagnostic steps (mammography, core needle biopsy, open biopsy or follow-up examinations) were conducted. The histology of all of the lesions was established with an ultrasound-guided, 14-gauge, automated gun core biopsy, or an open biopsy. Only biopsy results with a definitive diagnosis were accepted.
Statistical Analysis
Microsoft Office Excel 2003 (Microsoft Corporation) was used for data collection. Statistical analysis was performed using SPSS version 13 software (SSPS Inc, Chicago, IL) and MedCalc (MedCalc Software, Mariakerke, Belgium).
The best cutoff point of strain ratio was obtained by calculating the Youden’s index. Receiver operating characteristic curves (ROCs) were used to evaluate the diagnostic efficiency. The z-test was then used to compare the area under the curve (AUC). The accuracy, sensitivity, and specificity were compared by McNemar’s test. The Mann-Whitney U test was used to compare the medians of two independent groups for continuous data. The Fisher’s exact test was used to compare independent groups for categorical variables.
With a focus on BI-RADS-US 3 and 4 lesions, we calculated the pretest probability of disease (POD) as well as the posttest POD in the high-risk group and in the low-risk group. Pairwise comparisons were performed using the z-test. Statistical significance was assumed at P < 0.05 for all tests.
Results
1071 lesions in 976 patients were included in the final analysis ([Fig. 1]). The patients’ age, lesion size and strain ratio were compared between benign and malignant subgroups ([Table 1]). [Table 2] shows the histological diagnosis of the lesions.
Table 1
Participant characteristics of 976 patients with 1071 breast lesions.
characteristic
|
malignant
n = 424 (39.6 %)
|
benign
n = 647 (60.4 %)
|
p-value
|
all lesions
n = 1071
|
age (y)
|
median[1]
|
49 (14 – 88)
|
41 (17 – 81)
|
< 0.001
|
46.5 (14 – 88)
|
mean[2]
|
50.7 ± 12.0
|
41.6 ± 10.3
|
< 0.001
|
45.8 ± 14.7
|
lesion diameter (mm)
|
median1
|
12.0 (3.2 – 55)
|
9.1 (4.6 – 47.5)
|
< 0.001
|
11.6 (3.2 – 55)
|
mean2
|
13.5 ± 9.0
|
11.2 ± 7.7
|
< 0.001
|
12.9 ± 8.4
|
strain ratio
|
median1
|
5.41 (0.77 – 56.70)
|
1.63 (0.34 – 8.12)
|
< 0.001
|
2.87 (0.34 – 56.70)
|
mean2
|
7.97 ± 4.53
|
1.97 ± 1.24
|
< 0.001
|
4.53 ± 3.92
|
1 Data in parentheses represents the range.
2 Data presented as mean ± standard deviation.
Table 2
Histological diagnosis of 976 patients with 1071 breast lesions.
benign lesions (n = 647)
|
n
|
malignant lesions (n = 424)
|
n
|
fibroadenoma
|
406
|
invasive ductal carcinoma
|
329
|
fibroadenomatous hyperplasia
|
72
|
ductal carcinoma in situ
|
34
|
papilloma
|
62
|
invasive lobular carcinoma
|
11
|
adenopathy
|
34
|
mucinous carcinoma
|
9
|
phyllodes tumor
|
29
|
invasive papillocarcinoma
|
7
|
chronic inflammation
|
17
|
neuroendocrine carcinoma
|
7
|
atypical ductal hyperplasia
|
5
|
intraductal papillary carcinoma
|
6
|
tubular adenoma
|
5
|
invasive adenocarcinoma
|
5
|
juvenile hyaline fibromatosis
|
4
|
medullary carcinoma
|
3
|
lipoma
|
3
|
lymphoma
|
2
|
other benign lesions
|
10
|
other malignant lesions
|
11
|
The diagnostic performance of BI-RADS-US was calculated: the overall sensitivity of B-mode US was 99.0 % (420 of 424), with a 95 % confidence interval (CI) of 95.1 % to 99.3 %; the specificity was 53.3 % (345 of 647), with a 95 % CI of 54.2 % to 60.9 %; and the accuracy was 71.4 % (765 of 1071), with a 95 % CI of 68.6 % to 74.8 %. The area under the receiver operating characteristic curve (AUC) was 0.985 (95 % CI: 0.970, 0.991) for BI-RADS assessment alone.
