The New Soehendra Stent Retriever Makes Stent Exchange Much Easier

U.  Seitz 1 , S.  Seewald 1 , X.  He 1 , T.  Akaraviputh 1 , F.  Thonke 1 , N.  Soehendra 1
  • 1Dept. of Interdisciplinary Endoscopy, University Hospital Hamburg-Eppendorf, Germany
Further Information

U. Seitz, M.D.

Dept. of Interdisciplinary Endoscopy, University Hospital Hamburg-Eppendorf, University of Hospital

Martinistr. 52, 20246 Hamburg, Germany

Fax: + 49-40-428034420

Email: seitz@uke.uni-hamburg.de

Publication History

Publication Date:
02 January 2003 (online)

Table of Contents

    Biliary stent exchange is usually performed in two steps: 1) removal of the occluded stent using either a snare or a rat-tooth forceps, and 2) placement of a new biliary stent. However, in cases of complicated strictures, recannulation of the stricture may be difficult or even impossible. The Soehendra stent retriever was designed to avoid such problems and guarantee stent replacement. Those using this device for replacing on occluded stent may, however, encounter difficulties in inserting the guide wire into the stent and advancing the retriever over the guide wire to engage the distal end of the stent. This drawback has recently been eliminated by modifying the tip of the retriever (Figures [1] and [2].)

    Zoom Image

    Figure 1 A 2.5 cm long 5-French curved catheter is mounted on the tip of the retriever, to facilitate the engagement of the retriever with the distal end of the stent and thereby the insertion of the guide wire into the stent.

    Zoom Image

    Figure 2 Endoscopic view showing the coupling of the retriever with the distal end of the occluded stent. The guide wire is then advanced.

    In order to replace an occluded 10-French or 11.5-French plastic stent, the Soehendra stent retriever can now be inserted together with a 400-cm 0.035-inch Teflon-coated guide wire or other hydrophilic wire. The addition of a curved plastic tip significantly facilitates the cannulation of the occluded stent and the insertion of the guide wire (SSR-10-T or SSR-11.5-T; Wilson-Cook, Winston-Salem, North Carolina, USA). The retriever can then easily be coupled and screwed into the distal end of the stent. The stent is withdrawn leaving the guide wire in place, ensuring that a new stent can be placed.

    Plastic stents made of polyethylene tend to become softer after a few months, and so the tip of the retriever may not anchor properly in the distal end of the stent, causing failure of stent retrieval. In such cases, an 11.5-French retriever should be used for removal of a 10-French polyethylene stent.

    So far, we have used the modified stent retriever to replace occluded biliary stent in 20 procedures. In all cases, the procedure was successful and uneventful. The use of the new retriever for replacing 10-French or 11.5-French plastic stents has become an easy procedure, even for trainees.

      U. Seitz, M.D.

      Dept. of Interdisciplinary Endoscopy, University Hospital Hamburg-Eppendorf, University of Hospital

      Martinistr. 52, 20246 Hamburg, Germany

      Fax: + 49-40-428034420

      Email: seitz@uke.uni-hamburg.de

        U. Seitz, M.D.

        Dept. of Interdisciplinary Endoscopy, University Hospital Hamburg-Eppendorf, University of Hospital

        Martinistr. 52, 20246 Hamburg, Germany

        Fax: + 49-40-428034420

        Email: seitz@uke.uni-hamburg.de

        Zoom Image

        Figure 1 A 2.5 cm long 5-French curved catheter is mounted on the tip of the retriever, to facilitate the engagement of the retriever with the distal end of the stent and thereby the insertion of the guide wire into the stent.

        Zoom Image

        Figure 2 Endoscopic view showing the coupling of the retriever with the distal end of the occluded stent. The guide wire is then advanced.