Bispectral index in psychiatric patients: Is there a need for a separate monitor?

Rajiv Chawla
1   Department of Anaesthesiology and Intensive Care, Govind Ballabh Pant Hospital, New Delhi, India
,
Mritunjay Kumar
1   Department of Anaesthesiology and Intensive Care, Govind Ballabh Pant Hospital, New Delhi, India
› Author Affiliations

Subject Editor:
Further Information

Address for correspondence:

Dr. Rajiv Chawla
Department of Anaesthesiology and Intensive Care
Govind Ballabh Pant Hospital, University of Delhi, New Delhi - 110 002
India   

Publication History

Publication Date:
28 April 2018 (online)

 

The bispectral index (BIS) was conceptualised to provide a substitute for the complex electroencephalograph (EEG) analysis with an easy to understand ‘magic’ number, to measure hypnotic effect of anaesthetics and sedative drugs.[1] BIS monitoring helps in maintenance of adequate levels of hypnosis and prevents awakening and formation of implicit and explicit memory during general anaesthesia.[2] Intraoperative BIS monitoring has also been found to be useful in fast tracking day-care patients,[3] prevention of postoperative emergence delirium,[4] nausea and vomiting. Since its FDA approval in 1996, BIS monitor usage is extending outside the operating room. Quantitative BIS values correlation has been observed in many neurological conditions like cerebral ischaemia,[5] vasospasm,[6] syncope,[7] seizures,[8] cerebral palsy[9] and dementia.[10] It is also found to be useful in predicting efficacy of cardiopulmonary resuscitation,[11] outcome after hypoxic ischaemic injury,[12] and hypoglycaemia,[13] determining effect of hypothermia on the hypnotic state,[14] assessing brain death and level of consciousness in case of brain injury.[15]

Although BIS monitoring is being widely used in psychiatric patients during electroconvulsive therapy, its aberrance has still not been evaluated. The present issue of the journal carries an article, in which Ramesh et al.,[16] recorded BIS values of patients suffering from psychiatric conditions viz., psychosis, depression, organic and bipolar disorder and compared it with control group patients, who were not suffering from any psychiatric illness. In their study, they observed that BIS values in psychiatric patients are significantly lower compared to the control patients. Patients with psychosis and bipolar disorder were found to have lower BIS values than patients with depression. BIS being an empirical value derived from EEG pattern of approximately 2000 healthy volunteers/patients receiving a variety of commonly used general anaesthetics, sedative-hypnotics and opioids.[1] The question arises whether these values are applicable for psychiatric patients also? Do we need a separate monitor say, BISpsy derived from EEG recordings of these patients or a different range of BIS values for them? In an observational study on 18 patients undergoing electroconvulsive therapy (ECT), Gunawardane et al. found that a low BIS value does not correspond to a deep hypnotic level in patients after ECT.[17]

Technological advancements in anaesthesia have brought in monitors which provide continuous display of important variables in a non-invasive manner. These monitors are found to be cost effective, improve safety in perioperative period and have become essential components of our clinical practice. It includes the use of non-invasive blood pressure, pulse oximeter, end-tidal carbon dioxide and BIS. Each one of these monitors provides us with a ‘magic number’ which guides our intervention(s). However, one must appreciate that these monitors have their own limitations. New devices are known to introduce new informations (and errors!). BIS is no exception. It is known that pulse oximeter, (a reliable monitor displaying oxygen saturation in patients with normal haemoglobin) gives spurious 88% saturation despite high arterial oxygen saturation in patients with methaemoglobinaemia. On a similar analogy, BIS values (which are derived from normal individuals) might show spuriously low BIS values in psychiatric patients. The study by Ramesh et al. in psychiatric patients observed low BIS values and patients’ remain awake with no manifestation of sedation. The key question is whether this BIS value is a true reflection of sedation? Is this low BIS value because of the disease process, because of the psychiatric medications or a combination of these two is a matter which needs further investigation. The study by Ramesh et al. is only the beginning, and there are many more related issues to be studied. Simultaneous correlation of BIS with EEG monitoring, brain imaging and various clinical stages and manifestations of the particular psychiatric illness; choosing similar age group (to remove effect of age related cerebral changes) and numbers in each group, converting into a randomised, controlled trial are some of the ways the quality of such studies can be improved upon. There is a possibility that we might soon have an ideal non- invasive monitor, which can reflect the regression or progression of such illness and help the physicians titrate pharmacotherapy or other modalities of treatment in them. Association of poor outcomes of the patients with low BIS values, in presence or absence of anaesthestic drugs is another area, waiting to be explored.[18]

