CC BY-NC-ND 4.0 · Endosc Int Open 2024; 12(01): E52-E56
DOI: 10.1055/a-2221-7908
Original article

Through-the-scope clip retention rates and performance in a porcine model

Jad AbiMansour
1   Gastroenterology and Hepatology, Mayo Clinic, Rochester, United States (Ringgold ID: RIN6915)
,
Shunsuke Kamba
1   Gastroenterology and Hepatology, Mayo Clinic, Rochester, United States (Ringgold ID: RIN6915)
,
Louis-Michel Wong Kee Song
1   Gastroenterology and Hepatology, Mayo Clinic, Rochester, United States (Ringgold ID: RIN6915)
,
Elizabeth Rajan
1   Gastroenterology and Hepatology, Mayo Clinic, Rochester, United States (Ringgold ID: RIN6915)
› Institutsangaben
Gefördert durch: STERIS Endoscopy

Abstract

Background and study aims Limited comparative data exist to guide optimal through-the-scope (TTS) clip selection. The aim of this study was to compare the efficacy, retention, and safety of three industry-leading TTS clips on tissue that mimics common clinical scenarios.

Methods A survival study involving six domestic pigs was undertaken. Three commonly used clip models were selected: Assurance (STERIS, Mentor, Ohio, United States), Resolution (Boston Scientific, Boston, Massachusetts, United States), and SureClip (Micro-Tech, Ann Arbor, Michigan, United States). To mimic clinical practice, the following scenarios were assessed: (1) normal mucosa; (2) cold snare resection; and (3) hot mucosal resection simulating fibrotic ulcers. Deployment of clips was randomized to target sites. Repeat endoscopy was performed 2 weeks following placement. Endoscopists rated the ease of use of clip placement on a Likert scale of 1 to 5.

Results Fifty-four clips (18 Assurance, 18 Resolution, and 18 SureClip) were placed in six pigs. Mucosal healing was noted at all sites on follow up. Overall retention was nine of 18 (50.0%) SureClip, 10 of 18 (55.6%) Assurance, and 13 of 18 (72.2%) Resolution (P=0.369). There was no difference in clip retention on normal and cold snare resection sites; however, clip retention was significantly higher for Resolution clips on fibrotic ulcers (50.0% versus 0% for Assurance and 0% SureClip, P=0.03). No adverse events were reported. Ease of use was equivalent across all models.

Conclusions All clips were equivalent in efficacy and safety with successful clip deployment and mucosal healing. Overall retention rate was low for fibrotic tissue, with an improved retention rate observed with Resolution clips.



Publikationsverlauf

Eingereicht: 16. September 2023

Angenommen nach Revision: 17. November 2023

Accepted Manuscript online:
04. Dezember 2023

Artikel online veröffentlicht:
08. Januar 2024

© 2024. The Author(s). This is an open access article published by Thieme under the terms of the Creative Commons Attribution-NonDerivative-NonCommercial-License, permitting copying and reproduction so long as the original work is given appropriate credit. Contents may not be used for commercial purposes, or adapted, remixed, transformed or built upon. (https://creativecommons.org/licenses/by-nc-nd/4.0/).

