Cleaning and Disinfection of Radiology Equipment
General mandatory principles for cleaning and disinfecting radiology equipment are
listed in [[Table 1]].
X-ray equipment
Parts: Main equipment, touch user interface, monitors, keyboard, liquid crystal displays
(LCD) and detector.
Cover – Wrap the X-ray machine, image receptor and anatomical marker with disposable two-layer
double-bagged fluid-impermeable plastic material. The detector should be wrapped in
a disposable single-use water impermeable plastic bag and ideally this plastic bag
should be changed after every patient.
Cleaning – Only use water or a lukewarm diluted household cleaning agent solution. The use
of other than the recommended cleaning agents can result in damage in the equipment.
Wipe machine parts with a moist cloth (squeeze out wet cloth before using it) until
all contaminations are removed. Remove any watery residues immediately. The back of
the X-ray detector should be cleaned first so that it has more time to dry while the
front is being cleaned. Appropriate cleaning wipes with vendor-approved disinfectants
must be used to clean the detector at least twice daily.
Disinfection – Since the radiography equipment is a non-critical medical device, and also the
equipment is wrapped in a cover, it does not require disinfection. However, when there
is hazardous spill on the equipment, disinfection may be required. To disinfect surfaces,
aldehyde solutions, guanidine derivatives and peroxide compounds should be used. Alcohol-based
products should not be used for surface cleaning. After disinfection, wipe the system
with a dry soft cloth or let the system dry naturally. Sterilisation is not allowed
for X-ray equipment.
Ultrasound
Parts: Transducer, console, liquid crystal displays (LCD), touch user interface, monitors
and keyboard
Precautions:
-
Do not clean the system with chlorinated or aromatic solvents such as isopropyl alcohol
-
Do not use spray cleaners on the ultrasound as it may damage the internal components
-
Do not pour fluid on the ultrasound system
-
Disconnect transducers from the main system before cleaning
-
Take extra precautions to clean trackball, touchpad, keys, knobs and side controls
-
Ensure fluid does not seep into the openings of system.
Critical item: Ultrasound probe used in sterile body cavities (intraoperative) – should
be sterilised using glutaraldehyde or hydrogen peroxide.
Semi critical item: Endocavitory probes - should be treated with high-level disinfectant
like glutaraldehyde, hydrogen peroxide, ortho-phthalaldehyde, and peracetic acid with
hydrogen peroxide.
Noncritical item: probes used in intact skin – should be treated with low-level disinfectant.
Cleaning the transducer: Dip the transducer (not cable) in clean water or soap water to remove all foreign
material or wipe the transducer with moist sponge. Avoid using a brush. Dry the transducer
with a clean cloth.
Disinfecting the transducer: Immerse the transducer (not cable) into a solution (22% hydrogen peroxide or 4.5%
peroxyacetic acid or 2.4% glutaraldehyde) for 10–12 hours. After this, rinse the transducer
with water to remove all chemical residues. Wipe off the water with sterile cloth
after this.
Probes used only for external use (linear, curvilinear and phased array transducers)
on intact skin without contamination of blood or bodily fluids should be cleaned with
low-level disinfection (soap water, quaternary ammonia wipes) between each use.[[5]]
Probes used externally for percutaneous procedures (vascular access, thoracentesis,
paracentesis, arthrocentesis, pericardiocentesis, lumbar puncture, regional anaesthesia
and other procedures) should be covered with single-use protective covers and sterile
gel applied. They should subsequently be cleaned with low-level disinfection.
Probes used internally on mucous membranes and internal orifices ((endocavitory probe
for intraoral procedures/transvaginal examinations and transesophageal probes) should
be covered with high-quality single-use probe covers during each examination, followed
by high-level disinfection between each use. An alternative procedure for disinfecting
the endocavitory probes involves:[[6]]
-
Removing the gel from the transducer
-
Cleaning the transducer with soap and water
-
Wiping the transducer with 70% alcohol or soak it for 2 minutes in 500 ppm chlorine
-
Rinsing with tap water
-
Air drying and wiping with cloth.
Computed tomography (CT) scanner system
Parts: Gantry – gantry cover, operator panel, touch panel; Patient table – patient
table, mattress; monitor; other components – positioning aids, body straps, ECG cables,
foot switch.
Gantry - Before starting the cleaning process, the gantry should be switched off.
Remove all impurities like blood and contrast with damp cloth. Cleaning agents and
disinfectants recommended for gantry cover, operating panel and touch panel are surgical
spirit, alcohol (≤70%), aldehyde, quaternary compounds, guanidine derivatives, pyridine
derivatives, chlorine derivatives, benzene, alkylamines, organic acids and peroxide
compounds. For the touch panel in the gantry system, use an anti-static cleaner, alcohol-free
glass cleaning fluid or a solution of water and a pH-neutral cleaning agent. Paper
towels are not recommended as these fibrous materials can scratch the screen. Corrosive
agents should be avoided as it can damage the surface coating.
