Perfusion Quantification in Contrast-Enhanced Ultrasound (CEUS) – Ready for Research Projects and Routine Clinical Use

Perfusions-Quantifizierung mittels kontrastverstärkter Sonografie (CEUS) – einsatzbereit für Forschungsprojekte und die klinische Routineanwendung
F. Tranquart
,
L. Mercier
,
P. Frinking
,
E. Gaud
,
M. Arditi
Further Information

Publication History

Publication Date:
21 June 2012 (online)

Abstract

With contrast-enhanced ultrasound (CEUS) now established as a valuable imaging modality for many applications, a more specific demand has recently emerged for quantifying perfusion and using measured parameters as objective indicators for various disease states. However, CEUS perfusion quantification remains challenging and is not well integrated in daily clinical practice. The development of VueBox™ alleviates existing limitations and enables quantification in a standardized way. VueBox™ operates as an off-line software application, after dynamic contrast-enhanced ultrasound (DCE-US) is performed. It enables linearization of DICOM clips, assessment of perfusion using patented curve-fitting models, and generation of parametric images by synthesizing perfusion information at the pixel level using color coding. VueBox™ is compatible with most of the available ultrasound platforms (nonlinear contrast-enabled), has the ability to process both bolus and disruption-replenishment kinetics loops, allows analysis results and their context to be saved, and generates analysis reports automatically. Specific features have been added to VueBox™, such as fully automatic in-plane motion compensation and an easy-to-use clip editor. Processing time has been reduced as a result of parallel programming optimized for multi-core processors. A long list of perfusion parameters is available for each of the two administration modes to address all possible demands currently reported in the literature for diagnosis or treatment monitoring. In conclusion, VueBox™ is a valid and robust quantification tool to be used for standardizing perfusion quantification and to improve the reproducibility of results across centers.

Zusammenfassung

Beim kontrastverstärkten Ultraschall (CEUS), der sich jetzt in vielen Anwendungsbereichen als wertvolle darstellende Modalität durchgesetzt hat, ist mit der Quantifizierung der Perfusion und der Verwendung der ermittelten Parameter als objektive Indikatoren für verschiedene Erkrankungsstadien nun eine sehr spezielle Anforderung entstanden. Allerdings ist die Perfusionsquantifizierung mittels CEUS anspruchsvoll und noch nicht ausreichend in den klinischen Alltag integriert. Die Entwicklung von VueBoxTM verringert die bestehenden Einschränkungen und ermöglicht eine Quantifizierung in standardisierter Weise. VueBoxTM arbeitet als Anwender-Software im Off-line-Modus, nachdem eine dynamische kontrastverstärkte Sonografie (DCE-US) durchgeführt wurde. Sie ermöglicht die Linearisierung von DICOM-Clips, die Bewertung der Perfusion mittels patentierter Curve-fitting-Modelle und die Erzeugung parametrischer Darstellungen durch die Synthetisierung der Perfusionsinformation auf der Pixel-Ebene mittels Farbkodierung. VueBoxTM ist ist mit den meisten der verfügbaren Ultraschallplattformen (die nicht linearen Kontrast ermöglichen) kompatibel, kann sowohl kinetische Bolus- als auch Disruptions-Regenerations-Schleifen verarbeiten, ermöglicht die Speicherung der Analyseergebnisse und deren Inhalte und erstellt die Analyseprotokolle automatisch. Spezielle Funktionen wurden der VueBoxTM hinzugefügt, wie der vollautomatische Bewegungsausgleich in der Ebene und einen einfach zu bedienenden Clip-Editor. Die Verarbeitungszeit wurde aufgrund der Verbesserung der parallelen Programmierung der Multicore-Prozessoren verkürzt. Eine große Bandbreite an Perfusionsparametern ist bei jedem der beiden Administrationsmodi verfügbar, um alle möglichen Anforderungen, die derzeit für die Diagnose und die Therapieüberwachung in der Literatur berichtet werden, zu erfüllen. Abschließend lässt sich sagen, dass VueBoxTM eine nützliche und robuste Quantifizierungsmethode ist, um eine standardisierte Perfusionsquantifizierung durchzuführen und die Reproduzierbarkeit der Befunde zwischen den Zentren zu verbessern.

