Key words muscular - ultrasound - elastography
Introduction
Ultrasound (US) elastography is a method for evaluating tissue stiffness. Mainly two
different methods are commercially available – strain and shear wave elastography.
Shear wave elastography can be subdivided into shear wave speed measurements and shear
wave speed imaging according to the EFSUMB guidelines [1 ].
Strain elastography has mainly been evaluated for focal lesions in the breast and
in the thyroid [2 ]
[3 ] and is based on mechanical stress applied by the physician performing the examination.
The elasticity is shown visually on a color scale from blue to red and semiquantitative
values may be calculated. Shear wave elastography evaluates tissue stiffness by applying
an acoustic push pulse, which deforms the tissue. This deformation causes shear waves,
which occur perpendicularly to the US push pulse. The speed of these shear waves is
proportional to tissue stiffness and a quantitative value of stiffness is shown. The
method has mainly been evaluated for assessment of liver stiffness but also recently
in skeletal muscle [4 ]
[5 ].
Tension-type headache (TTH) is subdivided into tension-type headache associated with
or without pericranial muscle tenderness [6 ] and the one-year prevalence of neck pain is 88% in TTH [7 ]. Experimentally, pericranial muscles are significantly more tender in patients with
chronic tension-type headache than in healthy controls [8 ] and the degree of tenderness is strongly correlated to the frequency and intensity
of tension-type headaches [9 ]. In addition, pericranial muscles are significantly harder in patients with chronic
tension-type headache than in healthy controls [10 ], which correlate with muscle tenderness of the trapezius muscle and relate to the
headache state [11 ]. To date, there are no studies using elastography in primary headaches, but US elastography
may be an ideal noninvasive quantitative tool to uncover muscle tension in these patients.
The aim of the study was to establish reference values for shear wave elastography
for the pericranial muscles in healthy individuals, more specifically the m. trapezius,
m. splenius capitis, m. semispinalis capitis, m. sternocleidomastoideus and m. masseter,
and to evaluate day-to-day variations in shear wave speeds and evaluate the effect
of the pennation of the muscle fibers, i. e., scanning parallel or perpendicular to
the fibers.
Materials and Methods
Subjects
All subjects gave informed oral consent to participate in the study. The study was
approved by the Regional Committee on Health Research Ethics (ID: H-15014237). 10
healthy volunteers were included in the study – 5 males and 5 females, median age:
31 years (range: 22–46 years). All subjects were physically active, but were asked
not to do any sports within 48 h before the study. The median weight, height and BMI
were 70.5 kg (range: 48–105 kg), 177 cm (range: 157–195 cm) and 22.4 (range: 19.5–27.6),
respectively. All volunteers were scanned twice with one week separating the two sessions
in order to determine possible day-to-day variation.
The volunteers had the trapezius muscle, the semispinalis capitis muscle, the splenius
muscle, the sternocleidoid muscle and the masseter muscle scanned on both sides ([Fig. 1 ]). The dominant hand and possible sport activity were registered.
Fig. 1 Drawing of the muscles examined in two planes (Thien Phu Do is acknowledged for drawing
the figure)
Ultrasound
A GE Logic E9 system with a 9 MHz linear array transducer and MSK preset (GE, Chalfont
St. Giles) was used for all measurements. All muscles were scanned parallel and perpendicular
to the direction of the muscle fibers. For the trapezius, two measurements, one medial
measurement and one more lateral measurement, were performed. In the masseter muscle
two scan planes perpendicular to each other were obtained, but due to the pennation
of the muscle it was impossible to obtain images with only one direction of the pennation.
Elastography
The system shows real-time color coded shear-wave elastography images in a pre-defined
ROI of the B-mode image with a frame rate of approximately 1 Hz. Cine loops of ten
frames were stored in the system for later post-processing. All recordings were performed
by the same experienced physician with light pre-compression on the skin. The subjects
were examined sitting on a chair with their arms resting in their lap. All recordings
were done from the back of the patient. An example of an elastogram is shown in [Fig. 2 ]
[3 ]
[4 ].
Fig. 2 B-mode image of the trapezius muscle (left) and corresponding image with elastogram
(right). Shear wave speed is shown in the lower left corner for 5 different ROIs (only
number 5 shown).
Fig. 3 B-mode image of the masseter muscle (left) and corresponding image with elastogram
(right)
Fig. 4 B-mode image of the sternocleidomastoid muscle (left) and corresponding image with
elastogram (right)
After all recordings were performed, the shear wave speeds of the different muscles
in m/s were registered. ROIs were placed in each muscle in agreement by two observers.
