Keywords
digital PET-CT - NEMA - NEMA NU2–2018 - PET - PET-CT - resolution - sensitivity -
TOF
Introduction
The positron emission tomography-computed tomography (PET-CT) scanner is used as a
diagnostic imaging tool in the treatment of oncological and nononcological diseases.[1] The special feature is that it is a hybrid fusion modality. The PET component of
this modality plays a crucial role in functional imaging to measure the physiology
and metabolism of tumors in the human body.[2] The CT portion offers dual functionality in PET-CT imaging: one as a standard diagnostic
CT to provide diagnostic and anatomical information similar to the standalone diagnostic
CT scanner, and the other as a PET-assisted modality to provide the CT data for the
attenuation corrections of PET images.[3] Attenuation correction CT data can be diagnostic and nondiagnostic. The nondiagnostic
CT data are acquired with low-radiation exposure factors (low milliampere seconds
[mAs] and peak kilovolt [kVp]). Technological advances in PET-CT, such as improvements
in image quality, performance parameters, and the introduction of deep learning reconstructions
in PET, are leading to major changes in nuclear medicine practice and patient diagnosis
management.[4]
[5] The recent introduction of digital PET-CT technology has led to a reduction in absorbed
radiation dose to patients by PET imaging due to a reduction in injection of radionuclides
as well as a reduction in radiation dose by CT imaging of PET-CT.[4]
[6] The digital PET-CT systems are integrated with a semiconductor-based silicon photomultiplier
(SiPM). It is a replacement for the photomultiplier glass tube (PMT). At present,
most of the PET-CT manufacturers have already released their digital PET-CT systems
and only a few of them have replaced their traditional analog PET-CT system with a
digital PET-CT. Our facility recently installed the United Imaging Healthcare uMIvista
PET-CT system. This is the country's one of the initial uMIvista digital PET-CT system.
uMIvista has a 24 cm wide PET axial field of view (AFOV) with integrated digital SiPM
detector, lutetium yttrium oxyorthosilicate (LYSO) crystal with time-of-flight (TOF)
technology, and 160-slice CT scanner with the fastest rotation time of 0.3 second.[4]
[7]
This study includes the National Electrical Manufacturer Association (NEMA) NU2–2018
performance evaluation of the uMIvista PET-CT system. The latest NEMA NU2–2018 guidelines
have been followed for the evaluation performance parameters of this PET-CT scanner:
axial, tangential, and radial spatial resolution, sensitivity, counting losses, scatter,
randomness, random and counting loss correction, image quality, time and energy resolution,
image uniformity, and image registration alignment. All assay performance tests were
performed using the radiopharmaceuticals fluoro-18-sodium fluoride and 18F-fluorodeoxyglucose
(18F-FDG).[4]
[7]
[9]
[10]
[11]
[12]
Methods
uMIvista PET-CT System
The uMIvista PET-CT (Shanghai United Imaging Healthcare China) system is integrated
with LYSO crystals (Lu1.8Y.2SiO5:Ce) with SiPM detectors and AFOVs of 24 cm. This
PET-CT system is equipped with an 80-row, 160-slice CT scanner. The width of the CT
detector is 40 mm. As for the PET hardware, the system is integrated into a 64680
LYSO crystal connected to SiPM detector modules. The crystal size is 2.76 × 2.76.
The time window is 4 nanoseconds and the system is integrated with three-dimensional
TOF-mode acquisition.
Technical Performance Parameter Assessment: Spatial Resolution and Sensitivity, Count
Losses, Scatter, Randoms Fraction, Correction of Random and Counts Loss, Image Quality,
Timing and Energy Resolution, Image Uniformity, and Image Registration Alignment
The NEMA technical performance of the integrated TOF-PET digital system was calculated
using the NEMA NU2–2018 guidelines standard.[4]
[7]
[8]
[9] The NEMA parameter evaluation of the PET-CT system included spatial resolution (axial,
radial, and tangential at 10, 100, and 200 mm), NEMA sensitivity, NEMA image quality,
contrast recovery coefficient (CRC) and background variability (BV), image uniformity,
count loss correction, scatter fraction accuracy, TOF resolution, count accuracy,
count rate performance (noise-equivalent count rate [NECR]), single and coincidence
count corrections, dead time correction, image uniformity, and PET-CT image registration.
