The neonatal period has the greatest incidence of seizures in life, with 1.8-3.5 per
1,000 live births[1]. Seizures in the newborn are associated with high morbidity and mortality, which
make their detection and treatment essential[2],[3].
Seizure activity in neonates is often difficult to observe, making the detection of
seizures particularly challenging. Clinical observation alone can lead to underdiagnosis
of neonatal seizures, as nearly 80% of seizures can be occult[4]. It is for these reasons that effective methods for seizure detection are of fundamental
importance in neonatal care.
Electrophysiological brain activity, as measured by electroencephalography (EEG),
is well established as a tool for providing information regarding the functional and
metabolic state of the brain and the occurrence of epileptic seizure episodes[5]. In neonatal care, EEG has been used extensively for estimation of the degree of
cerebral maturation in preterm infants and for detection of abnormal patterns indicating
focal and global cerebral lesions[6],[7],[8]. In the neonatal setting, as well as in intensive care in general, the EEG is most
often recorded for a patient intermittently, at best serially and, on rare occasions
only, continuously[9],[10]. A limiting disadvantage with intermittent conventional EEG during neonatal care
is the difficulty in discriminating emerging trends of development of the electrocerebral
activity over hours and days. This interpretation, if possible at all, requires specialized
skills not usually available in the neonatal intensive care unit (NICU)[5]. Thus, the main disadvantage with intermittent conventional EEG during neonatal
care is the inability to diagnose seizures when they occur.
Amplitude-integrated EEG (aEEG) is a technique for simplified EEG monitoring that
has found an increasing clinical application in neonatal intensive care. Its main
value lies in allowing real-time detection of electrographic seizures, providing the
opportunity for treatment at the time they occur, and not requiring specialized staff
for their interpretation.
Several studies have evaluated the sensitivity and specificity of aEEG for the diagnosis
of neonatal seizures with variable results.[11]
–
[23] However, these studies have not addressed the importance of electrophysiological
interpretation of aEEG findings.
One advantage of aEEG over conventional EEG is that the former does not require specialized
staff for their interpretation. Nevertheless, for its correct interpretation, and
to avoid misdiagnosing rhythmic artifacts as seizures, it is necessary for members
of the care team to recognize the electrophysiological ictal patterns in the conventional
EEG trace available in current devices. Many false positives in the aEEG pattern can
be easily recognized when reviewing the raw trace[23].
Despite the importance of the recognition of these patterns during the use of aEEG,
electrophysiological diagnosis of seizures is seldom emphasized. Previous papers have
not explored the electrophysiological basis of seizure detection using current aEEG
devices. The aim of this paper is to discuss the electrophysiological basis of the
differentiation of epileptic seizures and extracranial artifacts to avoid misdiagnosis
with aEEG devices.
ELECTROCLINICAL ASPECTS OF SEIZURES IN THE NEONATAL PERIOD
The electrographic and clinical characteristics of seizures in the neonate are unique
compared with older children and adults[24]. In the neonate, electrographic seizure patterns vary widely, electrical seizure
activity does not accompany all behaviors considered to be seizures, and electrographic
seizures frequently occur without evident clinical seizures[25]
–
[27].
From the perspective of the neonatal EEG, neonatal seizures can be classified according
to the temporal relationship between the electrical event and the clinical event:
electroclinical seizures (a clinical seizure with ictal electrographic correlation),
clinical-only seizures (clinical seizure without concurrent electrographic correlate),
and electrographic seizures (ictal electrographic abnormalities without concurrent
clinical seizures)[28],[29].
Electroclinical seizures are characterized by a temporal overlap between clinical
seizures and electrical seizure activity on the EEG. In many instances, the electrical
seizure and clinical events are closely associated, with the onset and termination
of both events coinciding[25].
Focal clonic, focal tonic, and some myoclonic seizures and spasms are associated with
electrical seizure activity. These are epileptic in origin[30],[31].