Georg Thieme Verlag KG
Rüdigerstraße 14, 70469 Stuttgart, Germany

 
  • References

  • 1 Hayashi T. The study on stanch clips for the treatment by endoscopy. Gastroenterolog Endosc 1975; 17: 92-101
  • 2 Parsi MA, Schulman AR, Aslanian HR. et al. Devices for endoscopic hemostasis of nonvariceal GI bleeding (with videos). VideoGIE 2019; 4: 285-299
  • 3 Wang TJ, Aihara H, Thompson AC. et al. Choosing the right through-the-scope clip: a rigorous comparison of rotatability, whip, open/close precision, and closure strength (with videos). Gastrointest Endosc 2019; 89: 77-86.e1
  • 4 National Research Council (US) Committee for the Update of the Guide for theCare and Use of Laboratory Animals. Guide for the Care and Use of Laboratory Animals. 8th Aufl. Washington (DC): National Academies Press (US); 2011
  • 5 Qadeer MA, Dumot JA, Vargo JJ. et al. Endoscopic clips for closing esophageal perforations: case report and pooled analysis. Gastrointest Endosc 2007; 66: 605-611 DOI: 10.1016/j.gie.2007.03.1028. (PMID: 17725956)
  • 6 Udorah MO, Fleischman MW, Bala V. et al. Endoscopic clips prevent displacement of intestinal feeding tubes: a long-term follow-up study. Dig Dis Sci 2010; 55: 371-374 DOI: 10.1007/s10620-009-0726-3. (PMID: 19242799)
  • 7 Raymer GS, Sadana A, Campbell DB. et al. Endoscopic clip application as an adjunct to closure of mature esophageal perforation with fistulae. Clin Gastroenterol Hepatol 2003; 1: 44-50
  • 8 Voermans RP, Worm AM, van Berge Henegouwen MI. et al. In vitro comparison and evaluation of seven gastric closure modalities for natural orifice transluminal endoscopic surgery (NOTES). Endoscopy 2008; 40: 595-601
  • 9 Liaquat H, Rohn E, Rex DK. Prophylactic clip closure reduced the risk of delayed postpolypectomy hemorrhage: experience in 277 clipped large sessile or flat colorectal lesions and 247 control lesions. Gastrointest Endosc 2013; 77: 401-407 DOI: 10.1016/j.gie.2012.10.024. (PMID: 23317580)
  • 10 Kaltenbach T, Anderson JC, Burke CA. et al. Endoscopic removal of colorectal lesions—recommendations by the US Multi-Society Task Force on Colorectal Cancer. Gastrointest Endosc 2020; 91: 486-519 DOI: 10.1053/j.gastro.2019.12.018. (PMID: 32122632)
  • 11 Aslanian HR, Sethi A, Bhutani MS. et al. ASGE guideline for endoscopic full-thickness resection and submucosal tunnel endoscopic resection. VideoGIE 2019; 4: 343-350 DOI: 10.1016/j.vgie.2019.03.010. (PMID: 31388606)
  • 12 Daram SR, Tang S-J, Wu R. et al. Benchtop testing and comparisons among three types of through-the-scope endoscopic clipping devices. Surg Endosc 2013; 27: 1521-1529
  • 13 Jensen DM, Machicado GA, Hirabayashi K. Randomized controlled study of 3 different types of hemoclips for hemostasis of bleeding canine acute gastric ulcers. Gastrointest Endosc 2006; 64: 768-773 DOI: 10.1016/j.gie.2006.06.031. (PMID: 17055872)
  • 14 Shin EJ, Ko C-W, Magno P. et al. Comparative study of endoscopic clips: duration of attachment at the site of clip application. Gastrointest Endosc 2007; 66: 757-761 DOI: 10.1016/j.gie.2007.03.1049. (PMID: 17905019)
  • 15 Jensen DM, Machicado GA. Hemoclipping of chronic canine ulcers: a randomized, prospective study of initial deployment success, clip retention rates, and ulcer healing. Gastrointest Endosc 2009; 70: 969-975 DOI: 10.1016/j.gie.2009.04.052. (PMID: 19640519)
  • 16 Khashab M, Saxena P, Valeshabad A. et al. Which clip? A prospective comparative study of retention rates of endoscopic clips on normal mucosa and ulcers in a porcine model. Saudi J Gastroenterol 2014; 20: 360
  • 17 Swellengrebel HAM, Marijnen CAM, Vincent A. et al. Evaluating long-term attachment of two different endoclips in the human gastrointestinal tract. World J Gastrointest Endosc 2010; 2: 344-348 DOI: 10.4253/wjge.v2.i10.344. (PMID: 21160584)