According to NEMA NU2–2018 guidelines, axial, radial, and tangential spatial resolutions
were measured using an 18F-FDG point source.[4]
[7]
[11]
[12] The measurements of the spatial resolution were performed at 10, 100, and 200 mm
from the center. The 18F-FDG point source was manufactured in the radiopharmaceutical
laboratory of a PET-CT facility and has a radioactivity concentration of approximately
1 Megabecquerel (MBq) as specified in the PET-CT manufacturer's instructions for use.
The source was positioned on the scanner and data collection was performed according
to the NEMA protocol. Axial, tangential, and radial spatial resolutions were measured
at 1/2 and 1/8 AFOV of the PET, and the average resolution was calculated. The NEMA
sensitivity test was performed using a sensitivity phantom. The phantom was prepared
with 18F-FDG solution in the radiopharmaceutical laboratory behind the lead (Pb)-shielded
L bench by filling the phantom tubing with 18F-FDG radiopharmaceutical. Activity was
approximately 20 MBq during scan acquisition. The phantom was scanned at the center
and 100 mm from the AFOV. Scans were captured according to the scanner's default protocol.
A coincidence window was used to correct for random coincidence events. A NEMA phantom
with a length of 700 mm and a diameter of 200 mm was set up to assess the scatter
fraction, the count losses, and the random fraction. The phantom was prepared and
filled with approximately 450 MBq 18F-FDG. The prepared phantom was located isocentrically
on the PET-CT scan table and the scan was recorded. This was a long scan of around
12 hours that was done overnight. The collected data were processed according to NEMA
NU2–2018 to calculate the performance values. The acquired raw NEMA-PET data were
used by retrospective reconstruction, and counting losses and random corrections were
formulated. The Monte Carlo method and coincidence time window were used to assess
scatter fraction, random fraction, and count loss. The standard protocol was followed
with TOF, expectation maximization of ordered subsets, and point spread function reconstruction.
A NEMA phantom was prepared and used to calculate the image quality of the acquired
PET images. The phantom had spherical spaces with different diameters (10, 13, 17,
22, 28, and 37 mm); these were prepared with 18F-FDG radioactivity. The rest of the
Phantom had a capacity of 9.8 L. The radioactivity of the 18F-FDG concentration was
four times the background volume of the phantom. The radioactivity of the bead was
approximately 120 MBq during PET data acquisition. The total scan time was approximately
30 minutes, with the standard scan time being 5 minutes per image. The CT scan was
also acquired for attenuation correction of PET images. The captured images were used
to evaluate image quality. CRC, BV, and relative lung error were calculated. TOF resolution
and energy resolution were calculated using full-width half-maximum (FWHM) and full-width
tenths-maximum (FWTM).
Results
Spatial Resolution and Sensitivity
The spatial resolutions of the PET were measured as tangential, radial, and transaxial
spatial resolutions at 10 mm with 3.01 mm, 2.95 mm, and 2.93 mm, respectively; at
100 mm with 3.17 mm, 3.42 mm, and 3.05 mm, respectively; and at 200 mm with 3.65 mm,
4.54 mm, and 3.17 mm, respectively, at FWHM; while at FWTM, the values at 10 mm were
5.79 mm, 5.57 mm, and 5.69 mm, respectively; at 100 mm were 5.59 mm, 5.96 mm, and
5.91, respectively; and at 200 mm were 6.50 mm, 8.07 mm, and 5.95 mm, respectively.
All values are given in [Table 1]. The measured NEMA sensitivity of the uMIvista PET scanner was 12.053 cps/kBq in
the middle. [Fig. 1A] shows center axial sensitivity profiles. Axial, tangential, and radial spatial resolutions
were measured at ½ and ⅛ AFOV of the PET. Both are summarized in [Table 2] and [3].
