Effects of non-invasive respiratory support on sleep in preterm infants evaluated by actigraphy
Authors
Objective Few studies have evaluated sleep in preterm infants under non-invasive ventilatory support in neonatal intensive care units (NICU). The main objective of this study was to evaluate the influence of continuous positive airway pressure (CPAP) in the sleep of premature babies.
Material and Methods Crossover study in a NICU. We selected preterm infants with gestational age between 28 and 37 weeks using nasal CPAP. Eighteen preterm were included. Patients were monitored with actigraphy and with the Neonatal Behavioral Assessment Scale (NBAS).
Results Results showed a reduction in sleep effciency, total sleep time and total sleep period during the CPAP period when compared to the non-CPAP. NBAS demonstrated significantly greater time of deep sleep and light sleep in the period without CPAP.
Conclusion Our data suggests that the use of CPAP, during the first week of life, in preterm neonates, is associated with transitory alterations of sleep organization.
INTRODUCTION
Sleep architecture in premature infants is a complex phenomenon, since the maturation of their brain structures and functions are not always linear[1],[2].
Prematurity may have a major impact on the structure of sleep, including adverse intrauterine exposures and/or stressors in the neonatal period. This impact is explained by the maturation of the suprachiasmatic nucleus occurring in an unfavorable environment, including stressors such as hypoxia, malnutrition and constant luminosity in neonatal intensive care units (NICUs)[3].
Other studies also demonstrate that several stimuli can cause alterations in sleep architecture, such as handling by the assistant team, changes in body position and use of mechanical ventilation[4],[5].
Additionally, there is evidence that sleep deprivation during brain development results in a reduction of neuronal plasticity, altering learning processes and with long-term effects on behavior and in some brain functions[6]-[8].
Improved respiratory support is one of the major factors responsible for the decreased mortality in NICUs, since the beginning of the twentieth century, when the use of continuous positive airway pressure (CPAP) started to be routinely used to stabilize the respiratory disease in preterm infants[9]-[11]. Currently, randomized controlled trials suggest that the early use of nasal CPAP in very low birth weight infants is associated with a reduction in the need for invasive support[9].
Knowing the numerous interferences that a premature infant is exposed in the NICU, it is impressive how just a few studies evaluated sleep of premature infants in the use of non-invasive respiratory support[1]. Thus, the objective of this study was to evaluate subject and objective sleep in preterm newborns undergoing continuous positive airway pressure using nasal cannula.
MATERIAL AND METHODS
Crossover study performed at the NICU of ULBRA (Universidade Luterana do Brasil) University Hospital. The study was approved by the Ethics Committee of the Pontifícia Universidade Católica do Rio Grande do Sul (PUCRS) and of ULBRA (number: 66935717.9.1001.5336). A neonatologist calculated gestational age with the New Ballard Method, which includes neuromuscular and physical maturity[12]. All parents/guardians signed the informed consent form. Clinical characteristics of the sample are summarized in [Table 1].
|
Variables |
n=18 |
|---|---|
|
Gestational age (weeks) - mean ± SD |
31,7 ± 2,1 |
|
Days of life with CPAP - median (P25-P75) |
2 (1-3) |
|
Days of life without CPAP - median(P25-P75) |
5 (4-6) |
|
Sex - n (%) |
|
|
Female |
8 (44,4) |
|
Male |
10 (55,6) |
|
Weight (g) - mean ± SD |
1708 ± 416 |
|
Delivery - n (%) |
|
|
Vaginal |
9 (50,0) |
|
Cesarean |
9 (50,0) |
|
Premature labor - n (%) |
14 (77,8) |
|
Use of corticoid - n (%) |
16 (88,9) |
|
Use of surfactant - n (%) |
10 (55,6) |
|
Oxygen (FiO2) - mean ± DP |
0.319 ± 0.057 |
|
Apgar 1 - median (P25-P75) |
8 (7-8) |
|
Apgar 5 - median (P25-P75) |
8 (8-9) |
|
PEEP ( cmH2O) - mean ± SD |
5,33 ± 0,49 |
The actigraphy monitoring occurred within the first week of life in all patients, the first evaluation (with CPAP) occurred in the 2nd day of life (median) and the second (without nasal CPAP) in the 5th (median) day of life.
