Background Biliary tract emergencies are managed with endoscopic retrograde cholangiopancreatography (ERCP) using duodenoscopes, which are reusable devices that require high-level disinfection to minimize risk of cross-contamination. Recent reports about newly developed single-use duodenoscopes (SUDs) suggest equivalent performance with reusable duodenoscopes, but the effectiveness of SUDs in emergency ERCP has not yet been studied.

Patients and methods We conducted a prospective case series of emergency ERCP procedures using SUDs (EXALT model D-Boston Scientific, United States) in a real-life, tertiary care setting without any possibility of using a back-up reusable duodenoscope.

Results Twenty-one emergent ERCPs (acute cholangitis 48%, severe jaundice 38 %, others 14 %) were performed in 19 patients (mean age 49.5±15 years). Almost all procedures (20 of 21; 95 %) were technically and clinically successful, whereas selective cannulation failed in one case. Among the successful ERCPs, five (24 %) were achieved by a novice operator. The image was often considered skewed toward yellow tones (48 %), whereas stiffness and pushability for stent insertion were found suboptimal in 5 % of the procedures, without any impact on procedure success.

Conclusions SUDs are effective and appropriate devices for emergent situations in real life even in non-expert hands and even if a regular duodenoscope is unavailable as a backup.



Publikationsverlauf

Eingereicht: 09. Februar 2022

Angenommen nach Revision: 20. Mai 2022

Accepted Manuscript online:
09. August 2022

Artikel online veröffentlicht:
15. November 2022

© 2022. The Author(s). This is an open access article published by Thieme under the terms of the Creative Commons Attribution-NonDerivative-NonCommercial License, permitting copying and reproduction so long as the original work is given appropriate credit. Contents may not be used for commercial purposes, or adapted, remixed, transformed or built upon. (https://creativecommons.org/licenses/by-nc-nd/4.0/)

Georg Thieme Verlag KG
Rüdigerstraße 14, 70469 Stuttgart, Germany

 
  • References

  • 1 Rauwers AW, Kwakman JA, Vos MC. et al. Endoscope-associated infections: A brief summary of the current state and views toward the future. Tech Gastrointest Endosc 2019; 21: 150608
  • 2 Rauwers AW, Voor in’t holt AF, Buijs JG. et al. High prevalence rate of digestive tract bacteria in duodenoscopes: a nationwide study. Gut 2018; 67: 1637-1645
  • 3 Peter S, Bang JY, Varadarajulu S. Single-use duodenoscopes: where are we and where are we going?. Curr Opin Gastroenterol 2021; 37: 416-420
  • 4 Bang JY, Hawes R, Varadarajulu S. Equivalent performance of single-use and reusable duodenoscopes in a randomised trial. Gut 2021; 70: 838-844
  • 5 Bulut M, Hjørne F, Knuhtsen S. et al. Sterile laparoscopic transgastric ERCP with single-use disposable duodenoscope. Endoscopy 2021;
  • 6 Ehrlich D, Muthusamy VR. Device profile of the EXALT Model D single-use duodenoscope for endoscopic retrograde cholangiopancreatography: overview of its safety and efficacy. Expert Rev Med Devices 2021; 18: 421-427
  • 7 Muthusamy VR, Bruno MJ, Kozarek RA. et al. Clinical evaluation of a single-use duodenoscope for endoscopic retrograde cholangiopancreatography. Clin Gastroenterol Hepatol Off Clin Pract J Am Gastroenterol Assoc 2020; 18: 2108-2117.e3
  • 8 Reddy DN, Ramchandani M, Lakhtakia S. et al. Single-use duodenoscope in the management of an elderly patient with difficult bile duct stones: laser lithotripsy using a disposable cholangioscope. VideoGIE 2021; 6: 319-321
  • 9 Ross AS, Bruno MJ, Kozarek RA. et al. Novel single-use duodenoscope compared with 3 models of reusable duodenoscopes for ERCP: a randomized bench-model comparison. Gastrointest Endosc 2019; S0016510719322011
  • 10 Slivka A, Ross AS, Sejpal DV. et al. Single-use duodenoscope for ERCP performed by endoscopists with a range of experience in procedures of variable complexity. Gastrointest Endosc 2021;
  • 11 Napoléon B, Gonzalez J, Grandval P. et al. Evaluation of the performances of a single‐use duodenoscope: Prospective multi‐center national study. Dig Endosc 2021;
  • 12 Singer M, Deutschman CS, Seymour CW. et al. The Third International Consensus Definitions for Sepsis and Septic Shock (Sepsis-3). JAMA 2016; 315: 801-810
  • 13 Schutz SM, Abbott RM. Grading ERCPs by degree of difficulty: a new concept to produce more meaningful outcome data. Gastrointest Endosc 2000; 51: 535-539
  • 14 Baker N, Bromley-Dulfano R, Chan J. et al. COVID-19 solutions are climate solutions: lessons from reusable gowns. Front Public Health 2020; 8: 590275