Circular ROIs were placed to cover as much muscle as possible. Five measurements were
performed in each cine loop from each measuring point in each muscle – one for each
recorded frame. The size and depth of the ROI were registered for all measurements.
Statistics
The mean shear wave speeds for all muscle groups in the two orthogonal planes were
calculated.
The repeated measurements were analyzed in a mixed effects linear statistical model
including sequence, side, dominant side, muscle group, gender, muscle group-scanning
direction interaction, dominant side-scanning direction interaction, sequence-scanning
direction, and sequence-muscle group interaction as fixed effects and subject-dominant
side-muscle group-scanning direction as random effect.
The effects of sequence, dominant side and scanning direction were tested in the statistical
model using SAS 9.4 (SAS Institute). The fixed effects were tested in the full model
using two-sided type 3 tests of effects. A paired t-test was performed to analyze
the variation between days.
Results
A total of 20 repeated samples were measured for each of the 6 groups of muscles for
each subject for right and left sides, adding up to 240 measurements per subject.
The mean shear wave speeds (m/s) in the parallel plane were: 2.45 (SD:+/−0.25) for
masseter, 3.36 (SD:+/−0.75) for semispinalis, 3.04 (SD:+/−0.65) for splenius, 2.75
(SD:+/−0.23) for sternocleidomastoid, 3.20 (SD:+/−0.27) for trapezius and 3.87 (SD:+/−3.87)
for trapezius lateral. [Fig. 5 ] presents histograms for the combination of muscle groups and scanning direction
using subject mean values within each category.
Fig. 5 Probability histogram panel plot. X-axis shows mean shear wave speed and y-axis shows
proportion of subjects for all examined muscles in two planes.
The effect of sequence (day-to-day variation) was statistically significant (p=0.0258,
test for sequence-muscle group interaction). [Fig. 6 ] shows that even though the effect is statistically significant, the effect is not
in the same direction for all muscle groups. Further analysis showed no statistically
significant difference between the two days. The mean shear wave speed for each muscle,
each fiber direction, dominant and non-dominant hand for the two different days is
shown in [Table 1 ].
Fig. 6 Mean shear wave speed for all examined muscles for each scan plane
Table 1 Mean shear wave speed for each muscle, each fiber direction, dominant and non-dominant
hand for each of the two days including SD.
Dominant side
Muscle
Fiber direction
Day
Mean (m/s)
SD
Yes
Masseter
Parallel
1
2.38
0.31
Yes
2
2.38
0.26
Yes
Perpendicular
1
2.49
0.35
Yes
2
2.48
0.41
Yes
Semispinalis
Parallel
1
3.50
1.15
Yes
2
3.67
1.15
Yes
Perpendicular
1
2.36
0.47
Yes
2
2.09
0.53
Yes
Splenius
Parallel
1
2.78
0.57
Yes
2
3.30
1.08
Yes
Perpendicular
1
2.44
0.51
Yes
2
2.18
0.53
Yes
Sternocleidomastoid
Parallel
1
3.14
0.39
Yes
2
2.53
0.30
Yes
Perpendicular
1
2.29
0.54
Yes
2
2.37
0.30
Yes
Trapezius
Parallel
1
3.32
0.60
Yes
2
2.80
0.37
Yes
Perpendicular
1
3.04
0.73
Yes
2
3.03
0.72
Yes
Trapezius lateral
Parallel
1
3.93
0.65
Yes
2
3.71
0.41
Yes
Perpendicular
1
2.82
1.03
Yes
2
2.54
0.29
No
Masseter
Parallel
1
2.62
0.55
No
2
2.43
0.37
No
Perpendicular
1
2.70
0.53
No
2
2.51
0.31
No
Semispinalis
Parallel
1
3.07
0.84
No
2
3.22
0.58
No
Perpendicular
1
2.24
0.58
No
2
2.33
0.63
No
Splenius
Parallel
1
2.93
1.02
No
2
3.15
0.48
No
Perpendicular
1
2.13
0.43
No
2
2.40
0.69
No
Sternocleidomastoid
Parallel
1
2.82
0.47
No
2
2.50
0.30
No
Perpendicular
1
2.24
0.38
No
2
2.39
0.63
No
Trapezius
Parallel
1
3.42
0.78
No
2
3.25
0.49
No
Perpendicular
1
2.88
0.80
No
2
2.43
0.31
No
Trapezius lateral
Parallel
1
3.91
0.65
No
2
3.93
0.62
No
Perpendicular
1
2.75
0.68
No
2
2.33
0.60
The effect of dominant side was not statistically significant (p=0.69, test for dominant
side-scanning direction interaction and p=0.60 main effect of dominant side).
