Legal basis
Application
According to the German Protection against Infection Act (IfSG, § 23 (3)), the heads of medical facilities must ensure that the necessary measures as determined by current medical knowledge are taken to prevent nosocomial infections and the transmission of pathogens. Moreover, according to § 36 of the IfSG, in-house infection prevention procedures must be defined in hygiene plans [9].
In the event of damage, e. g. the occurrence of an infection after an ultrasound intervention, the examiner must show that the necessary measures according to current medical science have been met and the valid recommendations have been observed [33]. Therefore, according to IfSG § 23 (4): “It is assumed that the requirements according to current medical science in this area have been met if the published recommendations of the Commission for Hospital Hygiene and Infection Prevention have been observed” [9].
In 2011, this commission (KRINKO) published the “Hygiene Requirements for Puncture and Injection” [13], which are discussed in the following.
The KRINKO recommendations [13] require skin antiseptic with a sufficiently long application time for percutaneous puncture. Microorganisms must not be able to be transferred from the clothing of the person performing the examination to the patient. Four risk groups are defined on the basis of the puncture procedure, the puncture target, and the assumed infection risk ([Table 1]). Individual risk factors must also be taken into consideration.
Table 1
Puncture risk groups according to KRINKO [13].
risk group
|
definition
|
1
|
simple puncture procedure and low risk of puncture-associated infection
|
2
|
simple puncture procedure and low infection risk but serious consequences documented in the literature in the case of (rare) occurrence of infection and no need to temporarily set down sterile puncture tools.
|
3
|
puncture of organs or cavities or complex puncture procedure with need to temporarily set down sterile puncture tools, with or without an assistant.
|
4
|
complex puncture with the need to temporarily set down sterile puncture tools and aseptic handling of puncture tools by an assistant and/or the insertion of catheters or foreign objects into body cavities or deep clefts in tissue.
|
Based on the categorization of the risk groups, KRINKO makes detailed recommendations regarding materials, barrier precautions, and protective clothing with examples ([Table 2]).
Table 2
KRINKO recommendations to be taken into consideration in a hygiene plan, modified according to [13].
risk group
|
type of puncture
|
type of swab
|
cover
|
additional protective clothing
|
examiner
|
assistance
|
1
|
i. v. injection (peripheral)
|
low-germ
|
none
|
low-germ gloves
|
not required
|
2
|
pleural puncture, ascites puncture, bladder puncture (diagnostic)
|
sterile
|
none
|
sterile gloves, surgical mask
|
no special measures
|
3
|
organ puncture (e. g. kidney, liver, lymph node, spleen, thyroid)
amniocentesis
chorionic villus sampling
transvaginal/transrectal (ultrasound-guided) cyst or tissue puncture
|
sterile
|
sterile drape
|
sterile gloves
|
no special measures
|
joint puncture
|
sterile
|
sterile drape
|
sterile gloves, surgical mask in the case of puncture with syringe change
|
surgical mask in the case of puncture with syringe change
|
4
|
pleural drainage catheter placement in other body cavities or organs
|
sterile
|
sterile drape
|
sterile gloves, surgical mask, surgical cap, sterile long-sleeve gown
|
surgical mask
|
The KRINKO recommendations also relate to ultrasound-guided puncture [13]
[15]. The guidelines do concede that there are no randomized studies with a clinical end point and only minimal scientific data in this regard is available [13]
[15]:
In the case of ultrasound-guided puncture, in which the probe touches the puncture site or can come into contact with the puncture needle, the probe is to be provided with a sterile cover ([Fig. 1], [2], [3]).
Fig. 1 Incorrect puncture using the freehand technique with minimal needle distance from the ultrasound probe. The needle unintentionally comes in contact with the ultrasound probe that has only been disinfected. A sterile cover for the ultrasound probe would be needed in this case.
Fig. 2 A safety distance between the ultrasound probe and the sterile needle is maintained in this freehand biopsy. If the examiner can ensure that the needle will not come into contact with the ultrasound probe, a sterile ultrasound probe cover is not necessary.
Fig. 3 Correct application of the sterile ultrasound probe cover in a freehand biopsy with possible contact between the needle and the ultrasound probe. The examiner is wearing sterile gloves. A sterile drape/incise drape as a barrier precaution is obligatory when tools (e. g. syringe) must be changed/set down during puncture.
In the case of ultrasound-guided puncture for inserting a catheter, the sterile cover must also include the supply cable ([Fig. 4]).
Fig. 4 Placement of an abscess drainage tube. Ultrasound probe, cable, and keyboard are covered with sterile covers. Following hand disinfection, the examiner is dressed as in a surgical situation. A sterile drape/incise drape serves as the barrier precaution. A sterile ultrasound gel is used as the contact medium. A sterile work table is additionally available.
If a non-sterile ultrasound coupling medium is used, it must not be allowed to contaminate the needle or the puncture region. This is ensured, for example, when the ultrasound probe is applied outside the puncture region ([Fig. 2]).
If an ultrasound coupling medium is need directly at the puncture site, an alcohol skin disinfectant or sterile ultrasound gel is to be used ([Fig. 1], [3]).
Hand hygiene as the most effective single step in breaking the chain of infection in health care facilities is particularly important in (interventional) ultrasound. Hygienic hand disinfection must be performed immediately before contact with a patient, immediately prior to aseptic activities, immediately after contact with potentially infectious materials, after contact with a patient, and after contact with the patient’s immediate surroundings and is thus an essential and continually repeated part of diagnostic ultrasound imaging. Surgical hand disinfection (washing hands and lower arms with soap, followed by disinfection up to the elbows) is required for intraoperative ultrasound [13]
[34].
