Magnetic resonance imaging of the lung with a volumetric interpolated 3D-Gradient echo sequence

Magnetresonanztomographie der Lunge mit einer volumeninterpolierten 3D-GradientenechosequenzJ. Biederer1 , J. Graessner2 , M. Heller1
  • 1Klinik für Diagnostische Radiologie, Universitätsklinikum Kiel
  • 2Siemens Aktiengesellschaft, Hamburg
Further Information

Dr. med. Jürgen Biederer

Klinik für Diagnostische Radiologie
Universitätsklinikum Kiel

Arnold-Heller-Straße 9

24105 Kiel 1

Phone: 0431-597-3153

Fax: 0431-597-3151

Email: juergen.biederer@rad.uni-kiel.de

Publication History

Publication Date:
04 October 2001 (online)

Table of Contents #

Summary.

Objective: To evaluate the feasibility of a magnetic resonance imaging protocol for the lung using a volumetric interpolated 3D-gradient echo sequence (3D-GRE). Subjects and Methods: A T1-weighted 3D-GRE sequence was used for volumetric interpolated breath-hold examinations (“VIBE”) of the lung in twelve healthy volunteers (TR/TE 4.5/1.9 ms, flip-angle 12°, matrix 502 × 512 [interpolated]). Three coronal 80 mm slabs (16 partitions, 2.5 mm slices) were added to 3D data sets for multi-planar reformations. No contrast material was applied. Artifacts and resolution of vessel and airway structures in each lung segment were evaluated by two observers (12 volunteers, 228 lung segments). Results: The protocol provided excellent images of vascular and tracheo-bronchial structures with very moderate pulsation artifacts. 224/228 lung segments were imaged with “good” (146/228) or “sufficient” quality (78/228). Segmental and sub-segmental vessels (5th order) were identified due to bright flow signal without administration of contrast material. Walls of segmental bronchi (3rd order) were delineated in all parts of the lung. Conclusions: The high spatial resolution, the excellent visualization of lung anatomy, the low rate of artifacts without respiratory triggering and the short acquisition times are clear advantages of the 3D-GRE (VIBE) sequence compared to existing 2D-GRE strategies for MRI of the lung.

#

Magnetresonanztomographie der Lunge mit einer volumeninterpolierten 3D-Gradientenechosequenz.

Ziel: Bewertung eines MRT-Sequenzprotokolls zur nativen Untersuchung der Lunge mit einer volumeninterpolierten 3D-Gradientenechosequenz (3D-GRE). Methoden: 12 gesunde Probanden wurden mit einer T1-gewichteten 3D-Gradientenechosequenz (TR/TE 4,5/1,9 ms, Flip-Winkel 12°, Matrix 502 × 512 [interpoliert]) in Atemanhaltetechnik ohne Kontrastmittel untersucht (volumetric interpolated breath-hold examination = “VIBE”). Je drei koronare Schichtblöcke von 80 mm Dicke (16 Partitionen, 2,5 mm Schichtdicke) wurden zu dreidimensionalen Datensätzen der gesamten Lunge addiert. Detailauflösung und Artefakte wurden semiquantitativ von 2 Untersuchern im Konsens bewertet (12 Probanden, 228 Lungensegmente). Ergebnisse: Das Untersuchungsprotokoll ergibt eine hervorragende Darstellung der Lungengefäße und des Bronchialsystems. Bildartefakte durch Pulsation sind gering: 224/228 Lungensegmente wurden in guter oder unwesentlich beeinträchtigter Bildqualität erfasst (146/228 gut; 78/228 unw. beeintr.). Helles Flusssignal kennzeichnet segmentale und subsegmentale Gefäße 5. Ordnung auch ohne Kontrastmittelgabe. Wände von Segmentbronchien (3. Ordnung) werden in allen Lungenabschnitten dargestellt. Schlussfolgerungen: Hohe Detailauflösung mit hervorragender Darstellung der Lungenstruktur und geringen Artefakten auch unter Verzicht auf eine kardiale Triggerung bei kurzen Akquisitionszeiten kennzeichnen deutliche Vorteile volumeninterpolierter 3D-GRE (VIBE)-Sequenzen gegenüber der üblichen 2D-GRE Technik.

