Basics of Electrosurgery
Thermal Effect on Tissue
The thermal effect of heat on tissue initiates at ≥ 40°C. At this temperature the
effects include initial tissue damage, edema formation, and, depending on the duration
of application, the tissue can recover or die (devitalization). At ≥ 60°C, devitalization
(destruction) of the cells, shrinkage of the connective tissue through denaturation
occurs. The coagulation effect is produced when the tissue is heated more gently (>
60°C to < 100°C). Cutting effect is produced when the tissue is rapidly heated beyond
100°C resulting in rapid vaporization of the fluid and disruption of the cell structures.
Even higher temperatures result in either carbonization (about 150°C) or vaporization
of the tissue (about 300°C) ([Fig. 1]).
Fig. 1 Thermal effects on biological tissue at different temperatures.
Current, Resistance, and Voltage
Current (I) refers to the movement of electric charge (specifically, electrons) within a circuit
over a given time period.
Resistance (R) or impedance represents the degree to which the flow of current is hindered. Resistance
is significantly affected by the content of water and electrolytes in the tissue.
For example, tissues that contain a high amount of water, such as blood vessels, offer
lower resistance compared to those that are less hydrated, such as bone and fat. As
a result, cutting through a fatty lesion using a snare proves to be more challenging
than cutting through a nonfatty polyp under the same electrosurgical conditions. Furthermore,
the presence of fibrosis and scars can elevate the impedance of tissue, potentially
requiring modifications to the power and/or the waveform of the current to achieve
the intended outcome. The accumulation of char on the knife's tip can also obstruct
the flow of current, highlighting the importance of regularly cleaning the electrodes
during a procedure to ensure effective contact coagulation.
Voltage (V) is the force that pushes a current through a resistance and is measured in volts.
Therefore, higher voltages potentially increase the depth of thermal effect (desired
or undesired). Caution is advised while using modes with very high voltages like forced
coagulation, swift coagulation, and spray coagulation.
Current density is the amount of electrical current per unit area of cross-section. Small areas of
contact with the electrode (e.g., snare, sphincterotome, needle knife) produce higher
current density resulting in a rapid and sharp cut. No wonder, precut papillotomy
is still considered dangerous by some experts. While a clean cut without much coagulation
is the desired effect during endoscopic sphincterotomy (EST) to minimize thermal injury
to the pancreatic sphincter, the same may be counterproductive during other procedures
like endoscopic submucosal dissection (ESD) and polypectomy where hemostasis is equally
important. The bulbous tip of the commonly used electrosurgical knives for ESD and
braided nature of polypectomy snares ensure a broad area of contact with the tissue
balancing the speed of cut while maintaining hemostasis due to slow and controlled
cutting. Another practical example of current density is that during endoscopic cutting
or resection the current is very concentrated along the device (knife tip or snare
wire) but dispersed over the much larger area of the grounding pad when it exits the
patient's body. Therefore, while cutting or coagulation occurs at the resection site,
the same amount of current produces no noticeable rise in skin temperature.
Duty cycle refers to the percentage of total time that electrical current is actually delivered.
A duty cycle of 100% implies that the current is delivered continuously for the entire
activation period without any pauses. The ultimate effect on the tissue in pure waveforms
is determined by peak voltage (Vp). While a peak voltage of > 200 Vp produces pure
cutting effect, a peak voltage of ≤ 200 Vp produces pure coagulation effect. In clinical
practice, the pure waveforms are modulated by introducing interruptions or pauses
in the waveform to achieve both the effects, that is, cut as well as coagulation.
During the pauses, the target tissue has more opportunity to cool which in turn promotes
greater degrees of tissue coagulation over cutting effect. In general, lower is the
duty cycle (longer pause) more is the coagulation effect. Therefore, ESU modes with
predominant coagulation effects (Forced Coag, Spray Coag, Swift Coag) have low duty cycles ranging from 4 to 8% and higher voltages (> 650–4,300 Vp).