Preparation
Preparation includes preliminary cleaning, cleaning, disinfection, rinsing, drying and inspection of the cleanliness and intactness of the ultrasound probe. In 2012, KRINKO and the Federal Institute for Drugs and Medical Devices (BfArM) published “Hygiene Requirements for the Preparation of Medical Devices” [14].
All medical devices must be cleaned and disinfected using validated methods and sterilized if necessary [19]. The requirements [14] necessitate detailed description of a validated method for the preparation of devices to be used in a low-germ – i. e. disinfected – and sterile state with reference to the Medical Device Directive and DIN EN ISO 17664. The legal situation in Germany requires ultrasound probe manufacturers to describe these methods. However, this information is often missing in practice [28]
[35].
With respect to the type of application and the resulting risk, there are three categories of medical devices: non-critical, semi-critical, and critical. Ultrasound probes that only come in contact with healthy skin are “non-critical”. Cleaning and disinfection after every use is sufficient here. Disinfectants with only bactericidal and yeasticidal (yeast killing) activity are to be used.
Ultrasound probes that come into contact with the mucous membrane or diseased skin, e. g. during introitus or perineal ultrasound, are “semi-critical medical devices without special preparation requirements”. In this case, disinfection with bactericidal (including mycobacteria), fungicidal and virucidal activity must be performed. The check for virucidal activity must be performed according to the current guidelines of the Society for Virology in Germany and the Robert Koch Institute [14].
Rigid endoscopic ultrasound probes ([Fig. 5]) are generally semi-critical medical devices of group A (without special preparation requirements). In Appendix 7 of the indicated publication, BfArM and RKI define the requirement as follows: “Manufacturers of ultrasound probes for transvaginal use are required to provide specifications regarding at least one effective and suitable disinfection method with the above-mentioned spectrum of efficacy in the instructions for use. Proof of efficacy with recognized methods must be documented by an expert opinion (...)” [14]. This requirement was included in the ultrasound agreement in 2017 [36].
Fig. 5 Rigid endoscopic ultrasound probe with puncture adapter. The ultrasound probe is first covered with a low-germ cover that is partially filled with ultrasound gel. The disposable needle guide is mounted over this.
Flexible endoscopic ultrasound probes and rigid probes with an internal instrument channel are part of group B (with greater preparation requirements) and must therefore undergo preliminary cleaning directly after use and should preferably be prepared for use by machine [14].
As thermolabile devices, guide adapters that can be clamped to the ultrasound probe ([Fig. 6]) are to be disinfected at least once and covered with a sterile cover. Injection needles and needle guides are generally to be used as sterile disposable products ([Fig. 5], [6]). Puncture needles and drainage tube are “critical” medical devices that must be sterile. Ultrasound probes with a puncture channel ([Fig. 7]) must also be sterilized [14].
Fig. 6 Ultrasound probe with puncture adapter. The guide adapter is attached to the ultrasound probe under the sterile cover. The sterile disposable needle guide is clamped on top. The ultrasound probe cover also covers the cable.
Fig. 7 Ultrasound probe with puncture channel with removed needle guide. As a rule, these ultrasound probes and needle guides must be sterilized since the needle in the channel touches both parts.
In intraoperative ultrasound ([Fig. 8]), ultrasound probes are inserted into open body cavities and are thus also considered critical medical devices. In addition, they are thermolabile (“critical C”). The KRINKO/BfArM recommendations require machine cleaning and disinfection as well as sterilization (e. g. plasma or ethylene oxide sterilization). Moreover, preparation is subject to external quality control. This is to be verified by certification of the quality management system to guarantee that these requirements have been met [14].
Fig. 8 Intraoperative ultrasound with intervention. The sterilized ultrasound probe or probe treated with bactericidal (including mycobacteria), fungicidal, and virucidal immersion disinfection is packed in sterile packaging. The examiner is dressed like a surgeon.
In English and French-speaking countries, there is often a differentiation between “low”, “intermediate” and “high-level disinfection”. “Low-level disinfection” refers to the killing of most vegetative bacteria, some fungi, and viruses in a short period of time (< 10 min). “High-level disinfection” means that all microorganisms including spores are killed [32]
[37]. This categorization is not legally relevant in Germany.
Current data
Percutaneous ultrasound
The contamination rates of examined ultrasound probes from the clinical routine were mentioned in the introduction. Transmission of Staphylococcus aureus from patient to patient via only dry-wiped, non-disinfected ultrasound probes was already proven 30 years ago [39]. However, reports of infections due to transmission of microorganisms via contaminated ultrasound probes are not known. This would also be difficult to prove since at the time of manifestation of an infection the ultrasound probe would typically already have been used again numerous times.
Bacterial contamination of ultrasound gel
As noninvasive products for temporary use, ultrasound gels are class I medical devices according to European law and as such are equal to pharmaceuticals with respect to function, safety, and quality. In the EU it can therefore be assumed that commercially available ultrasound gel is primarily not contaminated with bacteria.
However, contaminated gels were found in up to 2.5 % of cases (Pseudomonas aeruginosa, USA; Burkholderia cepacia, Italy, Canada; mycobacteria, Taiwan) and were determined to be a source of nosocomial infections [40]
[41]
[42]
[43]
[44]. Endemic transmission of Achromobacter xylosoxidans after transrectal prostate biopsy was able to be traced back to contaminated gel in Israel [45]. Heated gel in pediatrics was identified as the source of infection with Staphylococcus aureus [46]. However, heating the gel to over 40 °C tends to reduce the rate of bacterial contamination [47].