#

Introduction

The unique anatomy of the lung with it’s low proton density and strong susceptibility artifacts at air-tissue interfaces remains a fascinating challenge for the development of suitable magnetic resonance imaging (MRI) techniques [1]. Fast 2D Fourier transform gradient echo sequences (2D-GRE) with short TE were among the first applicable protocols for breath-hold imaging. As a major disadvantage of 2D-GRE sequences, the use of very thin sections is limited by the reduction in signal to noise ratio [2] [3] [4]. The aim of this study was to create a lung MRI protocol using a T 1-weighted three-dimensional fast Fourier transform gradient recalled echo sequence (3D-GRE) for volumetric interpolated breath-hold examination (“VIBE”). Compared to conventional 2D-GRE sequences, we expected a higher spatial resolution and a reduction of cardiac pulsation artifacts with the particular advantages of 3D-acquisitions, such as contiguous slices and reconstruction capabilities. We suggest a complete protocol for MRI of the lung with 3D-GRE (VIBE) at maximum spatial resolution and with coronal slice orientation.

#

Subjects and Methods

All studies were performed on a commercial 1.5 T MRI unit (Magnetom Vision, Siemens, Erlangen, Germany) with a maximum gradient strength of 25 mT/m and rise time of 600 µs. Twelve healthy volunteers (8 m, 4 f, average age 33,8 years, age range 18 - 64 years) were examined with the body phased array coil. The requirements of the institutional review board were fulfilled and all volunteers gave their informed written consent for the study after being informed about the procedure. No contrast material was applied.

For imaging, we used a T1-weighted 3D-GRE (VIBE) sequence with a repetition time (TR) of 4.5 ms, an echo time (TE) of 1.9 ms and a flip angle of 12°. The technical features of the sequence structure were described in detail by Rofsky et al. [5] and in the manuals of the manufacturer (Siemens, Erlangen, Germany). Briefly, the 3D-GRE (VIBE) is a 3D radio-frequency-spoiled GRE sequence similar to sequences used for 3D-MR-angiograms, but with a reduced flip angle to enhance soft tissue contrast. The sequence uses symmetric echo in the read direction and partial in-plane Fourier sampling in the phase encoding direction. The matrix is interpolated in both directions resulting in images with a maximum in-plane resolution of 512 × 512 pixels. Acquisition time is further shortened by reducing the number of phase encoding steps along the partition direction with asymmetric echo sampling and sinc interpolation (zero filling) [5]. The sequence protocol also includes a quick fat saturation scheme with chemically selective fat-saturation pulses and optional cardiac triggering. The latter was not applied to the present series.

The rectangular field of view was adapted individually to each subject with sizes from 350 to 450 mm for images in coronal orientation. Slabs of 80 mm thickness with 16 partitions and an initial matrix of 251 × 256 (98 %) were obtained within a scan time of 20 s or less (depending on the rectangular field of view). Two additional partitions provided 14 % oversampling for elimination of mirroring artifacts. The effective slice thickness of 5 mm was reduced to 2.5 mm by volume interpolation. All images were displayed with an interpolated in-plane matrix of 502 × 512 pixels. For fat signal suppressed images, initial and interpolated matrices were reduced to 90 % to keep within the time frame of a breath hold. The other parameters remained unchanged. Three coronal slabs were acquired to cover the lung completely. The first slab was centered on the mid-mediastinum and the tracheal bifurcation. Additional slabs were positioned anterior or posterior to this slab with at least 10 mm overlap. Numeric slab positioning in 2.5 mm steps permitted secondary reformations of the whole volume. All images were obtained in deep breath-hold. No respiratory or cardiac-triggering was applied.

The presence of artifacts and the resolution of vessel- and airway structures for each of the 228 lung segments in twelve volunteers were evaluated by two observers, judgement by consensus. Depending on the presence of artifacts, the image quality for each segment was judged as “good”, “sufficient” or “not sufficient”. For image review, we used a commercial work station (MagicViewTM, Siemens, Erlangen, Germany). In addition to the obligatory review of the source images in coronal orientation, we reviewed secondary reformations of images in transverse and sagittal planes using the multiplanar reformation mode of the workstation (MPR).