On the other hand, modes with relatively high duty cycle (30–100%) like EndoCut, PulseCut, and DryCut produce mainly cutting effects with less intense coagulation. Duty cycle or nature
of waveform is a simple way to compare modes named differently across various ESUs
([Fig. 2]; [Table 1]).
Fig. 2 Different types of waveforms and their impact on the tissue.
Table 1
Basic characteristics of various electrosurgical modes
|
ERBE VIO 300D
|
ESG-300
|
|
Setting
|
Voltage
|
Duty cycle
|
Crest factor
|
Setting
|
Voltage
|
Duty cycle
|
|
Soft Coag
|
55–199
|
100%
|
1.4
|
Soft Coag
|
221
|
100%
|
|
EndoCut Q and I
|
0–770
|
100%
|
1.4
|
PulseCut (slow/fast)
|
770
|
100%
|
|
DryCut
|
650–1450
|
30%
|
3.0
|
Blend Cut
|
1400
|
50%
|
|
Swift Coag
|
660–2500
|
8%
|
5.4
|
Power Coag
|
2500
|
8%
|
|
Forced Coag
|
880–1800
|
8%
|
6.0
|
Forced Coag
|
2000
|
8%
|
|
Spray Coag
|
3800–4300
|
4%
|
7.4
|
Spray Coag
|
4300
|
4%
|
Crest factor is the ratio of peak to root mean square voltage. Pure sinus current waveforms have
a crest factor of 1.4, whereas modulated current waveforms have a crest factor ranging
from 1.5 to 8. In general, crest factor is proportional to the intensity of coagulation
provided by that particular waveform or mode on ESU. Crest factor is useful to compare
different modes across various ESUs ([Table 1]).
Monopolar and Bipolar Circuitry
In monopolar mode, the electrical current passes from the active electrode to the
target tissue, through the patient's body, to finally exit the patient through a large
dispersive neutral electrode (also called a patient plate or return electrode). The
vast majority of devices used in therapeutic GI endoscopy are monopolar including
electrosurgical knives, coagulation forceps, polypectomy snares, sphincterotome, etc.
In bipolar mode, the endoscopic device contains both the active and the neutral electrodes
in close proximity to each other. The electrical current passes directly from one
electrode to the other through a small amount of tissue in contact with both electrodes.
The current does not pass through the rest of the patient's body and no patient plate
is required. Only a few commercially available devices are bipolar in nature like
the SpeedBoat device (Creo Medical), heater probe or Gold Probe (Boston Scientific),
coagulation forceps (Hemostat Y, Pentax), and radiofrequency ablation probes. The
advantage of bipolar instruments is that they use less power to achieve similar effects.
Consequently, bipolar devices are associated with less tissue damage, with decreased
risk of collateral thermal injury and perforation when compared to monopolar devices.[1] Since the current does pass through the patient's body, bipolar instruments are
less likely to interfere with the implanted cardiac devices (ICDs). However, these
bipolar devices are not widely available, in part due to their more complex design
and manufacturing costs relative to their monopolar counterparts[2] ([Fig. 3]).
Fig. 3 Diagrammatic representation of monopolar (A) and bipolar (B) electrosurgery. (With permission from: Rey J F, Beilenhoff U, Neumann C S, Dumonceau
J M. European Society of Gastrointestinal Endoscopy (ESGE) guideline: the use of electrosurgical
units. Endoscopy 2010 42(9):764-72.)[2]
Major Modes in GI Endoscopy
There are two main electrosurgical effects on the tissue, that is, cut and coagulation.
In clinical practice, majority of the electrosurgical modes possess both the properties,
that is, cut and coagulation ([Fig. 4]).
Fig. 4 Common electrosurgical modes used in therapeutic gastrointestinal (GI) endoscopy.
Cutting Modes
Cutting modes commonly used during therapeutic GI endoscopy procedures include EndoCut Q and I (Erbe) and PulseCut slow or fast (Olympus). Although the nomenclature suggests a pure cutting effect,
these modes have blended effects (both cutting and coagulation) on the target tissue.