Till all these questions are answered, let us only accept that BIS values are low in psychiatric patients and also caution that it might not be a true indicator of the depth of anaesthesia in psychiatric patients.


#

No conflict of interest has been declared by the author(s).

  • REFERENCES

  • 1 Rampil IJ. A primer for EEG signals processing in anesthesia. Anesthesiology 1998; 89: 980-1002
  • 2 Liu J, Harbhej S, White PF. Electroencephalographic bispectral index correlates with intraoperative recall and depth of propofol- induced sedation. Anesth Analg 1997; 84: 185-9
  • 3 Liu SS. Effects of bispectral index monitoring on ambulatory anesthesia. A meta-analysis of randomized controlled trials and a cost analysis. Anesthesiology 2004; 101: 311-5
  • 4 Berger M, Nadler J, Mathew JP. Preventing delirium after cardiothoracic surgery: Provocative but preliminary evidence for bispectral index monitoring. Anesth Analg 2014; 118: 706-7
  • 5 Merat S, Levecque JP, Le Gulluche Y, Diraison Y, Brinquin L, Hoffmann JJ. BIS monitoring may allow the detection of severe cerebral ischemia. Can J Anaesth 2001; 48: 1066-9
  • 6 Brallier JW, Deiner SG. Use of the bilateral BIS monitor as an indicator of cerebral vasospasm in ICU patients. Middle East J Anesthesiol 2013; 22: 161-4
  • 7 Win NN, Kohase H, Miyamoto T, Umino M. Decreased bispectral index as an indicator of syncope before hypotension and bradycardia in two patients with needle phobia. Br J Anaesth 2003; 91: 749-52
  • 8 White PF, Rawal S, Recart A, Thornton L, Litle M, Stool L. Can the bispectral index be used to predict seizure time and awakening after electroconvulsive therapy?. Anesth Analg 2003; 96: 1636-9
  • 9 Choudhry DK, Brenn BR. Bispectral index monitoring: A comparison between normal children and children with quadriplegic cerebral palsy. Anesth Analg 2002; 95: 1582-5
  • 10 Renna M, Handy J, Shah A. Low baseline bispectral index of the electroencephalogram in patients with dementia. Anesth Analg 2003; 96: 1380-5
  • 11 Lasasso TJ, Muzzi DA, Meyer FB, Sharbrough FW. Electroencephalographic monitoring of cerebral function during asystole and successful cardiopulmonary resuscitation. Anesth Analg 1992; 75: 1021-4
  • 12 Myles P, Daly D, Silvers A, Cairo S. Prediction of neurological outcome using bispectral index monitoring in patients with severe ischemic-hypoxic brain injury undergoing emergency surgery. Anesthesiology 2009; 110: 1106-15
  • 13 Wu CC, Lin CS, Mok MS. Bispectral index monitoring during hypoglycemic coma. J Clin Anesth 2002; 14: 305-6
  • 14 Mathew JP, Weatherwax KJ, East CJ, White WD, Reves JG. Bispectral analysis during cardiopulmonary bypass: The effect of hypothermia on the hypnotic state. J Clin Anesth 2001; 13: 301-5
  • 15 Myles PS, Cairo S. Artifact in the bispectral index in a patient with severe ischemic brain injury. Anesth Analg 2004; 98: 706-7
  • 16 Ramesh VJ, Radhakrishnan MC, Thimmiah R, Muralidharan K, Thirthali J, Rao GS. et al. Lower Bispectral index values in psychiatric patients: A prospective, observational study. J Neuroanaesthesiol Crit Care. 2014: 2
  • 17 Gunawardane PO, Murphy PA, Sleigh JW. Bispectral index monitoring during electroconvulsive therapy. Br J Anesth 2002; 88: 184-7
  • 18 Leslie K, Short TG. Low bispectral index values and death: The unresolved causality dilemma. Anesth Analg 2011; 113: 660-3

