Klin Monbl Augenheilkd 2006; 223(11): 899-903
DOI: 10.1055/s-2006-927094
Klinische Studie

© Georg Thieme Verlag KG Stuttgart · New York

Augeninnendruckmessungen mit dem Selbsttonometer Proview® im Vergleich zur Goldmann-Applanationstonometrie bei Gesunden und Glaukompatienten

Intraocular Pressure Measurements with the ProviewTM Self-Tonometer in Comparison of Goldmann Applanation Tonometry in Healthy and Glaucomatous EyesM. W. Meyer1 , V. Yevelenko2 , L. Hoy3 , A. Meyer1 , C. Erb1, 4
  • 1Augenklinik der Medizinischen Hochschule Hannover (Direktor: Prof. Dr. med. R. Winter)
  • 2Augenklinik St.-Joseph-Stift, Bremen (Chefarzt: Dr. med. A. Mohr)
  • 3Institut für Biometrie der Medizinischen Hochschule Hannover (Direktor: Prof. Dr. med. H. Hecker)
  • 4Schlosspark-Klinik, Berlin (Chefarzt: PD Dr. med. C. Erb)
Further Information

Publication History

Eingegangen: 14.9.2005

Angenommen: 7.6.2006

Publication Date:
28 November 2006 (online)

Zusammenfassung

Hintergrund: Die Druckphosphen-Tonometrie (DPT) ist eine psychophysische Methode, die auf der entoptischen Wahrnehmung von Druckphosphenen bei einer Erhöhung des Augeninnendrucks basiert. Die Messmethode des Selbsttonometers Proview® basiert auf diesem Prinzip. In der vorliegenden Studie wurde die Zuverlässigkeit und Praxistauglichkeit dieses Tonometers im Vergleich zum Goldmann-Applanationstonometer (GAT) untersucht. Patienten und Methoden: Es wurden insgesamt 150 Augen von 62 gesunden Probanden und 88 Patienten mit chronischem Offenwinkelglaukom oder okulärer Hypertension in die Studie eingeschlossen. Nach einer Übungsphase erfolgte zunächst eine Messung des Augeninnendrucks mit dem GAT. Danach wurde von den Probanden eine Selbsttonometrie mit dem Proview® durchgeführt. Dabei erfolgten alle Messungen in sitzender Position. Ergebnisse: Für alle untersuchten Augen (n = 150) fielen die Messwerte mit dem Proview® durchschnittlich 5,5 mmHg höher aus als die mit dem GAT gemessenen Augeninnendruckwerte. Lediglich 34 % aller Messungen mit den beiden Verfahren lagen in einem Differenzbereich von ± 3 mmHg. Ein Vergleich zwischen Glaukompatienten mit nachgewiesenen Gesichtsfeldausfällen und gesunden Probanden bzw. Patienten mit okulärer Hypertension ohne Vorliegen von Gesichtsfelddefekten ergab nahezu identische mittlere Differenzwerte zwischen dem Proview® und dem GAT. In beiden Gruppen wurden mit dem Selbsttonometer im Durchschnitt ungefähr 5,0 mmHg höhere Tensionswerte ermittelt. Schlussfolgerungen: Die Messwerte des Proview®-Tonometers zeigten im Vergleich zum GAT nur eine geringe Übereinstimmung, sodass eine Selbsttonometrie, die auf der Basis einer subjektiven Wahrnehmung von Druckphosphenen beruht, in der Routinediagnostik nicht empfohlen werden kann. Beginnende Gesichtsfelddefekte scheinen keinen Einfluss auf die Augeninnendruckmessung mit dem Selbsttonometer zu haben.

Abstract

Background: The aim of this study was to compare the results of intraocular pressure (IOP) measurements obtained using the pressure phosphene tonometer Proview™ with those from Goldmann applanation tonometer (GAT) in normal and glaucomatous eyes. Patients and Methods: The IOP in 150 eyes of 62 healthy volunteers and 88 patients with glaucoma or ocular hypertension was measured prospectively in a sitting position. After being trained to use the Proview™ device, Goldmann applanation tonometry was performed first. Then the patient took a reading with the Proview™ self-tonometer. Results: For all investigated eyes the measurements with the Proview™ were on average 5.5 mmHg higher than those by GAT. Only 34 % of the readings from the two devices were within a difference range of ± 3 mmHg. On comparing the group of glaucomatous patients with slight visual field defects with the group of healthy subjects and patients with ocular hypertension without visual field defects we determined almost the same mean difference between the Proview™ and GAT (mean difference in the group with visual field defects = 4.7 ± 4.1 mmHg; without defects = 4.8 ± 2.9 mmHg). Conclusions: The tonometer Proview™ did not show a close agreement to GAT. Therefore, the PPT does not offer an alternative method for measuring IOP. We do not recommend the Proview™ for self-tonometry at home or for clinical management of patients with glaucoma. Slight visual field defects seem to have no influence on intraocular pressure measurement with the self-tonometer.