Examples of pure cut modes include AutoCut and HighCut (Erbe Vio 300D). The common indications for blended (EndoCut and PulseCut) modes
include snare polypectomy, endoscopic mucosal resection (EMR), EST, and mucosal incision
during ESD. The modifiable parameters on ESUs while using these modes include effect
(intensity of coagulation), duration (amount of tissue cut in one go), and interval
(duration between two cut cycles). Duration and interval determine the speed of cutting,
that is, higher duration and lower interval result in faster cutting. On the counter
side, the risk of intraprocedural bleeding may be higher with faster cutting as there
is little time for the coagulation to occur. Similarly, higher effect results in better
coagulation with the downside of increased risk of thermal injury. Therefore, the
settings on ESU are governed by several factors like vascularity of the tissue, wall
thickness, and the accessory used.
DryCut mode, when compared to EndoCut, has better coagulation capability owing to its lower duty cycle and higher peak
voltage. Therefore, this mode is often utilized for mucosal incision especially in
regions with high vascularity like large rectal polyps.
Coagulation Modes
Coagulation modes in the commercially available ESUs include Forced Coag (Erbe and Olympus), Swift Coag (Erbe), Spray Coag (Erbe and Olympus), Soft Coag (Erbe and Olympus), and Power Coag (Olympus). Among these modes, Soft Coag has the lowest peak voltage (≤ 200 V) with no cutting capacity (pure waveform, 100%
duty cycle), and therefore, intended to be used only for preemptive coagulation or
achieving hemostasis during active bleeding. Other coagulation modes have intense
coagulation capacity along with some tissue cutting capability. These modes are utilized
during ESD and third space endoscopy procedures. The modifiable factors in these modes
include power and effect. Lower settings (power and effect) are desirable in thin-walled
regions of GI tract like the duodenum and right colon. On the other hand, higher settings
may be preferred in regions with relatively thicker walls like the stomach and rectum
([Fig. 5]).
Fig. 5 Thermal sensitivity across different locations in the gastrointestinal (GI) tract.
Newer Modes
Several newer electrosurgical modes have been introduced to enhance the safety and
efficacy of endoscopic dissection procedures. One of these modes, preciseSECT, responds
very quickly and precisely to different tissue impedances by dynamically adjusting
the current curve. This mode has been proposed for precise submucosal dissection and
pronounced hemostasis with little thermal damage. The high impedance given by a small
contact surface (i.e., tip of the knife) enhances the dissecting effect. Conversely,
an effective coagulation effect can be achieved through lower impedance with the preciseSECT
mode. This sealing of the vessel is achieved by applying the instrument tip to blood
vessels in a more planar manner (large contact surface). Although several advantages
have been proposed, high-quality trials are required to evaluate the real-world impact
of this novel mode for ESD procedures.
Practical Applications of Electrosurgery
This section outlines the recommended modes and settings on ESUs for different therapeutic
GI endoscopy procedures. It is crucial to understand that these recommendations depend
on several factors, including the operator's personal preference, the specific tools
used (such as the type of knife), the ESU brand, the lesion's location, and the thermal
sensitivity of the GI tract being treated. As such, these recommendations should be
applied with careful consideration of the specific clinical situation and not adhered
to without thoughtful adaptation to the context ([Table 2]).