If gel is transferred from the disposable bottle, contamination and infection can occur. Eight cases of gel-associated infection with ESBL-producing Klebsiella pneumoniae were reported in France [48]. Contamination rates can be high: in two publications 28 % and 59.5 % of all gels were contaminated including with pathogenic bacteria (Stenotrophomonas maltophilia, Acinetobacter baumannii, Staphylococcus aureus, MRSA, Bacillus spp., among others) [47]
[49].
In a comment regarding bacterial contamination of ultrasound gel, North American authors cited case reports indicating contamination of the gel with Pseudomonas aeruginosa already during manufacture [50]. Despite the addition of preservatives, e. g. parabens, ultrasound gels are not always free of microorganisms [42]. In addition, ultrasound gel can enter the needle when it passes through the gel [51]. Various authors require the use of sterile ultrasound gel for every procedure requiring skin disinfection. This relates to all percutaneous biopsies [44]
[50].
-
Commercially available ultrasound gels are usually but not always free of microorganisms.
-
Refillable gel containers are often contaminated.
-
There are case histories and reports of local outbreaks of various infections caused by the transmission of microorganisms via contaminated ultrasound gels.
-
Infections caused by transmission of microorganisms via contaminated ultrasound probes are not known despite high contamination rates.
Ultrasound probe preparation
According to the international literature, a broad range of cleaning and disinfection methods are used for ultrasound probes.
Cleaning only with dry paper is common but is not sufficient from a hygienic standpoint: Of 50 ultrasound probe swabs examined in an Iranian study, 98 % were positive for aerobic bacteria and 52 % for anaerobic bacteria immediately after an examination. After wiping with paper towels, the rates were 42 % and 21 %, respectively [52]. A Nigerian study found positive swabs in 37 or 40 patients immediately after examination, including 34.6 % cases of MRSA. An average of 90 colony-forming units (CFUs) were detected immediately after examination, 12.9 CFUs after the probe was wiped once, and 3.3 CFUs after the probe was wiped twice [53]. This coincides with findings from the USA [25] and Wales [26].
A study from Wales prospectively examined ultrasound probes immediately after abdominal ultrasound in 40 patients (including 34 with MRSA colonization). 40 swabs were taken immediately and then another 40 after the probes were wiped with paper towels. Two groups were formed: 20 swabs were performed after disinfection with 70 % alcohol and 20 after another wiping only with dry paper. On average, there were 128 CFUs immediately after use, 21 CFUs after wiping with paper once, 2 CFUs after wiping with paper twice, and only 0.05 CFUs after cleaning with paper and alcohol. 41 % of the first 34 swabs taken from ultrasound probes after use on MRSA-positive patients were positive for MRSA [26].
In a Canadian study, 31 non-endocavitary ultrasound probes were contaminated with various quantities of MRSA and were then cleaned with 0.5 % hydrogen peroxide solution. After five minutes, contamination could no longer be detected on any of the probes [54]. An English prospective study examined 120 swabs of ultrasound probes and ultrasound units for regional anesthesia after cleaning and disinfection with 70 % isopropanol. No bacteria growth was seen in 85 % of cases and growth of apathogenic microorganisms of the skin was seen in 15 % of cases [55].
French authors compare three different methods for cleaning/disinfection: Dry wiping with paper towels, antiseptic cleaning with a cloth sprayed with disinfectant (mixed solution, Aniospray 29) and a ten-minute UV-C treatment. The median microbial reduction (in CFUs) was 100 % after UV-C treatment, 98.4 % after cleaning with an antiseptic wipe, and 87.5 % after wiping with dry paper. Bacteria-free swabs were seen in 88 %, 16 % and 4 % of cases [56].
A study from Turkey found comparable efficacy for the cleaning of ultrasound probes with 80 °C hot water for five minutes and cleaning with antiseptic wipes (Cleanisept® Wipes) based on 20 swabs in each case: Coagulase-negative staphylococci were detected in 5 % of cases (1/20) [57].
-
In daily use there are primarily microorganisms of normal skin flora on ultrasound probes.
-
These microorganisms and pathogenic bacteria like MRSA cannot be reliably eliminated by wiping.
-
There are numerous proposed and used cleaning and disinfection methods with varying success rates.
Percutaneous interventional ultrasound
Given the broad use of ultrasound for guiding invasive measures, there are astonishingly few publications that address hygiene in this regard. The available publications regarding infectious complications in ultrasound-guided interventions are case histories and largely retrospective analyses or surveys [58]
[59]
[60]
[61]
[62]. However, retrospective surveys have the risk of overlooking clinically inapparent infections or infections treated elsewhere. In addition, the hygiene conditions and antiseptic measures are usually not standardized and therefore cannot be tracked or compared in all cases.
A large retrospective study from Italy found only three infections in 16 648 biopsies and 3035 therapeutic interventions involving the liver over a period of 22 years, all after ethanol instillation. However, only 8 % of the 12 962 biopsied patients were reexamined with documentation after two to three weeks [63].
In the case of a retrospective analysis of a biopsy register of the Mayo Clinic in Rochester, infections were detected in 8 (0.1 %) of the 6613 primarily ultrasound-guided liver biopsies performed over 12 years. Unfortunately, the authors do not provide any information regarding hygiene measures. All eight infections (cholangitis, chologenic sepsis, infected biliary fistulas) occurred in patients with biliary obstruction or after liver transplantation. The authors discussed periinterventional antibiotic prophylaxis for this subgroup [64].
