Evaluation of a new flexible fiber CO2 laser for gastrointestinal cutting: NOTES and mucosectomy in a porcine model
S. AnandasabapathyMD
Unit 436
Department of Gastroenterology, Hepatology and
Nutrition
The University of Texas, M.D. Anderson Cancer Center
Houston
TX 77030
USA
Fax: +1-713-563-4398
Email: sananda@mdanderson.org
Publication History
Publication Date:
15 August 2008 (online)
The CO2 laser is one of the most widely used surgical tools in otolaryngology, offering precise cutting, hemostasis, predictable penetration depth, and minimal collateral damage, due to efficient absorption of CO2 laser energy by tissue water. Endoscopic use was previously limited by lack of an efficient, flexible delivery system. In 2002, Fink et al. described a hollow-core photonic band-gap fiber created with multilayered dielectric mirrors [1]. The fiber core acts as a channel for both helium flow (3 l/minute) and beam delivery, and allows improved power output. Although the precision and safety of the device have been described in the larynx and airway [2], its performance in the gastrointestinal tract has not been described. In a nonsurvival porcine model, we evaluated the performance of the flexible-fiber CO2 laser in two novel endoscopic applications: mucosectomy and natural orifice transluminal surgery (NOTES). Indeed, the hemostatic abilities of the laser and the established safety of CO2 in the peritoneal cavity make it ideal for these applications.
Four Yorkshire pigs underwent general anesthesia and gastroscopy with a double-channel endoscope. Using a 1.8 m fiber at 20 W pulsed current, 1.5 cm incisions were made along the greater curvature in two pigs. No blood was noted at the incision sites and the peritoneal cavity was accessed. In the first pig, biopsies were taken of the liver and spleen using standard forceps. Bleeding at the splenic biopsy site was successfully cauterized using the CO2 laser ([Fig. 1]). In the second pig, grasping forceps were used to pull a loop of jejunum into the stomach. The loop was then anchored to the interior gastric wall with clips, and an incision made to create a gastrojejunostomy ([Fig. 2]).
The remaining two pigs underwent endoscopy and free-hand mucosectomy in the esophagus and stomach after injection of submucosal saline. No bleeding was noted. A 2 cm segment of mucosa was removed from each site. Histopathology confirmed removal of the mucosa and part of the submucosa with minimal damage to the muscularis propria ([Fig. 3]).
#Acknowledgment
Funding for this study was provided by Omni-Guide, Inc. Endoscopic equipment and technical support were provided by the Pentax Corporation and Cook Endoscopy, Inc.
Endoscopy_UCTN_Code_TTT_1AO_2AG
Endoscopy_UCTN_Code_TTT_1AO_2AD
#References
- 1 Temelkuran B, Hart S D, Benoit G. et al . Wavelength-scalable hollow optical fibres with large photonic bandgaps for CO2 laser transmission. Nature. 2002; 420 650-653
- 2 Holsinger F C, Prichard C N, Shapira G. et al . Use of the photonic band gap fiber assembly CO2 laser system in head and neck surgical oncology. Laryngoscope. 2006; 116 1288-1290
S. AnandasabapathyMD
Unit 436
Department of Gastroenterology, Hepatology and
Nutrition
The University of Texas, M.D. Anderson Cancer Center
Houston
TX 77030
USA
Fax: +1-713-563-4398
Email: sananda@mdanderson.org
References
- 1 Temelkuran B, Hart S D, Benoit G. et al . Wavelength-scalable hollow optical fibres with large photonic bandgaps for CO2 laser transmission. Nature. 2002; 420 650-653
- 2 Holsinger F C, Prichard C N, Shapira G. et al . Use of the photonic band gap fiber assembly CO2 laser system in head and neck surgical oncology. Laryngoscope. 2006; 116 1288-1290
S. AnandasabapathyMD
Unit 436
Department of Gastroenterology, Hepatology and
Nutrition
The University of Texas, M.D. Anderson Cancer Center
Houston
TX 77030
USA
Fax: +1-713-563-4398
Email: sananda@mdanderson.org