Electrographic seizures (electrical seizure activity with no clinical accompaniment)
are very frequent in the newborn. Typically, no behavioral changes are associated
with seizure discharges[4],[32]. One frequent cause of electrographic seizures is the use of antiepileptic drugs[28]. The antiepileptic drugs may suppress the clinical component of the electroclinical
seizure but not the electrical component, thus the clinical seizure may be controlled
but electrical seizure activity may persist. However, electrographic seizures are
frequent in the neonatal period even without antiepileptic treatment. These seizures
will be undiagnosed without the use of EEG monitoring[33].
Some types of clinical seizures have no specific relation to electrical seizure activity[34]. Those that occur in the absence of any electrical seizure activity include generalized
tonic, motor automatisms, and some myoclonic seizures. These clinical events are initiated
and elaborated by nonepileptic mechanisms[35]. When these nonepileptic clinical episodes are erroneously diagnosed as epileptic
seizures, there is a risk of unnecessarily treating newborns with antiepileptic drugs
that are not required.
AMPLITUDE-INTEGRATED ELECTROENCEPHALOGRAPHY
General information
Amplitude-integrated EEG was first designed by Maynard as the cerebral function monitor
in the 1960s and originally applied in adult intensive care by Prior[36]. The method was first used in newborn babies in the late 1970s and early 1980s[37].
This method is based on a time-compressed (usually 6 cm/hour) semi-logarithmic (linear
0-10 mV, logarithmic 10-100 mV) display of the peak-to-peak amplitude values of EEG
passed through an asymmetrical bandpass filter that strongly enhances higher over
lower frequencies, with suppression of activity below 2 Hz and above 15 Hz. This approach
minimizes artifacts from sources such as sweating, movement, muscle activity, and
interference but may obscure some seizures. The current devices consist mainly of
two independent channels of aEEG and their corresponding two-channel conventional
EEGs available on the screen for simultaneous viewing.
The EEG trace of the neonate is characterized by intermittent bursts of high amplitude,
intermixed with lower amplitude continuous activity. The bandwidth of the aEEG curve
reflects variations in minimum and maximum EEG amplitude. The semi-logarithmic display
enhances identification of changes in the low voltage range and avoids overloading
the display at high amplitudes[5].
The aEEG technique extracts the cerebral activity and transforms it into five basal
activity patterns (interictal patterns) and one ictal (seizures) pattern. The background
pattern is defined as continuous (normal pattern with sleep-wake cycle), discontinuous
(moderate abnormal in full-term newborns), burst suppression (pathologic or drug induced),
low voltage and flat/isoelectric ([Figure 1 A-D]). This system is applicable to newborns of all ages and diagnoses[38]. An ictal pattern (seizures) is usually seen as an abrupt increase of maximal and
minimal aEEG amplitude, seldom only the minimal amplitude, often followed by a transient
postictal amplitude depression ([Figure 1 E])[5]. The aEEG devices show the conventional EEG trace simultaneously with the aEEG trace
to confirm the aEEG findings.
Figure 1 (A) Continuous background pattern, with prominent sleep-wake cycle: upper-margin
voltage is > 10 mV and lower margin voltage is > 5 mV (arrow). The sleep-wake cycle
is characterized by smooth cyclic variations, mainly of minimum amplitude, with periods
of broader bandwidth that represent quiet sleep (circle) and periods of narrow bandwidth
that correspond to wakefulness or active sleep (arrowhead). (B) Discontinuous background
pattern: upper margin is > 10 mV and lower margin is < 5 mV. The sleep- wake cycle
is not present. (C) Burst suppression pattern: upper and lower margin voltages are
< 10 mV and <5 mV respectively reflecting voltage suppression, vertical lines (arrow)
show periods of burst. Conventional EEG row shows a brief burst of cerebral activity
(circle) between periods of voltage suppression. (D) Isoelectric or flat tracing:
both margins are < 5 mV and prominent spikes are likely due to patient movement. (E)
Ictal pattern: seizures are shown as a sudden rise in lower and upper margin (circle).
The discovery that information displayed by the aEEG paradigm can predict outcome
after perinatal asphyxia has attracted medical attention and been the subject of numerous
publications each year. Currently, aEEG is being used as an inclusion criterion in
studies of therapeutic hypothermia[39]
–
[45]. However, an often-ignored value of aEEG is its utility for real-time detection
of seizures.