Literatur

  • 1 Alvarez T L. et al . The Proview phosphene tonometer fails to measure ocularpressure accurately in clinical practice.  Ophthalmology. 2004;  112 (5) 1077-1085
  • 2 Aristotle. De Anima; with translation, introduction and notes by Hicks RD. Amsterdam; Hakkert 1965: 77-83
  • 3 Brigatti L, Maguluri S. Reproducibility of self-measured intraocular pressure with the phosphene tonometer in patients with ocular hypertension and early to advanced glaucoma.  J Glaucoma. 2005;  14 (1) 36-39
  • 4 Danesh-Meyer H V. et al . Comparison of the Proview pressure phosphene tonometer performed by the patient and examiner with the Goldmann applanation tonometer.  Clin Experiment Ophthalmol. 2004;  32 (1) 29-32
  • 5 Draeger J. Simple hand applanation tonometer for use on the seated as well as on the supine patient.  Am J Ophthalmol. 1966;  62 (6) 1208-1210
  • 6 Draeger J. Intraocular pressure self-monitoring in glaucoma patients.  Klin Monatsbl Augenheilkd. 1998;  213 (1) aA9
  • 7 Fresco B B. A new tonometer-the pressure phosphene tonometer. Clinical comparison with Goldmann tonometry.  Ophthalmology. 1998;  105 2123-2126
  • 8 Goldmann H. Un nouveau tonomètre à applanation.  Bull Soc Franc Ophthalmol. 1955;  67 474-477
  • 9 Goldmann H. Applanation tonometry.  Ophthalmologica. 1957;  134 (4) 221-242
  • 10 Kingman S. Glaucoma is second leading cause of blindness globally.  Bull Health World Organ. 2004;  82 (11) 887-888
  • 11 Le A. et al . Risk factors associated with the incidence of open-angle glaucoma: the visual impairment project.  Invest Ophthalmol Vis Sci. 2003;  44 (9) 3783-3789
  • 12 Leydhecker W. The practical value of tonometry, gonioscopy & tonography.  Klin Monatsbl Augenheilkd. 1957;  130 (5) 587-600
  • 13 Li J. et al . Clinical comparison of the Proview eye pressure monitor with the Goldmann applanation tonometer and the Tonopen.  Arch Ophthalmol. 2004;  122 (8) 1117-1121
  • 14 Meyer M W. et al . Comparison of intraocular pressure measurements with the digital tonometer TGDc-01 ‘PRA’ and the Goldmann applanation tonometer.  Ophthalmic Res. 2004;  36 (5) 250-254
  • 15 Moreno M J, Sadaba L M, Heras H. Intraocular pressure measurement using the Proview tonometer.  Arch Soc Esp Oftalmol. 2005;  80 (2) 79-83
  • 16 Perkins E S. Hand-held applanation tonometer.  Br J Ophthalmol. 1965;  49 (11) 591-593
  • 17 Rai S. et al . Level of disagreement between Proview™ phosphene tonometer and Goldmann applanation tonometer intraocluar pressure readings.  J Glaucoma. 2005;  14 (2) 120-123
  • 18 Sampaolesi R. Diurnal variation of intraocular pressure in healthy, suspected and glaucomatous eyes.  Bibl Ophthalmol. 1968;  74 1-23
  • 19 Tai M C. et al . Clinical evaluation of the intraocular pressure in patients with glaucoma or ocular hypertension by a self-assessable tonometer.  J Ocul Pharmacol Ther. 2005;  21 (1) 55-61
  • 20 Thylefors B, Négrel A D. The global impact of glaucoma.  Bull World Health Organ. 1994;  72 (3) 323-326
  • 21 Von Purkinje J E. Subjektive Erscheinungen nach Wirkung der Sinne.  Zur Physiologie der Sinne. 1825;  134
  • 22 Voogd S D. et al . Incidence of open-angle glaucoma in a general elderly population. The Rotterdam Study.  Ophthalmology. 2005;  112 (9) 1487-1493
  • 23 Whitacre M, Stein R. Sources of error with use of Goldmann-type tonometers.  Surv Ophthalmol. 1993;  38 1-30