The endoscopic removal of large biliary stones often requires a mechanical lithotriptor
[1]
[2]
[3]
[4]
[5]. One must anticipate the likelihood that mechanical lithotripsy will be needed and
use a preloaded lithotriptor to capture the stone. Occasionally a stone has been judged
to be extractable using a standard basket but the basket with the captured stone then
becomes trapped in the distal bile duct. The standard approach to such a situation
is to cut off the basket handle and remove the endoscope [2]
[4]. A flexible spiral metal crushing sleeve (e. g. Soehendra lithotriptor, Wilson-Cook,
Winston-Salem, North Carolina, USA) is then advanced over the basket wires [2]
[4]. Traction is applied with a crank under fluoroscopy. Usually the stone will be crushed,
or the basket wires will break releasing the stone [2]
[4]. We describe a novel and simple ”through-the-endoscope” technique for removing a
noncrushing basket with a trapped stone.
The proposed approach involves leaving the endoscope in place. The plastic sheath
of the basket is removed and, if necessary, to accomplish this the plastic handle
is cut off with wire-cutters. A 7-Fr or 10-Fr Medi-Globe metal spiral sheath (Medi-Globe,
Grassau, Germany; formerly manufactured by Pentax, Orangeburg, New York State, USA)
is then advanced through the endoscope channel over the basket wire. Under direct
endoscopic view the metal sheath is passed over the basket wires through the papilla.
When the basket is engaged, the position of the basket and the metal sheath is checked
under fluoroscopy. The basket wire is then tightened with the handle of the Medi-Globe
mechanical lithotriptor under fluoroscopic control until the stone is crushed or the
basket breaks and disengages the stone (Figure [1]). We have attempted this ”though-the-endosope” technique in four cases and succeeded
in all of them.
Figure 1 The plastic sheath of the Dormia basket has been removed and the Medi-Globe metal
sleeve has been inserted via the channel of the endoscope over the basket wire. The
Medi-Globe crank is capturing the basket wire
If a stone has been judged to be extractable using the standard basket, but the basket
with the captured stone then becomes stuck in the distal bile duct, the only currently
available option is to use the Soehendra lithotriptor. When using the Soehendra lithotriptor
we prefer to leave the plastic sheath in place to create a three-layer system (basket
wire, plastic sheath and metal sleeve) because the wire and plastic sheath can be
cranked together. However, the Medi-Globe system handle can capture only the wire.
If the plastic sheath is left in place it will wrinkle and jam the metal sheath as
the wire is cranked. For this reason we recommend removal of the plastic sheath. A
great advantage of our proposed technique is that it can be used with any commercially
available ”noncrushing” baskets. This ”through-the-endoscope” approach avoids the
need for removing the endoscope and then re-intubating the patient to complete the
procedure. This should be of particular benefit in patients with difficult duodenal
access. It also allows for a direct endoscopic view of the passage of the metal sleeve
through the papilla, avoiding the risk of papillary tissue being caught between the
metal sleeve and basket wire. We believe that this technique is easier, faster and,
it is hoped, safer than the standard approach, and can replace the use of the cumbersome
Soehendra lithotriptor.