The effect of scan plane in relation to muscle pennation was statistically significant
(p<0.0001, test for muscle-scanning direction interaction). The mean shear-wave speed
when scanning parallel to the muscle fibers was significantly higher than the mean
shear-wave speed when scanning perpendicular to the fibers. The results are shown
as model estimate LSMeans in [Fig. 7 ]. Values differed significantly between muscles.
Fig. 7 Mean shear wave speed estimates when scanning parallel and perpendicular to the muscle
fibers
Discussion
We have examined factors influencing the elasticity of head and neck muscles in healthy
individuals and found that the pennation of the muscle fibers had an influence on
the shear wave speeds measured.
This has been confirmed in other studies. Gennisson et al. examined beef muscle fibers
and found that the shear waves propagated more easily parallel to the muscle fibers
that perpendicular to them and Eby et al. examined shear moduli with increasing strain
with different transducer positions - parallel, perpendicular and 45 degrees – and
shear waves propagated better in the parallel position [12 ]
[13 ].
The shear modulus of tissue may be calculated from the shear-wave speed using the
equation: μ=c2 ρ, where μ is the shear modulus, c is the shear wave speed and ρ is the density which
may be assumed to be 1000 kg/m3 . This formula assumes isotropy, and homogeneous materials, and it may only be accurate
to report shear modulus when the waves are propagating parallel to the underlying
muscle fibers. In other transducer orientations it may be more correct only to report
shear wave speeds [1 ]
[13 ]. We only reported the shear wave speeds in order to be able to compare the values
from scanning parallel and perpendicular to the muscle fibers. Also in the masseter
muscle fibers were interleaved in different directions. In musculoskeletal US, anisotropy
is a well-known artifact especially when scanning tendons. Skeletal muscle may be
considered transversely isotropic but in this transducer orientation anisotropy from
the examined tissue may be a challenge [14 ].
There was a statistically significant effect from the day the volunteers were scanned,
but we do not find it clinically relevant and it may be due to coincidence in this
relatively small population, also because the effect was not in the same direction
for all muscles. Niitsu et al. found a post-exercise hardness of the biceps muscle,
which lasted for two to four days [15 ]. All volunteers were asked not to perform vigorous exercise two days before our
measurements. This may also explain why we did not find a statistically significant
effect of dominant vs. non-dominant hand. Furthermore, the muscles were examined when
relaxed and not during exercise.
Another group has examined US elastography for evaluation of trigger points [16 ]. Although they only examined three persons with chronic neck pain and compared them
with 17 healthy volunteers, they found significantly stiffer trapezius muscles in
the persons suffering from chronic neck pain. They examined the muscle in the transverse
plane and the median shear wave speed was 2.02+/−0.42 m/s in healthy volunteers and
2.49+/−0.39 in persons with chronic neck pain. Our mean values in the trapezius medial
and lateral were: 2.84 +/−0.55 m/s and 2.61 +/−0.49 m/s, respectively, where the medial
position was comparable to the position in the paper. This difference may be due to
the elastography method not being exactly the same. We measured shear wave speeds
from an area sampled in real time and the authors of the other study used point shear
wave elastography from another vendor.
We chose the ROI to cover as much as possible of the muscle thickness to obtain representative
values. We noticed that the signal was dampened in persons with thick muscles and
in the muscles that were lying deeper (semispinalis muscle compared to splenius muscle).
Other studies have shown a dependency on depth in a phantom and in the liver for shear
wave elastography [17 ] and also in muscles [18 ]. Kot et al. found an effect of ROI size on the maximum value of the shear wave speed,
but not on the mean shear wave speed when scanning the thigh muscles [19 ].
Limitations
Our study had some limitations. Although the same radiologist examined the volunteers,
it was impossible to ensure that the measurement was performed at exactly the same
point in the two sessions. It was difficult to distinguish between the splenius and
the semispinalis muscle, which were both rather thin muscles. Also the scan planes
in the masseter were not only parallel or perpendicular to the fibers as the two heads
of the masseter muscles are interleaved. All persons were examined in the sitting
position resting, but even minor tension in the muscles may have influenced the measurements.
Conclusion
The shear wave speed variation depended on the direction of scanning. No difference
was seen between measurements between dominant and non-dominant side. The day-to-day
variation was statistically significant, but the differences were not clinically relevant.
Shear wave elastography may be a method to evaluate muscle stiffness in patients suffering
from chronic neck pain.