Table 1
Average resolution over both axial positions
Source location
|
(0, 10) mm
|
(0, 100) mm
|
(0, 200) mm
|
Direction
|
Tangential
|
Radial
|
Axial
|
Tangential
|
Radial
|
Axial
|
Tangential
|
Radial
|
Axial
|
FWHM (mm)
|
3.01
|
2.95
|
2.93
|
3.17
|
3.42
|
3.05
|
3.65
|
4.54
|
3.17
|
FWTM (mm)
|
5.79
|
5.57
|
5.69
|
5.59
|
5.96
|
5.91
|
6.50
|
8.07
|
5.95
|
Abbreviations: FWHM, full-width half-maximum; FWTM, full-width tenth-maximum.
Table 2
Axial, radial, and tangential resolution for 1/2 AFOV
Source location
|
(0, 10) mm
|
(0, 100) mm
|
(0, 200) mm
|
Direction
|
Tangential
|
Radial
|
Axial
|
Tangential
|
Radial
|
Axial
|
Tangential
|
Radial
|
Axial
|
FWHM (mm)
|
3.01
|
3.00
|
3.02
|
3.17
|
3.40
|
3.01
|
3.68
|
4.68
|
3.17
|
FWTM (mm)
|
5.81
|
5.46
|
5.71
|
5.56
|
5.98
|
5.81
|
5.94
|
8.06
|
5.93
|
Abbreviations: AFOV, axial field of view; FWHM, full-width half-maximum; FWTM, full-width
tenth-maximum.
Table 3
Axial, radial, and tangential resolution for 1/8th AFOV
Source location
|
(0, 10) mm
|
(0, 100) mm
|
(0, 200) mm
|
Direction
|
Tangential
|
Radial
|
Axial
|
Tangential
|
Radial
|
Axial
|
Tangential
|
Radial
|
Axial
|
FWHM (mm)
|
3.01
|
2.95
|
2.93
|
3.17
|
3.42
|
3.05
|
3.65
|
4.54
|
3.17
|
FWTM (mm)
|
5.79
|
5.57
|
5.69
|
5.59
|
5.96
|
5.91
|
6.50
|
8.07
|
5.95
|
Abbreviations: AFOV, axial field of view; FWHM, full-width half-maximum; FWTM, full-width
tenth-maximum.
Fig. 1 (A) NEMA sensitivity: (B, C) PET image of the image quality phantom, (B) transverse and coronal image, and (C) lung relative error; (D) scatter fraction, counts rate (total, true, random, and scatter), and NECR; and
(E) percentage error versus radioactivity concentration. NECR, noise-equivalent count
rate; NEMA, National Electrical Manufacturer Association; PET, positron emission tomography.
Count Losses, Scatter, Randoms Fraction, and Correction of Random and Counts Loss
The measured scatter fraction was 39.15% at a maximum NECR of 452.98 kcps at 29.95
kBq/cc and 37.66% at low activity of 119.95 kcps at 17.42 kBq /cc. The measured peak
coincidence count rate is 1,984 kcps and the single peak count rate is 64,495.73 kcps.
All values are described in [Table 5] and [Figs. 1D] and [2A,B]. The count rate errors at different activity concentrations are shown in [Fig. 1E]. The maximum percent error below the peak NECR activity at a peak NECR of 119.95
kcps at 17.42 kBq/cc was 4.8%.
Fig. 2 (A) Counts rate performance; (B) dead time loss estimation; (C) TOF timing resolution; (D) image uniformity; (E) PET-CT image registration alignment; and (F) energy resolution. PET-CT, positron emission tomography–computed tomography; TOF,
time-of-flight.