#

Results

The 20 s breath-hold time was tolerated by all subjects. Imaging times for the acquisition of 3 slabs ranged from 5 - 10 minutes.

The quality of images obtained with the 3D-GRE (VIBE) sequence was high. Lung vessels showed bright signal due to inflow-artifacts, a typical effect on gradient recalled echo sequences. Segmental and sub-segmental vessels (4th to 5th order) could be regularly identified in all parts of the lung without administration of contrast material. The signal intensity of these flow artifacts varied with velocity and direction of the flow towards the slab orientation. As a side-effect, pulmonary arteries and veins could be discriminated by differences in signal intensity on single images (Fig. [1]). Although motion artifacts led to blurring of mediastinal structures next to the heart, the trachea and the bronchial tree were well delineated. The walls from main bronchi, lobar bronchi and segmental bronchi (bronchi of 3rd order) could be identified in all subjects and in all parts of the lung. Even single cartilage rings of the trachea and main bronchi could be detected in some subjects (Fig. [1]).

Zoom Image

Fig. 1 Volumetric interpolated breath-hold 3D-GRE (VIBE) image in a 24 years old male volunteer. Clear delineation of the tracheal bifurcation and the hilar vasculature. NB: Different signal intensities of pulmonary arteries (1) and pulmonary veins (2) without the application of contrast material. Bright signal from cartilage rings of the left main bronchus (3). (coronal slice, TR 4.5 ms, TE 1.9 ms, FOV 400 mm, Matrix 502 × 512, 2.5 mm slice thickness).

Secondary reformations were found to be particularly useful to detect small structures within the mediastinum. Reconstructed images in the axial plane reached a reasonable quality (Fig. [2]).

Zoom Image
Zoom Image

Fig. 2 Pre-aortic lymphnode of 8 × 5 × 4 mm on a coronal primary image (a) and a secondary reconstructed axial slice (b) in a 32 years old male volunteer. Primary coronal images in 3D-GRE (VIBE) technique (TR/TE 4.5/1.9 ms, FOV 380 mm, Matrix 502 × 512, 2.5 mm slice thickness).

Overall, the influence of motion and pulsation artifacts on image quality was moderate. Imaging quality was judged as “good” or “sufficient” in 224 of the 228 lung segments from twelve volunteers (146/228 “good”; 78/228 “sufficient”; 4/228 “not sufficient”). Ghosting artifacts related to cardiac and vessel pulsation predominated. These bright bands were most prominent next to the heart and the aorta, as the phase-encoding was orientated in right-left direction (Fig. [3]). A significant contribution of breathing artifacts was not noted, as all volunteers tolerated the 20 s breath-hold periods.

Zoom Image
Zoom Image

Fig. 3 Moderate ghosting artifacts due to cardiac and vessel pulsation in a 23 years old female volunteer (coronal images in 3D-GRE (VIBE) technique, TR/TE 4.5/1.9 ms, FOV 400 mm, Matrix 502 × 512, 2.5 mm slice thickness).

#

Discussion

The long acquisition times and a lower spatial resolution compared to CT are major problems in MR imaging of the lung. New detector techniques have boosted imaging speed and spatial resolution of helical CT, leaving MRI of lung morphology and fine structure in a difficult situation [6]. Thus recent sequence developments have focused on functional aspects like ventilation or perfusion imaging and lesion characterization [2] [7] [8] [9] [10] [11] [12]. 3D-GRE (VIBE) sequences might open a new perspective for further developments in MR imaging of lung morphology.

3D-GRE sequences with short TR and TE were primarily developed for MR angiography using flip-angles of 20 - 60° to pronounce the signal from blood after the application of T 1-shortening contrast material. 3D-GRE (VIBE) protocols use a similar sequence structure, but with lower flip-angles of 12° - 15° to enhance soft tissue signal. These sequences combine angiographic capabilities and a high spatial resolution with reasonable soft tissue contrast. The first application of 3D-GRE (VIBE) sequences was abdominal MRI. Liver studies with axial slices used a limited in-plane resolution with a matrix of 256 × 256 (interpolated from an initial matrix of 160 × 160) and included a quick fat saturation scheme [5]. Semelka et al. have noticed good qualities of these images for lower parts of the chest. Images from patients referred for abdominal MRI clearly delineated the mediastinum and the lower parts of the lung, as far as they were covered. These results already suggested, that a high imaging quality and negligible image artifacts could be expected [13].

