Keywords prosthesis and implants - hearing loss - otitis externa
Palavras-chave próteses e omplantes - perda auditiva - otite externa
Introduction
The osseointegrated implants were firstly introduced to clinical practice in Sweden
in the 1970's, and they have ever since been widely accepted in such areas as dentistry,
maxillofacial surgeries, and reconstructing and orthopedic surgeries[1 ]. The Bone Anchored Hearing Aid (BAHA) system, which was originally used by Tjellstrom and his team in Sweden in
1977[2 ], made the prospect of osseointegrated implants renowned in the area of otology.
BAHA system is comprised of two major parts: a titanium screw with an abutment implanted
in the cortical part of the mastoid bone and one external unit named processor, which
is connected to the abutment. The processor is intended to receive the environmental
sounds and convert them into a mechanical energy, what is turned into vibration and
transmitted to the abutment, which, in turn, shall stimulate the cortical part of
the temporal bone. This vibration is absorbed by the skull and directly stimulates
the cochleas without involving the air hearing conduction, i.e., the external acoustic
meatus and the middle ear. In some special occasions, such as surgeries on small children,
the surgery can be performed in two different times: firstly, by placing the titanium
screw and, secondly, by coupling the abutment with this osseointegrated screw.
The titanium screw is surgically implemented, and this procedure can be performed
by way of a local anesthesia. The processor shall be inserted only after the osseointegration
period, which is three months for adults and six months for children. The processor
is easily connected and disconnected by the patient him/herself.
Although it has been long used in Europe and the United States, in Brazil BAHA system
has been recently used, and it has a scarce national publication and only a few surgical
cases. The team at the Audiological Diagnostic Center of Iguaçu Hospital (CDAH) in
Curitiba has been the first one to implant BAHA after it has been authorized by the
National Agency for Sanitary Surveillance.
In the meantime, the team faced some difficulties with respect to methodology of audiological
evaluation of the different hearing loss cases on BAHA candidates. This fact led to
understanding that it is urgent to develop a protocol to audiologically evaluate and
follow-up with patients who are candidates to the procedure.
The present study intends to show the first case of BAHA performed in Brazil, as duly
approved by ANVISA, having the surgery been performed by Hospital Iguaçu's team, as
well as to describe the protocol used during the service. The research was approved
by the Ethical Committee of Tuiuti University of Parana under the number CEP-UTP 047/09,
and the patient has signed a term of free and clarified consent, allowing the audiological
findings and images presented herein to be divulged.
Literature's Revision
BAHA system is indicated to patients with conductive and mixed hearing loss. The average
bone conduction in frequencies of 0.5, 1, 2 and 3 kHz must be equal to or above 45
cBNA for the abutment-coupled processor, or as high as 65 dBNA for the box processor.
A more recent indication is for patients with unilateral sensorineural hearing loss
or Single Sided Deafness (SSD), presenting contralateral ear with an average bone conduction equal to or above
20 dBNA[3 ].
Conductive and mixed hearing losses are highly prevailing disorders that can be either
treated with surgical techniques or rehabilitated by an individual sound-amplifier
apparatus (ISAA). There is, however, a subgroup of patients who are not entitled to
a surgical treatment or utilize ISAA by a number of reasons, some of which are agenesis
or stenosis of external acoustic meatus, rebel causes with regard to medical and/or
surgical treatments of chronic otitis media, chronic otitis externa and open cavity
mastoidectomy, among others.
The crown prostheses of bone conduction, mostly indicated for this population, show
such hassles as skin irritation due to unstopping pressure in the support place, extremely
bad esthetics, and difficulty in maintaining the crown well-positioned in children,
for it is easily removable[4 ]. BAHA can overcome all these hassles because it does not place any pressure on skin,
but it is discreet and esthetically well-accepted instead, besides being safely tightened
to the implant, always in the right place of stimulation[5 ].
BAHA system presents some outstanding results in cases of conductive and mixed hearing
disorders, despite having some restraints on the gain provided to the sensorial component
of mixed hearing loss[6 ].
Case Report
VLVM, 50-year-old female, presented a moderate left-ear and severe right-ear mixed
hearing loss, both of which with ascending configurations and progressive characteristics,
caused by bilateral osteosclerosis. The patient appeared at CDAHI after being submitted
to three surgeries of stapedotomy on the right ear and one surgery on the left one,
all of which were unsuccessful, and subsequent worsening of the hearing thresholds.
She mentioned a continual tinnitus in the right ear, which was more noticeable during
silence, deemed to have a medium impact according to the visual analogue scale (VAS)
and vertigo when placing and using ISAA.
She did not actually use ISAA, due to showing chronic otitis externa associated with
otorrhea and otalgia.
Pre-Surgical Evaluation
In the pre-surgical evaluation, medical consultation, audiological and psychological
evaluation, and an evaluation protocol for tinnitus were performed, consisting of:
visual analogue scale, THI (Tinnitus Handicap Inventory ), Tyler Activities (characteristics related to concentration, emotions and hearing)
and Tyler Handicap (factor 1 = social, emotional and behavioral, factor 2 = hearing
and factor 3 = prospect of life).
