CC BY-NC-ND 4.0 · Endosc Int Open 2021; 09(12): E1914-E1916
DOI: 10.1055/a-1629-1221
Innovation forum

Single-use Albarran module: A removable elevator system to enhance safety in biliary endoscopy?

Clara Benedetta Conti
Interventional Endoscopy – ASST Monza, Ospedale San Gerardo, Monza, Italy
,
Roberto Frego
Interventional Endoscopy – ASST Monza, Ospedale San Gerardo, Monza, Italy
,
Alessandro Ettore Redaelli
Interventional Endoscopy – ASST Monza, Ospedale San Gerardo, Monza, Italy
,
Marta Maino
Interventional Endoscopy – ASST Monza, Ospedale San Gerardo, Monza, Italy
,
Giacomo Mulinacci
Interventional Endoscopy – ASST Monza, Ospedale San Gerardo, Monza, Italy
,
Marco Emilio Dinelli
Interventional Endoscopy – ASST Monza, Ospedale San Gerardo, Monza, Italy
› Author Affiliations

Bacterial infections are a major burden in biliary endoscopy. Besides incomplete manual cleaning of endoscopes, hard-to-clean components of duodenoscopes are responsible for a substantial rate of bacterial contamination. A large meta-analysis showed that 15 % of duodenoscopes harbored microorganisms of gastrointestinal or oral origin, independent from the colony forming unit (CFU) count. This indicates that the current reprocessing and process control procedures, even with ethylene oxide (ETO), could be insufficient [1]. Therefore, US Food and Drug Administration (FDA) warnings led companies to redesign duodenoscopes with the aim of enhancing safety. Disposable caps or caps + elevator are available options on the market [2] [3]. Moreover, two disposable duodenoscopes models have recently been approved by the FDA: the Boston Scientifc EXALT Model D and the Ambu A/S Ambu aScope Duodeno. Studies on disposable duodenoscope feasibility and possible economic impact on the market yielded interesting results, although the data are still preliminary [4] [5] [6] [7] [8].

In this paper, we present an innovative solution of a duodenoscope with a removable cap, wire, and distal lever mechanism. This elevator system is called the single-use Albarran module (Karl Storz) ([Fig. 1]). With it, a large, open surface can be created that is easy to brush and clean, as is the case with any other instrument without a bifurcated channel. Basically, the module is assembled and removed before and after every procedure, to be reprocessed separately from the rest of the duodenoscope (even with ETO). Interestingly, the distal end of the duodenoscope belongs to the removable module ([Fig. 2]). This detail makes the cleaning both the channel and the tip easier. Karl Storz conducted some tests on duodenoscopes contaminated with organic fluids, such as blood. After removal of the single-use Albarran module and duodenoscope reprocessing, the instruments achieved the officially required level of cleaning (i. e. < 6.4 μg/cm2 for protein, < 2.4 μg/cm2 for hemoglobin and < 4 Log10/cm2 for bacteria) [9].

Zoom Image
Fig. 1 The image on the left shows the lever mechanism to remove the cap; the image on the right shows the wire that is inserted in the cap.
Zoom Image
Fig. 2 The image shows the distal portion of the endoscope, without wire and cap.

Ten endoscopic retrograde cholangiopancreatographies (ERCPs) were performed in nine consecutive patients using the single-use Albarran module. Data are summarized in  [Table 1]. The setting was urgent in one case and elective in nine. A native papilla was present in seven patients (70 %); the bile duct cannulation and successful procedure rates were both 100 %. Mean procedure time was 30 minutes. Indications for ERCP were biliary lithiasis in seven cases and cancer palliation in two cases. Sphincterotomy and balloon dilation were performed in five and four cases, respectively. Mechanical lithotripsy was performed only once, whereas stone clearing required balloon dilation in six cases. Brushing for cytology was done in one case and stenting for palliation in two.

Table 1

Elevator performance evaluation[1].

Patient (sex; age)

Indication

Bile duct cannulation

Operative procedure

Operative device

Stent characteristics

Procedure setting/duration (min)

Elevator performance

M,83

Biliary AP

Yes

Precut, plastic stent placement

Plastic, 10 Fr, 5 cm

Elective/31

Good

M,76

CBD stones

Yes (previous sphincterotomy)

Stent removal, balloon dilation, mechanic lithotripsy, stent placement

CRE Boston Scientific, balloon dilator 12–15 mm, Fogarty catheters

NA

Elective/34

Good

F,82

Biliary hilar stricture

Yes

Balloon dilation of main hepatic ducts

CONMED balloon dilator 6.00 mm

NA

Elective/43

Good

M,53

Cancer palliation

Yes

Stent placement

Partially covered SEMS 10 mm Ø, 6 cm

Elective/19

Good

M,59

CBD stones

Yes

Balloon dilation

CBD clearance

Dormia basket, CRE Boston Scientific, balloon dilator 12–15 mm

NA

Elective/20

Good

M,76

Biliary AP

Yes

Stent placement

Plastic 10 Fr, 5 cm

Urgent/43

Good

M;71

CBD stones

Yes

CBD clearance

NA

Elective/22

Good

F;42

Cancer palliation

Yes

Balloon dilation

Stent placement

CRE Boston Scientific, balloon dilator 10–12 mm

Partially covered SEMS

Elective/30

Good

F;82

2nd look

Yes (previous sphincterotomy)

Stent removal

Brushing

Stent placement

2 plastic stents, 10 Fr, 12 cm

Elective/43

Good

M;41

2nd look

Yes (previous sphincterotomy)

CBD clearance

NA

Elective/15

Good

NA, not applicable; AP, acute pancreatitis; CBD, common bile duct; PC, pancreatic cancer; SEMS, self-expandable metallic stent.

1 Special attention was given to ease of accessory insertion, grip, friction, strength, and width of the movement, rated as good = no problems nor difficulties; fair = any problem occurred with no interference with the outcome of the procedure; poor = any problem occurred significant enough to interrupt the procedure or change the elevator.


The aim of this small study was to gain insights into this novel device, with special attention directed towards identifying any technical problems related to use of the disposable systems. Notably, we did not find any issues with devices insertion, grip, friction, strength, or width of the elevator movements, in contrast to the study by Bang et al. [10] Because of the type of study (case series) and small sample size, it was not possible to compare the single-use Albarran maneuverability and mechanical features with the standard module. The only difference from the standard Albarran module was a shorter width during the elevator stroke, which did not hinder the planned interventions. In our experience, the disposable device allows the use of all devices commonly employed in biliary endoscopy. Importantly, no cases of sepsis or infectious events were reported. Unfortunately, a direct comparison of infectious event rate between the standard and single-use Albarran module is not possible at this time, due to the small sample size in the present case series. However, to our knowledge, this is the first description of single-use Albarran module use during routine endoscopy in real-life settings. Further studies and randomized controlled trials are needed to address comparisons with the standard duodenoscopes and to better evaluate its use in clinical practice.



Publication History

Received: 04 June 2021

Accepted: 24 August 2021

Article published online:
14 December 2021

© 2021. The Author(s). This is an open access article published by Thieme under the terms of the Creative Commons Attribution-NonDerivative-NonCommercial License, permitting copying and reproduction so long as the original work is given appropriate credit. Contents may not be used for commercial purposes, or adapted, remixed, transformed or built upon. (https://creativecommons.org/licenses/by-nc-nd/4.0/)