Inclusion/exclusion and sample criteria
Inclusion criteria were premature infants with less than 37 weeks and with birth weight over 800g, using oral or intravenous caffeine and undergoing continuous positive airway pressure using nasal cannula. Exclusion criteria were presence of suspected seizures, continuous use of sedative drugs, cerebral malformations, chromosomal disorders, cerebral hemorrhage. Eighteen preterm infants were included in the sample criteria (8 female, 10 male) with a gestational age between 29-36 weeks (31.7 ± 2.1) and birth weight from 840g to 2,340g (1708 ± 416); nine were delivered by a cesarean section.
Procedure and evaluations
Patients were evaluated in two moments, initially using nasal CPAP with the Nellcor Puritan Bennet machine respirator, using nasal cannula and later, the same patients, were evaluated without nasal CPAP. All analysis were performed during the first week of life.
At first, the actigraph was placed randomly in the right or in the left leg (the side oscillated according to the oximeter device) for 48 hours. The actigraph was attached with a cotton bandage. Within this period, the Neonatal Behavioral Assessment Scale (NBAS) as proposed by Brazelton and Nugent[13] was performed during four intercalated 30-minute episodes, totaling 2 hours of observation. The same researcher carried out this evaluation. Sleep diary was used to exclude staff handling and feeding periods.
The Micro Motionlogger from Ambulatory Monitoring Inc. was held for 48 hours in the leg, between the knee and ankle under a cotton bed. The observed actigraph patterns were total sleep period (TSP) - defined as the number of minutes from onset of sleep to awakening; total sleep time (TST) - characterized as the number of minutes of the sleep time excluding the awakenings; sleep efficiency - the ratio between TST and TSP, expressed as a percentage[14]-[16].
Immediately after CPAP removal, the same evaluation was repeated for a further 48 hours.
Sleep rating by the NBAS is divided in six stages such as: deep sleep (closed eyes, regular breathing, no spontaneous activity, no eye movement); light sleep (closed eyes, irregular breathing, low activity level, eye movements may be present, suctioning or occasional movements); sleepy (open eyes, heavy eyelids, variable activity level, responsive to sensory response with slow response, slight movements and not fully alert); awake (alert, ocular opening, minimal motor activity reactive to visual and sound stimulus) and fully awake (eyes open, increased motor activity, maximum reaction to external stimuli and crying stage).
The sleep diary was used just to register the time of feeding and handling by the medical and nursing staff.
Statistical analysis
The sample size calculation was performed in the WinPEPI program (Programs for Epidemiologists for Windows) version 11.43 and based on the study by Collins et al.[1]. To reach a significance level of 5%, power of 90% and an effect size of at least one standard deviation between the two evaluations, a minimum of 17 premature infants were necessary. Quantitative variables were described by mean and standard deviation or interquartile range and median. Qualitative variables were described by absolute and relative absolute frequencies. To compare means between moments, the t-student test for paired samples was applied. To evaluate the associations of clinical variables with the results of the actigraphy analysis, the Spearman and Mann-Whitney correlation tests were used. SPSS version 21.0 was used to analyze data.
RESULTS
[Table 2] compares the actigraphy findings with and without CPAP. TST, TSP and SE were significantly higher without CPAP. A significant alteration in the sleep macrostructure, with reduction of all the parameters studied (total sleep time, total sleep period and sleep efficiency) was observed, when data were compared to the non-CPAP period.
|
Variables # |
Without CPAP |
With CPAP |
Difference (IC 95%) |
p |
|---|---|---|---|---|
|
Mean ± SD |
Mean ± SD |
|||
|
TST (min) |
745,8 ± 125,7 |
431,8 ± 132,6 |
313,9 (229,8 to 398,1) |
<0,001 |
|
TSP (min) |
981,9 ± 170,3 |
691,2 ± 186,6 |
290,7 (174,5 to 406,9) |
<0,001 |
|
SE (%) |
76,2 ± 6,8 |
62,0 ± 13,2 |
14,2 (8,8 to 19,6) |
<0,001 |
[Table 3] shows the NBAS analysis. The periods in deep and light sleep were significantly higher without nasal CPAP. Drowsiness and waking periods were significantly higher with CPAP.
|
Variables |
With CPAP |
Without CPAP |
p |
|---|---|---|---|
|
Median (P25-P75) |
Median (P25-P75) |
||
|
Deep sleep (%) |
0 (0-12,8) |
25 (7,5-39,8) |
0,008 |
|
Light sleep (%) |
24 (19,8-28,5) |
42,5 (31,5-62,5) |
0,002 |
|
Sleepy (%) |
39,5 (29,5-42,8) |
12,5 (0-30) |
0,001 |
|
Awake (%) |
29 (12,8-37,5) |
0 (0-15) |
0,006 |
|
Crying (%) |
0 (0-1,3) |
0 (0-0) |
0,206 |
|
Fully awake (%) |
0 (0-10) |
0 (0-0) |
0,173 |
Results showed a reduction in total sleep time, total sleep period ([Figure 1]) and sleep efficiency ([Figure 2]) during the CPAP period when compared to the non-CPAP.