 
  • References

  • 1 Simpson DH, Chin CT, Burns PN. Pulse inversion Doppler: a new method for detecting nonlinear echoes from microbubble contrast agents. IEEE Trans Ultrason Ferroelectr Freq Control 1999; 46: 372-382
  • 2 Wilson S, Burns P. Microbubble-enhanced US in body imaging: what role?. Radiology 2010; 257: 24-38
  • 3 Claudon M, Cosgrove DO, Albrecht T et al. Guidelines and good clinical practice recommendations for contrast enhanced ultrasound (CEUS) – update 2008. Ultraschall in Med Biol 2008; 29: 28-44
  • 4 Piscaglia F, Nolsøe C, Dietrich CF et al. The EFSUMB Guidelines and Recommendations on the Clinical Practice of Contrast Enhanced Ultrasound (CEUS): Update 2011 on non-hepatic applications. Ultraschall in Med 2011; in press
  • 5 Correas JM, Claudon M, Tranquart F et al. The kidney: imaging with microbubble contrast agents. Ultrasound Q 2006; 22: 53-66
  • 6 Seidel G, Meairs S. Ultrasound contrast agents in ischemic stroke. Cerebrovasc Dis 2009; 27: 25-39
  • 7 Girlich C, Jung EM, Huber E et al. Comparison between preoperative quantitative assessment of bowel wall vascularization by contrast-enhanced ultrasound and operative macroscopic findings and results of histopathological scoring in Crohn’s disease. Ultraschall in Med 2011; 32: 154-159
  • 8 Caproni N, Marchisio F, Pecchi A et al. Contrast-enhanced ultrasound in the characterisation of breast masses: utility of quantitative analysis in comparison with MRI. Eur Radiol 2009; 20: 1384-1395
  • 9 Prantl L, Schreml S, Walter M et al. Evaluation of microcirculation of free flaps of the lower leg by contrast harmonic imaging (CHI) with time intensity curve (TIC) analysis. Clin Hemorheol Microcirc 2008; 39: 343-350
  • 10 Geis S, Prantl L, Gehmert S et al. TTP (time to PEAK) and RBV (regional blood volume) as valuable parameters to detect early flap failure. Clin Hemorheol Microcirc 2011; 48: 81-94
  • 11 Rognin N, Frinking P, Messager T et al. A New Method for Enhancing Dynamic Vascular Patterns of Focal Liver Lesions in Contrast Ultrasound. IEEE Ultrasonics Symp Proc 2007; 546-549
  • 12 Testa AC, Timmerman D, Van Belle V et al. Intravenous contrast ultrasound examination using contrast-tuned imaging (CnTI) and the contrast medium SonoVue for discrimination between benign and malignant adnexal masses with solid components. Ultrasound Obstet Gynecol 2009; 34: 699-710
  • 13 Lassau N, Koscielny S, Chami L et al. Advanced hepatocellular carcinoma: early evaluation of response to Bevacizumab therapy at dynamic contrast enhanced US with quantification – Preliminary results. Radiology 2011  258: 291-300
  • 14 Lamuraglia M, Escudier B, Chami L et al. To predict progression-free survival and overall survival in metastatic renal cancer treated with sorafenib: pilot study using dynamic contrast-enhanced Doppler ultrasound. Eur J Cancer 2006; 42: 2472-2479
  • 15 De Giorgi U, Aliberti C, Benea G et al. Effect of angiosonography to monitor response during imatinib treatment in patients with metastatic gastrointestinal stromal tumors. Clin Cancer Res 2005; 11: 6171-6176
  • 16 Lassau N, Koscielny S, Albiges L et al. Metastatic renal cell carcinoma treated with sunitinib: early evaluation of treatment response using dynamic contrast-enhanced ultrasonography. Clin Cancer Res 2010; 16: 1216-1225
  • 17 Cosgrove D, Lassau N. Imaging of perfusion using ultrasound. Eur J Nucl Med Mol Imaging 2010; 37: 65-85
  • 18 Wei K, Jayaweera AR, Firoozan S et al. Quantification of myocardial blood flow with ultrasound induced destruction of microbubbles administered as a constant venous infusion. Circulation 1998; 97: 473-483
  • 19 Peronneau P, Lassau N, Leguerney I et al. Contrast ultrasonography: necessity of linear data processing for the quantification of tumor vascularization. Ultraschall in Med 2010; 31: 370-378
  • 20 Rognin NG, Frinking P, Costa M et al. In-vivo Perfusion Quantification by Contrast Ultrasound:Validation of the Use of linearized Video Data vs. Raw RF Data. IEEE Ultrasonics Symposium 2008; 1-4: 1690-1693
  • 21 Krix M, Kiessling F, Farhan N et al. A multivessel model describing replenishment kinetics of ultrasound contrast agent for quantification of tissue perfusion. Ultrasound Med Biol 2003; 29: 1421-1430
  • 22 Arditi M, Frinking PJ, Zhou X et al. A new formalism for the quantification of tissue perfusion by the destruction-replenishment method in contrast ultrasound imaging. IEEE Trans Ultrason Ferroelect Freq Contr 2006; 53: 1118-1129
  • 23 Rognin N, Mercier L, Frinking P et al. Parametric imaging of dynamic vascular patterns of focal liver lesions in contrast-enhanced ultrasound. IEEE Ultrasonics Symp Proc 2009; 1282-1285
  • 24 Hoeffel C, Mulé S, Huwart L et al. Renal blood flow quantification in pigs using contrast-enhanced ultrasound: an ex vivo study. Ultraschall in Med 2010; 31: 363-369
  • 25 Seicean A, Badea R, Stan-Iuga R et al. Quantitative contrast-enhanced harmonic endoscopic ultrasonography for the discrimination of solid pancreatic masses. Ultraschall in Med 2010; 31: 571-576
  • 26 Averkiou M, Lampaskis M, Kyriakopoulou K et al. Quantification of tumor microvascularity with respiratory gated contrast enhanced ultrasound for monitoring therapy. Ultrasound Med Biol 2010; 36: 68-77
  • 27 Williams R, Hudson JM, Lloyd BA et al. Dynamic Microbubble Contrast-enhanced US to Measure Tumor Response to Targeted Therapy: A Proposed Clinical Protocol with Results from Renal Cell Carcinoma Patients Receiving Antiangiogenic Therapy. Radiology 2011; 260: 581-590