Rofo 2005; 177(2): 258-264
DOI: 10.1055/s-2004-813670
Medizinphysik und Technik

© Georg Thieme Verlag KG Stuttgart · New York

Vergleich von Strahlenexposition und Bildqualität eines Siremobil-IsoC3D mit einem 16-Zeilen-Spiral-CT bei Diagnostik und Intervention am humanen Becken

Comparison of Radiation Dose and Image Quality of Siremobil-IsoC3D with a 16-Slice Spiral CT for Diagnosis and Intervention in the Human Pelvic BoneG. Wieners1 , M. Pech1 , A. Beck1 , B. König2 , U. Erdmenger2 , U. Stöckle2 , P. Wust1 , R. Felix1 , R.-J Schröder1
  • 1Klinik für Strahlenheilkunde, Charité, Universitätsmedizin Berlin
  • 2Klinik für Unfallchirurgie, Charité, Universitätsmedizin Berlin
Further Information

Publication History

Publication Date:
09 November 2004 (online)

Zusammenfassung

Ziel: Vergleich der Bildqualität eines 16-Zeilen-Computertomographen mit einem Siremobil-IsoC3D am humanen Becken bei gleichzeitiger Messung der Strahlenexposition vor und nach Implantation einer Sakroiliakal-Schraube (SI-Schraube). Material und Methodik: Es wurde an acht Humanpräparaten die Beckenregion auf fünf verschiedenen Höhen durch Anfertigung von Einzelschichten mit dem Siremobil-IsoC3D untersucht. Anschließend erfolgte bei allen Präparaten eine Untersuchung des kompletten Beckens an einem 16-Zeilen-Spiral-CT im Standardprotokoll. Es erfolgte in mehreren Untersuchungsserien eine Überprüfung der Bildqualität sowohl vor und nach Implantation einer SI-Schraube als auch bei stufenweiser Reduktion des CT-Röhrenstroms, um festzustellen, wann es zu einer Angleichung der Bildqualität beider Modalitäten kommt. Bewertet wurde die Bildqualität anhand der Darstellung von Neuroforamina, Nervenwurzeln, Sakroiliakal-Fugen (SI-Fugen), Intervertebralräumen, Osteophyten, M. iliopsoas, Facies acetabuli, Hüftkopf, Fovea centralis, Hüftmuskulatur, Hüftgelenkspalt und Os pubis. Die visuelle Analyse erfolgte durch drei Radiologen und drei Unfallchirurgen anhand einer Bewertungsskala von 1 bis 5. Die Dosismessungen erfolgten für beide Modalitäten mit einem jeweils endorektal platzierten NOMEX-Dosimeter. Ergebnisse: Die durchschnittlich erreichten Bewertungsscores der jeweiligen Untersucher am Siremobil-IsoC3D lagen zwischen 3 und 4,3 Notenpunkten. Die entsprechenden CT-Scores betrugen bei einem Röhrenstrom von 250 mA zwischen 1,3 und 2,2. Die Reduktion der Stromstärke in der CT bis auf ca. 80 mA beeinflusste die Bildqualitätsbeurteilung bezüglich der analysierten Strukturen kaum. Bei Werten von 60 mA wurden die knöchernen Strukturen auch nach Implantation einer SI-Schraube als gut bewertet, die Weichteilstrukturen jedoch bereits deutlich schlechter. Das durchschnittliche Dosislängenprodukt einer Beckenuntersuchung des IsoC3D betrug 41,2 mGy × cm, die des CT bei einem Röhrenstrom von 250 mA 389 mGy × cm, von 80 mA 125 mGy × cm und von 60 mA 82 mGy × cm. Schlussfolgerungen: Das Siremobil-IsoC3D ist für diagnostische Zwecke nur eingeschränkt geeignet. Das höhere Dosislängenprodukt im CT scheint durch eine signifikant bessere Qualität und durch zu erwartende diagnostische Vorteile gerechtfertigt. Ein wesentlicher Vorteil des Siremobil-IsoC3D ist die intraoperative Verfügbarkeit bei für die Interventionskontrolle ausreichender Bildqualität mit dreidimensionaler Darstellung im Gegensatz zur konventionellen Durchleuchtung.