Our protocol for imaging the lung uses nearly the highest in-plane resolution of the sequence and thin slices to optimize spatial resolution. The slice thickness of 2.5 mm and the initial matrix of 251 × 256 (98 %) (502 × 512 after volume interpolation) provide interpolated voxel sizes between 2.5 × 0.70 × 0.65 mm3 (350 mm FOV) and 2.5 × 0.9 × 0.88 mm3 (450 mm FOV). A preliminary series of lung phantom studies confirmed best detection of small solid lesions within lung tissue at a flip angle of 12° and with a slice thickness between 2 and 3 mm. Slices thinner than 2 mm were not favorable, as the signal-to-noise ratio (SNR) decreased. The acceptable SNR of small voxels seems to be an obvious advantage of 3D-GRE (VIBE) sequences. Rofsky et al. described an inherent improvement of SNR in volumetric interpolated techniques due to filtering procedures [5]. However, using a fast 2D-GRE sequence with short TE, the reduction of SNR would have inhibited the use of very thin sections [2] [3] [4]. As a concession to the high spatial resolution of our 3D-GRE (VIBE) protocol, the slab thickness is limited to 80 mm and completion of the study requires the addition of three slabs.

The feasibility of the 3D-GRE (VIBE) imaging protocol could be demonstrated in twelve healthy subjects. The sequence provided good images of lung vasculature and segmentation without administration of contrast material. Small vessels were clearly detected due to their bright flow signal. As a side-effect, we noted different signal intensities of pulmonary arteries and veins on coronal slices. This is particularly interesting, as the overlay of pulmonary arteries and veins impairs the discrimination between the two vessel systems in contrast material studies [7] [8] [9] [14]. The tracheo-bronchial system was also well depicted with delineation of bronchial walls up to the 3rd order. This clear delineation of small lung structure details on 3D-GRE (VIBE) images reflects good qualities of image data processing. In particular, partial volume effects decrease with zero filling interpolation, like it was described for 3D-GRE sequences used in MR-angiography [5] [16].

The overall imaging quality was at least comparable to 2D-GRE studies of the lung [2] [3] [4] [16]. Ghosting artifacts from cardiac and vessel pulsation, which are generally very strong on coronal 2D-GRE images, were absent or only moderate on our 3D-GRE (VIBE) series [13] [16]. Pre-saturation pulses to suppress the influence from moving mediastinal structures - sometimes used for 2D-GRE protocols requiring additional imaging time [16] - were not necessary. The 3D-GRE (VIBE) protocol can be obviously applied without cardiac or respiratory triggering in sufficiently cooperative subjects, as breath-hold imaging minimizes artifacts from respiratory motion [16] [17]. It remains difficult to estimate, how far the still present pulsation artifacts would influence the diagnostic value of the sequence in patients, e. g. for detecting small lesions. As a possible solution, the sequence protocol includes optional cardiac triggering. It was successfully tested in two subjects, but not used for the present series (data not shown).

The possible clinical benefits of the 3D-GRE (VIBE) sequence protocol can be only estimated at this stage. The present study evaluated the sequence in healthy subjects and at standardized conditions. More difficult conditions are encountered in patients, i. e. subjects in respiratory distress. Thinner slabs of 40 mm (eight partitions) have to be tailored to keep within breath-hold time frames of 10 seconds or less. Multiplanar reformations (MPR) will be more difficult to perform, if the deepness of inspiration varies between the acquisitions. A sufficient overlap of the imaging volumes would be needed to avoid missing small lesions at the slab limits. Future hardware upgrades allowing for the acquisition of thicker slabs at maximum spatial resolution and with reduced reconstruction times will contribute to overcome these problems. Although the first experiences are promising (Fig. [4]), larger patient studies including different lung pathologies and with CT correlation are required to prove the diagnostic accuracy of 3D-GRE (VIBE).