Patient table - All chemicals used for cleaning gantry can be used for cleaning the
patient table. Mattress should always be covered with plastic sheet for maintaining
hygiene. Cleaning agents and disinfectants recommended for mattress are pH-neutral
agents, guanidine derivatives, chlorine derivatives and alkylamines. Soft brush may
be used for regular cleaning. Mattress should not be cleaned with surgical spirit
as this may lead to wrinkling. Alcohols and aldehydes should not be used for cleaning
mattress.
Monitor - Ensure monitor is switched off before cleaning. Clean the monitor with damp
cloth. Only an antistatic cleaner should be used for cleaning the monitor screen.
The monitor housing can be cleaned with same materials as used for gantry and surgical
spirit should be avoided.
Other components - Patient positioning aids made of foam material and body straps
can be cleaned by using pH-neutral agents, guanidine derivatives, chlorine derivatives
and alkylamines. Alcohol (≤70%) can be used for cleaning ECG cables.
Disinfection: Spray machine for fogging is not recommended for CT scanner. It should be applied
only on floor. Normally technicians/end users never wipe the equipment surface after
fumigation, which is the reason for not recommending fumigation.
Bacillocid rasant is a newer and effective compound in environmental decontamination
with good cost benefits, good material compatibility, excellent cleaning properties
and virtually no residues. It has the advantage of being a formaldehyde-free disinfectant
cleaner with low use concentration, provides complete asepsis within 30 to 60 minutes.
Air disinfection – Air exchange of minimum fresh air volume of 3 cum/hour/person and 3.75 cum/hour/sqm
is recommended. The recommendation is to maximize supply of outside air within the
limits of the system. In buildings without mechanical ventilation systems, it is recommended
to actively use operable windows. Installation of UVGI (Ultraviolet germicidal irradiation)
for larger ducted units and AHUs to keep coils continuously clean and disinfected
is highly recommended. It is advisable to regularly inspect the AHUs and ducts for
air tightness and low leakage. Minimum air changes of around 10–15 ACHP (air changes
per hour) is advised for good ventilation. The mechanical exhaust air should have
70% to 80% of fresh air. In case of recirculating system, it is recommended to limit
return air circulation. Air steriliser or ultraviolet lamp disinfection is being recommended
for air disinfection at least 4 times/day. It does not damage electronic or cable
components inside the gantry. Low-pressure mercury (Hg) discharge lamps are commonly
used in ultraviolet germicidal irradiation (UVGI) applications which emit shortwave
ultraviolet-C (UV-C, 100–280 nanometer) radiation, primarily at 254 nm. UV-C radiation
kills or inactivates microbes by damaging their deoxyribonucleic acid (DNA). Research
has shown that UV-C light has been effective in altering the cell structure and exposure
to UV-C light of 254 nm for a period of 20-30 minutes and is effective in deactivating
the virus. Usage of ozone and hydrogen peroxide is no longer recommended for air disinfection.[[7]]
Magnetic resonance imaging (MRI)
Follow the manufacturer’s instructions and use only approved disinfectants and cleaning
agents. Always use a damp cloth for cleaning. Do not pour chemicals in the scanner.
Cleaning patient table, straps, plugs and connectors: Approved agents are aldehydes,
quaternary compounds, guanidine derivatives, peroxide compounds, pyridine derivatives,
chlorine derivatives, alkylamines, organic acids and commercially available detergents.
Benzene, acetone and alcohol solutions are not recommended.
Receptors: All the chemicals mentioned above can be used for cleaning except chemicals
containing alcohol and ethers.
LCD monitors and video display: Ideally should be cleaned thoroughly once every two
months. Use commercially available window cleaners in diluted form.
Camera lens, data carrier: Use dry cotton or microfiber cloth for cleaning.
Mammography
Parts: Compression plates, biopsy unit, object table, patient handles and face shield.
All parts of the unit which come into contact with the patient must be cleaned prior
to examination. The plastic trays can be removed from the holder to simplify cleaning.
Wipe the system with a damp, lint-free, non-woven cloth or sponge. Moisten the cloth
only with water or lukewarm, diluted and aqueous solution using a household cleanser
without abrasives. Anaesthetics and skin disinfectants used during biopsies can damage
or discolour the plastic parts of the compression plates. If such substances get onto
the compression plates, wipe these off immediately with a moist cloth.
Compression troughs and holders should be disinfected by quaternary ammonium compounds.
Alcohol-based products should not be used. Disinfectants based on substituted phenols
and chlorine-splitting compounds have a slightly corrosive effect and therefore are
not recommended generally.
Hospital policies for disinfection
It is the responsibility of radiology departments, in collaboration with hospital
infection control committee and Medical Superintendent Office, to establish policies
for disinfecting radiology equipment to manage infection risk. In addition, radiology
departments should check with their equipment vendors to determine which disinfectants
are safe for radiology equipment. There should be clear cut written SOPs in Radiology
Department regarding (i) when to disinfect the facility, (ii) the personnel who will
be involved in cleaning, (iii) the specific agents to be used and (iv) the radiologist
designated for monitoring and ensuring the disinfection and cleaning.