Since infections can also occur with delay, infectious complications were not recorded in the multicenter prospective PIUS study by DEGUM published in 2015 [65]. The available prospective data is analyzed in the following.
Infections after ultrasound-guided percutaneous injection/biopsy
An older prospective study from Italy examined 719 diagnostic and therapeutic abdominal interventions in 558 patients. An antisepsis regime analogous to that of venipuncture was used in this context: Skin disinfection with an iodine solution, ultrasound probe disinfection (70 % alcohol), no sterile covers, no contact between needle and ultrasound probe. The coupling medium was sterile gel or Vaseline. Puncture was performed using the freehand and in-plane technique. The observation period was five days. Serologic testing for hepatitis was performed after three and six months in patients and examiners. No signs of infection were seen [66].
A prospective study from Romania examined infectious complications after ultrasound-guided interventions involving the musculoskeletal system and reported no infections in 988 interventions in 945 patients (0). Non-sterile gel was used, the probe was cleaned only with a dry, soft paper, and no sterile cover was used. The examiner had washed but not disinfected his hands and wore gloves, an apron and a mask coving the mouth. The patient’s skin was disinfected with 70 % alcohol and a povidone-iodine solution or povidone-iodine solution alone. A follow-up examination was performed after three weeks as planned. Otherwise, the patient was instructed to call in the event of infection [67].
In contrast to prospective studies, two large retrospective analyses from the Mayo Clinic in Rochester found a low incidence of intervention-associated infections:
An analysis of registry data of 13 534 interventions (thoracic and abdominal, biopsy and drainage, but no local ablative procedures) found a total of 14 infections (0.1 %) during the observation period of 12 months [68]. A sterile cover for the ultrasound probe was not used for transcutaneous biopsies but the patient’s skin was disinfected [69]. The infections included abscesses, bacteremia, and peritonitis. There were no infections after fine-needle aspiration and thoracocentesis, while infections occurred in 0.2 % of cases after cutting needle biopsy. Infections were observed more frequently after biopsy of the pancreas (0.9 %) and liver biopsy (0.3 %) than after kidney biopsy (0.09 %), for example [68].
In a further, mixed retrospective and prospective database analysis by the Mayo Clinic, 12 708 ultrasound-guided breast biopsies from the years 2005 – 2013 were evaluated. After skin disinfection, the puncture area was surrounded by sterile surgical drapes, and the ultrasound probes were disinfected with povidone iodine and a quaternary ammonium chloride solution or with isopropanol and a quaternary ammonium chloride solution. The gel was sterile. A cover was not used for the ultrasound probes. The intervention-associated infection rate was 0.11 %. All infections were localized and were able to be treated with short-term oral antibiotic therapy [70].
Infections after ultrasound-guided percutaneous placement of a drain or catheter
We could not find any prospective studies regarding infection rates after ultrasound-guided drain placement. It was problematic that drains are often placed in abscesses and empyemas. Therefore, it is difficult to determine the cause of infection (primary disease vs. intervention).
Two prospective studies examined the rate of catheter-associated bacteremia/septicemia after ultrasound-guided placement of central-venous catheters versus the “landmark technique”. Puncture was performed using barrier precautions and a sterile ultrasound probe cover. A non-randomized observational study from Switzerland found an infection rate of 1.7 % after ultrasound-guided placement of 844 catheters. The rate is similar to that of the “landmark technique” (2 %; 33/1639). The average catheter dwell time was six and five days, respectively [71]. In another randomized study (450 patients per arm), the infection rate was significantly higher. However, it was significantly lower in the ultrasound-guided group (10.4 %) than in the group without ultrasound guidance (16 %). The authors attribute the reduction to the lower number of failed puncture attempts due to ultrasound guidance. The dwell time of the catheters is not specified [72].
Infections after ultrasound-guided local ablation therapy
A large retrospective monocentric study from Italy found only 5 infections in 3035 therapeutic liver interventions in 22 years. All occurred after ethanol instillation: two early abscesses in a total of 315 cyst aspirations (0.6 %) and one case of cholangitis after a total of 372 ethanol instillations of tumors (0.27 %). Two “late” abscesses eight and twelve months after intervention are not to be considered procedure-related [63].
One case of ascites and sepsis (0.3 %) was registered prospectively by the same working group after radiofrequency ablation (RFA) in 375 sessions. The observation period was 24 hours [73].
-
Infections after ultrasound-guided freehand puncture without use of a sterile ultrasound probe cover (“no-touch” technique) could not be identified in the few available prospective studies.
-
Extensive evaluation of registry data found infections after percutaneous ultrasound-guided interventions in a total of 0.1 – 0.2 % of cases.
-
In the case of complex therapeutic interventions, the infection rate is slightly higher.
-
When placing central-venous catheters, ultrasound control seems to result in lower infection rates than the “landmark technique”.
Rigid endoscopic ultrasound
Endorectal ultrasound
Transrectal ultrasound (TRUS) is used to evaluate the prostate and its internal structures in that the ultrasound probe is inserted into the rectum and positioned in the immediate vicinity of the prostate. Rigid endorectal ultrasound is also used to stage rectal cancer, to evaluate fistulas, and to evaluate the pelvic floor and the sphincters.
Moreover, ultrasound-guided biopsy of the prostate, e. g. for the diagnosis of prostate cancer, is largely performed with endocavitary probes ([Fig. 5]), with a puncture needle penetrating the rectal mucosa which is populated by microorganisms.
Contamination of endorectal probes
45 % of 60 swabs of endorectal probes in the clinical routine over three months contained microorganisms from the environment, while 3 % contained potential pathogens [27]. A relatively current metaanalysis regarding contamination of endovaginal and endorectal probes included 32 studies from 1993 – 2011. The rate of pathogenic microorganisms was almost 13 % incl. 1 % viruses (HSV, HPV, CMV) [74].