Zoom Image
Zoom Image

Fig. 4 Bronchogenic carcinoma in the upper lobe of the right lung (47 years old female patient). Subpleural tumour localization, no chest wall invasion (1). Enlarged lymph node (2) at the right upper lobe bronchus next to the bright flow signal of the Azygos vein. Different signal intensities of pulmonary arteries (3) and veins (4). (3D-GRE [VIBE], coronal images, TR/TE 4.5/1.9 ms, FOV 410 mm, Matrix 502 × 512, 2.5 mm slice thickness, no contrast material applied).

However, short acquisition time, high spatial resolution, good visualization of lung anatomy and a low rate of artifacts are clear advantages of the 3D-GRE (VIBE) sequence. We expect, that 3D-volumetric interpolated sequences with these qualities will soon replace the existing 2D-GRE strategies for MRI of the thorax. Future effort will have to focus on the further reduction of artifacts and on the clinical applications.

#

References

  • 1 Su S, Saunders J K, Smith I CP. Resolving Anatomical Details in Lung Parenchyma: Theory and Experiment for a Structurally and Magnetically Inhomogeneous Lung Imaging Model.  Magn Reson Med. 1995;  33 760-765
  • 2 Hatabu H, Chen Q, Stock K W, Gefter G B, Itoh H. Fast magnetic reso- nance imaging of the lung.  Eur J Radiol. 1999;  29 114-132
  • 3 Alsop D C, Hatabu H, Bonnet M, Listerud J, Gefter W. Multi-Slice, Breathold Imaging of the Lung with Submillisecond Echo Times.  Magn Reson Med. 1995;  33 678-682
  • 4 Moody A R, Bolton S C, Horsfield M A. Optimization of a Breat-Hold Magnetic resonance Gradient Echo Technique for the Detection of Interstitial Lung Disease.  Invest Radiol. 1995;  30 730-737
  • 5 Rofsky N M, Lee F S, Laub G, Pollack M A, Krinsky G A, Thomasson D, Ambrosino M M, Weinreb J C. Abdominal MR imaging with a Volumetric Interpolated Breath-hold Examination.  Radiology. 1999;  212 876-884
  • 6 Thompson B H, Stanford W. MR imaging of pulmonary and mediastinal malignancies.  Magn Reson Imaging Clin N Am. 2000;  8 729-739
  • 7 Böck J C, Pison U, Kaufmann F, Felix R. Gadolinium-DTPA-Polylysine-enhanced Pulmonary Time-of-Flight MR Angiography.  J Magn Reson Imaging. 1994;  4 473-476
  • 8 Mai V M, Berr S S. MR Perfusion Imaging of Pulmonary Parenchyma Using Pulsed Arterial Spin Labeling Techniques: FAIRER and FAIR.  J Magn Reson Imaging. 1999;  9 483-487
  • 9 Bremerich J, Roberts T PL, Wendland M F, Wyttenbach R, Arheden H, Reddy G P, Shafagi N, Higgins C B, Saeed M. Three-Dimensional MR Imaging of Pulmonary Vessels and Parenchyma With NC100150 Injection (ClariscanTM).  J Magn Reson Imaging. 2000;  11 622-628
  • 10 Amundsen T, Torheim G, Waage A, Bjermer L, Stehen P A, Haraldseth O. Perfusion Magnetic Resonance Imaging of the Lung: Characterization of Pneumonia and Chronic Obstructive Pulmonary Disease. A Feasibility Study.  J Magn Reson Imaging. 2000;  12 224-231
  • 11 Edelman R R, Hatabu H, Tadamura E, Potthumarthi W L, Prasad V. Noninvasive assessment of regional ventilation in the human lung using oxygen-enhanced magnetic resonance imaging.  Nature Med. 1996;  11 1236-1239
  • 12 Biederer J, Reuter M, Winkler C, Both M, Richter C, Schnabel A, Gross W -L, Heller M. MR-Tomographie der Lunge bei Wegenerscher Granulomatose.  Fortschr Röntgenstr. 2001;  173 S32
  • 13 Semelka R C, Cem Balci N, Wilber K P, Fisher L L, Brown M A, Gomez-Caminero A, Molina P L. Breath-hold 3D gradient-echo MR imaging of the lung parenchyma: Evaluation of reproducibility of image quality in normals and preliminary observations in patients with disease.  J Magn Reson Imaging. 2000;  11 195-200
  • 14 Pielleul F, Naeem M erchant. MRI of the pulmonary veins: Comparison between 3D MR angiography and T1-weighted spin echo.  J Comput Assist Tomogr. 2000;  24 683-687
  • 15 Du Y P, Parker D L, Davis W L, Cao G. Reduction of partial-volume artifacts with zero-filled interpolation in three-dimensional MR angiography.  J Magn Reson Imaging. 1994;  4 733-741
  • 16 Biederer J, Reuter M, Steffens J -C, Schnabel A, Gross W -L, Heller M. MRT der Lunge mit T1-gewichteten Gradientenecho- und T2-gewichteten TSE-Sequenzen - Untersuchung gesunder Probanden und Vorstellung klinischer Fallbeispiele mit CT-Korrelation.  Fortschr Röntgenstr. 2000;  172 S 15
  • 17 Leutner C, Gieseke J, Lutterbey G, Kuhl C K, Flacke S, Glasmacher A, Theisen A, Wardelmann E, Grohe C, Schild H H. MRT versus CT in der Diagnostik von Pneumonien: Evaluation einer T2-gewichteten ultraschnellen Turbo-Spin-Echo-Sequenz (UTSE).  Fortschr Röntgenstr. 1999;  170 449-456

