A Potentially Serious Disadvantage of Multiband Device Over Conventional Ligation Using an Overtube
C. S. Leung, M.B., Ch.B.
B14N, Department of Medicine, Yan Chai Hospital
Yan Chai Street, Tsuen Wan, New Territories, Hong Kong SAR, China
Fax: + 852-2417-3307
Email: ambroseleung@hotmail.com
Publication History
Publication Date:
30 January 2003 (online)
A patient suffering from chronic alcoholic Child B cirrhosis underwent oesophagogastroduodenoscopy because of anemia by means of a Pentax EG-2730. Esophageal varices with cherry red spots were found. Multiband ligation (Saeed Multi-Band Ligator MBL-6; Wilson-Cook Medical, Winston-Salem, North Carolina, USA) was carried out. The Opti-Vu barrel with preloaded bands was attached to the tip of the endoscope, with the barrel advanced onto the tip as far as possible. The banding was performed successfully without difficulty. However, the Opti-Vu barrel was then found to be missing from the tip of the endoscope. The endoscope was inserted again, and the barrel was found in the upper end of the esophagus (Figure [1]). The barrel was finally retrieved by means of a foreign-body forceps (Figure [2]). There was no bleeding or trauma related to this detached barrel or its removal and the patient recovered uneventfully. We found later that the model we used (MBL-6) was suitable for an endoscope with an outer diameter of 9.5 - 13 mm. The endoscope we used, however, had an outer diameter of only 9 mm.
There are very few reports on complications with this new ligator. When this multiple-band ligator is used, the trigger cords that are attached to the transparent tubing are also detached when all six bands are fired. With a conventional band ligator, the string is still firmly attached to the barrel, especially after the band is fired, by further tightening of the trigger cord.
This case reminded us that multiband ligators should be properly checked for compatibility with the endoscope. It is also advisable to check for a firm fit between the barrel and the endoscope. When using multiband ligators, we should pay the same attention to checking whether the barrel is still on the tip of the endoscope after the procedure, as we do to checking the number of gauzes recovered after a laparotomy.
C. S. Leung, M.B., Ch.B.
B14N, Department of Medicine, Yan Chai Hospital
Yan Chai Street, Tsuen Wan, New Territories, Hong Kong SAR, China
Fax: + 852-2417-3307
Email: ambroseleung@hotmail.com
C. S. Leung, M.B., Ch.B.
B14N, Department of Medicine, Yan Chai Hospital
Yan Chai Street, Tsuen Wan, New Territories, Hong Kong SAR, China
Fax: + 852-2417-3307
Email: ambroseleung@hotmail.com