Address for correspondence:

Dr. Rajiv Chawla
Department of Anaesthesiology and Intensive Care
Govind Ballabh Pant Hospital, University of Delhi, New Delhi - 110 002
India   

  • REFERENCES

  • 1 Rampil IJ. A primer for EEG signals processing in anesthesia. Anesthesiology 1998; 89: 980-1002
  • 2 Liu J, Harbhej S, White PF. Electroencephalographic bispectral index correlates with intraoperative recall and depth of propofol- induced sedation. Anesth Analg 1997; 84: 185-9
  • 3 Liu SS. Effects of bispectral index monitoring on ambulatory anesthesia. A meta-analysis of randomized controlled trials and a cost analysis. Anesthesiology 2004; 101: 311-5
  • 4 Berger M, Nadler J, Mathew JP. Preventing delirium after cardiothoracic surgery: Provocative but preliminary evidence for bispectral index monitoring. Anesth Analg 2014; 118: 706-7
  • 5 Merat S, Levecque JP, Le Gulluche Y, Diraison Y, Brinquin L, Hoffmann JJ. BIS monitoring may allow the detection of severe cerebral ischemia. Can J Anaesth 2001; 48: 1066-9
  • 6 Brallier JW, Deiner SG. Use of the bilateral BIS monitor as an indicator of cerebral vasospasm in ICU patients. Middle East J Anesthesiol 2013; 22: 161-4
  • 7 Win NN, Kohase H, Miyamoto T, Umino M. Decreased bispectral index as an indicator of syncope before hypotension and bradycardia in two patients with needle phobia. Br J Anaesth 2003; 91: 749-52
  • 8 White PF, Rawal S, Recart A, Thornton L, Litle M, Stool L. Can the bispectral index be used to predict seizure time and awakening after electroconvulsive therapy?. Anesth Analg 2003; 96: 1636-9
  • 9 Choudhry DK, Brenn BR. Bispectral index monitoring: A comparison between normal children and children with quadriplegic cerebral palsy. Anesth Analg 2002; 95: 1582-5
  • 10 Renna M, Handy J, Shah A. Low baseline bispectral index of the electroencephalogram in patients with dementia. Anesth Analg 2003; 96: 1380-5
  • 11 Lasasso TJ, Muzzi DA, Meyer FB, Sharbrough FW. Electroencephalographic monitoring of cerebral function during asystole and successful cardiopulmonary resuscitation. Anesth Analg 1992; 75: 1021-4
  • 12 Myles P, Daly D, Silvers A, Cairo S. Prediction of neurological outcome using bispectral index monitoring in patients with severe ischemic-hypoxic brain injury undergoing emergency surgery. Anesthesiology 2009; 110: 1106-15
  • 13 Wu CC, Lin CS, Mok MS. Bispectral index monitoring during hypoglycemic coma. J Clin Anesth 2002; 14: 305-6
  • 14 Mathew JP, Weatherwax KJ, East CJ, White WD, Reves JG. Bispectral analysis during cardiopulmonary bypass: The effect of hypothermia on the hypnotic state. J Clin Anesth 2001; 13: 301-5
  • 15 Myles PS, Cairo S. Artifact in the bispectral index in a patient with severe ischemic brain injury. Anesth Analg 2004; 98: 706-7
  • 16 Ramesh VJ, Radhakrishnan MC, Thimmiah R, Muralidharan K, Thirthali J, Rao GS. et al. Lower Bispectral index values in psychiatric patients: A prospective, observational study. J Neuroanaesthesiol Crit Care. 2014: 2
  • 17 Gunawardane PO, Murphy PA, Sleigh JW. Bispectral index monitoring during electroconvulsive therapy. Br J Anesth 2002; 88: 184-7
  • 18 Leslie K, Short TG. Low bispectral index values and death: The unresolved causality dilemma. Anesth Analg 2011; 113: 660-3