Dr. med. Jürgen Biederer

Klinik für Diagnostische Radiologie
Universitätsklinikum Kiel

Arnold-Heller-Straße 9

24105 Kiel 1

Phone: 0431-597-3153

Fax: 0431-597-3151

Email: juergen.biederer@rad.uni-kiel.de

#

References

  • 1 Su S, Saunders J K, Smith I CP. Resolving Anatomical Details in Lung Parenchyma: Theory and Experiment for a Structurally and Magnetically Inhomogeneous Lung Imaging Model.  Magn Reson Med. 1995;  33 760-765
  • 2 Hatabu H, Chen Q, Stock K W, Gefter G B, Itoh H. Fast magnetic reso- nance imaging of the lung.  Eur J Radiol. 1999;  29 114-132
  • 3 Alsop D C, Hatabu H, Bonnet M, Listerud J, Gefter W. Multi-Slice, Breathold Imaging of the Lung with Submillisecond Echo Times.  Magn Reson Med. 1995;  33 678-682
  • 4 Moody A R, Bolton S C, Horsfield M A. Optimization of a Breat-Hold Magnetic resonance Gradient Echo Technique for the Detection of Interstitial Lung Disease.  Invest Radiol. 1995;  30 730-737
  • 5 Rofsky N M, Lee F S, Laub G, Pollack M A, Krinsky G A, Thomasson D, Ambrosino M M, Weinreb J C. Abdominal MR imaging with a Volumetric Interpolated Breath-hold Examination.  Radiology. 1999;  212 876-884
  • 6 Thompson B H, Stanford W. MR imaging of pulmonary and mediastinal malignancies.  Magn Reson Imaging Clin N Am. 2000;  8 729-739
  • 7 Böck J C, Pison U, Kaufmann F, Felix R. Gadolinium-DTPA-Polylysine-enhanced Pulmonary Time-of-Flight MR Angiography.  J Magn Reson Imaging. 1994;  4 473-476
  • 8 Mai V M, Berr S S. MR Perfusion Imaging of Pulmonary Parenchyma Using Pulsed Arterial Spin Labeling Techniques: FAIRER and FAIR.  J Magn Reson Imaging. 1999;  9 483-487
  • 9 Bremerich J, Roberts T PL, Wendland M F, Wyttenbach R, Arheden H, Reddy G P, Shafagi N, Higgins C B, Saeed M. Three-Dimensional MR Imaging of Pulmonary Vessels and Parenchyma With NC100150 Injection (ClariscanTM).  J Magn Reson Imaging. 2000;  11 622-628
  • 10 Amundsen T, Torheim G, Waage A, Bjermer L, Stehen P A, Haraldseth O. Perfusion Magnetic Resonance Imaging of the Lung: Characterization of Pneumonia and Chronic Obstructive Pulmonary Disease. A Feasibility Study.  J Magn Reson Imaging. 2000;  12 224-231
  • 11 Edelman R R, Hatabu H, Tadamura E, Potthumarthi W L, Prasad V. Noninvasive assessment of regional ventilation in the human lung using oxygen-enhanced magnetic resonance imaging.  Nature Med. 1996;  11 1236-1239
  • 12 Biederer J, Reuter M, Winkler C, Both M, Richter C, Schnabel A, Gross W -L, Heller M. MR-Tomographie der Lunge bei Wegenerscher Granulomatose.  Fortschr Röntgenstr. 2001;  173 S32
  • 13 Semelka R C, Cem Balci N, Wilber K P, Fisher L L, Brown M A, Gomez-Caminero A, Molina P L. Breath-hold 3D gradient-echo MR imaging of the lung parenchyma: Evaluation of reproducibility of image quality in normals and preliminary observations in patients with disease.  J Magn Reson Imaging. 2000;  11 195-200
  • 14 Pielleul F, Naeem M erchant. MRI of the pulmonary veins: Comparison between 3D MR angiography and T1-weighted spin echo.  J Comput Assist Tomogr. 2000;  24 683-687
  • 15 Du Y P, Parker D L, Davis W L, Cao G. Reduction of partial-volume artifacts with zero-filled interpolation in three-dimensional MR angiography.  J Magn Reson Imaging. 1994;  4 733-741
  • 16 Biederer J, Reuter M, Steffens J -C, Schnabel A, Gross W -L, Heller M. MRT der Lunge mit T1-gewichteten Gradientenecho- und T2-gewichteten TSE-Sequenzen - Untersuchung gesunder Probanden und Vorstellung klinischer Fallbeispiele mit CT-Korrelation.  Fortschr Röntgenstr. 2000;  172 S 15
  • 17 Leutner C, Gieseke J, Lutterbey G, Kuhl C K, Flacke S, Glasmacher A, Theisen A, Wardelmann E, Grohe C, Schild H H. MRT versus CT in der Diagnostik von Pneumonien: Evaluation einer T2-gewichteten ultraschnellen Turbo-Spin-Echo-Sequenz (UTSE).  Fortschr Röntgenstr. 1999;  170 449-456