Table 4
Summary of all contrast recovery coefficients and corresponding background variability
Sphere diameter (mm)
|
10
|
13
|
17
|
22
|
28
|
37
|
Hot contrast (%)
|
54.6
|
60.6
|
66.5
|
74.7
|
|
|
Cold contrast (%)
|
|
|
|
|
85.7
|
88.5
|
Background variability (%)
|
6.2
|
5.1
|
4.0
|
3.2
|
2.7
|
2.3
|
Mean lung relative error (%)
|
2.9
|
|
|
|
|
|
Table 5
Analysis report for scatter fraction, and count rate and loss performance
System scatter fraction @ peak NECR
|
39.15%
|
System scatter fraction @ low activity
|
37.66%
|
Peak true rate
|
452.98 (kcps) @ 29.95 (kBq/cc)
|
Peak NECR
|
119.95 (kcps) @17.42 (kBq/cc)
|
Peak coincidence count rate (kcps)
|
1,984.00
|
Peak single count rate (kcps)
|
64,495.73
|
Peak true count rate (kcps)
|
452.98
|
The max % error below the NECR peak activity
|
4.80%
|
TOF resolution
|
302.294 ps
|
Energy resolution
|
11.76%
|
Transverse uniformity (0.00–0.05)
|
0.036
|
Axial uniformity (0.00–0.05)
|
0.025
|
Accuracy in PET-CT registration (–1.5 to 1.5 mm)
|
|
X direction
|
–0.2510
|
Y direction
|
–0.2620
|
Z direction
|
0.4209
|
Abbreviations: kBq/cc, kilobecquerel per cubic centimeter; NECR, noise-equivalent
count rate; PET-CT, positron emission tomography-computed tomography; ps, TOF, time-of-flight.
Image Quality, TOF Timing Resolution, Energy Resolution, Image Uniformity, and PET-CT
Image Alignment
The CRC and BV ratio of all measured bullets was 4:1, and the CRCs of the radioactivity-filled
hot and cold bullets were 55.6, 60.6, 66.5, and 74.7% at 10, 13, 17, and 22 mm, respectively,
and 85.7 and 88.5% at 28 and 37 mm, respectively; while BV was 6.2, 5.1, 4.0, 3.2,
2.7, and 2.3% at 10, 13, 17, 22, 28, and 37 mm, respectively. The measured relative
lung error was 2.9% as shown in Fig. 1C. The hot and cold sphere CRC and BV in the
4:1 ratio is summarized in [Table 4]. The image section in transverse and coronal views is shown in [Fig. 1B]. The measured TOF timing resolution was 302.294 ps and the measured energy resolution
was 11.76% with FWHM and FWTM. The TOF resolution and energy resolution are summarized
in [Table 5] and [Fig. 2C] and F. The measured transverse and axial uniformity were 0.036 and 0.025, which
was found to be the acceptable limit and is summarized in [Table 5] and [Fig. 2D]. The alignment of the PET-CT registration was measured in the X, Y, and Z directions.
The measured values are within the permissible limits at –0.251, –0.262, and 0.421 mm.
It is summarized in [Table 5] and [Fig. 2E].
uMIvista and Equivalent PET-CT Scanner
The evaluated NEMA NU2–2018 performance parameters of the uMIvista PET-CT system were
compared with those of an equivalent PET-CT system. The comparison was performed with
uMI550 PET-CT, GE Discovery MI and IQ PET-CTs, and Siemens Biograph Vision 600 PET-CT
(as shown in [Table 6]).