After the first otorhinolaryngological consultation, in which otoscopy clearly showed
otitis externa, edema and purulent secretion in the external acoustic meatus, caused
by ISAA, and normal tympanic membranes, it has been indicated an ear tomography and
an audiological evaluation composed of pure-tone threshold audiometry ([Figure 1 ]), logoaudiometry, immitance audiometry, brainstem auditory evoked potential test
(AEP), vectoelectronystagmograph and an open-field audiometry. The latter was performed
in three test situations: 1) patient without amplifiers; 2) by using the BAHA tester
that is available jointly with the system Kit; 3) with the hearing prosthesis used
by the patient.
Figure 1. Pre-surgical pure-tone threshold audiometry – Better bone conduction in the left
ear, side preferable to use BAHA, according to the patient.
The image test found permeable cochleas and a medium metallic prosthesis in the right
vestibule.
The evaluation with BAHA tester was performed with the arch-coupled processor, providing
the patient with a hearing sensitivity caused by the processor vibration[3 ]. By way of an open-field audiometry, pure-tones thresholds issued by a lateral acoustic
box were evaluated, and a 60-dBNA monosyllable word-distinction test was performed
with a concurring 55-dBNA speech noise before the patient. An external test has also been performed, in which the patient
stayed with the tester for one day, using it on the right side during the morning
and on the left side during the afternoon.
After the patient's exams were analyzed, she has been indicated to use BAHA Cochlear
Divine on her left ear, which is the ear with best bone conduction thresholds and
the side preferred by the patient during the external test.
The evaluation results were explained and the patient was guided with respect to the
risks and benefits of the surgery. The patient signed both a term of consent and an
authorization for the images to be used in this work.
The surgery
The surgery was performed on February 28, 2008 by a one-time method, in which the
screw and the abutment are connected and placed in the same surgical time. A local
anesthesia was performed without any intercurrence.
Description of surgical technique
Patient under intravenous sedation, in dorsal decubitus, with head lateralization;
Trichotomy and local antisepsis;
Local infiltration with adrenaline and lidocaine solution 1:100000;
Previous location of the place where the implant shall be fixed, at 5–5.5 cm of the
external acoustic meatus over a line with a 45° angle to the horizontal axis of CAE
and skin marking with a surgical pen ([Figure 2 ]);
Drawing of a quadrangular flap with anterior pedicle deepening in the subcutaneous,
however without removing periosteum;
Incision in the central area of the periosteum and lateral dilatation of this periosteum;
Drilling with a 3-mm-deep drill bit, perpendicularly to the bone and irrigated by
saline solution;
Verification of the depth of drilling and likely contact with dura mater and, in case
there is still a texture of bone, drilling with a 4-mm-drill bit is performed;
Drilling with a drill 'countersink' to stretch the drilling onto the accurate diameter
of implant ([Figure 3 ]);
Scraping with a bistoury the pillar flap follicles and narrowing both the subcutaneous
flap with dermatomes and the areas adjoining the flap by using an ordinary bistoury;
Fixation of the titanium implant;
Flap repositioning by drilling its central area to exteriorize the external implant
area ([Figure 4 ]);
Flap suture with mononylon 4.0;
Curative with a silicon button and a petrolatum gauze;
External curative with a band.
Figure 2. Previous localization of the site where implant shall be installed.
Figure 3. Bone drilling to couple the abutment.
Figure 4. Flap repositioning by drilling its central area to exteriorize the external implant
area.
Figura 5. Abutment-coupled BAHA processor.
While waiting for osseointegration, i.e., three months, the patient has returned four
times to CDAHI for medical consultation, in which curative maintenance and phonoaudiological
consultation were performed for purposes of guiding her how to take care and sanitize
the pillar.
Processor placement and post-surgical evaluation
After three months after surgery, the patient has attended the phonoaudiological session
to adapt the process and be guided how to handle and take care of BAHA. Open-field
audiometry with pillar-coupled BAHA, external test and use of telephone were performed.
The distinction of words was evaluated after the surgery by using the same evaluation
methodology used before the surgery ([Table 1 ]).
Table 1.
Open-field threshold audiometry (dBNA) and percentile ratio of monosyllable recognition
with a 65/60 relation between signal and noise.
Open-field audiometry
500 Hz
1000Hz
2000 Hz
3000 Hz
4000 Hz
Average*
IPRF**
No amplifiers
70
50
50
20
20
47,50
84%
LE ISAA
35
25
25
25
30
27,5
92%
LE BAHA tester
20
20
25
15
20
21,25
100%
LE BAHA
30
20
20
15
15
21,25
100%
Legend: * Average between frequencies of 0.5, 1, 2 and 3 kHz.
** Percentile index of speech recognition for monosyllables at 60 dBNA on the left
side with a concurrent noise at 55 dBNA before the patient – 5 dBNA signal-noise relation.
With regard to the complaint about tinnitus, the following results were observed:
The tinnitus has started to have a low impact on her life, according to EVA's degree;
There was a 47% reduction on THI's total score, mainly in the hearing and emotional
aspects;
As regards the patient's activities, it has been observed a 30% reduction on the scores
related to concentration, emotion and hearing.