Table 2
Electrosurgical modes and settings on the electrosurgical generator for various therapeutic
interventions
|
Intervention
|
VIO 200/300 (ERBE)
|
ESG-300 (Olympus)
|
|
Polypectomy and EMR
|
EndoCut Q (E3, D1, I 4-6)
|
PulseCut Slow (E2, 120 W)
|
|
Large pedunculated/rectal polyps
|
EndoCut Q (E4, D1, I 6)
|
Forced Coag (E4, 120 W)
|
|
Right colon polypectomy
|
EndoCut Q (E1-2, D1, I 4-6)
|
PulseCut Slow (E2, 120 W)
|
|
Ablation to reduce recurrence (snare tip soft coagulation)
|
Soft Coag (E4, 80 W)
|
Soft Coag (E3, 50 W)
|
|
Endoscopic sphincterotomy
|
EndoCut I (E2, D3, I3)
|
PulseCut Fast (E2, 120 W)
|
|
Precut papillotomy
|
EndoCut I (E2, D3, I3)
|
NA
|
|
Endoscopic submucosal dissection[a]
|
|
|
|
Marking
|
Forced Coag (E2, 20 W), Soft Coag (E4, 50–80 W)
|
Forced Coag (E2, 20 W) or Soft Coag (E3, 50 W)
|
|
Mucosal incision
|
EndoCut I (E2, D3, I 1), DryCut (E2-3, 30 W)
|
PulseCut Fast (E2, 120 W)
|
|
Submucosal dissection
|
Swift Coag (E4, 30–40 W)
|
Power Coag (E2, 30 W)
|
|
Hemostasis (small vessels)
|
Swift Coag (E4, 30–40 W)
|
Power Coag (E2, 30 W)
|
|
Hemostasis (large vessels)
|
Soft Coag (E4-5, 60–80 W)
|
Soft Coag (E3, 50 W)
|
|
Peroral endoscopic myotomy[b]
|
|
|
|
Mucosal incision
|
EndoCut I (E2, D2, I 2)
|
PulseCut Fast (E2, 120 W)
|
|
Submucosal tunneling
|
Spray Coag (E2, 50 W)
|
Spray Coag (E2, 40 W)
|
|
Myotomy
|
Spray Coag (E2, 50 W), EndoCut I (E2, D2, I 2)
|
PulseCut Fast (E2, 120 W)
|
|
Hemostasis
|
Soft Coag (E4-5, 60–80 W)
|
Soft Coag (E3, 50 W)
|
|
Device-assisted EFTR
|
|
|
|
Marking
|
Forced Coag (E2, 20 W) or Soft Coag (E4, 80 W)
|
Forced Coag (E2, 20 W) or Soft Coag (E3, 50 W)
|
|
Cutting
|
High Cut (E4, 200 W), AutoCut (E5, 180 W)
|
Pure Cut (E1, 120 W)
|
Abbreviations: D, duration; E, effect; EFTR, endoscopic full-thickness resection;
EMR, endoscopic mucosal resection; I, interval; NA, not available.
a While using Dual knife from Olympus.
b While using Triangular tip knife from Olympus.
Endoscopic Retrograde Cholangiopancreatography (Sphincterotomy)
Several cutting currents have been utilized for EST including (1) pure cut, (2) coagulation,
(3) blended (cutting and coagulation), and (4) proprietary currents (EndoCut) which are characterized by an automatically controlled cut system.[3] These currents have been compared for main adverse events associated with EST, that
is, bleeding and perforation. Overall, the data suggest that mild bleeding is higher
with pure cutting modes. Whereas there is no major impact of the mode on perforation,
pancreatitis, and significant bleeding.[4] Although EST can be carried out safely with any of the aforementioned modes, cutting
modes capable of automatic control (EndoCut mode) are commonly used in clinical practice to reduce the risk of mild bleeding
and unexpected zipper cuts.[3] The recommended settings on the commonly used ESU are as follows: EndoCut I (Vio 300D: E2, D3, I3) and PulseCut Fast (ESG 300: E2, 20 W). Active bleeding during sphincterotomy is controlled using
sphincterotome by pressing the blue paddle and activating Forced Coag (E2, 60 W).