DISCUSSION
In this study, we have evaluated sleep characteristics of preterm infants through the analysis of two methods, a behavioral scale and the actigraph monitoring. Results showed a reduction in sleep efficiency, total sleep time and total sleep period during the CPAP period when compared to the non-CPAP. NBAS demonstrated significantly greater time of deep sleep and light sleep in the period without CPAP. Both methods identified that CPAP influences preterm sleep expressed as transient alterations in the sleep macrostructure, that were reverted after the withdrawal of the ventilatory support.
The use of nasal CPAP might alter sleep organization in different ways such as due to mechanical problems (incorrect choice of the size of the nasal cannula, problems with its fixation, both essential to provide the necessary pressure in the airways and to avoid lesions in the nasal septum) or lack of constant overview of the device (checking the heating, humidification and flow of oxygen and compressed air)[9]. Further aspects refer to careful changes of body position and the possibility of abdominal distension due to nasal flow and harming ventilator stabilization[9]. Sleep position might also alter sleep organization. In previous studies, Jarus et al.[4] observed that in the prone position more sleep patterns (deep sleep, light sleep and drowsy) were observed as opposed to more awake patterns (quiet awake, active awake and agitated) that were seen in the supine position. It is important to note that in the NICU where the data was collected there are protocols/routines, regarding care with sounds, lights, handling and position of the newborn that are followed by the assistant team and regularly monitored. It is important to note that, during the 2 hours of infant sleep assessment, staff handling and feeding periods were excluded.
The American Academy of Pediatrics advises that noise levels in the NICU should be less than 45dB. However, it is known that nasal CPAP generates noises greater than 45dB. These values may vary with the type of device and respirators used to deliver continuous positive airway pressure[17],[18].
Reinforcing the findings of our study, previous studies reiterate that any type of manipulation in the baby, lighting and sound can alter the quality of sleep in NICU[4],[5],[19]. We know that the use of CPAP may increase the manipulation due to an improperly coupled nasal cannula, noises or any physical discomfort suffered by prematurity. Levy et al.[19] suggests that reducing handling or adjusting the time of handling in active sleep may be beneficial.
Likewise, Collins et al.[1] compared sleep quality, through actigraphy and behavioral scale, in high-flow catheter and nasal CPAP. This paper observed decreased sleep efficiency and increased activity in the high-flow catheter due to respirator noise and manipulation of the assisting team.
It is important to emphasize that premature infants under CPAP are also under unfavorable clinical conditions, requiring oxygenation and pressure in the upper airway[11],[20]. Therefore, the question arises whether CPAP alone is responsible for the alterations on sleep organization or it is an additional factor with the underlying pathology. As we observed in our findings, CPAP withdrawal coincided with a significant improvement in sleep characteristics, but it is important to question whether such improvement is not secondary to a favorable clinical outcome. Further studies are needed to answer this specific question.
As limitations of our study, we can cite the small number of patients despite the power of 90%; another limitation of the study was that the treatment - non-treatment group was not randomized. The sound and handling characteristics of the attending staff should also be considered, since alarms are common at a NICU. In addition, although the NICU had positional routines when using CPAP (alternate prone and supine position), this aspect was not analyzed in this study. No information on circadian rhythmicity were provided or analyzed because actigraph does not show this data.
In conclusion, our study showed that premature infants exposed to nasal CPAP had alterations in the sleep architecture when objective and subjective measures were analyzed.
Conflict of Interests
The authors have no conflict of interests to declare.
FINANCIAL DISCLOSURE
MLN is supported by the Grant PQ 306338/2017-3 from CNPq - Brazil. This study was financed in part by the Coordenação de Aperfeiçoamento de Pessoal de Nível Superior - Brazil - (CAPES) - Finance Code 001, through a scholarship received by FS.
