Stent-in-stent, a safe and effective technique to remove fully embedded esophageal metal stents: case series and literature review

Authors

  • Alberto Aiolfi

    University of Milan, Department of Biomedical Sciences for Health, Division of General Surgery, IRCCS Policlinico San Donato, Milan, Italy
  • Davide Bona

    University of Milan, Department of Biomedical Sciences for Health, Division of General Surgery, IRCCS Policlinico San Donato, Milan, Italy
  • Chiara Ceriani

    University of Milan, Department of Biomedical Sciences for Health, Division of General Surgery, IRCCS Policlinico San Donato, Milan, Italy
  • Matteo Porro

    University of Milan, Department of Biomedical Sciences for Health, Division of General Surgery, IRCCS Policlinico San Donato, Milan, Italy
  • Luigi Bonavina

    University of Milan, Department of Biomedical Sciences for Health, Division of General Surgery, IRCCS Policlinico San Donato, Milan, Italy
Further Information

Publication History

submitted 23 October 2014

accepted after revision 07 January 2014

Publication Date:
24 June 2015 (online)

Background: Endoscopic stenting is a widely used method for managing esophageal anastomotic leaks and perforations. Self-expanding metal stents (SEMSs) have proved effective in sealing these defects, with a lower rate of displacement than that of self-expanding plastic stents (SEPSs) as a result of tissue proliferation and granulation tissue ingrowth at the uncovered portion of the stent, which anchor the prosthesis to the esophageal wall. Removal of a fully embedded stent is challenging because of the risk of bleeding and tears.

Materials and methods: Temporary placement of a new stent within the first stent (stent-in-stent technique) may facilitate the mobilization and safe removal of both stents by inducing pressure ischemia of the granulation tissue. We report our own experience with the stent-in-stent technique in five consecutive patients in whom a partially covered Ultraflex stent had previously been implanted and compare our results with those in the current literature.

Results: The first SEMSs remained in place for a median of 40 days (range 18 – 68) without displacement. Placement of the new stent was technically successful in all patients. All stents were left in place for a median of 9 days. The overall stent-in-stent success rate was 100 % for the removal of embedded stents. No serious adverse events related to the procedure occurred.

Conclusion: The procedure was safe, well tolerated, and effective. The use of a partially covered Ultraflex stent of the same size as the old stent for a limited time (≤ 6 days) was consistently successful.