Polypectomy and EMR
Polypectomy and EMR are among the most common therapeutic procedures performed in
a GI endoscopy unit. The settings and modes are essentially similar for both polypectomy
and EMR. EndoCut (Erbe) and PulseCut (Olympus) modes are frequently utilized for both the procedures. The effect can be
modulated according to the presumed vascularity of the polyp as well as the location
in GI tract. Since EMR in cases with laterally spreading polyps involves larger areas
of resection (vs. polypectomy in pedunculated polyps), the closure of snare should
be fast and complete to prevent deep thermal injury to the muscle bed. In contrast,
gradual closure of snare while the foot paddle is continuously activated is recommended
when resecting thick stalked polyps. Some experts prefer coagulation mode (Forced Coag) for performing polypectomies. In a well-conducted randomized trial including 928
patients with nonpedunculated colorectal polyps 20 mm or larger, EndoCut and Forced Coag modes were comparable with regards to serious adverse events, complete resection
rate, or polyp recurrence.[5] Therefore, electrosurgical settings can be selected based on the endoscopist's expertise
and preference. The settings for the commonly available ESUs in India have been outlined
in [Table 2].
Endoscopic Submucosal Dissection and Per-oral Endoscopic Myotomy
ESD and peroral endoscopic myotomy (POEM) procedures involve several steps each requiring
different modes and settings on ESU. It is important to note that the settings vary
according to the manufacturer of the ESU, location of the lesion, and the type of
device used. In addition, the use of a particular mode differs according to the preference
of the operator performing the procedure. In general, lower settings are preferred
at locations were the luminal wall is thin (like the duodenum and right colon). Conversely,
higher settings are required at locations with high vascularity and thick mucosa (rectum
and stomach).
Endoscopic Submucosal Dissection
ESD involves several steps including marking, mucosal incision, and submucosal dissection.
The decision to mark around the lesion is at the discretion of the endoscopist. In
general, marking is performed for lesions located in the esophagus and stomach, and
avoided in the colon mainly due to clear and discernible margins of colonic polyps.
Marking should be done using the same knife with lower settings in one of the coagulation
modes (e.g., Forced Coag E1, 20 W). Alternatively, argon plasma coagulation (APC) can also be used for marking
around the lesion. Care should be taken to avoid deep mucosal burns during marking
as it may lead to leakage of submucosal cushion fluid. Mucosal incision is performed
using one of the cutting currents (EndoCut or DryCut). Coagulation currents (Forced Coag or Swift Coag) are utilized for the subsequent steps, that is, precutting and submucosal dissection.
Some experts prefer cutting modes (EndoCut or DryCut) for submucosal dissection to minimize charring. Small intervening vessels are coagulated
with the same mode, whereas soft coagulation is preferred in cases with more than
mild bleeding or failure to achieve hemostasis with the previous coagulation modes.
It is important to note that the settings on ESU differ according to the manufacturer
as well as the accessory used for the procedure. Therefore, it is crucial to recognize
the type of knife and the recommendations from the manufacturer before implementing
similar settings for the procedure.
Peroral Endoscopic Myotomy
POEM procedure involves several steps including mucosal incision, submucosal tunneling,
and myotomy. In contrast to ESD for polyps, mucosa is preserved and muscle is severed
during POEM. In general, the modes and settings on ESU are similar to those utilized
during ESD. Some experts prefer spray coagulation mode for submucosal tunneling for
rapid dissection. Since spray coagulation mode is a noncontact mode with very high
peak voltage (∼4000 V) and strong coagulation effect, caution should be exercised
especially during dissection close to the gastroesophageal junction where the mucosa
and muscle layer are in close proximity to each other.
The authors of this article utilize the following settings while using triangular
tip knife and Erbe Vio 300D: mucosal incision (EndoCut, E2, D2, I2), submucosal tunneling (Spray Coag E2, 50 W), and myotomy (EndoCut or Spray Coag with similar settings).
Hemostasis
Various endoscopic instruments designed for hemostasis include heater probe (Olympus
and Boston Scientific), coagulation forceps, and APC probe.
Heater probe (e.g., BiCOAG Bipolar Probe, Olympus) and Gold Probe (Boston Scientific
Inc., Natick, Massachusetts, United States) are bipolar devices which achieve hemostasis
by mechanical compression (coaptive coagulation) and application of coagulation current
in short bursts (5–10 s). It is mainly used for bleeding peptic ulcers.