CONTRIBUTION STATEMENT
Fernanda Schmidt: Conceptualization, data curation, investigation, methodology, project administration, supervision, validation, visualization, writing - original draft, writing - review & editing. Felipe Kalil Neto: writing - original draft, writing - review & editing. Graciane Radaelli: writing - review & editing. Magda L. Nunes: data curation, investigation, methodology, project administration, supervision, validation, visualization, writing - original draft, writing - review & editing.
-
REFERENCES
- 1 Collins CL, Barfield C, Davis PG, Horne RSC. Randomized controlled trial to compare sleep and wake in preterm infants less than 32 weeks of gestation receiving two different modes of non-invasive respiratory support. Early Hum Dev. 2015 Dec;91(12):701-4.
- 2 Nunes ML, Costa JC. Sleep and epilepsy in neonates. Sleep Med. 2010 Aug;11(7):665-73.
- 3 Kennaway DJ. Programming of the fetal suprachiasma nucleus and subsequent adult rhythmicity. Trends Endocrinol Metab. 2002 Nov;13(9):398-402.
- 4 Jarus T, Bart O, Rabinovich G, Sadeh A, Bloch L, Dolfin T, et al. Effects of prone and supine positions on sleep state and stress responses in preterm infants. Infant Behav Dev. 2011 Apr;34(2):257-63.
- 5 Chang YJ, Anderson GC, Lin CH. Effects of prone and supine positions on sleep state and stress responses in mechanically ventilated preterm infants during the first postnatal week. J Adv Nurs. 2002 Oct;40(2):161-9.
- 6 Graven S. Sleep and brain development. Clin Perinatol. 2006 Sep;33(3):693-706.
- 7 Loughlin GM, Carrol JL, Marcus CL. Sleep and breathing children: a developmental approach. New York: Marcel Dekker; 2000.
- 8 Maquet P, Smith C, Stickgold R. Sleep and brain plasticity. Oxford/ New York: Oxford University Press; 2003.
- 9 Sahini R, Schiaratura M, Polin RA. Strategies for the prevention of continuous positive airway pressure failure. Semin Fetal Neonatal Med. 2016 Jun;21(3):196-203.
- 10 Gregory GA, Kitterman JA, Phibbs RH, Tooley WH, Harnilton WK. Treatment of the idiopathic respiratory distress syndrome with continuous positive airway pressure. N Engl J Med. 1971 Jun;284(24):1333-40.
- 11 Schmölzer GM, Kumar M, Pichler G, Aziz K, O’Reilly M, Cheung PY. Non-invasive versus invasive respiratory support in preterm infants at birth: systematic review and meta-analysis. BMJ. 2013 Oct;347:f5980.
- 12 Ballard JL, Khoury JC, Wedig K, Wang L, Eilers-Walsman BL, Lipp R. New Ballard Score, expanded to include extremely premature infants. J Pediatr. 1991 Sep;119(3):417-23.
- 13 Brazelton TB, Nugent JK. Neonatal Behavioral Assessment Scale (NBAS). 3rd ed. London: MacKeith Press; 1995.
- 14 Sadeh A, Acebo C. The role of actigraphy in sleep medicine. Sleep Med Rev. 2002 Apr;6(2):113-24.
- 15 Sadeh A. The role and validity of actigraphy in sleep medicine: an update. Sleep Med Rev. 2011 Aug;15(4):259-67.
- 16 Sung M, Adamson TM, Horne RSC. Validation of actigraphy for determining sleep and wake in preterm infants. Acta Paediatr. 2009;98:52-7.
- 17 American Academy of Pediatrics (AAP). Committee on Environmental Health. Noise: a hazard for the fetus and newborn. Pediatrics. 1997 Oct;100(4):724-7.
- 18 Swanson JR, Almarode S, Walsh BK, Attridge JT. Noise generation from non-invasive respiratory support equipment in a NICU [Abstract]. E-PAS [Internet]. 2008; [access in 30th september 2012]; 634458.6. Available from: www.abstractstoview.com
- 19 Levy J, Hassan F, Plague MA, Sokoloff MD, Kushwaha JS, Chervin RD, et al. Impact of hands-on care on infant sleep in the neonatal intensive care unit. Pediatr Pulmonol. 2017 Jan;52(1):84-90.
- 20 Gupta S, Donn SM. Continuous positive airway pressure: physiology and comparison of devices. Semin Fetal Neonatal Med. 2016 Jun;21(3):204-11.