There is controversy regarding the sensitivity and specificity of aEEG for electrographic
detection of seizures, especially in comparison with conventional EEG[11]
–
[23]. But even with less sensitivity and specificity, the aEEG has numerous advantages:
it allows for continuous monitoring over long periods of time; the simplicity of its
interpretation does not require extensive training; and its use permits immediate
detection and treatment of seizures.
Although aEEG is easily interpreted by personnel who have not trained in electrophysiology,
familiarity with some relatively simple electrophysiological characteristics is necessary
for the correct diagnosis of seizures.
The electrophysiological diagnosis of epileptic seizures and its differentiation with
extracranial artifacts
It is a challenge for all involved in neonatal care to accurately diagnose neonatal
epileptic seizures, as nearly 80% of them are subclinical[46],[47]. The use of EEG monitoring permits detection of seizures, but there are many artifacts
visible in both EEG and aEEG traces that can be misdiagnosed as seizures. Differentiating
a nonepileptic rhythmic pattern from a seizure in the aEEG trace can also be challenging
because both have a similar appearance: a sudden rise in minimal and maximal amplitude.
To differentiate between them, it is necessary to review the conventional EEG tracing
and to know the electrophysiological characteristics of epileptic seizures.
The hallmark of an electrographic seizure, visible in conventional EEG, is the sudden
appearance of repetitive discharge events consisting of a definite beginning, middle
and end, that evolve in frequency, morphological appearance and amplitude ([Figure 2])[10].
Figure 2 Conventional EEG of a newborn with an electrographic seizure. The beginning of a
rhythmic repetitive activity in left occipital area (01) is marked with the arrow
in [Figure 2A]. The rhythmic discharge evolves in morphology, amplitude and frequency (A and B).
[Figure 2B] shows diffusion to surrounding areas (circle) and the end of the seizure (arrow).
This pattern should be always reviewed in the conventional EEG trace available in
current devices ([Figure 3]). Infrequently, the ictal pattern does not include evolution and instead manifests
as regular repetitive spikes or regular rhythmic slowing.
Figure 3 A suspected seizure in the aEEG (circle in A) is reviewed. Each figure (A-E) shows
three hours of aEEG trace in the upper row and the corresponding 15 seconds of conventional
EEG in the lower row, both traces available simultaneously in current devices. The
vertical line in the aEEG shows the part of the trace displayed in the conventional
EEG. A repetitive high amplitude slow wave with a rhythmic pattern is noted in the
beginning of the seizure (A). In B, C, D and E the ictal pattern evolves progressively
in morphology, frequency and amplitude. F shows the end of the seizure (marked with
the arrow in E). The spikes are of lower amplitude and lower frequency and at the
end of the seizure, voltage flattening is observed (arrow). This figure shows the
epileptic seizure typically evolving in frequency, amplitude and morphology.
The ictal pattern recorded from the scalp at the start of the seizure can take different
forms. The predominant frequency in a given seizure can be in the delta (0-3 Hz),
theta (4-7 Hz), alpha (8-13 Hz) or beta (14-30 Hz) frequencies, or be a mixture of
these, with different morphology: spikes, sharp waves, slow waves, spike and waves,
electrodecremental activity (a diffuse flattening of brain rhythms), then evolving
in frequency, amplitude and/or morphology[30],[40],[48]
–
[50]. The voltages of the activity may also vary, from extremely low (usually when faster
frequencies are present) to very high (commonly seen when slow frequencies are present)
and evolve within the seizure[32].
The evolving pattern helps us to distinguish electrographic seizures from the wide
range of nonepileptic paroxysmal behavior that may occur in healthy or sick infants,
especially from extracerebral artifacts produced by the neonate or neonatal care and
procedures[51]. Some of these extracerebral artifacts are rhythmic, so they may be confused with
ictal activity. However, these artifacts are typically monomorphic and do not evolve
in frequency, morphological appearance or amplitude ([Figure 4]).
Figure 4 Each figure shows three hours of aEEG (upper row) with the corresponding 15 seconds
of conventional EEG in the lower row (arrowhead A marks one hour of trace). The vertical
line in the aEEG shows the portion of the trace displayed in the conventional EEG.