The perforation rate of the latex covers on endorectal probes is high and can be as much as 9 % after puncture [74]
[75].
Infections after endoscopic ultrasound-guided transrectal biopsy
After transrectal prostate biopsy, infectious complications are seen in 1 – 17.5 % of cases. These are primarily urinary tract infections and prostatitis. Sepsis occurs more rarely (0.3 % – 3.1 %) [74]
[76]
[77]
[78]
[79]
[80]
[81]
[82]
[83]. In a population study including 75 190 men, infections are the most common complication with a rate of 72 % [84].
According to data from the SEER registry, the risk for subsequent hospitalization due almost exclusively to infectious complications after transrectal prostate biopsy is 2.65 times higher compared to a randomly selected control population without prostate biopsy [85]. Infections with gram-negative bacteria (E. coli, klebsiella) and clostridium as well as with pathogens from venereal infections (e. g., Treponema pallidum, Neisseria gonorrhoeae, HIV) are primarily seen. Infections with multiresistant microorganisms not included in conventional periinterventional antibiotic regimens are detected in a relevant percentage of cases with sepsis [76]
[86].
In particular, the difficult-to-disinfect internal puncture channel in the probe [87] and the repeated use of the initially sterile biopsy needle are problematic here. Rectal colonization with fluoroquinolone-resistant E. coli, previous antibiotic treatment, a medical profession, and international travel were identified as risk factors in addition to multiple biopsy [76]
[80]
[88]
[89]
[90]
[91].
Value of antibiotic prophylaxis
Antibiotic prophylaxis lowers the rate of symptomatic infections after transrectal prostate biopsy even if it is still unclear which medications should be used and for how long [77]
[92]
[93]. Antibiotic prophylaxis for transrectal endoscopic ultrasound-guided prostate biopsy is included in the German and European guidelines [94]
[95].
Prospective randomized studies and metaanalyses show that targeted antibiotic prophylaxis on the basis of microbiological examination of rectal swabs is more effective than standard prophylaxis [96]
[97]
[98]
[99]. Empirical combination prophylaxis is also more effective than the standard approach [92]
[99]
[100].
Other preventative measures
A metaanalysis of 7 studies and 2049 patients found a significant relative reduction in the risk for fever and bacteremia when antibiotic administration was combined with rectal enemas with a povidone-iodine solution [101]. In a prospective randomized study not included in this analysis including 865 men, only a tendency but no significant reduction in infectious complications from 3.5 % to 2.6 % could be observed [102].
The literature also includes a recommendation for formalin disinfection of the biopsy needle in multiple punctures of the prostate for reducing the risk of sepsis [103].
-
The infection rate after transrectal prostate biopsy is high.
-
Rigid rectal endoscopic ultrasound probes are often contaminated after examination.
Endovaginal ultrasound
Transvaginal ultrasound probes are standard ultrasound equipment in gynecology and obstetrics. They are routinely used in gynecological diagnosis, in reproductive medicine, in early prenatal diagnosis, and for evaluating the fetal brain based on cephalic presentation and for evaluating the cervix and the internal cervix in advanced pregnancy. As a rule, the vaginal probe for ultrasound examination is covered with a latex cover partially filled with gel that is then removed and disposed of after the examination. Compared to simple commercial covers, condoms have a similar leakage rate [104].
The use of only a latex cover for a vaginal probe does not meet the necessary requirements for application and preparation [14]. Vaginal ultrasound probes are categorized as semi-critical medical devices of group A, i. e., they are to be prepared according to the manufacturer’s specifications after every examination of a patient [14]
[105]
[106].
Contamination of endovaginal probes
Cross-contamination with various pathogens (e. g. MRSA, HBV, HCV, HIV, HSV, HPV, CMV) can occur when handling the probe, ultrasound gel, and cover [6]
[14]
[74]
[107]. Two metaanalyses describe the contamination rates of endovaginal probes:
One found a pooled endovaginal probe contamination rate of 12.9 % for bacteria and 1 % for viruses after “low-level disinfection”. The latex covers were contaminated with viruses (HSV, CMV, HPV) in 19.4 % of cases. Immediately after removal of the latex cover, the contamination rate is 33.7 % [74].
The other metaanalysis found a contamination rate of the ultrasound probes of 50 % for bacteria and 4 % for viruses immediately after examination, with testing only for HPV being performed in most cases [104].
The ultrasound probe handle also presents a potential risk for the transmission of microorganisms. Therefore, it must always be ensured that the handle is sufficiently cleaned and disinfected [107]
[108]
[109].
Infections following endovaginal ultrasound
Cases of infection from in-vitro fertilization have been reported [110]
[111]
[112]. These cases were infections with the patient’s own vaginal flora and transmission of HCV. Large prospective studies also describe almost no infections. Therefore, there was only one case of fever and no documented infections in 1058 individual aspiration procedures [113].
-
Endovaginal probes are frequently contaminated with bacteria after examination even when used correctly.
-
Only individual cases of actual infection due to microorganisms transmitted via vaginal endoscopic ultrasound have been reported.
Ultrasound probe preparation
In the daily routine probes must be able to be cleaned and disinfected quickly and reliably to keep the time interval between examinations as low as possible. Ideally the ultrasound probe does not need to be disconnected from the ultrasound unit or brought to another room for preparation. The KRINKO-/BfArM recommendation requires a separate area for preparation which can however be in the examination room [14]. Additional important aspects are compatibility of the cleaner and disinfectant with the probe material and safety for patients and examiners.