Corresponding author:
Publication History
Received: 01 April 2020
Accepted: 06 July 2020
Article published online:
30 November 2023
© 2023. Brazilian Sleep Association. This is an open access article published by Thieme under the terms of the Creative Commons Attribution-NonDerivative-NonCommercial License, permitting copying and reproduction so long as the original work is given appropriate credit. Contents may not be used for commercial purposes, or adapted, remixed, transformed or built upon. (https://creativecommons.org/licenses/by-nc-nd/4.0/)
Thieme Revinter Publicações Ltda.
Rua do Matoso 170, Rio de Janeiro, RJ, CEP 20270-135, Brazil
-
REFERENCES
- 1 Collins CL, Barfield C, Davis PG, Horne RSC. Randomized controlled trial to compare sleep and wake in preterm infants less than 32 weeks of gestation receiving two different modes of non-invasive respiratory support. Early Hum Dev. 2015 Dec;91(12):701-4.
- 2 Nunes ML, Costa JC. Sleep and epilepsy in neonates. Sleep Med. 2010 Aug;11(7):665-73.
- 3 Kennaway DJ. Programming of the fetal suprachiasma nucleus and subsequent adult rhythmicity. Trends Endocrinol Metab. 2002 Nov;13(9):398-402.
- 4 Jarus T, Bart O, Rabinovich G, Sadeh A, Bloch L, Dolfin T, et al. Effects of prone and supine positions on sleep state and stress responses in preterm infants. Infant Behav Dev. 2011 Apr;34(2):257-63.
- 5 Chang YJ, Anderson GC, Lin CH. Effects of prone and supine positions on sleep state and stress responses in mechanically ventilated preterm infants during the first postnatal week. J Adv Nurs. 2002 Oct;40(2):161-9.
- 6 Graven S. Sleep and brain development. Clin Perinatol. 2006 Sep;33(3):693-706.
- 7 Loughlin GM, Carrol JL, Marcus CL. Sleep and breathing children: a developmental approach. New York: Marcel Dekker; 2000.
- 8 Maquet P, Smith C, Stickgold R. Sleep and brain plasticity. Oxford/ New York: Oxford University Press; 2003.
- 9 Sahini R, Schiaratura M, Polin RA. Strategies for the prevention of continuous positive airway pressure failure. Semin Fetal Neonatal Med. 2016 Jun;21(3):196-203.
- 10 Gregory GA, Kitterman JA, Phibbs RH, Tooley WH, Harnilton WK. Treatment of the idiopathic respiratory distress syndrome with continuous positive airway pressure. N Engl J Med. 1971 Jun;284(24):1333-40.
- 11 Schmölzer GM, Kumar M, Pichler G, Aziz K, O’Reilly M, Cheung PY. Non-invasive versus invasive respiratory support in preterm infants at birth: systematic review and meta-analysis. BMJ. 2013 Oct;347:f5980.
- 12 Ballard JL, Khoury JC, Wedig K, Wang L, Eilers-Walsman BL, Lipp R. New Ballard Score, expanded to include extremely premature infants. J Pediatr. 1991 Sep;119(3):417-23.
- 13 Brazelton TB, Nugent JK. Neonatal Behavioral Assessment Scale (NBAS). 3rd ed. London: MacKeith Press; 1995.
- 14 Sadeh A, Acebo C. The role of actigraphy in sleep medicine. Sleep Med Rev. 2002 Apr;6(2):113-24.
- 15 Sadeh A. The role and validity of actigraphy in sleep medicine: an update. Sleep Med Rev. 2011 Aug;15(4):259-67.
- 16 Sung M, Adamson TM, Horne RSC. Validation of actigraphy for determining sleep and wake in preterm infants. Acta Paediatr. 2009;98:52-7.
- 17 American Academy of Pediatrics (AAP). Committee on Environmental Health. Noise: a hazard for the fetus and newborn. Pediatrics. 1997 Oct;100(4):724-7.
- 18 Swanson JR, Almarode S, Walsh BK, Attridge JT. Noise generation from non-invasive respiratory support equipment in a NICU [Abstract]. E-PAS [Internet]. 2008; [access in 30th september 2012]; 634458.6. Available from: www.abstractstoview.com
- 19 Levy J, Hassan F, Plague MA, Sokoloff MD, Kushwaha JS, Chervin RD, et al. Impact of hands-on care on infant sleep in the neonatal intensive care unit. Pediatr Pulmonol. 2017 Jan;52(1):84-90.
- 20 Gupta S, Donn SM. Continuous positive airway pressure: physiology and comparison of devices. Semin Fetal Neonatal Med. 2016 Jun;21(3):204-11.