Strain ratio of breast lesions
The mean strain ratio of the 647 benign lesions was 1.97 ± 1.24, ranging from 0.34 to 8.12. The 1st quartile, median and 3 rd quartile strain ratios were 1.22, 1.63, 2.41, respectively. The mean strain ratio of the 424 malignant lesions was 7.97 ± 4.53, ranging from 0.77 to 56.70. The 1st quartile, median, and 3 rd quartile strain ratios were 3.64, 5.41, 9.24, respectively. The strain ratios for malignant lesions were higher than for benign ones (P < 0.001).
The ROC curve was used to assess the diagnosis performance of the strain ratio method. The AUC was 0.936, and the maximum Youden’s index was 0.76. As the best cutoff point is defined when the maximum Youden’s index value is attained, the best cutoff point in this study was 2.98. With this best cutoff point, the sensitivity, specificity, accuracy, PPV and NPV of the strain ratio method were 86.9 %, 86.6 %, 82.6 %, 78.8 %, and 92.0 %, respectively.
Analysis of the strain ratio distribution of each BI-RADS-US category lesion ([Table 3])
Table 3
Strain ratio distribution of each BI-RADS-US category lesion.
BIRADS
(n)
|
n
|
B
|
M
|
B
|
M
|
[1]B
|
[2]M
|
SR≥ 2.98
|
SR<2.98
|
SR≥ 2.98
|
SR<2.98
|
mean
|
median
|
mean
|
median
|
2 (46)
|
46
|
0
|
NA
|
46
|
NA
|
NA
|
1.20 ± 0.46
|
1.23
|
NA
|
NA
|
3 (303)
|
299
|
4
|
7
|
292
|
3
|
1
|
1.75 ± 0.79
|
1.63
|
[1]3.86 ± 1.02
|
3.68
|
4a (199)
|
182
|
17
|
27
|
155
|
10
|
7
|
2.05 ± 1.33
|
1.65
|
[2]5.13 ± 4.99
|
3.89
|
4b (177)
|
96
|
81
|
52
|
44
|
66
|
15
|
3.94 ± 2.41
|
3.21
|
[3]7.51 ± 7.14
|
4.36
|
4c (210)
|
22
|
188
|
17
|
5
|
169
|
19
|
4.57 ± 2.39
|
4.09
|
[4]7.21 ± 6.48
|
5.39
|
5 (136)
|
0
|
136
|
NA
|
NA
|
127
|
9
|
NA
|
NA
|
10.19 ± 8.97
|
6.76
|
1 B = benign.
2 M = malignant.
1 Mean SR of malignant lesions of BI-RADS-3 was higher than that of benign ones (P = 0.002).
2 Mean SR of malignant lesions of BI-RADS-4A was higher than that of benign ones (P< 0.001).
3 Mean SR of malignant lesions of BI-RADS-4B was higher than that of benign ones (P< 0.001).
4 Mean SR of malignant lesions of BI-RADS-4C was higher than that of benign ones (P< 0.001).
46 lesions were classified as BI-RADS-US 2. According to histology, all (100 %) of the 46 lesions were benign. The mean strain ratio of BI-RADS-US 2 lesions was 1.20 ± 0.46. The stain ratio of all BI-RADS-US 2 lesions were under 2.98.
303 lesions were classified as BI-RADS-US 3. According to histology, 98.7 % (299 of 303) were benign. The mean strain ratio was 1.75 ± 0.79. Among the 299 benign lesions, 7 lesions had a strain ratio higher than 2.98. While 1.3 % (4 of 303) of these lesions were malignant, the mean strain ratio was 3.86 ± 1.02. Among the 4 malignant lesions, 3 lesions had a strain ratio higher than 2.98.
586 lesions were classified as BI-RADS-US 4. According to histology, 51.2 % (300 of 586) of these lesions were benign. The mean strain ratio was 2.39 ± 1.22. Among the 300 benign lesions, 96 lesions had strain ratios higher than 2.98. While 48.8 % (286 of 586) of the lesions were malignant, the mean strain ratio was 7.06 ± 4.25. Among the 286 malignant lesions, 41 lesions had strain ratios lower than 2.98. Further analysis of these lesions is in the next part.
136 lesions were classified as BI-RADS-US 5. According to histology, all (100 %) of the 136 lesions were malignant. The mean strain ratio was 10.19 ± 8.97. Among the 136 malignant lesions, 9 lesions had a strain ratio lower than 2.98, including 7 IDC and 2 DCIS lesions.