Table 6
Comparison of NEMA technical parameters for PET scanners from different manufacturers
Model
|
UIH uMIvista
|
UIH uMI550
|
UIH uMI550
|
Siemens
Biograph Vision 600
|
GE Discovery IQ
|
GE Discovery MI
|
Authors
|
Present study
|
Singh et al (2023)[4]
|
Chen et al (2020)[7]
|
Reddin et al (2018)[10]
|
Reynés-Llompart et al (2017)[11]
|
Pan et al (2019)[12]
|
Manufacturer
|
United
Imaging Healthcare
|
United
Imaging Healthcare
|
United Imaging Healthcare
|
Siemens Healthineer
|
GE Healthcare
|
GE Healthcare
|
CT component
|
160 slices
|
80 slices
|
80 slices
|
128 slices
|
16 slices
|
128 slices
|
Specifications
|
NEMA NU2–2018
|
NEMA NU2–2018
|
NEMA NU2–2018
|
NEMA NU2–2018
|
NEMA NU2–2012
|
NEMA NU2–2012
|
AFOV (cm)
|
24
|
24
|
24
|
26.3
|
25
|
25
|
Crystal
|
LYSO
|
LYSO
|
LYSO
|
LSO
|
BGO
|
LYSO
|
Crystal dimension (mm)
|
2.76 × 2.76 × 16.3
|
2.76 × 2.76 × 16.3
|
2.76 × 2.76 × 16.3
|
3.2 × 3.2 × 20
|
6.3 × 6.3 × 30
|
4 × 4 × 25
|
Detector
|
SiPM
|
SiPM
|
SiPM
|
SiPM
|
PMT
|
SiPM
|
Sensitivity (cps/kBq)
|
Center
|
12.05
|
10.35
|
10.24
|
15.1
|
20.8
|
20.81
|
10 cm
|
11.82
|
9.74
|
10.32
|
15.6
|
20.4
|
20.21
|
Spatial resolution
|
Tan/rad/axial @ 1 cm
|
3.01/2.95/2.93
|
3/3/2.7
|
2.9/3/3
|
3.7/3.5/3.6
|
4.7/4.2/4.8
|
4.3/4.3/5
|
Tan/rad/axial @ 10 cm
|
3.17/3.42/3.05
|
3/3.4/3
|
3/3.3/3
|
3.9/4.6/4.3
|
5.1/5.6/4.8
|
4.6/5.5/6.5
|
Tan/rad/axial @ 20 cm
|
3.65/4.61/3.17
|
3.6/4.6/3.1
|
4/4.1/3.1
|
3.5/5.8/4.4
|
5.5/8.5/4.8
|
5/7.4/6.6
|
Image quality (%)
|
10 mm (CR/BV)
|
54.6/6.2
|
50.5/6
|
46.5/6.4
|
86.8/6
|
18/4.4
|
46/9.3
|
13 mm (CR/BV)
|
60.6/5.1
|
62.3/4.8
|
76.2/5.5
|
72.2/5
|
37/4
|
54/7.1
|
17 mm (CR/BV)
|
66.5/4.0
|
74.7/3.9
|
79.3/4.7
|
85/3.9
|
59/3.6
|
66/5.4
|
22 mm (CR/BV)
|
74.7/3.2
|
81.6/3.3
|
82.8/4
|
89.8/3.3
|
70/3.4
|
71/4.4
|
28 mm (CR/BV)
|
85.7/2.7
|
82/2.7
|
82.1/3.3
|
87.4/3.0
|
61/3.3
|
85/3.8
|
37 mm (CR/BV)
|
88.5/2.3
|
84.5/2.2
|
83.9/2.5
|
89.6/2.2
|
64/3.5
|
89/3.5
|
Mean lung relative error
|
2.9
|
5.6
|
3.1
|
3.5
|
26
|
5.9
|
Count rate
|
Scatter fraction (%)
|
37.66
|
38.5
|
36.65
|
39
|
36.2
|
40.2
|
Peak NECR (kcps)
|
452.98 (kcps) @ 29.95 (kBq/cc)
|
130.05 (kcps) @ 19.28 (kBq/cc)
|
124.4 kcps @ 18.85 kBq/mL
|
296 kcps @ 30.9 kBq/mL
|
123.6 kcps @ 9.1 kBq/mL
|
266.3 kcps @ 20.8 kBq/mL
|
Timing resolution (ps)
|
302
|
372
|
372
|
215
|
Non-TOF
|
381.7
|
Abbreviations: AFOV, axial field of view; BGO, bismuth germanate; cps/kBq, count per
second per kilobecquerel; CR/BV, contrast recovery/background variability; LYSO; lutetium
yttrium orthosilicate; NECR, noise-equivalent count rate; NEMA, National Electrical
Manufacturer Association; PET, positron emission tomography; PMT, photomultiplier
tube; ps, picosecond; SiPM, silicon photomultiplier; TOF, time-of-flight.