After removal of the cover, the probe must be cleaned as well as undergo bactericidal, fungicidal, and virucidal disinfection [14]
[105]
[108]
[114]
[115]
[116].
Various methods are used for disinfecting rigid endoscopic ultrasound probes: Treatment with virucidal wipes [117]
[118], wipes plus foam [105], immersion disinfection methods [117]
[119], a fully automatic disinfection system with hydrogen peroxide aerosol [107]
[109]
[120]
[121] or methods in which UV-C radiation is used [122]
[123]. While most bacteria but not always all viruses and no bacterial spores are killed when using wipes alone, all microorganisms, including some bacterial spores, are inactivated with the other methods. The current WFUMB guidelines regarding the preparation of endovaginal probes provides an overview [105].
It was shown that the reduction in test organisms with simple wipe disinfection is not always sufficient. This was due to joints, grooves, and corners that cannot be sufficiently reached and cleaned with a wipe [107].
In addition, it was able to be shown that various disinfection liquids used in the immersion method, e. g. glutaraldehyde, orthophthaldehyde, ethanol, isopropanol and phenol, were not able to inactivate HPV 16 [117]
[119]
[121]
[123]. Immersion disinfection methods take the longest treatment time (at least 15 min) and there is an inhalation health risk in poorly ventilated rooms [105]
[120].
-
Various methods can be used for disinfecting rigid endoscopic ultrasound probes.
-
Disinfection with simple wiping without ensuring that the disinfectant reaches joints and edges does not provide sufficient disinfection in the case of rigid endoscopic ultrasound probes.
Flexible endoscopic ultrasound
Since endoscopic ultrasound units are inserted through digestive and respiratory tracts populated by microorganisms, bacterial translocation of local flora (primarily alpha hemolytic streptococci and staphylococci) can occur. In an overview regarding the safety of transesophageal echocardiography (TEE), bacteremia rates of 0 – 17 % are specified [124]. In the case of gastrointestinal diagnostic endoscopic ultrasound without intervention, cases of transient bacteremia are registered in approximately 2 % of all cases without clinical signs of infection being observed in prospective studies [125]
[126]. Apparent infections as complications of noninvasive endoscopic ultrasound are rare and only described in case histories. For example for TEE, legionella infections from contaminated water [127] and E. coli pneumonia from an echoendoscope with a defective outer layer have been reported [128].
Endoscopic ultrasound interventions have become an indispensable part of visceral medicine and oncology diagnosis and treatment. According to data of the German Endoscopic Ultrasound Registry of DEGUM, endoscopic ultrasound-guided fine needle aspiration (EUS-FNA) or fine needle biopsy is performed in approximately 13 % of all endoscopic ultrasound examinations (EUS) and a therapeutic intervention is performed in almost 3 % of cases. All of these interventions have in common that an instrument inserted through the instrument channel contaminated with the bacterial flora of the tracheobronchial system, esophagus, stomach, duodenum or rectum is inserted transmurally into a primarily sterile compartment or also into an infected cavity (necrosis, abscess, empyema). In the case of various puncture and intervention targets, passage of the needle through the peritoneal or pleural cavity cannot be avoided.
A small study (n = 43) showed contamination of endobronchial ultrasound-guided transbronchial aspiration needles (EBUS-TBNA) with local oropharyngeal flora in 15 cases (35 %) [129]. Mechanical transmural translocation of local flora is thus a possible epiphenomenon of endoscopic ultrasound interventions. In contrast to percutaneous ultrasound-guided interventions, neither disinfection nor barrier precautions [8]
[21]
[130]
[131] are possible or productive.
Bacteremia after EUS-FNA and EBUS-TBNA
Three prospective studies with a total of 202 patients examined the rate of bacteremia and infections after EUS-FNA in the upper digestive tract. Bacteremia caused by pathogens that were probably not the result of contamination of the blood cultures was observed in five patients (2.5 %). Clinical signs of infection did not occur 125, 126, 132].
The bacteremia rate of 1 – 2 % in prospective studies was comparably low even after EUS-FNA of the lower digestive tract [125]
[133]
[134]. A bacteremia rate after endoscopic ultrasound of 15 % (12 % after EUS without FNA, 6 % after EUS-FNA) was seen in a small population of 41 patients with liver cirrhosis [135]. A bacteremia rate of 7 % was reported for EBUS-TBNA in a small study [129]. These bacteremia rates are in a range that was also reported for flexible bronchoscopy and flexible esophagogastroduodenoscopy (1 – 7 %) [129]. In a prospective animal experiment, the rate of bacteremia after EUS-FNA was not affected by the use of a 0.5 % povidone-iodine solution for moistening the mucosa or by temporary treatment with the proton pump inhibitor omeprazole (40 mg/day) [136].
Fever and infections after EUS-FNA and EBUS-TBNA
A systematic review of 51 studies published by December 2007 regarding EUS-FNA including 10 941 patients found an average complication rate of 0.98 % and even 1.72 % in prospective studies. Transient fever occurred in 12 cases, and infections in 5 cases (11.2 % and 4.7 % of all complications, respectively). This corresponds to an incidence of 0.1 % (fever) and 0.046 % (infection) [137]. In major case series, transient episodes of fever following transduodenal or transgastral EUS-FNA of solid pancreatic lesions were observed in 0.4 – 1 % of patients [138]. After EUS-FNA of the lower digestive tract, episodes of fever occurred in only 0.89 % of cases in a large case series [133]. Prospective studies did not find a relevant infection risk after transrectal EUS-FNA. However, the study protocols always included prior complete bowel cleansing as for a colonoscopy [133]
[134]
[139]
[140].