Analysis of the strain ratio method to further evaluate BI-RADS-US 3 and BI-RADS-US 4 lesions
The 4 malignant BI-RADS-US 3 lesions included 3 IDC lesions (strain ratios: 3.18, 4.88, 4.59) and 1 DCIS lesion (strain ratio: 2.82). All 4 malignant lesions had higher strain ratios than the 1st quartile, median and the 3 rd quartile strain ratios of all benign lesions (1.29, 1.68, and 2.41, respectively). According to the ROC curve, the AUC of the strain ratio method in BI-RADS-US 3 lesions was 0.977 ([Fig. 2]).
Fig. 2 Receiver operating characteristic (ROC) curves for the strain ratio method in BI-RADS-US 3 lesions.
199 lesions were classified as BI-RADS-US 4A, 91.5 % (182 of 199) of which were benign. The mean strain ratio was 1.75. While 8.5 % (17 of 199) of these lesions were malignant, the mean strain ratio was 5.13. Among these malignant lesions, all 17 lesions had higher strain ratios than 1.29 and 1.63, the 1st quartile and the median strain ratios of all benign lesions, respectively. 16 had higher strain ratios than 2.41, the 3 rd quartile strain ratio of all benign lesions. According to the ROC curve, the AUC of the strain ratio method in BI-RADS-US 4A lesions was 0.929 ([Fig. 3]).
Fig. 3 Receiver operating characteristic (ROC) curves for the strain ratio method in BI-RADS-US 4A lesions.
177 lesions were classified as BI-RADS-US 4B, 54.2 % (96 of 177) of which were benign. The mean strain ratio was 3.94. While 45.8 % (81 of 177) of these lesions were malignant, the mean strain ratio was 7.14. According to the ROC curve, the AUC of the strain ratio method in BI-RADS-US 4B lesions was 0.740 ([Fig. 4]).
Fig. 4 Receiver operating characteristic (ROC) curves for the strain ratio method in BI-RADS-US 4B lesions.
210 lesions were classified as BI-RADS-US 4C, 10.4 % (22 of 210) of which were benign. The mean strain ratio was 4.57. While 89.6 % (188 of 210) of these lesions were malignant, the mean strain ratio was 7.72. According to the ROC curve, the AUC of the strain ratio method in BI-RADS-US 4C lesions was 0.710 ([Fig. 5]).
Fig. 5 Receiver operating characteristic (ROC) curves for the strain ratio method in BI-RADS-US 4C lesions.
Probability of disease in BI-RADS-US category 3 and 4A lesions
The pretest probability of disease in BI-RADS-US 3 lesions was 1.3 % and increased significantly to 29.8 % with an abnormal strain ratio (higher than 2.98). The posttest probability of disease with a normal strain ratio (lower than 2.98) decreased significantly to 0.3 %.
The pretest probability of disease in BI-RADS-US 4A lesions was 8.5 % and increased significantly to 48.7 % with an abnormal strain ratio (higher than 2.98). The posttest probability of disease with a normal strain ratio (lower than 2.98) decreased significantly to 4.5 % ([Table 4]).
Table 4
Probability of disease in BI-RADS-US 3 and BI-RADS-US 4A lesions after including strain ratio.
|
BI-RADS 3 lesions
|
BI-RADS 4A lesions
|
|
SR ≥ 2.98
|
SR < 2.98
|
SR ≥ 2.98
|
SR < 2.98
|
pretest POD (%)
|
1.3
|
8.5
|
posttest POD (%)
|
29.8[1]
|
0.3b[2]
|
48.7[3]
|
4.5[4]
|
likelihood ratio
|
32.6
|
0.3
|
8.9
|
0.44
|
Comparison of preteset and posttest POD and likelihood ratios; POD = probability of disease.
1 p = 0.033 (comparison of POD before test and POD after positive test).
2 p = 0.039 (comparison of POD before test and POD after negative test).
3 p = 0.003 (comparison of POD before test and POD after positive test).
4 p = 0.014 (comparison of POD before test and POD after negative test).
Effect of SE on sensitivity and specificity: BI-RADS-US 3 and 4A lesions
Inclusion of strain ratio for BI-RADS 3 lesions improved the overall sensitivity without a loss of specificity. Inclusion of strain ratio for BI-RADS 4A lesions improved the overall specificity without a loss of sensitivity. By adding SE information to all lesions, the overall specificity improved from 53.3 % for BI-RADS assessment alone to 76.4 % ([Table 5]). Two strategies were used for combining the BI-RADS category and strain ratio. The aggressive combination: the lesion was diagnosed as malignant when either SR or BI-RADS yielded a positive result. The conservative combination: the lesion was diagnosed as malignant when both SR and BI-RADS yielded a positive result. The lesion was diagnosed as benign by both strategies only if both methods yielded a negative result. The aggressive combination provided better diagnostic performance than the conservative combination, with higher sensitivity in BI-RADS category 3 and higher specificity in 4a lesions and all lesions.