In the handicap questionnaire, there was only a 3% reduction, however when the scores
were separately compared regarding factor 1, there was a 40% improvement, and regarding
factor 2, a 60% improvement. Accordingly, a worsening occurred in factor 3 (57%).
The patient mentioned a lower perception of the tinnitus all the time when using the
system and a lower perception of the tinnitus in the morning when waking up.
Discussion of Results
In the cases of chronic otitis externa and open mastoid cavity, for occluding MAE,
using ISAA mold can induce or worsen the infectious feature, leading to continual
otorrhea[7 ]. This makes BAHA an advantageous and better accepted tool for not using MAE molds[8 ], providing an effective and ongoing usage of sound amplification.
At present, various audiological evaluation methods have been used, but they have
not been validated or verified for not having a standard evaluation protocol14 .
When comparing BAHA audiological results with those of ISAA, some authors show there
is no difference in the audiometric results between each other[9 ]
[10 ], however there are indications that the bigger the air-osseous interval (AOI), the
bigger BAHA advantage is in comparison with ISAA[11 ]. This advantage was observed in this case, mainly in 4,000-Hz frequency, where there
was a 15-dBNA difference when using BAHA in comparison with ISAA. Additionally, an
improvement occurred when distinguishing words, being 92% of right answers with ISAA
and 100% of right answers with BAHA. Another advantage BAHA has over ISAA was the
use of telephone. The patient mentioned a great hearing comfort, with a better understanding
on the phone, showing neither microphonia nor metallic voice caused by the hearing
apparatus.
As to esthetics, she declares to be more comfortable with BAHA, because the people
do not notice its presence as they used to notice ISAA.
When comparing the result of pre-surgical open-field audiometry, performed with BAHA
tester, with the post-surgical one, performed with the implant-coupled BAHA, Hakansson et al., 1990[9 ], reported to have better thresholds of airways with implant-coupled BAHA, and this
difference can be between 2 and 15 dBNA, especially in acute frequencies where skin
attenuation is bigger. In our evaluation, we observed it did not occur only in the
500-Hz frequency, we had a worse 10 dBNA threshold with implant-coupled BAHA. In the
1,000 and 3,000 Hz frequencies, there was no difference between them both, and at
2,000 and 4,000 Hz, our findings agree with the cited authors, as we observed better
5 dBNA thresholds with pillar-coupled BAHA, in comparison with the arch-coupled tester.
Maybe we could have better thresholds with an implanted BAHA, if the adjustments used
in the processor were the same as those in the pre-surgical evaluation, however it
did not occur because the patient was discomforted when we adjusted the processor
at a maximum gain, and we accordingly chose a lower gain adjustment, which would give
her more comfort in the post-surgical evaluation.
If we should analyze the average between the frequencies 50, 1,000, 2,000 and 3,000 Hz,
the pre-surgical and post-surgical thresholds are the same, what makes us trust the
pre-surgical evaluation as a preview of the result that shall be achieved in the post-surgical
period.
It has been observed that the airway thresholds on open-field audiometry with BAHA
have exceeded the patient's best pre-surgical bone conduction, what Lustig and colleagues[5 ] refer to as 'overclosure'. This was observed in 30% of the cases evaluated in their
studies.
Van Der Pouw e col. Van Der Pouw and colleagues[6 ] explain that Divine BAHA can also gain as much as 10 dB in the sensorial component
of the hearing loss, what was observed in our patient at 1,000 and 2,000-Hz frequencies,
in which she had a 10-dBNA gain, and at 3,000 and 4,000 Hz, a 5-dBNA gain. ([Table 2 ]).
Table 2.
Comparison between pre-surgical bone conduction (BC) thresholds and open-field air
conduction thresholds when using abutment-coupled BAHA.
500 Hz
1000Hz
2000 Hz
3000 Hz
4000 Hz
Average*
LE Bone conduction
20
30
30
20
20
25
LE BAHA
30
20
20
15
15
21,25
Legend: * Average between frequencies of 0.5, 1, 2 and 3 kHz.
The AOI for less than 10 dB occurs in 80% of the cases with BAHA[12 ].
Both the suspension of vertigo and a significant reduction of tinnitus perception
were observed in the questionnaires, complying with Sánchez -Camón and colleagues' findings, and showing a positive effect of BAHA system on the tinnitus
symptom[13 ].
Final Commentaries
The surgery of BAHA implantation is a safe and easily performed procedure. Since it
is a breakthrough in the Brazilian practice, a few difficulties were faced regarding
this patient's audiological evaluation, for there is not a standardization of the
methodology to be used. We suggest the exchange of experiences between the centers
providing this service, so that by standardizing the audiological evaluation, we can
compare results and pursue a better quality of service. The reliability in the pre-surgical
audiological evaluation as a preview of the post-surgical result, the outstanding
audiological results achieved, as well as the patient's high degree of satisfaction
provided by BAHA system, turns it into an advantageous option to rehabilitate conductive
and mixed hearing losses when ISAA is not possible to be used.