Abstract

Purpose: To compare the image quality of 16-slice computed tomography with the image quality of Siremobil-IsoC3D of the pelvic region and to measure simultaneously the radiation dose before and after implantation of a sacroiliac screw (SI-screw) Materials and Methods: The pelvic region of 8 human cadavers was examined in the Siremobil-IsoC3D at five different levels. We used a standard protocol for the 16-slice CT of the complete pelvic region before and after insertion of a pelvic screw, followed by stepwise reduction of the tube current to find the tube current that equalizes the image quality of both modalities. We controlled the image quality by judging important structures such as neuroforamen, nerves, sacroiliacal joint space, intervertebral space, osteophytes, iliopsoas muscle, acetabular surface, fovea centralis, hip joint and os pubis. The image quality was judged by three radiologists and three trauma surgeons using a ranking from 1 to 5. The dose was measured with an endorectally placed NOMEX Dosimeter, to obtain the gonadal dose. Results: The medium score for all viewers of the Siremobil-IsoC3D examinations was between 3 and 4.3. The medium score for all CT-examinations with a tube current of 250 mA was between 1.3 and 2.2. The reduction of tube current down to 80 mA hardly influenced the marks for the analyzed structures. Under 80 mA, bony structures, even after implantation of a SI-screw, were still marked as good, but soft tissue differentiation was getting worse. For the examination of the pelvis, the average dose-length product for the IsoC3D was 41.2 mGy × cm. The medium dose-length product for CT was 389 mGy × cm for 250 mA, 125 mGy × cm for 80 mA and 82 mGy × cm for 60 mA. Conclusion: The Siremobil-IsoC3D is sufficient for therapeutic intraoperative purpose, but the image quality is not sufficient for diagnostic purpose. The higher dose-length product of a CT examination is justifiable because of a better overview, shorter examination time and qualitative superiority. An advantage of the Siremobil-IsoC3D is the intraoperative availability with acceptable 3D image quality compared to conventional fluoroscopy.

Literatur

  • 1 Butler W, Piaggio C, Constantinou C. et al . A mobile computed tomographic scanner with intraoperative and intensive care unit applications.  Neurosurgery. 1998;  42 (6) 1304-1310
  • 2 Frodel J, Pacheco E. The use of intraoperative image-guided surgical techniques for reconstruction of orbital and zygomatic deformities.  Facial Plastic Surgery. 1999;  1511 83-89
  • 3 Starr A J. Percutaneous fixation of the columns of the acetabulum.  J Orthop Trauma. 1998;  12 51-58
  • 4 Gress H, Wolf H, Baum U. et al . Dosage reduction in computed tomography by anatomy-oriented attenuation-based tube-current modulation: first clinical results.  Fortschr Röntgenstr. 1999;  170 246-250
  • 5 Euler E, Wirth S, Linsenmaier U. et al . Vergleichende Untersuchung zur Qualität der C-Bogen-basierten 3D Bildgebung am Talus.  Unfallchirurg. 2002;  9 839-846
  • 6 Nolte L, Zamorano L, Jiang Z. et al . A novel approach to computer assisted spine surgery.  Proc 1 st Int Symp on Medical robotics and Computer assisted Surgery. 1994;  323-328
  • 7 Foley K, Smith M. Image guided spine surgery.  Neurosurg Clin N Am. 1996;  7 171-186
  • 8 Laine T, Lund T, Yilkoski M. et al . Accuracy of pedicle screw insertion with and without computer assistance: a randomised controlled clinical study in 100 consecutive patients.  Eur Spine J. 2000;  9 235-240
  • 9 Mayr E, Häuser H, Rüter A. et al . Minimalinvasive, intraoperativ Ct-gesteuerte Korrektur einer Kalkaneusosteosynthese.  Unfallchirurg. 1999;  102 239-244
  • 10 Stanley R. Use of intraoperative computed tomography during repair of orbito-zygomatic fractures.  Arch Facial Plast Surg. 1999;  1 19-24
  • 11 Taller S, Suchomel P, Lukas R. et al . CT-guided international fixation of a hangman’s fracture.  Euro Spine J. 2000;  9 393-397
  • 12 Stöckle U, König B, Dahne M. et al . Navigationsverfahren in der Becken und Azetabulumchirurgie.  Unfallchirurg 10. 2002;  105 886-892
  • 13 Coppenrath E, Schmid C, Brandl R. et al . Spiral CT des Abdomens: Gewichtsadaptierte Dosisreduktion.  Fortschr Röntgenstr. 2001;  173 52-56
  • 14 Scheck R, Coppenrath E, Bäuml A. et al . Radiation dose and image quality in spiral computed tomography: multicenter evaluation at six institutions.  BrJ Radiol. 1998;  71 734-744
  • 15 Weidner A, Wahler M, Chiu S. et al . Modification of C1-C2 transarticular screw fixation by image-guided surgery.  Spine. 2000;  25 2668-2673
  • 16 Ebraheim N, Savolaine E, Hoeflinger M. et al . Radiological diagnosis of screw penetration of the hip joint in a acetabular fracture reconstruction.  J Orthop Trauma. 1989;  3 196-201
  • 17 Wedegärtner U, Gatzka C, Rueger J M. et al . Multislice CT (MSCT) in der Detektion und Klassifikation von Becken- und Azetabulumfrakturen.  Fortschr Röntgenstr. 2003;  175 105-111

Dr. Gero Wieners

Klinik für Strahlenheilkunde, Charité, Universitätsmedizin Berlin, Campus Virchow Klinikum

Augustenburger Platz 1

13353 Berlin

Phone: ++ 49/30/4 50 55 70 02

Fax: ++ 49/30/4 50 55 79 01

Email: gero.wieners@charite.de