Mediastinal lymph nodes and tumors
For mediastinal endoscopic ultrasound-guided fine needle aspiration (EUS-FNA and EBUS-TBNA), the cumulative infection rate in a metaanalysis of 190 studies (16 181 patients) was 0.07 % with infectious complications being less common after EBUS-TBNA (0.02 %) than after EUS-FNA (0.17 %) [141]. However, a multicenter survey analysis from Japan showed a risk of infectious complications and fever after EBUS-TBNA of 0.24 %, which was comparable to that of EUS-FNA. The complications included mediastinitis in 0.1 %, pneumonia in 0.05 %, self-limiting fever in 0.05 %, and other infectious complications in 0.03 % [142]. A Korean study analyzing 552 EBUS-TBNA procedures reported infectious complications in 0.54 % of cases. However, transient fever was observed in a total of 20 % of all patients [143]. Comparable data with a rate of severe infectious complications after EBUS-TBNA of 0.16 % was reported by three Turkish centers [144].
An overview from the year 2012 included twelve published cases of severe infectious complications after transesophageal EUS-FNA of mediastinal lymph nodes, including ten cases of mediastinitis. Five of these cases occurred in patients with sarcoidosis [138]. The results of a survey in Holland including all 14 075 patients who underwent EUS-FNA of mediastinal lymph nodes between 1999 and 2011 in 89 Dutch hospitals provides a good starting point for the rate of severe infectious complications after EUS-FNA of mediastinal target structures. Severe complications occurred in 22 patients (0.16 %) including 15 (0.1 %) of an infectious nature (mediastinitis, mediastinal abscess, pleural empyema). There were three deaths as a result of infectious complications, exclusively in patients with a poor performance status [145]. Sarcoidosis may be a risk factor for mediastinal lymph node enlargement. Two centers in Holland reported five cases of mediastinal abscesses after EUS-FNA or EBUS-TBNA in 252 patients with mediastinal lymph node sarcoidosis (2 %). A systematic overview from the year 2017 included 17 published cases of mediastinitis and/or pericarditis after EBUS-TBNA [146].
Subepithelial tumors
The development of intratumoral abscesses after endoscopic ultrasound-guided biopsy of large gastrointestinal stromal tumors was reported in five cases with a fatality in one case [147]
[148]
[149]
[150].
Cystic lesions and edema
The infection risk after EUS-FNA of cystic lesions is significantly higher than after EUS-FNA of solid lesions. It was able to be shown in experiments that EUS-FNA of a container filled with a culture medium through beef tripe contaminated by bacteria resulted in growth in the culture medium in 100 % of cases. The transmission risk could be reduced to 20 % by treating the tripe with 5 % povidone iodine while cleaning with water, chlorhexidine, and high-proof alcohol was not effective [151]. A systematic review found fever and infectious complications in only 0.05 % of cases after EUS-FNA of solid pancreatic lesions, but in 0.55 % of cases after EUS-FNA of cystic pancreatic lesions even though almost 94 % of the included patients with cystic pancreatic lesions received periinterventional antibiotic prophylaxis [137]. In a current metaanalysis of the risk of EUS-FNA of cystic pancreatic lesions (40 studies including 56 124 patients), the infection rate was 0.44 % [152].
The risk of severe infectious complications after EUS-FNA and EBUS-TBNA of cystic mediastinal lesions seems particularly high. An overview from the year 2012 included eight cases in which mediastinitis occurred after EUS-FNA of mediastinal cysts. Three of these cases occurred in spite of the periinterventional administration of antibiotics [138]. Three additional cases were reported by 2015 [153]
[154]. Severe infections were also reported after EBUS-TBNA of mediastinal cysts [141]
[144]
[155]
[156]. It must be taken into consideration that bronchogenic cysts are often relatively dense and contain material that is not anechoic and therefore can be misinterpreted as solid mediastinal lesions on CT and endoscopic ultrasound [155]
[157]
[158].
EUS-FNA of ascites also has a particular infection risk that probably also applies to pleural and pericardial effusions [138]. One case of death due to sepsis was reported after EUS-FNA of a liver lesion in a patient with untreated obstructive cholestasis probably due to contamination of an obstructed intrahepatic bile duct [159].
Value of antibiotic prophylaxis
Professional societies such as the European Society for Gastrointestinal Endoscopy (ESGE), the American Society for Gastrointestinal Endoscopy (ASGE), the EFSUMB and the German Society for Gastroenterology, Digestive and Metabolic Diseases (DGVS) recommend periinterventional antibiotic prophylaxis prior to EUS-FNA of cystic pancreatic lesions or generally before transintestinal puncture of cysts and edema [160]
[161]
[162]
[163] and evaluate this as a quality indicator in some cases [161]
[164]
[165]. However, these recommendations are based less on verified evidence than on indirect conclusions and expert opinion. Prospective controlled comparisons are not available.
A retrospective comparative study and a metaanalysis did not show a difference in the rate of cyst infections with and without antibiotic prophylaxis [152]
[166]. Three- to five-day oral treatment with gyrase inhibitors or beta-lactam antibiotics is largely used after initial intravenous administration immediately before and during intervention. In retrospective studies, no cyst infection was observed even after intravenous single-shot application of ceftriaxone [167], ciprofloxacin and piperacillin/tazobactam [168] in larger patient populations.
EUS-FNA of mediastinal cysts is considered contraindicated by the DGVS and the EFSUMB guidelines (“should be avoided”) [161]
[162].