Dr. med. Jürgen Biederer

Klinik für Diagnostische Radiologie
Universitätsklinikum Kiel

Arnold-Heller-Straße 9

24105 Kiel 1

Phone: 0431-597-3153

Fax: 0431-597-3151

Email: juergen.biederer@rad.uni-kiel.de

Zoom Image

Fig. 1 Volumetric interpolated breath-hold 3D-GRE (VIBE) image in a 24 years old male volunteer. Clear delineation of the tracheal bifurcation and the hilar vasculature. NB: Different signal intensities of pulmonary arteries (1) and pulmonary veins (2) without the application of contrast material. Bright signal from cartilage rings of the left main bronchus (3). (coronal slice, TR 4.5 ms, TE 1.9 ms, FOV 400 mm, Matrix 502 × 512, 2.5 mm slice thickness).

Zoom Image
Zoom Image

Fig. 2 Pre-aortic lymphnode of 8 × 5 × 4 mm on a coronal primary image (a) and a secondary reconstructed axial slice (b) in a 32 years old male volunteer. Primary coronal images in 3D-GRE (VIBE) technique (TR/TE 4.5/1.9 ms, FOV 380 mm, Matrix 502 × 512, 2.5 mm slice thickness).

Zoom Image
Zoom Image

Fig. 3 Moderate ghosting artifacts due to cardiac and vessel pulsation in a 23 years old female volunteer (coronal images in 3D-GRE (VIBE) technique, TR/TE 4.5/1.9 ms, FOV 400 mm, Matrix 502 × 512, 2.5 mm slice thickness).

Zoom Image
Zoom Image

Fig. 4 Bronchogenic carcinoma in the upper lobe of the right lung (47 years old female patient). Subpleural tumour localization, no chest wall invasion (1). Enlarged lymph node (2) at the right upper lobe bronchus next to the bright flow signal of the Azygos vein. Different signal intensities of pulmonary arteries (3) and veins (4). (3D-GRE [VIBE], coronal images, TR/TE 4.5/1.9 ms, FOV 410 mm, Matrix 502 × 512, 2.5 mm slice thickness, no contrast material applied).