Preemptive coagulation and minor bleeding episodes during dissection can be managed
using the dissection knife. Low power settings are used for preemptive coagulation
(Forced Coag, effect 1, 10 W; Spray Coag, effect 1, 7 W; VIO300D). Conversely, for vessels with diameters ≥ 2 mm or for patients
at an increased risk of hemorrhage, hemostatic forceps is preferred to ensure complete
hemostasis by thermocoagulation in the soft coagulation mode (effect 4–5, 80 W; VIO300D).[6] After grasping with the forceps, it is important to gently pull back the vessel
to avoid transmission of current in the deeper layers of the GI wall. Since the colonic
walls are thinner as compared to the gastric wall, a smaller coagulation forceps with
lower opening width (4 vs. 5 and 6.5 mm) is preferred. Soft coagulation is preferred
over other coagulation modes for hemostasis due to its several properties including
slow and deep coagulation, low tissue sticking, and low carbonization. Other coagulation
modes (Swift Coag, Forced Coag, and Spray Coag) produce rapid heating and early charring which hinders subsequent transmission of
the current. As a result, the coagulation may be ineffective. In addition, these modes
have additional cutting properties which may be counterproductive.
During APC, the electrical energy is transferred from the APC probe to the tissue
via ionized, electrically conductive argon gas, that is, plasma. The main advantage
of APC is limited tissue penetration, that is, ≤ 5 mm. APC is mainly used in circumstances
of diffuse bleeding, such as angiodysplasia, gastric antral vascular ectasia, radiation
proctitis, and bleeding ulcers. Another use of APC is ablation of large tumor masses
(debulking) for which higher power settings with longer activation times are required.
In recent studies, APC has also been utilized to ablate the margins after EMR in cases
with colorectal polyps in order to prevent recurrences[7]
[8]
[9] ([Table 3]).
Table 3
Practical application of argon plasma coagulation in gastrointestinal tract
|
Mode
|
Properties
|
Clinical application
|
Settings[a]
|
|
Forced APC
|
Continuous beam with steady power output, fast and effective, large area of hemostasis;
provides effective coagulation and devitalization; used for tumor debulking and coagulation
of acute ulcer bleeding
|
Tumor ablation
Tumor bleeding
Dieulafoy's ulcer
Stent trimming
Bleeding ulcer
|
> 60 W, 20–50 W
30–60 W
30–60 W
30–60 W
30–60 W
|
|
Pulsed APC
|
Intermittent power output, more controlled effect on tissue;
suitable for hemostasis of diffuse and widespread bleeding (GAVE, angiodysplasias)
and for ablation Barrett's esophagus
|
Barrett's esophagus, residual adenomas, radiation proctitis, hemostasis stomach/colon,
stent ingrowth/overgrowth
|
30–50 W (Barrett's), Effect 2
10–30 W, Effect 1 or 2 (others including left colon and rectum)
|
|
Smart/precise APC
|
Automatic power adjustment according to distance from tissue;
works in the lower energy range, suitable for treating angiodysplasias in the right
colon, cecum, and small bowel
|
Hemostasis duodenum and right colon
|
Effect 4–5, 10–30 W
|
Abbreviations: APC, argon plasma coagulation; GAVE, gastric antral vascular ectasia.
a Flow rate recommended according to the site is: esophagus, stomach, rectum: 1.4 l/m;
sigmoid to transverse colon, small bowel: 1–1.4 l/m; cecum and right colon: 0.5 l/m.