Table 5
Estimates of effect of combining SR with BI-RADS 3 and 4a lesions.
|
category 3
|
category 4a
|
all lesions
|
SEN (%)
|
SPE (%)
|
SEN (%)
|
SPE (%)
|
SEN (%)
|
SPE (%)
|
BI-RADS
|
0
|
100
|
100
|
0
|
99.0
|
53.3
|
SR
|
75.0
|
97.6
|
85.2
|
58.8
|
86.9
|
86.6
|
[1]p
|
< 0.001
|
< 0.001
|
< 0.001
|
< 0.001
|
0.49
|
< 0.001
|
[2]BI-RADS+SR
|
75.0
|
97.6
|
100
|
85.2
|
96.9
|
76.4
|
[3]
p
|
< 0.001
|
0.56
|
0.81
|
< 0.001
|
0.74
|
< 0.001
|
[4]BI-RADS+SR
|
0
|
97.6
|
58.8
|
85.2
|
87.2
|
76.4
|
[5]
p
|
0.63
|
0.25
|
< 0.001
|
< 0.001
|
< 0.001
|
< 0.001
|
[6]p
|
< 0.001
|
NA
|
< 0.001
|
NA
|
< 0.001
|
NA
|
1 p Comparison of sensitivity/specificity of BI-RADS alone and SR alone
2 Aggressive combination of BI-RADS and SR: if either of the two methods yielded a positive result, the lesion was diagnosed as malignant. If results from both methods were negative, the lesion was diagnosed as benign.
3 p Comparison of sensitivity/specificity of BI-RADS alone and BI-RADS combined with SR aggressively.
4 Conservative combination of BI-RADS and SR: if the results of both methods were positive, the lesion was diagnosed as malignant. If both methods yielded negative results, the lesion was diagnosed as benign.
5 p Comparison of sensitivity/specificity of BI-RADS alone and BI-RADS combined with SR conservatively.
6 p Comparison of specificity of the aggressive combination and the conservative combination.
Discussion
Two main forms of elastography have become established in clinical practice. However, other important implementations are to be expected as the field matures [14]
[15]. The 2013 BI-RADS-US classification system stated that stiffness provides useful information that can be added to the diagnostic system. However, the combination of strain ratio and the BI-RADS-US classification system, especially with a focus on BI-RADS category 3 and category 4a, is relatively rare.
In clinical practice, SE examinations are usually performed after conventional ultrasound. The combination of these two methods is the most important point. Until now, there is minimal literature indicating any advantage from adding SE evaluation to lesions categorized BI-RADS-US 0, 1, 2 or 5. Our study is consistent with previous studies. However, improved diagnostic performance can be achieved for BI-RADS-US 3 and 4 lesions. Therefore, our study was focused on breast lesions with an unclear diagnosis and requiring more advanced assessment.
In our study, with a cutoff point of 2.98, the sensitivity and specificity of the strain ratio method were 86.9 % and 86.6 %, respectively. This is comparable to previous studies. The sensitivity ranged from 73.9 % to 92.4 % and the specificity from 86.4 % to 95.2 % [16]
[17]
[18]
[19]
[20]. Adding SR to BI-RADS feature analysis improved the specificity of breast US lesion assessment without a loss of sensitivity. Interestingly, this result is not only consistent with strain elastography but is also consistent with emerging shear wave elastography (SWE) technology [21].
In our study, the risk for malignancy in BI-RADS-US 3 lesions is 1.3 %, i. e., relatively low. With a suspicious strain ratio, the risk of malignancy increased significantly to a POD of 29.8 %. Usually BI-RADS-US 3 lesions only require follow-up. However, in light of the high posttest probability of malignancy, we encourage immediate biopsy of BI-RADS-US 3 lesions exhibiting a suspicious strain ratio ([Fig. 6]).
Fig. 6 UE image of a neuroendocrine carcinoma in a 34-year-old woman. The lesion was classified as BI-RADS-US 3. The strain ratio was 4.21.