A sudden rise in lower margin is noted and reviewed (circle in A). In the conventional
EEG, the pattern is monorhythmic all along the suspected ictal trace with the same
frequency, amplitude and morphology at the beginning (A); 60 minutes later (B); and
120 minutes later at the end of the event (C) refuting the ictal origin of the event.
Other characteristics that help to differentiate between a seizure and a rhythmic
artifact are the electric field and the spread of the seizure to other areas of the
brain, both visible only in the case of epileptic seizures ([Figure 1]). The cerebral activity should have an electric field: it should be reflected in
physically-adjacent electrodes with less amplitude with respect to the electrode closest
to the foci, and perhaps in synaptically-linked regions such as the contralateral
hemisphere[24],[52]. The electric field will not be visible in current aEEG devices because it has only
two independent (not chained) bipolar EEG channels, and the spread of the seizure
will be evidenced only in the case of propagation to the contralateral hemisphere,
where it will be recorded by the other (contralateral) EEG channel. However, the observation
of the evolving pattern is sufficient to ensure the diagnosis of epileptic seizures
in most cases.
The role of aEEG in the neonatal intensive care unit
The rationale for detecting electrographic seizures rests on the assumption that detection
and treatment will ultimately lead to improvement. It is currently unknown whether
electrographic seizures independently damage the brain or whether they are mere biomarkers
of an underlying brain injury, but evidence from animal studies suggests that seizures
may alter brain development and lead to long-term deficits in learning, memory, and
behavior[53],[54]. A growing body of literature demonstrates that the electrographic seizure burden
is independently associated with worse outcomes, and suggests that electrographic
seizures independently contribute to brain damage[54],[55]. It is well known that a significant association exists between seizure duration
and severity of brain injury found on cerebral magnetic resonance images in newborns
with hypoxic-ischemic encephalopathy[55].
If we assume that electrographic seizures are damaging the brain and anti-epileptic
drugs stop seizures and improve outcomes, then real-time electrographic seizure detection
is critical.
The aEEG performance depends on the degree of expertise and familiarity with aEEG
reading, but is easier than conventional EEG and allows bedside seizure detection
by practitioners with limited training in neurophysiology. This feature makes it the
ideal resource for continuous monitoring within the NICU, even above the gold-standard
conventional video EEG, which requires highly-specialized personnel, such that it
is impractical for continuous monitoring in real time.
More studies are needed comparing both techniques (conventional EEG versus aEEG),
strengths and weaknesses, reliability and reproducibility of the aEEG for the diagnosis
of neonatal seizures, as well as the improvement of outcome of newborns with the use
of aEEG to determine the real impact of its use in the NICU. But there is no doubt
that those who use aEEG should know the basic principles of the electrophysiological
diagnosis of seizures to optimize the correct diagnosis of these with aEEG.
BEST PRACTICE IN NEONATAL SEIZURE DETECTION WITH THE USE OF AEEG
For correct interpretation, the whole seizure observed in the aEEG trace must be reviewed
in the conventional EEG row to confirm the evolving pattern ([Figure 3]). Most rhythmic artifacts do not show the typical evolving pattern and are thus
easily identifiable in the conventional EEG but easily mistaken in the aEEG ([Figure 4]). Therefore, while the aEEG is useful to quickly visualize the possible ictal episode;
concurrent use of the conventional EEG will allow confirmation of the epileptic or
nonepileptic origin of the event.
Reviewing the conventional EEG row for each critical episode observed in the aEEG
trace optimizes our recognition of seizures such that we can avoid misdiagnosis of
artifacts and erroneous treatment of the nonepileptic neonate with antiepileptic drugs
that may have neurotoxic effects[56].
CONCLUSION
The main value of aEEG is that it permits real-time recognition of electrographic
seizures without the need for interpretation by specialized providers.
The hallmark of an electrographic seizure is the sudden appearance of repetitive discharge
events consisting of a definite beginning, middle and end, that evolve in frequency,
morphological appearance, and amplitude. Rhythmic extracerebral artifacts are typically
monomorphic and do not evolve in frequency, morphological appearance, or amplitude.
For proper interpretation of the aEEG, and to avoid misdiagnosing an artifact as a
seizure, it is necessary to recognize the ictal pattern in the conventional EEG tracing.