Other Procedures
Ampullectomy
The main adverse events associated with endoscopic ampullectomy include bleeding,
pancreatitis, and perforation. Settings with high coagulation capacity are likely
to be associated with lower bleeding, but greater risk of postprocedure pancreatitis
as a result of thermal injury to the pancreatic orifice. On the other hand, pure cutting
modes (e.g., AutoCut) may be associated with less thermal injury to the pancreatic sphincter but higher
risk of procedural bleeding. A randomized trial comparing pure cut (AutoCut, Erbe) and endocut modes for endoscopic papillectomy concluded no difference in the
incidences of delayed bleeding and pancreatitis, but higher rate of crush artifacts
and lower rate of immediate bleeding in the EndoCut group especially in tumors greater than 14 mm in diameter.[10] A recent systematic review concluded that both pure cut and blended endocut modes
are associated with high incidence of adverse events. Therefore, there is an unmet
need to establish a standard method of endoscopic papillectomy, including the electrosurgical
cutting mode.[11] In order to answer this question, Yamamoto et al conducted an elegant study in animals
followed by a small clinical study.[12] The authors concluded that the ideal settings while using Erbe ESUs are EndoCUT-I mode (effect 1, duration 4, interval 1) and a thinner snare wire (0.40 mm Snare
Master, Olympus or 0.48 mm Captivator, Boston Scientific). The authors of this manuscript
advocate similar settings for endoscopic papillectomy.
EUS drainage and Endoscopic full thickness resection
Endoscopic ultrasound-guided drainage using electrocautery enhanced stents (e.g.,
Hot Axios, Boston Scientific) is performed using pure cut mode (80–120 W).
Device-assisted endoscopic full-thickness resection (FTRD, Ovesco, Germany) is used
for full-thickness resection of mucosal as well as submucosal lesions in colon and
upper GI tract. This is one of the few procedures where a pure cutting current is
utilized, that is, AutoCut or HighCut (E4, 180–200 W). Since the clip is deployed before resection, the risk of bleeding
is minimal in this scenario while using a pure cutting current.
Special Considerations
Neutral or Return Electrode
In the current era, “split or dual sensing” grounding pads are used to avoid the issue
of overheating with the older generation grounding pads when placed inappropriately.
The new “neutral electrodes” are equipped with an equipotential ring requiring no
particular plate orientation.[2] Moreover, the newer generation ESUs include pad safety and tissue sensing technology
in their design. As such, the grounding pads are able to communicate with the generator's
computer and alarm system and produce both a visual and audible warning to the operator
that the pad is sensing an impedance situation that, if not corrected, could result
in a temperature rise on the patient's skin under the pad. The best practices for
the use of grounding pads are as follows.[13] The pads should be placed in the most suitable position that is muscular, well vascularized,
and close to the treatment site (e.g., flank and upper thigh). Pad should not be placed
over tattoos, implants, metal, bony protuberances, broken skin, or scar tissue. Although
the pad can be placed on the upper arm, this location should be avoided if the patient
has an ICD.
Implanted Cardiac Devices
ICDs such as pacemakers and cardioverter defibrillators are meant to sense electric
signals from the heart. However, noncardiac signals such as those from ESU may interfere
in the functionality of ICD. The response to this interference may be in the form
of temporary inhibition, inappropriate programming, reversion to asynchronous pacing
mode, and triggering an unintended shock to the patient. Therefore, caution is advised
while using electrosurgery in these patients. Although newer pacemakers are resistant
to electric interference, it is advisable to check the manufacturer's recommendation
for all the ICDs. It is especially important to identify patients who are dependent
on their pacemaker for moment-to-moment maintenance of adequate rhythm and hemodynamics.
The use of bipolar devices is preferred while performing therapeutic GI procedures
in these patients because the energy travels only the short distance between electrodes
and is less likely to pass near the implanted device. Alternatively, use of minimal
effective settings (low power and voltage) and avoidance of continuous activation
of the current for prolonged periods is recommended while using monopolar devices.
The active electrode should be at least 6 inches away from the cardiac device and
its lead system.[13]
Newer Electrosurgical Units
The new generation ESUs are equipped with microprocessor that controls the voltage
to a constant level to achieve a uniform, even incision and coagulation by detecting
discharge at approximately 350 times per second for the ICC (Integrated Cutting and
Coagulation) and 1,000 times per second for the VIO300D. A new ESU (VIO3, Erbe) is
available commercially which detects discharge at 25,000,000 times per second for
the VIO3.[12] This new system is equipped with an unparalleled discharge detection function, enabling
highly accurate voltage control. In the authors' experience, the performance (dissection
and coagulation) of this newly available ESU is distinctly better than the previous
versions.