BI-RADS-US 4 lesions cover a wide range of likelihood of malignancy (2~95 %). However, the majority of BI-RADS-US 4A lesions are in fact benign. Strain ratio allowed helpful categorization of lesions into a low-risk group (posttest POD 4.5 %) and a high-risk group (posttest POD 48.7 %). We suggest follow-up of BI-RADS-US 4A lesions with a normal strain ratio ([Fig. 7]) and immediate biopsy of BI-RADS-US 4A lesions with a suspicious strain ratio ([Fig. 8]). Biopsy in these cases could be expected to yield at least 1 carcinoma per 3 biopsies. Furthermore, in the BI-RADS-US 4A lesions in our study, all 17 malignant lesions had a higher strain ratio than 1.29 and 1.63, the 1st quartile and the median strain ratios of all benign lesions, respectively. If we choose the median strain ratio as the diagnostic criteria, the positive rate of biopsy will be improved because missed diagnoses will bring more serious consequences for patients.
Fig. 7 UE image of a fibroadenoma in a 46-year-old woman. The lesion was classified as BI-RADS-US 4A. The strain ratio was 2.11.
Fig. 8 UE image of an IDC in a 43-year-old woman. The lesion was classified as BI-RADS-US 4A. The strain ratio was 4.11.
However, we found that strain ratios did not provide beneficial information for BI-RADS-US 4B and 4C lesions. The 2013 BI-RADS-US classification system emphasized that elastography evaluation should not override the more predictive morphologic features of malignancy [11]. If a lesion was already highly suspicious on the mammogram and/or conventional ultrasound, the stiffness would not have an effect on the subsequent management of the patient. Therefore, on account of the high probability of malignancy, we suggest that a biopsy should be performed in every BI-RADS-US 4B or 4C lesion with a suspicious strain ratio.
Although the strain ratios of malignant lesions were significantly higher than those of benign ones, there is overlap in the elasticity coefficient for different tissues. Therefore, false-negative and false-positive results were unavoidable [22]
[23]. Most false-negative lesions were invasive ductal carcinomas (IDCs) and ductal carcinomas in situ (DCISs). In our study, the majority of malignant lesions were IDCs. Only 10 % (33 of 329) were misdiagnosed. DCIS has been shown to be softer than IDC ([Fig. 9]), which is consistent with Thomas et al. [24]. Meanwhile, most false-positive lesions were intraductal papillomas (38 of 144 false-positive results). This is similar to Ann et al. [25], who suggested that the mean elasticity score of intraductal papillomas was higher than that of fibrocystic changes ([Fig. 10]). The second most common false-positive results were fibroadenomas (27 of 144 false-positive results), 3 of which were complicated by calcification and 5 of which were lager than 20 mm.
Fig. 9 UE image of a papilloma in a 24-year-old woman. The lesion was classified as BI-RADS-US 3. The strain ratio was 7.62.
Fig. 10 UE image of a DCIS in a 39-year-old woman. The lesion was classified as BI-RADS-US 2. The strain ratio was 0.89.
In this study, the only malignant BIRADS-3 lesion misdiagnosed by SR with a strain ratio of 2.82 was a DCIS, while 5 of the 7 false-negative BIRADS 4a lesions misdiagnosed by SR were DCISs. 7 benign BIRADS-3 lesions had a strain ratio higher than 2.98, including 4 papillomas and 3 fibroadenomas (2 lesions lager than 30 mm, 1 lesion with calcification). 27 false-positive BIRADS 4a lesions contained 12 papillomas (3 lesions with calcification), 7 fibroadenomas (2 lesions lager than 30 mm, 4 lesions with calcification, 1 lesion with fibrous hyperplasia), 5 phyllodes tumors and 3 cases of chronic inflammation.
The main limitation of our study is the operator dependence. To a certain degree, this is inevitable in all studies of SE operation. Fortunately, all of the operators in our study were experienced. As a semi-quantitative method, strain ratio is more subjective and could avoid inter-operator variability. Another limitation is the lack of comparative studies. To prove our concept, a prospective, multi-center trial might be needed to fully evaluate the strain ratio method.
Conclusion
In our study, using the best cutoff point of 2.98, the stain ratio method successfully differentiated between benign and malignant breast masses. This semi-quantitative method not only improved the specificity of breast US lesion assessment without a loss of sensitivity but also yielded additional diagnostic information in the evaluation of BI-RADS-US 3 and 4 lesions. It should be integrated into the BI-RADS-US classification system and daily practice.