Endosc Int Open 2016; 04(01): E42-E50
DOI: 10.1055/s-0034-1393367
Original article
© Georg Thieme Verlag KG Stuttgart · New York

Performance of novel tissue harmonic echo imaging using endoscopic ultrasound for pancreatic diseases

Kazuyuki Matsumoto*
1   Center for Gastroenterology, Teine-Keijinkai Hospital, Hokkaido, Japan
,
Akio Katanuma*
1   Center for Gastroenterology, Teine-Keijinkai Hospital, Hokkaido, Japan
,
Hiroyuki Maguchi
1   Center for Gastroenterology, Teine-Keijinkai Hospital, Hokkaido, Japan
,
Kuniyuki Takahashi
1   Center for Gastroenterology, Teine-Keijinkai Hospital, Hokkaido, Japan
,
Manabu Osanai
1   Center for Gastroenterology, Teine-Keijinkai Hospital, Hokkaido, Japan
,
Kei Yane
1   Center for Gastroenterology, Teine-Keijinkai Hospital, Hokkaido, Japan
,
Toshifumi Kin
1   Center for Gastroenterology, Teine-Keijinkai Hospital, Hokkaido, Japan
,
Ryo Takaki
1   Center for Gastroenterology, Teine-Keijinkai Hospital, Hokkaido, Japan
,
Tomoaki Matsumori
1   Center for Gastroenterology, Teine-Keijinkai Hospital, Hokkaido, Japan
,
Katsushige Gon
1   Center for Gastroenterology, Teine-Keijinkai Hospital, Hokkaido, Japan
,
Akiko Tomonari
1   Center for Gastroenterology, Teine-Keijinkai Hospital, Hokkaido, Japan
,
Masanori Nojima
2   Center for Translational Research, The Institute of Medical Science, The University of Tokyo, Tokyo, Japan
› Author Affiliations
Further Information

Publication History

Publication Date:
19 November 2015 (online)

Background and study aims: Recently, tissue harmonic echo (THE) imaging has advanced with the development of a new endoscopic ultrasound (EUS) monitor/processing unit. With this new technology, penetration (THE-P) and resolution (THE-R) images can be obtained. The aim of this study was to investigate the performance of this novel THE imaging using a new processing unit for pancreatic diseases.

Patients and methods: Fifty patients with pancreatic lesions (38 cystic, 12 solid) were retrospectively analyzed. At each examination, 3 EUS images of the same pancreatic lesion were obtained using B-mode, THE-P mode, and THE-R mode imaging. Each set of EUS images was randomly arranged and evaluated independently by 4 physicians blinded to the imaging technique. Images were compared using a Likert scale 5-point grading system for each parameter.

Results: For cystic lesions, THE-P mode images were significantly superior to conventional B-mode images for visualizing the boundary, septum, nodules, and total image quality (P < 0.05). THE-R mode images were significantly superior to conventional B-mode images for visualizing the boundary, septum, and total image quality (P < 0.05). However, for solid lesions, there was no significant difference in all the evaluation points between THE-P and conventional B-mode images. THE-R mode images were inferior to conventional B-mode images for visualizing the boundary, internal structure, and total image quality (P < 0.05).

Conclusions: For pancreatic cystic lesions, THE mode images provided better lesion characterization than conventional B-mode images. Further research is required to determine if this improvement will result in improved EUS diagnostics.

* These authors contributed equally.


 
  • References

  • 1 Tranquart F, Grenier N, Eder V et al. Clinical use of ultrasound tissue harmonic imaging. Ultrasound Med Biol 1999; 256: 889-894
  • 2 Ward B, Baker AC, Humphrey VF. Nonlinear propagation applied to the improvement of resolution in diagnostic medical ultrasound. J Acoust Soc Am 1997; 101: 143-154
  • 3 Shapiro RS, Wagreich J, Rarsons RB et al. Tissue harmonic imaging sonography: evaluation of image quality compared with conventional sonography. Am J Roentgenol 1998; 171: 1203-1206
  • 4 Ishikawa H, Hirooka Y, Itoh A et al. A comparison of image quality between tissue harmonic imaging and fundamental imaging with an electronic radial scanning echoendoscope in the diagnosis of pancreatic diseases. Gastrointest Endosc 2003; 57: 931-936
  • 5 Kim SH, Lee JM, Kim KG et al. Comparison of fundamental sonography, tissue-harmonic sonography, fundamental compound sonography, and tissue-harmonic compound sonography for focal hepatic lesions. Eur Radiol 2006; 16: 2444-2453
  • 6 Blaivas M, DeBehnke D, Sierzenski PR et al. Tissue harmonic imaging improves organ visualization in trauma ultrasound when compared with standard ultrasound mode. Acad Emerg Med 2002; 9: 48-53
  • 7 Byrne MF, Jowell PS. Gastrointestinal imaging: Endoscopic ultrasound. Gastroenterology 2002; 122: 1631-48
  • 8 Kahl S, Glasbrenner B, Zimmermann S et al. Endoscopic ultrasound in pancreatic disease. Dig Dis 2002; 20: 120-126
  • 9 Meritz HR, Sechopoulos P, Delbeke D et al. EUS, PET, and CT scanning for evaluation of pancreatic adenocarcinoma. Gastrointest Endosc 2000; 52: 367-371
  • 10 Matsumoto K, Katanuma A. Novel tissue harmonic imaging clearly visualizes a case of intraductal papillary mucinous neoplasm with mural nodules. JOP 2014; 15 (03) 259-260
  • 11 Bang JY, Hawes RH, Varadarajule S. Objective evaluation of a new endoscopic ultrasound processor. Dig Endosc 2013; 25 (05) 554-555
  • 12 Katanuma A, Isayama H, Bapaye A. Endoscopic ultrasonography using new functions for pancreatobiliary disease: Current status and future perspectives. Dig Endosc 2015; 27: 47-54
  • 13 Thosani N, Thosani S, Qiao W et al. Role of EUS-FNA-based cytology in the diagnosis of mucinous pancreatic cystic lesions: a systematic review and meta-analysis. Dig Dis Sci 2010; 55 (10) 2756-2766
  • 14 Yamao K, Mizuno N, Takagi T et al. How I do it and when I use (and do not use) EUS-FNA. Gastrointest Endosc 2009; 69: 134-137
  • 15 Hirooka Y, Goto H, Itoh A et al. Case of intraductal papillary mucinous tumor in which endosonography-guided fine-needle aspiration biopsy caused dissemination. J. Gastroenterol Hepatol 2003; 18: 1323-1324