General antibiotic prophylaxis is not recommended in EUS-FNA, EBUS-TBNA and transrectal EUS-FNA of solid lesions [160]
[161]
[162]
[163]. Antibiotic prophylaxis can be considered in individual cases when the needle must pass through bile or pancreatic ducts [139], the needle will touch large necrotic areas within the tumor/lymph node, or the patient is immunocompromised.
Other preventative measures
By using contrast-enhanced endoscopic ultrasound, necrotic areas can be identified within solid lesions [169]
[170]
[171]
[172]. To minimize the risk of infection, it seems useful particularly in large, inhomogeneous solid lesions and in possibly cystic lesions containing material that is not anechoic to perform contrast-enhanced endoscopic ultrasound in order to be able to avoid transmural puncture of cystic or necrotic areas or to be able to initiate immediate postinterventional antibiotic therapy [138]
[162].
Since the risk of infection after EUS-FNA of solid lesions is very low even when performing multiple puncture procedures with a single needle (up to seven needle punctures in solid pancreatic lesions), it is not necessary to change the needle to prevent infection.
Local application of disinfecting substances prior to EUS-FNA, as proposed by an group of authors on the basis of experimental data [151], has not yet been evaluated in studies and does not play a role in clinical practice and in guideline recommendations.
Cystic lesions should be completely drained during puncture [139].
In centers with corresponding expertise, complete colon cleansing is generally performed prior to transrectal or transcolic EUS-FNA [139].
EUS-FNA of liver lesions in untreated obstructive cholestasis should be generally avoided [138]
[139].
In patients with clinically relevant bronchopulmonary infections, EBUS-TBNA should only be performed after the infection has been managed.
Fever and infections after endoscopic ultrasound-guided therapeutic interventions
Compared to endoscopic ultrasound-guided fine needle aspiration, endoscopic ultrasound-guided therapeutic interventions are associated with a higher infection risk. This is true for injection, implantation, and drainage procedures. In addition to procedure-related factors (implants remaining in the body, injection of immunosuppressive drugs and glucocorticoids and other fluids), patient factors are also a contributing factor because many of these methods are used particularly in tumor patients for palliative care [162]
[173].
Injection and implantation procedures
After endoscopic ultrasound-guided blockade of the celiac plexus with triamcinolone, retroperitoneal abscesses were described in four cases and pleural empyema in three, while endoscopic ultrasound-guided neurolysis of the plexus with high-proof alcohol does not seem to have a relevant infection risk [173]. Due to the relatively small number of cases in the available studies, definitive data regarding the risk of further EUS-guided injection and implantation procedures is not available [174]
[175].
Drainage procedures
In expert hands, transmural endoscopic ultrasound-guided drainage of pancreatic pseudocysts, biliomas, and of demarcated (peri-) pancreatic necroses and peri-intestinal abscesses is a relatively safe intervention. However, there is a relevant risk of secondary infections due to occlusion or migration of the implanted plastic or metal stents of 4 % (0 – 26 %) in pseudocyst drainage or necrosis drainage and of approx. 2 % in biliary drainage [173]
[174]
[175]
[176]. To prevent infections, the creation of sufficiently wide and stable transmural accesses and a sufficiently long drainage duration are therefore recommended in endoscopic ultrasound-guided drainage interventions [161]
[174].
Value of antibiotic prophylaxis
For endoscopic ultrasound-guided drainage procedures and blockade of the celiac plexus, the DGVS and EFSUMB generally recommend antibiotic prophylaxis due to the relevant infection risk even though there are no prospective studies showing a reduction of the infection risk as a result of the periinterventional administration of antibiotics [161]
[162]
[174].
-
Infectious complications of EUS-FNA and EBUS-TBNA are probably primarily the result of translocation of the oropharyngeal or rectal bacterial flora. They are very rare in fine needle puncture of solid lesions but occur approximately 10 times more often in the puncture of cysts, effusions, and edema. There is a relevant infection risk in selected endoscopic ultrasound-guided transmural interventions.
-
Endoscopic ultrasound-guided fine needle puncture of cystic mediastinal lesions and solid liver lesions in the case of untreated obstructive cholestasis can result in severe infections. Puncture of necrotic areas in solid lesions should be avoided.
-
When puncturing cystic/liquid lesions, complete draining of the lesion via complete aspiration of the lesion (diagnostic puncture) and ensuring sufficiently wide, stable, and long-term transmural accesses (therapy) should be targeted with the goal of minimizing the risk of infection.
-
There is no data regarding periinterventional antibiotic administration in diagnostic endoscopic ultrasound-guided puncture of cystic pancreatic lesions and in endoscopic ultrasound-guided drainage procedures (pseudocysts, abscesses, necroses, obstructed bile and pancreatic ducts). However, this is recommended by professional societies.
Intraoperative ultrasound
Ultrasound probes are inserted into open body cavities during intraoperative ultrasound. There are no known publications regarding infections in this regard. According to EFSUMB and the KRINKO/BfArM recommendations, intraoperative probes must be sterile [14]
[21]. The probes are thermolabile and can only be treated with low-temperature sterilization [16]. This limits their availability.
If these sterilization procedures are not available, scientists from Australia [177] recommend immersion disinfection of the ultrasound probe with a bactericidal (including mycobacteria), fungicidal, and virucidal product (“high-level disinfection”) followed by application of a sterile ultrasound probe cover. The guidelines of the French Radiology Society recommend disinfection with a bactericidal (including mycobacteria), fungicidal, virucidal, and sporicidal product when sterilization is not possible [38].