Discussion
The uMIvista is the latest model of a digital PET-CT scanner. Post-installation technical
performance evaluation was performed against NEMA NU2–2018 guidelines.[6]
[7]
[8]
[10] The collected NEMA NU2–2018 parameters were compared with existing equivalent PET-CT
scanners for performance evaluation and are summarized in [Table 6]. The compared PET-CT scanners had a similar physical PET field of view. The uMIvista
PET-CT has a 24 cm wide PET field of view, while the uMI550 has 24 cm, the Discovery
MI and IQ have 25 cm, and the Siemens Biograph Vision 600 has 26 cm.[4]
[7]
[10]
[11]
[12]
NEMA NU2–2018 Performance Parameters
The measured spatial resolutions in all three directions (10, 100, and 200 mm) of
uMIvista were similar to those of Singh et al[4] and Chen et al[7] who studied uMI550 PET-CT scanners. This was because the physical and technical
parameters of both PET-CT scanner models are similar, such as: crystal size, crystal
type, detector type, PET ring diameter, and same manufacturer, as described in [Table 6]. When compared with GE Discovery MI and IQ-PET-CT systems reported by Llompart et
al[11] and Pan et al,[12] the measured spatial resolution of the uMIvista PET-CT system was assessed to be
much higher. Similar higher spatial resolution compared with the Siemens Biograph
PET-CT system was reported by Reddin et al.[10] This difference was measured with the uMIvista PET-CT system due to the crystal
size.[10]
[11]
[12]
The sensitivity of the uMIvista PET-CT system was found to be higher than the uMI550
PET-CT system reported by Singh et al[4] and Chen et al.[7] The sensitivity of the Discovery MI and IQ systems was highest (14). This significant
difference was observed due to the thickness of the crystal. Thicker crystals achieve
higher sensitivity. The crystal thickness of the uMIvista and uMI550 systems is thinner
than the Siemens Biograph Vision and GE Discovery MI and IQ systems. This results
in the highest sensitivity of Discovery IQ and MI, followed by Biography Vision, uMIvista,
and uMI550.[4]
[7]
[10]
[11]
[12] The measured scatter fraction was 39.15% at maximum NECR and 37.66% at low activity.
The measured fraction of scatter was consistent with the rest of the PET-CT systems,
with minor deviations, as shown in [Table 6]. The standard values for the scatter fraction range from 35 to 40%. It was observed
that uMIvista and other PET-CT systems had shown similar results. The peak true count
rate and NECR measurements of uMIvista were found to be higher than those of Singh
et al[4] and Chen et al[7] who examined uMI550 system. In addition, it was also higher than the GE Discovery
MI and IQ systems and the Siemens Biograph Vision 600. The measured TOF timing resolution
of the uMIvista PET-CT system was 302.294 ps, reportedly superior to uMI550 studied
by Singh et al[4] and Chen et al[7] as well as to GE. The measured TOF time resolution of the uMIvista PET-CT system
was 302.294 ps and was reportedly superior to that of Singh et al[4] and Chen et al[7] who examined the uMI550 and the GE Discovery MI PET-CT system, while the TOF time
resolution of the Siemens Biograph Vision 600 is higher than that of the uMIvista
PET-CT. Image quality on the NEMA body phantom in terms of CRC, BV, and relative lung
error was measured and observed to be consistent with other digital PET-CT systems
(see [Table 6]) and the analog Discovery IQ system. This was due to the superior performance of
SiPM detectors over PMT, as summarized in [Table 6].
Conclusion
The NEMA NU2–2018 performances of this TOF-integrated digital PET-CT system are observed
to be superior for various parameters.