Subscribe to RSS
DOI: 10.3414/ME13-02-0053
Monitoring Nocturnal Heart Rate with Bed Sensor
Publication History
received:26 November 2013
accepted:13 May 2014
Publication Date:
20 January 2018 (online)
Summary
Introduction: This article is part of the Focus Theme of Methods of Information in Medicine on “Biosignal Interpretation: Advanced Methods for Studying Cardiovascular and Respiratory Systems”.
Objectives: The aim of this study is to assess the reliability of the estimated Nocturnal Heart Rate (HR), recorded through a bed sensor, compared with the one obtained from standard electrocardiography (ECG).
Methods: Twenty-eight sleep deprived patients were recorded for one night each through matrix of piezoelectric sensors, integrated into the mattress, through polysomnography (PSG) simultaneously. The two recording methods have been compared in terms of signal quality and differences in heart beat detection.
Results: On average, coverage of 92.7% of the total sleep time was obtained for the bed sensor, testifying the good quality of the recordings. The average beat-to-beat error of the inter-beat intervals was 1.06%. These results suggest a good overall signal quality, however, considering fast heart rates (HR > 100 bpm), performances were worse: in fact, the sensitivity in the heart beat detection was 28.4% while the false positive rate was 3.8% which means that a large amount of fast beats were not detected.
Conclusions: The accuracy of the measurements made using the bed sensor has less than 10% of failure rate especially in periods with HR lower than 70 bpm. For fast heart beats the uncertainty increases. This can be explained by the change in morphology of the bed sensor signal in correspondence of a higher HR.
-
References
- 1 Kelly JM, Strecker RE, Bianchi MT. Recent Developments in Home Sleep-Monitoring Devices. ISRN Neurology. 2012
- 2 Penzel T, Kantelhardt JW, Grote L, Peter J, Bunde A. Comparison of Detrended Fluctuation Analysis and Spectral Analysis for Heart Rate Variability in Sleep and Sleep Apnea. IEEE Transactions on Biomedical Engineering. 2003. 50 10
- 3 Alihanka J, Vaahtoranta K, Saarikivi I. A new method for longterm monitoring of the ballistocardiogram, heart rate, and respiration. Am J Physiol 1993; 240 (09) 289-297.
- 4 Lekkala J, Tuppurainen J, Paajanen M. Material and operational properties of large area membrane type sensors for smart environments. XVII IMEKO World Congress. 2003: 22-27.
- 5 Kortelainen JM, Mendez MO, Bianchi AM, Matteucci M, Cerutti S. Sleep Staging Based on Signals Acquired Through Bed Sensor. IEEE Transactions on Information Technology in Biomedicine 2010; 14 (03) 776-785.
- 6 Cornelussen R, Splett V, Klepfer R, Stegemann B, Kornet L, Prinzen F. Electrical modalities beyond pacing for the treatment of heart failure. Heart Failure Reviews 2011; 16 (03) 315-325.
- 7 Kortelainen JM, Virkkala J. FFT averaging of multichannel BCG signals from bed mattress sensor to improve estimation of heart beat interval. 29th Annual International Conference of the IEEE. 2007: 6685-6688.
- 8 Kortelainen JM, van Gils M, Pärkkä J. Multichannel Bed Pressure Sensor for Sleep monitoring. Computing in Cardiology; Krakow, Poland: Sep 2012: 9-12.
- 9 Kuna S. et al An Official ATS/AASM/ACCP/ERS Workshop Report: research properties in ambulatory management of adults with obstructive sleep apnea. Proc Am Thorac Soc 2011; 8: 1-16.